Standard Practice for Characterization and Verification of Linear Phased Array Ultrasonic Probes

SIGNIFICANCE AND USE
5.1 This practice is intended to provide standardized procedures for evaluating linear phased-array ultrasonic probes. It is not intended to define performance and acceptance criteria, but rather to provide data from which such criteria may be established.  
5.2 Implementation may require more detailed procedural instructions in a format of the using facility.  
5.3 The measurement data obtained may be employed by users of this guide to specify, describe, or provide performance criteria for procurement and quality assurance, or service evaluation of the operating characteristics of linear phased-array ultrasonic probes. All or portions of the standard practice may be used as determined by the user.  
5.4 The measurements are made primarily under pulse-echo conditions. To determine the relative performance of a probe element as either a transmitter or a receiver may require additional tests.  
5.5 While these procedures relate to many of the significant parameters, others that may be important in specific applications may not be treated. These might include power handling capability, breakdown voltage, wear properties of contact units, radio-frequency interference, and the like.  
5.6 Care must be taken to ensure that comparable measurements are made and that users of the standard practice follow similar procedures. The conditions specified or selected (if optional) may affect the test results and lead to apparent differences.  
5.7 Interpretation of some test results, such as the shape of the frequency response curve, may be subjective. Small irregularities may be significant. Interpretation of the test results is beyond the scope of this practice.  
5.8 Certain results obtained using the procedures outlined may differ from measurements made with phased-array ultrasonic test instruments. These differences may be attributed to differences in the nature of the experiment or the electrical characteristics of the instrumentation.  
5.9 The pulse generator used t...
SCOPE
1.1 This practice covers measurement procedures for evaluating certain characteristics of phased-array ultrasonic probes that are used with phased-array ultrasonic examination instrumentation.  
1.2 This practice describes means for obtaining performance data that may be used to define the acoustic and electric responses of phased-array ultrasonic probes including contact (with or without a wedge) and immersion linear phased-array probes used for ultrasonic nondestructive testing with central frequencies ranging from 0.5 MHz to 10 MHz. Frequencies outside of this range may use the same methods but the testing equipment may vary.  
1.3 When ultrasonic values dependent on material are specified in this practice, they are based on carbon steel with an ultrasonic wave propagation speed of 5920 m/s (±50 m/s) for longitudinal wave modes and 3255 m/s (±30 m/s) for transverse or shear wave modes.  
1.4 This practice describes some of the characterization and verification procedures that can be carried out at the end stage of the manufacturing process of linear phased array probes. This practice does not describe the methods or acceptance criteria used to verify the performance of the combined phased array ultrasonic instrument and probe system.  
1.5 While this practice is intended to provide standardized procedures for evaluating linear phased-array ultrasonic probes, it may, with suitable modifications, be used for evaluation of configurations other than linear; for example, 1.5D or 2D matrix array probes.  
1.6 Units—The values stated in SI units are to be regarded as the standard. The values given in parentheses after SI units are provided for information only and are not considered standard.  
1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practice...

General Information

Status
Published
Publication Date
30-Nov-2022
Technical Committee
Drafting Committee
Current Stage
Ref Project

Buy Standard

Standard
ASTM E2904-22 - Standard Practice for Characterization and Verification of Linear Phased Array Ultrasonic Probes
English language
6 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
REDLINE ASTM E2904-22 - Standard Practice for Characterization and Verification of Linear Phased Array Ultrasonic Probes
English language
6 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation:E2904 −22
Standard Practice for
Characterization and Verification of Linear Phased Array
1
Ultrasonic Probes
This standard is issued under the fixed designation E2904; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope* 1.7 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the
1.1 This practice covers measurement procedures for evalu-
responsibility of the user of this standard to establish appro-
ating certain characteristics of phased-array ultrasonic probes
priate safety, health, and environmental practices and deter-
that are used with phased-array ultrasonic examination instru-
mine the applicability of regulatory limitations prior to use.
mentation.
1.8 This international standard was developed in accor-
1.2 This practice describes means for obtaining perfor-
dance with internationally recognized principles on standard-
mance data that may be used to define the acoustic and electric
ization established in the Decision on Principles for the
responses of phased-array ultrasonic probes including contact
Development of International Standards, Guides and Recom-
(with or without a wedge) and immersion linear phased-array
mendations issued by the World Trade Organization Technical
probes used for ultrasonic nondestructive testing with central
Barriers to Trade (TBT) Committee.
frequencies ranging from 0.5 MHz to 10 MHz. Frequencies
outside of this range may use the same methods but the testing
2. Referenced Documents
equipment may vary. 2
2.1 ASTM Standards:
1.3 Whenultrasonicvaluesdependentonmaterialarespeci- E1065 Practice for Evaluating Characteristics of Ultrasonic
Search Units
fied in this practice, they are based on carbon steel with an
ultrasonic wave propagation speed of 5920 m/s (650 m/s) for E1316 Terminology for Nondestructive Examinations
longitudinal wave modes and 3255 m/s (630 m/s) for trans-
3. Terminology
verse or shear wave modes.
3.1 Definitions:
1.4 This practice describes some of the characterization and
3.1.1 For definitions of terms used in this practice, see
verification procedures that can be carried out at the end stage
Terminology E1316.
of the manufacturing process of linear phased array probes.
3.1.2 Technical Specification of the Phased Array Probes:
This practice does not describe the methods or acceptance
3.1.2.1 certification, n—defined as measured performance
criteria used to verify the performance of the combined phased
of an individual probe. This is the document that reports the
array ultrasonic instrument and probe system.
measured performance of a specific probe (specific to a serial
1.5 While this practice is intended to provide standardized
number).
procedures for evaluating linear phased-array ultrasonic
3.1.2.2 datasheet, n—defined as specification. This is the
probes, it may, with suitable modifications, be used for
document that defines the general geometry and expected
evaluation of configurations other than linear; for example,
performance of a specific model or part number.
1.5D or 2D matrix array probes.
3.1.2.3 probe shape and size, n—probe form and dimen-
1.6 Units—The values stated in SI units are to be regarded
sions.
as the standard. The values given in parentheses after SI units
3.1.2.4 probe type, n—contact or immersion.
are provided for information only and are not considered
standard. 3.1.3 Terminology for Array Probes:
3.1.3.1 active aperture, n—pitch times the number of ele-
ments.
1
This practice is under the jurisdiction of ASTM Committee E07 on Nonde-
structive Testing and is the direct responsibility of Subcommittee E07.06 on
2
Ultrasonic Method. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved Dec. 1, 2022. Published January 2023. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
approved in 2012. Last previous edition approved in 2017 as E2904 – 17. DOI: Standards volume information, refer to the standard’s Document Summary page on
10.1520/E2904-22. the ASTM website.
*A Summary of Changes section appears at the end of this standard
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
E2904−22
3.1.3.2 element length, n—in a rectangular element, the 4.2 Physical Aspects—Identification and physical dimen-
acoustic element
...

This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: E2904 − 17 E2904 − 22
Standard Practice for
Characterization and Verification of Linear Phased Array
1
Ultrasonic Probes
This standard is issued under the fixed designation E2904; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope*
1.1 This standard practice covers measurement procedures for evaluating certain characteristics of phased-array ultrasonic probes
that are used with phased-array ultrasonic examination instrumentation.
1.2 This standard practice describes means for obtaining performance data that may be used to define the acoustic and electric
responses of phased-array ultrasonic probes including contact (with or without a wedge) and immersion linear phased-array probes
used for ultrasonic nondestructive testing with central frequencies ranging from 0.5 MHz to 10 MHz. Frequencies outside of this
range may use the same methods but the testing equipment may vary.
1.3 When ultrasonic values dependent on material are specified in this document,practice, they are based on carbon steel with an
ultrasonic wave propagation speed of 5920 m/s (650 m/s) for longitudinal wave modes and 3255 m/s (630 m/s) for transverse
or shear wave modes.
1.4 This documentpractice describes some of the characterization and verification procedures that can be carried out at the end
stage of the manufacturing process of linear phased array probes. This documentpractice does not describe the methods or
acceptance criteria used to verify the performance of the combined phased array ultrasonic instrument and probe system.
1.5 While this practice is intended to provide standardized procedures for evaluating linear phased-array ultrasonic probes, it may,
with suitable modifications, be used for evaluation of configurations other than linear; for example, 1.5D or 2D matrix array probes.
1.6 Units—The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information
only.after SI units are provided for information only and are not considered standard.
1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of
regulatory limitations prior to use.
1.8 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
1
This practice/guidepractice is under the jurisdiction of ASTM Committee E07 on Nondestructive Testing and is the direct responsibility of Subcommittee E07.06 on
Ultrasonic Method.
Current edition approved Nov. 1, 2017Dec. 1, 2022. Published December 2017January 2023. Originally approved in 2012. Last previous edition approved in 20122017
as E2904E2904 – 17.–12. DOI: 10.1520/E2904-17.10.1520/E2904-22.
*A Summary of Changes section appears at the end of this standard
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
E2904 − 22
2. Referenced Documents
2
2.1 ASTM Standards:
E1065 Practice for Evaluating Characteristics of Ultrasonic Search Units
E1316 Terminology for Nondestructive Examinations
3. Terminology
3.1 Definitions:
3.1.1 For definitions of terms used in this practice, see Terminology E1316.
3.1.2 Technical Specification of the Phased Array Probes:
3.1.2.1 datasheet—defined as specification. This is the document that defines the general geometry and expected performance
of a specific model or part number.
3.1.2.1 certification—certification, n—defined as measured performance of an individual probe. This is the document that
reports the measured performance of a specific probe (specific to a serial number).
3.1.2.2 probedatasheet, type—n—contact or immersion.defined as specification. This is the document that defines the general
geometry and expected performance of a specific model or part number.
3.1.2.3 probe shape and size—size, n—probe form and dimensions.
3.1.2.4 probe type, n—contact or immersion.
3.1.3 Termin
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.