Amendment 2 - Medical electrical equipment - Part 1-10: General requirements for basic safety and essential performance - Collateral Standard: Requirements for the development of physiologic closed-loop controllers

Amendement 2 - Appareils électromédicaux - Partie 1-10 : Exigences générales pour la sécurité de base et les performances essentielles - Norme collatérale : Exigences pour le développement des régulateurs physiologiques en boucle fermée

General Information

Status
Published
Publication Date
21-Jul-2020
Current Stage
PPUB - Publication issued
Completion Date
22-Jul-2020
Ref Project

Buy Standard

Standard
IEC 60601-1-10:2007/AMD2:2020 - Amendment 2 - Medical electrical equipment - Part 1-10: General requirements for basic safety and essential performance - Collateral Standard: Requirements for the development of physiologic closed-loop controllers
English and French language
21 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (sample)

IEC 60601-1-10
Edition 1.0 2020-07
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
AMENDMENT 2
AMENDEMENT 2
Medical electrical equipment –
Part 1-10: General requirements for basic safety and essential performance –
Collateral Standard: Requirements for the development of physiologic
closed-loop controllers
Appareils électromédicaux –
Partie 1-10: Exigences générales pour la sécurité de base et les performances
essentielles – Norme collatérale: Exigences pour le développement des
régulateurs physiologiques en boucle fermée
IEC 60601-1-10:2007-11/AMD2:2020-07(en-fr)
---------------------- Page: 1 ----------------------
THIS PUBLICATION IS COPYRIGHT PROTECTED
Copyright © 2020 IEC, Geneva, Switzerland

All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form

or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from

either IEC or IEC's member National Committee in the country of the requester. If you have any questions about IEC

copyright or have an enquiry about obtaining additional rights to this publication, please contact the address below or

your local IEC member National Committee for further information.

Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite ni

utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie et

les microfilms, sans l'accord écrit de l'IEC ou du Comité national de l'IEC du pays du demandeur. Si vous avez des

questions sur le copyright de l'IEC ou si vous désirez obtenir des droits supplémentaires sur cette publication, utilisez

les coordonnées ci-après ou contactez le Comité national de l'IEC de votre pays de résidence.

IEC Central Office Tel.: +41 22 919 02 11
3, rue de Varembé info@iec.ch
CH-1211 Geneva 20 www.iec.ch
Switzerland
About the IEC

The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes

International Standards for all electrical, electronic and related technologies.
About IEC publications

The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the

latest edition, a corrigendum or an amendment might have been published.

IEC publications search - webstore.iec.ch/advsearchform Electropedia - www.electropedia.org

The advanced search enables to find IEC publications by a The world's leading online dictionary on electrotechnology,

variety of criteria (reference number, text, technical containing more than 22 000 terminological entries in English

committee,…). It also gives information on projects, replaced and French, with equivalent terms in 16 additional languages.

and withdrawn publications. Also known as the International Electrotechnical Vocabulary

(IEV) online.
IEC Just Published - webstore.iec.ch/justpublished

Stay up to date on all new IEC publications. Just Published IEC Glossary - std.iec.ch/glossary

details all new publications released. Available online and once 67 000 electrotechnical terminology entries in English and

a month by email. French extracted from the Terms and definitions clause of IEC
publications issued between 2002 and 2015. Some entries

IEC Customer Service Centre - webstore.iec.ch/csc have been collected from earlier publications of IEC TC 37, 77,

If you wish to give us your feedback on this publication or need 86 and CISPR.
further assistance, please contact the Customer Service
Centre: sales@iec.ch.
A propos de l'IEC

La Commission Electrotechnique Internationale (IEC) est la première organisation mondiale qui élabore et publie des

Normes internationales pour tout ce qui a trait à l'électricité, à l'électronique et aux technologies apparentées.

A propos des publications IEC

Le contenu technique des publications IEC est constamment revu. Veuillez vous assurer que vous possédez l’édition la

plus récente, un corrigendum ou amendement peut avoir été publié.

Recherche de publications IEC - Le premier dictionnaire d'électrotechnologie en ligne au monde,

webstore.iec.ch/advsearchform avec plus de 22 000 articles terminologiques en anglais et en

La recherche avancée permet de trouver des publications IEC français, ainsi que les termes équivalents dans 16 langues

en utilisant différents critères (numéro de référence, texte, additionnelles. Egalement appelé Vocabulaire

comité d’études,…). Elle donne aussi des informations sur les Electrotechnique International (IEV) en ligne.

projets et les publications remplacées ou retirées.
Glossaire IEC - std.iec.ch/glossary

IEC Just Published - webstore.iec.ch/justpublished 67 000 entrées terminologiques électrotechniques, en anglais

Restez informé sur les nouvelles publications IEC. Just et en français, extraites des articles Termes et définitions des

Published détaille les nouvelles publications parues. publications IEC parues entre 2002 et 2015. Plus certaines

Disponible en ligne et une fois par mois par email. entrées antérieures extraites des publications des CE 37, 77,

86 et CISPR de l'IEC.
Service Clients - webstore.iec.ch/csc
Si vous désirez nous donner des commentaires sur cette
publication ou si vous avez des questions contactez-nous:
sales@iec.ch.
Electropedia - www.electropedia.org
---------------------- Page: 2 ----------------------
IEC 60601-1-10
Edition 1.0 2020-07
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
A MENDMENT 2
AM ENDEMENT 2
Medical electrical equipment –
Part 1-10: General requirements for basic safety and essential performance –
Collateral Standard: Requirements for the development of physiologic
closed-loop controllers
Appareils électromédicaux –
Partie 1-10: Exigences générales pour la sécurité de base et les performances
essentielles – Norme collatérale: Exigences pour le développement des
régulateurs physiologiques en boucle fermée
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
INTERNATIONALE
ICS 11.040.01 ISBN 978-2-8322-8622-7

Warning! Make sure that you obtained this publication from an authorized distributor.

Attention! Veuillez vous assurer que vous avez obtenu cette publication via un distributeur agréé.

---------------------- Page: 3 ----------------------
– 2 – IEC 60601-1-10:2007/AMD2:2020
© IEC 2020
FOREWORD

This amendment has been prepared by subcommittee 62A: Common aspects of electrical

equipment used in medical practice, of IEC technical committee 62: Electrical equipment in

medical practice, and ISO subcommittee 3: Respiratory devices and related equipment used for

patient care, of ISO technical committee 121: Anaesthetic and respiratory equipment.

It is published as a double logo amendment.
The text of this amendment is based on the following documents of IEC:
FDIS Report on voting
62A/1394/FDIS 62A/1409/RVD

Full information on the voting for the approval of this amendment can be found in the report on

voting indicated in the above table. In ISO, the amendment has been approved by 17 P

members out of 17 having cast a vote.

The committee has decided that the contents of this amendment and the base publication will

remain unchanged until the stability date indicated on the IEC web site under

"http://webstore.iec.ch" in the data related to the specific publication. At this date, the

publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.

NOTE The attention of users of this document is drawn to the fact that equipment manufacturers and testing

organizations may need a transitional period following publication of a new, amended or revised IEC or

ISO publication in which to make products in accordance with the new requirements and to equip themselves for

conducting new or revised tests. It is the recommendation of the committee that the content of this publication be

adopted for mandatory implementation nationally not earlier than 3 years from the date of publication.

_____________
---------------------- Page: 4 ----------------------
IEC 60601-1-10:2007/AMD2:2020 – 3 –
© IEC 2020
INTRODUCTION TO AMENDMENT 2

The first edition of IEC 60601-1-10 was published in 2007 and amended in 2013. Since the

publication of IEC 60601-1-10:2007+A1:2013, the IEC Subcommittee (SC) 62A Secretariat has

been collecting issues from a variety of sources including comments from National Committees.

At the November 2015 meeting of IEC/SC 62A in Kobe, Japan, the subcommittee initiated a

process to identify high-priority issues that need to be considered in an amendment and should

not wait until the second edition of IEC 60601-1-10, which is presently targeted for publication

sometime after 2024.

Those issues selected for inclusion on the final "short list" to be addressed in Amendment 2

were those approved by a 2/3 majority of the National Committees present and voting at the

Frankfurt meeting of SC 62A. At the meeting held on 10 October 2016, 13 items were presented

to the National Committees present. All 13 items received the required 2/3 majority of the

National Committees present and voting and have been included in the "short list" for

consideration in preparing Amendment 2. All remaining issues have been placed on a "long list"

for consideration in the second edition of IEC 60601-1-10.

The "short list" of issues was documented in the design specification for Amendment 2. As

IEC 60601-1-10 was jointly developed with ISO/TC 121/SC 3, the work was assigned to

IEC/SC 62A-ISO/TC 121/SC 3 Joint Working Group (JWG) 5. JWG 5 was directed to consider

each issue described in Clause 6 of the design specification and develop an appropriate

solution for the identified problem. That final solution in this amendment can encompass any

technical solution proposed by the author of the issue or it can involve a different solution

developed by the expert group. The expert group can also have recommended that no change

to the standard was justified by the problem statement.

Because this is an amendment to IEC 60601-1-10:2007, the style in force at the time of

publication of IEC 60601-1-10 has been applied to this amendment. The style specified in

ISO/IEC Directives Part 2:2018 has only been applied when implementing the new style

guidance would not result in additional editorial changes. For example, references to

amendments take the following form: "IEC 60601-1:2005+A1:2012+A2:2020".

Users of this document should note that when constructing the dated references to specific

elements in a standard, such as definitions, amendments are only referenced if they modified

the text being cited. For example, if a reference is made to a definition that has not been

modified by an amendment, then the reference to the amendment is not included in the dated

reference.
---------------------- Page: 5 ----------------------
– 4 – IEC 60601-1-10:2007/AMD2:2020
© IEC 2020
1.1 * Scope

Replace the existing second paragraph and note with the following new paragraph and example:

This collateral standard specifies requirements for the development (analysis, design,

VERIFICATION and VALIDATION) of a PHYSIOLOGIC CLOSED-LOOP CONTROLLER (PCLC) as part of a

PHYSIOLOGIC CLOSED-LOOP CONTROL SYSTEM (PCLCS) to control at least one PATIENT VARIABLE (i.e.

a PHYSIOLOGIC VARIABLE) in ME EQUIPMENT and ME SYSTEMS.

EXAMPLE A PATIENT VARIABLE can be a measure of body chemistry (e.g. electrolytes or blood glucose value), a

physical property (e.g. body temperature, electrophysiologic characteristic, hemodynamic quantity), or a

pharmaceutical concentration.
1.3.1 IEC 60601-1

Replace, in the first two dashes of the existing second paragraph, the parentheses, added by

Amendment 1, with the words ", including any amendments".
2 Normative references

Replace the existing references to IEC 60601-1, IEC 60601-1-6, IEC 60601-1-8, IEC 62366 and

ISO 14971, modified by Amendment 1, by the following new references:

IEC 60601-1:2005, Medical electrical equipment – Part 1: General requirements for basic safety

and essential performance
Amendment 1:2012
Amendment 2:2020

IEC 60601-1-6:2010, Medical electrical equipment – Part 1-6: General requirements for basic

safety and essential performance – Collateral standard: Usability
Amendment 1:2013
Amendment 2:2020

IEC 60601-1-8:2006, Medical electrical equipment – Part 1-8: General requirements for basic

safety and essential performance – Collateral Standard: General requirements, tests and

guidance for alarm systems in medical electrical equipment and medical electrical systems

Amendment 1:2012
Amendment 2:2020

IEC 62366-1:2015, Medical devices – Part 1: Application of usability engineering to medical

devices
Amendment 1:2020

ISO 14971:2019, Medical devices – Application of risk management to medical devices

Add the following new normative reference:
ISO 9000:2015, Quality management systems – Fundamentals and vocabulary
3 Terms and definitions
Replace the existing first paragraph with the following new paragraph:
For the purposes of this document, the terms and definitions given in

IEC 60601-1:2005+A1:2012+A2:2020, IEC 60601-1-6:2010+A1:2013+A2:2020, IEC 60601-1-

8:2006+A1:2012+A2:2020, IEC 62366-1:2015+A1:2020, ISO 9000:2015 and the following

apply.
---------------------- Page: 6 ----------------------
IEC 60601-1-10:2007/AMD2:2020 – 5 –
© IEC 2020

Figure 1 – Functional diagram indicating typical components of a PHYSIOLOGIC CLOSED-

LOOP CONTROL SYSTEM (PCLCS) utilizing a PCLC

Add, to the definition of symbol "y" in the existing key, the word "controlled" before "PHYSIOLOGIC

VARIABLE".
3.4
* COMMAND VARIABLE
Add the following new note:
NOTE A COMMAND VARIABLE may be a range or a function (e.g. clinical protocol).
3.20
* PHYSIOLOGIC CLOSED-LOOP CONTROLLER
Add the following new notes:

NOTE 1 A PCLC may utilize multiple PHYSIOLOGIC VARIABLES and COMMAND VARIABLES as well as multiple

CONTROLLER OUTPUT VARIABLES.

NOTE 2 A PATIENT VARIABLE may be used as an input to adjust the parameters of the CONTROL TRANSFER ELEMENT.

3.21
PHYSIOLOGIC VARIABLE
Replace the existing definition with the following new definition:
PATIENT VARIABLE intended to be controlled
Delete the existing note.
3.23
RELATIVE OVERSHOOT

Add, after "PHYSIOLOGIC VARIABLE" in the existing Note 1, the words "or FEEDBACK VARIABLE".

Add the following new term and definition:
3.29
* PATIENT VARIABLE
PATIENT attribute, characteristic, quantity or condition that is measured

EXAMPLE A PATIENT VARIABLE can be a measure of body chemistry (e.g. electrolytes or blood glucose value), a

physical property (e.g. body temperature, electrophysiologic characteristic, hemodynamic quantity), or a

pharmaceutical concentration.
4 * General requirements

Add, before the word "substances" in the existing two bullets of the first paragraph, the words

"or removed" in two places.

Add, after the existing dash "PATIENT TRANSFER ELEMENT, including any hysteresis" of the first

paragraph, the following new dash:
– uncertainty of the model of the PATIENT TRANSFER ELEMENT;
5.1 * Instructions for use
Replace, in the existing second paragraph, "Table C.2" with "Table C.1".
---------------------- Page: 7 ----------------------
– 6 – IEC 60601-1-10:2007/AMD2:2020
© IEC 2020
5.2 Technical description
Add, before the first existing paragraph, the following new paragraph:

In addition to the requirements in 7.9.3 of the general standard, the technical description shall

contain the PCLCS block diagram and theory of operation.
Replace, in the existing first paragraph, "Table C.3" with "Table C.2".
Add, after the existing paragraph, the following new paragraph:
Compliance is checked by inspection of the technical description.
SABILITY
6.1 * U

Replace the three existing dashes after "the current value of:" with the following new dashes:

– COMMAND VARIABLE or REFERENCE VARIABLE;
– CONTROLLER OUTPUT VARIABLE or MANIPULATED VARIABLE;
– PHYSIOLOGIC VARIABLE or FEEDBACK VARIABLE; and
– any PATIENT VARIABLE that is used by the PCLC;
6.3 * PCLCS VARIABLE logging
Replace the existing first paragraph with the following new paragraphs:

ME EQUIPMENT or ME SYSTEMS that incorporate a PCLC shall provide a means to log the values

of at least
– the COMMAND VARIABLE or REFERENCE VARIABLE;
– the CONTROLLER OUTPUT VARIABLE or MANIPULATED VARIABLE;
– the PHYSIOLOGIC VARIABLE or FEEDBACK VARIABLE; and
– any PATIENT VARIABLE that is used by the PCLC.

The log is necessary to analyze the performance of the PCLCS. The resolution and duration of

the log shall be based on HAZARDS identified in Clause 4. The log should be capable of storing

the information for a reasonable period of time.
8.2.1 RECORDS and PROCESS scaling

Replace, in the existing first paragraph, "IEC 62304:2006" with "Clause 14 of the general

standard".
8.2.2.1 * Application specification

Replace the existing subclause, including its title, with the following new subclause and title:

8.2.2.1 * USE SPECIFICATION

The MANUFACTURER shall prepare the USE SPECIFICATION, as required in IEC 62366-1, of the

ME EQUIPMENT or ME SYSTEM that incorporates a PCLC.

A summary of the USE SPECIFICATION shall be included in the instructions for use.

Compliance is checked by inspection of the USABILITY ENGINEERING FILE and the instructions for

use.
---------------------- Page: 8 ----------------------
IEC 60601-1-10:2007/AMD2:2020 – 7 –
© IEC 2020
8.2.2.2 * State VARIABLES

Replace the existing subclause, including its title, with the following new subclause and title:

8.2.2.2 * PCLCS attributes
MANUFACTURER shall characterize attributes of the following:
The
– the COMMAND VARIABLE or REFERENCE VARIABLE;
– the CONTROLLER OUTPUT VARIABLE or MANIPULATED VARIABLE;
– the PHYSIOLOGIC VARIABLE or FEEDBACK VARIABLE; and
– the PATIENT VARIABLE that is used by the PCLC.

The MANUFACTURER shall characterize the limitations and ranges of the PATIENT TRANSFER

ELEMENT, including INTERPATIENT VARIABILITY and INTRAPATIENT VARIABILITY.

If practicable, the MANUFACTURER shall develop and use a mathematical model to characterize

the PATIENT TRANSFER ELEMENT. If using a MATHEMATICAL MODEL, the MANUFACTURER shall provide

a justification to support that the model represents the relevant PATIENT responses. The

INTERPATIENT VARIABILITY and INTRAPATIENT VARIABILITY and identify
justification shall consider

the limitations of that model, including a characterization of the uncertainty of that mathematical

model, such as by using a credibility assessment framework.
NOTE See reference [14] for additional information.

If developing a mathematical model of the PATIENT TRANSFER ELEMENT is not practicable, the

MANUFACTURER shall provide a justification for any assumptions made regarding the PATIENT

TRANSFER ELEMENT.
The MANUFACTURER shall describe the PCLC modes of operation.
Compliance is checked by inspection of the RISK MANAGEMENT FILE.
8.2.5.1 * VALIDATION plan

Replace, in the existing first paragraph, the fourth and fifth dashes with the following new

dashes:
– monitoring of the PHYSIOLOGIC VARIABLES or FEEDBACK VARIABLES;
– PATIENT TRANSFER ELEMENT, including INTERPATIENT VARIABILITY and INTRAPATIENT
VARIABILITY; and
Replace, in the existing note, the word "NOTE" with "NOTE 1".
Add, after the existing note, the following new paragraph and note:

When a mathematical model of the PATIENT TRANSFER ELEMENT is used as part of PCLCS

VALIDATION, the MANUFACTURER shall provide a rationale to support that the model is sufficiently

credible for its role in PCLCS VALIDATION. A credibility assessment framework may be used to

generate a rationale to support that a mathematical model is sufficiently credible for a proposed

context of use.
NOTE 2 See reference [14] for additional information.
A.1 General guidance
Add, after "delivery" in the existing first paragraph, the words "or removal".
---------------------- Page: 9 ----------------------
– 8 – IEC 60601-1-10:2007/AMD2:2020
© IEC 2020

Replace, in the existing note following the third paragraph, the word "NOTE" with "NOTE 1".

Replace the existing Example 1 with the following new example:

EXAMPLE 1 ME EQUIPMENT providing a safety interlock that stops (and does not titrate or restart) energy or

substance delivery based on physiologic limits is not a PCLCS because the PATIENT VARIABLE is not controlled (i.e. it

is not a PHYSIOLOGIC VARIABLE). For example, ME EQUIPMENT that stops, or decreases by a pre-determined rate, the

infusion of a sedative-hypnotic or opioid intravenous infusion when blood oxygen saturation (SpO ) or respiratory

rate decreases below threshold values is not a PCLCS. On the other hand it would be a PCLCS, if the ME EQUIPMENT

interrupts and restarts the infusion or increases the infusion based on a PHYSIOLOGIC VARIABLE rising above a

threshold or falling below a threshold.
Replace the existing Example 3 with the following new example:

EXAMPLE 3 A pressure-controlled (or pressure-regulated volume-controlled) ventilator that uses airway pressure

(or pressure and volume) for feedback to control breathing system pressure. The ventilator is a closed-loop control

system, but not a PCLCS, because while the breathing system pressure (or pressure and volume) is both the

'MANIPULATED VARIABLE' and 'FEEDBACK VARIABLE', it is not a quantity or condition measured from the PATIENT.

Replace, in the existing note following Example 3, the word "NOTE" with "NOTE 2".

Add, after the words "baby incubator uses" in the existing Example 4, the article "the" and add,

after "baby's temperature", the parenthetical "(e.g. skin or core)".
Add, after the existing Example 8, the following new example:

EXAMPLE 9 Sleep apnoea breathing therapy equipment (including auto CPAP or auto continuous positive airway

pressure equipment) uses air flow and pressure from inside the equipment for feedback control to control the

delivered pressure. CPAP equipment is a closed-loop control system, but not a PCLCS, because while the breathing

system pressure is both the MANIPULATED VARIABLE and FEEDBACK VARIABLE, it is not a quantity or condition measured

from the PATIENT.
A.2 Rationale for particular clauses and subclauses
Definition 3.8 – DISTRIBUTED PCLCS

Replace, in the existing first paragraph, the second sentence with the following new sentence:

A typical example would be a stand-alone anaesthesia workstation controlling the effect of the

anaesthetic (for example, blood pressure or EEG measurement).
Definition 3.20 – PHYSIOLOGIC CLOSED-LOOP CONTROLLER
Replace the existing text with the following new text:

A PHYSIOLOGIC CLOSED-LOOP CONTROLLER uses feedback from a PATIENT VARIABLE (i.e. a

PHYSIOLOGIC VARIABLE) to adjust that variable to the COMMAND VARIABLE. Such a controlled

PATIENT VARIABLE is called a PHYSIOLOGIC VARIABLE. Many such controllers are therapeutic

controllers because they are being used to control the flow of energy or substances to a PATIENT

for therapeutic purposes. The committee chose to not use "therapeutic" in the defined term

because these controllers could also be used to control the delivery of non-therapeutic

substances such as anaesthetic agents.

A PCLCS by design can incorporate multiple PCLCS, multiple PHYSIOLOGIC VARIABLES or multiple

COMMAND VARIABLES. A PATIENT VARIABLE that is not controlled can be measured from the

PATIENT and used to influence output of the CONTROL TRANSFER ELEMENT by adjusting algorithm

parameters.

EXAMPLE 1 Neonatal PATIENT oxygenation based on controlled SpO can measure pulse rate to adjust the gain in

the PCLC. The PATIENT VARIABLE, pulse rate, is not being controlled but is a variable directly affecting the output of

the PCLC.
---------------------- Page: 10 ----------------------
IEC 60601-1-10:2007/AMD2:2020 – 9 –
© IEC 2020

EXAMPLE 2 Hemodynamic control of mean arterial blood pressure (MAP) and cardiac output (CO) by fluid

resuscitation and vasopressor therapy. In this example, two PHYSIOLOGIC VARIABLES are being controlled (CO and

MAP) by infusion of fluid and vasopressors. The MANIPULATED VARIABLES are the rate of fluid and vasopressor infusion

and the COMMAND VARIABLES are the desired CO and MAP. This is an example of a multiple-input multiple-output

PCLC.
Add, after the existing rationale to 3.22, the following new rationale:
Definition 3.29 – PATIENT VARIABLE

The term PATIENT VARIABLE distinguishes between a controlled PATIENT VARIABLE (i.e.

PHYSIOLOGIC VARIABLE) and a PATIENT VARIABLE that is not directly controlled. A PATIENT VARIABLE

can be used to affect different elements of the PCLCS (e.g. is used to adjust a parameter in the

PCLC or FEEDBACK ELEMENT).

While environmental factors like ambient temperature, ambient light, etc. can also be recorded

and used for adjustments of the controller output variable, a measured PATIENT VARIABLE reflects

attribute, characteristic, quantity or condition of the PATIENT that is not just a reflection of those

environmental conditions.
Subclause 5.1 – Instructions for use

Replace, in the existing first paragraph, modified by Amendment 1, the reference "IEC 62366"

with "IEC 62366-1".

Replace, in the existing first sentence of the third paragraph, the words "end-tidal PATIENT

anaesthetic agent concentration" with "anaesthetic agent effect on the PATIENT".
Subclause 8.2.2.1 – Application specification

Replace the existing rationale, including the title, by the following new rationale and title:

Subclause 8.2.2.1 – USE SPECIFICATION

The ME EQUIPMENT or ME SYSTEM USE SPECIFICATION describes the important attributes that are

fundamental to their function. The ME EQUIPMENT or ME SYSTEM USE SPECIFICATION is the

foundation for defining the PCLCS.
VARIABLES
Subclause 8.2.2.2 – State
Replace the existing title with the following new title:
Subclause 8.2.2.2 – PCLCS attributes

Replace the existing penultimate dash "Limits of the range of the PATIENT TRANSFER ELEMENT",

including the rationale, with the following new dash:
– Model of the PATIENT TRANSFER ELEMENT

It might not be practicable to capture the full spectrum of human variation in a model of sufficient

fidelity of the PATIENT TRANSFER ELEMENT for either development or validation due to (for

example) ethical or experimental constraints. Establishing the limits of the model is fundamental

to the safe operation of a PCLCS as it impacts the approach taken to validate the PCLCS.

Understanding the limits of variation, including the INTERPATIENT VARIABILITY and INTRAPATIENT

VARIABILITY, of transfer function that is embodied by the model of the PATIENT TRANSFER ELEMENT

is necessary for the design of a safe and effective PCLCS.
---------------------- Page: 11 ----------------------
– 10 – IEC 60601-1-10:2007/AMD2:2020
© IEC 2020
C.1 ACCOMPANYING DOCUMENTS, instructions for use
Replace, in the existing paragraph, "Table C.2" with "Table C.1.
Table C.2 – ACCOMPANYING DOCUMENTS, instructions for use
Replace, in the existing title, "Table C.2" with "Table C.1".

Replace the existing penultimate requirement, "Summary of application specification" with

"Summary of USE SPECIFICATION".
C.2 ACCOMPANYING DOCUMENTS, technical description
Replace, in the existing paragraph, "Table C.3" with "Table C.2".
Table C.3 – ACCOMPANYING DOCUMENTS, technical description
Replace, in the existing title, "Table C.3" with "Table C.2".
Bibliography
Add the following new entry to the Bibliography:
[14] ASME V&V 40-2018, Assessing Credibility of Computational Modeling Through
Verification and Validation: Application to Medical Devices
Index of defined terms used in this collateral standard
Replace the following existing terms in the index, modified by Amendment 1:

ALARM CONDITION ................................................................. IEC 60601-1:2005+A1:2012, 3.141

ALARM SIGNAL ...................................................................... IEC 60601-1:2005+A1:2012, 3.142

ALARM SYSTEM .................................................................... IEC 60601-1-:2005+A1:2012, 3.143

HAZARD................................................................... IEC 60601-1:2005+A1:2012+A2:2020, 3.39

INFORMATION SIGNAL ............................................................ IEC 60601-1:2005+A2:2020, 3.150

INTENDED USE ........................................................................ IEC 60601-1:2005+A2:2020, 3.44

MANUFACTURER ...................................................................... IEC 60601-1:2005+A2:2020, 3.55

OBJECTIVE EVIDENCE ............................................... IEC 60601-1:2005+A1:2012+A2:2020, 3.72

OPERATOR PROFILE ................................................................ IEC 60601-1-6:2006+A2:2020, 3.2

PROCESS ............................................................................... IEC 60601-1:2005+A2:2020, 3.89

RECORD .................................................................. IEC 60601-1:2005+A1:2012+A2:2020, 3.98

RESIDUAL RISK ...................................................................... IEC 60601-1:2005+A2:2020, 3.100

RISK ...................................................................... IEC 60601-1:2005+A1:2012+A22020, 3.102

RISK ANALYSIS ........................................................ IEC 60601-1:2005+A1:2012+A22020, 3.103

RISK MANAGEMENT ................................................................ IEC 60601-1:2005+A2:2020, 3.107

RISK MANAGEMENT FILE .......................................... IEC 60601-1:2005+A1:2012+A2:

...