ISO/FDIS 8655-7
(Main)Piston-operated volumetric apparatus
Piston-operated volumetric apparatus
Appareils volumétriques à piston
General Information
RELATIONS
Standards Content (sample)
FINAL
INTERNATIONAL ISO/FDIS
DRAFT
STANDARD 8655-7
ISO/TC 48
Piston-operated volumetric
Secretariat: DIN
apparatus —
Voting begins on:
2021-11-17
Part 7:
Voting terminates on:
Alternative measurement procedures
2022-01-12
for the determination of volume
Appareils volumétriques à piston —
Partie 7: Modes opératoires de mesure alternatifs pour la
détermination de volumes
ISO/CEN PARALLEL PROCESSING
RECIPIENTS OF THIS DRAFT ARE INVITED TO
SUBMIT, WITH THEIR COMMENTS, NOTIFICATION
OF ANY RELEVANT PATENT RIGHTS OF WHICH
THEY ARE AWARE AND TO PROVIDE SUPPOR TING
DOCUMENTATION.
IN ADDITION TO THEIR EVALUATION AS
Reference number
BEING ACCEPTABLE FOR INDUSTRIAL, TECHNO-
ISO/FDIS 8655-7:2021(E)
LOGICAL, COMMERCIAL AND USER PURPOSES,
DRAFT INTERNATIONAL STANDARDS MAY ON
OCCASION HAVE TO BE CONSIDERED IN THE
LIGHT OF THEIR POTENTIAL TO BECOME STAN-
DARDS TO WHICH REFERENCE MAY BE MADE IN
NATIONAL REGULATIONS. © ISO 2021
---------------------- Page: 1 ----------------------
ISO/FDIS 8655-7:2021(E)
COPYRIGHT PROTECTED DOCUMENT
© ISO 2021
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on
the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address below
or ISO’s member body in the country of the requester.ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
© ISO 2021 – All rights reserved
---------------------- Page: 2 ----------------------
ISO/FDIS 8655-7:2021(E)
Contents Page
Foreword ..........................................................................................................................................................................................................................................v
Introduction .............................................................................................................................................................................................................................. vi
1 Scope ................................................................................................................................................................................................................................. 1
2 Normative references ..................................................................................................................................................................................... 1
3 Terms and definitions .................................................................................................................................................................................... 2
4 General requirements .................................................................................................................................................................................... 2
4.1 Metrological confirmation ........................................................................................................................................................... 2
4.2 Uncertainty of measurement ..................................................................................................................................................... 2
4.3 Operator qualification ..................................................................................................................................................................... 3
5 Performance requirements ........................................................................................................................................... ...........................3
5.1 Performance tolerances ................................................................................................................................................................. 3
5.2 Operator impact .................................................................................................................................................................................... 3
6 Test conditions .......................................................................................................................................................................................................3
6.1 General ........................................................................................................................................................................................................... 3
6.2 Test equipment ....................................................................................................................................................................................... 3
6.3 Test room, environmental conditions ................................................................................................................................ 4
6.4 Test volumes ............................................................................................................................................................................................. 4
6.4.1 Fixed volume POVA........................................................................................................................................................... 4
6.4.2 Adjustable volume POVA ............................................................................................................................................. 4
6.5 Number of measurements per test volume ................................................................................................................... 5
6.6 Test liquids ................................................................................................................................................................................................. 5
7 Evaluation ................................................................................................................................................................................................................... 5
7.1 Mean volume ............................................................................................................................................................................................ 5
7.2 Systematic error of measurement ......................................................................................................................................... 6
7.3 Random error of measurement ............................................................................................................................................... 6
8 Test methods ............................................................................................................................................................................................................ 7
8.1 General ........................................................................................................................................................................................................... 7
8.2 Gravimetric method ........................................................................................................................................... ................................ 7
8.3 Dual-dye ratiometric photometric method ................................................................................................................... 8
8.4 Single dye photometric method .............................................................................................................................................. 8
8.5 Hybrid photometric/gravimetric method for multichannel POVA .......................................................... 8
8.6 Titration method .................................................................................................................................................................................. 8
8.7 Batch testing ............................................................................................................................................................................................. 8
9 Dispense procedures .......................................................................................................................................................................................9
9.1 General ........................................................................................................................................................................................................... 9
9.2 Preparation ................................................................................................................................................................................................ 9
9.3 Single-channel air displacement pipettes (in accordance with ISO 8655-2) .................................. 9
9.3.1 General ........................................................................................................................................................................................ 9
9.3.2 Test cycle ................................................................................................................................................................................... 9
9.4 Multi-channel pipettes (in accordance with ISO 8655-2) ............................................................................. 10
9.5 Positive displacement pipettes (in accordance with ISO 8655-2) ......................................................... 11
9.6 Burettes (in accordance with ISO 8655-3) ................................................................................................................. 11
9.7 Dilutors (in accordance with ISO 8655-4) ..................................................................................................................12
9.7.1 General .....................................................................................................................................................................................12
9.7.2 Test cycle ................................................................................................................................................................................12
9.8 Dispensers (in accordance with ISO 8655-5) ........................................................................................................... 13
9.9 Syringes (in accordance with ISO 8655-9) ................................................................................................................. 13
9.9.1 General .....................................................................................................................................................................................13
9.9.2 Test cycle ................................................................................................................................................................................ 13
10 Reporting of results .......................................................................................................................................................................................14
iii© ISO 2021 – All rights reserved
---------------------- Page: 3 ----------------------
ISO/FDIS 8655-7:2021(E)
Annex A (normative) Gravimetric procedure .........................................................................................................................................16
Annex B (normative) Dual-dye ratiometric photometric procedure .............................................................................21
Annex C (normative) Single dye photometric procedure ...........................................................................................................29
Annex D (normative) Photometric/gravimetric hybrid procedure ................................................................................33
Annex E (normative) Titrimetric procedure............................................................................................................................................41
Annex F (normative) Conversion of liquid mass to volume ......................................................................................................45
Bibliography .............................................................................................................................................................................................................................48
© ISO 2021 – All rights reserved---------------------- Page: 4 ----------------------
ISO/FDIS 8655-7:2021(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of
any patent rights identified during the development of the document will be in the Introduction and/or
on the ISO list of patent declarations received (see www.iso.org/patents).Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and
expressions related to conformity assessment, as well as information about ISO's adherence to
the World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT), see
www.iso.org/iso/foreword.html.This document was prepared by Technical Committee ISO/TC 48, Laboratory equipment, in collaboration
with the European Committee for Standardization (CEN) Technical Committee CEN/TC 332, Laboratory
equipment, in accordance with the Agreement on technical cooperation between ISO and CEN (Vienna
Agreement).This second edition cancels and replaces the first edition (ISO 8655-7:2005), which has been technically
revised. It also incorporates the Technical Corrigendum ISO 8655-7:2005/Cor.1:2008.
The main changes compared to the previous edition are as follows:— a gravimetric test method was added (see 8.2);
— a photometric/gravimetric hybrid test method was added (see 8.5);
— a batch testing method was added (see 8.7);
— measurement procedures for all methods are given in normative Annexes A to E;
— standard dispense procedures for POVA described in ISO 8655-2, ISO 8655−3, ISO 8655−4,
ISO 8655−5, and ISO 8655−9 were added (see Clause 9);— requirements for operator qualification have been added (see 4.3);
— requirements for testing of multi-channel POVA is described in more detail, with specific procedures
given for these apparatus (see 8.5, and Annex D);— Annexes A, B, and C of the first edition have been deleted and replaced.
A list of all parts in the ISO 8655 series can be found on the ISO website.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www.iso.org/members.html.© ISO 2021 – All rights reserved
---------------------- Page: 5 ----------------------
ISO/FDIS 8655-7:2021(E)
Introduction
The ISO 8655 series addresses the needs of:
— manufacturers, as a basis for quality control including, where appropriate, the issuance of
manufacturer’s declarations;— calibration laboratories, test houses, users of the equipment and other bodies as a basis for
independent calibration, testing, verification, and routine tests.The tests specified in the ISO 8655 series are intended to be carried out by trained personnel.
© ISO 2021 – All rights reserved---------------------- Page: 6 ----------------------
FINAL DRAFT INTERNATIONAL STANDARD ISO/FDIS 8655-7:2021(E)
Piston-operated volumetric apparatus —
Part 7:
Alternative measurement procedures for the
determination of volume
1 Scope
This document specifies alternative measurement procedures for the determination of volume of
piston-operated volumetric apparatus.The procedures are applicable to complete systems comprising the basic apparatus and all parts
selected for use with the apparatus, disposable or reusable, involved in the measurement by delivery
process (Ex). Methods described in this document are suitable for various maximum nominal volumes
of piston-operated volumetric apparatus. It is the responsibility of the user to select the appropriate
method.2 Normative references
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments) applies.
ISO 1042, Laboratory glassware — One-mark volumetric flasksISO 2859-1, Sampling procedures for inspection by attributes — Part 1: Sampling schemes indexed by
acceptance quality limit (AQL) for lot-by-lot inspectionISO 3696:1987, Water for analytical laboratory use — Specification and test methods
ISO 3951-1, Sampling procedures for inspection by variables — Part 1: Specification for single sampling
plans indexed by acceptance quality limit (AQL) for lot-by-lot inspection for a single quality characteristic
and a single AQLISO 8655-1:— , Piston-operated volumetric apparatus — Part 1: Terminology, general requirements and
user recommendationsISO 8655-2:— , Piston-operated volumetric apparatus — Part 2: Pipettes
ISO 8655-3:2021, Piston-operated volumetric apparatus — Part 3: Burettes
ISO 8655-4:2021, Piston-operated volumetric apparatus — Part 4: Dilutors
ISO 8655-5:2021, Piston-operated volumetric apparatus — Part 5: Dispensers
ISO 8655-6, Piston-operated volumetric apparatus — Part 6: Gravimetric reference measurement
procedure for the determination of volumeISO 8655-8, Piston-operated volumetric apparatus — Part 8: Photometric reference measurement
procedure for the determination of volume1) Under preparation. Stage at the time of publication: ISO/FDIS 8655-1:2021.
2) Under preparation. Stage at the time of publication: ISO/FDIS 8655-2:2021.
© ISO 2021 – All rights reserved
---------------------- Page: 7 ----------------------
ISO/FDIS 8655-7:2021(E)
ISO 8655-9, Piston-operated volumetric apparatus — Part 9: Manually operated precision laboratory
syringesISO/IEC Guide 2, Standardization and related activities — General vocabulary
ISO/IEC Guide 99, International vocabulary of metrology — Basic and general concepts and associated
terms (VIM)3 Terms and definitions
For the purposes of this document, the terms and definitions given in ISO 8655-1, ISO/IEC Guide 2, and
ISO/IEC Guide 99 apply.ISO and IEC maintain terminological databases for use in standardization at the following addresses:
— IEC Electropedia: available at https:// www .electropedia .org/— ISO Online browsing platform: available at https:// www .iso .org/ obp
3.1
acceptance quality limit
AQL
worst tolerable quality level
Note 1 to entry: This concept only applies when a sampling scheme with rules for switching and for
discontinuation, such as in ISO 2859-1, ISO 3951-1 or ISO 3951-5 is used.Note 2 to entry: Although individual lots with quality as bad as the acceptance quality limit may be accepted
with fairly high probability, the designation of an acceptance quality limit does not suggest that this is a desirable
quality level. Sampling schemes found in International Standards such as ISO 2859-1, ISO 3951-1 or ISO 3951-5,
with their rules for switching and for discontinuation of sampling inspection, are designed to encourage suppliers
to have process averages consistently better than the AQL.4 General requirements
4.1 Metrological confirmation
Metrological confirmation of all POVA shall be performed on a regular basis to ensure the apparatus
conforms to requirements for its intended use. The requirements of the methods and procedures
described in this document are suitable to be used in the metrological confirmation of POVA. For
metrological confirmation, no less than ten replicate measurements per selected volume shall be
performed and the measurement procedures in this document shall be validated by comparison to one
of the reference measurement procedures described in ISO 8655-6 or ISO 8655-8.4.2 Uncertainty of measurement
When performing calibrations (ISO 8655-1:—, 6.4) according to measurement procedures described in
this document, the expanded measurement uncertainty of the mean delivered volume for each selected
volume shall be estimated and reported (see Clause 10 (m)).When performing testing (ISO 8655-1:2022, 6.4) or routine tests (ISO 8655-1:2022, 6.5), it is optional to
estimate and report the expanded measurement uncertainty.NOTE For further information on uncertainty for the photometric and gravimetric methods, refer to
[1] [2]ISO/TR 16153 and ISO/TR 20461 respectively.
© ISO 2021 – All rights reserved
---------------------- Page: 8 ----------------------
ISO/FDIS 8655-7:2021(E)
4.3 Operator qualification
An operator who uses POVA for volumetric transfers, performs metrological confirmation or routine
tests of POVA shall be adequately trained on the use of the type of POVA under test. Operator training
and competence should be documented.NOTE 1 Previously calibrated POVA can be used for the qualification of operators.
NOTE 2 Training and qualification requirements for operators of POVA are intended to be included in
ISO 8655-10.5 Performance requirements
5.1 Performance tolerances
Calibration and routine test results may be reported without comparison to performance tolerances.
If the results are verified against performance tolerances, these tolerances shall be stated on the test
report/certificate.Performance tolerances may be based on the user’s liquid handling process tolerances or the product
tolerances given in the part of ISO 8655 corresponding to the type of POVA under test or the tolerances
specified by the manufacturer, subject to them being fit for purpose.5.2 Operator impact
Measurement of volumetric performance includes random and systematic errors of the POVA, as well as
errors introduced by the device’s operator. The performance of a hand-held pipette is inseparable from
the performance of its operator.NOTE More information about operator impact is given in ISO 8655-10.
6 Test conditions
6.1 General
Test conditions described in this clause shall be validated for their suitability for the selected test
method and procedure. Test conditions, together with the test equipment and detailed test procedure,
impact the uncertainty of measurement. Examples for the calculation of the expanded uncertainty of
[1]the mean volume and of the uncertainty in use of a single delivered volume are given in ISO/TR 16153
[2]and ISO/TR 20461 .
6.2 Test equipment
All equipment used for the testing of POVA, including for the preparation of test solutions, shall be
chosen such that the required uncertainty of measurement can be obtained.All test equipment used shall be of suitable readability, accuracy, reproducibility and stability, consistent
with the required expanded uncertainty of measurement.Deviations from the test equipment given in this document shall be taken into account when calculating
the expanded measurement uncertainty and shall be proven to yield measurement results fit for the
intended purpose.© ISO 2021 – All rights reserved
---------------------- Page: 9 ----------------------
ISO/FDIS 8655-7:2021(E)
6.3 Test room, environmental conditions
The following applies:
a) The test room should be kept at a steady temperature throughout the entirety of the equilibration
time for the test equipment and POVA (±1 °C), and throughout the POVA testing time (±0,5 °C). All
test equipment, POVA, exchangeable parts (e. g. pipette tips), and reagents used shall be equilibrated
to the test room temperature.b) The air temperature, relative humidity, and barometric pressure at the time of the test shall be
recorded. At the start and at the end of the n replicate measurements, the temperature of the test
liquid shall be recorded.NOTE 1 Air temperature and barometric pressure are necessary for the conversion of liquid mass to
volume (see Annex F); the relative humidity is necessary for the stability of the room conditions and is
necessary for documentation in the test report.c) To aid evaluation of a POVA’s fitness for purpose, the test room conditions (temperature, relative
humidity, and barometric pressure) should reflect the environmental conditions under which the
POVA is used, within the constraints mentioned in a). This can be achieved when a POVA is tested
within the laboratory in which it is used. Other environmental and non-environmental factors can
influence a POVA’s fitness for purpose.d) The test environment should be draft free.
e) Prior to the test, the apparatus to be tested, all test equipment, and test solutions shall have stood
in the test room conditions for a sufficient time to reach equilibrium with the test room conditions.
f) The environmental conditions, air temperature and air humidity, shall be within the specified
limits for the test room for at least 2 h before starting the test (minimum equilibration time) and
during the test itself.NOTE 2 It is unlikely that this minimum equilibration time will be less than 2h and can be considerably
longer.NOTE 3 Calibration laboratories at test houses or pipette manufacturer’s quality control laboratories can
often precisely control environmental conditions to achieve a desired standard condition. It can be very
challenging to reproduce such results under different environmental conditions.6.4 Test volumes
6.4.1 Fixed volume POVA
In the case of a fixed-volume POVA, the selected volume V is the nominal volume V and is the only
S nomtest volume.
6.4.2 Adjustable volume POVA
a) For calibrations and testing, adjustable volume POVA shall be tested at least at three volumes:
— nominal volume,— 50 % of the nominal volume, or the closest possible (if equidistant, use the higher value);
— the lower limit of the useable volume range or 10 % of the nominal volume (whichever is the
greater).Measurement of further volumes is optional.
b) For routine tests, fewer than three volumes may be tested.
© ISO 2021 – All rights reserved
---------------------- Page: 10 ----------------------
ISO/FDIS 8655-7:2021(E)
In case the POVA is to be tested at only two volumes, the nominal volume and the lower limit of the
useable volume range, or 10 % of the nominal volume (whichever is the greater), shall be tested.
NOTE The linearity of delivered volumes between these two test points is unknown and is likely to increase
the risk for volumetric errors as compared to a test at three volumes.In case the POVA is tested only at one volume, it shall be tested at its nominal volume, or at the volume
at which it will be used.6.5 Number of measurements per test volume
The confidence of metrological confirmation increases with the number of replicate measurements
for each test volume. A minimum of 10 measurements per volume are required by the reference
measurement procedures specified in ISO 8655-6 and ISO 8655-8. For the test procedures specified in
this document, 10 measurements are recommended, but fewer replicate measurements may be made if
the expanded uncertainty of measurement for the POVA is fit for the intended purpose. The number of
replicates shall not be less than 4.After repair or adjustment of the POVA, a minimum of 10 measurements shall be performed.
The replicate volume measurements shall be used to calculate the systematic and the random errors of
measurement in accordance with Clause 8. When applicable, the reported uncertainty shall be based on
the number of replicates.6.6 Test liquids
POVA are typically supplied with adjustments using water. Calibrations or routine tests may be
performed using other liquids or solutions. For the purpose of this document, the term “test liquid”
is used for pure solvents, as well as for prepared chromophore or other solutions. The test liquid used
shall be described in sufficient detail to allow replication of the test and interpretation of the results.
The following characteristics of the test liquid shall be taken into account when determining the
measured volume: Z-factor when weighing, absorbances of the chromophores when using photometric
methods, and conductivity and reactivity when performing potentiometric titration.
Depending on the type of POVA, the following parameters can influence the amount of liquid aspirated
and/or dispensed: viscosity, density, chemical composition, and surface t...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.