Standard Test Method for Temperature Calibration of Dynamic Mechanical Analyzers

SIGNIFICANCE AND USE
5.1 Dynamic mechanical analyzers monitor changes in the viscoelastic properties of a material as a function of temperature and frequency, providing a means to quantify these changes. In most cases, the value to be assigned is the temperature of the transition (or event) under study. Therefore, the temperature axis (abscissa) of all DMA thermal curves must be accurately calibrated by adjusting the apparent temperature scale to match the actual temperature over the temperature range of interest.
SCOPE
1.1 This test method describes the temperature calibration of dynamic mechanical analyzers (DMA) from –150°C to 300°C.  
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Specific precautionary statements are given in Note 7.

General Information

Status
Historical
Publication Date
31-Mar-2013
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM E1867-13 - Standard Test Method for Temperature Calibration of Dynamic Mechanical Analyzers
English language
5 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
REDLINE ASTM E1867-13 - Standard Test Method for Temperature Calibration of Dynamic Mechanical Analyzers
English language
5 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: E1867 − 13
StandardTest Method for
1
Temperature Calibration of Dynamic Mechanical Analyzers
This standard is issued under the fixed designation E1867; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope materials. This is accomplished by loading melting point
referencematerialsintoapolymertube,orwrappingthemwith
1.1 This test method describes the temperature calibration
polymer tape and subjecting it to a mechanical oscillation at
of dynamic mechanical analyzers (DMA) from –150°C to
either fixed or resonant frequency. The extrapolated onset of
300°C.
melting is identified by a rapid decrease in the ordinate signal
1.2 The values stated in SI units are to be regarded as
(the apparent storage modulus, stress, inverse strain or probe
standard. No other units of measurement are included in this
position). This onset is used for temperature calibration with
standard.
two melting point reference materials.
1.3 This standard does not purport to address all of the
5. Significance and Use
safety concerns, if any, associated with its use. It is the
5.1 Dynamic mechanical analyzers monitor changes in the
responsibility of the user of this standard to establish appro-
viscoelastic properties of a material as a function of tempera-
priate safety and health practices and determine the applica-
ture and frequency, providing a means to quantify these
bility of regulatory limitations prior to use. Specific precau-
changes. In most cases, the value to be assigned is the
tionary statements are given in Note 7.
temperature of the transition (or event) under study. Therefore,
2. Referenced Documents the temperature axis (abscissa) of all DMA thermal curves
2
must be accurately calibrated by adjusting the apparent tem-
2.1 ASTM Standards:
perature scale to match the actual temperature over the
E473 Terminology Relating to Thermal Analysis and Rhe-
temperature range of interest.
ology
E1142 Terminology Relating to Thermophysical Properties
6. Interferences
E2161 Terminology Relating to Performance Validation in
6.1 An increase or decrease in heating rates or change in
Thermal Analysis
purge gas type or rate from those specified may alter results.
6.2 Once the temperature calibration procedure has been
3. Terminology
executed, the measuring temperature sensor position shall not
3.1 Definitions:
be changed, nor shall it be in contact with the specimen or
3.1.1 The technical terms used in this test method are
specimen holder in a way that would impede movement. If the
defined in Terminologies E473, E1142, and E2161, including
temperature sensor position is changed or is replaced, then the
dynamic mechanical analysis, frequency, stress, strain and
entire calibration procedure shall be repeated.
storage modulus.
6.3 Once the temperature calibration has been executed, the
4. Summary of Test Method
geometry deformation (bending study, versus tensile, and the
like) shall not be changed. If the specimen testing geometry
4.1 An equation is developed for the linear correlation of
differs significantly from that of the calibrants, then the
experimentally observed program or sensor temperature and
calibration shall be repeated in the geometry matching that of
the actual melting temperature for known melting reference
specimen testing.
6.4 This method does not apply to calibration for shear or
1
ThistestmethodisunderthejurisdictionofASTMCommitteeE37onThermal
compressive geometries of deformation.
Measurements and is the direct responsibility of Subcommittee E37.10 on
Fundamental, Statistical and Mechanical Properties.
7. Apparatus
Current edition approved April 1, 2013. Published May 2013. Originally
7.1 The function of the apparatus is to hold a specimen of
approved in 1997. Last previous edition approved in 2012 as E1867 – 12. DOI:
10.1520/E1867-13.
uniform dimension so that the specimen acts as the elastic and
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
dissipative element in a mechanically oscillated system. Dy-
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
namic mechanic analyzers typically operate in one of several
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website. modes as outlined in Table 1.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
E1867 − 13
TABLE 1 Dynamic Mechanical Analyzer Modes of Operation
7.4 PTFE Tape, for wrapping metal point standards.
Mechanical Response
7.5 Calibrati
...

This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: E1867 − 12 E1867 − 13
Standard Test Method for
1
Temperature Calibration of Dynamic Mechanical Analyzers
This standard is issued under the fixed designation E1867; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This test method describes the temperature calibration of dynamic mechanical analyzers (DMA) from –150 °C–150°C to
500 °C.300°C.
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
limitations prior to use. Specific precautionary statements are given in Note 7.
2. Referenced Documents
2
2.1 ASTM Standards:
E473 Terminology Relating to Thermal Analysis and Rheology
E1142 Terminology Relating to Thermophysical Properties
E2161 Terminology Relating to Performance Validation in Thermal Analysis
3. Terminology
3.1 Definitions:
3.1.1 The technical terms used in this test method are defined in Terminologies E473, E1142, and E2161, including dynamic
mechanical analysis, frequency, stress, strain and storage modulus.
4. Summary of Test Method
4.1 An equation is developed for the linear correlation of experimentally observed program or sensor temperature and the actual
melting temperature for known melting reference materials. This is accomplished by loading melting point reference materials into
a polymer tube, or wrapping them with polymer tape and subjecting it to a mechanical oscillation at either fixed or resonant
frequency. The extrapolated onset of melting is identified by a rapid decrease in the ordinate signal (the apparent storage modulus,
stress, inverse strain or probe position). This onset is used for temperature calibration with two melting point reference materials.
5. Significance and Use
5.1 Dynamic mechanical analyzers monitor changes in the viscoelastic properties of a material as a function of temperature and
frequency, providing a means to quantify these changes. In most cases, the value to be assigned is the temperature of the transition
(or event) under study. Therefore, the temperature axis (abscissa) of all DMA thermal curves must be accurately calibrated by
adjusting the apparent temperature scale to match the actual temperature over the temperature range of interest.
6. Interferences
6.1 An increase or decrease in heating rates or change in purge gas type or rate from those specified may alter results.
6.2 Once the temperature calibration procedure has been executed, the measuring temperature sensor position shall not be
changed, nor shall it be in contact with the specimen or specimen holder in a way that would impede movement. If the temperature
sensor position is changed or is replaced, then the entire calibration procedure shall be repeated.
1
This test method is under the jurisdiction of ASTM Committee E37 on Thermal Measurements and is the direct responsibility of Subcommittee E37.10 on Fundamental,
Statistical and Mechanical Properties.
Current edition approved Sept. 1, 2012April 1, 2013. Published October 2012May 2013. Originally approved in 1997. Last previous edition approved in 20112012 as
E1867 – 11.E1867 – 12. DOI: 10.1520/E1867-12.10.1520/E1867-13.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
E1867 − 13
6.3 Once the temperature calibration has been executed, the geometry deformation (bending study, versus tensile, and the like)
shall not be changed. If the specimen testing geometry differs significantly from that of the calibrants, then the calibration shall
be repeated in the geometry matching that of specimen testing.
6.4 This method does not apply to calibration for shear or compressive geometries of deformation.
7. Apparatus
7.1 The function of the apparatus is to ho
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.