Capability of detection - Part 6: Methodology for the determination of the critical value and the minimum detectable value in Poisson distributed measurements by normal approximations

This document presents methods for determining the critical value of the response variable and the minimum detectable value in Poisson distribution measurements. It is applicable when variations in both the background noise and the signal are describable by the Poisson distribution. The conventional approximation is used to approximate the Poisson distribution by the normal distribution consistent with ISO 11843-3 and ISO 11843-4. The accuracy of the normal approximation as compared to the exact Poisson distribution is discussed in Annex B.

Capacité de détection — Partie 6: Méthodologie pour la détermination de la valeur critique et de la valeur minimale détectable pour les mesures distribuées selon la loi de Poisson approximée par la loi Normale

General Information

Status
Published
Publication Date
22-Oct-2025
Current Stage
6060 - International Standard published
Start Date
23-Oct-2025
Due Date
12-Jul-2026
Completion Date
23-Oct-2025

Relations

Effective Date
13-Jul-2024

Overview

ISO 11843-6:2025 - Capability of detection, Part 6 - defines a statistical methodology for determining the critical value and the minimum detectable value (MDV) when both signal and background follow a Poisson distribution. The standard uses the conventional normal approximation to the Poisson law (consistent with ISO 11843-3 and ISO 11843-4) to compute decision thresholds, and it discusses the accuracy of that approximation (Annex B).

Key topics and requirements

  • Scope and assumptions
    • Applicable when variations in background and signal are describable by the Poisson distribution.
    • Uses the normal approximation to derive variances, critical values and MDVs.
  • Decision criteria
    • Defines the critical value (yC) for the response variable to control the probability of false detection (α, error of the first kind).
    • Defines the MDV (gx) tied to the probability of failing to detect a true signal (β, error of the second kind).
  • Computation and reporting
    • Provides formulae for computing yC using standard-normal quantiles and observed means/variances (Clause 6).
    • Includes guidance on estimating mean and variance under normal approximation (Annex A) and examples (Annex D).
    • Specifies reporting requirements for capability assessments and applications (Clauses 7 and 8).
  • Practical measurement constraints
    • Both signal and background must be raw, unprocessed counts.
    • Prefer longer single measurements over many short ones (improves normal approximation).
    • Repeat measurements are required to estimate means reliably.
    • Detector must operate in its linear counting range; number of channels and FWHM must match between blank and sample (Annex C).

Applications and users

  • Targeted at laboratories and instrument operators using pulse-counting detectors and spectroscopy techniques where counts follow Poisson statistics:
    • X-ray fluorescence (XRF), X-ray diffraction (XRD), X-ray photoelectron spectroscopy (XPS)
    • Electron and ion spectroscopy (AES, SIMS), and mass spectrometry (GC–MS)
  • Typical uses:
    • Determining detection limits for hazardous-substance screening (e.g., RoHS testing).
    • Environmental monitoring, contamination control, material analysis.
    • Establishing decision thresholds in quality control and compliance testing.
  • Users include metrology labs, analytical chemists, instrument manufacturers, and accreditation bodies.

Related standards

  • ISO 11843 series (Parts 1–4) - general capability of detection principles and linear calibration cases.
  • ISO 3534-1 - statistical vocabulary and symbols.
  • ISO Guide 30 - reference materials terms.

Keywords: ISO 11843-6:2025, capability of detection, Poisson distribution, normal approximation, critical value, minimum detectable value, detection limit, pulse-counting, spectroscopy, XRF, XRD, GCMS.

Standard

ISO 11843-6:2025 - Capability of detection — Part 6: Methodology for the determination of the critical value and the minimum detectable value in Poisson distributed measurements by normal approximations Released:23. 10. 2025

English language
21 pages
sale 15% off
Preview
sale 15% off
Preview

Frequently Asked Questions

ISO 11843-6:2025 is a standard published by the International Organization for Standardization (ISO). Its full title is "Capability of detection - Part 6: Methodology for the determination of the critical value and the minimum detectable value in Poisson distributed measurements by normal approximations". This standard covers: This document presents methods for determining the critical value of the response variable and the minimum detectable value in Poisson distribution measurements. It is applicable when variations in both the background noise and the signal are describable by the Poisson distribution. The conventional approximation is used to approximate the Poisson distribution by the normal distribution consistent with ISO 11843-3 and ISO 11843-4. The accuracy of the normal approximation as compared to the exact Poisson distribution is discussed in Annex B.

This document presents methods for determining the critical value of the response variable and the minimum detectable value in Poisson distribution measurements. It is applicable when variations in both the background noise and the signal are describable by the Poisson distribution. The conventional approximation is used to approximate the Poisson distribution by the normal distribution consistent with ISO 11843-3 and ISO 11843-4. The accuracy of the normal approximation as compared to the exact Poisson distribution is discussed in Annex B.

ISO 11843-6:2025 is classified under the following ICS (International Classification for Standards) categories: 03.120.30 - Application of statistical methods; 17.020 - Metrology and measurement in general. The ICS classification helps identify the subject area and facilitates finding related standards.

ISO 11843-6:2025 has the following relationships with other standards: It is inter standard links to ISO 11843-6:2019. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

You can purchase ISO 11843-6:2025 directly from iTeh Standards. The document is available in PDF format and is delivered instantly after payment. Add the standard to your cart and complete the secure checkout process. iTeh Standards is an authorized distributor of ISO standards.

Standards Content (Sample)


International
Standard
ISO 11843-6
Third edition
Capability of detection —
2025-10
Part 6:
Methodology for the determination
of the critical value and the
minimum detectable value in
Poisson distributed measurements
by normal approximations
Capacité de détection —
Partie 6: Méthodologie pour la détermination de la valeur
critique et de la valeur minimale détectable pour les mesures
distribuées selon la loi de Poisson approximée par la loi Normale
Reference number
© ISO 2025
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on
the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address below
or ISO’s member body in the country of the requester.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii
Contents Page
Foreword .iv
Introduction .v
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 1
4 Symbols . 2
5 Measurement system and data handling . 2
6 Computation by approximation . 3
6.1 The critical value based on the normal distribution .3
6.2 Determination of the critical value of the response variable .4
6.3 Sufficient capability of the detection criterion .5
6.4 Confirmation of the sufficient capability of detection criterion .6
7 Reporting the results from an assessment of the capability of detection . 7
8 Reporting the results from an application of the method . 7
Annex A (informative) Estimating the mean value and variance when the Poisson distribution
is approximated by the normal distribution . 8
Annex B (informative) Accuracy of approximations . 9
Annex C (informative) Selecting the number of channels for the detector .15
Annex D (informative) Examples of calculations .16
Bibliography .21

iii
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out through
ISO technical committees. Each member body interested in a subject for which a technical committee
has been established has the right to be represented on that committee. International organizations,
governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely
with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are described
in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the different types
of ISO document should be noted. This document was drafted in accordance with the editorial rules of the
ISO/IEC Directives, Part 2 (see www.iso.org/directives).
ISO draws attention to the possibility that the implementation of this document may involve the use of (a)
patent(s). ISO takes no position concerning the evidence, validity or applicability of any claimed patent
rights in respect thereof. As of the date of publication of this document, ISO had not received notice of (a)
patent(s) which may be required to implement this document. However, implementers are cautioned that
this may not represent the latest information, which may be obtained from the patent database available at
www.iso.org/patents. ISO shall not be held responsible for identifying any or all such patent rights.
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and expressions
related to conformity assessment, as well as information about ISO's adherence to the World Trade
Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www.iso.org/iso/foreword.html.
This document was prepared by Technical Committee ISO/TC 69, Application of statistical methods,
Subcommittee SC 6, Measurement methods and results.
This third edition cancels and replaces the second edition (ISO 11843-6:2019), which has been technically
revised.
The main changes are as follows:
— the symbols were modified to conform to ISO 11843-1;
— a list of symbols was moved from Annex A to Clause 4;
— in 6.3, explanatory text of how to determine the minimum detectable value was added;
— Clause 8 was revised to provide a description of the appropriate approach for determining whether or
not the target substance has been detected;
— typographic and obvious errors were corrected.
A list of all parts in the ISO 11843 series can be found on the ISO website.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www.iso.org/members.html.

iv
Introduction
Many types of instruments use the pulse-counting method for detecting signals. X-ray, electron and ion-
spectroscopy detectors, such as X-ray diffractometers (XRD), X-ray fluorescence spectrometers (XRF), X-ray
photoelectron spectrometers (XPS), Auger electron spectrometers (AES), secondary ion mass spectrometers
(SIMS) and gas chromatograph mass spectrometers (GCMS) are of this type. These signals consist of a series
of pulses produced at random and irregular intervals. They can be understood statistically using a Poisson
distribution and the methodology for determining the minimum detectable value can be deduced from
statistical principles.
Determining the minimum detectable value of signals is sometimes important in practical work. The
value provides a criterion for deciding when “the signal is certainly not detected”, or when “the signal is
[1]to[8]
significantly different from the background noise level” . For example, it is valuable when measuring the
presence of hazardous substances or surface contamination of semi-conductor materials. RoHS (Restrictions
on Hazardous Substances) sets limits on the use of six hazardous materials (hexavalent chromium, lead,
mercury, cadmium and the flame retardant agents, perbromobiphenyl, PBB, and perbromodiphenyl ether,
PBDE) in the manufacturing of electronic components and related goods sold in the EU. For that application,
XRF and GCMS are the testing instruments used. XRD is used to measure the level of hazardous asbestos
and crystalline silica present in the environment or in building materials.
Although the methodology employed to determine the minimum detection values has long been established
in the field of chemical analysis, it has hitherto remained undefined within the domain of pulse count
measurements. The necessity of establishing a methodology for determining the minimum detectable value
[9]
in this field is duly acknowledged .
In this document the Poisson distribution is approximated by the normal distribution, ensuring consistency
with the IUPAC approach laid out in the ISO 11843 series. The conventional approximation is used to generate
the variance, the critical value of the response variable, the capability of detection criteria and the minimum
[10]
detectability level .
In this document:
— α is the probability of erroneously detecting that a system is not in the basic state, when really it is in
that state;
— β is the probability of erroneously not detecting that a system is not in the basic state when the value of
the state variable is equal to the minimum detectable value(x ).
D
v
International Standard ISO 11843-6:2025(en)
Capability of detection —
Part 6:
Methodology for the determination of the critical value
and the minimum detectable value in Poisson distributed
measurements by normal approximations
1 Scope
This document presents methods for determining the critical value of the response variable and the
minimum detectable value in Poisson distribution measurements. It is applicable when variations in
both the background noise and the signal are describable by the Poisson distribution. The conventional
approximation is used to approximate the Poisson distribution by the normal distribution consistent with
ISO 11843-3 and ISO 11843-4.
The accuracy of the normal approximation as compared to the exact Poisson distribution is discussed in
Annex B.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes
requirements of this document. For dated references, only the edition cited applies. For undated references,
the latest edition of the referenced document (including any amendments) applies.
ISO Guide 30, Reference materials — Selected terms and definitions
ISO 3534-1, Statistics — Vocabulary and symbols — Part 1: General statistical terms and terms used in
probability
ISO 11843-1, Capability of detection — Part 1: Terms and definitions
ISO 11843-2, Capability of detection — Part 2: Methodology in the linear calibration case
ISO 11843-3, Capability of detection — Part 3: Methodology for determination of the critical value for the
response variable when no calibration data are used
ISO 11843-4, Capability of detection — Part 4: Methodology for comparing the minimum detectable value with
a given value
3 Terms and definitions
For the purposes of this document, the terms and definitions given in ISO 3534-1, ISO 11843-1, ISO 11843-2,
ISO 11843-3, ISO 11843-4, and ISO Guide 30 apply.
ISO and IEC maintain terminology databases for use in standardization at the following addresses:
— ISO Online browsing platform: available at https:// www .iso .org/ obp
— IEC Electropedia: available at https:// www .electropedia .org/

4 Symbols
X
state variable
Y
response variable
number of replications of measurements on the reference material representing the value of the basic
J
state variable (blank sample)
K
number of replications of measurements on the actual state (test sample)
number of replications of measurements of each reference material in assessment of the capability
N
of detection
x
a value of state variable
y
a value of response variable
y
critical value of the response variable defined by ISO 11843-1 and ISO 11843-3
C
x
given value which is tested to determine whether it is greater than the minimum detectable value
g
x
minimum detectable value of the state variable
D
σ
standard deviation under actual performance conditions for the response in the basic state
b
σ
standard deviation under actual performance conditions for the response in a sample with the state
g
variable equal to x
g
η
expected value under the actual performance conditions for the response in the basic state
b
η
expected value under the actual performance conditions for the response in a sample with the state
g
variable equal to x
g
y
the arithmetic mean of the actual measured response in the basic state
b
y the arithmetic mean of the actual measured response in a sample with the state variable equal to x
g g
y minimum detectable response value with the state variable equal to x
D d
λ
mean value corresponding to the expected number of events in Poisson distribution
α
the probability that an error of the first kind has occurred
β
the probability that an error of the second kind has occurred
1−α
confidence level
1−β
confidence level
z ()1−α -quantile of the standard normal distribution
1−α
z ()1−β -quantile of the standard normal distribution
1−β
T
lower confidence limit
5 Measurement system and data handling
The conditions under which Poisson counts are made are usually specified by the experimental set-up. The
number of pulses that are detected increases with both the time and with the width of the region over which

the spectrum is observed. These two parameters should be noted and not changed during the course of the
measurement.
The following restrictions should be observed if the minimum detectable value is to be determined reliably:
a) Both the signal and the background noise should follow the Poisson distributions. The signal is the mean
value of the gross count.
b) The raw data should not receive any processing or treatment, such as smoothing.
c) Time interval: Measurement over a long period of time is preferable to several shorter measurements.
A single measurement taken for over one second is better than 10 measurements over 100 ms
(milliseconds) each. The approximation of the Poisson distribution by the normal distribution is more
reliable with higher mean values.
NOTE In the case of a measurement system that obtains an average signal of 2 000 counts/s, a single
measurement taken over 2 s yields to a measured value of 4 000 counts. In pulse count measurement, the standard
deviation is given by the square root of the measured value (see Annex A), which in this case is 63,20 counts, and
the coefficient of variation, C , is 0,016. On the other hand, if multiple measurements are taken with a measurement
V
time of 100 ms (milliseconds), an average measured value of 200 counts is obtained, with a standard deviation of
14,1 counts and a C of 0,071. Higher measured values provide greater accuracy than lower measured values.
V
d) The number of measurements: Since only mean values are used in the approximations presented here,
repeated measurements are needed to determine them. The power of test increases with the number of
measurements.
e) Number of channels used by the detector: There should be no overlap of neighbouring peaks. The
number of channels that are used to measure the background noise and the sample spectra should be
identical (see Figure C.1).
f) Peak width: The full width at half maximum (FWHM) is the recommended coverage for monitoring a
single peak. It is preferable the measurements are based on the top, the bottom or both, of a noisy peak.
The appropriate FWHM should be assessed beforehand by measuring a standard sample. An identical
value of the FWHM should be used for both the background noise and the sample measurements.
Additional factors are:
— the instrument should work correctly;
— the detector should be operating within its linear counting range;
— both the ordinate and the abscissa axes should be calibrated;
— there should be no signal that cannot be clearly identified as not being noise; degradation of the specimen
during measurement should be negligibly small;
— at least one signal or peak belonging to the element under consideration should be observable.
6 Computation by approximation
6.1 The critical value based on the normal distribution
The decision on whether a measured signal is significant or not can be made by comparing the arithmetic
mean y of the actual measured values with a suitably chosen value y . The value y , which is referred to
g C C
as the critical value, satisfies the requirement:
Py()>=yx 0 ≤α (1)
gC
where the probability is computed under the condition that the system is in the basic state (x = 0) and α is a
pre-selected probability of false decision.

Formula (1) gives the probability that yy> under the condition that:
gC
yy=+z σ + (2)
Cb 1−α b
JK
where
z is the (1 − α)-quantile of the standard normal distribution where 1 − α is the confidence level;
1−α
σ is the standard deviation under actual performance conditions for the response in the basic state;
b
y is the arithmetic mean of the actual measured response in the basic state;
b
J is the number of repeat measurements of the blank reference sample. This represents the value
of the basic state variable;
K is the number of repeat measurements of the test sample. This gives the value of the actual
state variable.
NOTE "The only + sign is used in Formula (2). In the pulse counting method the response variable is positive
integer and always increases as the state variable increases.
The definition of the critical value follows ISO 11843-1 and ISO 11843-3. Its relationship to the measured
values in the active and basic states is illustrated in Figure 1.
Key
X state variable
Y response variable
α probability that an error of the first kind has occurred
β probability that an error of the second kind has occurred
Figure 1 — Conceptual diagram showing the relative position of the critical value and the measured
values of the active and basic states
6.2 Determination of the critical value of the response variable
If the response variable follows a Poisson distribution with a sufficiently large mean value, the standard
deviation of the repeated measurements of the response variable in the basic state is estimated as y .
b
This is an estimate of σ . The standard deviation of the repeated measurements of the response variable in
b
the actual state of the sample is y , giving an estimate of σ (see Annex A).
g g
The critical value, y , of a response variable that follows the Poisson distribution approximated by the
C
normal distribution generally satisfies Formula (3):
11 11
yy=+z σ +≈ yz++y (3)
Cb 11−−ααbb b
JK JK
where y is the arithmetic mean of the actual measured response in the basic state.
b
6.3 Sufficient capability of the detection criterion
The sufficient capability of detection criterion enables decisions to be made about the detection of a signal
by comparing the critical value probability with a specified value of the confidence level, 1−β . If the
criterion is satisfied, it can be concluded that the minimum detectable value, x , is less than or equal to the
D
value of the state variable, x . The minimum detectable value then defines the smallest value of the response
g
variable, η , for which an incorrect decision occurs with a probability, β . At this value, there is no signal,
g
only background noise, and an ‘error of the second kind’ has occurred.
If the standard deviation of the response for a given value x is σ , the criterion for the probability to be
g g
greater than or equal to 1−β is set by Formula (4), from which Formulae (5) and (6) can be derived:
η ≥+yz σσ+ (4)
g Cb1−β g
JK
If y is replaced by yz=+ησ + , defined in Formulae (2) and (3), then:
C Cb 1−α b
JK
11 11
2 2
η −≥ησz ++z σσ+ (5)
gb 11−−αβbb g
JK JK
where
α is the probability that an error of the first kind has occurred;
β is the probability that an error of the second kind has occurred;
η is the expected value under the actual performance conditions for the response in the basic state;
b
η is the expected value under the actual performance conditions for the response in a sample
g
with the state variable
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...

ISO 11843-6:2025は、ポアソン分布測定における応答変数のクリティカルバリューおよび最小検出値の決定に関する方法論を提示します。この標準の主な目的は、背景雑音と信号の変動がポアソン分布で記述される場合における検出能力の標準化を図ることです。 この標準の強みは、ポアソン分布を正規分布で近似する従来の手法を利用しており、ISO 11843-3およびISO 11843-4との整合性を保っています。これにより、研究者や技術者は、測定結果の解釈において一貫した基準を持てるため、信頼性の高いデータ解析が可能になります。 さらに、附属書Bでは、正規近似と正確なポアソン分布との比較について議論されています。この分析により、近似手法の精度を理解し、実際の測定における限界を認識することができるため、非常に価値のある情報を提供しています。 したがって、ISO 11843-6:2025は、ポアソン分布に基づく測定法の標準化において重要な役割を果たしており、関連する研究や産業の分野で非常に有用です。この標準に従うことで、検出能力の評価や測定結果の信頼性向上が期待できるため、関連者にとって必須の文書と言えるでしょう。

La norme ISO 11843-6:2025 présente une méthodologie essentielle pour la détermination de la valeur critique et de la valeur minimale détectable dans des mesures distribuées selon la loi de Poisson. Cette standardisation est particulièrement pertinente pour les domaines où des variations du bruit de fond et du signal peuvent être modélisées par cette distribution. En intégrant des méthodes conventionnelles d'approximation de la distribution de Poisson à une distribution normale, cette norme assure une transition entretenue avec les ISO 11843-3 et ISO 11843-4, consolidant ainsi son cadre théorique. Les principaux atouts de l'ISO 11843-6:2025 résident dans son approche rigoureuse à définir des critères quantitatifs pour la détection. La norme permet non seulement d'évaluer les seuils critiques dans des contextes variés, mais elle garantit également une fiabilité accrue des méthodes d'analyse. En discutant dans l'Annexe B la précision de l'approximation normale par rapport à la distribution de Poisson exacte, le document offre des indications précieuses sur les limites des approches statistiques en fonction des conditions expérimentales. La pertinence de l'ISO 11843-6:2025 s'étend à des applications variées dans les sciences physiques et l'ingénierie, où des mesures précises et fiables sont cruciales. Son emploi contribue à améliorer la qualité des données recueillies, en permettant aux professionnels de faire des choix éclairés basés sur des méthodologies prouvées et normalisées. Par conséquent, cette norme se positionne comme un élément incontournable pour les praticiens souhaitant optimiser l'analyse des mesures basées sur des distributions de type Poisson.

Die Norm ISO 11843-6:2025 bietet eine umfassende Methodologie zur Bestimmung des kritischen Wertes und des minimalen nachweisbaren Wertes in von Poisson verteilten Messungen. Der Geltungsbereich dieser Norm ist besonders relevant für Anwendungen, in denen sowohl Hintergrundrauschen als auch das Signal durch die Poisson-Verteilung beschrieben werden können. Dies stellt sicher, dass die Norm in einer Vielzahl von praktischen Szenarien anwendbar ist, in denen statistische Zuverlässigkeit entscheidend ist. Ein wesentlicher Stärke der ISO 11843-6:2025 liegt in der Verwendung der herkömmlichen Annäherung, die die Poisson-Verteilung durch die Normalverteilung approximiert. Diese Annäherung ist konsistent mit den zuvor veröffentlichten Normen ISO 11843-3 und ISO 11843-4, was eine nahtlose Integration und Anwendung in bestehenden Standards ermöglicht. Insbesondere wird im Anhang B der Norm die Genauigkeit dieser Normalannäherung im Vergleich zur exakten Poisson-Verteilung diskutiert, was zusätzliche Klarheit über das Verhalten der statistischen Modelle bietet. Die Relevanz dieser Norm erstreckt sich über verschiedene Fachgebiete, insbesondere in der Messtechnik und Qualitätskontrolle, wo die Ermittlung von kritischen Werten und nachweisbaren Grenzwerten von großer Bedeutung ist. Durch die Bereitstellung einer klaren und strukturierten Methodik zur Analyse und Verifizierung von Messdaten trägt ISO 11843-6:2025 zur Verbesserung der Genauigkeit und Zuverlässigkeit in der Datenauswertung bei. Zusammenfassend lässt sich sagen, dass die ISO 11843-6:2025 eine wichtige Ressource für Fachleute darstellt, die sich mit Messungen in Poisson-verteilten Umgebungen beschäftigen. Ihre methodischen Ansätze sind sowohl praktisch als auch theoretisch fundiert und bieten einen wertvollen Leitfaden für die Bestimmung von kritischen und nachweisbaren Werten in der wissenschaftlichen und industriellen Praxis.

The standard ISO 11843-6:2025 provides a comprehensive methodology for the determination of critical values and minimum detectable values in measurements that are Poisson distributed, making it a vital resource for professionals working in fields that require high statistical accuracy in their data assessments. One of the key strengths of ISO 11843-6:2025 is its focus on utilizing conventional approximations to simplify the analysis of Poisson-distributed data by leveraging the normal distribution, thus aligning with prior standards ISO 11843-3 and ISO 11843-4. This integration not only ensures consistency across methodologies but also enhances the practicality of the standard for practitioners. Additionally, the document meticulously addresses how to apply these approximations effectively, broadening its applicability to situations where both the background noise and signal variations are governed by the Poisson distribution. This focus on the common characteristics of the data will resonate with users looking for reliable guidance in determining the critical value and the minimum detectable value in their experimental measurements. Furthermore, Annex B of the standard enhances its relevance by discussing the accuracy of the normal approximation compared to the exact Poisson distribution. This critical comparison enables users to understand the limitations and advantages of employing the normal approximation, thus informing their decisions and enhancing the overall robustness of their measurement interpretations. In summary, ISO 11843-6:2025 stands out as a significant standard in the field of measurement, providing essential methodologies that improve the detection capabilities for Poisson distributed measurements through carefully articulated statistical approximations. Its alignment with existing standards and the comprehensive discussion of accuracy solidify its position as a key reference for practitioners aiming to optimize their measurement approaches.

ISO 11843-6:2025 표준은 포아송 분포 측정에서 반응 변수의 한계값과 최소 검출 가능 값을 결정하기 위한 방법론을 제시합니다. 이 표준의 주요 범위는 포아송 분포에 의해 설명 가능한 배경 잡음과 신호의 변동이 있는 상황에서 적용됩니다. 특히, ISO 11843-3 및 ISO 11843-4와 일관성을 유지하며 포아송 분포를 정규 분포로 근사하는 데 필요한 전통적인 근사 방식을 소개합니다. 이 표준의 강점은 포아송 측정의 정확한 결과를 도출할 수 있도록 하는 확립된 방법론을 제공한다는 점입니다. 또한, 부록 B에서는 정규 근사가 정확한 포아송 분포와 비교할 때의 정확성에 대해 상세히 논의하고 있어, 실무자들이 해당 방법론의 신뢰성을 평가하는 데 도움을 줄 수 있습니다. 이러한 정보들은 연구 및 실험 설계 시 중요한 참고자료가 됩니다. ISO 11843-6:2025는 이러한 측정의 정밀도와 신뢰성을 향상시킬 수 있는 기초 자료로, 포아송 분포에 대한 이해를 돕고 관련 작업 과정의 효율을 증가시킬 수 있는 이러한 방법론은 현재와 미래의 연구 및 산업 적용에 있어 매우 중요한 표준입니다.