ASTM E902-05
(Practice)Standard Practice for Checking the Operating Characteristics of X-Ray Photoelectron Spectrometers (Withdrawn 2011)
Standard Practice for Checking the Operating Characteristics of X-Ray Photoelectron Spectrometers (Withdrawn 2011)
SIGNIFICANCE AND USE
This practice should first be used to establish the operating characteristics of a particular X-ray photoelectron spectrometer at a time when the spectrometer performance is known to be optimum. Hence, the spectrometer settings in Section 5 and the expected performance figures given in Section 7 are to be taken only as guides, to be supplanted by the behavior of the user’actual spectrometer.
Subsequently, this practice should be used as a routine check, performed at frequent intervals with the same instrument settings, and the results compared with those obtained in 4.1. Significant deviation from optimum performance may indicate that the spectrometer requires recalibration or other maintenance.
Typical analysis settings should be used with this practice. The use of settings not specified by this practice is left to the discretion of the user, however, the settings should be recorded in accordance with Practice E 996 and the same settings should be used consistently whenever this practice is repeated, so that the results obtained will be directly comparable to previous results.
SCOPE
1.1 This practice covers a procedure for checking some of the operating characteristics of an X-ray photoelectron spectrometer. Tests herein provide checks of the repeatability of intensity measurements and the drift of the intensities with time. This practice may be conducted at the same time as the spectrometer energy calibration using Practice E 2108.
1.2 LimitationsThis practice is meant to augment, and not to replace, the calibration procedures recommended by the manufacturer of the spectrometer. This practice is also not meant to be used as a means of comparison between X-ray photoelectron spectrometers, but only as a self-consistent check of the operating characteristics of an individual spectrometer.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
WITHDRAWN RATIONALE
DESIG: E902 05 ^RATIONALE: Rationale
This practice covers a procedure for checking some of the operating characteristics of an X-ray photoelectron spectrometer.
Formerly under the jurisdiction of Committee E42 on Auger Electron Spectroscopy and X-Ray Photoelectron Spectroscopy, this practice was withdrawn without replacement in November 2011 due to its limited use by the industry.
General Information
Relations
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation:E902–05
Standard Practice for
Checking the Operating Characteristics of X-Ray
1
Photoelectron Spectrometers
This standard is issued under the fixed designation E902; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope ISO 21270 Surface ChemicalAnalysis — X-ray Photoelec-
tron and Auger Electron Spectrometers – Linearity of
1.1 This practice covers a procedure for checking some of
Intensity Scale
the operating characteristics of an X-ray photoelectron spec-
ISO 24237 Surface Chemical Analysis— X-ray Photoelec-
trometer. Tests herein provide checks of the repeatability of
tron Spectroscopy – Repeatability and Constancy of
intensity measurements and the drift of the intensities with
Intensity Scale
time. This practice may be conducted at the same time as the
spectrometer energy calibration using Practice E2108.
3. Terminology
1.2 Limitations—Thispracticeismeanttoaugment,andnot
3.1 Definitions—Terms used in X-ray photoelectron spec-
to replace, the calibration procedures recommended by the
troscopy are defined in Terminology E673.
manufacturer of the spectrometer. This practice is also not
3.2 Additionally, the following terms and abbreviations are
meant to be used as a means of comparison between X-ray
used throughout this practice:
photoelectron spectrometers, but only as a self-consistent
check of the operating characteristics of an individual spec-
trometer.
c = counts
1.3 This standard does not purport to address all of the
ch = channel
safety concerns, if any, associated with its use. It is the
cps = counts per second
responsibility of the user of this standard to establish appro-
eV = electron volts
priate safety and health practices and determine the applica-
i = numberofdatachannelsacquiredforthepeakof
bility of regulatory limitations prior to use. interest
n = number of channels
2. Referenced Documents
no. = number of
2
S/B = signal-to-background ratio
2.1 ASTM Standards:
sc = span
E673 Terminology Relating to Surface Analysis
Dx = step size, eV, between successive data channels
E996 Practice for Reporting Data in Auger Electron Spec-
2
A = peak area above background, mm
troscopy and X-ray Photoelectron Spectroscopy
B = background height, mm
E1078 Guide for Specimen Preparation and Mounting in
H = maximum peak height above background, mm
Surface Analysis
I = peak area intensity above background, c-eV/s
A
E2108 Practice for Calibration of the Electron Binding-
I = maximum signal intensity above background,
H
Energy Scale of an X-Ray Photoelectron Spectrometer
cps
2.2 ISO Standards
P = peak position on the binding energy scale, eV
FWHM = full width at half maximum
1 4. Significance and Use
This practice is under the jurisdiction of ASTM Committee E42 on Surface
AnalysisandisthedirectresponsibilityofSubcommitteeE42.03onAugerElectron
4.1 This practice should first be used to establish the
Spectroscopy and X-Ray Photoelectron Spectroscopy.
operating characteristics of a particular X-ray photoelectron
Current edition approved April 1, 2005. Published April 2005. Originally
spectrometer at a time when the spectrometer performance is
approved in 1982. Last previous edition approved in 1999 as E902–94(1999).
DOI: 10.1520/E0902-05.
known to be optimum. Hence, the spectrometer settings in
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Section 5 and the expected performance figures given in
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Section 7 are to be taken only as guides, to be supplanted by
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website. the behavior of the user’s actual spectrometer.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
1
---------------------- Page: 1 ----------------------
E902–05
4.2 Subsequently, this practice should be used as a routine on the binding energy scale (Note 1). In either case, acquire
check, performed at frequent intervals with the same instru- enough scans to collect at least 10000 counts at the peak
ment settings, and the results compared with those obtained in maximum.
4.1. Significant deviation from optimum performance may
NOTE 1—Forinstrumentswheretheminimumscanwidthislargerthan
indicate that the spectrometer requires recalibration or other
recommended, use the minimum allowable scan width. For instruments
maintenance.
where the energy interval cannot be set up in integer steps, use the closest
4.3 Typical analysis settings should be used with this allowed energy.
practice.Theuseof
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.