Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems

2020-11-16_JO_BT167/DG11868/DV to remove the link to (MDD 93/42/EEC) and M/432

Medizinische elektrische Geräte - Teil 1-8: Allgemeine Festlegungen für die Sicherheit einschließlich der wesentlichen Leistungsmerkmale - Ergänzungsnorm: Alarmsysteme - Allgemeine Festlegungen, Prüfungen und Richtlinien für Alarmsysteme in medizinischen elektrischen Geräten und in medizinischen elektrischen Systemen

Appareils électromédicaux - Partie 1-8 : exigences générales pour la sécurité de base et les performances essentielles - Norme collatérale: exigences générales, essais et guide pour les systèmes d'alarme des appareils et des systèmes électromédicaux

Medicinska električna oprema - 1-8. del: Splošne zahteve za osnovno varnost in bistvene lastnosti - Spremljevalni standard: Splošne zahteve, preskušanje in navodila za alarmne sisteme v medicinski električni opremi in medicinskih električnih sistemih - Dopolnilo A2 (IEC 60601-1-8:2006/A2:2020)

General Information

Status
Published
Publication Date
15-Jul-2021
Current Stage
6060 - Document made available - Publishing
Start Date
16-Jul-2021
Completion Date
16-Jul-2021

Relations

Buy Standard

Amendment
EN 60601-1-8:2008/A2:2021 - BARVE
English language
61 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)


SLOVENSKI STANDARD
01-september-2021
Medicinska električna oprema - 1-8. del: Splošne zahteve za osnovno varnost in
bistvene lastnosti - Spremljevalni standard: Splošne zahteve, preskušanje in
navodila za alarmne sisteme v medicinski električni opremi in medicinskih
električnih sistemih - Dopolnilo A2 (IEC 60601-1-8:2006/A2:2020)
Medical electrical equipment - Part 1-8: General requirements for basic safety and
essential performance - Collateral Standard: General requirements, tests and guidance
for alarm systems in medical electrical equipment and medical electrical systems (IEC
60601-1-8:2006/A2:2020)
Medizinische elektrische Geräte - Teil 1-8: Allgemeine Festlegungen für die Sicherheit
einschließlich der wesentlichen Leistungsmerkmale - Ergänzungsnorm: Alarmsysteme -
Allgemeine Festlegungen, Prüfungen und Richtlinien für Alarmsysteme in medizinischen
elektrischen Geräten und in medizinischen Systemen (IEC 60601-1-8:2006/A2:2020)
Appareils électromédicaux - Partie 1-8: Exigences générales pour la sécurité de base et
les performances essentielles - Norme collatérale: Exigences générales, essais et guide
pour les systèmes d'alarme des appareils et des systèmes électromédicaux (IEC 60601-
1-8:2006/A2:2020)
Ta slovenski standard je istoveten z: EN 60601-1-8:2007/A2:2021
ICS:
11.040.01 Medicinska oprema na Medical equipment in general
splošno
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

EUROPEAN STANDARD EN 60601-1-8:2007/A2

NORME EUROPÉENNE
EUROPÄISCHE NORM
July 2021
ICS 11.040.01
English Version
Medical electrical equipment - Part 1-8: General requirements for
basic safety and essential performance - Collateral Standard:
General requirements, tests and guidance for alarm systems in
medical electrical equipment and medical electrical systems
(IEC 60601-1-8:2006/A2:2020)
Appareils électromédicaux - Partie 1-8 : exigences Medizinische elektrische Geräte - Teil 1-8: Allgemeine
générales pour la sécurité de base et les performances Festlegungen für die Sicherheit einschließlich der
essentielles - Norme collatérale: exigences générales, wesentlichen Leistungsmerkmale - Ergänzungsnorm:
essais et guide pour les systèmes d'alarme des appareils et Alarmsysteme - Allgemeine Festlegungen, Prüfungen und
des systèmes électromédicaux Richtlinien für Alarmsysteme in medizinischen elektrischen
(IEC 60601-1-8:2006/A2:2020) Geräten und in medizinischen Systemen
(IEC 60601-1-8:2006/A2:2020)
This amendment A2 modifies the European Standard EN 60601-1-8:2007; it was approved by CENELEC on 2020-08-27. CENELEC
members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this amendment the
status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC
Management Centre or to any CENELEC member.
This amendment exists in three official versions (English, French, German). A version in any other language made by translation under the
responsibility of a CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the same status as
the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the
Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and the United Kingdom.

European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2021 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
Ref. No. EN 60601-1-8:2007/A2:2021 E

European foreword
The text of document 62A/1392/FDIS, future IEC 60601-1-8/A2, prepared by SC 62A "Common
aspects of electrical equipment used in medical practice" of IEC/TC 62 "Electrical equipment in
medical practice" was submitted to the IEC-CENELEC parallel vote and approved by CENELEC as
The following dates are fixed:
• latest date by which the document has to be implemented at national (dop) 2022-01-16
level by publication of an identical national standard or by endorsement
• latest date by which the national standards conflicting with the (dow) 2024-07-16
document have to be withdrawn
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC shall not be held responsible for identifying any or all such patent rights.
Any feedback and questions on this document should be directed to the users’ national committee. A
complete listing of these bodies can be found on the CENELEC website.
Endorsement notice
The text of the International Standard IEC 60601-1-8:2006/A2:2020 was approved by CENELEC as a
European Standard without any modification.
In the official version, for Bibliography, the following notes have to be added for the standards
indicated:
IEC 80001-1 NOTE Harmonized as EN 80001-1
ISO 9000:2015 NOTE Harmonized as EN ISO 9000:2015 (not modified)
Annex ZA
(normative)
Normative references to international publications
with their corresponding European publications
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments)
applies.
NOTE 1 Where an International Publication has been modified by common modifications, indicated by (mod),
the relevant EN/HD applies.
NOTE 2 Up-to-date information on the latest versions of the European Standards listed in this annex is
available here: www.cenelec.eu.
Replace Annex ZA by the following one:
Publication Year Title EN/HD Year
IEC 60417 - Graphical symbols for use on equipment. - -
Index, survey and compilation of the single
sheets.
IEC 60601-1 2005 Medical electrical equipment - Part 1: EN 60601-1 2006
General requirements for basic safety and
essential performance
- -  + corrigendum Mar. 2010
+ A1 2012  + A1 2013
- -  + A12 2014
+ A2 2020  + A2 2021
IEC 61672-1 2013 Electroacoustics - Sound level meters - EN 61672-1 2013
Part 1: Specifications
IEC 62366-1 2015 Medical devices - Part 1: Application of EN 62366-1 2015
usability engineering to medical devices
- -  + AC 2015
+ A1 2020  + A1 2020
ISO 3744 2010 Acoustics - Determination of sound power EN ISO 3744 2010
levels and sound energy levels of noise
sources using sound pressure -
Engineering methods for an essentially
free field over a reflecting plane
ISO 7000 - Graphical symbols for use on equipment - - -
Registered symbols
IEC 60601-1-8
Edition 2.0 2020-07
INTERNATIONAL
STANDARD
colour
inside
AMENDMENT 2
Medical electrical equipment –

Part 1-8: General requirements for basic safety and essential performance –

Collateral standard: General requirements, tests and guidance for alarm

systems in medical electrical equipment and medical electrical systems

INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
ICS 11.040.01 ISBN 978-2-8322-8631-9

– 2 – IEC 60601-1-8:2006/AMD2:2020
© IEC 2020
FOREWORD
This amendment has been prepared by subcommittee 62A: Common aspects of electrical
equipment used in medical practice, of IEC technical committee 62: Electrical equipment in
medical practice, and ISO subcommittee 3: Respiratory devices and related equipment used for
patient care, of ISO technical committee 121: Anaesthetic and respiratory equipment.
It is published as a double logo amendment.
The text of this amendment is based on the following documents of IEC:
FDIS Report on voting
62A/1392/FDIS 62A/1407/RVD
Full information on the voting for the approval of this amendment can be found in the report on
voting indicated in the above table. In ISO, the amendment has been approved by 15 P
members out of 15 having cast a vote.
The committee has decided that the contents of this amendment and the base publication will
remain unchanged until the stability date indicated on the IEC web site under
"http://webstore.iec.ch" in the data related to the specific publication. At this date, the
publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
NOTE The attention of users of this document is drawn to the fact that equipment manufacturers and testing
organizations may need a transitional period following publication of a new, amended or revised IEC or
ISO publication in which to make products in accordance with the new requirements and to equip themselves for
conducting new or revised tests. It is the recommendation of the committee that the content of this publication be
adopted for mandatory implementation nationally not earlier than 3 years from the date of publication.

IMPORTANT – The 'colour inside' logo on the cover page of this publication indicates
that it contains colours which are considered to be useful for the correct understanding
of its contents. Users should therefore print this document using a colour printer.

_____________
IEC 60601-1-8:2006/AMD2:2020 – 3 –
© IEC 2020
INTRODUCTION TO AMENDMENT 2
The second edition of IEC 60601-1-8 was published in 2006 and amended in 2012. Since the
publication of IEC 60601-1-8:2006+A1:2012, the IEC Subcommittee (SC) 62A Secretariat has
been collecting issues from a variety of sources including comments from National Committees.
At the November 2015 meeting of IEC/SC 62A in Kobe, Japan, the subcommittee initiated a
process to identify high-priority issues that need to be considered in an amendment and should
not wait until the third edition of IEC 60601-1-8, which is presently targeted for publication
sometime after 2024.
Those issues selected for inclusion on the final "short list" to be addressed in Amendment 2
were those approved by a 2/3 majority of the National Committees present and voting at the
Frankfurt meeting of SC 62A. At the meeting held on 10 October 2016, 20 items were presented
to the National Committees present. All 20 items received the required 2/3 majority of the
National Committees present and voting and have been included in the "short list" for
consideration in preparing Amendment 2. All remaining issues have been placed on a "long list"
for consideration in the third edition of IEC 60601-1-8.
The "short list" of issues was documented in the design specification for Amendment 2. As
IEC 60601-1-8 was jointly developed with ISO/TC 121/SC 3, the work was assigned to
IEC/SC 62A-ISO/TC 121/SC 3 Joint Working Group (JWG) 2. JWG 2 was directed to consider
each issue described in Clause 6 of the design specification and develop an appropriate
solution for the identified problem. That final solution in this amendment can encompass any
technical solution proposed by the author of the issue or it can involve a different solution
developed by the expert group. The expert group can also have recommended that no change
to the standard was justified by the problem statement.
Because this is an amendment to IEC 60601-1-8:2006, the style in force at the time of
publication of IEC 60601-1-8 has been applied to this amendment. The style specified in
ISO/IEC Directives Part 2:2018 has only been applied when implementing the new style
guidance would not result in additional editorial changes. For example, notes to definitions are
designated as "NOTE" rather than "Note to entry" in Clause 3.
Users of this document should note that when constructing the dated references to specific
elements in a standard, such as definitions, amendments are only referenced if they modified
the text being cited. For example, if a reference is made to a definition that has not been
modified by an amendment, then the reference to the amendment is not included in the dated
reference.
– 4 – IEC 60601-1-8:2006/AMD2:2020
© IEC 2020
INTRODUCTION
Replace, in the second sentence of the existing second paragraph, "source" with "origin".
1.3.1 IEC 60601-1
Replace the first two existing dashes with the following new dashes:
– "the general standard" designates IEC 60601-1 alone, including any amendments;
– "this collateral standard" designates IEC 60601-1-8 alone, including any amendments;
2 Normative references
Replace the existing references to IEC 60601-1, IEC 61672-1 and IEC 62366-1 by the following
new references:
IEC 60601-1:2005, Medical electrical equipment – Part 1: General requirements for basic safety
and essential performance
Amendment 1:2012
Amendment 2:2020
IEC 61672-1:2013, Electroacoustics – Sound level meters – Part 1: Specifications
IEC 62366-1:2015, Medical devices – Part 1: Application of usability engineering to medical
devices
Amendment 1:2020
3 Terms and definitions
Replace the existing first paragraph with the following new paragraph:
For the purposes of this document, the terms and definitions given in
IEC 60601-1:2005+A1:2012+A2:2020, IEC 62366-1:2015+A1:2020, and the following
definitions apply.
3.17
* DISTRIBUTED ALARM SYSTEM
Replace the existing term and definition with the following new entry:
3.17
* DISTRIBUTED ALARM SYSTEM
DAS
ALARM SYSTEM that involves more than one item of equipment in a ME SYSTEM intended for delivery
of ALARM CONDITIONS with tec
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.