This document specifies a chromatographic method for the determination of levoglucosan in aqueous or organic extracts of filter samples collected in accordance with EN 12341:2023 [5]. The method has been tested for concentrations of ca. 10 ng/m3 up to ca. 3 000 ng/m3 with a sampling duration of 24 h. The procedure is also suitable for the determination of galactosan and mannosan.
Depending on the analysis instrumentation used, the carbohydrates inositol, glycerol, threitol/erythritol, xylitol, arabitol, sorbitol, mannitol, threalose, mannose, glucose, galactose and fructose can also be determined. However, no performance characteristics are given for these compounds in this document.

  • Draft
    42 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies the sampling and analysis of phthalates in indoor air and describes the sampling and analysis of phthalates in house dust and in solvent wipe samples of surfaces by means of gas chromatography-mass spectrometry (GC-MS). Two alternative sampling, sample preparation and sample introduction methods, whose comparability has been proven in an interlaboratory test, are specified for indoor air[1]: — sorbent tubes sampling with subsequent thermal desorption GC-MS, and — sampling by adsorption and subsequent solvent extraction and injection to GC-MS. Additional adsorbents that can be used are described in Annex B. Depending on the sampling method, the compounds dimethyl phthalate to diisoundecylphthalate can be analysed in house dust as described in Annex D. The investigation of house dust samples is only appropriate as a screening method. This investigation only results in indicative values and is not acceptable for a final assessment of a potential need for action. Dimethyl phthalate to diisoundecylphthalate can be analysed in solvent wipe samples as described in Annex C. Solvent wipe samples are suitable for non-quantitative source identification. NOTE In principle, the method is also suitable for the analysis of other phthalates, adipates and cyclohexane dicarboxylic acid esters, but this is confirmed by determination of the performance characteristics in each case. General information on phthalates are given in Annex A.

  • Standard
    45 pages
    English language
    sale 15% off

This document specifies a standard method for determining the particle number concentration in ambient air in a range up to about 107 cm–3 for averaging times equal to or larger than 1 min. The standard method is based on a Condensation Particle Counter (CPC) operated in the counting mode and an appropriate dilution system for concentrations exceeding the counting mode range. It also defines the performance characteristics and the minimum requirements of the instruments to be used. The lower and upper sizes considered within this document are 10 nm and a few micrometres, respectively. This document gives guidance on sampling, operation, data processing and QA/QC procedures including calibration parameters.

  • Draft
    53 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document provides definitions of the quantities measured by various candidate methods, their basic principles, and their advantages and disadvantages.
Currently no traceable primary reference materials are available for EC and OC analyses. This document provides guidance to test the equivalence between candidate methods and EN 16909 for EC and/or OC determination(s), based on EN 16450.

  • Draft
    17 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies a general laboratory test method for the determination of the area specific emission rate of volatile organic compounds (VOCs) from samples of newly produced building products or furnishing under defined climate conditions. The method can also, in principle, be applied to samples of aged products. The emission data obtained can be used to calculate concentrations in a model room (see Table B.1).
This document is applicable to various emission test chambers used for the determination of the emission of VOCs from building products or furnishing.
This document is also applicable to samples of wood-based panels and other building products, in order to determine the emission rate of formaldehyde.
NOTE            In principle, this document can be applied to the study of any gas phase emissions from samples of building products and furnishing.

  • Draft
    23 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies the sampling procedures, transport conditions, storage and substrate used that can affect emissions of volatile organic compounds for three types of building products or furnishing: solid, liquid and combined. For individual products, the preparation of a test specimen for each type is specified.

  • Draft
    20 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies a general laboratory test method for the determination of the area specific emission rate of volatile organic compounds (VOCs) from samples of newly produced building products or furnishing under defined climate conditions. The method can also, in principle, be applied to samples of aged products. The emission data obtained can be used to calculate concentrations in a model room (see Table B.1). This document is applicable to various emission test chambers used for the determination of the emission of VOCs from building products or furnishing. This document is also applicable to samples of wood-based panels and other building products, in order to determine the emission rate of formaldehyde. NOTE In principle, this document can be applied to the study of any gas phase emissions from samples of building products and furnishing.

  • Standard
    16 pages
    English language
    sale 15% off
  • Standard
    17 pages
    French language
    sale 15% off

This document specifies the sampling procedures, transport conditions, storage and substrate used that can affect emissions of volatile organic compounds for three types of building products or furnishing: solid, liquid and combined. For individual products, the preparation of a test specimen for each type is specified.

  • Standard
    14 pages
    English language
    sale 15% off
  • Standard
    15 pages
    French language
    sale 15% off

This document describes a test procedure for sampling, elution, detection, and quantification of N-nitrosamines in air samples derived from a test chamber according to EN 16516:2017+A1:2020. The following N-nitrosamines are covered:
-Nitrosodimethylamine, CAS No. 62-75-9,
- N-Nitrosomethylethylamine, CAS No. 10595-95-6,
- N-Nitrosodiethylamine, CAS No. 55-18-5,
- N-Nitrosodipropylamine, CAS No. 621-64-7,
- N-Nitrosodiisopropylamine, CAS No. 601-77-4,
- N-Nitrosodibutylamine, CAS No. 924-16-3,
- N-Nitrosopiperidine, CAS No. 100-75-4,
- N-Nitrosopyrrolidine, CAS No. 930-55-2 and
- N-Nitrosomorpholine, CAS No. 59-89-2.

  • Technical specification
    19 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document describes the pros and cons for the different methods for reporting the potential release of dangerous substances into soil, groundwater or surface water and indoor air, which are:
—   level (or declared values); and
—   classes;
as defined in the Construction Products Regulation (CPR).
In addition, the pros and cons of additional methods based on discussion in CEN/TCs and WGs are described, which are:
—   categories; and
—   manufacturer’s declaration.

  • Technical report
    30 pages
    English language
    sale 10% off
    e-Library read for
    1 day
  • Standard
    1 page
    English language
    sale 15% off
  • Standard
    1 page
    French language
    sale 15% off

This document describes the pros and cons for the different methods for reporting the potential release of dangerous substances into soil, groundwater or surface water and indoor air, which are:
—   level (or declared values); and
—   classes;
as defined in the Construction Products Regulation (CPR).
In addition, the pros and cons of additional methods based on discussion in CEN/TCs and WGs are described, which are:
—   categories; and
—   manufacturer’s declaration.

  • Technical report
    30 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This part of EN 14662 gives general guidance for the sampling and analysis of benzene in air by pumped sampling, thermal desorption and capillary gas chromatography.
This part of EN 14662 is in accordance with the generic methodology selected as the basis of the European Union reference method for the determination of benzene in ambient air [1] for the purpose of comparison of measurement results with limit values with a one-year reference period.
This part of EN 14662 is valid for the measurement of benzene in a concentration range of approximately 0,5 μg/m3 to 50 μg/m3 in an air sample typically collected over a period of 24 hours.
The upper limit of the useful range is set by the sorptive capacity of the sorbent and by the linear dynamic range of the gas chromatograph column and detector or by the sample splitting capacity of the analytical instrumentation used. The lower limit of the useful range depends on the noise level of the detector and on blank levels of benzene and/or interfering artefacts on the sorbent. Artefacts are typically sub ng for sorbents, but higher levels of aromatic hydrocarbons have been noted in other sorbents. The detection limit will be approximately 1/10 of the lower concentration range.
Expansion of Scope:
This part of EN 14662 provides general guidance for the sampling of benzene using either a single sampler, which is changed manually after every exposure period, or a multi-tube sampler capable of storing and exposing multiple samples sequentially without user intervention.

  • Standard
    72 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document gives general guidance for the sampling and analysis of benzene in air by pumped sampling, thermal desorption and capillary gas chromatography.
This document is in accordance with the generic methodology selected as the basis of the European Union reference method for the determination of benzene in ambient air [1] for the purpose of comparison of measurement results with limit values with a one-year reference period.
This document is valid for the measurement of benzene in a concentration range of approximately 0,5 μg/m3 to 50 μg/m3. Air samples are typically collected over periods ranging from a few hours to 7 days.
The upper limit of the useful range is set by the sorptive capacity (the safe sampling volume) of the sorbent and by the linear dynamic range of the gas chromatograph column and detector or by the sample splitting capacity of the analytical instrumentation used. The lower limit of the useful range depends on the noise level of the detector and on blank levels of benzene and/or interfering artefacts on the sorbent. Artefacts are typically sub ng for graphitised carbon sorbents, but higher levels of aromatic hydrocarbons have been noted in other sorbents - e.g. porous polymers. The detection limit will be approximately 1/10 of the lower concentration range.
This document provides general guidance for the sampling of benzene using either a single sampler, which is changed manually after every exposure period, or a multi-sampler capable of storing and exposing multiple samples without user intervention.

  • Standard
    72 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies a laboratory test method for measuring perceived air quality using human subjects that can be used for assessing the performance of air cleaners removing gas-phase pollutants. The method describes the performance of gas-phase air cleaners with respect to removal of pollutants that can be sensed by human subjects. The method has a reference to sensory tests specified in ISO 16000-28. Air cleaners removing particles and aerosols (mechanical or electronic filters) can also remove pollutants responsible for sensory response. The method described in this document does not apply to testing of these air cleaners.

  • Standard
    16 pages
    English language
    sale 15% off

This document describes a test procedure for sampling, elution, detection, and quantification of N-nitrosamines in air samples derived from a test chamber according to EN 16516:2017+A1:2020. The following N-nitrosamines are covered:
-Nitrosodimethylamine, CAS No. 62-75-9,
- N-Nitrosomethylethylamine, CAS No. 10595-95-6,
- N-Nitrosodiethylamine, CAS No. 55-18-5,
- N-Nitrosodipropylamine, CAS No. 621-64-7,
- N-Nitrosodiisopropylamine, CAS No. 601-77-4,
- N-Nitrosodibutylamine, CAS No. 924-16-3,
- N-Nitrosopiperidine, CAS No. 100-75-4,
- N-Nitrosopyrrolidine, CAS No. 930-55-2 and
- N-Nitrosomorpholine, CAS No. 59-89-2.

  • Technical specification
    19 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies a determination of formaldehyde (HCHO) and other carbonyl compounds (aldehydes and ketones) in air. The method is specific to formaldehyde but, with modification, at least 12 other aromatic as well as saturated and unsaturated aliphatic carbonyl compounds can be detected and quantified. It is suitable for determination of formaldehyde and other carbonyl compounds in the approximate concentration range 1 µg/m3 to 1 mg/m3. The sampling method gives a time-weighted average (TWA) sample. It can be used for long-term (1 h to 24 h) or short-term (5 min to 60 min) sampling of air for formaldehyde.
This document specifies a sampling and analysis procedure for formaldehyde and other carbonyl compounds that involves collection from air on to adsorbent cartridges coated with 2,4-dinitrophenylhydrazine (DNPH) and subsequent analysis of the hydrazones formed by high performance liquid chromatography (HPLC) with detection by ultraviolet absorption[12],[16]. The method is not suitable for longer chained or unsaturated carbonyl compounds.

  • Standard
    32 pages
    English language
    sale 10% off
    e-Library read for
    1 day
  • Standard
    27 pages
    English language
    sale 15% off
  • Standard
    29 pages
    French language
    sale 15% off

This European Standard describes a standard method for determining the PM10 or PM2,5 mass concentrations of suspendedparticulate matter in ambient air by sampling the particulate matter on filters and weighing them by means of a balance.
Measurements are performed with samplers with inlet designs as specified in Annex A, operating at a nominal flow rate of 2,3 m3/h,over a nominal sampling period of 24 h. Measurement results are expressed in μg/m3, where the volume of air is the volume atambient conditions near the inlet at the time of sampling.
The range of application of this European Standard is for 24 h measurements from approximately 1 μg/m3 (i.e. the limit of detection ofthe standard measurement method expressed as its uncertainty) up to 150 μg/m3 for PM10 and 120 μg/m3 for PM2,5.
This European Standard describes procedures and gives requirements for the testing and use of so-called sequential samplers,equipped with a filter changer, suitable for extended stand-alone operation. Sequential samplers are commonly used throughout theEuropean Union for the measurement of concentrations in ambient air of PM10 or PM2,5. However, this European Standard does notexclude the use of single-filter samplers.
This European Standard represents an evolution of earlier European Standards (EN 12341:1998 and 2014, EN 14907:2005). Newequipment procured shall comply fully with this European Standard.
Older versions of these samplers, including those described in EN 12341:2014 Annex B, have a special status in terms of their use. These samplers can still be used for monitoring purposes and for ongoing quality control, provided that a well justified additionalallowance is made to their uncertainties
This European Standard also provides guidance for the selection and testing of filters with the aim of reducing the measurementuncertainty of the results obtained when applying this European Standard.

  • Standard
    61 pages
    English language
    sale 10% off
    e-Library read for
    1 day

To provide a concise overview of the following aspects of the application of reference rooms for the evaluation of emissions from products in indoor environments;
European dimension of the scope (regulations and schemes)
Evaluation of VOC emissions from building products: principles
Background history
Implementation in national regulations
Implementation in voluntary schemes
Broader application of the reference room (in addition to construction products)
Other possible dimensions of a reference room
Conclusion and references

  • Technical report
    20 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies the measurement methods and strategies for determining the total number of airborne particles per unit volume of air indoor, using a condensation particle counter (CPC) for particles approximately between 10 nm to 3 µm. NOTE As the particle number concentration is usually dominated by the ultrafine particle (UFP) fraction, the obtained result can be used as an approximation of the UFP concentration. Quality assurance, determination of the measurement uncertainty and minimal reporting information are also discussed in this document. This document is applicable to indoor environments as specified in ISO 16000-1. This document does not address the determination of bioaerosols or the chemical characterization of particles. Nevertheless, some bioaerosols can be detected by the CPC and then contribute to the measured count of particles.

  • Standard
    34 pages
    English language
    sale 15% off

This European Standard describes a standard method for determining the PM10 or PM2,5 mass concentrations of suspendedparticulate matter in ambient air by sampling the particulate matter on filters and weighing them by means of a balance.
Measurements are performed with samplers with inlet designs as specified in Annex A, operating at a nominal flow rate of 2,3 m3/h,over a nominal sampling period of 24 h. Measurement results are expressed in μg/m3, where the volume of air is the volume atambient conditions near the inlet at the time of sampling.
The range of application of this European Standard is for 24 h measurements from approximately 1 μg/m3 (i.e. the limit of detection ofthe standard measurement method expressed as its uncertainty) up to 150 μg/m3 for PM10 and 120 μg/m3 for PM2,5.
This European Standard describes procedures and gives requirements for the testing and use of so-called sequential samplers,equipped with a filter changer, suitable for extended stand-alone operation. Sequential samplers are commonly used throughout theEuropean Union for the measurement of concentrations in ambient air of PM10 or PM2,5. However, this European Standard does notexclude the use of single-filter samplers.
This European Standard represents an evolution of earlier European Standards (EN 12341:1998 and 2014, EN 14907:2005). Newequipment procured shall comply fully with this European Standard.
Older versions of these samplers, including those described in EN 12341:2014 Annex B, have a special status in terms of their use. These samplers can still be used for monitoring purposes and for ongoing quality control, provided that a well justified additionalallowance is made to their uncertainties
This European Standard also provides guidance for the selection and testing of filters with the aim of reducing the measurementuncertainty of the results obtained when applying this European Standard.

  • Standard
    61 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies a procedure for the assessment of the indoor air quality that is valid for all interior rooms in residential and non-residential buildings with natural or mechanical ventilation, in which people do not only stay temporarily. This document is applicable to indoor environments as defined in ISO 16000-1. The assessment of working materials in workrooms or workplaces in buildings, that are subject to statutory occupational safety specifications, are excluded from this document. In these rooms, only air constituents that do not originate from working materials can be assessed according to this document. It is not possible to define classes with exact values for the individual pollutants, as the corresponding limit and guide values differ in individual countries. In addition, the values relate to different observation periods. Aspects concerning electromagnetic fields, noise and vibrations and their effect on the indoor air quality are not the object of this document. The classification of further consequences and measures, such as organisational steps, structural engineering measures, renovation proposals, further human medicine appraisals and the like, are not the object of this document. NOTE This document applies to of all types of indoor environments occupied by all kinds of persons, including regular users, clients and workers.

  • Standard
    25 pages
    English language
    sale 15% off

To provide a concise overview of the following aspects of the application of reference rooms for the evaluation of emissions from products in indoor environments;
European dimension of the scope (regulations and schemes)
Evaluation of VOC emissions from building products: principles
Background history
Implementation in national regulations
Implementation in voluntary schemes
Broader application of the reference room (in addition to construction products)
Other possible dimensions of a reference room
Conclusion and references

  • Technical report
    20 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies a method for determination of volatile organic compounds (VOC) in indoor air and in air sampled for the determination of the emission from products or materials used in indoor environments (according to ISO 16000‑1) using test chambers and test cells. The method uses sorbent sampling tubes with subsequent thermal desorption (TD) and gas chromatographic (GC) analysis employing a capillary column and a mass spectrometric (MS) detector with or without an additional flame ionisation detector (FID)[13].
The method is applicable to the measurement of most GC-compatible vapour-phase organic compounds at concentrations ranging from micrograms per cubic metre to several milligrams per cubic metre. Many very volatile organic compounds (VVOC) and semi-volatile organic compounds (SVOC) can be analysed depending on the sorbents used.

  • Standard
    41 pages
    English language
    sale 10% off
    e-Library read for
    1 day
  • Standard
    36 pages
    English language
    sale 15% off
  • Standard
    36 pages
    English language
    sale 15% off
  • Standard
    39 pages
    French language
    sale 15% off
  • Standard
    39 pages
    French language
    sale 15% off

This document specifies the whole vehicle test chamber, the vapour sampling assembly and the operating conditions for the determination of volatile organic compounds (VOCs), and carbonyl compounds in vehicle cabin air. There are three measurements performed: one (for VOCs and carbonyl compounds) during the simulation of ambient conditions (ambient mode) at standard conditions of 23 °C - 25 °C with no air exchange; a second only for the measurement of formaldehyde at elevated temperatures (parking mode); and a third for VOCs and carbonyl compounds simulating driving after the vehicle has been parked in the sun starting at elevated temperatures (driving mode). For the simulation of the mean sun irradiation, a fixed irradiation in the whole vehicle test chamber is employed.
The VOC method is valid for measurement of non-polar and slightly polar VOCs in a concentration range of sub-micrograms per cubic metre up to several milligrams per cubic metre. Using the principles specified in this method, some semi-volatile organic compounds (SVOC) can also be analysed. Compatible compounds are those which can be trapped and released from the Tenax TA®[1] sorbent tubes described in ISO 16000‑6, which includes VOCs ranging in volatility from n-C6 to n-C16.
The sampling and analysis procedure for formaldehyde and other carbonyl compounds is performed by collecting air on to cartridges coated with 2,4-dinitrophenylhydrazine (DNPH) and subsequent analysis by high performance liquid chromatography (HPLC) with detection by ultraviolet absorption. Formaldehyde and other carbonyl compounds can be determined in the approximate concentration range 1 µg/m3 to 1 mg/m3.
The method is valid for passenger cars, as defined in ECE-TRANS-WP.29/1045.
This document gives guidelines for:
a) transport and storage of the test vehicles until the start of the test;
b) conditioning for the surroundings of the test vehicle and the test vehicle itself as well as the whole vehicle test chamber;
c) conditioning of the test vehicle prior to measurements;
d) simulation of ambient air conditions (ambient mode);
e) formaldehyde sampling at elevated temperatures (parking mode);
f) simulation of driving after the test vehicle has been parked in the sun (driving mode).
  
[1] Tenax TA® is the trade name of a product supplied by Buchem. This information is given for the convenience of users of this document and does not constitute an endorsement by ISO of the product named. Equivalent products may be used if they can be shown to lead to the same results.

  • Standard
    34 pages
    English language
    sale 10% off
    e-Library read for
    1 day
  • Standard
    29 pages
    English language
    sale 15% off
  • Standard
    29 pages
    English language
    sale 15% off

This document specifies a determination of formaldehyde (HCHO) and other carbonyl compounds (aldehydes and ketones) in air. The method is specific to formaldehyde but, with modification, at least 12 other aromatic as well as saturated and unsaturated aliphatic carbonyl compounds can be detected and quantified. It is suitable for determination of formaldehyde and other carbonyl compounds in the approximate concentration range 1 µg/m3 to 1 mg/m3. The sampling method gives a time-weighted average (TWA) sample. It can be used for long-term (1 h to 24 h) or short-term (5 min to 60 min) sampling of air for formaldehyde. This document specifies a sampling and analysis procedure for formaldehyde and other carbonyl compounds that involves collection from air on to adsorbent cartridges coated with 2,4-dinitrophenylhydrazine (DNPH) and subsequent analysis of the hydrazones formed by high performance liquid chromatography (HPLC) with detection by ultraviolet absorption[12],[16]. The method is not suitable for longer chained or unsaturated carbonyl compounds.

  • Standard
    32 pages
    English language
    sale 10% off
    e-Library read for
    1 day
  • Standard
    27 pages
    English language
    sale 15% off
  • Standard
    29 pages
    French language
    sale 15% off

This document describes a procedure for measuring the performance of general ventilation air cleaning
devices in their end use installed configuration. The performance measurements include removal
efficiency by particle size and the resistance to airflow. The test procedures include the definition and
reporting of the system airflow.
The procedure describes a method of counting ambient air particles of 0,3 μm to 5,0 μm upstream and
downstream of the in-place air cleaner(s) in a functioning air handling system. The procedure describes
the reduction of particle counter data to calculate removal efficiency by particle size.
Since filter installations vary dramatically in design and shape, a protocol for evaluating the suitability
of a site for filter evaluation and for system evaluation is included. When the evaluated site conditions
meet the minimum criteria established for system evaluation, the performance evaluation of the system
can also be performed according to this procedure.
This document also describes performance specifications for the testing equipment and defines
procedures for calculating and reporting the results. This document is not intended for measuring
performance of portable or movable room air cleaners or for evaluation of filter installations with an
expected filtration efficiency at or above 99 % or at or below 30 % when measured at 0,4 μm.

  • Standard
    51 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies a method for assessing the performance of photocatalytic inorganic materials contained in cement mortars and/or limes or ceramic-based matrices, paints or materials deposited as thin films or coatings on a variety of substrates for the photocatalytic abatement of nitric oxide in the gas phase. This method does not apply to the assessment of samples to be applied with flow perpendicular to the surface or flow permeating the surface itself as polymeric and paper filters, honeycomb structures and suchlike.
The performance for the photocatalytic sample under test is evaluated by measuring the degradation rate of nitric oxide (NO) using the method specified herein. The photocatalytic abatement rate is calculated from the observed rate by eliminating the effects of mass transfer. The intrinsic photocatalytic abatement rate is an intrinsic property of the material tested and makes it possible to distinguish the photocatalytic activities of various products with an absolute scale defined with physical and engineering meaning.
For the measurements and calculations described in this document the concentration of nitrogen oxides (NOx) is defined as the stoichiometric sum of nitric oxide (NO) and nitrogen dioxide (NO2).

  • Standard
    28 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This Technical Specification (TS) describes the general principles, including testing procedures and requirements, for the evaluation of performances of low-cost sensor systems for the monitoring of gaseous compounds in ambient air at fixed sites. The evaluation of sensor systems includes tests that shall be performed under prescribed laboratory and/or field conditions.
This TS is not intended for the test of sensors systems used for mobile devices, for the testing of networks of sensor nodes, or for indoor air monitoring although their potential importance is recognized and they could be the subjects of future TS documents.
Low-cost sensors are based on several principles of operations, e. g. amperometric sensors, metal oxides, optical sensors (Infra-Red absorption) etc. However, sensors share some common features, regarding their portability and low-cost compared to traditional reference methods. Typically, sensors are able to continuously monitor air pollution, with low response time ranging between a few tens of seconds and a few minutes.
The described procedure is applicable to the determination of the mass concentration of air pollutants. The pollutants that are considered in this TS consists of:
-the gaseous pollutants regulated under Directives 2008/50/EC: O3, NO2 and NO, CO, SO2 and benzene, in the range of concentrations expected in outdoor ambient air;
-CO2 as proxy for activities involving combustion processes or for CO2 evaporation from soil or water.
When applying the current Technical Specifications, the evaluation of sensors considers the thresholds, limits and averaging times that are defined into the Air Quality Directive (2008/50/EC)[1]. Generally, the Directive sets Limit Values consists of an annual average that is computed by averaging hourly values. For sensors, it can be useful to select shorter averaging time.
In order to rely on the results of tests this protocol, future users shall make sure that sensors will be implemented with the same configuration as the sensor submitted to this protocol. This can include: the same power supply, data acquisition, data processing, identical sampling/ protective box and periodicity of calibration. The sensor shall be submitted to the same regime of QA/QC and maintenance operation as during tests. In addition, it is strongly recommended that sensors measurements are periodically compared side-by-side with the reference method.
For the purpose of this technical specification sensor systems are significantly less expensive than reference methods for the same pollutant.

  • Technical specification
    88 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies the general principles, including testing procedures and requirements, for the classification of performance of low-cost sensor systems for the monitoring of gaseous compounds in ambient air at fixed sites. The classification of sensor systems includes tests that are performed under prescribed laboratory and field conditions.
The procedure described is applicable to the determination of the mass concentration of air pollutants. The pollutants that are considered in this document are the gaseous pollutants regulated under Directive 2008/50/EC (O3, NO/NO2/NOx, CO, SO2 and benzene) in the range of concentrations expected in ambient air.
This document provides a classification that is consistent with the requirements for indicative measurements and objective estimation defined in Directive 2008/50/EC. In addition, it provides a classification for applications (non-regulatory measurements) that require more relaxed performance criteria.
This document applies to sensor systems used as individual systems. It does not apply to sensor systems as part of a sensor network. However, for some applications (e.g. in cities) sensor systems are deployed as part of a sensor network. Annex A gives information on the use of sensor systems as nodes in a sensor network.
This document gives guidance on the testing of CO2 sensor systems in Annex B since, although not listed in Directive 2008/50/EC, CO2 is an interesting indicator as proxy for activities involving combustion processes or for CO2 evaporation from soil or water.

  • Technical specification
    88 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document provides specific rules for the assessment of the release on dangerous substances from glass products into indoor air of buildings in complement to the horizontal rules given in EN 16516.
This document addresses specifically products as mentioned in TC 129 Mandate - M135 Amendment 1 EN (2012), i.e. products covered by the following European Standards: EN 1036 2 and EN 16477 21. However, this document can also be applied to other glass products containing volatile organic compounds (VOC) such as: EN 1279 5, EN 15755 1 and EN 14449. Glass products that do not contain organic compounds are not in the scope of this document (see Annex A).
This document addresses the release of dangerous substances into indoor air from construction products, although it can also be applied to glass products used in other applications such as furniture.

  • Standard
    14 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document describes a method for assessing the performance of photocatalytic inorganic materials contained in cement mortars and/or limes or ceramic-based matrices, paints or materials deposited as thin films or coatings on a variety of substrates for the photocatalytic abatement of nitric oxide in the gas phase. This method is not suitable for the assessment of samples to be applied with flow perpendicular to the surface or flow permeating the surface itself as polymeric and paper filters, honeycomb structures and suchlike.
The performance for the photocatalytic sample under test is evaluated by measuring the degradation rate of nitric oxide (NO) using the method described herein. The photocatalytic abatement rate is calculated from the observed rate by eliminating the effects of mass transfer. The intrinsic photocatalytic abatement rate is an intrinsic property of the material tested and makes it possible to distinguish the photocatalytic activities of various products with an absolute scale defined with physical and engineering meaning.
For the measurements and calculations described in this document the concentration of nitrogen oxides (NOx) is defined as the stoichiometric sum of nitric oxide (NO) and nitrogen dioxide (NO2).

  • Standard
    28 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies the whole vehicle test chamber, the vapour sampling assembly and the operating conditions for the determination of volatile organic compounds (VOCs), and carbonyl compounds in vehicle cabin air. There are three measurements performed: one (for VOCs and carbonyl compounds) during the simulation of ambient conditions (ambient mode) at standard conditions of 23 °C - 25 °C with no air exchange; a second only for the measurement of formaldehyde at elevated temperatures (parking mode); and a third for VOCs and carbonyl compounds simulating driving after the vehicle has been parked in the sun starting at elevated temperatures (driving mode). For the simulation of the mean sun irradiation, a fixed irradiation in the whole vehicle test chamber is employed. The VOC method is valid for measurement of non-polar and slightly polar VOCs in a concentration range of sub-micrograms per cubic metre up to several milligrams per cubic metre. Using the principles specified in this method, some semi-volatile organic compounds (SVOC) can also be analysed. Compatible compounds are those which can be trapped and released from the Tenax TA®[1] sorbent tubes described in ISO 16000‑6, which includes VOCs ranging in volatility from n-C6 to n-C16. The sampling and analysis procedure for formaldehyde and other carbonyl compounds is performed by collecting air on to cartridges coated with 2,4-dinitrophenylhydrazine (DNPH) and subsequent analysis by high performance liquid chromatography (HPLC) with detection by ultraviolet absorption. Formaldehyde and other carbonyl compounds can be determined in the approximate concentration range 1 µg/m3 to 1 mg/m3. The method is valid for passenger cars, as defined in ECE-TRANS-WP.29/1045. This document gives guidelines for: a) transport and storage of the test vehicles until the start of the test; b) conditioning for the surroundings of the test vehicle and the test vehicle itself as well as the whole vehicle test chamber; c) conditioning of the test vehicle prior to measurements; d) simulation of ambient air conditions (ambient mode); e) formaldehyde sampling at elevated temperatures (parking mode); f) simulation of driving after the test vehicle has been parked in the sun (driving mode). [1] Tenax TA® is the trade name of a product supplied by Buchem. This information is given for the convenience of users of this document and does not constitute an endorsement by ISO of the product named. Equivalent products may be used if they can be shown to lead to the same results.

  • Standard
    34 pages
    English language
    sale 10% off
    e-Library read for
    1 day
  • Standard
    29 pages
    English language
    sale 15% off
  • Standard
    29 pages
    English language
    sale 15% off

This document specifies a method for determination of volatile organic compounds (VOC) in indoor air and in air sampled for the determination of the emission from products or materials used in indoor environments (according to ISO 16000‑1) using test chambers and test cells. The method uses sorbent sampling tubes with subsequent thermal desorption (TD) and gas chromatographic (GC) analysis employing a capillary column and a mass spectrometric (MS) detector with or without an additional flame ionisation detector (FID)[13]. The method is applicable to the measurement of most GC-compatible vapour-phase organic compounds at concentrations ranging from micrograms per cubic metre to several milligrams per cubic metre. Many very volatile organic compounds (VVOC) and semi-volatile organic compounds (SVOC) can be analysed depending on the sorbents used.

  • Standard
    41 pages
    English language
    sale 10% off
    e-Library read for
    1 day
  • Standard
    36 pages
    English language
    sale 15% off
  • Standard
    36 pages
    English language
    sale 15% off
  • Standard
    39 pages
    French language
    sale 15% off
  • Standard
    39 pages
    French language
    sale 15% off

This document describes and specifies the whole vehicle test chamber, the vapour sampling assembly and the operating conditions for the determination of volatile organic compounds (VOCs; for more information see Annex E), and carbonyl compounds in vehicle cabin air. There are three measurements performed: one (for VOCs and carbonyl compounds) during the simulation of ambient conditions (ambient mode) at standard conditions of 23 °C with no air exchange; a second only for the measurement of formaldehyde at elevated temperatures (parking mode); and a third for VOCs and carbonyl compounds simulating driving after the vehicle has been parked in the sun starting at elevated temperatures (driving mode). For the simulation of the mean sun irradiation, fixed irradiation in the whole vehicle test chamber is employed.
The VOC method is valid for measurement of non-polar and slightly polar VOCs in a concentration range of sub-micrograms per cubic metre up to several milligrams per cubic metre. Using the principles described in this method, some semi-volatile organic compounds (SVOC) can also be analysed. Compatible compounds are those which can be trapped and released from the Tenax TA®1) sorbent tubes described in ISO 16000‑6, which includes VOCs ranging in volatility from n-C6 to n-C16.
The sampling and analysis procedure for formaldehyde and other carbonyl compounds is performed by collecting air on to cartridges coated with 2,4-dinitrophenylhydrazine (DNPH) and subsequent analysis by high performance liquid chromatography (HPLC) with detection by ultraviolet absorption. Formaldehyde and other carbonyl compounds can be determined in the approximate concentration range 1Â ÎĽg/m3 to 1Â mg/m3.
This method applicable to trucks and buses, as defined in ISOÂ 3833:1977 3.1.1 to 3.1.6.
This document describes:
a) Transport and storage of the test vehicle until the start of the test.
b) Conditioning of the surroundings of the test vehicle and the test vehicle itself as well as the whole vehicle test chamber.
c) Conditioning of the test vehicle prior to measurements.
d) Simulation of ambient air conditions (ambient mode).
e) Formaldehyde sampling at elevated temperatures (parking mode).
f) Simulation of driving after the test vehicle has been parked in the sun (driving mode).
1)Tenax TA® is the trade name of a product supplied by Buchem. This information is given for the convenience of users of this document and does not constitute an endorsement by ISO of the product named. Equivalent products may be used if they can be shown to lead to the same results.

  • Standard
    24 pages
    English language
    sale 10% off
    e-Library read for
    1 day
  • Standard
    19 pages
    English language
    sale 15% off
  • Standard
    19 pages
    English language
    sale 15% off

This document describes and specifies the whole vehicle test chamber, the vapour sampling assembly and the operating conditions for the determination of volatile organic compounds (VOCs; for more information see Annex E), and carbonyl compounds in vehicle cabin air. There are three measurements performed: one (for VOCs and carbonyl compounds) during the simulation of ambient conditions (ambient mode) at standard conditions of 23 °C with no air exchange; a second only for the measurement of formaldehyde at elevated temperatures (parking mode); and a third for VOCs and carbonyl compounds simulating driving after the vehicle has been parked in the sun starting at elevated temperatures (driving mode). For the simulation of the mean sun irradiation, fixed irradiation in the whole vehicle test chamber is employed. The VOC method is valid for measurement of non-polar and slightly polar VOCs in a concentration range of sub-micrograms per cubic metre up to several milligrams per cubic metre. Using the principles described in this method, some semi-volatile organic compounds (SVOC) can also be analysed. Compatible compounds are those which can be trapped and released from the Tenax TA®1) sorbent tubes described in ISO 16000‑6, which includes VOCs ranging in volatility from n-C6 to n-C16. The sampling and analysis procedure for formaldehyde and other carbonyl compounds is performed by collecting air on to cartridges coated with 2,4-dinitrophenylhydrazine (DNPH) and subsequent analysis by high performance liquid chromatography (HPLC) with detection by ultraviolet absorption. Formaldehyde and other carbonyl compounds can be determined in the approximate concentration range 1 μg/m3 to 1 mg/m3. This method applicable to trucks and buses, as defined in ISO 3833:1977 3.1.1 to 3.1.6. This document describes: a) Transport and storage of the test vehicle until the start of the test. b) Conditioning of the surroundings of the test vehicle and the test vehicle itself as well as the whole vehicle test chamber. c) Conditioning of the test vehicle prior to measurements. d) Simulation of ambient air conditions (ambient mode). e) Formaldehyde sampling at elevated temperatures (parking mode). f) Simulation of driving after the test vehicle has been parked in the sun (driving mode). 1)Tenax TA® is the trade name of a product supplied by Buchem. This information is given for the convenience of users of this document and does not constitute an endorsement by ISO of the product named. Equivalent products may be used if they can be shown to lead to the same results.

  • Standard
    24 pages
    English language
    sale 10% off
    e-Library read for
    1 day
  • Standard
    19 pages
    English language
    sale 15% off
  • Standard
    19 pages
    English language
    sale 15% off

This document describes methods for determining air speed and flow direction, CO, NO and NO2 concentrations and visibility in road tunnels using direct-reading instruments. This document specifically excludes requirements relating to instrument conformance testing.

  • Standard
    37 pages
    English language
    sale 10% off
    e-Library read for
    1 day
  • Standard
    31 pages
    English language
    sale 15% off
  • Standard
    31 pages
    English language
    sale 15% off
  • Standard
    33 pages
    French language
    sale 15% off
  • Standard
    33 pages
    French language
    sale 15% off

This document specifies a laboratory test method using test chambers defined in ISO 16000-9 and further specified in EN 16516 and evaluation procedures for the determination of odours emitted from building products and materials.
Sampling, transport and storage of materials under test, as well as preparation of test specimens are described in ISO 16000-11 and further specified in EN 16516.

  • Standard
    42 pages
    English language
    sale 10% off
    e-Library read for
    1 day
  • Standard
    35 pages
    English language
    sale 15% off
  • Standard
    35 pages
    English language
    sale 15% off
  • Standard
    37 pages
    French language
    sale 15% off
  • Standard
    37 pages
    French language
    sale 15% off

This document describes methods for determining air speed and flow direction, CO, NO and NO2 concentrations and visibility in road tunnels using direct-reading instruments. This document specifically excludes requirements relating to instrument conformance testing.

  • Standard
    37 pages
    English language
    sale 10% off
    e-Library read for
    1 day
  • Standard
    31 pages
    English language
    sale 15% off
  • Standard
    31 pages
    English language
    sale 15% off
  • Standard
    33 pages
    French language
    sale 15% off
  • Standard
    33 pages
    French language
    sale 15% off

This document contains specifications for active sampling of bioaerosols from exhaust air flowing through a defined cross-section of a stack. It defines general principles that have to be taken into account during an isokinetic sampling campaign for bioaerosols by bubbling the exhaust air through a specific impinger designed for emission measurements.
In this document the application with culturable organisms is specified but the same principle might be applicable for non-cultural based methods (e.g. molecular and/or enzyme-based methods).
The impinger is designed to allow a sample volume flow of 1 m3/h to 1,8 m3/h, or 16 l/min to 30 l/min, respectively, and has been tested with regard to various microorganisms within broad concentration ranges [1; 2; 3; 4]

  • Standard
    57 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This European Standard specifies a horizontal reference method for the determination of emissions of regulated dangerous substances from construction products into indoor air. This method is applicable to volatile organic compounds, semi-volatile organic compounds, and very volatile aldehydes. It is based on the use of a test chamber and subsequent analysis of the organic compounds by GC-MS or HPLC.
NOTE 1   Supplemental information is given on indirect test methods (see Annex B) and on measuring very volatile organic compounds (see Annex C).
NOTE 2   This European Standard describes the overall procedure and makes use of existing standards mainly by normative reference, complemented when necessary with additional or modified normative requirements.

  • Standard
    61 pages
    English language
    sale 10% off
    e-Library read for
    1 day

The European Air Quality Directive (Directive on ambient air quality and cleaner air for Europe) identifies different uses for modelling: Assessment, planning, forecast and source apportionment (SA). This CEN/TS addresses source apportionment modelling and specifies performance tests to check whether given criteria for receptor oriented source apportionment (RM) are met. The scope of the tests set out in this CEN/TS is performance assessment of SA of particulate matter using RM in the context of the European Directives 2004/107/EC and 2008/50/EC (AQD) including the Commission Implementing Decision 2011/850/EU of 12 December 2011. The application of RM does not quantify the spatial origin of particulate matter hence this CEN/TS does not test spatial SA.
This CEN/TS addresses RM users: participants and organisers of source apportionment intercomparison studies as well as practitioners of individual source apportionment studies. This CEN/TS is suitable for the evaluation of results of a specific SA modelling system with respect to intercomparison reference values (a-priori known or calculated on the basis of participants' values, see 3.12) in the following application areas:
- Assessment of performance and uncertainties of a modelling system or modelling system set up using the indicators laid down in this CEN/TS.
- Testing and comparing different source apportionment outputs in a specific situation (applying an evaluation dataset) using the indicators laid down in this CEN/TS.
- QA/QC tests every time practitioners run a modelling system.
It should be noted for clarity that the procedures and calculations presented in this CEN/TS cannot be used to check the performance of a specific SA modelling result without having any a-priori reference information about the contributions of sources/source categories.
The principles of receptor oriented models are summarised in Annex A. An overview of uncertainty sources and recommendations about steps to follow in SA studies are provided in Annex B and Annex C.
There are different methodologies than RM widely used to accomplish SA, e.g. source oriented models. These other methodologies cover aspects of SA which are required in the AQD and are not addressed by RM. Performance assessment of such methodologies is out of the scope of this CEN/TS.

  • Technical specification
    29 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document describes procedures to assess the applicability of the standard method EN 16909 (determination of OC and EC deposited on filters) to particle size fractions up to 10 µm in aerodynamic diameter (50 % cut off).

  • Technical report
    21 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies the requirements and test methods of bamboo charcoal for air-purification applications.

  • Standard
    5 pages
    English language
    sale 15% off
  • Standard
    5 pages
    English language
    sale 15% off

This document specifies a method for the sampling and analysis of NH3 in ambient air using diffusive sampling.
It can be used for NH3 measurements at ambient levels, but the concentration range and exposure time are sampler dependent, and the end user is therefore advised to match the sampler type to the measurement requirement and to follow the operating instructions provided by the manufacturer.

  • Standard
    45 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies a laboratory test method using test chambers defined in ISO 16000-9 and further specified in EN 16516 and evaluation procedures for the determination of odours emitted from building products and materials. Sampling, transport and storage of materials under test, as well as preparation of test specimens are described in ISO 16000-11 and further specified in EN 16516.

  • Standard
    42 pages
    English language
    sale 10% off
    e-Library read for
    1 day
  • Standard
    35 pages
    English language
    sale 15% off
  • Standard
    35 pages
    English language
    sale 15% off
  • Standard
    37 pages
    French language
    sale 15% off
  • Standard
    37 pages
    French language
    sale 15% off

This document describes procedures to assess the applicability of the standard method EN 16909 (determination of OC and EC deposited on filters) to particle size fractions up to 10 µm in aerodynamic diameter (50 % cut off).

  • Technical report
    21 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This European Standard specifies a horizontal reference method for the determination of emissions of regulated dangerous substances from construction products into indoor air. This method is applicable to volatile organic compounds, semi-volatile organic compounds, and very volatile aldehydes. It is based on the use of a test chamber and subsequent analysis of the organic compounds by GC-MS or HPLC.
NOTE 1   Supplemental information is given on indirect test methods (see Annex B) and on measuring very volatile organic compounds (see Annex C).
NOTE 2   This European Standard describes the overall procedure and makes use of existing standards mainly by normative reference, complemented when necessary with additional or modified normative requirements.

  • Standard
    61 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This European Standard specifies a method for the sampling and analysis of NH3 in ambient air using
diffusive sampling.
It can be used for NH3 measurements at ambient levels but the concentration range and exposure time are sampler dependent and the end user shall use the working conditions for the various devices as recommended by the manufacturer.
Denuders may be used as a surrogate reference method, and for this reason their use is also described in this European Standard.

  • Standard
    45 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document describes a standard method for determining particle number size distributions in ambient air in the size range from 10 nm to 800 nm at total concentrations up to approximately 105 cm–3 with a time resolution of a few minutes. The standard method is based on a Mobility Particle Size Spectrometer (MPSS) used with a bipolar diffusion charger and a Condensation Particle Counter (CPC) as the detector. The document describes the performance characteristics and minimum requirements of the instruments and equipment to be used, and describes sampling, operation, data processing and QA/QC procedures, including calibration.

  • Technical specification
    63 pages
    English language
    sale 10% off
    e-Library read for
    1 day