ASTM E2428-15
(Practice)Standard Practice for Calibration of Torque-Measuring Instruments for Verifying the Torque Indication of Torque Testing Machines
Standard Practice for Calibration of Torque-Measuring Instruments for Verifying the Torque Indication of Torque Testing Machines
SIGNIFICANCE AND USE
4.1 Testing machines that apply and indicate torque are in general use in many industries. Practice E2624 has been written to provide a practice for the torque verification of these machines. A necessary element in Practice E2624 is the use of devices whose torque characteristics are known to be traceable to the International System of Units (SI). Practice E2428 describes how these devices are to be calibrated. The procedures are useful to users of testing machines, manufacturers and providers of torque measuring instruments, calibration laboratories that provide calibration services and documents of metrological traceability, and service organizations using devices to verify testing machines.
SCOPE
1.1 This practice is to specify procedure for the calibration of elastic torque-measuring instruments.
Note 1: Verification by deadweight and a lever arm is an acceptable method of verifying the torque indication of a torque testing machine. Tolerances for weights used are tabulated in Practice E2624; methods for calibration of the weights are given in NIST Technical Note 577, Methods of Calibrating Weights for Piston Gages.2
1.2 The values stated in SI units are to be regarded as standard. Other metric and inch-pound values are regarded as equivalent when required.
1.3 This practice is intended for the calibration of static torque measuring instruments. The practice is not applicable for dynamic or high-speed torque calibrations or measurements, nor can the results of calibrations performed in accordance with this practice be assumed valid for dynamic or high speed torque measurements.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
General Information
Relations
Buy Standard
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: E2428 − 15
StandardPractice for
Calibration of Torque-Measuring Instruments for Verifying
1
the Torque Indication of Torque Testing Machines
This standard is issued under the fixed designation E2428; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 2.2 American National Standard:
4
B46.1Surface Texture
1.1 This practice is to specify procedure for the calibration
5
2.3 BIPM Standard
of elastic torque-measuring instruments.
JCGM 200International vocabulary of meterology–Basic
NOTE 1—Verification by deadweight and a lever arm is an acceptable
and general concepts and associated terms (VIM)
method of verifying the torque indication of a torque testing machine.
Tolerances for weights used are tabulated in Practice E2624; methods for
ELASTIC TORQUE-MEASURING INSTRUMENTS
calibrationoftheweightsaregiveninNISTTechnicalNote577,Methods
2
of Calibrating Weights for Piston Gages.
3. Terminology
1.2 The values stated in SI units are to be regarded as
standard. Other metric and inch-pound values are regarded as 3.1 Definitions:
3.1.1 elastic torque-measuring device—a device or system
equivalent when required.
consisting of an elastic member combined with a device for
1.3 This practice is intended for the calibration of static
indicating the measured values (or a quantity proportional to
torque measuring instruments. The practice is not applicable
the measured value) of deformation of the member under an
for dynamic or high-speed torque calibrations or
applied torque.
measurements, nor can the results of calibrations performed in
3.1.2 primary torque standard—a deadweight force applied
accordance with this practice be assumed valid for dynamic or
through a lever arm or wheel, all displaying metrological
high speed torque measurements.
traceability to the International System of Units (SI).
1.4 This standard does not purport to address all of the
3.1.2.1 Discussion—for further definition of the term met-
safety concerns, if any, associated with its use. It is the
rological traceability, refer to the latest revision of JCGM:200.
responsibility of the user of this standard to establish appro-
3.1.3 secondary torque standard—an instrument or
priate safety and health practices and determine the applica-
mechanism, that has been calibrated by a comparison with a
bility of regulatory limitations prior to use.
primary torque standard(s).
3.1.4 torque—a vector product of force and length, ex-
2. Referenced Documents
pressed in terms of N-m, lbf-in., etc.
3
2.1 ASTM Standards:
3.2 Definitions of Terms Specific to This Standard:
E29Practice for Using Significant Digits in Test Data to
3.2.1 calibration equation—a mathematical relationship be-
Determine Conformance with Specifications
tween deflection and torque established from the calibration
E2624Practice for Torque Calibration of Testing Machines
data for use with the torque transducer in service, sometimes
and Devices
called the calibration curve.
3.2.1.1 Discussion—Torque transducers have torque-to-
deflection relationships that can be fitted to polynomial equa-
1
ThispracticeisunderthejurisdictionofASTMCommitteeE28onMechanical
Testing and is the direct responsibility of Subcommittee E28.01 on Calibration of tions.
Mechanical Testing Machines and Apparatus.
3.2.2 continuous-reading device—a class of instruments
Current edition approved Feb. 1, 2015. Published May 2015. Originally
whose characteristics permit interpolation of torque values
approved in 2008. Last previous edition approved in 2014 as E2428–14. DOI:
between calibrated torque values.
10.1520/E2428-15.
2
Available from National Institute of Standards and Technology (NIST), 100
Bureau Dr., Stop 1070, Gaithersburg, MD 20899-1070, http://www.nist.gov.
3
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
4
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Available fromAmerican National Standards Institute (ANSI), 25 W. 43rd St.,
Standards volume information, refer to the standard’s Document Summary page on 4th Floor, New York, NY 10036, http://www.ansi.org.
5
the ASTM website. BIPM, Pavillon de Breteuil, F-92312 Sèvres Cedex. http://www.bipm.org
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1
---------------------- Page: 1 ----------------------
E2428 − 15
3.2.2.1 Discussion—Such instruments usually have torque- 3.2.9.1 Discussion—Such instruments, usually those in
to-deflection relationships that can be fitted to polynomial which the reading is taken from a dial indicator, are used only
...
This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: E2428 − 14 E2428 − 15
Standard Practice for
Calibration of Torque-Measuring Instruments for Verifying
1
the Torque Indication of Torque Testing Machines
This standard is issued under the fixed designation E2428; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This practice is to specify procedure for the calibration of elastic torque-measuring instruments.
NOTE 1—Verification by deadweight and a lever arm is an acceptable method of verifying the torque indication of a torque testing machine. Tolerances
for weights used are tabulated in Practice E2624; methods for calibration of the weights are given in NIST Technical Note 577, Methods of Calibrating
2
Weights for Piston Gages.
1.2 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each
system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the
two systems may result in non-conformance with the standard.Other metric and inch-pound values are regarded as equivalent when
required.
1.3 This practice is intended for the calibration of static torque measuring instruments. The practice is not applicable for
dynamic or high-speed torque calibrations or measurements, nor can the results of calibrations performed in accordance with this
practice be assumed valid for dynamic or high speed torque measurements.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
limitations prior to use.
2. Referenced Documents
3
2.1 ASTM Standards:
E29 Practice for Using Significant Digits in Test Data to Determine Conformance with Specifications
E2624 Practice for Torque Calibration of Testing Machines and Devices
2.2 American National Standard:
4
B46.1 Surface Texture
5
2.3 BIPM Standard
JCGM 200 International vocabulary of meterology–Basic and general concepts and associated terms (VIM)
ELASTIC TORQUE-MEASURING INSTRUMENTS
3. Terminology
3.1 Definitions:
3.1.1 elastic torque-measuring device—a device or system consisting of an elastic member combined with a device for
indicating the measured values (or a quantity proportional to the measured value) of deformation of the member under an applied
torque.
1
This practice is under the jurisdiction of ASTM Committee E28 on Mechanical Testing and is the direct responsibility of Subcommittee E28.01 on Calibration of
Mechanical Testing Machines and Apparatus.
Current edition approved Oct. 15, 2014Feb. 1, 2015. Published May 2015. Originally approved in 2008. Last previous edition approved in 20082014 as E2428–08.–14.
DOI: 10.1520/E2428-14.10.1520/E2428-15.
2
Available from National Institute of Standards and Technology (NIST), 100 Bureau Dr., Stop 1070, Gaithersburg, MD 20899-1070, http://www.nist.gov.
3
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
4
Available from American National Standards Institute (ANSI), 25 W. 43rd St., 4th Floor, New York, NY 10036, http://www.ansi.org.
5
BIPM, Pavillon de Breteuil, F-92312 Sèvres Cedex. http://www.bipm.org
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1
---------------------- Page: 1 ----------------------
E2428 − 15
3.1.2 primary torque standards—standard—a deadweight force applied through a lever arm or wheel, with a calibrated length
or radius of a known uncertainty, that is traceable to national standards.all displaying metrological traceability to the International
System of Units (SI).
3.1.2.1 Discussion—
for further definition of the term metrological traceability, refer to the latest revision of JCGM:200.
3.1.3 secondary torque standard—an instrument or mechanism, that has been calibrated by a comparison with a primary torque
standard(s).
3.1.4 torque—a vector product of force and length, expressed in terms of N-m, lbf-in., etc.
3.2 Definitions of Terms Specific to This Standard:
3.2
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.