Medical electrical equipment - Characteristics of digital X-ray imaging devices - Part 1: Determination of the detective quantum efficiency

Specifies the method for the determination of the detective quantum efficiency (DQE) of digital X-ray imaging devices as a function of exposure and of spatial frequency for the working conditions in the range of the medical application as specified by the manufacturer. Is applicable to projection digital X-ray imaging devices producing images in digital format that are used for medical diagnosis. It is restricted to digital X-ray imaging devices that are used for radiographic imaging, such as CR systems, selenium-based systems, flat panel detectors, optically coupled CCD detectors, and digital X-ray image intensifiers used for single exposures.

General Information

Status
Replaced
Publication Date
22-Oct-2003
Current Stage
DELPUB - Deleted Publication
Completion Date
12-Mar-2015
Ref Project

Relations

Buy Standard

Standard
IEC 62220-1:2003 - Medical electrical equipment - Characteristics of digital X-ray imaging devices - Part 1: Determination of the detective quantum efficiency Released:10/23/2003 Isbn:2831872006
English language
29 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


INTERNATIONAL IEC
STANDARD
62220-1
First edition
2003-10
Medical electrical equipment –
Characteristics of digital X-ray imaging devices –
Part 1: Determination of the detective quantum
efficiency
Appareils électromédicaux –
Caractéristiques des appareils d'imagerie à rayonnement X –
Partie 1: Détermination de l'efficacité quantique de détection
Reference number
Publication numbering
As from 1 January 1997 all IEC publications are issued with a designation in the
60000 series. For example, IEC 34-1 is now referred to as IEC 60034-1.
Consolidated editions
The IEC is now publishing consolidated versions of its publications. For example,
edition numbers 1.0, 1.1 and 1.2 refer, respectively, to the base publication, the
base publication incorporating amendment 1 and the base publication incorporating
amendments 1 and 2.
Further information on IEC publications
The technical content of IEC publications is kept under constant review by the IEC,
thus ensuring that the content reflects current technology. Information relating to
this publication, including its validity, is available in the IEC Catalogue of
publications (see below) in addition to new editions, amendments and corrigenda.
Information on the subjects under consideration and work in progress undertaken
by the technical committee which has prepared this publication, as well as the list
of publications issued, is also available from the following:
• IEC Web Site (www.iec.ch)
• Catalogue of IEC publications
The on-line catalogue on the IEC web site (www.iec.ch/searchpub) enables you to
search by a variety of criteria including text searches, technical committees
and date of publication. On-line information is also available on recently issued
publications, withdrawn and replaced publications, as well as corrigenda.
• IEC Just Published
This summary of recently issued publications (www.iec.ch/online_news/ justpub)
is also available by email. Please contact the Customer Service Centre (see
below) for further information.
• Customer Service Centre
If you have any questions regarding this publication or need further assistance,
please contact the Customer Service Centre:

Email: custserv@iec.ch
Tel: +41 22 919 02 11
Fax: +41 22 919 03 00
INTERNATIONAL IEC
STANDARD
62220-1
First edition
2003-10
Medical electrical equipment –
Characteristics of digital X-ray imaging devices –
Part 1: Determination of the detective quantum
efficiency
Appareils électromédicaux –
Caractéristiques des appareils d'imagerie à rayonnement X –
Partie 1: Détermination de l'efficacité quantique de détection
 IEC 2003  Copyright - all rights reserved
No part of this publication may be reproduced or utilized in any form or by any means, electronic or
mechanical, including photocopying and microfilm, without permission in writing from the publisher.
International Electrotechnical Commission, 3, rue de Varembé, PO Box 131, CH-1211 Geneva 20, Switzerland
Telephone: +41 22 919 02 11 Telefax: +41 22 919 03 00 E-mail: inmail@iec.ch  Web: www.iec.ch
PRICE CODE
Commission Electrotechnique Internationale
U
International Electrotechnical Commission
Международная Электротехническая Комиссия
For price, see current catalogue

– 2 – 62220-1  IEC:2003(E)
CONTENTS
FOREWORD . 3
INTRODUCTION .5
1 Scope . 6
2 Normative references. 6
3 Terminology and definitions . 7
4 Requirements . 8
4.1 Operating conditions. 8
4.2 X-RAY EQUIPMENT . 8
4.3 RADIATION QUALITY . 9
4.4 TEST DEVICE .10
4.5 Geometry.11
4.6 IRRADIATION conditions.13
4.6.1 General conditions.13
4.6.2 Exposure measurement .13
4.6.3 Avoidance of LAG EFFECTS .14
4.6.4 IRRADIATION to obtain the CONVERSION FUNCTION.14
4.6.5 IRRADIATION for determination of the NOISE POWER SPECTRUM .14
4.6.6 IRRADIATION with TEST DEVICE in the RADIATION BEAM .15
5 Corrections of RAW DATA .16
6 Determination of the DETECTIVE QUANTUM EFFICIENCY.16
6.1 Definition and formula of DQE(u,v) .16
6.2 Parameters to be used for evaluation .17
6.3 Determination of different parameters from the images .17
6.3.1 Linearization of data .17
6.3.2 The NOISE POWER SPECTRUM (NPS).18
6.3.3 Determination of the MODULATION TRANSFER FUNCTION (MTF).20
7 Format of conformance statement.20
8 Accuracy.21
Annex A (normative) Determination of LAG EFFECTS .23
A.1 Test of additive LAG EFFECTS .23
A.2 Test of multiplicative LAG EFFECTS.24
Annex B (normative) Terminology – Index of defined terms.26
Annex C (informative) Calculation of the input NOISE POWER SPECTRUM.27
Bibliography.28

62220-1  IEC:2003(E) – 3 –
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
MEDICAL ELECTRICAL EQUIPMENT –
CHARACTERISTICS OF DIGITAL X-RAY IMAGING DEVICES –
Part 1: Determination of the detective quantum efficiency
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC provides no marking procedure to indicate its approval and cannot be rendered responsible for any
equipment declared to be in conformity with an IEC Publication.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International Standard IEC 62220-1 has been prepared by subcommittee 62B: Diagnostic
imaging equipment, of IEC technical committee 62: Electrical equipment in medical practice.
The text of this standard is based on the following documents:
FDIS Report on voting
62B/493/FDIS 62B/506/RVD
Full information on the voting for the approval of this standard can be found in the report on
voting indicated in the above table.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.

– 4 – 62220-1  IEC:2003(E)
In this standard, terms printed in SMALL CAPITALS are used as defined in IEC 60788, in
Clause 3 of this standard or other IEC publications referenced in Annex B. Where a defined
term is used as a qualifier in another defined or undefined term it is not printed in SMALL
CAPITALS, unless the concept thus qualified is defined or recognized as a “derived term
without definition”.
NOTE Attention is drawn to the fact that, in cases where the concept addressed is not strongly confined to the
definition given in one of the publications listed above, a corresponding term is printed in lower-case letters.
The committee has decided that the contents of this publication will remain unchanged until
2006-12. At this date, the publication will be
• reconfirmed;
• withdrawn;
• replaced by a revised edition, or
• amended.
62220-1  IEC:2003(E) – 5 –
INTRODUCTION
DIGITAL X-RAY IMAGING DEVICES are increasingly used in medical diagnosis and will widely
replace conventional (analogue) imaging devices such as screen-film systems or analogue X-
RAY IMAGE INTENSIFIER television systems in the future. It is necessary, therefore, to define
parameters that describe the specific imaging properties of these DIGITAL X-RAY IMAGING
DEVICES and to standardize the measurement procedures employed.
There is growing consensus in the scientific world that the DETECTIVE QUANTUM EFFICIENCY
(DQE) is the most suitable parameter for describing the imaging performance of an X-ray
imaging device. The DQE describes the ability of the imaging device to preserve the signal-to-
NOISE ratio from the radiation field to the resulting digital image data. Since in X-ray imaging,
the NOISE in the radiation field is intimately coupled to the exposure level, DQE values can
also be considered to describe the dose efficiency of a given imaging device.
NOTE 1 In spite of the fact that the DQE is widely used to describe the performance of imaging devices, the
connection between this physical parameter and the decision performance of a human observer is not yet
1)
completely understood [1], [3].
NOTE 2 The standard IEC 61262-5 specifies a method to determine the DQE of X-RAY IMAGE INTENSIFIERS at
nearly zero SPATIAL FREQUENCY. It focuses only on the electro-optical components of X-RAY IMAGE INTENSIFIERS, not
on the imaging properties as this standard does. As a consequence, the output is measured as an optical quantity
(luminance), and not as digital data. Moreover, IEC 61262-5 prescribes the use of a RADIATION SOURCE ASSEMBLY,
whereas this standard prescribes the use of an X-RAY TUBE. The scope of IEC 61262-5 is limited to X-RAY IMAGE
INTENSIFIERS and does not interfere with the scope of this standard.
The DQE is already widely used by manufacturers to describe the performance of their
equipment. The specification of the DQE is also required by regulatory agencies (such as the
Food and Drug Administration (FDA)) for admission procedures. However, there is presently
no standard governing either the measurement conditions or the measurement procedure with
the consequence that values from different sources may not be comparable.
This standard has therefore been developed in order to specify the measurement procedure
together with the format of the conformance statement for the DETECTIVE QUANTUM EFFICIENCY
of DIGITAL X-RAY IMAGING DEVICES.
In the DQE calculations proposed in this standard, it is assumed that system response is
measured for objects that attenuate all energies equally (task-independent) [5].
The standard will be beneficial for manufacturers, users, distributors and regulator
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.