Standard Test Method for Conducting Cyclic Potentiodynamic Polarization Measurements for Localized Corrosion Susceptibility of Iron-, Nickel-, or Cobalt-Based Alloys

SIGNIFICANCE AND USE
3.1 An indication of the susceptibility to initiation of localized corrosion in this test method is given by the potential at which the anodic current increases rapidly. The more noble this potential, obtained at a fixed scan rate in this test, the less susceptible is the alloy to initiation of localized corrosion. The results of this test are not intended to correlate in a quantitative manner with the rate of propagation that one might observe in service when localized corrosion occurs.  
3.2 In general, once initiated, localized corrosion can propagate at some potential more electropositive than that at which the hysteresis loop is completed. In this test method, the potential at which the hysteresis loop is completed is determined at a fixed scan rate. In these cases, the more electropositive the potential at which the hysteresis loop is completed the less likely it is that localized corrosion will occur.  
3.3 If followed, this test method will provide cyclic potentiodynamic anodic polarization measurements that will reproduce data developed at other times in other laboratories using this test method for the two specified alloys discussed in 3.4. The procedure is used for iron-, nickel-, or cobalt-based alloys in a chloride environment.  
3.4 A standard potentiodynamic polarization plot is included. These reference data are based on the results from five different laboratories that followed the standard procedure, using specific alloys of Type 304 stainless steel, UNS S30400 and Alloy C-276, UNS N10276.3 Curves are included which have been constructed using statistical analysis to indicate the acceptable range of polarization curves.  
3.5 The availability of a standard test method, standard material, and standard plots should make it easy for an investigator to check his techniques to evaluate susceptibility to localized corrosion.
SCOPE
1.1 This test method covers a procedure for conducting cyclic potentiodynamic polarization measurements to determine relative susceptibility to localized corrosion (pitting and crevice corrosion) for iron-, nickel-, or cobalt-based alloys in a chloride environment. This test method also describes an experimental procedure which can be used to check one's experimental technique and instrumentation.  
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

General Information

Status
Historical
Publication Date
30-Apr-2014
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM G61-86(2014) - Standard Test Method for Conducting Cyclic Potentiodynamic Polarization Measurements for Localized Corrosion Susceptibility of Iron-, Nickel-, or Cobalt-Based Alloys
English language
5 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
REDLINE ASTM G61-86(2014) - Standard Test Method for Conducting Cyclic Potentiodynamic Polarization Measurements for Localized Corrosion Susceptibility of Iron-, Nickel-, or Cobalt-Based Alloys
English language
5 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: G61 − 86 (Reapproved 2014)
Standard Test Method for
Conducting Cyclic Potentiodynamic Polarization
Measurements for Localized Corrosion Susceptibility of
Iron-, Nickel-, or Cobalt-Based Alloys
ThisstandardisissuedunderthefixeddesignationG61;thenumberimmediatelyfollowingthedesignationindicatestheyearoforiginal
adoptionor,inthecaseofrevision,theyearoflastrevision.Anumberinparenthesesindicatestheyearoflastreapproval.Asuperscript
epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope whichtheanodiccurrentincreasesrapidly.Themorenoblethis
potential, obtained at a fixed scan rate in this test, the less
1.1 This test method covers a procedure for conducting
susceptible is the alloy to initiation of localized corrosion.The
cyclic potentiodynamic polarization measurements to deter-
resultsofthistestarenotintendedtocorrelateinaquantitative
mine relative susceptibility to localized corrosion (pitting and
manner with the rate of propagation that one might observe in
crevicecorrosion)foriron-,nickel-,orcobalt-basedalloysina
service when localized corrosion occurs.
chloride environment. This test method also describes an
experimental procedure which can be used to check one’s 3.2 Ingeneral,onceinitiated,localizedcorrosioncanpropa-
experimental technique and instrumentation. gate at some potential more electropositive than that at which
the hysteresis loop is completed. In this test method, the
1.2 The values stated in SI units are to be regarded as
potential at which the hysteresis loop is completed is deter-
standard. No other units of measurement are included in this
mined at a fixed scan rate. In these cases, the more electro-
standard.
positive the potential at which the hysteresis loop is completed
1.3 This standard does not purport to address all of the
the less likely it is that localized corrosion will occur.
safety concerns, if any, associated with its use. It is the
3.3 If followed, this test method will provide cyclic poten-
responsibility of the user of this standard to establish appro-
tiodynamic anodic polarization measurements that will repro-
priate safety and health practices and determine the applica-
duce data developed at other times in other laboratories using
bility of regulatory limitations prior to use.
this test method for the two specified alloys discussed in 3.4.
2. Referenced Documents
The procedure is used for iron-, nickel-, or cobalt-based alloys
in a chloride environment.
2.1 ASTM Standards:
D1193Specification for Reagent Water
3.4 A standard potentiodynamic polarization plot is in-
G3Practice for Conventions Applicable to Electrochemical
cluded.These reference data are based on the results from five
Measurements in Corrosion Testing
different laboratories that followed the standard procedure,
G5Reference Test Method for Making Potentiodynamic
using specific alloys of Type 304 stainless steel, UNS S30400
Anodic Polarization Measurements
and Alloy C-276, UNS N10276. Curves are included which
2.2 ASTM Adjuncts:
have been constructed using statistical analysis to indicate the
Standard Samples (set of two)
acceptable range of polarization curves.
3.5 The availability of a standard test method, standard
3. Significance and Use
material, and standard plots should make it easy for an
3.1 An indication of the susceptibility to initiation of local-
investigator to check his techniques to evaluate susceptibility
ized corrosion in this test method is given by the potential at
to localized corrosion.
This test method is under the jurisdiction of ASTM Committee G01 on
4. Apparatus
Corrosion of Metals and is the direct responsibility of Subcommittee G01.11 on
Electrochemical Measurements in Corrosion Testing.
4.1 The polarization cell should be similar to the one
Current edition approved May 1, 2014. Published May 2014. Originally
described in Reference Test Method G5. Other polarization
approvedin1986.Lastpreviouseditionapprovedin2009asG61–86(2009).DOI:
cells may be equally suitable.
10.1520/G0061-86R14.
4.1.1 The cell should have a capacity of about 1 L and
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
should have suitable necks or seals to permit the introduction
Standards volume information, refer to the standard’s Document Summary page on
of electrodes, gas inlet and outlet tubes, and a thermometer.
the ASTM website.
3 The Luggin probe-salt bridge separates the bulk solution from
Available from ASTM International Headquarters. Order Adjunct No.
ADJG0061. Original adjunct produced before 1995. thesaturatedcalomelreferenceelectrode.Theprobetipshould
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
G61 − 86 (2014)
TABLE 1 Chemical Composition of Alloys Used in the Round
Robin, Weight %
Type 304
Alloy C-276
Element Stainless Steel
(UNS N10276)
(UNS S30400)
Carbon 0.003 0.060
Chromium 15.29 18.46
Cobalt 2.05 .
Columbium . 0.11
Copper . 0.17
Iron 5.78 balance
Manganese 0.48 1.43
Molybdenum 16.03 0.17
Nickel balance 8.74
Phosphorus 0.018 0.029
Silicon 0.05 0.60
Sulfur 0.006 0.014
Vanadium 0.20 .
Tungsten 3.62 .
4.5 Current-Measuring Instruments (Note 1)—An instru-
mentthatiscapableofmeasuringacurrentaccuratelytowithin
1%oftheabsolutevalueoveracurrentrangebetween1.0and
10 µAshouldbeused.Manycommercialunitshaveabuild-in
instrument with an output as a voltage, which is preferred for
recording purposes. For the purpose of the present test a
FIG. 1 Schematic Diagram of Specimen Holder (see Footnotes 3
logarithmic output is desirable.
and 4)
4.6 Anodic Polarization Circuit—Ascanning potentiostat is
used for potentiodynamic measurements. Potential and current
are plotted continuously using an X-Y recorder and a logarith-
beadjustablesothatitcanbebroughtintocloseproximitywith
the working electrode. mic converter (contained in the potentiostat or incorporated
into the circuit) for the current. Commercially available units
4.2 Specimen Holder:
are suitable.
4.2.1 Specimens should be mounted in a suitable holder
4.7 Electrodes:
designed for flat strip, exposing 1 cm to the test solution (Fig.
1). Such specimen holders have been described in the litera- 4.7.1 The standard Type 304 stainless steel (UNS S30400)
andAlloy C-276 (UNS N10276) should be machined into flat
ture. It is important that the circular TFE-fluorocarbon gasket
0.625-in. (14-mm) diameter disks. The chemical compositions
be drilled and machined flat in order to minimize crevices.
of the alloys used in the round robin are listed in Table 1.
4.3 Potentiostat (Note 1)—Apotentiostat that will maintain
4.7.2 Counter Electrodes—The counter electrodes may be
an electrode potential within 1 mV of a preset value over a
prepared as described in ReferenceTest Method G5 or may be
widerangeofappliedcurrentsshouldbeused.Forthetypeand
prepared from high-purity platinum flat stock and wire. A
size of standard specimen supplied, the potentiostat should
suitable method would be to seal the platinum wire in glass
have a potential range of−1.0 to+1.6 V and an anodic current
tubing and introduce the platinum electrode assembly through
output range of 1.0 to 10 µA. Most commercial potentiostats
a sliding seal. Counter electrodes should have an area at least
meet the specific requirements for these types of measure-
twice as large as the test electrode.
ments.
4.7.3 Reference Electrode —A saturated calomel electrode
NOTE 1—These instrumental requirements are based upon values
with a controlled rate of leakage (about 3 µL/h) is recom-
typical of the instruments in the five laboratories that have provided the
mended. This type of electrode is durable, reliable, and
data used in determining the standard polarization plot.
commercially available. Precautions should be taken to ensure
4.4 Potential-Measuring Instruments (Note 1)—The
that it is maintained in the proper condition. The potential of
potential-measuring circuit should have a high input imped-
thecalomelelectrodeshouldbecheckedatperiodicintervalsto
11 14
ance on the order of 10 to 10 Ω to minimize current drawn
ensure the accuracy of the electrode.
from the system during measurements. Instruments should
5. Reagents and Materials
have sufficient sensitivity and accuracy to detect a change in
potential of 61 mV, usually included in commercial poten-
5.1 Purity of Reagents—Reagent grade chemicals shall be
...


This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: G61 − 86 (Reapproved 2009) G61 − 86 (Reapproved 2014)
Standard Test Method for
Conducting Cyclic Potentiodynamic Polarization
Measurements for Localized Corrosion Susceptibility of
Iron-, Nickel-, or Cobalt-Based Alloys
This standard is issued under the fixed designation G61; the number immediately following the designation indicates the year of original
adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A superscript
epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This test method covers a procedure for conducting cyclic potentiodynamic polarization measurements to determine relative
susceptibility to localized corrosion (pitting and crevice corrosion) for iron-, nickel-, or cobalt-based alloys in a chloride
environment. This test method also describes an experimental procedure which can be used to check one’s experimental technique
and instrumentation.
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
limitations prior to use.
2. Referenced Documents
2.1 ASTM Standards:
D1193 Specification for Reagent Water
G3 Practice for Conventions Applicable to Electrochemical Measurements in Corrosion Testing
G5 Reference Test Method for Making Potentiodynamic Anodic Polarization Measurements
2.2 ASTM Adjuncts:
Standard Samples (set of two)
3. Significance and Use
3.1 An indication of the susceptibility to initiation of localized corrosion in this test method is given by the potential at which
the anodic current increases rapidly. The more noble this potential, obtained at a fixed scan rate in this test, the less susceptible
is the alloy to initiation of localized corrosion. The results of this test are not intended to correlate in a quantitative manner with
the rate of propagation that one might observe in service when localized corrosion occurs.
3.2 In general, once initiated, localized corrosion can propagate at some potential more electropositive than that at which the
hysteresis loop is completed. In this test method, the potential at which the hysteresis loop is completed is determined at a fixed
scan rate. In these cases, the more electropositive the potential at which the hysteresis loop is completed the less likely it is that
localized corrosion will occur.
3.3 If followed, this test method will provide cyclic potentiodynamic anodic polarization measurements that will reproduce data
developed at other times in other laboratories using this test method for the two specified alloys discussed in 3.4. The procedure
is used for iron-, nickel-, or cobalt-based alloys in a chloride environment.
3.4 A standard potentiodynamic polarization plot is included. These reference data are based on the results from five different
laboratories that followed the standard procedure, using specific alloys of Type 304 stainless steel, UNS S30400 and Alloy C-276,
UNS N10276. Curves are included which have been constructed using statistical analysis to indicate the acceptable range of
polarization curves.
This test method is under the jurisdiction of ASTM Committee G01 on Corrosion of Metals and is the direct responsibility of Subcommittee G01.11 on Electrochemical
Measurements in Corrosion Testing.
Current edition approved May 1, 2009May 1, 2014. Published May 2009May 2014. Originally approved in 1986. Last previous edition approved in 20032009 as
ε1
G61–86(2003)G61 . DOI: 10.1520/G0061-86R09.– 86 (2009). DOI: 10.1520/G0061-86R14.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Available from ASTM International Headquarters. Order Adjunct No. ADJG0061. Original adjunct produced before 1995.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
G61 − 86 (2014)
FIG. 1 Schematic Diagram of Specimen Holder (see Footnotes 3 and 4)
3.5 The availability of a standard test method, standard material, and standard plots should make it easy for an investigator to
check his techniques to evaluate susceptibility to localized corrosion.
4. Apparatus
4.1 The polarization cell should be similar to the one described in Reference Test Method G5. Other polarization cells may be
equally suitable.
4.1.1 The cell should have a capacity of about 1 L and should have suitable necks or seals to permit the introduction of
electrodes, gas inlet and outlet tubes, and a thermometer. The Luggin probe-salt bridge separates the bulk solution from the
saturated calomel reference electrode. The probe tip should be adjustable so that it can be brought into close proximity with the
working electrode.
4.2 Specimen Holder:
4.2.1 Specimens should be mounted in a suitable holder designed for flat strip, exposing 1 cm to the test solution (Fig. 1). Such
specimen holders have been described in the literature. It is important that the circular TFE-fluorocarbon gasket be drilled and
machined flat in order to minimize crevices.
4.3 Potentiostat (Note 1) —A potentiostat that will maintain an electrode potential within 1 mV of a preset value over a wide
range of applied currents should be used. For the type and size of standard specimen supplied, the potentiostat should have a
potential range of −1.0 to +1.6 V and an anodic current output range of 1.0 to 10 μA. Most commercial potentiostats meet the
specific requirements for these types of measurements.
NOTE 1—These instrumental requirements are based upon values typical of the instruments in the five laboratories that have provided the data used
in determining the standard polarization plot.
4.4 Potential-Measuring Instruments (Note 1)—The potential-measuring circuit should have a high input impedance on the
11 14
order of 10 to 10 Ω to minimize current drawn from the system during measurements. Instruments should have sufficient
sensitivity and accuracy to detect a change in potential of 61 mV, usually included in commercial potentiostats. An output as a
voltage is preferred for recording purposes.
4.5 Current-Measuring Instruments (Note 1)—An instrument that is capable of measuring a current accurately to within 1 %
of the absolute value over a current range between 1.0 and 10 μA should be used. Many commercial units have a build-in
instrument with an output as a voltage, which is preferred for recording purposes. For the purpose of the present test a logarithmic
output is desirable.
France, W. D., Jr., Journal of the Electrochemical Society, Vol 114, 1967, p. 818; and Myers, J. R., Gruewlar, F. G., and Smulezenski, L. A., Corrosion, Vol 24, 1968,
p. 352.
G61 − 86 (2014)
TABLE 1 Chemical Composition of Alloys Used in the Round
Robin, Weight %
Type 304
Alloy C-276
Element Stainless Steel
(UNS N10276)
(UNS S30400)
Carbon 0.003 0.060
Chromium 15.29 18.46
Cobalt 2.05 .
Columbium . 0.11
Copper . 0.17
Iron 5.78 balance
Manganese 0.48 1.43
Molybdenum 16.03 0.17
Nickel balance 8.74
Phosphorus 0.018 0.029
Silicon 0.05 0.60
Sulfur 0.006 0.014
Vanadium 0.20 .
Tungsten 3.62 .
4.6 Anodic Polarization Circuit—A scanning potentiostat is used for potentiodynamic measurements. Potential and current are
plotted continuously using an X-Y recorder and a logarithmic converter (contained in the potentiostat or incorporated into the
circuit) for the current. Commercially available units are suitable.
4.7 Electrodes:
4.7.1 The standard Type 304 stainless steel (UNS S30400) and Alloy C-276 (UNS N10276) should be machined into flat
0.625-in. (14-mm) diameter disks. The chemical compositions of the alloys used in the round robin are listed in Table 1.
4.7.2 Counter Electrodes—The counter electrodes may be prepared as described in Reference Test Method G5 or may be
prepared from high-purity platinum flat stock and wire. A suitable method would be to seal the platinum wire in glass tubing and
introduce the platinum electrode assembly through a sliding seal. Counter electrodes should have an area at least twice as large
as the test
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.