ASTM F311-08(2013)
(Practice)Standard Practice for Processing Aerospace Liquid Samples for Particulate Contamination Analysis Using Membrane Filters
Standard Practice for Processing Aerospace Liquid Samples for Particulate Contamination Analysis Using Membrane Filters
SIGNIFICANCE AND USE
4.1 This practice provides for the processing of liquid samples obtained in accordance with Practice F302 and Practices F303. It will provide the optimum sample processing for visual contamination methods such as Method F312, and Test Method F314.
SCOPE
1.1 This practice covers the processing of liquids in preparation for particulate contamination analysis using membrane filters and is limited only by the liquid-to-membrane filter compatibility.
1.2 The practice covers the procedure for filtering a measured volume of liquid through a membrane filter. When this practice is used, the particulate matter will be randomly distributed on the filter surface for subsequent contamination analysis methods.
1.3 The practice describes procedures to allow handling particles in the size range between 2 and 1000 μm with minimum losses during handling.
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
General Information
Relations
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: F311 − 08 (Reapproved 2013)
Standard Practice for
Processing Aerospace Liquid Samples for Particulate
Contamination Analysis Using Membrane Filters
ThisstandardisissuedunderthefixeddesignationF311;thenumberimmediatelyfollowingthedesignationindicatestheyearoforiginal
adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.Asuperscript
epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope F302 Practice for Field Sampling of Aerospace Fluids in
Containers
1.1 This practice covers the processing of liquids in prepa-
F303 Practices for Sampling for Particles in Aerospace
ration for particulate contamination analysis using membrane
Fluids and Components
filters and is limited only by the liquid-to-membrane filter
F312 Test Methods for Microscopical Sizing and Counting
compatibility.
Particles from Aerospace Fluids on Membrane Filters
1.2 The practice covers the procedure for filtering a mea-
sured volume of liquid through a membrane filter. When this
3. Definition
practice is used, the particulate matter will be randomly
3.1 filtered liquid dispenser—as used in this practice, a
distributed on the filter surface for subsequent contamination
dispensercapableofdeliveringrinseliquidthroughafilterwith
analysis methods.
pore size no larger than half the size of the smallest particle
1.3 The practice describes procedures to allow handling
being considered for measurement.
particles in the size range between 2 and 1000 µm with
4. Significance and Use
minimum losses during handling.
4.1 This practice provides for the processing of liquid
1.4 The values stated in SI units are to be regarded as
samples obtained in accordance with Practice F302 and Prac-
standard. No other units of measurement are included in this
tices F303. It will provide the optimum sample processing for
standard.
visual contamination methods such as Method F312, and Test
2. Referenced Documents
Method F314.
2.1 ASTM Standards:
5. Apparatus and Materials
D287 Test Method for API Gravity of Crude Petroleum and
Petroleum Products (Hydrometer Method) 5.1 Filtration Funnel—The funnel opening in contact with
the membrane shall be approximately 35.0 mm in inside
D1078 Test Method for Distillation Range of Volatile Or-
ganic Liquids diameter. The effective area shall be calibrated. If the funnel is
to be used for measuring the sample volume, the funnel shall
D1193 Specification for Reagent Water
D1353 Test Method for Nonvolatile Matter in Volatile Sol- be calibrated within 62 % of the required volume.
vents for Use in Paint, Varnish, Lacquer, and Related
5.2 Membrane Filter Support—Either a fritted-glass,
Products
sintered-metal, polyphenyl-sulfone or stainless steel screen
D1836 Specification for Commercial Hexanes
may be used. The support shall be so designed as to enable
D2021 Specification for Neutral Detergent, 40 PercentAlky-
attachment to a vacuum flask.
lbenzene Sulfonate Type (Withdrawn 2000)
5.3 Vacuum Flask.
5.4 Funnel-Holding Device—A provision should be made
This practice is under the jurisdiction of ASTM Committee E21 on Space
for the dissipation of static electricity from the funnel.
Simulation andApplications of Space Technology and is the direct responsibility of
Subcommittee E21.05 on Contamination. 5.5 A clean bench or hood, supplied with unidirectional
Current edition approved Nov. 1, 2013. Published November 2013. Originally
flow filtered air, in which uncovered components may be
approved in 1969. Last previous edition approved in 2008 as F311 – 08. DOI:
placed.
10.1520/F0311-13.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
5.6 Vacuum Source—Minimum vacuum gage reading of 61
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
kPa (or other metric units acceptable to ASTM).
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website.
5.7 Forceps, unserrated tips.
The last approved version of this historical standard is referenced on
www.astm.org. 5.8 Filtered Liquid Dispenser.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
F311 − 08 (2013)
used to avoid inhalation of the vapors.
5.9 Membrane Filter, pore size no greater than half the size
of the smallest particle being considered for measurement.The
6.6 Rinse Liquid—This liquid shall be specified by the user
filter shall have an imprinted grid on 3.10 6 0.02-mm centers.
and shall be compatible with the sample liquid the plastic film
The color shall be chosen to provide maximum contrast with
(see section 5.12) and the membrane filter.
the particulate contamination to be observed.
7. Preparation of Apparatus
NOTE 1—When testing military hydraulic fluids, a white cellulose
membrane of 0.45-µm pore size, capable of filtering 80 mL of distilled 7.1 Wash the apparatus thoroughly in a solution of liquid
water per square centimeter per minute at 93 kPa (or other metric units
detergent and hot water and rinse with hot tap water (Note 3).
acceptable to ASTM) and 25°C is required.
Rinse with filtered isopropyl alcohol to remove the water; then
5.10 Petri Dishes, glass. Plastic Petri dishes may be used rinse twice with filtered solvent.
only if the selected plastic is known to be compatible with the
NOTE 3—Distilled or deionized water shall be used in areas where
filtered liquid.
hardness or contamination increase the blank count over the allowable
level.
5.11 Glass Bottles, small-mouth, screw-capped, perma-
nently marked to indicate sample volume.
7.2 Sample Bottle Preparation—Repeattheprocedureof7.1
for both the sample bottles and caps, then allow them to drip
5.12 Plastic Film, 0.002-in. (0.05-mm) minimum thickness.
dry. Place a piece of plastic film that has been rinsed with
5.13 Contrasting Pigment Suspension, to be used in calibra-
filtered liquid over the mouth of the bottle. Hold the film while
tion of the funnel.
screwing on the cap to prevent the film from rotating.
5.14 Slides, glass, 50 by 50-mm.
NOTE 4—It is important to hold the film when applying and removing
the cap to prevent serration.
5.15 Tape, transparent, pressure-sensitive.
8. Funnel Calibration
6. Reagents and Materials
8.1 Disperse a small quantity of pigment into 100 mL of
6.1 Purity of Reagents—Reagent grade chemicals shall be
liquid in a glass bottle. Prepare for a filtration as described in
used in all tests. Unless otherwise indicated, it is intended that
9.1. Filt
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.