Tracheobronchial tubes — Sizing and marking

ISO 16628:2008 provides requirements for size designation and labelling of tracheobronchial tubes, including colour coding of the bronchial cuff and its associated pilot balloon. Tracheobronchial tubes that include bronchus blockers are excluded from the scope of ISO 16628:2008.

Sondes trachéo-bronchiques — Dimensionnement et marquage

General Information

Status
Withdrawn
Publication Date
04-Nov-2008
Current Stage
9599 - Withdrawal of International Standard
Completion Date
16-Jun-2022
Ref Project

RELATIONS

Buy Standard

Standard
ISO 16628:2008 - Tracheobronchial tubes -- Sizing and marking
English language
7 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (sample)

INTERNATIONAL ISO
STANDARD 16628
First edition
2008-11-15
Tracheobronchial tubes — Sizing and
marking
Sondes trachéo-bronchiques — Dimensionnement et marquage
Reference number
ISO 16628:2008(E)
ISO 2008
---------------------- Page: 1 ----------------------
ISO 16628:2008(E)
PDF disclaimer

This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but

shall not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In

downloading this file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat

accepts no liability in this area.
Adobe is a trademark of Adobe Systems Incorporated.

Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation

parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In

the unlikely event that a problem relating to it is found, please inform the Central Secretariat at the address given below.

COPYRIGHT PROTECTED DOCUMENT
© ISO 2008

All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means,

electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or

ISO's member body in the country of the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland
ii © ISO 2008 – All rights reserved
---------------------- Page: 2 ----------------------
ISO 16628:2008(E)
Foreword

ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies

(ISO member bodies). The work of preparing International Standards is normally carried out through ISO

technical committees. Each member body interested in a subject for which a technical committee has been

established has the right to be represented on that committee. International organizations, governmental and

non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the

International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.

International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.

The main task of technical committees is to prepare International Standards. Draft International Standards

adopted by the technical committees are circulated to the member bodies for voting. Publication as an

International Standard requires approval by at least 75 % of the member bodies casting a vote.

Attention is drawn to the possibility that some of the elements of this document may be the subject of patent

rights. ISO shall not be held responsible for identifying any or all such patent rights.

ISO 16628 was prepared by Technical Committee ISO/TC 121, Anaesthetic and respiratory equipment,

Subcommittee SC 2, Tracheal tubes and other equipment.

This first edition cancels and replaces the Technical Specification ISO/TS 16628:2003.

© ISO 2008 – All rights reserved iii
---------------------- Page: 3 ----------------------
ISO 16628:2008(E)
Introduction

Originally a Technical Specification, this document is being revised as an International Standard to provide

requirements for size designation and labelling of tracheobronchial tubes. Such requirements will have the

benefit of enabling clinicians to select the most appropriate device for their patient population and to select

accessories such as fiberoptic bronchoscopes based on the measured internal diameter of the bronchial

segment of the tracheobronchial tube.

This International Standard provides requirements for manufacturers in establishing a standard method of size

designation for tracheobronchial tubes and their parts. A tracheobronchial tube is a double lumen tracheal

tube that facilitates selective ventilation to one or both lungs and allows isolation of one lung from the other. It

is designed for either right or left mainstem bronchus placement, and has both a tracheal and bronchial cuff.

iv © ISO 2008 – All rights reserved
---------------------- Page: 4 ----------------------
INTERNATIONAL STANDARD ISO 16628:2008(E)
Tracheobronchial tubes — Sizing and marking
1 Scope

This International Standard provides requirements for size designation and labelling of tracheobronchial tubes,

including colour coding of the bronchial cuff and its associated pilot balloon. Tracheobronchial tubes that

include bronchus blockers are excluded from the scope of this International Standard.

2 Normative References

The following referenced documents are indispensable for the application of this document. For dated

references, only the edition cited applies. For undated references, the latest edition of the referenced

document (including any amendments) applies.
ISO 4135, Anaesthetic and respiratory equipment — Vocabulary
ISO 7000, Graphical symbols for use on equipment — Index and synopsis

ISO 15223-1, Medical devices — Symbols to be used with medical device labels, labelling and information to

be supplied — Part 1: General requirements
EN 1041, Information supplied by the manufacturer with medical devices
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
3.1
tracheobronchial tube

double-lumen tube designed for insertion into the trachea and a main bronchus to separate the right and left

lungs
4 General requirements
4.1 Outside diameter of bronchial segment

The outside diameter of the bronchial segment of the tube shall be the outside diameter measured at the mid-

point of the deflated bronchial cuff (see Key 9 of Figure 1) in millimetres, determined in accordance with

Annex A.
© ISO 2008 – All rights reserved 1
---------------------- Page: 5 ----------------------
ISO 16628:2008(E)
Key
1 machine end of the tracheobronchial tube
2 bronchial segment
3 cross-section of tracheal lumen (not necessarily circular)
4 cross-section of bronchial lumen (not necessarily circular)
5 patient end of the bronchial segment
6 patient end of the tracheal segment
7 tracheal cuff
8 bronchial cuff
9 point of measurement of the outside diameter of the bronchial segment
Figure 1 — Example of a tracheobronchial tube
4.2 Effective inside diameter

The inside diameter of the lumena of the bronchial and tracheal segments of the tube in millimetres shall be

determined in accordance with Annex B.

NOTE The effective inside diameter is intended as a guide to selecting the appropriate diameter bronchoscope or

other device to be inserted into the tracheobronchial tube.
4.3 Colour coding

The bronchial cuff and its associated pilot balloon shall be entirely coloured blue.

4.4 Segment
...