This document specifies requirements for reference measurement procedures (RMP) for measurands used in laboratory medicine.
This document applies to:
RMPs providing values of differential or rational quantities where each quantity value is a numerical value multiplied by a measurement unit. Annex A provides information on ordinal quantities and nominal properties;
any person, body or institution developing RMPs for measurands used in laboratory medicine.

  • Standard
    37 pages
    English language
    e-Library read for
    1 day

This document specifies requirements for certified reference materials (CRMs) of higher metrological order and the content of the supporting documentation and the calibration hierarchies as described in ISO 17511:2020, 5.2.1, 5.3.1, 5.4.1, 5.5.1, 5.6.1, 5.7.1. It is applicable to CRMs intended for use as either primary reference materials (PRMs), secondary calibrators or international conventional calibrators within calibration hierarchies appropriate for measurands used in laboratory medicine, or for applications as trueness controls. It also specifies requirements for determining the certified value of a CRM, including evaluation, and reporting of the assigned uncertainty.
This document is applicable primarily to CRMs with assigned property values where the property has a magnitude that can be expressed as a quantitative scalar number or ratio to a reference or refers to a counting scale as also described in ISO 17511:2020, Clause 1.
When a CRM includes multiple measurands, this document is applied to each of the certified quantity values present in the CRM.
Although intended to be applicable to producers of CRMs, this document is also useful for reference materials (RMs) that are not in conformity with the full metrological requirements of CRMs. For example, this document does not apply to an RM created by an in vitro diagnostic medical device (IVD MD) manufacturer for use as working calibrator or end-user calibrator within a calibration hierarchy traceable to a CRM, although some content can be useful in assessing its performance.

  • Standard
    32 pages
    English language
    e-Library read for
    1 day

This document specifies requirements for certified reference materials (CRMs) of higher metrological order and the content of the supporting documentation and the calibration hierarchies as described in ISO 17511:2020, 5.2.1, 5.3.1, 5.4.1, 5.5.1, 5.6.1, 5.7.1. It is applicable to CRMs intended for use as either primary reference materials (PRMs), secondary calibrators or international conventional calibrators within calibration hierarchies appropriate for measurands used in laboratory medicine, or for applications as trueness controls. It also specifies requirements for determining the certified value of a CRM, including evaluation, and reporting of the assigned uncertainty.
This document is applicable primarily to CRMs with assigned property values where the property has a magnitude that can be expressed as a quantitative scalar number or ratio to a reference or refers to a counting scale as also described in ISO 17511:2020, Clause 1.
When a CRM includes multiple measurands, this document is applied to each of the certified quantity values present in the CRM.
Although intended to be applicable to producers of CRMs, this document is also useful for reference materials (RMs) that are not in conformity with the full metrological requirements of CRMs. For example, this document does not apply to an RM created by an in vitro diagnostic medical device (IVD MD) manufacturer for use as working calibrator or end-user calibrator within a calibration hierarchy traceable to a CRM, although some content can be useful in assessing its performance.

  • Standard
    32 pages
    English language
    e-Library read for
    1 day

This document specifies requirements for reference measurement procedures (RMP) for measurands used in laboratory medicine.
This document applies to:
RMPs providing values of differential or rational quantities where each quantity value is a numerical value multiplied by a measurement unit. Annex A provides information on ordinal quantities and nominal properties;
any person, body or institution developing RMPs for measurands used in laboratory medicine.

  • Standard
    37 pages
    English language
    e-Library read for
    1 day

1.1       This document specifies requirements for the development, validation and routine control of a radiation sterilization process for medical devices.
NOTE         Although the scope is limited to medical devices, this document can be applicable to other products and equipment.
This document covers radiation processes employing irradiators using:
a)      the radionuclide 60Co or 137Cs;
b)      a beam from an electron generator; or
c)       a beam from an X-ray generator.
1.2       This document is not applicable to processes for inactivating viruses or the causative agents of spongiform encephalopathies, such as scrapie, bovine spongiform encephalopathy and Creutzfeldt-Jakob disease.
NOTE         For information on such processes, see ISO 22442-1, ISO 22442-2, ISO 22442-3, ISO 13022 and ICH Q5A.
1.2.1       This document does not specify requirements for designating a medical device as sterile.
NOTE         Regional and national requirements can designate medical devices as sterile. See, for example, EN 556-1 or ANSI/AAMI ST67.
1.2.2       This document does not specify a quality management system for the control of all stages of production of medical devices.
NOTE         It is not a requirement of this document to have a complete quality management system during manufacture, but the elements of a quality management system that are the minimum necessary to control the sterilization process are normatively referenced at appropriate places in the text (see, in particular, Clause 4). Attention is drawn to the standards for quality management systems (see ISO 13485) that control all stages of production of medical devices, including the sterilization process. Regional and national regulations for the provision of medical devices can require implementation of a complete quality management system and the assessment of that system by a third party.
1.2.3       This document does not require that biological indicators be used for validation or monitoring of radiation sterilization, nor does it require that a pharmacopoeial test for sterility be carried out for product release.
1.2.4       This document does not specify requirements for occupational safety associated with the design and operation of irradiation facilities.
NOTE         Regulations on safety requirements for occupational safety related to radiation can exist in some countries.
1.2.5       This document does not specify requirements for the sterilization of used or reprocessed devices.

  • Standard
    61 pages
    English language
    e-Library read for
    1 day

1.1       This document specifies requirements for the development, validation and routine control of a radiation sterilization process for medical devices.
NOTE         Although the scope is limited to medical devices, this document can be applicable to other products and equipment.
This document covers radiation processes employing irradiators using:
a)      the radionuclide 60Co or 137Cs;
b)      a beam from an electron generator; or
c)       a beam from an X-ray generator.
1.2       This document is not applicable to processes for inactivating viruses or the causative agents of spongiform encephalopathies, such as scrapie, bovine spongiform encephalopathy and Creutzfeldt-Jakob disease.
NOTE         For information on such processes, see ISO 22442-1, ISO 22442-2, ISO 22442-3, ISO 13022 and ICH Q5A.
1.2.1       This document does not specify requirements for designating a medical device as sterile.
NOTE         Regional and national requirements can designate medical devices as sterile. See, for example, EN 556-1 or ANSI/AAMI ST67.
1.2.2       This document does not specify a quality management system for the control of all stages of production of medical devices.
NOTE         It is not a requirement of this document to have a complete quality management system during manufacture, but the elements of a quality management system that are the minimum necessary to control the sterilization process are normatively referenced at appropriate places in the text (see, in particular, Clause 4). Attention is drawn to the standards for quality management systems (see ISO 13485) that control all stages of production of medical devices, including the sterilization process. Regional and national regulations for the provision of medical devices can require implementation of a complete quality management system and the assessment of that system by a third party.
1.2.3       This document does not require that biological indicators be used for validation or monitoring of radiation sterilization, nor does it require that a pharmacopoeial test for sterility be carried out for product release.
1.2.4       This document does not specify requirements for occupational safety associated with the design and operation of irradiation facilities.
NOTE         Regulations on safety requirements for occupational safety related to radiation can exist in some countries.
1.2.5       This document does not specify requirements for the sterilization of used or reprocessed devices.

  • Standard
    61 pages
    English language
    e-Library read for
    1 day

NOTE 1        There is guidance or rationale for this Clause in A.2.1.
This document specifies the requirements for information supplied by the manufacturer for a medical device or an accessory, as defined in 3.1. This document includes the generally applicable requirements for identification and labels on a medical device or accessory, the packaging, marking of a medical device or accessory, and accompanying information. This document does not specify the means by which the information is to be supplied.
NOTE 2        Some authorities having jurisdiction impose different requirements for the identification, marking and documentation of a medical device or accessory.
Specific requirements of medical device product standards or group standards take precedence over requirements of this document.

  • Standard
    60 pages
    English language
    e-Library read for
    1 day

NOTE 1        There is guidance or rationale for this Clause in A.2.1.
This document specifies the requirements for information supplied by the manufacturer for a medical device or an accessory, as defined in 3.1. This document includes the generally applicable requirements for identification and labels on a medical device or accessory, the packaging, marking of a medical device or accessory, and accompanying information. This document does not specify the means by which the information is to be supplied.
NOTE 2        Some authorities having jurisdiction impose different requirements for the identification, marking and documentation of a medical device or accessory.
Specific requirements of medical device product standards or group standards take precedence over requirements of this document.

  • Standard
    60 pages
    English language
    e-Library read for
    1 day

IEC 61326-2-6:2025 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of IN VITRO DIAGNOSTIC MEDICAL ELECTRICAL EQUIPMENT (IVD MEE). This part of IEC 61326 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of IVD MEE in the presence of electromagnetic disturbances and to electromagnetic disturbances emitted by IVD MEE.
BASIC SAFETY with regard to electromagnetic disturbances is applicable to all IVD MEE.
NOTE 1 Performance with respect to electromagnetic disturbances other than ESSENTIAL PERFORMANCE is the subject of IEC 61326-1:2020
NOTE 2 IT equipment can be a part of an IVD MEE, if it is required to maintain BASIC SAFETY or ESSENTIAL PERFORMANCE.
This edition includes the following significant technical changes with respect to the previous edition:
- Update of the document with respect to test levels and documentation.

  • Standard
    36 pages
    English language
    e-Library read for
    1 day

IEC 61326-2-6:2025 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of IN VITRO DIAGNOSTIC MEDICAL ELECTRICAL EQUIPMENT (IVD MEE). This part of IEC 61326 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of IVD MEE in the presence of electromagnetic disturbances and to electromagnetic disturbances emitted by IVD MEE. BASIC SAFETY with regard to electromagnetic disturbances is applicable to all IVD MEE. NOTE 1 Performance with respect to electromagnetic disturbances other than ESSENTIAL PERFORMANCE is the subject of IEC 61326-1:2020 NOTE 2 IT equipment can be a part of an IVD MEE, if it is required to maintain BASIC SAFETY or ESSENTIAL PERFORMANCE. This edition includes the following significant technical changes with respect to the previous edition: - Update of the document with respect to test levels and documentation.

  • Standard
    36 pages
    English language
    e-Library read for
    1 day

This document specifies general requirements for non-active surgical implants, hereafter referred to as implants.
This document is not applicable to dental implants, dental restorative materials, transendodontic and transradicular implants, intra-ocular lenses and implants utilizing viable animal or human tissue.
With regard to safety, this document specifies requirements for intended performance, design attributes, materials, design evaluation, manufacture, sterilization, packaging and information supplied by the manufacturer, and tests to demonstrate compliance with these requirements.
Additional requirements applicable to specific implants or implant families are given or referred to in Level 2 and Level 3 standards.
NOTE 1        This document does not require that the manufacturer have a quality management system in place. However, many regulatory authorities require the application of a quality management system, such as that described in ISO 13485, to ensure that the implant achieves its intended performance and safety.
NOTE 2        In this document, when not otherwise specified, the term "implant" refers to each individual component of a system or a modular implant, provided separately or as a set of components, as well as to all ancillary implants or associated implants designed for improving the intended performance.

  • Standard
    36 pages
    English language
    e-Library read for
    1 day

This document specifies general requirements for non-active surgical implants, hereafter referred to as implants.
This document is not applicable to dental implants, dental restorative materials, transendodontic and transradicular implants, intra-ocular lenses and implants utilizing viable animal or human tissue.
With regard to safety, this document specifies requirements for intended performance, design attributes, materials, design evaluation, manufacture, sterilization, packaging and information supplied by the manufacturer, and tests to demonstrate compliance with these requirements.
Additional requirements applicable to specific implants or implant families are given or referred to in Level 2 and Level 3 standards.
NOTE 1        This document does not require that the manufacturer have a quality management system in place. However, many regulatory authorities require the application of a quality management system, such as that described in ISO 13485, to ensure that the implant achieves its intended performance and safety.
NOTE 2        In this document, when not otherwise specified, the term "implant" refers to each individual component of a system or a modular implant, provided separately or as a set of components, as well as to all ancillary implants or associated implants designed for improving the intended performance.

  • Standard
    36 pages
    English language
    e-Library read for
    1 day

This document specifies the requirements for an aseptically processed medical device to be designated “STERILE”.

  • Standard
    17 pages
    English language
    e-Library read for
    1 day

This document specifies the requirements for an aseptically processed medical device to be designated “STERILE”.

  • Standard
    17 pages
    English language
    e-Library read for
    1 day

This document specifies the requirements for a terminally sterilized medical device to be designated ‘STERILE’. Part 2 of this European standard specifies the requirements for an aseptically processed medical device to be designated "STERILE".
NOTE   For the purpose of the EU Directive(s) for medical devices (see Bibliography), designation of a medical device as ‘STERILE’ is only permissible when a validated sterilization process has been applied. Requirements for validation and routine control of processes for the sterilization of medical devices are specified in EN ISO 11135, EN ISO 11137, EN ISO 14160, EN ISO 14937, EN ISO 17665-1, EN ISO 20857, EN ISO 25424 and ISO 22441.

  • Standard
    14 pages
    English language
    e-Library read for
    1 day

This document specifies requirements for information supplied by the manufacturer of in vitro diagnostic (IVD) reagents, calibrators and controls intended for professional use.
This document can also be applicable to accessories.
This document is applicable to the labels for outer and immediate containers and to the instructions for use.
This document does not apply to:
a) IVD instruments or equipment;
b) IVD reagents for self-testing.

  • Standard
    24 pages
    English language
    e-Library read for
    1 day

This document specifies requirements for information supplied by the manufacturer of in vitro diagnostic (IVD) instruments intended for self-testing.
This document is also applicable to apparatus and equipment intended to be used with IVD instruments for self-testing.
This document can also be applicable to accessories.
This document does not apply to:
a) instructions for instrument servicing or repair;
b) IVD reagents, including calibrators and control materials for use in control of the reagent;
c) IVD instruments for professional use.

  • Standard
    21 pages
    English language
    e-Library read for
    1 day

This document specifies requirements for information supplied by the manufacturer of in vitro diagnostic (IVD) reagents, calibrators, and controls intended for self-testing.
This document can also be applicable to accessories.
This document is applicable to the labels for outer and immediate containers and to the instructions for use.
This document does not apply to:
a) IVD instruments or equipment;
b) IVD reagents for professional use.

  • Standard
    24 pages
    English language
    e-Library read for
    1 day

This document defines concepts, establishes general principles, and specifies essential requirements for information supplied by the manufacturer of IVD medical devices.
This document does not address language requirements since that is the domain of national laws and regulations.
This document does not apply to:
a) IVD medical devices for performance evaluation (e.g. for investigational use only);
b) shipping documents;
c) material safety data sheets / Safety Data Sheets;
d) marketing information (consistent with applicable legal requirements).
.

  • Standard
    65 pages
    English language
    e-Library read for
    1 day

This document specifies the requirements for a terminally sterilized medical device to be designated ‘STERILE’. Part 2 of this European standard specifies the requirements for an aseptically processed medical device to be designated "STERILE".
NOTE   For the purpose of the EU Directive(s) for medical devices (see Bibliography), designation of a medical device as ‘STERILE’ is only permissible when a validated sterilization process has been applied. Requirements for validation and routine control of processes for the sterilization of medical devices are specified in EN ISO 11135, EN ISO 11137, EN ISO 14160, EN ISO 14937, EN ISO 17665-1, EN ISO 20857, EN ISO 25424 and ISO 22441.

  • Standard
    14 pages
    English language
    e-Library read for
    1 day

This document specifies requirements for information supplied by the manufacturer of in vitro diagnostic (IVD) instruments intended for professional use.
This document also applies to apparatus and equipment intended to be used with IVD instruments for professional use.
This document can also be applicable to accessories.
This document does not apply to:
a) instructions for instrument servicing or repair;
b) IVD reagents, including calibrators and control materials for use in control of the reagent;
c) IVD instruments for self-testing.

  • Standard
    22 pages
    English language
    e-Library read for
    1 day

This document specifies requirements for information supplied by the manufacturer of in vitro diagnostic (IVD) reagents, calibrators and controls intended for professional use.
This document can also be applicable to accessories.
This document is applicable to the labels for outer and immediate containers and to the instructions for use.
This document does not apply to:
a) IVD instruments or equipment;
b) IVD reagents for self-testing.

  • Standard
    24 pages
    English language
    e-Library read for
    1 day

This document defines concepts, establishes general principles, and specifies essential requirements for information supplied by the manufacturer of IVD medical devices.
This document does not address language requirements since that is the domain of national laws and regulations.
This document does not apply to:
a) IVD medical devices for performance evaluation (e.g. for investigational use only);
b) shipping documents;
c) material safety data sheets / Safety Data Sheets;
d) marketing information (consistent with applicable legal requirements).
.

  • Standard
    65 pages
    English language
    e-Library read for
    1 day

This document specifies requirements for information supplied by the manufacturer of in vitro diagnostic (IVD) reagents, calibrators, and controls intended for self-testing.
This document can also be applicable to accessories.
This document is applicable to the labels for outer and immediate containers and to the instructions for use.
This document does not apply to:
a) IVD instruments or equipment;
b) IVD reagents for professional use.

  • Standard
    24 pages
    English language
    e-Library read for
    1 day

This document specifies requirements for information supplied by the manufacturer of in vitro diagnostic (IVD) instruments intended for professional use.
This document also applies to apparatus and equipment intended to be used with IVD instruments for professional use.
This document can also be applicable to accessories.
This document does not apply to:
a) instructions for instrument servicing or repair;
b) IVD reagents, including calibrators and control materials for use in control of the reagent;
c) IVD instruments for self-testing.

  • Standard
    22 pages
    English language
    e-Library read for
    1 day

This document specifies requirements for information supplied by the manufacturer of in vitro diagnostic (IVD) instruments intended for self-testing.
This document is also applicable to apparatus and equipment intended to be used with IVD instruments for self-testing.
This document can also be applicable to accessories.
This document does not apply to:
a) instructions for instrument servicing or repair;
b) IVD reagents, including calibrators and control materials for use in control of the reagent;
c) IVD instruments for professional use.

  • Standard
    21 pages
    English language
    e-Library read for
    1 day

This document provides requirements for the development, validation and routine control of moist heat sterilization processes for medical devices. It also contains guidance which is intended to explain the requirements set forth in the normative sections. The guidance given is intended to promote good practice related to moist heat sterilization processes according to this document. The application within industrial and health care settings is considered.

  • Standard
    172 pages
    English language
    e-Library read for
    1 day

This document specifies the general requirements for, and offers guidance on, processes, programs and procedures for development, validation and routine control of aseptic processing of health care products.
This document includes requirements and guidance relative to the overall topic of aseptic processing.
Specific requirements and guidance on various specialized processes and methods related to sterilizing filtration, lyophilization, clean-in place (CIP) technologies, sterilization in place (SIP) and isolator systems are given in the other parts of the ISO 13408 series.

  • Standard
    91 pages
    English language
    e-Library read for
    1 day

This document provides requirements for the development, validation and routine control of moist heat sterilization processes for medical devices. It also contains guidance which is intended to explain the requirements set forth in the normative sections. The guidance given is intended to promote good practice related to moist heat sterilization processes according to this document. The application within industrial and health care settings is considered.

  • Standard
    172 pages
    English language
    e-Library read for
    1 day

This document specifies the general requirements for, and offers guidance on, processes, programs and procedures for development, validation and routine control of aseptic processing of health care products.
This document includes requirements and guidance relative to the overall topic of aseptic processing.
Specific requirements and guidance on various specialized processes and methods related to sterilizing filtration, lyophilization, clean-in place (CIP) technologies, sterilization in place (SIP) and isolator systems are given in the other parts of the ISO 13408 series.

  • Standard
    91 pages
    English language
    e-Library read for
    1 day

This document defines good study practice for the planning, design, conduct, recording and reporting of clinical performance studies carried out to assess the clinical performance and safety of in vitro diagnostic (IVD) medical devices for regulatory purposes.
NOTE 1    The purpose of these studies is to assess the ability of an IVD medical device in the hands of the intended user, to yield results pertaining to a particular medical condition or physiological/pathological state, in the intended population.
The document is not intended to describe whether the technical specifications of the IVD medical device in question are adequately addressed by the clinical performance study.
This document identifies the principles that underpin clinical performance studies and specifies general requirements intended to
—          ensure the conduct of the clinical performance study will lead to reliable and robust study results,
—          define the responsibilities of the sponsor and principal investigator,
—          assist sponsors, clinical research organization, investigators, ethics committees, regulatory authorities and other bodies involved in the conformity assessment of IVD medical devices, and
—          protect the rights, safety, dignity and well-being of the subjects providing specimens for use in clinical performance studies.
Analytical performance studies are out of the scope of this document.
NOTE 2    When the collection of specimens specifically for the analytical performance study creates an additional collection risk for subjects, some of the elements of this document (particularly the annexes) can be useful for ensuring subject safety.
Clinical performance studies that are performed for reasons other than pre- and post-market regulatory purposes, such as for re-imbursement purposes, are out of the scope of this document.
NOTE 3    Some of the elements of this document can be useful for the design of such studies, including subject safety and data integrity.
This document does not include safety information for laboratory workers or other personnel collecting the study specimens.
NOTE 4    Such information is included in other publications[1][12][13].
NOTE 5    Users of this document can consider whether other standards and/or requirements also apply to the IVD medical device which is the subject of the clinical performance study, for instance, in the situation for which there is an IVD medical device and a medical device used in an integrated system (e.g. a lancet, an IVD test strip, and a glucose meter), aspects of both this document and ISO 14155 can be considered.

  • Standard
    72 pages
    English language
    e-Library read for
    1 day

This document defines good study practice for the planning, design, conduct, recording and reporting of clinical performance studies carried out to assess the clinical performance and safety of in vitro diagnostic (IVD) medical devices for regulatory purposes.
NOTE 1    The purpose of these studies is to assess the ability of an IVD medical device in the hands of the intended user, to yield results pertaining to a particular medical condition or physiological/pathological state, in the intended population.
The document is not intended to describe whether the technical specifications of the IVD medical device in question are adequately addressed by the clinical performance study.
This document identifies the principles that underpin clinical performance studies and specifies general requirements intended to
—          ensure the conduct of the clinical performance study will lead to reliable and robust study results,
—          define the responsibilities of the sponsor and principal investigator,
—          assist sponsors, clinical research organization, investigators, ethics committees, regulatory authorities and other bodies involved in the conformity assessment of IVD medical devices, and
—          protect the rights, safety, dignity and well-being of the subjects providing specimens for use in clinical performance studies.
Analytical performance studies are out of the scope of this document.
NOTE 2    When the collection of specimens specifically for the analytical performance study creates an additional collection risk for subjects, some of the elements of this document (particularly the annexes) can be useful for ensuring subject safety.
Clinical performance studies that are performed for reasons other than pre- and post-market regulatory purposes, such as for re-imbursement purposes, are out of the scope of this document.
NOTE 3    Some of the elements of this document can be useful for the design of such studies, including subject safety and data integrity.
This document does not include safety information for laboratory workers or other personnel collecting the study specimens.
NOTE 4    Such information is included in other publications[1][12][13].
NOTE 5    Users of this document can consider whether other standards and/or requirements also apply to the IVD medical device which is the subject of the clinical performance study, for instance, in the situation for which there is an IVD medical device and a medical device used in an integrated system (e.g. a lancet, an IVD test strip, and a glucose meter), aspects of both this document and ISO 14155 can be considered.

  • Standard
    72 pages
    English language
    e-Library read for
    1 day

2021-11-09-JO-Following roundtable discussion on 2021-11-09 between EC desk officer, CCMC and CEN/TC 102, it was agreed that the assessment at ENQ will be performed by the Medical Device EC desk officer for a finale assessment at Enquiry. If positive, CEN/TC 102 may decide to skip the Formal Vote and proceed to publication.

  • Amendment
    33 pages
    English language
    e-Library read for
    1 day

2021-11-09-JO-Following roundtable discussion on 2021-11-09 between EC desk officer, CCMC and CEN/TC 102, it was agreed that the assessment at ENQ will be performed by the Medical Device EC desk officer for a finale assessment at Enquiry. If positive, CEN/TC 102 may decide to skip the Formal Vote and proceed to publication.

  • Amendment
    23 pages
    English language
    e-Library read for
    1 day

2021-11-09-JO-Following roundtable discussion on 2021-11-09 between EC desk officer, CCMC and CEN/TC 102, it was agreed that the assessment at ENQ will be performed by the Medical Device EC desk officer for a finale assessment at Enquiry. If positive, CEN/TC 102 may decide to skip the Formal Vote and proceed to publication.

  • Amendment
    23 pages
    English language
    e-Library read for
    1 day

2021-11-09-JO-Following roundtable discussion on 2021-11-09 between EC desk officer, CCMC and CEN/TC 102, it was agreed that the assessment at ENQ will be performed by the Medical Device EC desk officer for a finale assessment at Enquiry. If positive, CEN/TC 102 may decide to skip the Formal Vote and proceed to publication.

  • Amendment
    33 pages
    English language
    e-Library read for
    1 day

This document specifies requirements for the development and validation of processes for packaging medical devices that are terminally sterilized. These processes include forming, sealing and assembly of preformed sterile barrier systems, sterile barrier systems and packaging systems.
It is applicable to industry, to health care facilities, and to wherever medical devices are packaged and sterilized.
It does not cover all requirements for packaging medical devices that are manufactured aseptically. Additional requirements can be necessary for drug/device combinations.

  • Amendment
    11 pages
    English language
    e-Library read for
    1 day

This document specifies requirements and test methods for materials, preformed sterile barrier systems, sterile barrier systems and packaging systems that are intended to maintain sterility of terminally sterilized medical devices until the point of use.
It is applicable to industry, to health care facilities, and to wherever medical devices are placed in sterile barrier systems and sterilized.
It does not cover all requirements for sterile barrier systems and packaging systems for medical devices that are manufactured aseptically. Additional requirements can be necessary for drug/device combinations.
It does not describe a quality assurance system for control of all stages of manufacture.
It does not apply to packaging materials and/or systems used to contain a contaminated medical device during transportation of the item to the site of reprocessing or disposal.

  • Amendment
    11 pages
    English language
    e-Library read for
    1 day

The scope of the Amendment is the same as EN 61010-2-101:2017. It applies to equipment intended for in vitro diagnostic (IVD) medical purposes, including self-test IVD medical purposes.

  • Amendment
    17 pages
    English language
    e-Library read for
    1 day

NEW!IEC 61010-2-101:2018 is available as IEC 61010-2-101:2018 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.

IEC 61010-2-101:2018 applies to equipment intended for in vitro diagnostic (IVD) medical purposes, including self-test IVD medical purposes. It has the status of a group safety publication, as specified in IEC Guide 104. This document has been prepared in close collaboration with Working Group CENELEC BTTF 88.1. This third edition cancels and replaces the second edition published in 2015. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition:
- adaptation of changes introduced by Amendment 1 of IEC 61010-1;
- added tolerance for stability of AC voltage test equipment to Clause 6.
This Part 2-101 is intended to be used in conjunction with IEC 61010-1. It was established on the basis of the third edition (2010) and its Amendment 1 (2016).

  • Standard
    19 pages
    English language
    e-Library read for
    1 day

NEW!IEC 61010-2-101:2018 is available as IEC 61010-2-101:2018 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC 61010-2-101:2018 applies to equipment intended for in vitro diagnostic (IVD) medical purposes, including self-test IVD medical purposes. It has the status of a group safety publication, as specified in IEC Guide 104. This document has been prepared in close collaboration with Working Group CENELEC BTTF 88.1. This third edition cancels and replaces the second edition published in 2015. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: - adaptation of changes introduced by Amendment 1 of IEC 61010-1; - added tolerance for stability of AC voltage test equipment to Clause 6. This Part 2-101 is intended to be used in conjunction with IEC 61010-1. It was established on the basis of the third edition (2010) and its Amendment 1 (2016).

  • Standard
    19 pages
    English language
    e-Library read for
    1 day

The scope of the Amendment is the same as EN 61010-2-101:2017. It applies to equipment intended for in vitro diagnostic (IVD) medical purposes, including self-test IVD medical purposes.

  • Amendment
    17 pages
    English language
    e-Library read for
    1 day

This document specifies requirements and test methods for materials, preformed sterile barrier systems, sterile barrier systems and packaging systems that are intended to maintain sterility of terminally sterilized medical devices until the point of use.
It is applicable to industry, to health care facilities, and to wherever medical devices are placed in sterile barrier systems and sterilized.
It does not cover all requirements for sterile barrier systems and packaging systems for medical devices that are manufactured aseptically. Additional requirements can be necessary for drug/device combinations.
It does not describe a quality assurance system for control of all stages of manufacture.
It does not apply to packaging materials and/or systems used to contain a contaminated medical device during transportation of the item to the site of reprocessing or disposal.

  • Amendment
    11 pages
    English language
    e-Library read for
    1 day

This document specifies requirements for the development and validation of processes for packaging medical devices that are terminally sterilized. These processes include forming, sealing and assembly of preformed sterile barrier systems, sterile barrier systems and packaging systems.
It is applicable to industry, to health care facilities, and to wherever medical devices are packaged and sterilized.
It does not cover all requirements for packaging medical devices that are manufactured aseptically. Additional requirements can be necessary for drug/device combinations.

  • Amendment
    11 pages
    English language
    e-Library read for
    1 day

This document specifies terminology, principles and a process for risk management of medical devices, including software as a medical device and in vitro diagnostic medical devices. The process described in this document intends to assist manufacturers of medical devices to identify the hazards associated with the medical device, to estimate and evaluate the associated risks, to control these risks, and to monitor the effectiveness of the controls.
The requirements of this document are applicable to all phases of the life cycle of a medical device. The process described in this document applies to risks associated with a medical device, such as risks related to biocompatibility, data and systems security, electricity, moving parts, radiation, and usability.
The process described in this document can also be applied to products that are not necessarily medical devices in some jurisdictions and can also be used by others involved in the medical device life cycle.
This document does not apply to:
— decisions on the use of a medical device in the context of any particular clinical procedure; or
— business risk management.
This document requires manufacturers to establish objective criteria for risk acceptability but does not specify acceptable risk levels.
Risk management can be an integral part of a quality management system. However, this document does not require the manufacturer to have a quality management system in place.
NOTE Guidance on the application of this document can be found in ISO/TR 24971[9].

  • Amendment
    9 pages
    English language
    e-Library read for
    1 day

Frequently Asked Questions

An EU Regulation is a binding legislative act that must be applied in its entirety across the European Union. Unlike directives, regulations do not need to be transposed into national law and are directly applicable in all member states. Regulations are used when uniform application across all EU countries is essential.

Regulation 2017/746 covers "Regulation (EU) 2017/746 of the European Parliament and of the Council of 5 April 2017 on in vitro diagnostic medical devices and repealing Directive 98/79/EC and Commission Decision 2010/227/EU". This legislation has harmonized European standards (hENs) that provide a presumption of conformity with its essential requirements. There are 106 standards associated with this regulation.

Harmonized standards under 2017/746 are European standards (ENs) developed by CEN, CENELEC, or ETSI in response to a mandate from the European Commission. When these standards are cited in the Official Journal of the European Union, products manufactured in conformity with them benefit from a presumption of conformity with the essential requirements of 2017/746, facilitating CE marking and free movement within the European Economic Area.