Standard Test Method for Elapsed Time Calibration of Thermal Analyzers

SIGNIFICANCE AND USE
5.1 Most thermal analysis experiments are carried out under increasing temperature conditions where temperature is the independent parameter. Some experiments, however, are carried out under isothermal temperature conditions where the elapsed time to an event is measured as the independent parameter. Isothermal Kinetics (Test Methods E2070), Thermal Stability (Test Method E487), Oxidative Induction Time (OIT) (Test Methods D3895, D4565, D5483, E1858, and Specification D3350 and Loss-on-Drying (Test Method E1868) are common examples of these kinds of experiments.  
5.2 Modern scientific instruments, including thermal analyzers, usually measure elapsed time with excellent precision and accuracy. In such cases, it may only be necessary to confirm the performance of the instrument by comparison to a suitable reference. Only rarely will it may be required to correct the calibration of an instrument's elapsed time signal through the use of a calibration factor.  
5.3 It is necessary to obtain elapsed time signal conformity only to 0.1 times the repeatability relative standard deviation (standard deviation divided by the mean value) expressed as a percent for the test method in which the thermal analyzer is to be used. For those test methods listed in Section 2 this conformity is 0.1 %.
SCOPE
1.1 This test method describes the calibration or performance confirmation of the elapsed-time signal from thermal analyzers.  
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.3 There is no ISO standard equivalent to this test method.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

General Information

Status
Historical
Publication Date
14-Sep-2013
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM E1860-13 - Standard Test Method for Elapsed Time Calibration of Thermal Analyzers
English language
3 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
REDLINE ASTM E1860-13 - Standard Test Method for Elapsed Time Calibration of Thermal Analyzers
English language
3 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: E1860 − 13
Standard Test Method for
1
Elapsed Time Calibration of Thermal Analyzers
This standard is issued under the fixed designation E1860; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope E1858 Test Method for Determining Oxidation Induction
Time of Hydrocarbons by Differential Scanning Calorim-
1.1 This test method describes the calibration or perfor-
etry
mance confirmation of the elapsed-time signal from thermal
E1868 Test Method for Loss-On-Drying by Thermogravim-
analyzers.
etry
1.2 The values stated in SI units are to be regarded as
E2161 Terminology Relating to Performance Validation in
standard. No other units of measurement are included in this
Thermal Analysis
standard.
3. Terminology
1.3 There is no ISO standard equivalent to this test method.
3.1 Definitions:
1.4 This standard does not purport to address all of the
3.1.1 The technical terms used in this test method are
safety concerns, if any, associated with its use. It is the
defined in Terminologies E473, E1142, and E2161, including
responsibility of the user of this standard to establish appro-
calibration, conformance, relative standard deviation, and ther-
priate safety and health practices and determine the applica-
mal analysis.
bility of regulatory limitations prior to use.
4. Summary of Test Method
2. Referenced Documents
4.1 The elapsed time signal generated by a thermal analyzer
2
2.1 ASTM Standards:
is compared to a clock (or timer) whose performance is known
D3350 Specification for Polyethylene Plastics Pipe and Fit-
and traceable to a national metrology institute. The thermal
tings Materials
analyzer may be said to be in conformance, if the performance
D3895 Test Method for Oxidative-Induction Time of Poly-
of the thermal analyzer is within established limits.
olefins by Differential Scanning Calorimetry
Alternatively, the elapsed time signal may be calibrated using
D4565 Test Methods for Physical and Environmental Per-
a two point calibration method.
formance Properties of Insulations and Jackets for Tele-
5. Significance and Use
communications Wire and Cable
D5483 Test Method for Oxidation Induction Time of Lubri-
5.1 Most thermal analysis experiments are carried out under
catingGreasesbyPressureDifferentialScanningCalorim-
increasing temperature conditions where temperature is the
etry
independent parameter. Some experiments, however, are car-
E473 Terminology Relating to Thermal Analysis and Rhe-
ried out under isothermal temperature conditions where the
ology
elapsed time to an event is measured as the independent
E487 Test Method for Constant-Temperature Stability of
parameter.IsothermalKinetics(TestMethodsE2070),Thermal
Chemical Materials
Stability (Test Method E487), Oxidative Induction Time (OIT)
E691 Practice for Conducting an Interlaboratory Study to
(Test Methods D3895, D4565, D5483, E1858, and Specifica-
Determine the Precision of a Test Method
tion D3350 and Loss-on-Drying (Test Method E1868) are
E1142 Terminology Relating to Thermophysical Properties
common examples of these kinds of experiments.
5.2 Modern scientific instruments, including thermal
analyzers, usually measure elapsed time with excellent preci-
1
ThistestmethodisunderthejurisdictionofASTMCommitteeE37onThermal
sion and accuracy. In such cases, it may only be necessary to
Measurements and is the direct responsibility of Subcommittee E37.10 on
Fundamental, Statistical and Mechanical Properties.
confirm the performance of the instrument by comparison to a
Current edition approved Sept. 15, 2013. Published September 2013. Originally
suitable reference. Only rarely will it may be required to
approved in 1997. Last previous edition approved in 2007 as E1860 – 07. DOI:
correct the calibration of an instrument’s elapsed time signal
10.1520/E1860-13.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or through the use of a calibration factor.
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
5.3 It is necessary to obtain elapsed time signal conformity
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website. only to 0.1 times the repeatability relative standard deviation
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
E1860 − 13
(standard deviation divided by the mean value) expressed as a t 5 t S (1)
o
percent for the test method in which the thermal analyzer is to
where:
be used. For those test methods listed in Section 2 thi
...

This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: E1860 − 07 E1860 − 13
Standard Test Method for
1
Elapsed Time Calibration of Thermal Analyzers
This standard is issued under the fixed designation E1860; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This test method describes the calibration or performance confirmation of the elapsed-time signal from thermal analyzers.
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3 There is no ISO standard equivalent to this test method.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
limitations prior to use.
2. Referenced Documents
2
2.1 ASTM Standards:
D3350 Specification for Polyethylene Plastics Pipe and Fittings Materials
D3895 Test Method for Oxidative-Induction Time of Polyolefins by Differential Scanning Calorimetry
D4565 Test Methods for Physical and Environmental Performance Properties of Insulations and Jackets for Telecommunications
Wire and Cable
D5483 Test Method for Oxidation Induction Time of Lubricating Greases by Pressure Differential Scanning Calorimetry
E473 Terminology Relating to Thermal Analysis and Rheology
E487 Test Method for Constant-Temperature Stability of Chemical Materials
E691 Practice for Conducting an Interlaboratory Study to Determine the Precision of a Test Method
E1142 Terminology Relating to Thermophysical Properties
E1858 Test Method for Determining Oxidation Induction Time of Hydrocarbons by Differential Scanning Calorimetry
E1868 Test Method for Loss-On-Drying by Thermogravimetry
E2161 Terminology Relating to Performance Validation in Thermal Analysis
3. Terminology
3.1 Definitions:
3.1.1 The technical terms used in this test method are defined in Terminologies E473, E1142, and E2161, including calibration,
conformance, relative standard deviation, and thermal analysis.
4. Summary of Test Method
4.1 The elapsed time signal generated by a thermal analyzer is compared to a clock (or timer) whose performance is known and
traceable to a national metrology institute. The thermal analyzer may be said to be in conformance, if the performance of the
thermal analyzer is within established limits. Alternatively, the elapsed time signal may be calibrated using a two point calibration
method.
5. Significance and Use
5.1 Most thermal analysis experiments are carried out under increasing temperature ramp conditions where temperature is the
independent parameter. Some experiments, however, are carried out under isothermal temperature conditions where the elapsed
1
This test method is under the jurisdiction of ASTM Committee E37 on Thermal Measurements and is the direct responsibility of Subcommittee E37.10 on Fundamental,
Statistical and Mechanical Properties.
Current edition approved March 1, 2007Sept. 15, 2013. Published April 2007September 2013. Originally approved in 1997. Last previous edition approved in 20022007
as E1860 – 02.E1860 – 07. DOI: 10.1520/E1860-07.10.1520/E1860-13.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
E1860 − 13
time to an event is measured as the independent parameter. Isothermal Kinetics (Test Methods E2070), Thermal Stability (Test
Method E487), Oxidative Induction Time (OIT) (Test Methods D3895, D4565, D5483, E1858, and Specification D3350 and
Loss-on-Drying (Test Method E1868) are common examples of these kinds of experiments.
5.2 Modern scientific instruments, including thermal analyzers, usually measure elapsed time with excellent precision and
accuracy. In such cases, it may only be necessary to confirm the performance of the instrument by comparison to a suitable
reference. Only rarely will it may be required to correct the calibration of an instrument’s elapsed time signal through the use of
a calib
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.