Plastics — Determination of average molecular mass and molecular mass distribution of polymers using size-exclusion chromatography — Part 2: Universal calibration method

ISO 16014-2:2003 specifies a method for determining the average molecular mass and the molecular mass distribution of polymers using size-exclusion chromatography (SEC). The average molecular mass and the molecular mass distribution are calculated using a universal calibration curve instead of the conventional calibration curve. This test method is classified as a relative method as described in ISO 16014-1, but the average molecular masses and molecular mass distributions calculated by the method are equal to, or nearly equal to, the absolute values.

Plastiques — Détermination de la masse moléculaire moyenne et de la répartition des masses moléculaires des polymères par chromatographie d'exclusion stérique — Partie 2: Méthode d'étalonnage universelle

General Information

Status
Withdrawn
Publication Date
06-May-2003
Withdrawal Date
06-May-2003
Current Stage
9599 - Withdrawal of International Standard
Completion Date
03-Jul-2012
Ref Project

Relations

Buy Standard

Standard
ISO 16014-2:2003 - Plastics -- Determination of average molecular mass and molecular mass distribution of polymers using size-exclusion chromatography
English language
12 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

INTERNATIONAL ISO
STANDARD 16014-2
First edition
2003-04-15


Plastics — Determination of average
molecular mass and molecular mass
distribution of polymers using
size-exclusion chromatography —
Part 2:
Universal calibration method
Plastiques — Détermination de la masse moléculaire moyenne et de la
répartition des masses moléculaires des polymères par
chromatographie d'exclusion stérique —
Partie 2: Méthode d'étalonnage universelle





Reference number
ISO 16014-2:2003(E)
©
ISO 2003

---------------------- Page: 1 ----------------------
ISO 16014-2:2003(E)
PDF disclaimer
This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but
shall not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In
downloading this file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat
accepts no liability in this area.
Adobe is a trademark of Adobe Systems Incorporated.
Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation
parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In
the unlikely event that a problem relating to it is found, please inform the Central Secretariat at the address given below.


©  ISO 2003
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means,
electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or
ISO's member body in the country of the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland

ii © ISO 2003 — All rights reserved

---------------------- Page: 2 ----------------------
ISO 16014-2:2003(E)
Contents Page
Foreword. iv
1 Scope. 1
2 Normative references . 1
3 Terms and definitions. 2
4 Principle of universal calibration method . 3
5 Reagents . 3
6 Apparatus. 3
7 Procedure. 4
8 Data acquisition and processing. 4
9 Expression of results. 4
10 Precision . 7
11 Test report. 7
Annex A (informative) Supplementary information. 8
Annex B (informative) K and a in the Mark-Houwink-Sakurada equation. 11
Bibliography . 12

© ISO 2003 — All rights reserved iii

---------------------- Page: 3 ----------------------
ISO 16014-2:2003(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies
(ISO member bodies). The work of preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which a technical committee has been
established has the right to be represented on that committee. International organizations, governmental and
non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the
International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.
The main task of technical committees is to prepare International Standards. Draft International Standards
adopted by the technical committees are circulated to the member bodies for voting. Publication as an
International Standard requires approval by at least 75 % of the member bodies casting a vote.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. ISO shall not be held responsible for identifying any or all such patent rights.
ISO 16014-2 was prepared by Technical Committee ISO/TC 61, Plastics, Subcommittee SC 5, Physical-
chemical properties.
ISO 16014 consists of the following parts, under the general title Plastics — Determination of average
molecular mass and molecular mass distribution of polymers using size-exclusion chromatography:
 Part 1: General principles
 Part 2: Universal calibration method
 Part 3: Low-temperature method
 Part 4: High-temperature method

iv © ISO 2003 — All rights reserved

---------------------- Page: 4 ----------------------
INTERNATIONAL STANDARD ISO 16014-2:2003(E)

Plastics — Determination of average molecular mass and
molecular mass distribution of polymers using size-exclusion
chromatography —
Part 2:
Universal calibration method
1 Scope
This part of ISO 16014 specifies a method for determining the average molecular mass and the molecular
mass distribution of polymers using size-exclusion chromatography (SEC). The average molecular mass and
the molecular mass distribution are calculated using a universal calibration curve instead of the conventional
calibration curve. The concept of the universal calibration method is based on a relationship by which the
retention time in SEC depends on the size of the polymer molecule (the hydrodynamic volume) or on the
product of the intrinsic viscosity [η] and the molecular mass M. Many polymers classified as random-coil
polymers (regardless of their chemical structure, degree of branching, composition or tacticity) have been
experimentally confirmed as following this relationship. The universal calibration curve is constructed by
plotting the logarithm of [η]M versus the elution time t or elution volume V using polymer standards such as
polystyrene with narrow molecular mass distributions. The molecular mass M of an unknown polymer sample
i
can then be calculated from the value of [η]M at each retention time using the universal calibration curve and
a a+1
[η] or K and a in the Mark-Houwink-Sakurada equation ([η] = KM → [η]M = KM ), where K is a constant
and a is an exponent depending on the experimental conditions. Therefore this test method is classified as a
relative method as described in ISO 16014-1, but the average molecular masses and molecular mass
distributions calculated by the method are equal to, or nearly equal to, the absolute values.
A list of documents related to this part of ISO 16014 is given in the Bibliography.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.
ISO 472:1999, Plastics — Vocabulary
ISO 16014-1:2003, Plastics — Determination of average molecular mass and molecular mass distribution of
polymers using size-exclusion chromatography — Part 1: General principles
ISO 16014-3:2003, Plastics — Determination of average molecular mass and molecular mass distribution of
polymers using size-exclusion chromatography — Part 3: Low-temperature method
ISO 16014-4:2003, Plastics — Determination of average molecular mass and molecular mass distribution of
polymers using size-exclusion chromatography — Part 4: High-temperature method
© ISO 2003 — All rights reserved 1

---------------------- Page: 5 ----------------------
ISO 16014-2:2003(E)
3 Terms and definitions
For the purposes of this document, the terms and definitions given in ISO 472 and in ISO 16014-1 apply. The
definitions of the various types of average molecular mass are reproduced here for convenience.
Four types of average molecular mass are defined by the following equations, where N is the number of
i
molecules of species i of molecular mass M and a is the exponent of the Mark-Houwink-Sakurada equation.
i
NOTE Although the definitions are the same as in ISO 16014-1, the values of M in the equations are calculated
i
using a universal calibration curve as described in Clause 4.
3.1
number-average molecular mass
M
n

()NM×
ii

i=1
M = (1)
n

N
∑ i
i=1
3.2
mass-average molecular mass
M
w

2
()NM×
ii

i=1
M = (2)
w

()NM×
∑ ii
i=1
3.3
z-average molecular mass
M
z

3
()NM×
ii

i=1
M = (3)
z

2
()NM×
∑ ii
i=1
3.4
viscosity-average molecular mass
M
v
1/a


a+1
()NM×
ii


i=1
M = (4)
v


()NM×
∑ ii


i=1
2 © ISO 2003 — All rights reserved

---------------------- Page: 6 ----------------------
ISO 16014-2:2003(E)
4 Principle of universal calibration method
In this part of ISO 16014, the experimental SEC procedures are the same as in ISO 16014-1, ISO 16014-3
and ISO 16014-4, but the molecular mass M at each elution time is calculated based on the universal
i
calibration method. According to the theory, this method provides values for the average molecular mass and
the molecular mass distribution that are equal to, or nearly equal to, the absolute values.
According to theory, the size of a polymer in solution, the hydrodynamic volume V , is proportional to the
h
product of the intrinsic viscosity [η] and the molecular mass M as given by Equation (5):
[]ηη∝→VM/ [ ]M∝V (5)
hh
In SEC, many random-coil polymers, regardless of their chemical structure, degree of branching, composition,
or tacticity, follow the same relationship, in which the retention time is proportional to the product of [η] and M
under the specific experimental conditions, including the column system, solvent and temperature. The
relationship between the logarithm of [η]M and the elution time is called the “universal calibration curve”, and
the SEC technique for the determination of average molecular masses and molecular mass distributions using
this calibration curve is called the “universal calibration method”.
As a first step, the universal calibration curve, retention time t versus the logarithm of [η]M, is constructed
using polymer standards with narrow molecular mass distributions and the Mark-Houwink-Sakurada equation:
a
s
[]η =KM (6)
sss
where the subscript “s” denotes “polymer standard”.
Direct measurement of [η] is also possible.
s
In the next step, the molecular mass M at each retention time t is calculated using the universal calibration
i i
curve and the Mark-Houwink-Sakurada equation for the polymer sample:
a
[]η = KM (7)
a+1
[]ηηMM==[ ] KM (8)
s,iis, i i i
Again, direct measurement of [η] is possible.
Lastly, the average molecular mass and the molecular mass distribution are calculated from the molecular
mass M and peak height H at each retention time, using the equations given in 9.2, 9.3, and 9.4.
i i
For additional information on the universal calibration curve and the use of a conventional calibration curve
based on universal calibration theory, see 9.1 and A.2.1.
For additional information on the intrinsic viscosity [η], see A.2.2.
The values of K and a in the Mark-Houwink-Sakurada equation for several polymers are given in Annex B.
5 Reagents
See ISO 16014-1:2003, Clause 5.
6 Apparatus
See ISO 16014-1:2003, Clause 6.
© ISO 2003 — Al
...