Safety requirements for electrical equipment for measurement, control and laboratory use - Part 031: Safety requirements for hand-held and hand-manipulated probe assemblies for electrical test and measurement

1.1.1 Probe assemblies included in scope This part of IEC 61010 specifies safety requirements for hand-held and hand-manipulated probe assemblies of the types described below, and their related accessories. These probe assemblies are for direct electrical connection between a part and electrical test and measurement equipment. They may be fixed to the equipment or be detachable accessories for the equipment. a) Type A: low-voltage and high-voltage, non-attenuating probe assemblies. Non-attenuating probe assemblies that are RATED for direct connection to voltages exceeding 30 V r.m.s., 42,4 V peak, or 60 V d.c., but not exceeding 63 kV. They do not incorporate components which are intended to provide a voltage divider function or a signal conditioning function, but they may contain non-attenuating components such as fuses (see Figure 1). b) Type B: high-voltage attenuating or divider probe assemblies. Attenuating or divider probe assemblies that are RATED for direct connection to secondary voltages exceeding 1 kV r.m.s or 1,5 kV d.c. but not exceeding 63 kV r.m.s or d.c. The divider function may be carried out wholly within the probe assembely, or partly within the test or measurement equipment to be used with the probe assembly (see Figure 2). c) Type C: low-voltage attenuating or divider probe assemblies. Attenuating or divider probe assemblies for direct connection to voltages not exceeding 1 kV r.m.s. or 1,5 kV d.c. The signal conditioning function may be carried out wholly within the probe assembly, or partly within the test or measurement equipment intended to be used with the probe assembly (see Figure 3). d) Type D: low-voltage attenuating, non-attenuating or other signal conditioning probe assemblies, that are RATED for direct connection only to voltages not exceeding 30 V r.m.s., or 42,4 V peak, or 60 V d.c., and are suitable for currents exceeding 8 A (see Figure 4). 1.1.2 Probe assemblies excluded from scope This standard does not apply to current sensors within the scope of IEC 61010-2-032 (hand-held and hand-manipulated current sensors), but may apply to their input measuring circuit leads and accessories.

Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte - Teil 031: Sicherheitsbestimmungen für handgehaltenes Messzubehör zum elektrischen Messen und Prüfen

Exigences de sécurité pour appareils électriques de mesurage, de régulation et de laboratoire - Partie 031: Exigences de sécurité pour sondes équipées portatives et manipulées à la main pour mesurage et essais électriques

Varnostne zahteve za električno opremo za meritve, nadzor in laboratorijsko uporabo - 031. del: Varnostne zahteve za ročne sonde za električne meritve in preskušanje - Dopolnilo A11

Ta del standarda IEC 61010 določa varnostne zahteve za spodaj navedene tipe ročnih in ročno vodenih sond ter pripadajočo dodatno opremo. Te sonde so namenjene neposredni električni povezavi med delom in električnimi preskušanji ter merilno opremo. Pritrdijo se lahko na opremo ali so odstranljivi dodatki za opremo.
a) Tip A: nizkonapetostne in visokonapetostne neslabilne sonde. Neslabilne sonde, ki so NAZNAČENE za neposredno povezavo na napetosti nad 30 V efektivne izmenične napetosti, 42,4 V vršne vrednosti ali 60 V enosmerne napetosti, vendar ne presegajo 63 kV. Ne zajemajo komponent, ki so namenjene zagotavljanju funkcije razdelitve napetosti ali kondicioniranja signala, vendar lahko zajemajo neslabilne komponente, kot so varovalke (glej sliko 1).
b) Tip B: visokonapetostne slabilne ali razdelitvene sonde. Slabilne ali razdelitvene sonde, ki so NAZNAČENE za neposredno povezavo na drugotne napetosti nad 1 kV efektivne izmenične napetosti ali 1,5 kV enosmerne napetosti, vendar ne presegajo 63 kV efektivne izmenične napetosti ali enosmerne napetosti. Funkcija razdelitve se lahko v celoti izvede v sondi ali delno v preskusni ali merilni opremi, ki se uporablja s sondo (glej sliko 2).
c) Tip C: nizkonapetostne slabilne ali razdelitvene sonde. Slabilne ali razdelitvene sonde za neposredno povezano na napetosti, ki ne presegajo 1 kV efektivne izmenične napetosti ali 1,5 kV enosmerne napetosti. Funkcija kondicioniranja signala se lahko v celoti izvede v sondi ali delno v preskusni ali merilni opremi, ki je namenjena uporabi s sondo (glej sliko 3).
d) Tip D: nizkonapetostne slabilne, neslabilne ali druge sonde za kondicioniranje signala, ki so NAZNAČENE za neposredno povezavo le na napetosti pod 30 V efektivne izmenične napetosti ali 42,4 V vršne vrednosti ali 60 V enosmerne napetosti in so primerne za tok nad 8 A (glej sliko 4). Ta standard se ne uporablja za tokovne senzorje znotraj področja uporabe standarda IEC 61010-2-032 (Ročni in ročni vodeni tokovni senzorji), vendar se lahko uporabljajo za njihove vhodne merilne tokovne vode in dodatno opremo.

General Information

Status
Published
Publication Date
11-Nov-2021
Current Stage
6060 - Document made available - Publishing
Start Date
12-Nov-2021
Due Date
07-Mar-2022
Completion Date
12-Nov-2021

Relations

Buy Standard

Amendment
EN 61010-031:2015/A11:2022
English language
8 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)


SLOVENSKI STANDARD
01-marec-2022
Varnostne zahteve za električno opremo za meritve, nadzor in laboratorijsko
uporabo - 031. del: Varnostne zahteve za ročne sonde za električne meritve in
preskušanje - Dopolnilo A11
Safety requirements for electrical equipment for measurement, control and laboratory
use - Part 031: Safety requirements for hand-held probe assemblies for electrical
measurement and test
Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte - Teil
031: Sicherheitsbestimmungen für handgehaltenes Messzubehör zum elektrischen
Messen und Prüfen
Exigences de sécurité pour appareils électriques de mesurage, de régulation et de
laboratoire - Partie 2-031: Exigences de sécurité pour sondes équipées tenues à la main
pour mesurage et essais électriques
Ta slovenski standard je istoveten z: EN 61010-031:2015/A11:2021
ICS:
19.080 Električno in elektronsko Electrical and electronic
preskušanje testing
71.040.10 Kemijski laboratoriji. Chemical laboratories.
Laboratorijska oprema Laboratory equipment
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

EUROPEAN STANDARD EN 61010-031:2015/A11

NORME EUROPÉENNE
EUROPÄISCHE NORM
November 2021
ICS 19.080
English Version
Safety requirements for electrical equipment for measurement,
control and laboratory use - Part 031: Safety requirements for
hand-held and hand-manipulated probe assemblies for electrical
test and measurement
Exigences de sécurité pour appareils électriques de Sicherheitsbestimmungen für elektrische Mess-, Steuer-,
mesurage, de régulation et de laboratoire - Partie 031: Regel- und Laborgeräte - Teil 031:
Exigences de sécurité pour sondes équipées portatives et Sicherheitsbestimmungen für handgehaltenes
manipulées à la main pour mesurage et essais électriques Messzubehör zum elektrischen Messen und Prüfen
This amendment A11 modifies the European Standard ; it was approved by CENELEC on 2021-08-02. CENELEC members are bound to
comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this amendment the status of a national
standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC
Management Centre or to any CENELEC member.
This amendment exists in three official versions (English, French, German). A version in any other language made by translation under the
responsibility of a CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the same status as
the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the
Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and the United Kingdom.

European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2021 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
Ref. No. EN 61010-031:2015/A11:2021 E

European foreword
This document (EN 61010-031:2015/A11:2021) has been prepared by CLC/TC 66X “Safety of
measuring, control, and laboratory equipment”.
The following dates are fixed:
• latest date by which this document has to be (dop) 2022-05-12
implemented at national level by publication of
an identical national standard or by
endorsement
• latest date by which the national standards (dow) 2024-11-12
conflicting with this document have to be
withdrawn
This document amends EN 61010-031:2015/A1:2021.
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC shall not be held responsible for identifying any or all such patent rights.
This document has been prepared under a Standardization Request given to CENELEC by the
European Commission and the European Free Trade Association, and supports essential requirements
of EU Directive(s) / Regulation(s).
For relationship with EU Directive(s) / Regulation(s), see informative Annex ZZ, which is an integral part
of this document.
Any feedback and questions on this document should be directed to the users’ national committee. A
complete listing of these bodies can be found on the CENELEC website.
1 Modifications to Clause 2, “Normative references”
Replace IEC 60027 (all parts), IEC 61180-1:1992 and IEC 61180-2 with the following documents in
Clause 2 and throughout the document:

EN 60027-1:2006, Letters symbols to be used in electrical technology- Part 1: General (IEC 60027-
1:1992+A1:1997)
EN IEC 60027-2:2019, Letter symbols to be used in electrical technology - Part 2: Telecommunications
and electronics (IEC 60027-2:2019)
EN 60027-3:2007, Letter symbols to be used in electrical technology - Part 3: Logarithmic and related
quantities, and their units (IEC 60027-3:2002)
EN 60027-4:2007, Letter symbols to be used in electrical technology - Part 4: Rotating electrical
machines (IEC 60027-4:2006)
EN 60027-6:2007, Letter symbols to be used in elecrical technology - Part 6: Control technology
(IEC 60027-6:2006)
EN 60027-7:2010, Letter symbols to be used in electrical technology - Part 7: Power generation,
transmission and distribution (IEC 60027-7:2010)
EN 61180:2016, High-voltage test techniques for low-voltage equipment - Definitions, test and
procedure requirements, test equipment (IEC 61180:2016)
2 Modifications to subclause 6.5.2.5.1.1
Replace the second paragraph with the following note:

NOTE 1 The manufacturer is supposed to take the expected life of the probe assembly into account when
selecting insulating materials.”
As a result of this, renumber the existing note as NOTE 2.
3 Modifications to 6.5.2.6, “Solid insulation for probe assemblies which are
not RATED for MEASUREMENT CATEGORIES”
Replace the second paragraph with the following note:

NOTE 1 The manufacturer is supposed to take the expected life of the probe assembly into account when
selecting insulating materials.”
As a result of this, renumber the existing note as NOTE 2.
4 Modifications to subclause 6.6.1, “General”
Replace the first sentence of the second paragraph with the following sentence:
“Test equipment for the voltage tests is specified in EN 61180:2016.”