ASTM E948-16(2020)
(Test Method)Standard Test Method for Electrical Performance of Photovoltaic Cells Using Reference Cells Under Simulated Sunlight
Standard Test Method for Electrical Performance of Photovoltaic Cells Using Reference Cells Under Simulated Sunlight
SIGNIFICANCE AND USE
5.1 This test method provides a procedure for testing and reporting the electrical performance of photovoltaic cells.
5.2 The test results may be used for comparison of cells among a group of similar cells or to compare diverse designs, such as different manufacturers' products. Repeated measurements of the same cell may be used to study changes in device performance.
5.3 This test method determines the electrical performance of a photovoltaic cell at a single instant of time and the results do no imply any past or future performance.
5.4 This test method requires a linear reference cell calibrated with respect to an appropriate reference spectral irradiance distribution, such as Tables E490, or G173. It is the responsibility of the user to determine which reference spectral irradiance distribution is appropriate for a particular application.
SCOPE
1.1 This test method covers the determination of the electrical performance of a photovoltaic cell under simulated sunlight by means of a calibrated reference cell procedure.
1.2 Electrical performance measurements are reported with respect to a select set of standard reporting conditions (SRC) (see Table 1) or to user-specified reporting conditions. In either case, the chosen reporting conditions are abbreviated as RC.
1.2.1 The RC include the cell temperature, the total irradiance, and the reference spectral irradiance distribution.
1.3 This test method is applicable only to photovoltaic cells with a linear short-circuit current versus total irradiance response up to and including the total irradiance used in the measurement.
1.4 The cell parameters determined by this test method apply only at the time of test, and imply no past or future performance level.
1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
General Information
Relations
Standards Content (Sample)
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: E948 − 16 (Reapproved 2020) An American National Standard
Standard Test Method for
Electrical Performance of Photovoltaic Cells Using
Reference Cells Under Simulated Sunlight
This standard is issued under the fixed designation E948; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 2. Referenced Documents
2.1 ASTM Standards:
1.1 This test method covers the determination of the elec-
E490Standard Solar Constant and Zero Air Mass Solar
trical performance of a photovoltaic cell under simulated
Spectral Irradiance Tables
sunlight by means of a calibrated reference cell procedure.
E491Practice for Solar Simulation for Thermal Balance
1.2 Electrical performance measurements are reported with
Testing of Spacecraft
respect to a select set of standard reporting conditions (SRC)
E691Practice for Conducting an Interlaboratory Study to
(seeTable1)ortouser-specifiedreportingconditions.Ineither Determine the Precision of a Test Method
case, the chosen reporting conditions are abbreviated as RC. E772Terminology of Solar Energy Conversion
E927Classification for Solar Simulators for Electrical Per-
1.2.1 The RC include the cell temperature, the total
formance Testing of Photovoltaic Devices
irradiance, and the reference spectral irradiance distribution.
E973Test Method for Determination of the Spectral Mis-
1.3 This test method is applicable only to photovoltaic cells
match Parameter Between a Photovoltaic Device and a
with a linear short-circuit current versus total irradiance
Photovoltaic Reference Cell
response up to and including the total irradiance used in the
E1125 Test Method for Calibration of Primary Non-
measurement. ConcentratorTerrestrial Photovoltaic Reference Cells Us-
ing a Tabular Spectrum
1.4 The cell parameters determined by this test method
E1362Test Methods for Calibration of Non-Concentrator
apply only at the time of test, and imply no past or future
Photovoltaic Non-Primary Reference Cells
performance level.
G173TablesforReferenceSolarSpectralIrradiances:Direct
Normal and Hemispherical on 37° Tilted Surface
1.5 The values stated in SI units are to be regarded as
standard. No other units of measurement are included in this
3. Terminology
standard.
3.1 Definitions—Definitions of terms used in this test
1.6 This standard does not purport to address all of the
method may be found in Terminology E772 and in Specifica-
safety concerns, if any, associated with its use. It is the
tion E927.
responsibility of the user of this standard to establish appro-
3.2 Definitions of Terms Specific to This Standard:
priate safety, health, and environmental practices and deter- 3.2.1 effective irradiance, n—the irradiance that a solar
mine the applicability of regulatory limitations prior to use. simulatorproducesasmeasuredbyacell’sshort-circuitcurrent
relative to a reference value for the cell’s short-circuit current
1.7 This international standard was developed in accor-
at a particular RC.
dance with internationally recognized principles on standard-
3.2.1.1 Discussion—This reference value typically corre-
ization established in the Decision on Principles for the
sponds to a different spectral irradiance distribution than the
Development of International Standards, Guides and Recom-
solar simulator.
mendations issued by the World Trade Organization Technical
3.2.2 reporting conditions, RC, n—the reference spectral
Barriers to Trade (TBT) Committee.
irradiance distribution, total irradiance, and cell temperature to
which the photovoltaic current-voltage performance is mea-
sured and corrected.
This test method is under the jurisdiction of ASTM Committee E44 on Solar,
GeothermalandOtherAlternativeEnergySourcesandisthedirectresponsibilityof
Subcommittee E44.09 on Photovoltaic Electric Power Conversion. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
CurrenteditionapprovedJune1,2020.PublishedJuly2020.Originallyapproved contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
in 1993. Last previous edition approved in 2016 as E948–16. DOI: 10.1520/ Standards volume information, refer to the standard’s Document Summary page on
E0948-16R20. the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
E948 − 16 (2020)
TABLE 1 Standard Reporting Conditions
4.4 The data from the measurements are corrected to the
Reference Spectral Irradiance Total Irradiance, E Cell Temperature, T desired RC. Three possible SRC are defined in Table 1.
0 0
−2
Distribution (Wm ) (°C)
4.4.1 Measurement error in test cell current caused by
Tables G173 Direct Normal 900 25
deviationsoftheirradianceconditionsfromtheRCiscorrected
Tables G173 Hemispherical 1000 25
using the effective irradiance measured with the reference cell
Tables E490 1366.1 25
and the spectral mismatch parameter, M, which is determined
in accordance with Test Method E973.
3.2.3 test cell, n—the photovoltaic cell to be tested, or cell 4.4.1.1 This test method does not apply corrections to cell
under test, using the method described herein. voltage for irradiance deviations, thus the solar simulator
irradiance must be sufficiently well controlled to accurately
3.3 Symbols—The following symbols and units are used in
determine other parameters under RC, especially maximum
this test method:
power and open-circuit voltage. To this end, the effective
3.3.1 0—as a subscript, denotes a value under the specified
irradiance during the measurement is restricted to be within
RC.
62% of the RC irradiance. However, there will still be
3.3.2 A—area of the test cell, (m ).
measurement uncertainty due to irradiance variations in this
3.3.3 A —area of the reference cell, (m ).
range.
R
2 −1
4.4.2 Measurementerrorcausedbydeviationofthetest-cell
3.3.4 C —calibration constant of reference cell, (Am W ).
R
and reference-cell temperatures from the RC is minimized by
3.3.5 C —transfer calibration ratio, (dimensionless).
T
maintaining the cell temperatures sufficiently close to the
−2
3.3.6 E—total irradiance, Wm .
requiredRCvalue.Tothisend,thetestcelltemperatureduring
3.3.7 FF—fill factor, (%).
the measurement is restricted to be within 61°CoftheRC
temperature.
3.3.8 I—current of the test cell (A).
4.4.2.1 TestMethodE973providesforcorrectionoftestcell
3.3.9 I —current of the test cell at maximum power in the
MP
current through a temperature-dependent spectral mismatch
power-producing quadrant (A).
parameter, M(T); however, Test Method E973 allows the
3.3.10 I —short-circuit current of the test cell (A).
SC
temperature correction to be bypassed if the temperature is
3.3.11 I —short-circuit current of the reference cell (A).
within 61 °C.
SC,R
4.4.2.2 This test method does not apply corrections to cell
3.3.12 I —short-circuit current of the monitor cell (A).
SC,M
voltage for temperature deviations, thus the test-cell tempera-
3.3.13 M—spectral mismatch parameter (dimensionless).
ture must be sufficiently well controlled to accurately deter-
3.3.14 P —maximum power of the test cell in the power-
MP
mine other parameters under RC, especially maximum power
producing quadrant (W).
and open-circuit voltage. However, there will still be measure-
3.3.15 R —series resistance of the test cell (Ω).
ment uncertainty due to temperature variations in this range.
S
4.4.3 The measurement procedure employs a reference
3.3.16 S—current correction factor due to spatial non-
cell-testcellsubstitutiontechniquethatisdesignedtominimize
uniformity of irradiance (dimensionless).
errors in short-circuit current caused by spatial non-uniformity
3.3.17 T—temperature of the test cell (°C).
of the solar simulator irradiance. A correction for spatial
3.3.18 T —temperature of the reference cell (°C).
R
non-uniformity of irradiance may be applied to measured
3.3.19 U —ordered set of test cell current, voltage, and
current data if the reference cell and test cell have different
power values at RC (A, V, W).
areas; the correction is defined as the ratio of the effective
irradianceinthesolarsimulatorovertheareaofthetestcellto
3.3.20 V—voltage of the test cell (V).
the effective irradiance over the area of the reference cell.
3.3.21 V —voltage of the test cell at maximum power in
MP
the power-producing quadrant (V).
5. Significance and Use
3.3.22 V —open-circuit voltage of the test cell (V).
OC
5.1 This test method provides a procedure for testing and
3.3.23 η—efficiency (%).
reporting the electrical performance of photovoltaic cells.
4. Summary of Test Method
5.2 The test results may be used for comparison of cells
among a group of similar cells or to compare diverse designs,
4.1 The performance test of a photovoltaic cell consists of
such as different manufacturers’ products. Repeated measure-
measuring the electrical current versus voltage (I-V) charac-
ments of the same cell may be used to study changes in device
teristic of the cell while illuminated by a solar simulator and
performance.
with its temperature sufficiently controlled.
5.3 This test method determines the electrical performance
4.2 Acalibratedphotovoltaicreferencecell(see6.1)isused
to determine the effective irradiance during the test. of a photovoltaic cell at a single instant of time and the results
do no imply any past or future performance.
4.3 Simulatedsunlightisusedfortheelectricalperformance
measurement, and solar simulation requirements are defined in 5.4 This test method requires a linear reference cell cali-
Specification E927 (terrestrial applications) and Practice E491 brated with respect to an appropriate reference spectral irradi-
(space applications). ance distribution, such as Tables E490,or G173.Itisthe
E948 − 16 (2020)
responsibilityoftheusertodeterminewhichreferencespectral
irradiance distribution is appropriate for a particular applica-
tion.
6. Apparatus
6.1 Reference Cell—A linear, calibrated, photovoltaic solar
cell used to determine the total irradiance during the electrical
performance measurement.
6.1.1 Reference cells may be calibrated in accordance with
TestMethodsE1125orE1362,asisappropriateforaparticular
application.
NOTE 1—No reference cell calibration standards presently exist for
space applications, although procedures using high-altitude balloon and
low-earth orbit flights are being used to calibrate such reference cells.
6.1.2 Thecalibrationconstant, C ,ofthereferencecellshall
R
be with respect to the reference spectral irradiance distribution
FIG. 1 I-V Measurement Schematic
of the desired RC (see 1.2).
6.1.3 A current measurement instrument (see 6.3) shall be
used to determine the short-circuit current of the reference cell
6.4 Voltage Measurement Equipment—Electricalinstrumen-
under the solar simulator.
tationusedtomeasurethevoltageacrossthetestcellduringthe
6.1.4 Special Case—If the test cell also qualifies as a
performance measurement. The instrumentation shall have a
reference cell in that its I or calibration constant at the RC is
SC
resolution of at least 0.02 % of the maximum voltage
known prior to test, the test cell may be used to measure
encountered, and shall have a total error of less than 0.1% of
irradiancebyitselfandtheseparatereferencecellomitted.The
the maximum voltage encountered.
self-irradiance measurement technique is typically used to
6.4.1 The voltage measurement equipment shall measure
determinethefillfactorofareferencecellpost-calibration,and
data points simultaneously with the current (see 6.3) and
as a check for damage or degradation.
short-circuit current (see 6.9) measurement equipment, to
6.2 Test Fixture—Boththetestcellandthereferencecellare
within 10 µs.
mounted in a fixture that meets the following requirements:
6.5 Variable Load—An electronic load, such as a variable
6.2.1 Thetestfixtureshallensureauniformlateraltempera-
resistor or a programmable power supply, used to operate the
ture distribution to within 60.5°C during the performance
test cell at different points along its I-V characteristic.
measurement.
6.5.1 Thevariableloadshallbecapableofoperatingthetest
6.2.2 The test fixture shall include a provision for maintain-
cell at an I-V point where the voltage is within 1% of V in
OC
ing a constant cell temperature for both the reference cell and
the power-producing quadrant.
the test cell (see 7.11).
6.5.2 Thevariableloadshallbecapableofoperatingthetest
NOTE2—Whenusingpulsedorshutteredsolarsimulators,itispossible
cellatanI-Vpointwherethecurrentiswithin1%of I inthe
sc
thatthecelltemperaturewillincreaseuponinitialillumination,evenwhen
power-producing quadrant.
the cell temperature is controlled.
6.5.3 The variable load shall allow an output power (the
6.2.3 The test fixture, when placed in the solar simulator,
product of cell current and cell voltage) resolution of at least
shall ensure that the fields-of-view of both the reference cell
0.2% of P .
MP
and the test cell are identical.
6.5.4 The electrical response time of the variable load shall
NOTE 3—Some solar simulators may have significant amounts of be fast enough to sweep the range of I-V operating points
irradiation from oblique or non-perpendicular angles to the test plane. In
during the measurement period.
these cases, it is important that the test cell and the reference cell have
similar reflectance and angular-response characteristics. NOTE 4—It is possible that the response time of the test cell may limit
how fast the range of I-V operating points can be swept, especially when
6.2.4 A four-terminal connection (also known as a Kelvin
pulsed solar simulators are used. For these cases, it may be necessary to
connection, see Fig. 1) from the test cell to the I-V measure-
measure smaller ranges of the I-V curve using multiple measurements to
ment instrumentation (see 6.3 – 6.5) shall be used.
obtain the entire range required.
6.3 Current Measurement Equipment—Electrical instru-
6.6 Solar Simulator—Requirements of the solar simulator
mentation used to measure the current through the test cell
usedtoilluminatethetestcellaredefinedinSpecificationE927
during the performance measurement. The instrumentation
(terrestrial applications) and Practice E491 (space applica-
shall have a resolution of at least 0.02% of the maximum
tions).
current encountered, and shall have a total error of less than
6.6.1 The effective irradiance during the performance mea-
0.1% of t
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.