IEC 60601-2-2:2017
(Main)Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
IEC 60601-2-2:2017 is also available as IEC 60601-2-2:2017 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.
IEC 60601-2-2:2017 applies to the basic safety and essential performance of HF surgical equipment and HF surgical accessories.
HF surgical equipment having a rated output power not exceeding 50 W (for example for micro-coagulation, or for use in dentistry or ophthalmology) is exempt from certain of the requirements of this particular standard. These exemptions are indicated in the relevant requirements.
The object of this particular standard is to establish particular basic safety and essential performance requirements for HF surgical equipment and HF surgical accessories. This sixth edition cancels and replaces the fifth edition published in 2009. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
- refinement and additions to the defined terms;
- additional separation of the requirements for HF surgical equipment and HF surgical accessories;
- a new requirement for adult neutral electrodes to be contact quality monitoring neutral electrodes;
- new requirements for devices that have or use a high current mode.
Appareils électromédicaux - Partie 2-2: Exigences particulières pour la sécurité de base et les performances essentielles des appareils d'électrochirurgie à courant haute fréquence et des accessoires d'électrochirurgie à courant haute fréquence
IEC 60601-2-2:2017 est disponible sous forme de IEC 60601-2-2:2017 RLV qui contient la Norme internationale et sa version Redline, illustrant les modifications du contenu technique depuis l'édition précédente.
L'IEC 60601-2-2:2017 s’applique à la sécurité de base et aux performances essentielles des appareils d’électrochirurgie HF et des accessoires d'électrochirurgie HF. Les appareils d’électrochirurgie HF dont la puissance de sortie assignée est inférieure ou égale à 50 W (destinés, par exemple, à la micro coagulation, à l’ophtalmologie ou à l’usage dentaire) sont exemptés de certaines exigences de la présente norme particulière. Ces exemptions sont indiquées dans les exigences correspondantes.
La présente norme particulière a pour objet d’établir des exigences particulières relatives à la sécurité de base et aux performances essentielles des appareils d’électrochirurgie HF et des accessoires d’électrochirurgie HF.
Cette sixième édition annule et remplace la cinquième édition parue en 2009. Cette édition constitue une révision technique. Cette édition inclut les modifications techniques majeures suivantes par rapport à l'édition précédente:
- des précisions et des ajouts aux termes définis;
- une séparation supplémentaire des exigences relatives aux appareils d'électrochirurgie à courant haute fréquence (HF) et aux accessoires d'électrochirurgie à courant haute fréquence (HF);
- une nouvelle exigence concernant les électrodes neutres adultes devant servir d'électrodes neutres de surveillance de la qualité du contact;
- de nouvelles exigences relatives aux appareils ayant ou utilisant un mode de courant élevé.
General Information
Buy Standard
Standards Content (Sample)
IEC 60601-2-2
Edition 6.0 2017-03
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
Medical electrical equipment –
Part 2-2: Particular requirements for the basic safety and essential performance
of high frequency surgical equipment and high frequency surgical accessories
Appareils électromédicaux –
Partie 2-2: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils d'électrochirurgie à courant haute fréquence et des
accessoires d'électrochirurgie à courant haute fréquence
IEC 60601-2-2:2017-03(en-fr)
---------------------- Page: 1 ----------------------
THIS PUBLICATION IS COPYRIGHT PROTECTED
Copyright © 2017 IEC, Geneva, Switzerland
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form
or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from
either IEC or IEC's member National Committee in the country of the requester. If you have any questions about IEC
copyright or have an enquiry about obtaining additional rights to this publication, please contact the address below or
your local IEC member National Committee for further information.Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite
ni utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie
et les microfilms, sans l'accord écrit de l'IEC ou du Comité national de l'IEC du pays du demandeur. Si vous avez des
questions sur le copyright de l'IEC ou si vous désirez obtenir des droits supplémentaires sur cette publication, utilisez
les coordonnées ci-après ou contactez le Comité national de l'IEC de votre pays de résidence.
IEC Central Office Tel.: +41 22 919 02 113, rue de Varembé Fax: +41 22 919 03 00
CH-1211 Geneva 20 info@iec.ch
Switzerland www.iec.ch
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigenda or an amendment might have been published.IEC Catalogue - webstore.iec.ch/catalogue Electropedia - www.electropedia.org
The stand-alone application for consulting the entire The world's leading online dictionary of electronic and
bibliographical information on IEC International Standards, electrical terms containing 20 000 terms and definitions in
Technical Specifications, Technical Reports and other English and French, with equivalent terms in 16 additional
documents. Available for PC, Mac OS, Android Tablets and languages. Also known as the International Electrotechnical
iPad. Vocabulary (IEV) online.IEC publications search - www.iec.ch/searchpub IEC Glossary - std.iec.ch/glossary
The advanced search enables to find IEC publications by a 65 000 electrotechnical terminology entries in English and
variety of criteria (reference number, text, technical French extracted from the Terms and Definitions clause of
committee,…). It also gives information on projects, replaced IEC publications issued since 2002. Some entries have been
and withdrawn publications. collected from earlier publications of IEC TC 37, 77, 86 and
CISPR.IEC Just Published - webstore.iec.ch/justpublished
Stay up to date on all new IEC publications. Just Published IEC Customer Service Centre - webstore.iec.ch/csc
details all new publications released. Available online and If you wish to give us your feedback on this publication or
also once a month by email. need further assistance, please contact the Customer Service
Centre: csc@iec.ch.A propos de l'IEC
La Commission Electrotechnique Internationale (IEC) est la première organisation mondiale qui élabore et publie des
Normes internationales pour tout ce qui a trait à l'électricité, à l'électronique et aux technologies apparentées.
A propos des publications IECLe contenu technique des publications IEC est constamment revu. Veuillez vous assurer que vous possédez l’édition la
plus récente, un corrigendum ou amendement peut avoir été publié.Catalogue IEC - webstore.iec.ch/catalogue Electropedia - www.electropedia.org
Application autonome pour consulter tous les renseignements
Le premier dictionnaire en ligne de termes électroniques et
bibliographiques sur les Normes internationales,
électriques. Il contient 20 000 termes et définitions en anglais
Spécifications techniques, Rapports techniques et autres
et en français, ainsi que les termes équivalents dans 16
documents de l'IEC. Disponible pour PC, Mac OS, tablettes
langues additionnelles. Egalement appelé Vocabulaire
Android et iPad.
Electrotechnique International (IEV) en ligne.
Recherche de publications IEC - www.iec.ch/searchpub
Glossaire IEC - std.iec.ch/glossary
65 000 entrées terminologiques électrotechniques, en anglais
La recherche avancée permet de trouver des publications IEC
en utilisant différents critères (numéro de référence, texte, et en français, extraites des articles Termes et Définitions des
comité d’études,…). Elle donne aussi des informations sur les publications IEC parues depuis 2002. Plus certaines entrées
projets et les publications remplacées ou retirées. antérieures extraites des publications des CE 37, 77, 86 et
CISPR de l'IEC.IEC Just Published - webstore.iec.ch/justpublished
Service Clients - webstore.iec.ch/csc
Restez informé sur les nouvelles publications IEC. Just
Published détaille les nouvelles publications parues. Si vous désirez nous donner des commentaires sur cette
Disponible en ligne et aussi une fois par mois par email. publication ou si vous avez des questions contactez-nous:
csc@iec.ch.---------------------- Page: 2 ----------------------
IEC 60601-2-2
Edition 6.0 2017-03
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
Medical electrical equipment –
Part 2-2: Particular requirements for the basic safety and essential performance
of high frequency surgical equipment and high frequency surgical accessories
Appareils électromédicaux –
Partie 2-2: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils d'électrochirurgie à courant haute fréquence et des
accessoires d'électrochirurgie à courant haute fréquence
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
INTERNATIONALE
ICS 11.040.30 ISBN 978-2-8322-4008-3
Warning! Make sure that you obtained this publication from an authorized distributor.
Attention! Veuillez vous assurer que vous avez obtenu cette publication via un distributeur agréé.
® Registered trademark of the International Electrotechnical CommissionMarque déposée de la Commission Electrotechnique Internationale
---------------------- Page: 3 ----------------------
– 2 – IEC 60601-2-2:2017 IEC 2017
CONTENTS
FOREWORD ........................................................................................................................... 4
INTRODUCTION ..................................................................................................................... 7
201.1 Scope, object and related standards ......................................................................... 8
201.2 Normative references ................................................................................................ 9
201.3 Terms and definitions .............................................................................................. 10
201.4 General requirements .............................................................................................. 14
201.5 General requirements for testing of ME EQUIPMENT ................................................... 15
201.6 Classification of ME EQUIPMENT and ME SYSTEMS ...................................................... 15
201.7 ME EQUIPMENT identification, marking and documents .............................................. 15
201.8 Protection against electrical HAZARDS from ME EQUIPMENT ........................................ 20
201.9 Protection against MECHANICAL HAZARDS of ME EQUIPMENT and ME SYSTEMS .............. 36
201.10 Protection against unwanted and excessive radiation HAZARDS ................................ 36
201.11 Protection against excessive temperatures and other HAZARDS ................................ 36
201.12 Accuracy of controls and instruments and protection against hazardousoutputs .................................................................................................................... 38
201.13 HAZARDOUS SITUATIONS and fault conditions for ME EQUIPMENT ................................. 43
201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) ............................................... 44
201.15 Construction of ME EQUIPMENT ................................................................................. 44
201.16 ME SYSTEMS ............................................................................................................ 49
201.17 Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS ............................. 49
202 * ELECTROMAGNETIC DISTURBANCES – Requirements and tests .................................. 49
208 General requirements, tests and guidance for alarm systems in medicalelectrical equipment and medical electrical systems ................................................ 50
Annexes ............................................................................................................................... 51
Annex AA (informative) Particular guidance and rationale .................................................... 52
Annex BB (informative) ELECTROMAGNETIC DISTURBANCES created by HF SURGICALEQUIPMENT ............................................................................................................................ 78
Bibliography .......................................................................................................................... 87
Index of defined terms used in this particular standard .......................................................... 89
Figure 201.101 – Symbol used with an EARTH REFERENCED PATIENT CIRCUIT ........................... 16
Figure 201.102 – Symbol used with a HF ISOLATED PATIENT CIRCUIT ....................................... 16
Figure 201.103 – Circuit suitable for testing compliance to 201.8.4.101 ................................ 22
Figure 201.104 – Measurement of HF LEAKAGE CURRENT for EARTH REFERENCED PATIENTCIRCUITS and load between electrodes .................................................................................. 25
Figure 201.105 – Measurement of HF LEAKAGE CURRENT for EARTH REFERENCED PATIENTCIRCUITS and a load resistance from ACTIVE ELECTRODE to earth ............................................ 26
Figure 201.106 – Measurement of HF LEAKAGE CURRENT for HF ISOLATED PATIENTCIRCUITS ................................................................................................................................ 27
Figure 201.107 – Measurement of HF LEAKAGE CURRENT from a BIPOLAR ACCESSORY .............. 28
Figure 201.108 – Test apparatus for anchorages of cords of ACTIVE ACCESSORY ................... 34
Figure 201.109 – Measurement of output power – MONOPOLAR output ................................... 39
Figure 201.110 – Measurement of output power – BIPOLAR output ......................................... 40
---------------------- Page: 4 ----------------------IEC 60601-2-2:2017 IEC 2017 – 3 –
Figure 201.111 – Method of testing feedback from one active output to another in
simultaneous activation ......................................................................................................... 43
Figure AA.1 – Examples of various parts of an HF surgical ME SYSTEM ................................... 54
Figure AA.2 – Example of MONOPOLAR method of HF surgery using a NEUTRALELECTRODE ............................................................................................................................ 54
Figure AA.3 – Example of BIPOLAR method of HF surgery ...................................................... 55
Figure AA.4 – CREST FACTOR vs. peak voltage ...................................................................... 60
Figure AA.5 – Example of PATIENT circuit with NEUTRAL ELECTRODE referenced to earth
at operating frequencies ....................................................................................................... 64
Figure BB.1 – E-FIELD EMISSIONS test setup ........................................................................... 81
Figure BB.2 – H-FIELD EMISSIONS test setup .......................................................................... 82
Figure BB.3 – Conducted EMISSIONS test setup ..................................................................... 83
Figure BB.4 – Unit ad hoc test .............................................................................................. 85
Figure BB.5 – Power cord ad hoc test ................................................................................... 86
Figure BB.6 – ACCESSORY cord ad hoc test ........................................................................... 86
Table 201.101 – Colours of indicator lights and their meaning for HF SURGICALEQUIPMENT ............................................................................................................................ 16
Table 201.102 – Maximum output powers in SINGLE FAULT CONDITIONS .................................. 42
Table 201.103 – Test currents by weight range ..................................................................... 46
Table AA.1 – Summary of measured current and durations for 25 TUR procedures ............... 73
Table AA.2 – Summary of measured currents and durations for general surgicalprocedures............................................................................................................................ 74
Table BB.1 – Worst case EMISSIONS of spark gap type HF SURGICAL EQUIPMENT ..................... 84
Table BB.2 – Worst case EMISSIONS of non-spark gap (modern) HF SURGICAL EQUIPMENT ....... 84
---------------------- Page: 5 ----------------------– 4 – IEC 60601-2-2:2017 IEC 2017
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-2: Particular requirements for the basic safety and essential
performance of high frequency surgical equipment and
high frequency surgical accessories
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.5) IEC provides no marking procedure to indicate its approval and cannot be rendered responsible for any
equipment declared to be in conformity with an IEC Publication.6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International standard IEC 60601-2-2 has been prepared by IEC subcommittee 62D:Electromedical equipment, of IEC technical committee 62: Electrical equipment in medical
practice.This sixth edition cancels and replaces the fifth edition published in 2009. This edition
constitutes a technical revision. This edition includes the following significant technical
changes with respect to the previous edition:– refinement and additions to the defined terms;
– additional separation of the requirements for HF surgical equipment and HF surgical
accessories;– a new requirement for adult neutral electrodes to be contact quality monitoring neutral
electrodes;– new requirements for devices that have or use a high current mode.
---------------------- Page: 6 ----------------------
IEC 60601-2-2:2017 IEC 2017 – 5 –
The text of this particular standard is based on the following documents:
FDIS Report on voting
62D/1427/FDIS 62D/1442/RVD
Full information on the voting for the approval of this particular standard can be found in the
report on voting indicated in the above table.This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
In this standard, the following print types are used:– requirements and definitions: roman type;
– test specifications: italic type;
– informative material appearing outside of tables, such as notes, examples and references: in smaller type.
Normative text of tables is also in a smaller type;– TERMS DEFINED IN CLAUSE 3 OF THE GENERAL STANDARD, IN THIS PARTICULAR STANDARD OR AS
NOTED: SMALL CAPITALS.In referring to the structure of this standard, the term
– “clause” means one of the seventeen numbered divisions within the table of contents,
inclusive of all subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
– “subclause” means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all
subclauses of Clause 7).References to clauses within this standard are preceded by the term “Clause” followed by the
clause number. References to subclauses within this standard are by number only.In this standard, the conjunctive “or” is used as an “inclusive or” so a statement is true if any
combination of the conditions is true.The verbal forms used in this standard conform to usage described in Clause 7 of the ISO/IEC
Directives, Part 2. For the purposes of this standard, the auxiliary verb:– “shall” means that compliance with a requirement or a test is mandatory for compliance
with this standard;– “should” means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this standard;– “may” is used to describe a permissible way to achieve compliance with a requirement or
test.An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title
indicates that there is guidance or rationale related to that item in Annex AA.A list of all parts of the IEC 60601 series, published under the general title Medical electrical
equipment, can be found on the IEC website.---------------------- Page: 7 ----------------------
– 6 – IEC 60601-2-2:2017 IEC 2017
The committee has decided that the contents of this publication will remain unchanged until
the stability date indicated on the IEC website under "http://webstore.iec.ch" in the data
related to the specific publication. At this date, the publication will be• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
---------------------- Page: 8 ----------------------
IEC 60601-2-2:2017 IEC 2017 – 7 –
INTRODUCTION
The minimum safety requirements specified in this particular standard are considered to
provide for a practical degree of safety in the operation of HIGH FREQUENCY SURGICAL
EQUIPMENT.This particular standard amends and supplements IEC 60601-1:2005 and Amendment 1:2012,
Medical electrical equipment – Part 1: General requirements for basic safety and essential
performance, hereinafter referred to as the general standard (see 201.1.4).The requirements are followed by specifications for the relevant tests.
A "Particular guidance and rationale" section giving some explanatory notes, where
appropriate, about the more important requirements is included in Annex AA.Clauses or subclauses for which there are explanatory notes in Annex AA are marked with an
asterisk (*).It is considered that a knowledge of the reasons for these requirements will not only facilitate
the proper application of the standard but will, in due course, expedite any revision neces-
sitated by changes in clinical practice or as a result of developments in technology. However,
this annex does not form part of the requirements of this document.---------------------- Page: 9 ----------------------
– 8 – IEC 60601-2-2:2017 IEC 2017
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-2: Particular requirements for the basic safety and essential
performance of high frequency surgical equipment and
high frequency surgical accessories
201.1 Scope, object and related standards
Clause 1 of the general standard applies, except as follows:
201.1.1 * Scope
Replacement:
This part of IEC 60601 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of HF
SURGICAL EQUIPMENT and HF SURGICAL ACCESSORIES as defined in 201.3.224 and 201.3.223.
HF SURGICAL EQUIPMENT having a RATED OUTPUT POWER not exceeding 50 W (for example for
micro-COAGULATION, or for use in dentistry or ophthalmology) is exempt from certain of the
requirements of this particular standard. These exemptions are indicated in the relevant
requirements.201.1.2 Object
Replacement:
The object of this particular standard is to establish particular BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements for HF SURGICAL EQUIPMENT and HF SURGICAL ACCESSORIES as
defined in 201.3.224 and 201.3.223.201.1.3 Collateral standards
Addition:
This particular standard refers to those applicable collateral standards that are listed in
Clause 2 of the general standard and Clause 201.2 of this particular standard.IEC 60601-1-2:2014 and IEC 60601-1-8:2006 apply as modified in Clauses 202 and 208
respectively. IEC 60601-1-3, IEC 60601-1-10 and IEC 60601-1-11 do not apply. All other
published collateral standards in the IEC 60601-1 series apply as published.201.1.4 Particular standards
Replacement:
In the IEC 60601 series, particular standards may modify, replace or delete requirements
contained in the general standard and collateral standards as appropriate for the particular
ME EQUIPMENT under consideration, and may add other BASIC SAFETY and ESSENTIALPERFORMANCE requirements.
______________
The general standard is IEC 60601-1:2005/AMD1:2012, Medical electrical equipment – Part 1: General
requirements for basic safety and essential performance.---------------------- Page: 10 ----------------------
IEC 60601-2-2:2017 IEC 2017 – 9 –
A requirement of a particular standard takes priority over the general standard.
For brevity, IEC 60601-1 is referred to in this particular standard as the general standard.
Collateral standards are referred to by their document number.The numbering of clauses and subclauses of this particular standard corresponds to that of
the general standard with the prefix “201” (e.g. 201.1 in this document addresses the content
of Clause 1 of the general standard) or applicable collateral standard with the prefix “20x”
where x is the final digit(s) of the collateral standard document number (e.g. 202.4 in this
particular standard addresses the content of Clause 4 of the IEC 60601-1-2 collateral
standard, 203.4 in this particular standard addresses the content of Clause 4 of the
IEC 60601-1-3 collateral standard, etc.). The changes to the text of the general standard are
specified by the use of the following words:"Replacement" means that the clause or subclause of the general standard or applicable
collateral standard is replaced completely by the text of this particular standard.
"Addition" means that the text of this particular standard is additional to the requirements of
the general standard or applicable collateral standard."Amendment" means that the clause or subclause of the general standard or applicable
collateral standard is amended as indicated by the text of this particular standard.
Subclauses, figures or tables which are additional to those of the general standard are
numbered starting from 201.101. However, due to the fact that definitions in the general
standard are numbered 3.1 through 3.147, additional definitions in this document are
numbered beginning from 201.3.201. Additional annexes are lettered AA, BB, etc., and
additional items aa), bb), etc.Subclauses, figures or tables which are additional to those of a collateral standard are
numbered starting from 20x, where “x” is the number of the collateral standard, e.g. 202 for
IEC 60601-1-2, 203 for IEC 60601-1-3, etc.The term "this document" is used to make reference to the general standard, any applicable
collateral standards and this particular standard taken together.Where there is no corresponding clause or subclause in this particular standard, the clause or
subclause of the general standard or applicable collateral standard, although possibly not
relevant, applies without modification; where it is intended that any part of the general
standard or applicable collateral standard, although possibly relevant, is not to be applied, a
statement to that effect is given in this particular standard.201.2 Normative references
NOTE Informative references are listed in the bibliography beginning on page 87.
Clause 2 of the general standard applies, except as follows:
Replacement:
IEC 60601-1-2:2014, Medical electrical equipment – Part 1-2: General requirements for basic
safety and essential performance – Collateral Standard: Electromagnetic disturbances –
Requirements and testsIEC 60601-1-8:2006, Medical electrical equipment – Part 1-8: General requirements for basic
safety and essential performance – Collateral Standard: General requirements, tests and
guidance for alarm systems in medical electrical equipment and medical electrical systems
---------------------- Page: 11 ----------------------– 10 – IEC 60601-2-2:2017 IEC 2017
Addition:
CISPR 11:2015, Industrial, scientific and medical equipment – Radio-frequency disturbance
characteristics – Limits and methods of measurementIEC 61000-4-3:2006, Electromagnetic compatibility (EMC) – Part 4-3: Testing and
measurement techniques – Radiated, radio-frequency electromagnetic field immunity test
IEC 61000-4-6:2013, Electromagnetic compatibility (EMC) – Part 4-6: Testing andmeasurement techniques – Immunity to conducted disturbances, induced by radio-frequency
fields201.3 Terms and definitions
For the purposes of this document, the terms and definitions given in IEC 60601-1 and the
following apply.ISO and IEC maintain terminological databases for use in standardization at the following
addresses:• IEC Electropedia: available at http://www.electropedia.org/
• ISO Online browsing platform: available at http://www.iso.org/obp
Replace NOTE 1 with the following:
NOTE 1 Where the terms “voltage” and “current” are used in this document, they mean the RMS values of an
alternating, direct or composite voltage or current averaged over 1 s unless stated otherwise.
Addition:201.3.201
ACTIVE ACCESSORY
HF SURGICAL ACCESSORY intended for manipulation by the OPERATOR to produce an effect by
electrical conduction adjacent to the ACTIVE ELECTRODE at the intended site on the PATIENT,
generally comprising an ACTIVE HANDLE, the cord of an ACTIVE ACCESSORY, ACTIVE CONNECTOR
and ACTIVE...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.