ISO/TS 16766:2024
(Main)Manufacturers’ considerations for in vitro diagnostic medical devices in a public health emergency
Manufacturers’ considerations for in vitro diagnostic medical devices in a public health emergency
This document provides guidance to manufacturers on the minimum requirements for the lifecycle management of in vitro diagnostic (IVD) medical devices that are developed in preparation for and in response to a public health emergency involving infectious agents requiring immediate availability of authorized IVD devices. NOTE This document does not replace existing national (or regional) regulatory pathway requirements for IVD medical devices under non-emergency situations. The regulatory authorization process of emergency use-IVD medical devices is country-specific and it includes: — following a risk management process; — monitoring the device’s post-market performance and quality assurance; — implementing a communication system.
Aspects à prendre en compte par les fabricants de dispositifs médicaux de diagnostic in vitro en situation d’urgence de santé publique
General Information
Standards Content (Sample)
Technical
Specification
ISO/TS 16766
First edition
Manufacturers’ considerations for
2024-11
in vitro diagnostic medical devices
in a public health emergency
Aspects à prendre en compte par les fabricants de dispositifs
médicaux de diagnostic in vitro en situation d’urgence de santé
publique
Reference number
© ISO 2024
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on
the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address below
or ISO’s member body in the country of the requester.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii
Contents Page
Foreword .iv
Introduction .v
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 1
4 General considerations for the design and development process . 5
4.1 Safety and performance requirements .5
4.2 Quality management and risk management .6
4.3 Target condition and scientific validity .6
4.4 Assay technology .6
4.5 Intended use and risks and benefits of the device .6
4.6 Analytical performance .7
4.7 Stability .7
4.8 Clinical performance .7
5 General considerations for the risk management process . 8
5.1 General .8
5.2 Risk reduction .8
6 Monitoring the device’s post-market performance and quality assurance . 9
6.1 General .9
6.2 Monitoring post-market performance.9
6.3 Quality assurance .9
7 Implementing a communication process . 10
7.1 Establishing a communication process .10
7.2 Manufacturer responsibility .10
Bibliography .11
iii
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out through
ISO technical committees. Each member body interested in a subject for which a technical committee
has been established has the right to be represented on that committee. International organizations,
governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely
with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are described
in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the different types
of ISO documents should be noted. This document was drafted in accordance with the editorial rules of the
ISO/IEC Directives, Part 2 (see www.iso.org/directives).
ISO draws attention to the possibility that the implementation of this document may involve the use of (a)
patent(s). ISO takes no position concerning the evidence, validity or applicability of any claimed patent
rights in respect thereof. As of the date of publication of this document, ISO had not received notice of (a)
patent(s) which may be required to implement this document. However, implementers are cautioned that
this may not represent the latest information, which may be obtained from the patent database available at
www.iso.org/patents. ISO shall not be held responsible for identifying any or all such patent rights.
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and expressions
related to conformity assessment, as well as information about ISO's adherence to the World Trade
Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www.iso.org/iso/foreword.html.
This document was prepared by Technical Committee ISO/TC 212, Medical laboratories and in vitro diagnostic
systems.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www.iso.org/members.html.
iv
Introduction
During a pandemic, accurate identification and isolation of infected individuals is an effective initial response
to secure public health and safety before vaccines are available. The coronavirus disease of 2019 (COVID-19)
posed an unprecedented public health emergency, causing many countries to impose restrictions on travel
and daily activities to slow the spread of infection. An example where the spread of COVID-19 infection was
[1]
demonstrated to have been slowed before vaccines became available has been published. Here, a series of
interventions, such as the urgent introduction of appropriate emergency use in vitro diagnostic (emergency
use-IVD) medical devices, aggressive testing, rigorous contact tracing, etc., were applied in the early stage
of the pandemic. Such a series of interventions effectively slowed the spread of infections and succeeded in
maintaining public health and safety without the collapse of intensive care capabilities.
Often, regulatory authorization of in vitro diagnostic (IVD) medical devices takes months to a year or more to
review and approve under a traditional regulatory pathway. Following a global infectious disease outbreak
such as severe acute respiratory syndrome (SARS) or Middle East respiratory syndrome (MERS), the need
for an accelerated regulatory pathway to facilitate the introduction of emergency use-IVD medical devices
was recognised. While such pathways (i.e. emergency use authorization processes) have been implemented,
[2-14]
the processes for authorization are neither well established nor harmonized.
While some international guidance is available for the minimum requirements for an IVD medical device in a
public health emergency, the regulatory requirements can differ from one jurisdiction to another. For example,
information on the quality system (e.g. ISO 13485), developmental history, or raw materials/manufacturing
methods are required when a manufacturer applies for the accelerated regulatory pathway in some countries
but not in others. Also, some countries require a stability shelf life claim, but the level of evidence required
[2,11,16,17]
to demonstrate stability during the initial application varies by region. In an urgent situation such
as a pandemic, the application of non-standardized requirements can impede implementation of the use of
emergency use-IVD medical devices that are critical in protecting global public health.
This document provides minimum requirements, which span pre-market to post-market activities, to
accelerate the availability of IVD medical devices in a public health emergency.
v
Technical Specification ISO/TS 16766:2024(en)
Manufacturers’ considerations for in vitro diagnostic medical
devices in a public health emergency
1 Scope
This document provides guidance to manufacturers on the minimum requirements for the lifecycle
management of in vitro diagnostic (IVD) medical devices that are developed in preparation for and in
response to a public health emergency involving infectious agents requiring immediate availability of
authorized IVD devices.
NOTE This document does not replace existing national (or regional) regulatory pathway requirements for IVD
medical devices under non-emergency situations. The regulatory authorization process of emergency use-IVD medical
devices is country-specific and it includes:
— following a risk management process;
— monitoring the device’s post-market performance and quality assurance;
— implementing a communication system.
2 Normative references
There are no normative references in this document.
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
ISO and IEC maintain terminology databases for use in standardization at the following addresses:
— ISO Online browsing platform: available at https:// www .iso .org/ obp
— IEC Electropedia: available at https:// www .electropedia .org/
3.1
adverse event
untoward medical occurrence, inappropriate patient management decision, unintended disease or injury, or
untoward clinical sign in subjects, users, or other persons, with any connection to study related activities,
whether or not related to the IVD medical device (3.9) under investigation
Note 1 to entry: Adverse events can be caused by, for instance, insufficient or inadequate instructions for use,
deployment, installation, operation, or any malfunction of the IVD medical device under investigation.
Note 2 to entry: This definition includes the malfunction or deterioration of a device which has not yet caused death or
serious injury, but which can lead to death or serious injury.
Note 3 to entry: This definition is not intended to be used in determining whether an event is reportable to a regulatory
authority.
Note 4 to entry: For users or other persons, this definition is restricted to events related to investigational IVD medical
devices.
Note 5 to entry: False negative or false positive results are not considered adverse events unless, in an interventional
study, inappropriate patient management decisions are made based on those false results.
[SOURCE: ISO 20916:2019, 3.2]
3.2
analytical performance
ability of an IVD medical device (3.9) to detect or measure a particular analyte or measurand
Note 1 to entry: In metrological terms, this is referred to as performance of a measuring instrument or measuring system.
[SOURCE: ISO 18113-1:2022, 3.2.3, modified — in the definition, “or measurand” was added; Note 1 to entry
was added.]
3.3
analytical sensitivity
quotient of the change in a measurement indication and the corresponding change in a value of a quantity
being measured
Note 1 to entry: The sensitivity of a measurement procedure can depend on the value of the quantity being measured.
Note 2 to entry: The change considered in the value of the quantity being measured shall be large compared with the
resolution.
Note 3 to entry: The analytical sensitivity of a measuring system is the slope of the calibration curve.
Note 4 to entry: Analytical sensitivity should not be used to mean detection limit or quantitation limit and should not
be confused with diagnostic sensitivity.
Note 5 to entry: In metrological terms, this is referred to as measurement sensitivity.
[SOURCE: ISO 18113-1:2022, 3.2.4, modified — the preferred term “sensitivity of a measurement procedure”
was removed; Note 5 to entry added.]
3.4
clinical performance of an IVD medical device
clinical performance
ability of an IVD medical device (3.9) to yield results that are correlated with a particular clinical condition or
physiological/pathological process/state in accordance with the intended use (clinical test purpose, target
population and intended user)
Note 1 to entry: In accordance with intended use, clinical performance can include expected values, diagnostic
sensitivity and diagnostic specificity based on the known clinical condition or physiological/pathological process/
state of the individual, and negative and positive predictive values based on the prevalence of the disease.
[SOURCE: ISO 20916:2019, 3.10, modified — the preferred term “clinical performance” was added.]
3.5
clinical performance study
study undertaken to establish or confirm the clinical performance of an IVD medical device (3.4)
Note 1 to entry: Testing performed pre-market that is not designed to address clinical performance of an IVD medical
device is not considered a clinical performance study (e.g. customer feedback studies, external analytical performance
studies, research studies).
[SOURCE: ISO 20916:2019, 3.11]
3.6
complaint
electronic or oral communication that alleges deficiencies related to the identity, quality, durability,
reliability, usability, safety or performance of a medical device that has been released from the organization’s
control or related to a service that affects the performance of such medical devices
Note 1 to entry: This definition of “complaint” differs from the definition given in ISO 9000:2015.
[SOURCE: ISO 13485:2016, 3.4]
3.7
infectious agent
pathogen
infectious micro-organism or agent, such as a virus, bacterium, protozoan, prion, viroid, or fungus that can
cause disease
[SOURCE: ISO/TS 16975-4:2022, 3.16, modified — Note 1 to entry has been removed.]
3.8
intended use
objective intent of an IVD manufacturer (3.11) regarding the use of a product, process or service as reflected
in the specifications, instructions and information supplied by the IVD manufacturer
Note 1 to entry: Intended use statements for IVD labelling can include two components: a description of the
functionality of the IVD medical device (3.9) (e.g. an immunochemical measurement procedure for the detection
of analyte “x” in serum or plasma), and a statement of the intended medical use of the examination results.
Note 2 to entry: The intended use can include the indications for use.
[SOURCE: ISO 18113-1:2022, 3.1.37]
3.9
in vitro diagnostic medical device
IVD medical device
medical device, whether used alone or in combination, intended by the manufacturer (3.11) for the in vitro
examination of specimens derived from the human body solely or principally to provide information for
diagnostic, monitoring or compatibility purposes
Note 1 to entry: IVD medical devices include reagents, calibrators, control materials, specimen receptacles,
software, and related instruments or apparatus or other articles and are used, for example, for the following test
purposes: diagnosis, aid to diagnosis, screening, monitoring, predisposition, prognosis, prediction, determination of
physiological state.
Note 2 to entry: In some jurisdictions, certain IVD medical devices can be covered by other regulations.
[SOURCE: ISO 18113-1:2022, 3.1.33]
3.10
leftover specimen
leftover sample
unadulterated remnants of human derived specimens collected as part of routine clinical practice and after
all standard analyses have been performed
Note 1 to entry: Such specimens/samples would be otherwise discarded as there is no remaining clinical need for them.
Note 2 to entry: This can include specimens collected for research or other purposes not connected to the clinical
performance study (3.5) in question.
[SOURCE: ISO 20916:2019, 3.25]
3.11
manufacturer
natural or legal person with responsibility for design and/or manufacture of a medical device with the
intention of making the medical device available for use under that person’s name, whether or not such a
medical device is designed and/or manufactured by that person or on that person’s behalf by another
person(s)
Note 1 to entry: Provisions of national or regional regulations can apply to the definition of manufacturer.
Note 2 to entry: ‘Design and/or manufacture’ can include specification development, production, fabrication, assembly,
processing, packaging, repackaging, labelling, relabelling, sterilization, installation, or remanufacturing of a medical
device; or putting a collection of devices, and possibly other product, together for a medical purpose.
Note 3 to entry: An authorized representative, distributor or importer who only adds its own address and contact
details to the medical device or the packaging, without covering or changing the existing labelling, is not considered a
manufacturer.
Note 4 to entry: Any person who assembles or adapts a medical device that has already been supplied by another
person for an individual patient, in accordance with the instructions for use, is not the manufacturer, provided the
assembly or adaptation does not change the intended use of the medical device.
Note 5 to entry: to the extent that an accessory is subject to the regulatory requirements of the IVD medical device
(3.9), the person responsible for the design and/or manufacture of that accessory is considered to be a manufacturer.
[SOURCE: ISO 18113-1:2022, 3.1.42]
3.12
pandemic
epidemic occurring worldwide, or over a very wide area, crossing international boundaries and usually
affecting a large number of people
[SOURCE: ISO 6028:2023, 3.3]
3.13
post-market surveillance
systematic process to collect and analyse experience gained from medical devices that have been placed on
the market
[SOURCE: ISO/TR 20416:2020, 3.2]
3.14
post-production
part of the life cycle of the medical device after the design has been completed and the medical device has
been manufactured
EXAMPLE Transportation, storage, installation, product use, maintenance, repair, product changes,
decommissioning and disposal.
[SOURCE: ISO 14971:2019, 3.12]
3.15
public health emergency
extraordinary event which is determined to constitute a public health risk to regions or countries through
the international spread of disease and to potentially require a coordinated international response
[18]
[SOURCE: WHO/IHR:2005 ]
3.16
quality management
management with regard to quality
Note 1 to entry: Quality management can include establishing quality policies and quality objectives, and processes to
achieve these quality objectives through quality planning, quality assurance, quality control, and quality improvement.
[SOURCE: ISO 9000:2015, 3.3.4]
3.17
risk management
systematic application of management policies, procedures and practices to the tasks of analysing,
evaluating, controlling and monitoring risk
[SOURCE: ISO/IEC Guide 63:2019, 3.15]
3.18
stability
ability of an IVD medical device (3.9) to maintain its performance characteristics within the limits specified
by the manufacturer (3.11)
Note 1 to entry: Stability applies to:
— IVD reagents, calibrators and controls, when stored, transported and used in the conditions specified by the
manufacturer;
— reconstituted lyophilised materials, working solutions and materials removed from sealed containers, when
prepared, used and stored according to the manufacturer’s instructions for use;
— measuring instruments or measuring systems after calibration.
Note 2 to entry: Stability of an IVD reagent or measuring system is normally quantified with respect to time
— in terms of the duration of a time interval over which a metrological property changes by a stated amount, or
— in terms of the change of a property over a stated time interval.
[SOURCE: ISO 18113-1:2022, 3.1.85]
3.19
stakeholder
person or organization that can affect, be affected by, or perceive themselves to be affected by a decision or
activity
Note 1 to entry: A decision maker can be a stakeholder.
[SOURCE: ISO 22367:2020, 3.39]
3.20
surrogate sample
material or combination of materials used as a substitute for body fluid or tissue taken for examination from
a single human subject to study the characteristic of interest
Note 1 to entry: Surrogate samples include but are not limited to:
— pooled patient samples of biological origin;
— materials supplemented (e.g. spiked) with an analyte of interest;
— material created to have properties similar to or representative of the body fluid or tissue of interest;
— material composed of a combination of an analyte that simulates the analyte of interest and a matrix created to
have properties similar to or representative of the body fluid or tissue or of the patient or subject;
— more-complex combinations of fabricated analyte and matrix.
[SOURCE: CLSI EP39:2021, 1.4.1]
4 General considerations for the design and development process
4.1 Safety and performance requirements
Manufacturers should follow a risk-based approach to define and justify the applicable essential general
safety and performance requirements.
NOTE In traditional regulatory pathways, to ensure the safety and performance of an IVD medical device,
manufacturers often comply with national regulatory requirements and guidelines such as the International
Medical Device Regulators Forum (IMDRF) harmonized Essential Principles (Reference [19]), when designing and
manufacturing a device, taking into account the intended use, intended patient population, intended user, and intended
use setting of the device.
If available, manufacturers should be aware of national procedures for the design, development and
registration of emergency use-IVD medical devices.
4.2 Quality management and risk management
Manufacturers should have pre-established quality management and risk management systems that can be
leveraged during the manufacturing of emergency use-IVD medical devices.
NOTE Standards for quality management and risk management systems that apply to the manufacturing of
medical devices include ISO 13485, good manufacturing practice (medical devices) regulations for each country, and
ISO 14971.
4.3 Target condition and scientific validity
Manufact
...








Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...