Standard Practice for Calibration of Laboratory Volumetric Apparatus

SCOPE
1.1 This practice covers procedures for use in the calibration of volumetric ware, in accordance with E694 such as is in common use in chemical and clinical laboratories. It is based on the gravimetric determination of the quantity of water either contained or delivered, and the conversion of this value to true volume at the standard temperature of 20oC by means of suitable equations and standard tables. Calibration using mercury is excluded. Calibration may be performed using alternative gravimetric methodology, provided that it is demonstrated and documented that the results obtained are equivalent to those obtained using the methodology described herein.
1.2 This practice is intended to encompass capacity ware between the limits of 0.1 cm3 and 2000 cm3. Typical products falling within the purview of this practice are burets graduated" to deliver", graduated cylinders, volumetric flasks, specific gravity flasks, measuring and dilution pipets, and transfer and capacity pipets.
1.3 The procedures are not recommended for calibration of apparatus with capacities below 0.1 cm3, such as microglassware.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

General Information

Status
Historical
Publication Date
09-Oct-2001
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM E542-00 - Standard Practice for Calibration of Laboratory Volumetric Apparatus
English language
7 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


NOTICE: This standard has either been superseded and replaced by a new version or discontinued.
Contact ASTM International (www.astm.org) for the latest information.
Designation: E 542 – 00
Standard Practice for
Calibration of Laboratory Volumetric Apparatus
This standard is issued under the fixed designation E 542; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
This standard has been approved for use by agencies of the Department of Defense.
1. Scope manipulating the apparatus, weighing, and converting the
weight to the appropriate standard volume.
1.1 This practice covers procedures for use in the calibration
of volumetric ware, in accordance with E694 such as is in
4. Significance and Use
common use in chemical and clinical laboratories. It is based
4.1 The primary purpose of this practice is to provide
on the gravimetric determination of the quantity of water either
uniform procedures that may be used to accurately calibrate a
contained or delivered, and the conversion of this value to true
wide variety of volumetric ware. The techniques are simple in
volume at the standard temperature of 20°C by means of
concept and can provide reliable results, provided the proce-
suitable equations and standard tables. Calibration using mer-
dures are followed faithfully. Accordingly, the practice should
cury is excluded. Calibration may be performed using alterna-
provide a means for checking the original calibration of
tive gravimetric methodology, provided that it is demonstrated
glassware and similar apparatus and for periodic rechecks as
and documented that the results obtained are equivalent to
the need should arise.
those obtained using the methodology described herein.
4.2 Borosilicate volumetric glassware will hold its calibra-
1.2 This practice is intended to encompass capacity ware
3 3
tion indefinitely provided that it is not exposed to hydrofluoric
between the limits of 0.1 cm and 2000 cm . Typical products
acid, hot phosphoric acid, or strong, hot alkalis, and that it is
falling within the purview of this practice are burets graduated“
not heated above 150°C when dry. A frosting of the glass
to deliver”, graduated cylinders, volumetric flasks, specific
surface (viewed when dry) indicates that chemical attack has
gravity flasks, measuring and dilution pipets, and transfer and
occured, and recalibration may be in order. As a precaution,
capacity pipets.
however, it is recommended that the glassware be recalibrated
1.3 The procedures are not recommended for calibration of
3 after ten years of service regardless of its appearance.
apparatus with capacities below 0.1 cm , such as microglass-
4.3 Soda-lime volumetric glassware will become frosted
ware.
with time because of attack from moisture in the atmosphere as
1.4 This standard does not purport to address all of the
well as from the chemicals mentioned above. In addition, it
safety concerns, if any, associated with its use. It is the
should not be heated above 90°C when dry. It is recommended,
responsibility of the user of this standard to establish appro-
therefore, that it be recalibrated after five years of service
priate safety and health practices and determine the applica-
unless frosting (viewed when dry) is observed sooner.
bility of regulatory limitations prior to use.
5. Units of Measurement
2. Referenced Documents
5.1 Capacity—The basic SI unit for volume is the cubic
2.1 ASTM Standards:
metre, m . Due to its large size, it is rarely used in volumetric
E 1 Specification for ASTM Thermometers
calibration. Rather, the cubic centimetre, cm , is used and will
SI 10 Standard for Use of the International System of Units
be employed in this practice. The unit, millilitre, mL, may be
(SI) (the Modernized Metric System)
3 considered as equivalent to the cubic centimetre.
E 694 Specification for Volumetric Ware
5.2 Standard Temperature—Volumetric ware is almost uni-
3. Summary of Practice versally calibrated at 20°C. The procedures described provide
for such a calibration. When it is necessary to work at higher
3.1 This practice is based upon a determination of the
ambient temperatures, such as is the case in tropical countries,
volume of water either contained in or delivered by the vessel.
calibration may be required at the International Standards
Procedures are given for cleaning, setting the meniscus,
Organization (ISO) recommended temperature of 27°C. Equa-
tions are given to correct the calibrated volume to other
This practice is under the jurisdiction of ASTM Committee E41 on Laboratory
temperatures as this may be required or desired.
Apparatus, and is the direct responsibility of Subcommittee E41.01 on Apparatus.
Current edition approved Nov. 10, 2000. Published November 2000. Originally
6. Apparatus
published as E 542 – 79. Last previous edition E 542 – 94(1999).
Annual Book of ASTM Standards, Vol 14.03.
6.1 Balance, having sufficient capacity to weigh the loaded
Annual Book of ASTM Standards, Vol 14.04.
Copyright © ASTM, 100 Barr Harbor Drive, West Conshohocken, PA 19428-2959, United States.
E 542
vessel. The sensitivity of the balance will be a limiting factor 7.3.1 Reading the Meniscus—For all apparatus calibrated
in the accuracy of the measurements. Either a single-pan, by this procedure, the reading is made on the lowest point of
direct-reading balance or an equal-arm balance of adequate the meniscus. In order that the lowest point may be observed,
sensitivity and capacity may be used. In either case the weights it is necessary to place a shade of some dark material
must be calibrated with adequate accuracy. Ordinarily, weights immediately below and behind the meniscus, which renders the
with NBS Class S-1 tolerances are required. The balance must profile of the meniscus dark and clearly visible against a light
have physical dimensions to accept the size of the vessels background. A convenient device for this purpose is a collar-
which need to be weighed. shaped section of thick black rubber tubing, cut open at one
6.2 Thermometer, for measuring the temperature of the side and of such size as to clasp the tube firmly. Alternatively,
water. The accuracy of this calibration will depend upon the black paper may be used. “Short line” graduated vessels where
accuracy requirement of the volumetric calibration (see Section the lines are less than one half of the circumference may be
14 for tolerances). read more accurately by placing a mirror against the rear of the
6.3 Barometer, capable of providing atmospheric pressure vessel to reflect the front line.
measurements, consistent with the tolerances given in Section 7.3.2 Setting the Meniscus—Setting of the meniscus shall
14. Alternatively, the existing barometric pressure may be be performed by one of the two methods detailed below.
obtained from the local weather service. Wherever practical the meniscus should descend to the position
6.4 Distilled or Deionized Water, suitable for general labo- of setting.
ratory purposes. 7.3.2.1 The position of the lowest point of the meniscus
with reference to the graduation line is horizontally tangent to
7. General Considerations
the plane of the upper edge of the graduation line. The position
7.1 This section describes manipulative techniques required
of the meniscus is obtained by having the eye in the same plane
to obtain accurate and reproducible volumetric measurements.
of the upper edge of the graduation line.
7.2 Cleaning Procedures:
7.3.2.2 The position of the lowest point of the meniscus
7.2.1 It is important that volumetric ware be thoroughly
with reference to the graduation line is such that it is in the
cleaned before being tested or used. Glass apparatus must be
plane of the middle of the graduation line. This position of the
sufficiently clean to permit uniform wetting of the surface.
meniscus is obtained by making the setting in the center of the
When clean, the walls will be uniformly wetted and the water
ellipse formed by the graduation line on the front and the back
will adhere to the glass surface in a continuous film. The clean
of the tube as observed by having the eye slightly below the
walls of some plastic apparatus, however, will not be wetted.
plane of the graduation line. The setting is accurate if, as the
(Follow the instructions of the manufacturer. Do not use
eye is raised and the ellipse narrows, the lowest point of the
materials which will attack, discolor, or swell the plasticware.)
meniscus remains midway between the front and rear portions
Lack of cleanliness causes irregularities in capacity by distort-
of the graduation line. By this method it is possible to observe
ing the water surface. The liquids usually used for cleaning
the approach of the meniscus from either above or below the
glassware are sodium dichromatic-sulfuric acid cleaning solu-
line to its proper setting.
tion (commercially available from laboratory supply houses),
NOTE 1—The difference between meniscus positions resulting from the
nitric acid, fuming sulfuric acid, alcohol, and water. The choice
alternative methods of adjustment is the volume equivalent of one half the
of cleaning agent to be used depends on the nature of the
thickness of the graduation line. In the case of articles where the capacity
contaminant. After cleaning with the cleaning solution and
is read as the difference between two meniscus readings (for example on
thoroughly rinsing with tap water, the vessel should be rinsed
a buret), then no error results if the article is manufactured using one
method of adjustment and is later used by the other method.
with distilled water.
Even in the most unfavorable cases of single-mark articles (for example
7.2.2 After cleaning, the vessel should be rinsed with ethyl
large flasks), when working to the highest attainable accuracy, the
alcohol and dried with clean air at room temperature. It is not
difference between the two methods of adjustment is unlikely to exceed
necessary to dry any vessel marked “to deliver.” When
30 % of the Class A (precision apparatus) limit of error and a correction
cleaning small articles such as pipets, it is usually easier to fill
can be calculated where necessary.
them with cleaning solution by suction, using a vacuum line, if
8. Calibration Procedure for Burets
available, or a small rubber bulb, but never by mouth. The
solution should be drawn through the pipet several times until 8.1 Clamp the buret vertically on a support stand; also
the entire inside surface is evenly coated. Rinse thoroughly clamp a plain glass test tube, large enough to hold a thermom-
with tap water and then with distilled water. For cleaning eter, near the buret if the buret is of such a size that it is not
flasks, pour in enough cleaning solution while rotating so that large enough to insert a thermometer in the top for observing
a film of solution will cover the entire interior surface. A break the temperature of the water. Fill the buret from a reservoir or
in the film indicates a contaminated area. For filling a buret storage bottle, in which the water has reached equilibrium with
with cleaning solution, it should be held in a vertical position room temperature, and check to verify that there is neither
and filled by pouring into the top. Open the stopcock to drain. leakage from the tip nor from the stopcock plug. Drain and
Regardless of the type of vessel, always rinse thoroughly, first record the delivery time. Delivery time is determined by the
with tap water and then with distilled water. Plastic volumetric unrestricted outflow of the water from the zero mark to the
ware should be cleaned in an appropriate manner before lowest graduation mark with the stopcock fully open. Refill the
calibration. buret to approximately 10 mm above the zero mark and fill the
7.3 Reading and Setting a Liquid Meniscus: test tube that holds the thermometer; record the temperature.
E 542
Set the meniscus on the zero mark using the buret stopcock to one which has been filled from the water supply. The tempera-
lower the liquid level and touch the tip with the wetted wall of ture may be taken after final weighing by placing a thermom-
a beaker to remove any excess water. A weighing flask that has eter directly in the flask, provided the flask is of sufficient size
been tightly stoppered and weighed empty is placed with the to accommodate it.
inside of the neck in contact with the tip of the buret (the flask
11. Calibration of Flasks (to Deliver)
will be at a slight angle).
8.2 Fully open the stopcock until the water is only a few
11.1 Do not dry flasks that are calibrated to deliver prior to
millimetres above the line being tested and then the stream is
the test. Fill the flask to approximately the index line and
slowed so as to make an accurate setting. When the setting has
empty rapidly by gradually inclining the flask so as to avoid
been completed, move the flask horizontally, breaking the
splashing on the walls as much as possible. When the main
contact with the buret. Recheck the setting.
drainage stream has ceased, the flask will be nearly vertical.
8.3 Then stopper and weigh the flask a second time, after
Hold in this position for 30 s and touch off the drop of water
which refill the thermometer tube and test the next interval in
adhering to the top of the flask. Place a watch glass or plastic
the same manner as the first one—from the zero mark to the
cap on the flask to reduce evaporation and weigh immediately.
next interval needed.
Take a water temperature reading, fill the flask, and make the
8.4 For burets with a specified waiting time, use the
meniscus setting on the index line, taking care not to splash
following procedure: after adjustment to the zero mark, fully
water on the walls. Place the same cap or watch glass on the
open the stopcock until the meniscus has reached a position a
filled flask and weigh. The reverse of this procedure may be
few millimetres above the graduation line for calibration. After
used, if desirable.
the specified waiting time (for example, 30 s), adjust the
meniscus to the graduation line, remove the flask, and weigh.
12. Calibration of Other Volumetric Glassware
12.1 Measuring Pipets—Measuring pipets may be cali-
9. Calibration Procedure for Pipets (One Mark)
brated by a similar procedure as described in Section 8, except
9.1 Fill the pipet with distilled water by suction to the index
that the tip must be in contact with the wet wall of a beaker or
mark and measure the delivery time with the tip in contact with
other vessel when the setting is made on the zero line of a
the glass surface of the internal side of a beaker. Refill by
measuring pipet.
suction slightly above the index line
...