Integrated circuits - Measurement of electromagnetic immunity - Part 8: Measurement of radiated immunity - IC stripline method

IEC 62132-8:2012 specifies a method for measuring the immunity of an integrated circuit (IC) to radio frequency (RF) radiated electromagnetic disturbances over the frequency range of 150 kHz to 3 GHz.

Integrierte Schaltungen - Messung der elektromagnetischen Störfestigkeit - Teil 8: Messung der Störfestigkeit bei Einstrahlungen - IC-Streifenleiterverfahren

Circuits intégrés - Mesure de l'immunité électromagnétique - Partie 8: Mesure de l'immunité rayonnée - Méthode de la ligne TEM à plaques pour circuit intégré

La CEI 62132-8:2012 définit une méthode de mesure de l'immunité d'un circuit intégré (CI) aux perturbations électromagnétiques rayonnées aux fréquences radioélectriques sur la gamme de fréquences comprise entre 150 kHz et 3 GHz.

Integrirana vezja - Meritve elektromagnetne odpornosti - 8. del: Meritev odpornosti proti sevanju - Metoda z IC na tračnem valovodu

Ta del standarda IEC 62132 določa metodo za merjenje imunosti integriranega vezja (IC) na sevane elektromagnetne motnje radijske frekvence (RF) čez frekvenčni razpon od 150 kHz do 3 GHz.

General Information

Status
Published
Publication Date
20-Sep-2012
Withdrawal Date
09-Aug-2015
Current Stage
6060 - Document made available - Publishing
Start Date
21-Sep-2012
Completion Date
21-Sep-2012

Relations

Standard
EN 62132-8:2012
English language
25 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)


SLOVENSKI STANDARD
01-december-2012
,QWHJULUDQDYH]MD0HULWYHHOHNWURPDJQHWQHRGSRUQRVWLGHO0HULWHYRGSRUQRVWL
SURWLVHYDQMX0HWRGD],&QDWUDþQHPYDORYRGX
Integrated circuits - Measurement of electromagnetic immunity - Part 8: Measurement of
radiated immunity - IC stripline method
Circuits intégrés - Mesure de l'immunité électromagnétique - Partie 8: Mesure de
l'immunité rayonnée - Méthode de la ligne à ruban pour circuit intégré
Ta slovenski standard je istoveten z: EN 62132-8:2012
ICS:
31.200 Integrirana vezja, Integrated circuits.
mikroelektronika Microelectronics
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

EUROPEAN STANDARD
EN 62132-8
NORME EUROPÉENNE
September 2012
EUROPÄISCHE NORM
ICS 31.200
English version
Integrated circuits -
Measurement of electromagnetic immunity -
Part 8: Measurement of radiated immunity -
IC stripline method
(IEC 62132-8:2012)
Circuits intégrés -  Integrierte Schaltungen -
Mesure de l'immunité électromagnétique - Messung der elektromagnetischen
Partie 8: Mesure de l'immunité rayonnée - Störfestigkeit -
Méthode de la ligne TEM à plaques pour Teil 8: Messung der Störfestigkeit bei
circuit intégré Einstrahlungen -
(CEI 62132-8:2012) IC-Streifenleiterverfahren
(IEC 62132-8:2012)
This European Standard was approved by CENELEC on 2012-08-10. CENELEC members are bound to comply
with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European Standard
the status of a national standard without any alteration.

Up-to-date lists and bibliographical references concerning such national standards may be obtained on
application to the CEN-CENELEC Management Centre or to any CENELEC member.

This European Standard exists in three official versions (English, French, German). A version in any other
language made by translation under the responsibility of a CENELEC member into its own language and notified
to the CEN-CENELEC Management Centre has the same status as the official versions.

CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus,
the Czech Republic, Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany,
Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the Netherlands, Norway, Poland,
Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and the United Kingdom.

CENELEC
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung

Management Centre: Avenue Marnix 17, B - 1000 Brussels

© 2012 CENELEC - All rights of exploitation in any form and by any means reserved worldwide for CENELEC members.
Ref. No. EN 62132-8:2012 E
Foreword
The text of document 47A/882/FDIS, future edition 1 of IEC 62132-8, prepared by SC 47A, "Integrated
circuits", of IEC TC 47, "Semiconductor devices" was submitted to the IEC-CENELEC parallel vote and
approved by CENELEC as EN 62132-8:2012.

The following dates are fixed:
(dop) 2013-05-10
• latest date by which the document has
to be implemented at national level by
publication of an identical national
standard or by endorsement
(dow) 2015-08-10
• latest date by which the national
standards conflicting with the
document have to be withdrawn
This standard is to be used in conjunction with EN 62132-1.

Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC [and/or CEN] shall not be held responsible for identifying any or all such patent
rights.
Endorsement notice
The text of the International Standard IEC 62132-8:2012 was approved by CENELEC as a European
Standard without any modification.

- 3 - EN 62132-8:2012
Annex ZA
(normative)
Normative references to international publications
with their corresponding European publications

The following documents, in whole or in part, are normatively referenced in this document and are
indispensable for its application. For dated references, only the edition cited applies. For undated
references, the latest edition of the referenced document (including any amendments) applies.

NOTE  When an international publication has been modified by common modifications, indicated by (mod), the relevant EN/HD
applies.
Publication Year Title EN/HD Year

IEC 60050 Series International Electrotechnical Vocabulary - -

IEC 61000-4-20 2010 Electromagnetic compatibility (EMC) - EN 61000-4-20 2010
Part 4-20: Testing and measurement
techniques - Emission and immunity testing in
transverse electromagnetic (TEM)
waveguides
IEC 62132-1 2006 Integrated circuits - Measurement of EN 62132-1 2006
electromagnetic immunity, + corr. November 2006
150 kHz to 1 GHz -
Part 1: General conditions and definitions

IEC 62132-8 ®
Edition 1.0 2012-07
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
Integrated circuits – Measurement of electromagnetic immunity –

Part 8: Measurement of radiated immunity – IC stripline method

Circuits intégrés – Mesure de l'immunité électromagnétique –

Partie 8: Mesure de l’immunité rayonnée – Méthode de la ligne TEM à plaques

pour circuit intégré
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
PRICE CODE
INTERNATIONALE
CODE PRIX T
ICS 31.200 ISBN 978-2-83220-206-7

– 2 – 62132-8  IEC:2012
CONTENTS
FOREWORD . 4
1 Scope . 6
2 Normative references . 6
3 Terms and definitions . 6
4 General . 7
5 Test conditions . 7
5.1 General . 7
5.2 Supply voltage. 8
5.3 Frequency range . 8
6 Test equipment . 8
6.1 General . 8
6.2 Cables . 8
6.3 Shielding . 8
6.4 RF disturbance generator . 8
6.5 IC stripline . 8
6.6 50 Ω termination . 8
6.7 DUT monitor . 8
7 Test setup . 9
7.1 General . 9
7.2 Test configuration . 9
7.3 EMC test board (PCB) . 9
8 Test procedure . 9
8.1 General . 9
8.2 Operational check . 10
8.3 Immunity measurement . 10
8.3.1 General . 10
8.3.2 RF disturbance signal . 10
8.3.3 Test frequency steps and ranges . 10
8.3.4 Test levels and dwell time . 10
8.3.5 DUT monitoring . 10
8.3.6 Detail procedure . 11
9 Test report. 11
10 RF immunity acceptance level . 11
Annex A (normative) Field strength determination . 12
Annex B (normative) IC stripline descriptions . 15
Annex C (informative) Closed stripline geometrical limitations . 18
Bibliography . 22

Figure 1 – IC stripline test setup . 9
Figure A.1 – Definition of height (h) and width (w) of IC stripline . 12
Figure A.2 – EM field distribution . 13
Figure B.1 – Cross section view of an example of an open IC stripline . 15
Figure B.2 – Cross section view of an example of a closed IC stripline . 16

62132-8  IEC:2012 – 3 –
Figure B.3 – Example of IC stripline with housing . 17
Figure C.1 – Calculated H-field reduction of closed version referenced to referring
open version as a function of portion of active conductor width of closed version to
open version . 20
Figure C.2 – Location of currents and mirrored currents at grounded planes used for
calculation of fields . 21

Table 1 – Frequency step size versus frequency range . 10
Table B.1 – Maximum DUT dimensions for 6,7 mm IC stripline (Open version) . 16
Table B.2 – Maximum DUT dimensions for 6,7 mm IC stripline (Closed version) . 16
Table C.1 – Height of shielding, simulated at h = 6,7mm to achieve practically
bottom
50 Ω system . 19

– 4 – 62132-8  IEC:2012
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
INTEGRATED CIRCUITS –
MEASUREMENT OF ELECTROMAGNETIC IMMUNITY –

Part 8: Measurement of radiated immunity –
IC stripline method
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International Standard IEC 62132-8 has been prepared by subcommittee 47A: Integrated
circuits, of IEC technical committee 47: Semiconductor devices.
The text of this standard is based on the following documents:
FDIS Report on voting
47A/882/FDIS 47A/886/RVD
Full information on the voting for the approval of this standard can be found in the report on
voting indicated in the above table.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
This part of IEC 62132 is to be read in conjunction with IEC 62132-1.

62132-8  IEC:2012 – 5 –
A list of all the parts in the IEC 62132 series, published under the general title Integrated
circuits – Measurement of electromagnetic immunity, can be found on the IEC website.
Future standards in this series will carry the new general title as cited above. Titles of existing
standards in this series will be updated at the time of the next edition.
The committee has decided that the contents of this publication will remain unchanged until
the stability date indicated on the IEC web site under "http://webstore.iec.ch" in the data
related to the specific publication. At this date, the publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
IMPORTANT – The 'colour inside' logo on the cover page of this publication indicates
that it contains colours which are considered to be useful for the correct
understanding of its contents. Users should therefore print this document using a
colour printer.
– 6 – 62132-8  IEC:2012
INTEGRATED CIRCUITS –
MEASUREMENT OF ELECTROMAGNETIC IMMUNITY –

Part 8: Measurement of radiated immunity –
IC stripline method
1 Scope
This part of IEC 62132 specifies a method for measuring the immunity of an integrated circuit
(IC) to radio frequency (RF) radiated electromagnetic disturbances over the frequency range
of 150 kHz to 3 GHz.
2 Normative references
The following documents, in whole or in part, are normatively referenced in this document and
are indispensable for its application. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any
amendments) applies.
IEC 60050 (all parts), International Electrotechnical Vocabulary (available at
http://www.electropedia.org)
IEC 62132-1:2006, Integrated circuits – Measurement of electromagnetic immunity, 150 kHz
to 1 GHz – Part 1: General conditions and definitions
IEC 61000-4-20, Electromagnetic compatibility (EMC) – Part 4-20: Testing and measurement
techniques – Emission and immunity testing in transverse electromagnetic (TEM) waveguides
3 Terms and definitions
For the purpose of this document, the terms and definitions given in IEC 62132-1:2006,
Clause 3, IEC 60050-131 and IEC 60050-161, and the following, apply.
3.1
transverse electromagnetic mode
TEM
waveguide mode in which the components of the electric and magnetic fields in the
propagation direction are much less than the primary field components across any transverse
cross-section
Note 1 to entry: This note only applies to the French language.
3.2
TEM waveguide
open or closed transmission line system, in which a wave is propagating in the transverse
electromagnetic mode to produce a specified field for testing purposes
3.3
IC stripline
TEM waveguide, consisting of an active conductor placed on a defined spacing over an
enlarged ground plane, connected to a port structure on each end and an optional shielded
enclosure
62132-8  IEC:2012 – 7 –
Note 1 to entry: This arrangement guides a wave propagation in the transverse electromagnetic mode to produce a
specific field for testing purposes between the active conductor and the enlarged ground plane. The ground plane
of the standard EMC test board according to IEC 62132-1:2006, Annex B, should be used. An optional shielding
enclosure may be used for fixing the IC stripline configuration and for shielding purposes. This leads to a closed
version of the IC stripline in opposite to the open version without shielding enclosure. For further information see
Annex A.
3.4
two-port TEM waveguide
TEM waveguide with input/output measurement ports at both ends
3.5
characteristic impedance
magnitude of the ratio of the voltage between the active conductor and the corresponding
ground plane to the current on either conductor for any constant phase wave-front
Note 1 to entry: The characteristic impedance is independent of the voltage/current magnitudes and depends only
on the cross sectional geometry of the transmission line. TEM waveguides are typically designed to have a 50 Ω
characteristic impedance. For further information and equation to stripline arrangements see Annex A.
3.6
primary field component
primary component
electric field component aligned with the intended test polarization
Note 1 to entry: For example, in IC stripline, the active conductor is parallel to the horizontal floor, and the primary
mode electric field vector is vertical at the transverse centre of the IC stripline.
4 General
An IC to be evaluated for EMC performance is referred to as a device under test (DUT). The
DUT should be mounted on an EMC test board according to IEC 62132-1. The EMC test
board is provided with the appropriate measurement or monitoring points at which the DUT
response parameters can be measured. It controls the geometry and orientation of the DUT
relative to the active conductor and eliminates in the case of a closed version of the IC
stripline any connecting leads within the housing (these are on the backside of the board,
which is outside the housing).
For the IC stripline, one of the 50 Ω ports is terminated with a 50 Ω load. The other 50 Ω port
is connected to the output of an RF disturbance generator. The injected RF disturbance signal
exposes the DUT to an electromagnetic field determined by the injected power, the typical
impedance and the distance between the ground plane of the EMC test board and the active
conductor of the IC stripline. The relationship is given in Annex A.
Rotating the EMC test board in the four possible orientations in the aperture to accept EMC
test board of the IC stripline will affect the sensitivity of the DUT. Dependent upon the DUT,
the response parameters of the DUT may vary (e.g. a change of current consumption,
deterioration in function performance, waveform jitter). The intent of this test method is to
provide a quantitative measure of the RF immunity of DUTs for comparison or other purposes.
For further information see Annex A.
5 Test conditions
5.1 General
The test conditions shall meet the requirements as described in IEC 62132-1:2006, Clause 4.
In addition, the following test conditions shall apply.

– 8 – 62132-8  IEC:2012
5.2 Supply voltage
The supply voltage shall be as specified by the IC manufacturer. If the users of this procedure
agree to other values, they shall be documented in the test report.
5.3 Frequency range
The effective frequency range of this radiated immunity procedure is 150 kHz to 3 GHz.
6 Test equipment
6.1 General
The test equipment shall meet the requirements described in IEC 62132-1:2006, Clause 5. In
addition, the following test equipment requirements shall apply.
6.2 Cables
Double shielded or semi-rigid coaxial cable may be required depending on the local ambient
conditions.
6.3 Shielding
Testing in a shielded room is only necessary for the open IC stripline version. The closed
version of the IC stripline is shielded by its housing.
6.4 RF disturbance generator
An RF disturbance generator with sufficient power handling capabilities shall be used. The RF
disturbance generator may comprise an RF signal source with a modulation function, an RF
power amplifier. The voltage standing wave ratio (VSWR) at the output of the RF disturbance
generator shall be less than 1,5 over the frequency range being measured.
The gain (or attenuation) of the RF disturbance generating equipment, without the IC stripline,
shall be known with an accuracy ±0,5 dB.
6.5 IC stripline
The IC stripline (open or closed version) used for this test procedure shall be fitted with an
aperture to mate with the EMC test board. The IC stripline shall not exhibit higher order
modes over the frequency range being measured. For this procedure, the IC stri
...

Questions, Comments and Discussion