Environmental testing - Part 3-11: Supporting documentation and guidance - Calculation of uncertainty of conditions in climatic test chambers

Demonstrates how to estimate the uncertainty of steady-state temperature and humidity conditions in temperature and humidity chambers. Since this is inextricably linked to the methods of measurement, these are also described. This standard is equally applicable to all environmental enclosures, including rooms or laboratories. The methods used apply both to temperature chambers and combined temperature and humidity chambers. This standard is meant to help everyone using climatic test chambers. Those already familiar with uncertainty of measurement will find it useful for guidance on typical sources of uncertainty and how they should be quantified and combined. It is also intended to assist the first-time or occasional user who has little or no knowledge of the subject. To discuss uncertainty, it is important first to understand what is being measured or characterized. The calibration or characterization of the performance of a chamber is concerned with the humidity and temperature of the air in the chamber, as experienced by the item under test, at a given set point. This should not be confused with characterizing or calibrating the chamber sensor, which is a separate matter.

Umgebungseinflüsse - Teil 3-11: Unterstützende Dokumentation und Leitfaden - Berechnung der Messunsicherheit von Umgebungsbedingungen in Klimaprüfkammern

Essais d'environnement - Partie 3-11: Documentation d’accompagnement et guide - Calcul de l’incertitude des conditions en chambres d’essais climatiques

Montre comment estimer l'incertitude des conditions de température et d'humidité à l'état stable dans une chambre d'essais de température et d'humidité. Sachant que cela est inextricablement lié aux méthodes de mesures, celles-ci sont également décrites. La présente norme est également applicable à tout type d'enceintes d'environnement, et même les salles ou laboratoires. Les méthodes sont usuellement appliquées aux chambres d'essais régulées en température et aux chambres d'essais régulées en température et humidité. La présente norme cherche à aider toute personne utilisant des chambres d'essais climatiques. Les personnes étant déjà familières avec l'incertitude de mesure la trouveront utile pour les conseils sur les composantes types d'incertitudes et sur la manière dont il convient de les quantifier et de les combiner. Elle est également destinée à aider le nouvel utilisateur ou l'utilisateur occasionnel qui n'a qu'une faible connaissance du sujet ou pas de connaissance du tout. Afin d'aborder les incertitudes il est important tout d'abord de comprendre ce qui est mesuré ou caractérisé. L'étalonnage ou la caractérisation de la performance d'une chambre d'essais fait appel à la détermination de l'humidité et de la température de l'air de la chambre, vu par l'objet en essai, à une valeur de consigne donnée. Il convient que cela ne soit pas confondu avec le calibrage ou l'étalonnage du capteur de la chambre d'essai, qui est une opération différente.

Okoljski preskusi - 3-11.del: Spremljajoča dokumentacija in vodila - Izračunavanje nezanesljivosti pogojev v komorah za klimatsko preskušanje (IEC 60068-3-11:2007)

General Information

Status
Published
Publication Date
27-Jun-2007
Withdrawal Date
31-May-2010
Current Stage
6060 - Document made available - Publishing
Start Date
28-Jun-2007
Completion Date
28-Jun-2007

Relations

Overview

EN 60068-3-11:2007 (CLC) - Environmental testing - Part 3-11 provides supporting documentation and guidance for the calculation of uncertainty of steady‑state temperature and humidity conditions in climatic test chambers. It explains what is being measured (the air conditions experienced by the item under test, not the chamber sensor itself), describes measurement methods, and gives guidance on identifying, quantifying and combining uncertainty contributions. The guidance is applicable to temperature chambers, combined temperature‑humidity chambers and other environmental enclosures (including rooms and laboratories).

Key topics

  • Concept of uncertainty: definitions, relationship to error and “true value”, and how to report uncertainty.
  • Measurement methods: recommended approaches for temperature and relative humidity sensing and when to characterise an empty chamber versus a loaded chamber.
  • Sources of uncertainty: typical contributions such as sensor accuracy, spatial gradients, temporal fluctuations, calibration, mounting and measurement technique.
  • Combining uncertainties: methods for quantifying individual components and combining them (combined standard uncertainty and expanded uncertainty).
  • Analysis and presentation: average‑case and worst‑case analyses, dealing with anomalous data, and documenting results.
  • Worked examples and data: illustrative measurement data sets and step‑by‑step combination of uncertainty components (contained in annexes).

Practical applications

This standard helps organizations and individuals who need to determine whether climatic test conditions meet specified tolerances and to document measurement confidence. Typical uses:

  • Laboratory engineers validating climatic test chamber performance.
  • Calibration and metrology teams preparing uncertainty budgets for temperature and humidity measurements.
  • Test managers and quality personnel assessing compliance with environmental test specifications.
  • Designers of environmental tests who need to report measurement uncertainty in test reports.

Benefits include clearer traceability of chamber characterization, consistent uncertainty reporting, and improved confidence when declaring chamber conformance to test limits.

Who should use this standard

  • Environmental test technicians and engineers
  • Calibration laboratories and calibration authorities
  • Quality managers and compliance officers in industries performing environmental testing (electronics, automotive, aerospace, medical devices)
  • First‑time or occasional users needing practical guidance on uncertainty of measurement

Related standards

  • IEC/EN 60068 series (environmental testing)
  • EN 60068-3-5 - Confirmation of performance of temperature chambers
  • EN 60068-3-6 - Confirmation of performance of temperature/humidity chambers
  • ISO Guide to the Expression of Uncertainty in Measurement (GUM) and ISO 3534 statistical vocabularies

Keywords: environmental testing, climatic test chambers, uncertainty of measurement, temperature chambers, humidity chambers, measurement uncertainty, chamber characterization, calibration.

Standard
EN 60068-3-11:2007
English language
37 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Frequently Asked Questions

EN 60068-3-11:2007 is a standard published by CLC. Its full title is "Environmental testing - Part 3-11: Supporting documentation and guidance - Calculation of uncertainty of conditions in climatic test chambers". This standard covers: Demonstrates how to estimate the uncertainty of steady-state temperature and humidity conditions in temperature and humidity chambers. Since this is inextricably linked to the methods of measurement, these are also described. This standard is equally applicable to all environmental enclosures, including rooms or laboratories. The methods used apply both to temperature chambers and combined temperature and humidity chambers. This standard is meant to help everyone using climatic test chambers. Those already familiar with uncertainty of measurement will find it useful for guidance on typical sources of uncertainty and how they should be quantified and combined. It is also intended to assist the first-time or occasional user who has little or no knowledge of the subject. To discuss uncertainty, it is important first to understand what is being measured or characterized. The calibration or characterization of the performance of a chamber is concerned with the humidity and temperature of the air in the chamber, as experienced by the item under test, at a given set point. This should not be confused with characterizing or calibrating the chamber sensor, which is a separate matter.

Demonstrates how to estimate the uncertainty of steady-state temperature and humidity conditions in temperature and humidity chambers. Since this is inextricably linked to the methods of measurement, these are also described. This standard is equally applicable to all environmental enclosures, including rooms or laboratories. The methods used apply both to temperature chambers and combined temperature and humidity chambers. This standard is meant to help everyone using climatic test chambers. Those already familiar with uncertainty of measurement will find it useful for guidance on typical sources of uncertainty and how they should be quantified and combined. It is also intended to assist the first-time or occasional user who has little or no knowledge of the subject. To discuss uncertainty, it is important first to understand what is being measured or characterized. The calibration or characterization of the performance of a chamber is concerned with the humidity and temperature of the air in the chamber, as experienced by the item under test, at a given set point. This should not be confused with characterizing or calibrating the chamber sensor, which is a separate matter.

EN 60068-3-11:2007 is classified under the following ICS (International Classification for Standards) categories: 19.040 - Environmental testing. The ICS classification helps identify the subject area and facilitates finding related standards.

EN 60068-3-11:2007 has the following relationships with other standards: It is inter standard links to prEN IEC 60068-3-11:2024. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

You can purchase EN 60068-3-11:2007 directly from iTeh Standards. The document is available in PDF format and is delivered instantly after payment. Add the standard to your cart and complete the secure checkout process. iTeh Standards is an authorized distributor of CLC standards.

Standards Content (Sample)


SLOVENSKI STANDARD
01-oktober-2007
2NROMVNLSUHVNXVLGHO6SUHPOMDMRþDGRNXPHQWDFLMDLQYRGLOD,]UDþXQDYDQMH
QH]DQHVOMLYRVWLSRJRMHYYNRPRUDK]DNOLPDWVNRSUHVNXãDQMH ,(&

Environmental testing -- Part 3-11: Supporting documentation and guidance - Calculation
of uncertainty of conditions in climatic test chambers
Umgebungseinflüsse - Teil 3-11: Unterstützende Dokumentation und Leitfaden -
Berechnung der Messunsicherheit von Umgebungsbedingungen in Klimaprüfkammern
Essais d'environnement - Partie 3-11: Documentation d’accompagnement et guide -
Calcul de l’incertitude des conditions en chambres d’essais climatiques
Ta slovenski standard je istoveten z: EN 60068-3-11:2007
ICS:
19.040 Preskušanje v zvezi z Environmental testing
okoljem
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

EUROPEAN STANDARD
EN 60068-3-11
NORME EUROPÉENNE
June 2007
EUROPÄISCHE NORM
ICS 19.040
English version
Environmental testing -
Part 3-11: Supporting documentation and guidance -
Calculation of uncertainty of conditions in climatic test chambers
(IEC 60068-3-11:2007)
Essais d'environnement -  Umgebungseinflüsse -
Partie 3-11: Documentation Teil 3-11: Unterstützende Dokumentation
d’accompagnement et guide - und Leitfaden -
Calcul de l’incertitude des conditions Berechnung der Messunsicherheit
en chambres d’essais climatiques von Umgebungsbedingungen
(CEI 60068-3-11:2007) in Klimaprüfkammern
(IEC 60068-3-11:2007)
This European Standard was approved by CENELEC on 2007-06-01. CENELEC members are bound to comply
with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European Standard
the status of a national standard without any alteration.

Up-to-date lists and bibliographical references concerning such national standards may be obtained on
application to the Central Secretariat or to any CENELEC member.

This European Standard exists in three official versions (English, French, German). A version in any other
language made by translation under the responsibility of a CENELEC member into its own language and notified
to the Central Secretariat has the same status as the official versions.

CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Cyprus, the
Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia,
Lithuania, Luxembourg, Malta, the Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain,
Sweden, Switzerland and the United Kingdom.

CENELEC
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung

Central Secretariat: rue de Stassart 35, B - 1050 Brussels

© 2007 CENELEC - All rights of exploitation in any form and by any means reserved worldwide for CENELEC members.
Ref. No. EN 60068-3-11:2007 E
Foreword
The text of document 104/409/FDIS, future edition 1 of IEC 60068-3-11, prepared by IEC TC 104,
Environmental conditions, classification and methods of test, was submitted to the IEC-CENELEC parallel
vote and was approved by CENELEC as EN 60068-3-11 on 2007-06-01.
The following dates were fixed:
– latest date by which the EN has to be implemented
at national level by publication of an identical
national standard or by endorsement (dop) 2008-03-01
– latest date by which the national standards conflicting
with the EN have to be withdrawn (dow) 2010-06-01
Annex ZA has been added by CENELEC.
__________
Endorsement notice
The text of the International Standard IEC 60068-3-11:2007 was approved by CENELEC as a European
Standard without any modification.
__________
- 3 - EN 60068-3-11:2007
Annex ZA
(normative)
Normative references to international publications
with their corresponding European publications

The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.

NOTE  When an international publication has been modified by common modifications, indicated by (mod), the relevant EN/HD
applies.
Publication Year Title EN/HD Year
1) 2)
IEC 60068-3-5 - Environmental testing - EN 60068-3-5 2002
Part 3-5: Supporting documentation and
guidance - Confirmation of the performance
of temperature chambers
1) 2)
IEC 60068-3-6 - Environmental testing - EN 60068-3-6 2002
Part 3-6: Supporting documentation and
guidance - Confirmation of the performance
of temperature/humidity chambers

ISO 3534-1 2006 Statistics - Vocabulary and symbols - - -
Part 1: General statistical terms and terms
used in probability
ISO 3534-2 2006 Statistics - Vocabulary and symbols - - -
Part 2: Applied statistics
ISO Guide 99 1996 International vocabulary of basic and general - -
terms in metrology
ISO/IEC Guide 98 1995 Guide to the expression of uncertainty in - -
measurement
1)
Undated reference.
2)
Valid edition at date of issue.

INTERNATIONAL IEC
STANDARD
CEI
60068-3-11
NORME
First edition
INTERNATIONALE
Première édition
2007-05
Environmental testing –
Part 3-11:
Supporting documentation and guidance –
Calculation of uncertainty of conditions
in climatic test chambers
Essais d’environnement –
Partie 3-11:
Documentation d’accompagnement et guide –
Calcul de l’incertitude des conditions
en chambres d’essais climatiques
PRICE CODE
V
CODE PRIX
Commission Electrotechnique Internationale
International Electrotechnical Commission
МеждународнаяЭлектротехническаяКомиссия
For price, see current catalogue
Pour prix, voir catalogue en vigueur

– 2 – 60068-3-11 © IEC:2007
CONTENTS
FOREWORD.4
INTRODUCTION.6

1 Scope.7
2 Normative references .7
3 Terms and definitions .8
4 Concept of uncertainty.11
4.1 Uncertainty, error and “true value”.11
4.2 Statements of uncertainty.12
4.3 Combining uncertainties .13
5 Tolerance .13
6 Humidity and temperature measurement .13
7 Methods for determining climatic test chamber uncertainties .14
7.1 Empty chamber .16
7.2 Typical load.16
7.3 Measurement of conditions in the chamber during the test .17
7.4 Conditions to measure.17
7.5 Measurements required.18
7.6 Sources of uncertainty.19
7.7 Essential contributions of uncertainty .20
8 Estimation of uncertainty components and their combination .24
9 Overall uncertainty of temperature measurement.24
9.1 General .24
9.2 Further considerations.26
10 Overall uncertainty of relative humidity measurement .26
10.1 Uncertainty of temperature measurement at each sensor point.27
10.2 Uncertainty of the relative humidity measurement.27
11 Anomalous data and presentation of results .30
11.1 Average case analysis.30
11.2 Worst case analysis .30

Annex A (informative) Measurement data sets – Loaded chamber .32

Bibliography.34

60068-3-11 © IEC:2007 – 3 –
Figure 1 – Approaches to calibration method and uncertainty calculation.15
Figure 2 – Illustration of the fluctuation of a temperature sensor .23

Table 1 – Combination of temperature uncertainties .24
Table 2 – Combination of temperature uncertainties at each point .27
Table 3 – Combination of humidity uncertainties .28
Table A.1 – Typical temperature measurement data set and it’s analysis and refs .32
Table A.2 – Humidity measurements analysis based on Table A.1 temperatures.33

– 4 – 60068-3-11 © IEC:2007
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
ENVIRONMENTAL TESTING –
Part 3-11: Supporting documentation and guidance –
Calculation of uncertainty of conditions in climatic test chambers

FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC provides no marking procedure to indicate its approval and cannot be rendered responsible for any
equipment declared to be in conformity with an IEC Publication.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International Standard IEC 60068-3-11 has been prepared by IEC technical committee 104:
Environmental conditions, classification and methods of test.
The text of this standard is based on the following documents:
FDIS Report on voting
104/409/FDIS 104/415/RVD
Full information on the voting for the approval of this standard can be found in the report on
voting indicated in the above table.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.

60068-3-11 © IEC:2007 – 5 –
A list of all parts in the IEC 60068 series, under the general title Environmental testing can be
found on the IEC website.
The committee has decided that the contents of this publication will remain unchanged until
the maintenance result date indicated on the IEC web site under "http://webstore.iec.ch" in
the data related to the specific publication. At this date, the publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
– 6 – 60068-3-11 © IEC:2007
INTRODUCTION
This part of IEC 60068 provides guidance for analysing uncertainties of temperature and
humidity in climatic test chambers. It has been written for technicians, engineers and
managers in environmental testing, and for anyone who needs to understand the results of
environmental tests.
The performance of climatic test chambers is a key concern in environmental test engineer-
ing. To comply with any test specification, the performance of the chamber needs to be
characterized to decide whether the generated conditions fall within the specified limits. This
characterization can be a difficult task, and the analysis of uncertainties in chamber
performance is often surrounded by confusion. This publication is intended to ease that
process.
In what follows, the concept of uncertainty of measurement is introduced first and then the
significance of tolerance discussed. Aspects of humidity and temperature measurement are
considered, followed by methods for determining and combining uncertainties. The cases of
both calibrating an empty chamber and of measuring conditions in a loaded chamber are
considered. Finally, detailed guidance and worked examples are given for analysing results to
give estimates of uncertainty in the measured performance.

60068-3-11 © IEC:2007 – 7 –
ENVIRONMENTAL TESTING –
Part 3-11: Supporting documentation and guidance –
Calculation of uncertainty of conditions in climatic test chambers

1 Scope
This part of IEC 60068 demonstrates how to estimate the uncertainty of steady-state
temperature and humidity conditions in temperature and humidity chambers. Since this is
inextricably linked to the methods of measurement, these are also described.
This standard is equally applicable to all environmental enclosures, including rooms or
laboratories. The methods used apply both to temperature chambers and combined
temperature and humidity chambers.
This standard is meant to help everyone using climatic test chambers. Those already familiar
with uncertainty of measurement will find it useful for guidance on typical sources of
uncertainty and how they should be quantified and combined. It is also intended to assist the
first-time or occasional user who has little or no knowledge of the subject.
To discuss uncertainty, it is important first to understand what is being measured or
characterized. The calibration or characterization of the performance of a chamber is
concerned with the humidity and temperature of the air in the chamber, as experienced by the
item under test, at a given set point. This should not be confused with characterizing or
calibrating the chamber sensor, which is a separate matter.
2 Normative references
The following referenced documents are indispensable for the application of this document.
For dated references, only the edition cited applies. For undated references, the latest edition
of the referenced document (including any amendments) applies.
IEC 60068-3-5: Environmental testing – Part 3-5: Supporting documentation and guidance –
Confirmation of the performance of temperature chambers
IEC 60068-3-6: Environmental testing – Part 3-6: Supporting documentation and guidance –
Confirmation of the performance of temperature/humidity chambers
ISO 3534-1:2006, Statistics – Vocabulary and symbols – Part 1: General statistical terms and
terms used in probability
ISO 3534-2:2006, Statistics – Vocabulary and symbols – Part 2:Applied statistics
International Vocabulary of basic and general standard terms in metrology. ISO, Geneva,
Switzerland 1993 (ISBN 92-67-10175-1) – VIM

– 8 – 60068-3-11 © IEC:2007
Guide to the expression of uncertainty in measurement. ISO, Geneva, Switzerland 1993.
(ISBN 92-67-10188-9) – GUM
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
3.1
calibration authority
laboratory or other organization that performs calibrations and is itself accredited by the
appropriate national accreditation body
3.2
climatic test chamber
enclosure
chamber or enclosed space where the internal temperature or temperature and humidity can
be controlled within specified limits
3.3
combined standard uncertainty
standard uncertainty of the result of a measurement when that result is obtained from the
values of a number of other quantities, equal to the positive square root of a sum of terms, the
terms being the variances or covariances of these other quantities weighted according to how
the measurement result varies with changes in these quantities
See also GUM.
3.4
correction
value added algebraically to the result of a measurement to compensate determinable
systematic error
See also VIM.
3.5
confidence level
value of probability associated with a confidence interval
NOTE The confidence level is the likelihood that the “true value” lies within the stated range of uncertainty usually
expressed as a percentage, e.g. 95 %.
See also ISO 3534-1.
3.6
coverage factor
numerical factor to multiply the combined standard uncertainty to obtain an expanded
uncertainty
NOTE A coverage factor of k=2 corresponds to a confidence level of approximately 95 % if normally distributed
and if the number of degrees of freedom is sufficiently large.
See also GUM.
60068-3-11 © IEC:2007 – 9 –
3.7
dew point
temperature at which the partial pressure of the water vapour is equal to the saturation vapour
pressure over water or ice
NOTE The temperature to which the air would have to cool (at constant pressure and constant water vapour
content) in order to reach saturation. A state of saturation exists when the air is holding the maximum amount of
water vapour possible at the existing temperature and pressure.
3.8
dispersion
spread of repeated measurements of a quantity
3.9
drift
change in the indication of a measuring system not related to a change in the quantity being
measured
See also VIM.
NOTE The drift since the last calibration can be estimated and a correction applied to measured values.
3.10
error
difference between result of a measurement and the true value
3.11
expanded uncertainty
quantity defining an interval about the result of a measurement that may be expected to
encompass a large fraction of the distribution of values that could reasonably be attributed to
the measurand
See also VIM.
3.12
fluctuation
change (from the mean) in the temperature or humidity after stabilization from time to time at
a point in space
NOTE It may be measured by standard deviation or maximum deviation.
3.13
gradient
maximum difference in mean value, after stabilization, at any moment in time between two
separate points in the working space
3.14
incident air
conditioned airstream which flows into the working space
3.15
partial vapour pressure
contribution of water vapour in a given volume of air at a constant pressure and temperature
of the atmosphere
3.16
reference instrument
previously calibrated instrument used to measure the conditions within the enclosure

– 10 – 60068-3-11 © IEC:2007
3.17
relative humidity
ratio of actual partial vapour pressure to the saturation vapour pressure at any given
temperature and pressure, expressed as a percentage (% RH)
3.18
repeatability
closeness of agreement between independent results obtained in the normal and correct
operation of the same method on identical test material, in a short space of time, and under
the same test conditions (such as the same operator, same apparatus, same laboratory)
3.19
resolution
smallest changes between indications of the chamber controller display that can be
meaningfully distinguished
3.20
saturation vapour pressure
when a given volume of air, at a constant temperature, has water vapour present and is
incapable of holding more water vapour it is said to be saturated
3.21
stabilization
achievement of the state of temperature/humidity in the chamber when all mean values in the
working space are constant and have maintained temperature/humidity within a given
tolerance
3.22
standard deviation
measure of the dispersion of a set of measurements
NOTE The standard deviation, s, is the best estimate of sigma (the population standard deviation).
See also GUM and/or VIM.
3.23
standard uncertainty
uncertainty of the result of a measurement expressed as a standard deviation
See also GUM.
3.24
tolerance
acceptance limit specified or chosen for a process or product
See also ISO 3534-2.
3.25
traceability
property of the result of a measurement or the value of a standard whereby it can be related
to stated references, usually national or international standards, through an unbroken chain of
comparisons, all having stated uncertainties
NOTE The unbroken chain of comparisons is called a traceability chain.
See also ISO 3534-1 and VIM.
60068-3-11 © IEC:2007 – 11 –
3.26
true value
value which characterizes a quantity, perfectly defined in the conditions which exist when that
quantity is considered
NOTE The true value of a quantity is a theoretical concept and, in general, cannot be known exactly but is
estimated by measurement.
See also ISO 3534-1.
3.27
uncertainty
parameter, associated with the result of a measurement, which characterizes the dispersion
of the values that could reasonably be attributed to it
3.28
uncertainty budget
list of sources of uncertainty compiled with a view to evaluating a combined standard
uncertainty associated with a measurement result
3.29
uncertainty contribution
input to an uncertainty budget
3.30
working space
part of the chamber in which the specified conditions can be maintained within the specified
tolerances
4 Concept of uncertainty
4.1 Uncertainty, error and “true value”
In every measurement – no matter how careful – there is always a margin of doubt about the
result. In simple terms, the uncertainty of a measurement is a quantification of the doubt
about the measurement result.
While discussing uncertainty we often also need to consider a related but separate concept,
“error”. A measurement “error” is the difference between the measured value and the “true
value” of the thing being measured.
The “true value” of any quantity is in principle unknowable. This leads to a problem since the
“error” is defined as the result of a measurement minus the “true value”. Sometimes this
difference can be estimated. Both terms are best avoided as much as possible and, when
necessary, should be used with care. Discussion of “error analysis”, which used to be
included in many scientific papers, should have been entitled “analysis of the probable limits
of error”, or more properly, ”analysis of uncertainty”. In older publications the term “error” was
widely used when ‘uncertainty’ would have been the correct term.
Uncertainty is not the same as error. If the conditions in a test chamber are measured with a
calibrated instrument and the result is 75 % RH when the chamber controller says 90 % RH,
that does not mean the uncertainty is 15 % RH. It is known that the relative humidity is 75 %
RH. One is aware that either the controller reading is wrong or the chamber is operating
incorrectly. It has an error estimated to be 15 % RH. The uncertainty is a characteristic of the
measurement that gave the answer 75 % RH. Could that be wrong and, if so, by how much?

– 12 – 60068-3-11 © IEC:2007
When considering “true value”, uncertainty, and error, one of the most important sources of
this type of information for a measuring instrument is its calibration certificate. It is vital to use
all of the information provided by the calibration certificate to ensure that the best estimate of
the test uncertainties are obtained.
4.2 Statements of uncertainty
4.2.1 General
When reporting the results of a measurement, three numbers are necessary for a
metrologically correct and complete statement of the result of each measurement point. For
example, the complete statement could be:
The “true value” is: 39,1 °C ± 0,3 K with 95 % confidence:
• 39,1 °C is the best estimate of the true value;
• ±0,3 K is the confidence interval;
• 95 % is the confidence level.
An explanation of these three components follows.
4.2.2 Best estimate of the true value of the measured quantity
Often this will simply be the reading on the calibrated reference instrument which, in the case
of a climatic test, could be the temperature measurement system and/or hygrometer reading,
or if the chamber has been calibrated it could be the chamber controller display. If the
calibration shows either for an instrument or for a chamber controller that an error exists
(which is not an uncertainty), this should be used to apply a correction. For example, if the
calibration of a thermometer shows that it reads 1 K high, 1 K should be subtracted from the
reading to obtain the best estimate of the true value.
4.2.3 Confidence interval
This is the range of measured values within which the “true value” lies with a given level of
confidence. In our example this interval is ±0,3 K.
4.2.4 Confidence level
The “confidence level” of a measurement is a number (e.g. 95 %) expressing the degree of
confidence in the result. This is the probability that the real “true value” lies in the given
range. Most sets of data are normally distributed and about 68 % of the values will fall within
plus or minus one standard deviation of the mean. About 95 % of the values can be expected
to fall within plus or minus 2 standard deviations (95 % confidence level). Put another way,
when many such measurements are performed not more than 1 in 20 will lie outside the
stated limits. Hence multiplying the standard deviation by 2 is an accepted way of
encompassing 95 % of the range of values. With a 95 % confidence level, we are 95 % sure
that the “true value” lies in the stated range.
It is conventional to work at the 95 % confidence level. Higher confidence levels can be used
but the confidence interval will increase.
4.2.5 Statement of uncertainty
In the above example the statement of uncertainty is that the temperature was 39,1 °C± 0,3 K
with 95 % confidence. 39,1 °C was the best estimate of the temperature but because of the
uncertainties there is a possibility of it being in the range 38,8 °C to 39,4 °C with a confidence
of 95 %.
60068-3-11 © IEC:2007 – 13 –
4.3 Combining uncertainties
Uncertainty contributions shall be expressed in similar terms before they are combined. They
shall be in the same units and at the same level of confidence. All contributions should be
converted into standard uncertainties (i.e. having a confidence level of plus or minus one
standard deviation). This is discussed further in Clauses 9 and 10.
5 Tolerance
When a test item is to be conditioned one of the first questions asked is, “Will the chamber
achieve and maintain the required conditions?” This is asked since the test specification will
often set a tolerance for the required condition e.g. ±2 °C and ±5 % RH. In deciding whether a
tolerance is met, the uncertainty in the measured chamber performance shall be taken into
account.
Tolerances are not the same as uncertainties. Tolerances are acceptance limits which are
chosen for a process or product. Most often the aim of knowing the uncertainty in a chamber’s
performance is to decide whether a tolerance is met. In deciding this, the deviation from the
required condition, together with the uncertainty, shall be considered. Using the values cited
in 4.2.5, it is certain to within 95 % that the true temperature is between 38,8 °C and 39,4 °C.
If the required condition is 40 °C ± 2 K, then the probability that the true temperature lies
within the tolerance is considerably better than 95 % because the entire confidence interval
lies within the range of the tolerance.
If the measured humidity is 81,7 % RH, and the confidence interval is ±3,6 % RH at a 95 %
confidence level, then it is certain to within 95 % that the true humidity is between 78,1 % RH
and 85,3 % RH. If the required condition is 85 ± 5 % RH, even though the measured condition
is within this range, the probability that the true humidity is within ±5 % RH of the set point is
significantly less at a 95 % confidence level, because the entire confidence interval does not
lie within the range of the tolerance. However, from the uncertainty, there are statistical
methods for making a good estimate of how likely this is.
6 Humidity and temperature measurement
When taking humidity measurements there are many ways of approaching the situation. It is
generally assumed that the water vapour content of the air is uniform throughout the chamber.
This is a reasonable assumption, and people who have performed measurements of humidity
at multiple points in a chamber can confirm that this is normally the case. However, this does
not mean that the relative humidity is uniform.
Dew point, being directly related to vapour pressure, can be assumed to be uniform across
the chamber and is not affected by temperature. It may be that during routine tests, humidity
measurement is only made in one place. However, at some point, either during the test or
when the chamber is operating under similar conditions, humidity measurements shall be
made in at least two places so that an uncertainty can be assigned to the assumption that the
vapour content of the air is uniform.
For most environmental tests, the required humidity is specified in terms of relative humidity.
The importance of relative humidity arises because the behaviour of most organic materials
depends on this parameter. Factors such as physical expansion of plastics and wood,
biological activity, electrical impedance and corrosion rates are examples of processes that
are affected by the relative humidity.
In a chamber the vapour pressure is often nearly uniform.

– 14 – 60068-3-11 © IEC:2007
Even when the air is thoroughly stirred there are often temperature differences from place to
place in chambers and, although the water vapour pressure is often nearly uniform, the
temperature differences cause differences in the relative humidity. A single humidity
measurement at only one location is often sufficient to tell us about the vapour pressure in the
rest of the chamber. The single measurement should be made at a central point or on the
incident air side of the object under test.
The measurement can be made with any hygrometer, but normally it will be one of three
types:
− a dew-point (dp) hygrometer (mirror condensation);
− a psychrometer (wet/dry); or
− a relative humidity probe.
Examples are shown in Annex A.
7 Methods for determining climatic test chamber uncertainties
There are three basic methods for determining conditions in a climatic chamber. These three
methods reflect the different requirements in different types of testing and there are good
reasons for each approach. These methods are illustrated in Figure 1.

60068-3-11 IEC:2007                 – 15 –
©
1 2
Chamber status
Empty chamber Chamber typically Conditions measured
loaded at time of test
Humidity sensor
Multipoint Multipoint
Multipoint positioned
positioning
predefined positions positioned around the
around typical load
load
Single point. e.g. Single point. e.g. Multipoint
Single point. e.g. Multipoint around Multipoint
centrally located or the working space centrally located positioned around centrally located positioned around
Temperature sensor
adjacent to control on the incident air the load on the incident air the load
positioning
sensor side side
The suitability of the None
The effect of the load
load must be
Loading considerations must be considered
considered
Chamber repeatability Chamber repeatability Not applicable
Facility repeatability
Combine to show worst Combine to show
Combination
case condition averaged condition
techniques
Calibration equipment Uncertainty calculation
uncertainty
Uncertainty of
conditions
IEC  670/07
Figure 1 – Approaches to calibration method and uncertainty calculation

– 16 – 60068-3-11 © IEC:2007
7.1 Empty chamber
7.1.1 Advantages
Advantages are as follows:
a) The entire working space is calibrated.
b) Calibration need be carried out only once or twice a year.
c) Re-calibration is not required when the test sample is changed.
d) The suitability of the chamber can be assessed without subjecting the test sample to
conditioning.
e) Relatively low cost. Only one set of calibrated instruments required for many chambers.
7.1.2 Disadvantages
Disadvantages include:
a) The effect of the test sample is difficult to quantify, although it may be negligible for
samples that are very small compared with the chamber. It is very difficult to assign an
uncertainty to the effect of the load.
b) The effect of heat-dissipating test samples is very hard to quantify.
c) Drift, resolution, and repeatability of the chamber controller shall be assessed and their
contributions to the uncertainty calculations shall be included.
7.2 Typical load
Calibration of the chamber with a typical load is ideal where very similar tests are repeated.
7.2.1 Advantages
Advantages are as follows:
a) The affect of the load on the control of the chamber can be accurately assessed without
subjecting the test sample to an unknown stress.
b) The smallest suitable chamber that produces satisfactory conditions can be chosen
prior to test.
c) Careful positioning of the sensors can give detailed information about critical parts of
load.
d) Anomalies from dissipating loads can be quantified.
e) Relatively low cost. Only one set of calibrated instruments required for many chambers.
7.2.2 Disadvantages
Disadvantages include:
a) Re-calibration is required when the test sample is changed significantly.
b) Drift, resolution and repeatability of the chamber controller shall be assessed and their
contributions to the uncertainty calculations shall be included.

60068-3-11 © IEC:2007 – 17 –
7.3 Measurement of conditions in the chamber during the test
7.3.1 Advantages
Advantages are as follows:
a) This method gives the best estimate in the measured value of the conditions
experienced by the item under test. It is ideal when different kinds of loads and different
tests are being performed.
b) The effect of the load on the control of the chamber can be accurately assessed.
c) History of chamber calibration drift need not be assessed.
d) Careful positioning of the sensors can give detailed information about critical parts of
the load.
e) Anomalies from heat dissipating loads can be quantified.
f) This method can be economical because the chamber is not calibrated for conditions
that are not required.
7.3.2 Disadvantages
Disadvantages include:
a) Measurement equipment is required for every test.
b) Uncertainty calculations shall be made for every test.
c) Can be the most expensive method because measurement equipment is required all the
time.
7.4 Conditions to measure
If measurements are made at the time of the test, then an uncertainty can be calculated for
that condition. Alternatively, a calibration of the chamber could be performed for each
condition for which the chamber is to be used. However, in practice it is not always necessary
to perform a calibration at every possible condition.
If measurements are not made at the time of the test, the whole of the measurement
sequence and the analysis shall be repeated for a set of conditions that cover at least a range
of use. For evaluation an example is given in IEC 60068-3-6.
For temperature only (i.e. humidification OFF) this should include sufficient measurement
points to cover:
− the highest temperature;
− the lowest temperature;
− at least two temperatures with the cooling ON;
− at least two temperatures with the heating ON.
In addition to the temperature-only measurements above, measurements should be performed
for at least two humidity values, covering the range, for any of the above conditions where
humidity tests are to be performed.
It is necessary to perform so many measurements because each of the humidity and
temperature control systems can cause the chamber to have different gradients and
fluctuations. The temperature control is often much worse when the humidity system is on.

– 18 – 60068-3-11 © IEC:2007
If the chamber is only used at a few specific set points then only these need to be calibrated.
When the test is not performed at one of the calibrated levels, it is necessary to interpolate
between two calibrated levels. Interpolation should be used with caution and preferably only if
− the calibrated levels are reasonably close to the test level,
− the services used for each calibrated level (refrigeration, dehumidifiers, heaters, etc.) are
the same.
7.5 Measurements required
The measurements required are the same for all the methods; it is simply a matter of when
the measurements are made and how the results are analysed.
7.5.1 Temperature
For temperature measurement, an array of temperature sensors is used to measure the
temperature at points distributed around the chamber. There are other standards
IEC 60068-3-5 that give measurement methods but they do not address the consequences of
uncertainty and only refer to empty chambers.
For an empty chamber, eight sensors are normally used at the corners of the working space,
and a ninth in the centre. For large chambers more sensors may be required.
For a typical load, or an item under test, eight sensors, one at each corner of the object, are
usually used. For very small test items fewer sensors may be sufficient, but at least four
should be used. For large or unusually shaped objects, or where some particular point on the
test item is of special interest, extra sensors should be employed as appropriate.
For a heat dissipating test item, the measurement of the incident air temperature is usually
considered to be the condition of interest for the report but the other sensors should still be
used so that the local effects of the heat from the test item can be quantified.
7.5.2 Humidity
For humidity measurement, a hygrometer is positioned centrally on the incident air side of the
test item or in the centre of an empty chamber. This can be any kind of hygrometer, but is
most likely to be a relative humidity sensor, a psychrometer, or a condensation (chilled mirror)
hygrometer. The vapour pressure can be computed from the humidity and temperature
measurements. The vapour pressure is assumed to be the same everywhere in the chamber
and the relative humidity is computed from this vapour pressure and the temperature at each
of the temperature measurement sensors.
For each condition, a measurement of vapour pressure gradients shall be made so that an
uncertainty due to this variation can be calculated. This can be done using several
hygrometer probes of any type. However, relative humidity probes and psychrometers are
also sensitive to temperature so usually the estimate obtained using these instruments will be
larger than the true value.
60068-3-11 © IEC:2007 – 19 –
Another method is to sample from several points through tubing routed to a single hygrometer
and switched in alternation.
The vapour pressure gradient is normally small and need only be evaluated occasionally.
7.5.3 Recording procedure
To ensure that a valid assessment can be made,at least 5, and preferably 20 or more
recordings should be taken from each sensor at each set condition. The recordings should be
taken over a sufficient period of time so that several control fluctuations of the chamber can
be recorded. A period of 30 min is normally sufficient.
Recordings are taken from the array of sensors after the chamber has stabilized at each set
condition.
Measurements should be taken frequently from each sensor throughout the test period. Table
A.1 shows a typical data set, together with some of the analysis.
It is essential to ensure that the intervals between the measurements do not coincide with the
cycling interval of the chamber.
7.6 Sources of uncertainty
In any measurement, uncertainties arise from four basic sources.
7.6.1 Calibration uncertainties
Calibration uncertainties for the instruments used are listed on the calibration certificate.
These are normally at the 95 % confidence level. When interpreting uncertainties stated on a
calibration certificate, care shall be taken to consider all aspects of uncertaint
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...

La norme EN 60068-3-11:2007, intitulée "Essais environnementaux - Partie 3-11 : Documentation de support et guide - Calcul de l'incertitude des conditions dans les chambres d'essai climatiques", offre une approche systématique pour estimer l'incertitude des conditions de température et d'humidité à l'état stable dans les chambres de test. Son champ d'application est large, car elle est applicable à tous les types d'environnements clos, y compris des salles ou des laboratoires, ce qui en fait un document essentiel pour toute installation utilisant des chambres climatiques. Les forces de cette norme résident dans sa capacité à fournir des méthodes claires pour calculer l'incertitude de mesure, ce qui est crucial pour la fiabilité des résultats d'essai. Les utilisateurs réguliers des chambres climatiques, déjà familiarisés avec l'incertitude de mesure, bénéficieront d'une guidance sur les sources typiques d'incertitude, ainsi que sur la manière de les quantifier et de les combiner. Cela représente une valeur ajoutée pour les laboratoires qui cherchent à améliorer leur méthodologie d'essai. De plus, la norme s'adresse également aux utilisateurs occasionnels ou débutants, en leur offrant des fondations solides sur la compréhension des concepts d'incertitude. Elle aborde les mesures qui concernent directement les conditions ambiantes à l'intérieur de la chambre, comme la température et l'humidité de l'air, ce qui est fondamental pour évaluer correctement la performance des échantillons en test. En intégrant les aspects pratiques de la calibration des performances des chambres climatiques et en clarifiant la distinction entre la calibration des capteurs et celle des conditions mesurées, la norme se positionne comme un outil précieux. Elle favorise ainsi une meilleure compréhension et application des processus de mesure de l'incertitude dans les tests climatiques, renforçant ainsi la confiance dans les résultats obtenus. La norme EN 60068-3-11:2007 est donc d'une grande pertinence pour tous les acteurs impliqués dans les tests environnementaux, contribuant à une standardisation des pratiques et à une amélioration continue de la qualité des essais réalisés.

SIST EN 60068-3-11:2007 문서는 기후 시험 챔버에서 온도와 습도의 변동성을 추정하는 방법을 제시하고 있습니다. 이 표준은 온도 및 습도 챔버에서의 정밀한 측정을 위한 지침을 제공함으로써, 기후 시험을 수행하는 모든 사용자가 활용할 수 있도록 돕습니다. 표준의 중요한 강점 중 하나는 온도 및 습도의 상태를 보장하기 위해 사용해야 하는 측정 방법에 대한 명확한 설명을 포함하고 있다는 점입니다. 이러한 정보는 실험실 환경에서 사용할 수 있는 다양한 환경 밀폐 구조물에도 적용 가능하여, 유연성을 제공합니다. 이는 기후 시험 챔버를 사용하는 첫 번째 사용자나 간헐적으로 사용하는 사용자에게도 유용하며, 측정 불확실성에 대한 이해를 돕는 자료가 됩니다. 기준 문서는 특정 조건에서의 온도 및 습도를 명확히 이해하고 조정하는 방법에 대한 구체적인 가이드라인을 제공합니다. 사용자는 시험 중의 정확한 측정값을 보장하기 위해, 소스의 불확실성을 정량화하고 결합할 수 있는 방법론을 익힐 수 있습니다. 이는 기후 스펙을 충족시키면서, 측정의 신뢰성을 증가시키는데 중요한 요소입니다. 종합적으로, SIST EN 60068-3-11:2007 표준은 기후 시험 챔버를 사용하는 모든 이들에게 필수적인 도구로 자리 잡고 있으며, 이 표준을 통해 사용자들은 보다 향상된 실험 환경을 구축할 수 있을 것입니다. 이렇게 구체화된 표준은 기후 시험의 표준화와 신뢰성을 높이는 데 중요한 역할을 합니다.

標準規格 EN 60068-3-11:2007 は、環境試験に関する重要な資料として位置付けられています。この標準は、温度および湿度試験室における安定状態の温度および湿度条件の不確かさを推定する方法を示しており、測定手法との関連も詳しく説明しています。このように、幅広い環境エンクロージャーに適用可能であり、試験室やそのほかの環境条件を持つ場所でも有用です。 この標準の強みは、気候試験室を利用するすべての人々に向けて、不確かさの計算方法を明示している点にあります。すでに測定の不確かさについての知識を持つ人にとっては、不確かさの典型的な要因やそれをどのように定量化し、結合するかについてのガイダンスとして非常に役立つでしょう。また、初めてまたはたまに使用する人々にもとっては、基礎的な情報を提供し、理解を深める手助けとなります。 さらに、この標準は、試験対象物における空気の湿度と温度のパフォーマンスをキャリブレーションまたは特性評価する点に焦点を当てており、試験室センサーのキャリブレーションとは異なる重要なテーマを扱っています。これにより、使用者は試験室の条件を的確に把握し、それが試験対象物に与える影響を理解することができるようになります。 全体として、SIST EN 60068-3-11:2007 は、気候試験室における温度と湿度条件の不確かさの計算に関して、幅広く適用される実用的かつ重要なガイドラインを提供しており、その関連性は非常に高いと言えます。

Die Norm EN 60068-3-11:2007 bietet umfassende Richtlinien zur Berechnung der Unsicherheit von Temperatur- und Feuchtigkeitsbedingungen in Klimakammern. Sie richtet sich an eine Vielzahl von Nutzern, von erfahrenen Fachleuten im Bereich der Messtechnik bis hin zu Anfängern, die sich zum ersten Mal mit klimatischen Prüfkammern auseinandersetzen. Der Hauptaugenmerk dieser Norm liegt darauf, eine klare Methode zur Schätzung der Unsicherheit in stabilen Temperatur- und Feuchtigkeitsumgebungen bereitzustellen, was für präzise und reproduzierbare Testergebnisse von äußerster Bedeutung ist. Ein herausragendes Merkmal dieser Norm ist ihre umfassende Anwendbarkeit. Sie ist nicht nur für Temperaturkammern geeignet, sondern auch für kombinierte Temperatur- und Klimakammern sowie für jede Art von umwelttechnischen Prüfkammern, einschließlich Labore und spezielle Testräume. Diese generelle Anwendbarkeit unterstreicht die Relevanz der Norm in unterschiedlichen Industrien, die Umwelttests durchführen. Die Norm liefert auch wertvolle Informationen über typische Unsicherheitsquellen und deren Quantifizierung und Kombination. Dies ist besonders nützlich für diejenigen, die bereits Vertrautheit mit der Unsicherheit von Messungen haben und ihre Kenntnisse vertiefen möchten. Gleichzeitig bietet sie Einsteigern eine notwendige Anleitung, um die grundlegenden Konzepte der Unsicherheit im Kontext von Klimakammern zu verstehen. Die klare Abgrenzung zwischen der Kalibrierung der Kammer und der Kalibrierung des Sensors sorgt zudem für ein tieferes Verständnis und verhindert Verwirrung über verschiedene Aspekte des Testprozesses. Diese präzise Unterscheidung ist wichtig, da sie den Anwendern hilft, die Performance der Klimakammer korrekt zu charakterisieren und somit präzise Testergebnisse zu gewährleisten. Insgesamt stellt die EN 60068-3-11:2007 einen wertvollen Beitrag zur Standardisierung im Bereich der Umwelttests dar und ist ein unverzichtbares Dokument für jeden, der sicherstellen möchte, dass die Bedingungen in Klimakammern korrekt und zuverlässig bestimmt werden. Mit ihrer breiten Anwendbarkeit und klaren Richtlinien ist diese Norm ein wichtiges Hilfsmittel für die Industrie.

The EN 60068-3-11:2007 standard provides essential guidance for estimating the uncertainty of conditions in climatic test chambers, specifically focusing on steady-state temperature and humidity. Its scope is comprehensive, addressing both temperature chambers and combined temperature and humidity chambers, making it relevant for various environmental enclosures, including rooms and laboratories. One of the key strengths of this standard is its thorough approach to measuring and quantifying uncertainty. By detailing methods of measurement alongside uncertainty estimation, it caters to a wide audience, from those already versed in measurement uncertainty to first-time users with limited knowledge. This inclusivity enhances its applicability across diverse testing environments. Furthermore, the standard emphasizes the significance of understanding what is being measured or characterized, particularly in relation to the calibration and characterization of the conditions inside the chambers. This distinction is crucial, as it delineates the focus on the conditions experienced by the test item from the technical aspects of sensor calibration. In summary, EN 60068-3-11:2007 stands out as a robust resource for professionals working with climatic test chambers, equipping them with the necessary knowledge to effectively manage and quantify uncertainty in temperature and humidity conditions. Its detailed methodology and practical relevance make it a cornerstone document for those involved in environmental testing.