IEC 60601-2-64:2025
(Amendment)Medical electrical equipment - Part 2-64: Particular requirements for the basic safety and essential performance of light ion beam medical electrical equipment
Medical electrical equipment - Part 2-64: Particular requirements for the basic safety and essential performance of light ion beam medical electrical equipment
IEC 60601-2-64:2025 applies to the BASIC SAFETY and essential performance of LIGHT ION BEAM ME EQUIPMENT, hereafter referred to as ME EQUIPMENT, used for treatment of patients. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
This document, with the inclusion of TYPE TESTS and SITE TESTS, applies respectively to the manufacturer and specified installation aspects of LIGHT ION BEAM ME EQUIPMENT
– intended for RADIOTHERAPY in human medical practice, including those in which the selection and DISPLAY of operating parameters can be controlled automatically by PROGRAMMABLE ELECTRONIC SUBSYSTEMS (PESS),
– that, in NORMAL USE, deliver a RADIATION BEAM of LIGHT IONS having ENERGY PER NUCLEON in the range 10 MeV/n to 500 MeV/n,
and
– intended to be
• for NORMAL USE, operated under the authority of appropriately licensed or QUALIFIED PERSONS by OPERATORS having the required skills for a particular medical application, for particular SPECIFIED clinical purposes maintained in accordance with the recommendations given in the INSTRUCTIONS FOR USE,
• subject to regular quality assurance performance and calibration checks by a QUALIFIED PERSON.
IEC 60601-2-64:2025 cancels and replaces the first edition published in 2014. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a) harmonization with IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2011 and IEC 60601-1:2005/AMD2:2020;
b) harmonization with IEC 62667:2017 for defined terms and definitions;
c) address revision to neutrons outside the field of irradiation.
Appareils électromédicaux - Partie 2-64: Exigences particulières pour la sécurité de base et les performances essentielles des appareils électromédicaux par faisceau d'ions légers
L'IEC 60601-2-64:2025 s'applique à la SECURITE DE BASE et aux PERFORMANCES ESSENTIELLES des APPAREILS ELECTROMEDICAUX PAR FAISCEAU D'IONS LEGERS, désignés ci-après par le terme APPAREILS EM, utilisés pour le traitement des PATIENTS. Si un article ou un paragraphe est spécifiquement destiné à être applicable uniquement aux APPAREILS EM, ou uniquement aux SYSTEMES EM, le titre et le contenu de cet article ou de ce paragraphe l'indiquent. Si cela n'est pas le cas, l'article ou le paragraphe s'applique à la fois aux APPAREILS EM et aux SYSTEMES EM, selon le cas.
Le présent document, y compris les ESSAIS DE TYPE et les ESSAIS SUR LE SITE, s'applique respectivement au FABRICANT et à certains aspects d'installation des APPAREILS ELECTROMEDICAUX PAR FAISCEAU D'IONS LEGERS
– destinés à être utilisés à des fins de RADIOTHERAPIE dans la pratique médicale des patients, y compris les appareils dont le choix et l'AFFICHAGE des paramètres de fonctionnement peuvent être contrôlés automatiquement par des SOUS-SYSTEMES ELECTRONIQUES PROGRAMMABLES (PESS),
– qui, en UTILISATION NORMALE, délivrent un FAISCEAU DE RAYONNEMENT d'IONS LEGERS dont l'ENERGIE PAR NUCLEON se situe dans la plage 10 MeV/n à 500 MeV/n,
et
– destinés à être
• pour une UTILISATION NORMALE, manipulés, sous la responsabilité de PERSONNES QUALIFIEES ou dûment autorisées par des OPERATEURS qui disposent des compétences exigées pour une application médicale particulière, à des fins cliniques SPECIFIEES particulières, entretenus conformément aux recommandations données dans les INSTRUCTIONS D'UTILISATION,
• soumis à des contrôles périodiques de performance de l'assurance qualité et d'étalonnage par une PERSONNE QUALIFIEE.
L'IEC 60601-2-64:2025 annule et remplace la première édition parue en 2014. Cette édition constitue une révision technique.
Cette édition inclut les modifications techniques majeures suivantes par rapport à l'édition précédente:
a) harmonisation avec l'IEC 60601-1:2005, l'IEC 60601-1:2005/AMD1:2011 et l'IEC 60601-1:2005/AMD2:2020;
b) harmonisation avec l'IEC 62667:2017 pour les termes définis et les définitions;
c) révision de la section relative aux neutrons en dehors du champ d'irradiation.
General Information
Relations
Overview
IEC 60601-2-64:2025 - "Medical electrical equipment - Part 2-64" - defines the basic safety and essential performance requirements for light ion beam medical electrical equipment used in human radiotherapy. This second edition (2025) replaces the 2014 edition and is a technical revision harmonized with recent updates to the IEC 60601‑1 series and IEC 62667. The standard applies to ME equipment and ME systems that, in normal use, deliver light-ion radiation beams with energy per nucleon from 10 MeV/n to 500 MeV/n, and it covers both manufacturer (type tests) and installation (site tests) responsibilities.
Key topics and technical requirements
- Scope and applicability
- Covers devices intended for clinical radiotherapy operated by licensed or qualified personnel and maintained by qualified persons.
- Safety and essential performance
- Requirements derived from IEC 60601‑1 general clauses, adapted for light ion beam hazards and system complexities.
- Type tests and site tests
- Distinguishes tests performed by manufacturers (type tests) and those required at installation (site tests) with associated documentation and reporting.
- Programmable Electronic Subsystems (PESS)
- Addresses automated control and display of operating parameters, including requirements for PESS used in clinical control and safety functions.
- Radiation protection and interlocks
- Specifies limits and protective measures for unwanted or excessive radiation; Clause 201.10 defines conditions where interlocks must interrupt or terminate irradiation to preserve essential performance.
- Neutron considerations
- Includes revised guidance on neutrons outside the treatment field (technical change from the previous edition).
- Labeling, documentation and usability
- Identification, markings, movement designations (per IEC 61217 guidance), and instructions for use to support safe operation and clinical workflows.
- Quality assurance
- Equipment is required to be subject to regular QA performance and calibration checks by qualified personnel.
Practical applications - who uses this standard
- Medical device manufacturers - to design, test, and certify light ion beam therapy systems for compliance.
- Clinical engineers and medical physicists - for commissioning, site testing, QA, and calibration protocols.
- Hospital procurement and radiation safety officers - to evaluate equipment safety and installation requirements.
- Regulators and notified bodies - for conformity assessment and approval of radiotherapy ME equipment.
- System integrators and installers - to perform and document required site tests and safety interlocks.
Related standards
- IEC 60601‑1:2005 and amendments (general requirements for basic safety and essential performance)
- IEC 62667:2017 (terminology and test/reporting formats for light ion beam equipment)
- IEC 61217 (guidance on equipment movements and marking)
- IEC 60601‑2‑68 (x‑ray image guidance - under development)
This standard is essential reading for stakeholders responsible for the safe design, installation, operation and regulatory compliance of light ion beam radiotherapy systems.
Frequently Asked Questions
IEC 60601-2-64:2025 is a standard published by the International Electrotechnical Commission (IEC). Its full title is "Medical electrical equipment - Part 2-64: Particular requirements for the basic safety and essential performance of light ion beam medical electrical equipment". This standard covers: IEC 60601-2-64:2025 applies to the BASIC SAFETY and essential performance of LIGHT ION BEAM ME EQUIPMENT, hereafter referred to as ME EQUIPMENT, used for treatment of patients. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. This document, with the inclusion of TYPE TESTS and SITE TESTS, applies respectively to the manufacturer and specified installation aspects of LIGHT ION BEAM ME EQUIPMENT – intended for RADIOTHERAPY in human medical practice, including those in which the selection and DISPLAY of operating parameters can be controlled automatically by PROGRAMMABLE ELECTRONIC SUBSYSTEMS (PESS), – that, in NORMAL USE, deliver a RADIATION BEAM of LIGHT IONS having ENERGY PER NUCLEON in the range 10 MeV/n to 500 MeV/n, and – intended to be • for NORMAL USE, operated under the authority of appropriately licensed or QUALIFIED PERSONS by OPERATORS having the required skills for a particular medical application, for particular SPECIFIED clinical purposes maintained in accordance with the recommendations given in the INSTRUCTIONS FOR USE, • subject to regular quality assurance performance and calibration checks by a QUALIFIED PERSON. IEC 60601-2-64:2025 cancels and replaces the first edition published in 2014. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) harmonization with IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2011 and IEC 60601-1:2005/AMD2:2020; b) harmonization with IEC 62667:2017 for defined terms and definitions; c) address revision to neutrons outside the field of irradiation.
IEC 60601-2-64:2025 applies to the BASIC SAFETY and essential performance of LIGHT ION BEAM ME EQUIPMENT, hereafter referred to as ME EQUIPMENT, used for treatment of patients. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. This document, with the inclusion of TYPE TESTS and SITE TESTS, applies respectively to the manufacturer and specified installation aspects of LIGHT ION BEAM ME EQUIPMENT – intended for RADIOTHERAPY in human medical practice, including those in which the selection and DISPLAY of operating parameters can be controlled automatically by PROGRAMMABLE ELECTRONIC SUBSYSTEMS (PESS), – that, in NORMAL USE, deliver a RADIATION BEAM of LIGHT IONS having ENERGY PER NUCLEON in the range 10 MeV/n to 500 MeV/n, and – intended to be • for NORMAL USE, operated under the authority of appropriately licensed or QUALIFIED PERSONS by OPERATORS having the required skills for a particular medical application, for particular SPECIFIED clinical purposes maintained in accordance with the recommendations given in the INSTRUCTIONS FOR USE, • subject to regular quality assurance performance and calibration checks by a QUALIFIED PERSON. IEC 60601-2-64:2025 cancels and replaces the first edition published in 2014. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) harmonization with IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2011 and IEC 60601-1:2005/AMD2:2020; b) harmonization with IEC 62667:2017 for defined terms and definitions; c) address revision to neutrons outside the field of irradiation.
IEC 60601-2-64:2025 is classified under the following ICS (International Classification for Standards) categories: 11.040.60 - Therapy equipment. The ICS classification helps identify the subject area and facilitates finding related standards.
IEC 60601-2-64:2025 has the following relationships with other standards: It is inter standard links to IEC 60601-2-64:2014. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
You can purchase IEC 60601-2-64:2025 directly from iTeh Standards. The document is available in PDF format and is delivered instantly after payment. Add the standard to your cart and complete the secure checkout process. iTeh Standards is an authorized distributor of IEC standards.
Standards Content (Sample)
IEC 60601-2-64 ®
Edition 2.0 2025-12
INTERNATIONAL
STANDARD
REDLINE VERSION
Medical electrical equipment -
Part 2-64: Particular requirements for the basic safety and essential performance
of light ion beam medical electrical equipment
ICS 11.040.60 ISBN 978-2-8327-0929-0
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or
by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from either
IEC or IEC's member National Committee in the country of the requester. If you have any questions about IEC copyright
or have an enquiry about obtaining additional rights to this publication, please contact the address below or your local
IEC member National Committee for further information.
IEC Secretariat Tel.: +41 22 919 02 11
3, rue de Varembé info@iec.ch
CH-1211 Geneva 20 www.iec.ch
Switzerland
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.
About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigendum or an amendment might have been published.
IEC publications search - IEC Products & Services Portal - products.iec.ch
webstore.iec.ch/advsearchform Discover our powerful search engine and read freely all the
The advanced search enables to find IEC publications by a publications previews, graphical symbols and the glossary.
variety of criteria (reference number, text, technical With a subscription you will always have access to up to date
committee, …). It also gives information on projects, content tailored to your needs.
replaced and withdrawn publications.
Electropedia - www.electropedia.org
The world's leading online dictionary on electrotechnology,
IEC Just Published - webstore.iec.ch/justpublished
Stay up to date on all new IEC publications. Just Published containing more than 22 500 terminological entries in English
details all new publications released. Available online and and French, with equivalent terms in 25 additional languages.
once a month by email. Also known as the International Electrotechnical Vocabulary
(IEV) online.
IEC Customer Service Centre - webstore.iec.ch/csc
If you wish to give us your feedback on this publication or
need further assistance, please contact the Customer
Service Centre: sales@iec.ch.
CONTENTS
FOREWORD . 2
INTRODUCTION . 1
201.1 Scope, object and related standards . 6
201.2 Normative references . 8
201.3 Terms and definitions . 8
201.4 General requirements . 16
201.5 General requirements for testing ME EQUIPMENT . 16
201.6 Classification of ME EQUIPMENT and ME SYSTEMS . 17
201.7 ME EQUIPMENT identification, marking and documents . 18
201.8 Protection against electrical HAZARDS from ME EQUIPMENT . 21
201.9 Protection against MECHANICAL HAZARDS of ME EQUIPMENT and ME SYSTEMS . 21
201.10 Protection against unwanted and excessive radiation HAZARDS . 27
201.11 Protection against excessive temperatures and other HAZARDS . 49
201.12 Accuracy of controls and instruments and protection against hazardous
outputs . 49
201.13 HAZARDOUS SITUATIONS and fault conditions for ME EQUIPMENT . 49
201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) . 50
201.15 Construction of ME EQUIPMENT . 50
201.16 ME SYSTEMS. 50
201.17 ELECTROMAGNETIC DISTURBANCES of ME EQUIPMENT and ME SYSTEMS . 52
201.101 ELECTRONIC IMAGING DEVICES (EID) . 53
201.102 Operation of ME EQUIPMENT parts from outside the facility . 53
206 Usability . 53
Annexes . 57
Annex B (informative) Sequence of testing . 57
Annex I (informative) ME SYSTEMS aspects . 57
Bibliography . 58
Index of defined terms used in this document . 59
Figure 201.101 – PATIENT SUPPORT POSITIONER movements . 54
Figure 201.102 – Diagram illustrating example RADIATION HEAD components and
possible PATIENT position for NON-PRIMARY RADIATION REQUIREMENTS . 55
Figure 201.103 – Diagram illustrating distance along PATIENT plane to measure
NON-PRIMARY RADIATION ABSORBED DOSE . 56
Table 201.101 – Data required in the technical description to support Clause 201.10
SITE TEST compliance . 19
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
Medical electrical equipment -
Part 2-64: Particular requirements for the basic safety and essential
performance of light ion beam medical electrical equipment
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote international
co-operation on all questions concerning standardization in the electrical and electronic fields. To this end and
in addition to other activities, IEC publishes International Standards, Technical Specifications, Technical Reports,
Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC Publication(s)”). Their
preparation is entrusted to technical committees; any IEC National Committee interested in the subject dealt with
may participate in this preparatory work. International, governmental and non-governmental organizations liaising
with the IEC also participate in this preparation. IEC collaborates closely with the International Organization for
Standardization (ISO) in accordance with conditions determined by agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence between
any IEC Publication and the corresponding national or regional publication shall be clearly indicated in the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) IEC draws attention to the possibility that the implementation of this document may involve the use of (a)
patent(s). IEC takes no position concerning the evidence, validity or applicability of any claimed patent rights in
respect thereof. As of the date of publication of this document, IEC had not received notice of (a) patent(s), which
may be required to implement this document. However, implementers are cautioned that this may not represent
the latest information, which may be obtained from the patent database available at https://patents.iec.ch. IEC
shall not be held responsible for identifying any or all such patent rights.
This redline version of the official IEC Standard allows the user to identify the changes made
to the previous edition IEC 60601-2-64:2014. A vertical bar appears in the margin wherever a
change has been made. Additions are in green text, deletions are in strikethrough red text.
IEC 60601-2-64 has been prepared by subcommittee 62C: Equipment for radiotherapy, nuclear
medicine and radiation dosimetry, of IEC technical committee 62: Medical equipment, software,
and systems. It is an International Standard.
This second edition cancels and replaces the first edition published in 2014. This edition
constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous
edition:
a) harmonization with IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2011 and
IEC 60601-1:2005/AMD2:2020;
b) harmonization with IEC 62667:2017 for defined terms and definitions;
c) address revision to neutrons outside the field of irradiation.
The text of this International Standard is based on the following documents:
Draft Report on voting
62C/954/FDIS 62C/964/RVD
Full information on the voting for its approval can be found in the report on voting indicated in
the above table.
The language used for the development of this International Standard is English.
This document was drafted in accordance with ISO/IEC Directives, Part 2, and developed in
accordance with ISO/IEC Directives, Part 1 and ISO/IEC Directives, IEC Supplement, available
at www.iec.ch/members_experts/refdocs. The main document types developed by IEC are
described in greater detail at www.iec.ch/publications.
In this document, the following print types are used:
– requirements and definitions: roman type.
– test specifications: italic type;
– informative material appearing outside of tables, such as notes, examples and references: in smaller type.
Normative text of tables is also in a smaller type;
– TERMS DEFINED IN CLAUSE 3 OF IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 AND
IEC 60601-1:2005/AMD2:2020, IN THIS PARTICULAR STANDARD OR AS NOTED: SMALL CAPITALS.
In referring to the structure of this document, the term
– "clause" means one of the numbered divisions within the table of contents, inclusive of all
subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
– "subclause" means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all
subclauses of Clause 7).
References to clauses within this document are preceded by the term "Clause" followed by the
clause number. References to subclauses within this particular standard are by number only.
In this document, the conjunctive "or" is used as an "inclusive or" so a statement is true if any
combination of the conditions is true.
The verbal forms used in this document conform to usage described in Clause 7 of the ISO/IEC
Directives, Part 2. For the purposes of this document, the auxiliary verb:
– "shall" means that compliance with a requirement or a test is mandatory for compliance with
this document;
– "should" means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this document;
– "may" is used to describe a permissible way to achieve compliance with a requirement or
test.
A list of all parts of the IEC 60601 series, published under the general title Medical electrical
equipment, can be found on the IEC website.
The committee has decided that the contents of this document will remain unchanged until the
stability date indicated on the IEC website under webstore.iec.ch in the data related to the
specific document. At this date, the document will be
– reconfirmed,
– withdrawn, or
– revised.
INTRODUCTION
The use of LIGHT ION BEAM ME EQUIPMENT for RADIOTHERAPY purposes may can expose PATIENTS
to danger if the ME EQUIPMENT fails to deliver the required dose to the PATIENT, or if the ME
EQUIPMENT design does not satisfy standards of electrical and mechanical safety. The ME
EQUIPMENT may can also cause danger to persons in the vicinity if the ME EQUIPMENT itself fails
to contain the RADIATION adequately or if there are inadequacies in the design of the TREATMENT
ROOM.
This particular standard establishes requirements to be complied with by MANUFACTURERS in the
design and construction of LIGHT ION BEAM ME EQUIPMENT for use in RADIOTHERAPY; it does not
attempt to define their optimum performance requirements. Its purpose is to identify those
features of design that are regarded, at the present time, as essential for the safe operation of
ME EQUIPMENT; it places limits on the degradation of ME EQUIPMENT performance beyond
such
which it can be presumed that a fault condition exists and where an INTERLOCK then operates
to prevent continued operation of the ME EQUIPMENT.
Clause 201.10 contains limits beyond which INTERLOCKS prevent, INTERRUPT or TERMINATE
IRRADIATION, cause INTERRUPTION OF IRRADIATION or cause TERMINATION OF IRRADIATION in order
to insure that ESSENTIAL PERFORMANCE is maintained and to avoid an unsafe condition. TYPE
that are performed by the MANUFACTURER, or SITE TESTS, which are not necessarily
TESTS
performed by the MANUFACTURER, are SPECIFIED for each requirement. It should be understood
that, before installation, a MANUFACTURER can provide a compliance certificate relating only to
TYPE TESTS. Data available from SITE TESTS should be incorporated in the ACCOMPANYING
DOCUMENTS DOCUMENTATION, in the form of a SITE TEST report, by those who test the ME
EQUIPMENT at installation.
Closely related to this document is IEC 62667 which is currently being developed. It specifies
test methods and reporting formats for performance tests of LIGHT ION BEAM ME EQUIPMENT for
use in RADIOTHERAPY, with the aim of providing uniform methods of doing so. The annex Annex
A of IEC 62667:2017 provides forms for presenting performance values, measured per the
methods SPECIFIED.
201.1 Scope, object and related standards
Clause 1 of the general standard IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and
IEC 60601-1:2005/AMD2:2020 applies, except as follows:
201.1.1 Scope
Replacement:
This document applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of LIGHT ION BEAM ME
EQUIPMENT, hereafter referred to as ME EQUIPMENT, used for treatment of PATIENTS.
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME
SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case,
the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
This document, with the inclusion of TYPE TESTS and SITE TESTS, applies respectively to the
manufacture MANUFACTURER and some SPECIFIED installation aspects of LIGHT ION BEAM ME
EQUIPMENT
– intended for RADIOTHERAPY in human medical practice, including those in which the selection
and DISPLAY of operating parameters can be controlled automatically by PROGRAMMABLE
ELECTRONIC SUBSYSTEMS (PESS),
– that, in NORMAL USE, deliver a RADIATION BEAM of LIGHT IONS having ENERGY PER NUCLEON in
the range 10 MeV/n to 500 MeV/n,
and
– intended to be
• for NORMAL USE, operated under the authority of appropriately licensed or QUALIFIED
PERSONS by OPERATORS having the required skills for a particular medical application, for
particular SPECIFIED clinical purposes maintained in accordance with the
recommendations given in the INSTRUCTIONS FOR USE,
• subject to regular quality assurance performance and calibration checks by a QUALIFIED
PERSON.
NOTE 1 In this document, all references to installation refer to installation in the RESPONSIBLE ORGANIZATION's
premises.
NOTE 2 In this document, all references to ABSORBED DOSE refer to ABSORBED DOSE in water.
NOTE 3 Information regarding x-ray image guidance can be found in IEC 60601-2-68 (under development).
NOTE 4 IEC 61217 gives guidance on the designation of ME EQUIPMENT movements, the marking of scales, their
zero positions and the direction of movement with increasing value (see 201.7.4.101).
201.1.2 Object
Replacement:
___________
The general standard is IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, Medical electrical equipment –
Part 1: General requirements for basic safety and essential performance
The object of this document is to establish particular BASIC SAFETY and ESSENTIAL PERFORMANCE
requirements for LIGHT ION BEAM ME EQUIPMENT in the range 10 MeV/n to 500 MeV/n and to
SPECIFY tests to check compliance to those requirements.
NOTE The adoption of this document helps to ensure that the ME EQUIPMENT
− maintains PATIENT safety during ME EQUIPMENT movements and failure of the SUPPLY MAINS;
− delivers the pre-selected RADIATION TYPE, ENERGY PER NUCLEON, LIGHT ION species, and ABSORBED DOSE;
− delivers pre-selected LIGHT ION BEAMS to the PATIENT, by utilizing LIGHT ION BEAM modifying devices, etc.,
without causing unnecessary risk to the PATIENT, the OPERATOR, other persons or the environment.
201.1.3 Collateral standards
Addition:
Collateral standards published after the date of publication of this standard shall only apply
subject to further amendment to this standard.
This particular standard refers to those applicable collateral standards that are listed in
Clause 2 of the general standard IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and
IEC 60601-1:2005/AMD2:2020 and Clause 201.2 of this document.
IEC 60601-1-6:2010, IEC 60601-1-6:2010/AMD1:2013 and IEC 60601-1-6:2010/AMD2:2020
applies as modified in Clause 206. IEC 60601-1-3, IEC 60601-1-8, IEC 60601-1-9 [1] and
3 4
IEC 60601-1-10 [2] do not apply. All other published collateral standards in the IEC 60601-
1 series apply as published.
NOTE Collateral standards published after the date of publication of this document will only apply subject to further
amendment to this document.
201.1.4 Particular standards
Replacement:
In the IEC 60601 series, particular standards may modify, replace or delete requirements
contained in the general standard IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and
IEC 60601-1:2005/AMD2:2020 or and collateral standards as appropriate for the particular
ME EQUIPMENT under consideration, and may add other BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements.
A requirement of a particular standard takes priority over the general standard IEC 60601-
1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020.
For brevity, IEC 60601-1 is referred to in this particular standard as the general standard.
Collateral standards are referred to by their document number.
The numbering of clauses and subclauses of this document corresponds to that of the general
standard IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and
IEC 60601-1:2005/AMD2:2020 with the prefix "201" (e.g. 201.1 in this document addresses the
content of Clause 1 of the general standard IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012
and IEC 60601-1:2005/AMD2:2020) or applicable collateral standard with the prefix "20x" where
x is the final digit(s) of the collateral standard document number (e.g. 202.4 addresses the
content of Clause 4 of the IEC 60601-1-2 collateral standard, 203.4 addresses the content of
___________
IEC 60601-1-9, Medical electrical equipment – Part 1-9: General requirements for basic safety and essential
performance – Collateral Standard: Requirements for environmentally conscious design
IEC 60601-1-10, Medical electrical equipment – Part 1-10: General requirements for basic safety and essential
performance – Collateral Standard: Requirements for the development of physiologic closed-loop controllers
Numbers in square brackets refer to the Bibliography.
Clause 4 of the 60601-1-3 collateral standard, etc.). The changes to the text of the general
standard IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and
IEC 60601-1:2005/AMD2:2020 are specified by the use of the following words:
"Replacement" means that the clause or subclause of the general standard IEC 60601-1:2005,
IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 or applicable collateral
standard is replaced completely by the text in this document.
"Addition" means that the text of this document is additional to the requirements of the general
standard IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-
1:2005/AMD2:2020 or applicable collateral standard.
"Amendment" means that the clause or subclause of the general standard IEC 60601-1:2005,
IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 or applicable collateral
standard is amended as indicated by the text of this document.
Subclauses, figures or tables which are additional to those of the general standard IEC 60601-
1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 are numbered
starting from 201.101. However, due to the fact that definitions in the general standard
IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 are
numbered 3.1 through 3.139154, additional definitions in this document are numbered
beginning from 201.3.201. Additional annexes are lettered AA, BB, etc., and additional items
aa), bb), etc.
Subclauses, figures or tables which are additional to those of a collateral standard are
numbered starting from 20x, where "x" is the number of the collateral standard, e.g. 202 for
IEC 60601-1-2, 203 for IEC 60601-1-3, etc.
The term “this standard” is used to make reference to the general standard, any applicable
collateral standards and this particular standard taken together.
Where there is no corresponding clause or subclause in this document, the clause or subclause
of the general standard IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and
IEC 60601-1:2005/AMD2:2020 or applicable collateral standard, although possibly not relevant,
applies without modification; where it is intended that any part of the general standard
IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 or
applicable collateral standard, although possibly relevant, is not to be applied, a statement to
that effect is given in this document.
201.2 Normative references
NOTE Informative references are listed in the bibliography.
IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020, Clause
2 of the general standard applies, except as follows:
Replacement:
IEC 60601-1-2:2014, Medical electrical equipment – Part 1-2: General requirements for basic
safety and essential performance – Collateral Standard: Electromagnetic compatibility –
Requirements and tests
Addition:
IEC 60601-1:2005, Medical electrical equipment - Part 1: General requirements for basic safety
and essential performance
IEC 60601-1:2005/AMD1:2012
IEC 60601-1:2005/AMD2:2020
IEC 60601-2-1:20092020, Medical electrical equipment - Part 2-1: Particular requirements for
the basic safety and essential performance of electron accelerators in the range 1 MeV to 50
MeV
IEC 60601-2-11:2013, Medical electrical equipment – Part 2-11: Particular requirements for the
basic safety and essential performance of gamma beam therapy equipment
IEC 60601-2-68:2014, Medical electrical equipment - Part 2-68: Particular requirements for the
basic safety and essential performance of x-ray-based image guided radiotherapy equipment
for use with electron accelerators, light ion beam therapy equipment and radionuclide beam
therapy equipment
IEC TR 60788:2004, Medical electrical equipment - Glossary of defined terms
IEC 61217:2011, Radiotherapy equipment - Coordinates, movements and scales
ISO/IEC 14165-321:2009, Information technology – Fibre channel – Part 321: Audio video (FC-
AV)
CISPR 11, Industrial, scientific and medical equipment - Radio-frequency disturbance
characteristics - Limits and methods of measurement
201.3 Terms and definitions
For the purposes of this document, the terms and definitions given in IEC 60601-1:2005+A1:
2012 IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020,
IEC 60601-2-1: 20092020, and IEC TR 60788:2004 apply, except as follows:
NOTE An index of defined terms is given after the Bibliography.
Additional terms and definitions:
201.3.201
APPLICATOR CARRIAGE
most distal part of the RADIATION HEAD that cannot be removed without using tools to which
interchangeable LIGHT ION BEAM APPLICATORS are attached and which may extend toward and
retract away from the ISOCENTRE or ERP
Note 1 to entry: Colloquially the APPLICATOR CARRIAGE has sometimes been called a snout.
201.3.202
BEAM FLUENCE DISTRIBUTION MONITOR
system to monitor directly or indirectly the FLUENCE distribution of the beam to provide beam
steering or lateral spreading information
Note 1 to entry: This monitor may be used as a surrogate monitor for the DOSE distribution delivered to the patient.
Note 2 to entry: Examples of BEAM FLUENCE DISTRIBUTION MONITORS include quadrant foil ionization chambers,
concentric ring ionization chambers, multi-strip ionization chambers, scintillator plates, and scanning magnet field
probes.
201.3.203
BEAM FLUX MONITOR
system to monitor the FLUX of the beam
___________
A newer edition of IEC 60601-2-68 was published in 2025.
Note 1 to entry: This monitor may be used as a surrogate monitor for the ABSORBED DOSE RATE delivered to the
patient.
201.3.204
BEAM GATING
allowance or inhibition of IRRADIATION and related equipment movements according to the status
provided by a BEAM GATING SIGNAL
201.3.205
BEAM GATING SIGNAL
signal generated for the purpose of BEAM GATING
EXAMPLE Examples include a respiratory spirometer, electrocardiogram, optical sensor, etc.
201.3.206
CONTROLLING TIMER
device to measure the time during which IRRADIATION occurs and, if a predetermined time is
reached, to TERMINATE cause TERMINATION OF IRRADIATION
[SOURCE: IEC 60601-2-1: 20092020, 201.3.202210]
201.3.207
DOSE MONITOR UNIT
parameter, reported by the DOSE MONITORING SYSTEM, from which, through a calibration
procedure and with additional information, the ABSORBED DOSE delivered can be calculated
201.3.208
DOSE MONITOR UNIT RATE
DOSE MONITOR UNIT per unit time
201.3.209
DOSE MONITOR UNIT RATE MONITORING SYSTEM
system of devices for the measurement and DISPLAY of a radiation quantity related to DOSE
MONITOR UNIT RATE
201.3.210
DOSE MONITORING SYSTEM
system of devices for the measurement and DISPLAY of a radiation quantity related to the
ABSORBED DOSE
201.3.211
ENERGY PER NUCLEON
total kinetic energy of the ion divided by the number of nucleons in the nucleus at the point
where the ion enters the RADIATION HEAD before passing through any beam modifiers
201.3.212
EQUIPMENT REFERENCE POINT
ERP
point in space used for referencing dimensions of equipment and performing dosimetry
measurements
Note 1 to entry: Typically, the reference point is coincident with the ISOCENTRE. If the beam delivery equipment is
not ISOCENTRIC, then the centre of the PATIENT alignment systems may be used.
Note 2 to entry: The corresponding note to entry in the French text indicates that the abbreviation "ERP" stands for
"EQUIPMENT REFERENCE POINT" in English.
201.3.213
FLUENCE
particles per unit area
quotient of dN by dA, where dN is the number of particles incident on a sphere of cross-sectional
area dA, thus
Φ = dN/dA
Note 1 to entry: See ICRU 33, ICRU 85a. Definition from ICRU 85a:2011, 3.1.3.
201.3.214
FLUX
particles per unit time
quotient of dN by dt, where dN is the increment of the particle number in the time interval dt,
thus
.
N = dN/dt
Note 1 to entry: See ICRU 33, ICRU 85a. Definition from ICRU 85a:2011, 3.1.2.
201.3.215
HARD-WIRED
term used condition of a system where the its features of a system can be modified only by
physically removing and re-routing linked wires
[SOURCE: IEC 60601-2-1: 20092020, 201.3.208224]
201.3.216
INTERRUPTION OF IRRADIATION
TO INTERRUPT IRRADIATION
stopping of/to stop IRRADIATION and movements with the possibility of continuing without
reselecting operating conditions
[SOURCE: IEC 60601-2-1: 20092020, 201.3.210227]
201.3.217
LATERAL SPREADING DEVICE
LSD
device used to increase the lateral (x , y Xg, Yg) dimensions of a small diameter LIGHT ION BEAM
b b
produced by an accelerator
EXAMPLE Examples of spreading devices include a thin metal foil for scattering the ions or a magnet to defocus
the beam or to scan the beam laterally across the intended TARGET VOLUME.
Note 1 to entry: X Xg and y Yg are defined in IEC 61217:2011.
b b
201.3.218
LIGHT ION
species of ion with an atomic number less than or equal to that of neon (Z ≤ 10) and SPECIFIED
by its number of protons, number of nucleons and ionization state
201.3.219
LIGHT ION BEAM
collection of LIGHT IONS travelling in the same general direction
201.3.220
LIGHT ION BEAM APPLICATOR
device for holding a BEAM LIMITING DEVICE or ACCESSORY close to the PATIENT's skin during
delivery of LIGHT ION BEAM
Note 1 to entry: Several LIGHT ION BEAM APPLICATORS may be available to reduce the weight of APERTURES lifted by
therapists the OPERATOR, decrease the aperture/bolus to skin distance from the skin to ACCESSORIES and reduce
leakage radiation.
201.3.221
LIGHT ION BEAM DISTRIBUTION SYSTEM
system of components and a control system used to transport the RADIATION from a RADIATION
SOURCE to several TREATMENT stations, experimental stations or beam dumps
EXAMPLE Examples of components include vacuum pipes, magnets, and steering coils.
201.3.222
LIGHT ION RANGE
PHANTOM most distant from its surface at which the ABSORBED DOSE is a SPECIFIED
depth in a
value, given in the ACCOMPANYING DOCUMENTS DOCUMENTATION, of the dose at the nominal
centre-of-modulation depth or of the dose maximum for a non-range-modulated beam, which is
measured on the RADIATION BEAM AXIS in a SPECIFIED RADIATION FIELD and with the surface of the
PHANTOM at a SPECIFIED distance from the ERP without RANGE SHIFTERS or ACCESSORIES installed
in the RADIATION HEAD downstream of the ENERGY PER NUCLEON or range monitoring system
201.3.223
MODULATED SCANNING
SCANNING MODE wherein a small diameter LIGHT ION BEAM is scanned across a target to create
a field large enough to cover the target such that the intended FLUENCE delivered to the PATIENT
is different at different lateral locations
Note 1 to entry: Various spatial and temporal scanning patterns may be used to generate the modulated FLUENCE
distribution.
201.3.224
NON-PRIMARY RADIATION
RADIATION emitted from the LIGHT ION BEAM ME EQUIPMENT that is not intended to treat the PATIENT
201.3.225
NORMAL USE
operation, including routine inspection and adjustments by the OPERATOR, and STAND-BY,
according to the instructions for use
Note 1 to entry: NORMAL USE includes PATIENT TREATMENT, equipment calibration, quality assurance procedures,
maintenance, and other procedures performed by the OPERATOR required in preparation for PATIENT TREATMENTS.
[SOURCE: IEC TR 60788:2004, NG, 10.08, modified – A note to entry has been added.]
201.3.226
OPERATOR
person utilizing an EQUIPMENT individually with or without the aid of an assistant, who controls
some or all functions of the EQUIPMENT in his presence
[SOURCE: IEC TR 60788:2004, rm-85-02]
201.3.227
PATIENT SUPPORT
assembly of ME EQUIPMENT that supports the PATIENT
[SOURCE: IEC TR 60601-2-1:2009, 201.3.215]
201.3.225
PATIENT POSITIONER
assembly of equipment upon which the PATIENT is placed and positioned for
RADIOTHERAPY ME EQUIPMENT
[SOURCE: IEC 60601-2-1:2020, 201.3.235, modified – Addition of “ME EQUIPMENT”.]
201.3.226
PATIENT POSITIONER TOP
component of the PATIENT POSITIONER to which registration and immobilization devices are
attached and upon which PATIENT is placed
Note 1 to entry: PATIENT POSITIONER TOPS are often exchangeable to hold PATIENTS or equipment in different
orientations, for instance, a flat board or seat.
[SOURCE: IEC 60601-2-1:2020, 201.3.254, modified – Main term “table top” replaced with
“patient positioner top”; beginning of the definition “device attached to the table top support”
replaced with “component of the PATIENT POSITIONER”; Note 1 replaced.]
201.3.228227
PORTAL
collection of one or more pre-programmed TREATMENT segments treated automatically with a
single PATIENT set-up
Note 1 to entry: Segments may consist of treatment beam IRRADIATION, motion of devices, or imaging.
201.3.229228
PRE-PROGRAMMED MOVEMENT
movement of ME EQUIPMENT parts that takes place according to a previously planned programme
without intervention by the OPERATOR during a PATIENT treatment or imaging
201.3.230229
PRIMARY/SECONDARY DOSE MONITORING COMBINATION
utilization of two DOSE MONITORING SYSTEMS where one is arranged to be the PRIMARY DOSE
MONITORING SYSTEM and the other the SECONDARY DOSE MONITORING SYSTEM
201.3.231230
PROGRAMMABLE RANGE MODULATED PORTAL
PRMP
LIGHT ION PORTAL in which a program is used, with or without a RANGE MODULATION DEVICE (RMD)
or program is used, to tailor the DEPTH DOSE distribution by varying the penetration and
weighting factors of several component segments
Note 1 to entry: Typically the DEPTH DOSE distribution is tailored to give a uniform dose distribution over the depths
where the TARGET (tumour) VOLUME resides.
Note 2 to entry: The corresponding note to entry in the French text indicates that the abbreviation "PRMP" stands
for "PROGRAMMABLE RANGE MODULATED PORTAL" in English.
201.3.231
QUALITY FACTOR
Q
QUALITY FACTOR at a point in tissue given by:
∞
Q = Q L D dL
( )
∫ L
L=0
D
where D is the ABSORBED DOSE at that point, D is the distribution of D in unrestricted linear
L
energy transfer L at the point of interest, and Q(L) is the QUALITY FACTOR as a function of L
Note 1 to entry: The integration is to be performed over D , due to all charged particles, excluding their secondary
L
electrons.
Note 2 to entry: The QUALITY FACTOR Q at a point in tissue relates the DOSE EQUIVALENT H at that point to the
ABSORBED DOSE D at that point, using the formula: H = Q × D.
[SOURCE: ICRP Publication 116:2010]
201.3.232
RADIATION HEAD
structure from which the RADIATION BEAM emerges
[SOURCE: IEC TR 60788:2004, rm-20-06]
201.3.233
RANGE MODULATION DEVICE
RMD
device used to modulate the penetration of a RADIATION BEAM into a PATIENT to tailor the DEPTH
DOSE distribution during the delivery of one PORTAL
Note 1 to entry: The device may consist of a propellor shaped material that spins in the beam, a filter containing a
repeating pattern of metallic ridges (e.g. ridge filter, mini-ridge filter, ripple filter), a cone or set of cones, or a set of
uniform thickness blocks programmable in a binary fashion. Sub-types include discrete and programmable.
Note 2 to entry: The corresponding note to entry in the French text indicates that the abbreviation "RMD" stands for
"RANGE MODULATION DEVICE" in English.
201.3.234
RANGE SHIFTER
range modifying device that has a constant thickness at all positions lateral to the central axis
of the beam
201.3.235
REDUNDANT DOSE MONITORING COMBINATION
utilization of two DOSE MONITORING SYSTEMS where both systems are arranged to TERMINATE
cause TERMINATION OF IRRADIATION according to the pre-selected number of DOSE MONITOR UNITS
201.3.236
SCANNING MODE
method of delivering a scanned RADIATION BEAM to generate a laterally broad field
Note 1 to entry: Types of SCANNING MODES include: none, UNIFORM SCANNING and MODULATED SCANNING.
201.3.237
SITE TEST
after installation, test of an individual device or ME EQUIPMENT to establish compliance with
SPECIFIED criteria
Note 1 to entry: It is understood that SITE TESTS shall be performed but may or may not be performed by the
MANUFACTURER, per the agreement between the MANUFACTURER and the end USER.
[SOURCE: IEC 60601-2-1:2009, 201.3.221, modified – The note to entry has been added
IEC 60601-2-1:2020, 201.3.253, modified – In Note 1 to entry, replacement of “responsible
organization” with “end user”.]
201.3.238
TABLE TOP
an exchangeable device attached to the patient positioner to which registration and
immobilization devices are attached and upon which the patient is placed.
201.3.239238
TECHNICAL DOCUMENTATION
documentation that enables the conformity of the product with the requirements of the
standard(s) to be assessed
Note 1 to entry: This may include schedule drawings when a certification body is involved.
Note 2 to entry: It covers the design, manufacture and operation of the product and may contain:
– a general description;
– design and manufacturing drawings and layouts of components, sub-assemblies, circuits, etc.;
– descriptions and explanations necessary for the understanding of drawings and layouts and the operation of the
product;
– a list of the standards referred to in the Ex certificate, applied in full or in part, and descriptions of the solutions
adopted to meet the requirements of the standards;
– results of design calculations made, examinations carried out, etc.;
– test reports
Note 3 to entry: The technical description includes information derived from the TECHNICAL DOCUMENTATION.
[SOURCE: ISO/IEC 80079-34:20112018, 3.1210, modified – A third note to entry has been
added Note 1 to entry to Note 3 to entry are removed.]
201.3.240239
TERMINATION OF IRRADIATION
TERMINATE IRRADIATION
stopping of/to stop IRRADIATION and movements, with no possibility of restarting without
returning to the PREPARATORY STATE
Note 1 to entry: Examples of events that may TERMINATE cause TERMINATION OF IRRADIATION and stop movements
include:
− when the pre-selected value of DOSE MONITOR UNITS is reached,
− when the pre-selected value is reached in the termination system independent of the DOSE MONITORING
SYSTEM,
− by a deliberate manual act,
− by the operation of an INTERLOCK, or
− in MODULATED SCANNING, when pre-programmed scanning pattern is completed.
[SOURCE: IEC 60601-2-11:2013, 201.3.225, modified – The wording of both the definition and
Note 1 to entry is slightly different.]
201.3.241240
TRANSMISSION DETECTOR
RADIATION DETECTOR through which the LIGHT ION BEAM passes
[SOURCE: IEC 60601-2-1:20092020, 201.3.226257, modified – The term "RADIATION BEAM" has
been replaced by "LIGHT ION BEAM".]
201.3.242241
TYPE TEST
test on a representative sample of the equipment with the objective of determining if the
equipment, as designed and manufactured, can meet the requirements of this document
[SOURCE: IEC 60601-1:2005, 3.135]
201.3.243242
UNIFORM SCANNING
SCANNING MODE wherein a RADIATION BEAM is scanned laterally in a pre-defined pattern to
produce a field large enough to treat the target such that the FLUX of the RADIATION BEAM is not
intentionally changed throughout the lateral scan so that the resultant dose distribution within
the field is uniform
Note 1 to entry: The scanning may be accomplished by changing the current in electromagnets through which the
RADIATION BEAM passes or the use of permanent magnets mechanically rotated or translated.
201.3.244243
USER
when used in an IEC standard on electromedical EQUIPMENT, organization or individual
responsible for the use and maintenance of the EQUIPMENT
[SOURCE: IEC TR 60788:2004, rm-85-01]
201.4 General requirements
IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2011 and IEC 60601-1:2005/AMD2:2020,
Clause 4 of the general standard applies.
201.5 General requirements for testing of ME EQUIPMENT
IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2011 and IEC 60601-1:2005/AMD2:2020,
Clause 5 of the general standard applies, except as follows:
201.5.1 TYPE TESTS
Replacement of the title:
201.5.1 TYPE TESTS and SITE TESTS
Additional subclause:
201.5.1.101 Test grades
Three grades of TYPE TEST and two grades of SITE TEST procedures are SPECIFIED in this
document. Their requirements are as follows:
– TYPE TEST grade A: An analysis of ME EQUIPMENT design, as related to the SPECIFIED
RADIATION safety provisions, which shall result in a statement included in the technical
description, regarding the working principles or constructional means by which the
requirement is fulfilled.
– TYPE TEST/SITE TEST grade B: Visual inspection or functional test or measurement of the ME
EQUIPMENT. The test shall be in accordance with the procedure SPECIFIED in this document
and shall be based on operating states, including fault condition states, which are
achievable only without interference with the circuitry or construction of the ME EQUIPMENT.
–
...
IEC 60601-2-64:2025 provides comprehensive guidelines concerning the BASIC SAFETY and essential performance of LIGHT ION BEAM MEDICAL ELECTRICAL EQUIPMENT (ME EQUIPMENT) used in patient treatment, specifically in the radiotherapy context. The standard delineates critical requirements that ensure the safe and effective operation of such equipment, underlining its significance to both manufacturers and healthcare providers. One of the primary strengths of IEC 60601-2-64:2025 is its meticulous scope, which encompasses both TYPE TESTS and SITE TESTS linked to the manufacturer and installation aspects of LIGHT ION BEAM ME EQUIPMENT. The document emphasizes the need for operation under licensed authorities and qualified personnel, which helps to ensure that the use of this medical equipment adheres to stringent safety standards. The standard addresses vital performance criteria that are relevant for LIGHT ION BEAM ME EQUIPMENT, detailing the specific energy range of the radiation beam and the operational requirements for proper medical application. This clarity in outlining the expected conditions of NORMAL USE, along with stringent quality assurance measures, enhances the reliability and safety of such equipment in clinical settings. Additionally, the 2025 revision introduces technical changes that harmonize the standard with previous editions and related standards, such as IEC 60601-1 and IEC 62667. This harmonization facilitates a cohesive framework for users, ensuring that terminologies and definitions are consistent, which is crucial for effective communication and implementation among operators and manufacturers alike. Furthermore, revisions addressing neutrons outside the field of irradiation reflect a proactive approach in adapting to advancements and ensuring up-to-date safety measures. Overall, IEC 60601-2-64:2025 stands as a vital document that enhances the understanding, safety, and performance of LIGHT ION BEAM ME EQUIPMENT, confirming its relevance in contemporary medical practices, particularly in radiotherapy.
표준 IEC 60601-2-64:2025는 의료 전기 장비, 특히 빛 이온 빔 의료 전기 장비에 대한 기본 안전성과 필수 성능 요건을 규정하고 있습니다. 이 문서는 환자 치료에 사용되는 ME 장비에 적용되며, 해당 장비가 어떻게 안전하고 효과적으로 작동해야 하는지를 명확히 합니다. 이 표준의 강점 중 하나는 LIGHT ION BEAM ME EQUIPMENT의 기본 안전성 및 성능 요구사항을 포괄적으로 지정하고 있다는 점입니다. 특히, 이 문서는 TYPE TESTS와 SITE TESTS를 포함함으로써 제조자와 설치 측면 모두를 체계적으로 고려하고 있습니다. 또한, PROGRAMMABLE ELECTRONIC SUBSYSTEMS (PESS)에 의해 자동으로 제어될 수 있는 운영 매개변수의 선택 및 표시를 다루는 점은 최신 기술 동향을 반영하고 있어 매우 유용합니다. IEC 60601-2-64:2025는 10 MeV/n에서 500 MeV/n의 에너지를 갖는 LIGHT IONS의 방사선을 전달할 수 있는 장비에 적용됩니다. 이는 방사선 치료를 위한 중요한 기준이 될 수 있으며, 해당 장비가 NORMAL USE에서 적절하게 작동할 수 있도록 요구 사항을 명확히 합니다. 또한, 면허가 있는 적절한 자격을 갖춘 운전자가 특정 임상 목적에 맞게 장비를 운용하고, 정기적인 품질 보증 성능 및 교정 체크를 포함하도록 요구하는 점도 주목할 만합니다. 이 표준은 2014년에 발행된 첫 번째 판을 취소하고 대체하며, 기술적 개정을 포함하고 있습니다. 특히 IEC 60601-1:2005 및 그 개정판들과의 조화를 이루고, IEC 62667:2017의 정의와 용어에 대한 일치를 이루는 점은 표준의 일관성과 신뢰성을 확보하는 데 기여합니다. 또한, 조사 범위를 넘어서는 중성자에 대한 개정이 이루어진 점도 중요한 발전입니다. 결론적으로, IEC 60601-2-64:2025는 임상에서의 안전하고 효과적인 빛 이온 빔 의료 장비의 사용을 위한 필수적이고 포괄적인 기준을 제공하는 중요한 문서로 자리 잡고 있습니다.
IEC 60601-2-64:2025は、光イオンビーム医療機器の基本的な安全性と本質的な性能に関する特定要件を規定しており、本標準は特に患者の治療に使用される医療電気機器(ME EQUIPMENT)に焦点を当てています。この標準は、光イオンビーム医療機器における基本的な安全性の確保を目的としており、医療分野におけるリスク管理と品質保証において重要な役割を果たします。 本標準の適用範囲は明確であり、特に放射線治療に関連する光イオンビーム医療機器に関して、製造業者と設置の観点から、タイプテストとサイトテストを含めています。これにより、ME EQUIPMENTの運用に必要な条件が詳細に示され、オペレーターは適切な資格を持つことが求められるため、安全性が高まります。 IEC 60601-2-64:2025の強みの一つは、IEC 60601-1:2005およびその改訂版との調和を図っている点です。これにより、国際的な標準に従った信頼性の高い安全基準が確立されるため、医療機器の使用中のリスクを軽減することが可能になります。また、IEC 62667:2017との調和により、定義用語が明確にされているため、技術的な理解と運用が容易になります。 さらに、教育と訓練の重要性を強調することで、医療機器が正常に使用され、定期的な品質保証および校正チェックが適切に実施されることを保障しています。このように、本標準は、正常使用において資格を持つ人によって運用されることを前提としており、利用者の安全性を確保するための十分な基盤を提供しています。 全体として、IEC 60601-2-64:2025は、光イオンビーム医療機器の基本的な安全性に関する重要な枠組みを提供しており、医療現場での安全性と性能の確保において極めて関連性の高い標準であると言えます。
La norme IEC 60601-2-64:2025 porte sur la sécurité de base et les performances essentielles des équipements médicaux à faisceau d'ions légers, spécifiquement conçus pour la radiothérapie. Ce document est d'une grande importance pour assurer la sécurité des patients lors des traitements, car il établit des exigences spécifiques applicables à ces équipements médicaux. L'étendue de la norme inclut non seulement l'équipement médical, mais également les systèmes médicaux, garantissant une approche intégrée pour la sécurité des dispositifs utilisés en pratique médicale. Il est essentiel de noter que chaque clause ou sous-clause de la norme est clairement indiquée comme s'appliquant uniquement aux équipements médicaux ou aux systèmes médicaux, ce qui facilite la compréhension et l'application des exigences. Parmi les forces de la norme IEC 60601-2-64:2025, on peut souligner l'inclusion des tests de type et des tests sur site, qui sont cruciaux pour garantir la conformité et la performance des équipements. De plus, la norme insiste sur l'opération des appareils par des personnes qualifiées, ce qui contribue à la sécurité des procédures médicales impliquant des faisceaux d'ions légers. Cette édition apporte des changements techniques significatifs, comme l'harmonisation avec les normes antérieures, ce qui assure une continuité et une cohérence dans les exigences de sécurité. La mise à jour des définitions et des termes techniques renforce également la clarté et l'uniformité des pratiques au sein du secteur. Le fait qu’elle aborde la révision des neutrons en dehors du champ d'irradiation montre une réactivité aux besoins d'une pratique médicale évolutive. L'importance de la norme IEC 60601-2-64:2025 ne peut être sous-estimée, car elle forme le fondement pour la sécurité et l'efficacité des équipements de radiothérapie, assurant ainsi la protection des patients tout en répondant aux exigences réglementaires actuelles.
Die Norm IEC 60601-2-64:2025 legt besondere Anforderungen an die grundlegende Sicherheit und die wesentliche Leistung medizinischer elektrischer Geräte fest, die Lichtionenstrahlen verwenden. Diese Norm ist besonders relevant für die Radiotherapie in der menschlichen Medizin, da sie genau auf die Sicherheit und Funktionalität von Lichtionenstrahl-Medizinischen Geräten (ME EQUIPMENT) ausgerichtet ist. Der Anwendungsbereich dieser Norm ist klar definiert und erstreckt sich auf ME EQUIPMENT, das für die Behandlung von Patienten vorgesehen ist. Die Norm stellt sicher, dass alle in der Radiotherapie verwendeten Geräte, die beispielsweise von qualifiziertem Personal betrieben werden, höchsten Sicherheitsstandards entsprechen. Sie umfasst grundlegende Sicherheitsanforderungen und Leistungsanforderungen, die für die sichere Anwendung in medizinischen Einrichtungen entscheidend sind. Eine der Stärken der Norm IEC 60601-2-64:2025 ist die Einbeziehung von Typprüfungen und Standortprüfungen, die sich auf Herstellungs- und Installationsaspekte von Lichtionenstrahl-Medizintechnik konzentrieren. Dies stellt sicher, dass die Geräte in der Praxis ordnungsgemäß getestet und validiert werden, bevor sie im klinischen Umfeld eingesetzt werden. Zudem werden klare Anweisungen für den Betrieb und für die Durchführung regelmäßiger Qualitätssicherungsmaßnahmen durch qualifiziertes Personal gegeben. Ein wesentlicher Aspekt dieser überarbeiteten Norm ist ihre Harmonisierung mit anderen wichtigen Normen wie IEC 60601-1:2005 und IEC 62667:2017. Diese Harmonisierung verbessert die Kohärenz innerhalb des Normenwerks und sorgt dafür, dass die grundlegenden Begriffe und Definitionen einheitlich sind. Des Weiteren wird durch die Überarbeitung auf die Thematik der Neutronen außerhalb des Bestrahlungsfeldes eingegangen, was in der vorherigen Ausgabe nicht ausreichend berücksichtigt wurde. Insgesamt stellt IEC 60601-2-64:2025 eine wichtige Weiterentwicklung in der Sicherheit und der funktionalen Anforderung an Lichtionenstrahl-Medizinische Geräte dar. Die Norm trägt wesentlich dazu bei, die Patientensicherheit zu gewährleisten und die Qualität der medizinischen Behandlung durch präzise, gut getestete Geräte zu verbessern.








Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...