ASTM E2023-99(2004)
(Practice)Standard Practice for Fabrication of Neutron Radiographic Sensitivity Indicators
Standard Practice for Fabrication of Neutron Radiographic Sensitivity Indicators
SIGNIFICANCE AND USE
The only truly valid image quality indicator is a material or component, equivalent to the part being neutron radiographed, with a known standard discontinuity, inclusion, omission or flaw (reference standard comparison part). The SI is designed to substitute for the reference standard, providing qualitative information on hole and gap sensitivity in a single unit. Fabrication in accordance with this practice is vital for accurate and consistent measurements.
This practice shall be followed for the fabrication of all SIs to be used with Method E 545 to determine image quality in direct thermal neutron radiography.
SCOPE
1.1 This practice covers the fabrication of Sensitivity Indicators (SI), which can be used to determine the relative quality of radiographic images produced by direct, thermal neutron radiographic examination.
1.2 The values stated in inch-pound units are regarded as the standard. The SI equivalents given in parentheses in the text are for information only.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
General Information
Relations
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information.
Designation:E2023–99 (Reapproved 2004)
Standard Practice for
Fabrication of Neutron Radiographic Sensitivity Indicators
This standard is issued under the fixed designation E2023; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope holes of known dimensions. Visual inspection of the image of
this device provides subjective information regarding total
1.1 This practice covers the fabrication of Sensitivity Indi-
radiographic sensitivity with respect to the step-block material,
cators (SI), which can be used to determine the relative quality
aswellasoptionalsubjectivedata,regardingdetrimentallevels
of radiographic images produced by direct, thermal neutron
of gamma exposure.
radiographic examination.
4.2 NeutronradiographypracticesarediscussedinPractices
1.2 The values stated in inch-pound units are regarded to be
E748. The neutron radiograph used to determine image quality
standard. The SI equivalents given in brackets in the text are
using the SI shall meet the requirements of Method E545.
for information only.
1.3 This standard does not purport to address all of the
5. Significance and Use
safety concerns, if any, associated with its use. It is the
5.1 Theonlytrulyvalidimagequalityindicatorisamaterial
responsibility of the user of this standard to establish appro-
or component, equivalent to the part being neutron radio-
priate safety and health practices and determine the applica-
graphed, with a known standard discontinuity, inclusion, omis-
bility of regulatory limitations prior to use.
sion or flaw (reference standard comparison part). The SI is
2. Referenced Documents designed to substitute for the reference standard, providing
qualitative information on hole and gap sensitivity in a single
2.1 ASTM Standards:
unit. Fabrication in accordance with this practice is vital for
E543 Specification for Agencies Performing Nondestruc-
accurate and consistent measurements.
tive Testing
5.2 This practice shall be followed for the fabrication of all
E545 Test Method for Determining Image Quality in Direct
SIstobeusedwithMethodE545todetermineimagequalityin
Thermal Neutron Radiographic Examination
direct thermal neutron radiography.
E748 Practices for Thermal Neutron Radiography of Mate-
rials
6. Basis of Application
E1316 Terminology for Nondestructive Examinations
6.1 Qualification of Nondestructive Agencies—If specified
3. Terminology in the contractual agreement, NDT agencies shall be qualified
and evaluated as described in Practice E543. The applicable
3.1 Definitions—For definitions of terms used in this prac-
edition of Practice E543 shall be specified in the contractual
tice, see Terminology E1316, Section H.
agreement.
4. Summary of Practice 6.2 Procedures and Techniques—The procedures and tech-
niquestobeutilizedshallbeasdescribedinthispracticeunless
4.1 The Sensitivity Indicator (SI) is used for qualitative
otherwisespecified.Specifictechniquesmaybespecifiedinthe
determination of the sensitivity of detail visible on the neutron
contractual agreement.
radiograph. It consists of a step wedge containing gaps and
6.3 Reporting Criteria/Acceptance Criteria—Reporting cri-
teria for the examination results shall be in accordance with
This practice is under the jurisdiction of ASTM Committee E07 on Nonde-
Sections 9 and 10 unless otherwise specified. Acceptance
structive Testing and is the direct responsibility of Subcommittee E07.05 on
criteria, for example, reference radiographs, shall be specified
Radiology (Neutron) Method.
in the contractual agreement.
Current edition approved May 1, 2004. Published June 2004. Originally
approved in 1999. Last previous edition approved in 1999 as E2023 - 99. DOI:
6.4 Reexamination of repaired/reworked items is not ad-
10.1520/E2023-99R04.
dressedinthispracticeand,ifrequired,shallbespecifiedinthe
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contractual document.
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
E2023–99 (2004)
Material — Methylmethacrylate
Shim Thickness Hole Diameter
A 0.005 0.005
B 0.010 0.010
C 0.020 0.020
D 0.010 0.010
NOTE 1—All dimensions are in inches.
NOTE 2—The lead step may be replaced with a methylmethacrylate strip with the D shim eliminated.
FIG. 1 Sensitivity Indicator
7. Sensitivity Indicator (SI) 8.1.1 Mill a Channel, 0.850-in. (21.6-mm) wide by 6.5-in.
[165-mm]longfromanaluminumblock,1-in.[25.4-mm]wide
7.1 The Sensitivity Indicator (SI) shall be constructed of
by at least 0.303-in. [7.70-mm] high by 6.5-in. [165-mm] long.
castacrylicresin,lead(optional),andaluminum.Theconstruc-
The channel should begin 0.075-in. [1.90-mm] from edge A
tion and dimensions are shown in Fig. 1.
and leave 0.103-in. [2.62-cm] aluminum in the bottom of the
7.2 The optional lead step in the SI may be replaced with a
channel (see Fig. 2).
blank, cast acrylic resin step. The lead provides a visual
indication of beam gamma content; however, the lead image is 8.1.2 Mill the A Channel (see Fig. 1), within this channel,
0.125-in. [3.18-mm] wide by 0.005-in. [0.127-mm] deep by
not used for any of the SI calculations of Method E545.
7.3 The acrylic resin shall be methylmethacrylate. 6.5-in. [165-mm] long. The near edge of channel A should be
0.450-in. [11.4-mm] from edge A (see Fig. 3).
7.4 All dimensional tolerances are as noted on the figures.
7.5 Aluminum shims and strips shall be 99.9 % pure el-
8.1.3 Mill the B Channel, adjacent to the A channel,
emental material.
0.125-in. [3.18-mm] wide by 0.010-in. [0.254-mm] deep by
7.6 The SI may be encased in a 6061 aluminum dust cover,
6.5-in. [165-mm] long. The near edge of channel B should be
0.012-in. [0.305-mm] thick.
0.325-in. [8.26-mm] from edge A (see Fig. 3).
7.7 When used, the optional lead shim shall be at least
8.1.4 Mill the C Channel, adjacent to the B channel,
99.9 % pure elemental material.
0.125-in. [3.18-mm] wide by 0.020-in. [0.508-mm] deep by
6.5-in. [165-mm] long. The near edge of channel C should be
8. Fabrication
0.200 in. [5.08 mm] from edge A (see Fig. 3).
8.1 Components:
8.1.5 Optional—Mill the D channel adjacent to the C
channel, 0.125-in. [3.18-mm] wide by 0.010-in. [0.254-mm]
deep by 6.5-in. [165-mm] long. The near edge of channel D
The instructions in Section 8 assume the simultaneous fabrication of five units
for practical reasons. Units may be fabricated singly, if desired. should be 0.075-in. [1.90-mm] from edge A.
E2023–99 (2004)
NOTE 1—Unless otherwise specified, use the following:
Dimensions are in inches.
Tolerances on machined dimensions: .XX = 6 .01 .XXX = 6 .002.
FIG. 2 Main Channel in Aluminum Block
NOTE 1—Unless otherwise specified, use the following:
Dimensions are in inches.
Tolerances on machined dimensions: .XX = 6 .01 .XXX = 6 .002.
FIG. 3 Channels A Through D in Main Channel
E2023–99 (2004)
NOTE 1—The D channel is not required and shall not be milled if the
To verify that the various shims have the proper holes drilled
lead shim is not to be used.
into them, a certified hole measurement report is required.
8.1.6 Prepare five methylmethacrylate strips, 0.060-in. 8.1.11 To keep the SI intact during use, it is highly recom-
mended that a dust cover be kept on the unit. One dust cover
[1.52-mm] thick by at least 0.200-in. [5.08-mm] wide by
6.5-in. [165-mm] long. These strips will be used in 8.2.11. for each SI may be prepared from aluminum shim stock,
2 2
0.012-in. [0.305-mm] thick by 1.50-in. [38.1-mm ] by cutting
8.1.7 Prepare four methylmethacrylate strips, 0.125-in.
2 2
[3.18-mm] thick by at least 0.200-in. [5.08-mm] wide by out a 0.250-in. [6.35-mm ] from each corner and folding the
2 2
aluminum to form a cover, 1-in. [25.4-mm ] by 0.250-in.
6.5-in. [165-mm] long. One of the methylmethacrylate strips
may be replaced with an optional lead strip of the same [6.35-mm] high (see Fig. 1).
dimensions. These strips will be used in 8.2.5. 8.2 Assembly:
8.1.8 Prepare one each strip from aluminum shim stock, at
8.2.1 Insert Shim C into the C channel milled in 8.1.4 (see
least 0.200-in. [5.08-mm] wide by 6.5-in. [165-mm] long, with
Fig. 5).
the following thicknes
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.