EN ISO 8041:2005/prA1:2015
(Amendment)Human response to vibration - Measuring instrumentation - Amendment 1 (ISO 8041:2005/DAM 1:2015)
Human response to vibration - Measuring instrumentation - Amendment 1 (ISO 8041:2005/DAM 1:2015)
2016-01-21: WI stopped to follow ISO decision - revision started (see WI 00231111).
Schwingungseinwirkung auf den Menschen - Messeinrichtung - Änderung 1 (ISO 8041:2005/DAM 1:2015)
Réponse des individus aux vibrations - Appareillage de mesure - Amendement 1 (ISO 8041:2005/DAM 1:2015)
Odzivanje človeka na vibracije - Merilni instrumentarij - Dopolnilo A1 (ISO 8041:2005/DAM 1:2015)
General Information
Relations
Standards Content (Sample)
SLOVENSKI STANDARD
01-julij-2015
2G]LYDQMHþORYHNDQDYLEUDFLMH0HULOQLLQVWUXPHQWDULM'RSROQLOR$,62
'$0
Human response to vibration - Measuring instrumentation - Amendment 1 (ISO
8041:2005/DAM 1:2015)
Schwingungseinwirkung auf den Menschen - Messeinrichtung - Änderung 1 (ISO
8041:2005/DAM 1:2015)
Réponse des individus aux vibrations - Appareillage de mesure - Amendement 1 (ISO
8041:2005/DAM 1:2015)
Ta slovenski standard je istoveten z: EN ISO 8041:2005/prA1:2015
ICS:
13.160 Vpliv vibracij in udarcev na Vibration and shock with
ljudi respect to human beings
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
DRAFT INTERNATIONAL STANDARD
ISO/DIS 8041
ISO/TC 108/SC 3 Secretariat: DS
Voting begins on: Voting terminates on:
2015-04-02 2015-07-02
Human response to vibration — Measuring
instrumentation
AMENDMENT 1
Réponse des individus aux vibrations — Appareillage de mesure
AMENDEMENT 1
ICS: 13.160
ISO/CEN PARALLEL PROCESSING
This draft has been developed within the International Organization for
Standardization (ISO), and processed under the ISO lead mode of collaboration
as defined in the Vienna Agreement.
This draft is hereby submitted to the ISO member bodies and to the CEN member
bodies for a parallel five month enquiry.
Should this draft be accepted, a final draft, established on the basis of comments
received, will be submitted to a parallel two-month approval vote in ISO and
THIS DOCUMENT IS A DRAFT CIRCULATED
formal vote in CEN.
FOR COMMENT AND APPROVAL. IT IS
THEREFORE SUBJECT TO CHANGE AND MAY
NOT BE REFERRED TO AS AN INTERNATIONAL
STANDARD UNTIL PUBLISHED AS SUCH.
To expedite distribution, this document is circulated as received from the
IN ADDITION TO THEIR EVALUATION AS
committee secretariat. ISO Central Secretariat work of editing and text
BEING ACCEPTABLE FOR INDUSTRIAL,
composition will be undertaken at publication stage.
TECHNOLOGICAL, COMMERCIAL AND
USER PURPOSES, DRAFT INTERNATIONAL
STANDARDS MAY ON OCCASION HAVE TO
BE CONSIDERED IN THE LIGHT OF THEIR
POTENTIAL TO BECOME STANDARDS TO
WHICH REFERENCE MAY BE MADE IN
Reference number
NATIONAL REGULATIONS.
ISO/DIS 8041:2015(E)
RECIPIENTS OF THIS DRAFT ARE INVITED
TO SUBMIT, WITH THEIR COMMENTS,
NOTIFICATION OF ANY RELEVANT PATENT
RIGHTS OF WHICH THEY ARE AWARE AND TO
©
PROVIDE SUPPORTING DOCUMENTATION. ISO 2015
ISO/DIS 8041:2015(E)
© ISO 2015
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized otherwise in any form
or by any means, electronic or mechanical, including photocopying, or posting on the internet or an intranet, without prior
written permission. Permission can be requested from either ISO at the address below or ISO’s member body in the country of
the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland
ii © ISO 2015 – All rights reserved
ISO/DIS 8041:2015(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of any
patent rights identified during the development of the document will be in the Introduction and/or on
the ISO list of patent declarations received (see www.iso.org/patents).
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation on the meaning of ISO specific terms and expressions related to conformity
assessment, as well as information about ISO’s adherence to the WTO principles in the Technical Barriers
to Trade (TBT) see the following URL: Foreword - Supplementary information
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation on the meaning of ISO specific terms and expressions related to conformity
assessment, as well as information about ISO’s adherence to the WTO principles in the Technical Barriers
to Trade (TBT) see the following URL: Foreword - Supplementary information
The committee responsible for this document is ISO/TC 108, Mechanical vibration, shock and condition
monitoring, Subcommittee SC 3, Use and calibration of vibration and shock measuring instruments.
Page v, fifth indent
Add after pattern evaluation “or validation”.
Page vi
Add the following Introduction:
ISO/DIS 8041:2015(E)
Human response to vibration — Measuring
instrumentation
AMENDMENT 1
Introduction
Up to 2005, when this International Standard was published, measuring instrumentation for
human response to vibration (vibration meters) normally consisted of a signal processing unit and
a detachable vibration transducer. According to recent developments, however, part of the signal
processing steps can be integrated in the transducer unit, so that the signal coming out of the
transducer’s sensing element and going into the signal conditioning unit is not accessible any more.
These transducer units include e.g. IEPE and MEMS transducers.
Some of the test procedures specified in this International Standard, however, presume that this point in
the signal path is accessible (electrical input). Since such an input is not mandatory these tests can only
be performed on a vibration meter having an electrical input or after some technical modifications to
the instrumentation, e.g. internal access to signals. Or those tests can only be performed mechanically,
which in certain cases requires modifications to some test procedures. Such modifications to test
procedures, however, are beyond the present scope of this International Standard.
Some of the test procedures specified in this International Standard can only be performed if an
electrical output is available, e.g. 5.13. Since such an output is not mandatory these tests can only be
performed on a vibration meter having an electrical output or after some technical modifications to
the instrumentation, e.g. internal access to signals.
The verification test specified in this Amendment is practicable and has the objective of identifying
an instrument which is adequately calibrated for the intended applications and is suitable for the
purpose, at a cost reasonable for the calibration laboratory and affordable for the end user. Therefore
the verification test is strongly reduced in its extend compared to the full pattern evaluation or
validation and does only test the most relevant characteristcs of a vibration meter.
Page 1, Clause 1, first paragraph
Add after pattern evaluation “or validation”.
Renumber a) to be a1) and add the following:
a2) validation of one-off instruments, i.e. a limited set of tests of an individual vibration measuring
system against the relevant specifications defined in this International Standard;
Page 4, 3.1.5.7
Replace Formula (7) by the following:
222 2 2 2
ak=+ak ak+ a (7)
v xxw yyw zzw
Page 6, Figure 1
Replace in the figure capture “Overview” by “Example”.
ISO/DIS 8041:2015(E)
Page 7 and 8, 5.1
Add after the first and the third indent, respectively, the following symbols:
a
w
T
Page 17, 5.10, fifth paragraph, fifth line
Replace “a further” by “at least”.
Page 17, 5.10
Add at the end of the subclause the following Note:
NOTE — It is useful that the vibration meter is capable of indicating how long (in relation to the
measurement duration) in each channel an overload took place.
Page 19, 5.16
Add at the end of the subclause the following Note:
NOTE — This electrical cross-talk between measurement channels should be distinguished
from transducer transverse sensitivity which typically is greater than 0,5 %, see Clause E.2. The
transverse sensitivity can become relevant especially with multi-axial transducers.
Page 23, Clause 11
Replace a) by the following:
a1) Pattern evaluation (targeted at manufacturers) is a full set of tests to be performed on samples of
an instrument type. This test is used for product type testing or pattern approval of a production
series of vibration measuring instruments. The objective of this test is to demonstrate an instrument
design can meet the specifications defined in this International Standard;
a2) Validation of one-off instruments (targeted at laboratory users or research users) is a limited set
of tests to demonstrate that an individual vibration measuring system complies with the relevant
specifications defined in this International Standard;
Replace in b), first indent, “every 1 or” by “at least every”.
Delete in b) the second indent.
Add in b) in the last indent after modification “(e.g. use of another transducer)”.
Page 24, Table 13
Replace the table by the following:
2 © ISO 2015 – All rights reserved
ISO/DIS 8041:2015(E)
Table 13 — Summary of performance characteristics and test requirements
Specification Test type Test clause
Pattern
Validation
evaluation Periodic
Mechani- (one-off In-situ
Clause Characteristic Electrical (pro- verifica-
cal instru- check
duction tion
ments)
series)
5.1 General characteristics 12.5, 12.6 13.6, 13.7 14.5 15.2
Display of signal magni-
5.2 12.5 13.6
tude
5.3 Electrical output • 12.17
5.4 Vibration sensitivity • 12.7 13.9, 13.11 14.9 15.3
Accuracy of indication at
5.5 reference frequency under • 12.7 13.9
reference conditions
Frequency weightings and 12.11
5.6 • • 13.11 14.9
frequency responses Annex H
5.7 Amplitude linearity • • 12.10 13.11 14.9
5.8 Instrument noise • 12.12 13.12
5.9 Signal-burst response • 12.13
5.10 Overload indication • 12.14 13.14
5.11 Under-range indication • • 12.10
5.12 Time averaging • 12.13
Running r.m.s. accelera-
5.13 • 12.13
tion
5.14 Reset 12.15
5.15 Timing facilities 12.18 13.15
5.16 Electrical cross-talk • 12.8 13.13
5.2 Combined axis outputs • 12.16
5.17 Vibration transducer
• 12.9
(Annex E) characteristics
5.18 Power supply 12.19
6 Mounting Annex F
Environmental and elec-
7 12.20
tromagnetic criteria
Provision for use with
8 • 12.5,12.17
auxiliary devices
9 Instrument marking 12.4 13.5 14.6
Instrument documenta-
10 12.4 13.5 14.6
tion
Page 25, 12.3
Add the following second and third paragraphs:
The instrument shall be submitted for testing with equipment (e.g. adaptors) suitable for enabling the
input and output of electrical signals.
A vibration calibrator should be supplied with the vibration instrument.
ISO/DIS 8041:2015(E)
Page 26, 12.6, first paragraph
Replace the second sentence by “The instrument, transducer and the vibration calibrator (if supplied)
shall be visually inspected and all controls operated to ensure they are in working order.”
Page 26, 12.7
Replace the third paragraph by the following:
The reference vibration transducer shall be used to measure the value of the mechanical vibration
input generated at the reference vibration value and at the reference frequency, before measuring the
vibration magnitude with the vibration meter. For these measurements, the vibration meter shall be set
to the reference measurement range, band-limiting frequency-weighting and linear time averaging and
with a measurement duration sufficient for the indicated values to stabilize (for low-frequency testing,
this duration can be as much as 30 s for hand-arm vibration, 1 min for whole-body, and 5 min for low-
frequency whole-body applications). The value of the input signal plus background noise shall be at least
10 times the value of the background noise as measured by the vibration meter.
Page 28, 12.10.1, third paragraph from top of page
Add in the first sentence after specified lower boundary “[see G.2 j)]”.
Page 28, 12.10.2, first paragraph
Replace the last sentence by “The vibration transducers shall be mounted in accordance with the
calibration procedure specified in ISO 16063-21.”
Page 29, 12.10.2, paragraph in between the six indents
Replace “from the signal value that caused the first indication of overload” by “from that maximum
signal value”.
Page 29, 12.10.2, last but two paragraph
Replace “until the first indication of overload” by “until the maximum signal value specified above”.
Page 29, 12.10.2, last paragraph
Delete “and until the first indications of overload”.
Page 30, 12.11.2, first paragraph
Replace the last sentence by “The vibration transducers shall be mounted in accordance with the
calibration procedure specified in ISO 16063-21.”
Page 30, 12.11.2
Replace the second paragraph by the following:
At the reference frequency, the input mechanical vibration shall be adjusted such that the indication
of band-limited vibration on the instrument under test is 20 dB above the lower limit of the specified
, shall be used as a
linearity range. The band-limited acceleration value of this input signal, a
in
reference input value for subsequent tests.
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.