Optics and optical instruments — Focimeters — Part 1: General purpose instruments

ISO 8598-1:2014 specifies requirements and test methods for general purpose focimeters designed for the measurement of vertex powers, cylinder axis, prismatic power and prism base setting within a restricted area at a specified location of a lens. This excludes instruments that can only measure the whole lens at once. It is applicable to instruments typically intended for use by the ophthalmic community, with the capability to demonstrate conformity of spectacle lens products with the International Standards existing for these lenses.

Optique et instruments d'optique — Frontofocomètres — Partie 1: Instruments pour cas généraux

L'ISO 8598-1:2014 spécifie les exigences et méthodes d'essai des frontofocomètres pour cas généraux destinés à mesurer les puissances frontales, l'axe de cylindre, la puissance prismatique et les réglages de la base du prisme sur une surface définie d'un point spécifique d'un verre. Elle exclut donc tout instrument qui ne peut mesurer que la totalité du verre à la fois. Elle s'applique aux instruments typiquement destinés à la communauté ophtalmologique, dans le but de démontrer la conformité des verres de lunettes avec les Normes internationales en vigueur pour ces produits.

General Information

Status
Published
Publication Date
15-Sep-2014
Current Stage
9093 - International Standard confirmed
Completion Date
08-Oct-2019
Ref Project

Relations

Buy Standard

Standard
ISO 8598-1:2014 - Optics and optical instruments -- Focimeters
English language
27 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
ISO 8598-1:2014 - Optics and optical instruments -- Focimeters
English language
27 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
ISO 8598-1:2014 - Optique et instruments d'optique -- Frontofocomètres
French language
27 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
ISO 8598-1:2014 - Optique et instruments d'optique -- Frontofocometres
French language
27 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

INTERNATIONAL ISO
STANDARD 8598-1
First edition
2014-09-15
Optics and optical instruments —
Focimeters —
Part 1:
General purpose instruments
Optique et instruments d’optique — Frontofocomètres —
Partie 1: Instruments pour cas généraux
Reference number
ISO 8598-1:2014(E)
©
ISO 2014

---------------------- Page: 1 ----------------------
ISO 8598-1:2014(E)

COPYRIGHT PROTECTED DOCUMENT
© ISO 2014
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized otherwise in any form
or by any means, electronic or mechanical, including photocopying, or posting on the internet or an intranet, without prior
written permission. Permission can be requested from either ISO at the address below or ISO’s member body in the country of
the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland
ii © ISO 2014 – All rights reserved

---------------------- Page: 2 ----------------------
ISO 8598-1:2014(E)

Contents Page
Foreword .iv
Introduction .v
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 1
4 Technical requirements for general purpose focimeters . 3
5 Metrological requirements . 5
5.1 General . 5
5.2 Reference wavelength . 5
5.3 Performance requirement . 5
6 Test procedures . 8
6.1 General . 8
6.2 Checking the indication errors . 8
6.3 Checking the axis marker for the optical centre of lens . 8
6.4 Checking the alignment of the axis marker . 9
6.5 Checking the adjusting rail . 9
6.6 Checking the non-symmetric error for automated focimeters . 9
6.7 Checking the repeatability of vertex power measurement for automated focimeters .10
6.8 Checking for the centration error .10
6.9 Checking the capability of focimeters to measure tinted lenses .10
6.10 Checking the astigmatic axis repeatability for low-powered cylinder lens .10
6.11 Special procedures for eyepiece focimeters .11
6.12 Criterion for image focusing in manual focusing focimeters .11
7 Marking .11
7.1 Reference to ISO 8598-1 .11
7.2 General information to be supplied by the manufacturer .11
7.3 Additional information to be supplied by the manufacturer .12
Annex A (informative) Use of correction values when measuring spectacle lenses .13
Annex B (informative) Example for evaluation of uncertainty of measurement for automated
focimeters for general use .18
Annex C (normative) Specifications of special reference lenses .24
Annex D (informative) Information for users on the performance of general purpose focimeters
covered by this part of ISO 8598 .26
Bibliography .27
© ISO 2014 – All rights reserved iii

---------------------- Page: 3 ----------------------
ISO 8598-1:2014(E)

Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of
any patent rights identified during the development of the document will be in the Introduction and/or
on the ISO list of patent declarations received (see www.iso.org/patents).
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation on the meaning of ISO specific terms and expressions related to conformity
assessment, as well as information about ISO’s adherence to the WTO principles in the Technical Barriers
to Trade (TBT) see the following URL: Foreword - Supplementary information
The committee responsible for this document is ISO/TC 172, Optics and photonics, Subcommittee SC 7,
Ophthalmic optics and instruments.
This first edition of ISO 8598-1 cancels and replaces ISO 8598:1996, of which it constitutes a technical
revision. It also incorporates the Technical Corrigendum ISO 8598:1996/Cor.1:1998.
ISO 8598 consists of the following parts, under the general title Optics and optical instruments —
Focimeters:
— Part 1: General purpose instruments
iv © ISO 2014 – All rights reserved

---------------------- Page: 4 ----------------------
ISO 8598-1:2014(E)

Introduction
General purpose focimeters are intended for measurement of contact lenses, single-vision, multifocal
and progressive-power or degressive-power spectacle lenses, both uncut and mounted in spectacle
frames, and for the orientation and marking of spectacle lenses.
© ISO 2014 – All rights reserved v

---------------------- Page: 5 ----------------------
INTERNATIONAL STANDARD ISO 8598-1:2014(E)
Optics and optical instruments — Focimeters —
Part 1:
General purpose instruments
1 Scope
This part of ISO 8598 specifies requirements and test methods for general purpose focimeters designed
for the measurement of vertex powers, cylinder axis, prismatic power and prism base setting within a
restricted area at a specified location of a lens. This excludes instruments that can only measure the
whole lens at once.
It is applicable to instruments typically intended for use by the ophthalmic community, with the
capability to demonstrate conformity of lens products with the International Standards existing for
these lenses.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.
ISO 7944, Optics and optical instruments — Reference wavelengths
ISO 8429, Optics and optical instruments — Ophthalmology — Graduated dial scale
ISO 9342-1, Optics and optical instruments — Test lenses for calibration of focimeters — Part 1: Test lenses
for focimeters used for measuring spectacle lenses
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
3.1
adjusting rail
movable rail or bar used as the reference axis for spectacles during measurement and which is aligned
perpendicularly to the optical axis of the focimeter and parallel to the axis direction 0° to 180°
Note 1 to entry: This is also called the lens table or frame rest.
3.2
capability
ability of a system or process to achieve the required performance
3.3
general purpose focimeter
instrument that is used to measure vertex powers, cylinder axis and prismatic effects of spectacle
and contact lenses, to orientate and mark uncut lenses, and to verify the correct mounting of lenses in
spectacle frames
© ISO 2014 – All rights reserved 1

---------------------- Page: 6 ----------------------
ISO 8598-1:2014(E)

3.3.1
manual focusing focimeter
instrument that allows the operator to view images formed by rays of light passing through a lens and,
by manually focusing and adjusting, to measure the vertex power and identify the principal meridians
Note 1 to entry: For lenses with cylindrical power, the cylinder axis is found using the method provided to locate
the principal meridians of the lens in the area defined by the focimeter aperture. Prismatic power is measured
separately for this type of focimeter.
Note 2 to entry: There are sub-classes of manual focusing focimeter. One type has an eyepiece whereas the other
has a projection screen. With the eyepiece type, the measurement target is focused and viewed through an
eyepiece.
3.3.2
automated focimeter
instrument that measures the vertex power of a lens, in the area defined by the focimeter aperture, in a
single measurement without operator adjustment
3.3.3
continuously indicating focimeter
focimeter with a continuous scale
Note 1 to entry: For the purposes of this part of ISO 8598, this includes both automated instruments when set to
0,06 D or 0,01 D steps and conventional manual focusing instruments.
3.3.4
digitally rounding focimeter
focimeter which displays measured values rounded to the nearest 0,25 D or 0,12(5) D incremental value
3.4
centration error of the instrument
residual prismatic error of the instrument with no lens in place
3.5
indication
〈of a focimeter〉 quantitative value provided as the output of the focimeter
3.6
indication error
difference between the value indicated by the focimeter and the true value of the reference lens
Note 1 to entry: Here the true value of the reference lens, the back vertex power, is calculated using the four
known basic lens parameters: radii of curvature of the front and back surface (r and r ), the centre thickness (t),
1 2
and the refractive index (n) of the reference lens material, using the formulas listed in ISO 9342-1.
Note 2 to entry: When using a measuring instrument, the influence of the uncertainty and indication error of the
device should be considered.
3.7
lens support
mechanical interface of the instrument against which the spectacle lens or the contact lens is placed for
measurement
Note 1 to entry: The focimeter measures the vertex power related to the surface placed against the lens support.
3.8
near portion power
vertex power measured at the near design reference point, as specified by the manufacturer, of a
multifocal, progressive-power or degressive-power lens
2 © ISO 2014 – All rights reserved

---------------------- Page: 7 ----------------------
ISO 8598-1:2014(E)

3.9
non-symmetric error for cylindrical power and cylinder axis
residual error in the indicated cylindrical power and/or the indicated cylinder axis of a spherocylindrical
lens for an automated focimeter after calibration
3.10
non-symmetric prism error of a focimeter
difference in the prismatic power readings when a plano-prism is measured first with its base setting
in one direction and then in the opposite direction, for example, base settings of 180° and 360°, or 90°
and 270°
3.11.1
back vertex power
reciprocal of the paraxial value of the back vertex focal length measured in metres
3.11.2
front vertex power
reciprocal of the paraxial value of the front vertex focal length measured in metres
Note 1 to entry: Conventionally, the back vertex power, in dioptres, is specified as the “power” of a spectacle lens or
a contact lens, although the front vertex power is required for certain purposes, for example in the measurement
of some multifocal lenses or progressive-power lenses.
−1
Note 2 to entry: The unit for expressing vertex power is the reciprocal metre (m ). The name for this unit is the
“dioptre”, for which the symbol is “D”.
4 Technical requirements for general purpose focimeters
4.1 The measuring range shall include vertex powers with a range from at least −20 D to +20 D and
prismatic powers from 0 Δ to at least 5 Δ.
The instrument shall be capable of measuring the axis direction (see ISO 8429) of cylindrical lenses
between 0° and 180°. For prisms, it shall be possible to determine the direction of the base setting
between 0° and 360°.
4.2 For manual focusing focimeters with non-digital display, the dioptric power scale shall have an
interval not greater than 0,25 D and shall be clear enough for interpolations to be made to the nearest
0,12 D or less.
For axis directions (see ISO 8429) the scale interval shall not exceed 5° and shall be clear enough for
interpolations to be made to the nearest degree.
For prismatic power readings, the scale interval shall not exceed 1 Δ and shall be clear enough for
interpolations to be made to the nearest 0,50 Δ.
4.3 For focimeters with digital display in the range from +20 D to −20 D, the minimum step of the
display shall be equal to or lower than 0,06 D. The display shall show at least two decimal digits.
For axis directions, the increment of the digital display shall be 1°.
For prismatic power readings, the minimum step of the digital display shall be equal to or lower than
0,06 Δ.
4.4 The instrument designed to measure spectacle lenses shall be able to measure lenses with a
diameter of at least 80 mm and a thickness of at least 20 mm. Translational movements of the lenses
on the lens support of not less than 30 mm in a direction perpendicular to the optical axis and to the
adjusting rail shall be possible, starting from no more than 10 mm below, in front of, or behind, the optical
axis of the instrument, as applicable. The adjusting rail shall also be capable of movement of not less than
30 mm in a direction perpendicular to its length and the optical axis of the instrument. See Figure 1.
© ISO 2014 – All rights reserved 3

---------------------- Page: 8 ----------------------
ISO 8598-1:2014(E)

Dimensions in millimetres
Key
1 lens support
2 adjusting rail
Figure 1 — Minimum required movement of the adjusting rail
4.5 The lens support should be designed so that it does not cause any damage to the lens under test
when used in the manner recommended by the manufacturer.
It shall be designed and manufactured so that it will support a lens placed against it, holding the surface
that is in contact with it perpendicular to the optical axis of the focimeter.
The lens support shall not adversely affect the accuracy of measurement by introducing sagittal error.
Examples of suitable lens supports are shown in Figure 2.
a) b)
4 © ISO 2014 – All rights reserved

---------------------- Page: 9 ----------------------
ISO 8598-1:2014(E)

Key
d , d internal diameter of support
1 c
d d external diameter of support
2, e
d = (d + d )/2
3 1 2
h , h height of support
s c
r = (d – d )/4
1 2 1
r = (d – d )/4
2 e c
Figure 2 — Examples of lens supports for spectacle lenses (a) and contact lenses (b)
For spectacle lenses (a), d should be in the range 5 mm to 8 mm while for contact lenses (b), d should be
1 c
4,5 mm ± 0,5 mm. Because of the increased sagittal depth of steeply-curved contact lenses, the contact
lens support is usually smaller in diameter and slightly shorter. The height difference (h – h ) should be
s c
0,55 mm ± 0,02 mm.
NOTE The dimensions of the contact lens support are as specified in ISO 18369-3.
The internal diameter (d , d ) of the lens support for focimeters used for spectacle lens measurement or
1 c
for contact lens measurement shall be stated by the manufacturer.
4.6 The pin mark printed by the axis marker shall be small enough to enable the distance between a
first and a second mark to be distinguished.
NOTE A diameter of 0,7 mm is recommended.
4.7 The instrument shall be designed so that it gives stable measured values in a normal environment
(i.e. a temperature of 23 °C ± 5 °C, and a relative humidity of less than 85 %).
5 Metrological requirements
5.1 General
When tested with the reference lenses specified in ISO 9342-1, both manual focusing and automated
focimeters shall give indication readings for vertex power and prismatic power over their entire claimed
measuring range, and the indication error for the respective true value shall not exceed the limits given
in Tables 1 and 2.
5.2 Reference wavelength
The instrument shall be manufactured to indicate dioptric powers with reference to the green mercury
line λ = 546,07 nm or the yellow helium line λ = 587,56 nm, as given in ISO 7944, and the manufacturer
e d
shall indicate explicitly which wavelength has been selected.
NOTE If the light source used in the focimeter is not centred on the green mercury line λ = 546,07 nm or the
e
yellow helium line λ = 587,56 nm, the user should be able either to set the instrument, or to apply corrections to
d
the indicated values, for lenses made from materials other than the one that was used to calibrate the instrument.
5.3 Performance requirement
5.3.1 Indication error
When tested in accordance with 6.2.1, the instrument, when fitted with the spectacle lens support, shall
not exceed the permissible indication errors given in Tables 1 and 2 when used with the spectacle form
reference lenses specified in ISO 9342-1.
© ISO 2014 – All rights reserved 5

---------------------- Page: 10 ----------------------
ISO 8598-1:2014(E)

The maximum permissible indication errors given in Tables 1 and 2 correspond to the use of the reference
lenses specified in ISO 9342-1 with the respective true values.
Table 1 — Maximum permissible indication error on measured vertex power for general
purpose instruments
Values in dioptres (D)
Maximum permissible indication error
Instruments with Digitally rounding instruments
Measuring range of vertex power
continuous scale when set to increments of
a
0,25 0,12(5)
< 0 ≥ −5 > 0 ≤ +5 ±0,06 0,0 0,0
< −5 ≥ −10 > +5 ≤ +10 ±0,09 0,0 ±0,12(5)
< −10 ≥ −15 > +10 ≤ +15 ±0,12 0,0 ±0,12(5)
< −15 ≥ −20 > +15 ≤ +20 ±0,18 ±0,25 ±0,12(5)
< −20 > +20 ±0,25 ±0,25 ±0,25
NOTE 1 Calibrating to tighter tolerances (e.g. 10 % of the product tolerance) will improve instrument equivalence and
reduce possible differences between two instruments.
NOTE 2 The user should be aware that when a focimeter is set to read in steps of 0,01 D, the readings may not be true to this
level of precision.
a
The meaning of 0,12(5) is that the instruments has been set to 1/8th dioptre steps, displayed as 0,12 D.
Table 2 — Maximum permissible indication errors on measured prismatic power for general
purpose instruments
Values in prism dioptres (Δ)
Maximum permissible indication error
Measuring range of prismatic Instruments with Digitally rounding instruments
power continuous scale when set to increments of
a
0,25 0,12(5)
> 0 ≤ 5 0,25 0 0,12(5)
> 5 ≤ 10 0,25 0,25 0,25
> 10 ≤ 15 0,50 0,50 0,50
> 15 ≤ 20 0,75 0,75 0,75
> 20 1,00 1,00 1,00
NOTE Calibrating to tighter tolerances (e. g. to 10 % of the product tolerance) will improve instrument equivalence and
reduce possible differences between two instruments.
a
The meaning of 0,12(5) is that the instruments has been set to 1/8th dioptre steps, displayed as 0,12 D.
For automated focimeters, when any of the spherical reference lenses is measured after correct
centration, it shall indicate less than 0,06 D of astigmatic power.
5.3.2 Axis marker for the optical centre of lens
When tested in accordance with 6.3, the position of the axis marker for the optical centre of the lens
shall not deviate from the optical axis of the focimeter by more than 0,4 mm.
5.3.3 Axis marker alignment
When tested in accordance with 6.4, any misalignment of the orientation of the axis marker shall not
exceed the tolerance of ±1° from the 0° to 180° direction of the dial scale for manual focusing instruments
or at the 0° (or 180°) value of the indicated axis for automated focimeters.
6 © ISO 2014 – All rights reserved

---------------------- Page: 11 ----------------------
ISO 8598-1:2014(E)

5.3.4 Adjusting rail
When tested in accordance with 6.5, any misalignment of the adjusting rail shall not exceed the
tolerance of 1° from the position parallel to the 0° to 180° directions of the dial scale for manual focusing
instruments or the 0° (or 180°) value of the indicated axis for automated focimeters.
5.3.5 Capability of measuring tinted lenses
When tested in accordance with 6.9, focimeters should be capable of measuring tinted lenses with
luminous transmittance τ ≥ 18 %. The indicated value from measuring the reference lens and neutral
V
reference filter as specified in Clause C.4 shall remain within the maximum permissible indication
errors of Table 1. The limited capability of measuring tinted lenses shall be stated if the instrument
cannot meet the requirement of measuring tinted lenses with luminous transmittance τ ≥ 18 %.
V
NOTE In practice it may be difficult to measure lenses whose colour departs significantly from neutral if the
transmittance of the lens at the operating wavelength of the focimeter is less than 18 %.
5.3.6 Non-symmetric errors for automated focimeters
5.3.6.1 General
The cylindrical power, cylinder axis and prism error requirements are for automated focimeters that
measure sphere, cylinder, cylinder axis and prism simultaneously.
5.3.6.2 Non-symmetric error for cylindrical power
When tested in accordance with 6.6.2, the non-symmetric error for the indicated power readings of the
spherocylindrical-power reference lens specified in Clause C.3 shall not exceed 0,06 D sph/0,09 D cyl.
5.3.6.3 Non-symmetric error for cylinder axis
When tested in accordance with 6.6.3, the non-symmetric error for the cylinder axis from each reading
of the spherocylindrical-power reference lens specified in Clause C.3 for each orientation shall not
exceed ±1°.
5.3.6.4 Non-symmetric prism error
When tested in accordance with 6.6.4, the non-symmetric error for prismatic power for an automated
focimeter shall not exceed 0,06 Δ. At the same time, the indicated values for the spherical and cylindrical
powers shall not exceed ± 0,06 D.
5.3.7 Repeatability for the indication reading of automated focimeters
When tested in accordance with 6.7, the repeatability for indication readings for an automated focimeter
shall not exceed 0,06 D within the spherical power range −10 D to +10 D, and shall not exceed 0,12 D
within the measuring range over ±10 D.
5.3.8 Centration error for manual focusing focimeters
When tested in accordance with 6.8, the residual prismatic error of the focimeter with no lens in place
shall not exceed 0,1 Δ.
5.3.9 Astigmatic axis repeatability for low-powered cylindrical lenses with manual focimeters
When tested in accordance with 6.10, the standard deviation of the axis readings for a 0,25 D cylindrical
lens shall be less than 4°.
© ISO 2014 – All rights reserved 7

---------------------- Page: 12 ----------------------
ISO 8598-1:2014(E)

6 Test procedures
6.1 General
The reference lenses described in ISO 9342-1 shall be used to verify that the requirements of 5.1 to 5.3
are met.
The initial calibration of the focimeter and the metrological verification shall be carried out using all of
those reference lenses in the reference lens set specified in ISO 9342-1 that are within the measuring
range of the instrument.
Manual focusing focimeters should, before use, be set up according to 6.11 and the target focused
according to 6.12.
6.2 Checking the indication errors
6.2.1 Checking the indication errors for vertex power
The spherical reference lenses shall be used to check whether the vertex powers measured by the
focimeter meet the permissible indication errors shown in Table 1.
Ensure that the lens support used is the correct one for the reference lens set selected according to 6.1.
Place each individual reference lens with its back surface against the lens support, and centre it on the
optical axis of the focimeter.
Measure and record the vertex power.
It is recommended that three independent readings be taken for each reference lens and that the mean
value be calculated and taken as the actual measured value of the lens.
NOTE The term “independent reading” means that the reference lens is removed from the support and
replaced again between each reading.
6.2.2 Checking the indication errors for prismatic power
The prismatic reference lenses shall be used to check whether the prismatic powers measured by the
focimeter meet the permissible indication errors shown in Table 2.
For prismatic power measurement, it is recommended that for each of the reference prisms, an
independent reading be taken corresponding to each of the 180° and 360° base setting directions. Take
the larger reading of the two measured values at 180° and 360° directions as the actual measured value
of the reference lens in order to calculate the indication errors.
6.3 Checking the axis marker for the optical centre of lens
A spherical reference lens of at least +15 D shall be used to check whether the axis marker for the optical
centre of lens meets the requirement in 5.3.3.
Centre the lens so that the indicated prism power is zero, and then mark it with the axis marker. Rotate
the reference lens through approximately 180° and re-centre to zero indicated prism power. Re-mark the
lens. Half the distance between the centres of the central marks from the first and second measurements
is the distance between the position of the axis marker for the optical centre of the lens and the optical
axis of the focimeter.
8 © ISO 2014 – All rights reserved

---------------------- Page: 13 ----------------------
ISO 8598-1:2014(E)

6.4 Checking the alignment of the axis marker
The 5 D plano-cylindrical reference lens as specified by ISO 9342-1 shall be used to check whether the
alignment of the axis marker meets the requirement in 5.3.2. The axis marker shall be checked using the
horizontal centreline on the reference lens.
For manual focusing focimeters, place the 5 D plano-cylindrical reference lens on t
...

FINAL
INTERNATIONAL ISO/FDIS
DRAFT
STANDARD 8598-1
ISO/TC 172/SC 7
Optics and optical instruments —
Secretariat: DIN
Focimeters —
Voting begins on:
2012-09-13
Part 1:
Voting terminates on:
General purpose instruments
2012-11-13
Optique et instruments d’optique — Frontofocomètres —
Partie 1: Instruments pour cas généraux
Please see the administrative notes on page iii
RECIPIENTS OF THIS DRAFT ARE INVITED TO
SUBMIT, WITH THEIR COMMENTS, NOTIFICATION
OF ANY RELEVANT PATENT RIGHTS OF WHICH
THEY ARE AWARE AND TO PROVIDE SUPPOR TING
DOCUMENTATION.
IN ADDITION TO THEIR EVALUATION AS
Reference number
BEING ACCEPTABLE FOR INDUSTRIAL, TECHNO-
ISO/FDIS 8598-1:2012(E)
LOGICAL, COMMERCIAL AND USER PURPOSES,
DRAFT INTERNATIONAL STANDARDS MAY ON
OCCASION HAVE TO BE CONSIDERED IN THE
LIGHT OF THEIR POTENTIAL TO BECOME STAN-
DARDS TO WHICH REFERENCE MAY BE MADE IN
©
NATIONAL REGULATIONS. ISO 2012

---------------------- Page: 1 ----------------------
ISO/FDIS 8598-1:2012(E)
Copyright notice
This ISO document is a Draft International Standard and is copyright-protected by ISO. Except as
permitted under the applicable laws of the user’s country, neither this ISO draft nor any extract
from it may be reproduced, stored in a retrieval system or transmitted in any form or by any means,
electronic, photocopying, recording or otherwise, without prior written permission being secured.
Requests for permission to reproduce should be addressed to either ISO at the address below or ISO’s
member body in the country of the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Reproduction may be subject to royalty payments or a licensing agreement.
Violators may be prosecuted.
ii © ISO 2012 – All rights reserved

---------------------- Page: 2 ----------------------
ISO/FDIS 8598-1:2012(E)
ISO/CEN PARALLEL PROCESSING
This final draft has been developed within the International Organization for Standardization (ISO), and pro-
cessed under the ISO-lead mode of collaboration as defined in the Vienna Agreement. The final draft was
established on the basis of comments received during a parallel enquiry on the draft.
This final draft is hereby submitted to the ISO member bodies and to the CEN member bodies for a parallel
two-month approval vote in ISO and formal vote in CEN.
Positive votes shall not be accompanied by comments.
Negative votes shall be accompanied by the relevant technical reasons.
© ISO 2012 – All rights reserved iii

---------------------- Page: 3 ----------------------
ISO/FDIS 8598-1:2012(E)
Contents Page
Foreword .v
Introduction .vi
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 1
4 Technical requirements for general purpose focimeters . 3
5 Metrological requirements . 5
5.1 General . 5
5.2 Reference wavelength . 5
5.3 Performance requirement . 5
6 Test procedures . 8
6.1 General . 8
6.2 Checking the indication errors . 8
6.3 Checking the axis marker for the optical centre of lens . 8
6.4 Checking the alignment of the axis marker . 9
6.5 Checking the adjusting rail . 9
6.6 Checking the non-symmetric error for automated focimeters . 9
6.7 Checking the repeatability of vertex power measurement for automated focimeters .10
6.8 Checking for the centration error .10
6.9 Checking the capability of automated focimeters to measure tinted lenses .10
6.10 Checking the astigmatic axis repeatability for low-powered cylinder lens .10
6.11 Special procedures for eyepiece focimeters .11
6.12 Criterion for image focusing in manual focusing focimeters .11
7 Marking .11
7.1 Reference to ISO 8598-1 .11
7.2 General information to be supplied by the manufacturer .11
7.3 Additional information to be supplied by the manufacturer .12
Annex A (informative) Use of correction values when measuring spectacle lensesor
contact lenses .13
Annex B (informative) Example for evaluation of uncertainty of measurement for automated
focimeters for general use .18
Annex C (normative) Specifications of special reference lenses .24
Annex D (informative) Information for users on the performance of general purpose focimeters
covered by this part of ISO 8598 .26
Bibliography .27
iv © ISO 2012 – All rights reserved

---------------------- Page: 4 ----------------------
ISO/FDIS 8598-1:2012(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.
The main task of technical committees is to prepare International Standards. Draft International
Standards adopted by the technical committees are circulated to the member bodies for voting.
Publication as an International Standard requires approval by at least 75 % of the member bodies
casting a vote.
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights.
ISO 8598-1 was prepared by Technical Committee ISO/TC 172, Optics and photonics, Subcommittee SC 7,
Ophthalmics optics and instruments.
This first edition of ISO 8598-1 cancels and replaces ISO 8598:1996, as well as ISO 8598:1996/Cor.1:1998,
of which this constitutes a technical revision.
ISO 8598 consists of the following parts, under the general title Optics and optical instruments — Focimeters:
— Part 1: General purpose instruments
© ISO 2012 – All rights reserved v

---------------------- Page: 5 ----------------------
ISO/FDIS 8598-1:2012(E)
Introduction
General purpose focimeters are intended for measurement of contact lenses, single-vision, multifocal
and progressive-power or degressive-power spectacle lenses, both uncut and mounted in spectacle
frames, and for the orientation and marking of spectacle lenses.
vi © ISO 2012 – All rights reserved

---------------------- Page: 6 ----------------------
FINAL DRAFT INTERNATIONAL STANDARD ISO/FDIS 8598-1:2012(E)
Optics and optical instruments — Focimeters —
Part 1:
General purpose instruments
1 Scope
This part of ISO 8598 specifies requirements and test methods for general purpose focimeters designed
for the measurement of vertex powers, cylinder axis, prismatic power and prism base setting within a
restricted area at a specified location of a lens.
It is applicable to instruments typically intended for use by the ophthalmic community, with the
capability to demonstrate conformity of lens products with the International Standards existing for
these lenses.
NOTE This excludes instruments that can only measure the whole lens at once.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.
ISO 7944, Optics and optical instruments — Reference wavelengths
1)
ISO 9342-1 , Optics and optical instruments — Test lenses for calibration of focimeters — Part 1: Test
lenses for focimeters used for measuring spectacle lenses
ISO 9342-2, Optics and optical instruments — Test lenses for calibration of focimeters — Part 2: Test lenses
for focimeters used for measuring contact lenses
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
3.1
adjusting rail
movable rail or bar used as the reference axis for spectacles during measurement and which is aligned
perpendicularly to the optical axis of the focimeter and parallel to the axis direction 0° to 180°
NOTE This is also called the lens table or frame rest.
3.2
capability
ability of a system or process to determine whether a product meets the requirements for its intended use
3.3
general purpose focimeter
instrument that is used to measure vertex powers and prismatic effects of spectacle and contact lenses,
to orientate and mark uncut lenses, and to verify the correct mounting of lenses in spectacle frames
1) ISO 9342-1 and ISO 9342-2 will, on the occasion of their next revision, be combined into one single standard
and published as Part 2 to ISO 8598, i.e. ISO 8598-2.
© ISO 2012 – All rights reserved 1

---------------------- Page: 7 ----------------------
ISO/FDIS 8598-1:2012(E)
3.3.1
manual focusing focimeter
instrument that allows the operator to view images formed by rays of light passing through a lens and,
by manually focusing and adjusting, to measure the vertex power and identify the principal meridians
NOTE 1 For lenses with cylinder power, the cylinder axis is found using the method provided to locate the principal
meridians of the lens in the area defined by the focimeter aperture. Prismatic power is measured separately for
this type of focimeter.
NOTE 2 There are sub-classes of manual focusing focimeter. One type has an eyepiece whereas the other has a
projection screen. With the eyepiece type, the measurement target is focused and viewed through an eyepiece.
3.3.2
automated focimeter
instrument that measures the vertex power of a lens, in the area defined by the focimeter aperture, in a
single measurement without operator adjustment
3.3.3
continuously indicating focimeter
focimeter with a continuous scale
NOTE For the purposes of this part of ISO 8598, this includes both automated instruments when set to 0,06 D or
0,01 D steps and conventional manual focusing instruments.
3.3.4
digitally rounding focimeter
focimeter which displays measured values rounded to the nearest 0,25 D or 0,12(5) D incremental value
3.4
centration error of the instrument
residual prismatic error of the instrument with no lens in place
3.5
indication
〈of a focimeter〈 quantitative value provided as the output of the focimeter
3.6
indication error
difference between the value indicated by the focimeter and the true value of the reference lens
NOTE 1 Here the true value of the reference lens, the back vertex power, is calculated using the four known basic
lens parameters: radii of curvature of the front and back surface (r and r ), the central thickness (t), and the
1 2
refractive index (n) of optical glass, using the formulas listed in ISO 9342-1 and ISO 9342-2.
NOTE 2 When using a measuring instrument, the influence of the uncertainty and indication error of the device
should be considered.
3.7
lens support
mechanical interface of the instrument against which the spectacle lens or the contact lens is placed
for measurement
NOTE The focimeter measures the vertex power related to the surface placed against the lens support.
3.8
near portion power
vertex power measured at the near reference point, as specified by the manufacturer, of a multifocal,
progressive-power or degressive-power lens
2 © ISO 2012 – All rights reserved

---------------------- Page: 8 ----------------------
ISO/FDIS 8598-1:2012(E)
3.9
non-symmetric error for cylindrical power and cylinder axis
residual error in the indicated cylindrical power and/or the indicated cylinder axis of a spherocylindrical
lens for an automated focimeter after calibration
3.10
non-symmetric prism error of a focimeter
difference in the prismatic power readings when a plano-prism is measured first with its base setting in one
direction and then in the opposite direction, for example, base settings of 180° and 360°, or 90° and 270°
3.11 Vertex power
3.11.1
back vertex power
reciprocal of the paraxial value of the back vertex focal length measured in metres
3.11.2
front vertex power
reciprocal of the paraxial value of the front vertex focal length measured in metres
NOTE 1 Conventionally, the back vertex power, in dioptres, is specified as the “power” of a spectacle lens or a
contact lens, although the front vertex power is required for certain purposes, for example in the measurement of
some multifocal lenses or progressive-power lenses.
−1
NOTE 2 The unit for expressing vertex power is the reciprocal metre (m ). The name for this unit is the “dioptre”,
for which the symbol is “D”.
4 Technical requirements for general purpose focimeters
4.1 The measuring range shall include vertex powers with a range from at least −20 D to +20 D and
prismatic powers from 0 Δ to at least 5 Δ.
The instrument shall be capable of measuring the axis direction (see ISO 8429) of cylindrical lenses
between 0° and 180°. For prisms, it shall be possible to determine the direction of the base setting
between 0° and 360°.
4.2 For manual focusing focimeters with non-digital display, the dioptric power scale shall have an
interval not greater than 0,25 D and shall be clear enough for interpolations to be made to the nearest
0,12 D or less.
For axis directions (see ISO 8429) the scale interval shall not exceed 5° and shall be clear enough for
interpolations to be made to the nearest degree.
For prismatic power readings, the scale interval shall not exceed 1 Δ and shall be clear enough for
interpolations to be made to the nearest 0,50 Δ.
4.3 For focimeters with digital display in the range from +20 D to −20 D, the minimum step of the
display shall be equal to or lower than 0,06 D. The display shall show at least two decimal digits.
For axis directions, the increment of the digital display shall be 1°.
For prismatic power readings, the minimum step of the digital display shall be equal to or lower than 0,06 Δ.
4.4 The instrument designed to measure spectacle lenses shall be able to measure lenses with a
diameter of at least 80 mm and a thickness of at least 20 mm. Translational movements of the lenses
on the lens support of not less than 30 mm in a direction perpendicular to the optical axis and to the
adjusting rail shall be possible, starting from no more than 10 mm below, in front of, or behind, the optical
axis of the instrument, as applicable. The adjusting rail shall also be capable of movement of not less than
30 mm in a direction perpendicular to its length and the optical axis of the instrument. See Figure 1.
© ISO 2012 – All rights reserved 3

---------------------- Page: 9 ----------------------
ISO/FDIS 8598-1:2012(E)
Dimensions in millimetres
Key
1 lens support
2 adjusting rail
Figure 1 — Minimum required movement of the adjusting rail
4.5 The lens support should be designed so that it does not cause any damage to the lens under test
when used in the manner recommended by the manufacturer.
It shall be designed and manufactured so that it will support a lens placed against it, holding the surface
that is in contact with it perpendicular to the optical axis of the focimeter.
The lens support shall not adversely affect the accuracy of measurement by introducing sagittal error.
Examples of suitable lens supports are shown in Figure 2.
Key
d , d internal diameter of support
1 c
d d external diameter of support
2, e
d = (d + d )/2
3 1 2
h , h height of support
s c
r = (d - d )/4
1 2 1
r = (d – d )/4
2 e c
Figure 2 — Examples of lens supports for spectacle lenses (a) and contact lenses (b)
4 © ISO 2012 – All rights reserved

---------------------- Page: 10 ----------------------
ISO/FDIS 8598-1:2012(E)
For spectacle lenses (a), d should be in the range 5 mm to 8 mm while for contact lenses (b), d should be
1 c
4,5 mm ± 0,5 mm. Because of the increased sagittal depth of steeply-curved contact lenses, the contact
lens support is usually smaller in diameter and slightly shorter. The height difference (h – h ) should be
s c
0,55 mm ± 0,02 mm.
NOTE The dimensions of the contact lens support are as specified in ISO 18369-3.
The internal diameter (d , d ) of the lens support for focimeters used for spectacle lens measurement or
1 c
for contact lens measurement shall be stated by the manufacturer.
4.6 The pin mark printed by the axis marker shall be small enough to enable the distance between a
first and a second mark to be distinguished.
NOTE A diameter of 0,7 mm is recommended.
4.7 The instrument shall be designed so that it gives stable measured values in a normal environment
(i.e. a temperature of 23 °C ± 5 °C, and a relative humidity of less than 85 %).
5 Metrological requirements
5.1 General
When tested with the reference lenses specified in ISO 9342-1 and/or ISO 9342-2, both manual focusing
and automated focimeters shall give indication readings for vertex power and prismatic power over
their entire claimed measuring range, and the indication error for the respective true value shall not
exceed the limits given in Tables 1 and 2.
5.2 Reference wavelength
The instrument shall be manufactured to indicate dioptric powers with reference to the green mercury
line λ = 546,07 nm or the yellow helium line λ = 587,56 nm, as given in ISO 7944 and the manufacturer
e d
shall indicate explicitly which wavelength has been selected.
NOTE If the light source used in the focimeter is not centred on the green mercury line λ = 546,07 nm or the
e
yellow helium line λ = 587,56 nm, the user should be able either to set the instrument, or to apply corrections to
d
the indicated values, for lenses made from materials other than the one that was used to calibrate the instrument.
5.3 Performance requirement
5.3.1 Indication error
When tested in accordance with 6.2.1, the instrument, when fitted with the spectacle lens support, shall
not exceed the permissible indication errors given in Tables 1 and 2 when used with the spectacle form
reference lenses specified in ISO 9342-1.
The instrument, if intended by the manufacturer solely for contact lens measurement, shall meet
the requirements for permissible indication errors given in Table 1 when used with a support of the
diameter specified in 4.5 by Figure 2 b) and, preferably, the contact lens form reference lenses specified
in ISO 9342-2.
NOTE 1 The requirement mentioned above is for checking the indication error of focimeters when used for
contact lens measurement, but not for the permissible error of commercial contact lenses.
NOTE 2 If the instrument is used for contact lens measurement but calibrated by the spectacle form reference
lenses given in ISO 9342-1, this should be stated and the user should be aware that the calibration results may be
significantly different from those obtained with the contact lens form reference lenses given in ISO 9342-2.
© ISO 2012 – All rights reserved 5

---------------------- Page: 11 ----------------------
ISO/FDIS 8598-1:2012(E)
The maximum permissible indication errors given in Tables 1 and 2 correspond to the use of the reference
lenses specified in ISO 9342-1 and/or ISO 9342-2 with the respective true values.
Table 1 — Maximum permissible indication error on measured vertex power for general
purpose instruments
Values in dioptres (D)
Maximum permissible indication error
Instruments with con- Digitally rounding instruments when set to
Measuring range of vertex power
tinuous scale increments of
a
0,25 0,12(5)
< 0 ≥ −5 > 0 ≤ +5 ±0,06 0,0 0,0
< −5 ≥ −10 > +5 ≤ +10 ±0,09 0,0 ±0,12(5)
< −10 ≥ −15 > +10 ≤ +15 ±0,12 0,0 ±0,12(5)
< −15 ≥ −20 > +15 ≤ +20 ±0,18 ±0,25 ±0,12(5)
< −20 > +20 ±0,25 ±0,25 ±0,25
NOTE 1 Calibrating to tighter tolerances (e.g. 10 % of the product tolerance) will improve instrument equivalence and
reduce possible differences between two instruments.
NOTE 2 The user should be aware that when a focimeter is set to read in steps of 0,01 D, the readings may not be true to this
level of precision.
a
The meaning of 0,12(5) is that the instruments has been set to 1/8th dioptre steps, displayed as 0,12 D.
Table 2 — Maximum permissible indication errors on measured prismatic power for general
purpose instruments
Values in prism dioptres (Δ)
Maximum permissible indication error
Measuring range of prismatic Instruments with con- Digitally rounding instruments when set to incre-
power tinuous scale ments of
a
0,25 0,12(5)
> 0 ≤ 5 0,25 0 0,12(5)
> 5 ≤ 10 0,25 0,25 0,25
> 10 ≤ 15 0,50 0,50 0,50
> 15 ≤ 20 0,75 0,75 0,75
> 20 1,00 1,00 1,00
NOTE Calibrating to tighter tolerances (e. g. to 10 % of the product tolerance) will improve instrument equivalence and
reduce possible differences between two instruments.
a
The meaning of 0,12(5) is that the instruments has been set to 1/8th dioptre steps, displayed as 0,12 D.
If fitted with interchangeable supports, the instrument should be able to provide correct results for
spectacle form reference lenses when fitted with the spectacle lens support. When using a calibration
conversion table provided by the instrument manufacturer, contact lens form reference lenses measured
with the contact lens support should also provide correct results by using the correction value from
the conversion table. The correction value from the conversion table shall not exceed 0,25 D for the
measuring range −10,00 D to +10,00 D, and it shall not exceed 0,50 D for the measuring range over ± 10 D.
For automated focimeters, when any of the spherical reference lenses is measured after correct
centration, it shall indicate less than 0,06 D of astigmatic power.
5.3.2 Axis marker for the optical centre of lens
When tested in accordance with 6.3, the position of the axis marker for the optical centre of the lens
shall not deviate from the optical axis of the focimeter by more than 0,4 mm.
6 © ISO 2012 – All rights reserved

---------------------- Page: 12 ----------------------
ISO/FDIS 8598-1:2012(E)
5.3.3 Axis marker alignment
When tested in accordance with 6.4, any misalignment of the orientation of the axis marker shall not
exceed the tolerance of ±1° from the 0° to 180° directions of the dial scale or the 0° (or 180°) value of the
indicated axis for automated focimeters.
5.3.4 Adjusting rail
When tested in accordance with 6.5, any misalignment of the adjusting rail shall not exceed the
tolerance of 1° from the position parallel to the 0° to 180° directions of the dial scale for manual focusing
instruments or the 0° (or 180°) value of the indicated axis for automated focimeters.
5.3.5 Non-symmetric errors for automated focimeters
5.3.5.1 General
These cylindrical powers, cylinder axis and prism error requirements are for automated focimeters that
measure sphere, cylinder, cylinder axis and prism simultaneously.
5.3.5.2 Non-symmetric error for cylindrical power
When tested in accordance with 6.6.2, the non-symmetric error for the indicated power readings of the
spherocylindrical-power reference lens specified in Clause C.3 shall not exceed 0,06 D sph/0,09 D cyl.
5.3.5.3 Non-symmetric error for cylinder axis
When tested in accordance with 6.6.3, the non-symmetric error for the cylinder axis from each reading of
the spherocylindrical-power reference lens specified in Clause C.3 for each orientation shall not exceed ±1°.
5.3.5.4 Non-symmetric error for prismatic power
When tested in accordance with 6.6.4, the non-symmetric error for prismatic power for an automated
focimeter shall not exceed 0,06 Δ. At the same time, the indicated values for the spherical and cylindrical
powers shall not exceed ± 0,06 D.
5.3.6 Repeatability for the indication reading of automated focimeters
When tested in accordance with 6.7, the repeatability for indication readings for an automated focimeter
shall not exceed 0,06 D within the spherical power range −10 D to +10 D, and shall not exceed 0,12 D
within the measuring range over ±10 D.
5.3.7 Centration error for manual focusing focimeters
When tested in accordance with 6.8, the residual prismatic error of the focimeter with no lens in place
shall not exceed 0,1 Δ.
5.3.8 Capability of measuring tinted lenses
When tested in accordance with 6.9, focimeters should be capable of measuring tinted lenses with
luminous transmittance τ ≥ 18 %. The indicated value from measuring the reference lens and neutral
V
reference filter as specified in Clause C.4 shall remain within the maximum permissible indication
errors of Table 1. The limited capability of measuring tinted lenses shall be stated if the instrument
cannot meet the requirement of measuring tinted lenses with luminous transmittance τ ≥ 18 %.
V
NOTE In practice it may be difficult to measure lenses whose colour departs significantly from neutral if the
transmittance of the lens at the operating wavelength of the focimeter is less than 18 %.
© ISO 2012 – All rights reserved 7

---------------------- Page: 13 ----------------------
ISO/FDIS 8598-1:2012(E)
5.3.9 Astigmatic axis repeatability for low-powered cylindrical lenses with manual focimeters
When tested in accordance with 6.10, the standard deviation of the axis readings for a 0,25 D cylindrical
lens shall be less than 4°.
6 Test procedures
6.1 General
The reference lenses described in ISO 9342-1 and ISO 9342-2 shall be used to verify that the requirements
of 5.1 to 5.3 are met.
NOTE It is recommended that the reference lenses of ISO 9342-2 should be used if a focimeter is fitted
...

NORME ISO
INTERNATIONALE 8598-1
Première édition
2014-09-15
Optique et instruments d’optique —
Frontofocomètres —
Partie 1:
Instruments pour cas généraux
Optics and optical instruments — Focimeters —
Part 1: General purpose instruments
Numéro de référence
ISO 8598-1:2014(F)
©
ISO 2014

---------------------- Page: 1 ----------------------
ISO 8598-1:2014(F)

DOCUMENT PROTÉGÉ PAR COPYRIGHT
© ISO 2014
Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite ni utilisée
sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie, l’affichage sur
l’internet ou sur un Intranet, sans autorisation écrite préalable. Les demandes d’autorisation peuvent être adressées à l’ISO à
l’adresse ci-après ou au comité membre de l’ISO dans le pays du demandeur.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Publié en Suisse
ii © ISO 2014 – Tous droits réservés

---------------------- Page: 2 ----------------------
ISO 8598-1:2014(F)

Sommaire Page
Avant-propos .iv
Introduction .v
1 Domaine d’application . 1
2 Références normatives . 1
3 Termes et définitions . 1
4 Exigences techniques pour les frontofocomètres pour cas généraux .3
5 Exigences métrologiques . 5
5.1 Généralités . 5
5.2 Longueur d’onde de référence . . 5
5.3 Exigence de performance . 6
6 Modes opératoires . 8
6.1 Généralités . 8
6.2 Vérification des erreurs d’indication . 8
6.3 Vérification du marqueur d’axe pour le centre optique du verre . 9
6.4 Vérification de l’alignement du marqueur d’axe . 9
6.5 Vérification du rail de réglage . 9
6.6 Vérification de l’erreur non symétrique pour les frontofocomètres automatiques . 9
6.7 Vérification de la répétabilité des mesures de puissance frontale pour
frontofocomètres automatiques .10
6.8 Vérification de l’erreur de centrage .10
6.9 Vérification de la capacité des frontofocomètres à mesurer des verres teintés .10
6.10 Vérification de la répétabilité de l’axe astigmatique pour les verres cylindriques à
faible puissance .11
6.11 Modes opératoires spéciaux pour les frontofocomètres à oculaire .11
6.12 Critère de mise au point d’image avec un frontofocomètre manuel .11
7 Marquage .12
7.1 Référence à l’ISO 8598-1 .12
7.2 Informations générales que le fabricant doit fournir .12
7.3 Informations supplémentaires que le fabricant doit fournir .12
Annexe A (informative) Utilisation de valeurs de correction lors du mesurage de verres
de lunettes .13
Annexe B (informative) Exemple d’évaluation de l’incertitude de mesure des frontofocomètres
automatiques pour cas généraux .18
Annexe C (normative) Spécifications pour les verres de référence spéciaux .24
Annexe D (informative) Informations destinées aux utilisateurs relatives aux performances des
frontofocomètres pour cas généraux couverts par la présente partie de l’ISO 8598 .26
Bibliographie
.27
© ISO 2014 – Tous droits réservés iii

---------------------- Page: 3 ----------------------
ISO 8598-1:2014(F)

Avant-propos
L’ISO (Organisation internationale de normalisation) est une fédération mondiale d’organismes
nationaux de normalisation (comités membres de l’ISO). L’élaboration des Normes internationales est
en général confiée aux comités techniques de l’ISO. Chaque comité membre intéressé par une étude
a le droit de faire partie du comité technique créé à cet effet. Les organisations internationales,
gouvernementales et non gouvernementales, en liaison avec l’ISO participent également aux travaux.
L’ISO collabore étroitement avec la Commission électrotechnique internationale (CEI) en ce qui concerne
la normalisation électrotechnique.
Les procédures utilisées pour élaborer le présent document et celles destinées à sa mise à jour sont
décrites dans les Directives ISO/CEI, Partie 1. Il convient, en particulier de prendre note des différents
critères d’approbation requis pour les différents types de documents ISO. Le présent document a été
rédigé conformément aux règles de rédaction données dans les Directives ISO/CEI, Partie 2 (voir www.
iso.org/directives).
L’attention est appelée sur le fait que certains des éléments du présent document peuvent faire l’objet de
droits de propriété intellectuelle ou de droits analogues. L’ISO ne saurait être tenue pour responsable
de ne pas avoir identifié de tels droits de propriété et averti de leur existence. Les détails concernant les
références aux droits de propriété intellectuelle ou autres droits analogues identifiés lors de l’élaboration
du document sont indiqués dans l’Introduction et/ou sur la liste ISO des déclarations de brevets reçues
(voir www.iso.org/patents).
Les éventuelles appellations commerciales utilisées dans le présent document sont données pour
information à l’intention des utilisateurs et ne constituent pas une approbation ou une recommandation.
Pour une explication de la signification des termes et expressions spécifiques de l’ISO liés à l’évaluation de
la conformité, aussi bien que pour des informations au-sujet de l’adhésion de l’ISO aux principes de l’OMC
concernant les obstacles techniques au commerce (OTC) voir le lien suivant: Foreword - Supplementary
information
Le comité chargé de l’élaboration du présent document est l’ISO 172, Optique et photonique, sous-comité
SC 7, Optique et instruments ophtalmiques.
La première édition de l’ISO 8598-1 annule et remplace l’ISO 8598:1996 qui ont fait l’objet d’une révision
technique. Elle incorpore également l’ISO 8598:1996/Cor.1:1998.
L’ISO 8598 comprend les parties suivantes, présentées sous le titre général Optique et instruments
d’optique — Frontofocomètres:
— Partie 1: Instruments pour cas généraux
iv © ISO 2014 – Tous droits réservés

---------------------- Page: 4 ----------------------
ISO 8598-1:2014(F)

Introduction
Les frontofocomètres pour cas généraux sont conçus pour le mesurage des verres de lunettes unifocaux,
multifocaux et progressifs ou dégressifs, non détourés ou sur montures, ainsi que pour l’orientation et
le marquage des verres de lunettes.
© ISO 2014 – Tous droits réservés v

---------------------- Page: 5 ----------------------
NORME INTERNATIONALE ISO 8598-1:2014(F)
Optique et instruments d’optique — Frontofocomètres —
Partie 1:
Instruments pour cas généraux
1 Domaine d’application
La présente partie de l’ISO 8598 spécifie les exigences et méthodes d’essai des frontofocomètres pour
cas généraux destinés à mesurer les puissances frontales, l’axe de cylindre, la puissance prismatique
et les réglages de la base du prisme sur une surface définie d’un point spécifique d’un verre. Elle exclut
donc tout instrument qui ne peut mesurer que la totalité du verre à la fois.
Elle s’applique aux instruments typiquement destinés à la communauté ophtalmologique, dans le but
de démontrer la conformité des verres de lunettes avec les Normes internationales en vigueur pour ces
produits.
2 Références normatives
Les documents suivants, en totalité ou en partie, sont référencés de manière normative dans le présent
document et sont indispensables pour son application. Pour les références datées, seule l’édition citée
s’applique. Pour les références non datées, la dernière édition du document de référence s’applique (y
compris les éventuels amendements).
ISO 7944, Optique et instruments d’optique — Longueurs d’onde de référence
ISO 8429, Optique et instruments d’optique — Ophtalmologie — Échelle graduée
ISO 9342-1, Optique et instruments d’optique — Verres étalons pour l’étalonnage des frontofocomètres —
Partie 1: Verres étalons pour frontofocomètres pour le mesurage des verres de lunettes
3 Termes et définitions
Pour les besoins du présent document, les termes et définitions suivants s’appliquent.
3.1
rail de réglage
rail ou barre mobile, perpendiculaire à l’axe optique du frontofocomètre et parallèle à l’axe 0° à 180°, et
servant d’axe de référence pour les lunettes pendant le mesurage
Note 1 à l’article: Il peut également être appelé platine verre ou appui de monture.
3.2
capacité
aptitude d’un système ou d’un processus pour son utilisation prévue
3.3
frontofocomètre pour cas généraux
instrument servant à mesurer les puissances frontales, l’axe du cylindre et les effets prismatiques des
verres de lunettes, à orienter et à marquer les verres non détourés et à vérifier le montage correct des
verres sur des montures de lunettes
© ISO 2014 – Tous droits réservés 1

---------------------- Page: 6 ----------------------
ISO 8598-1:2014(F)

3.3.1
frontofocomètre à mise au point manuelle
instrument qui permet à l’opérateur de visualiser les images formées par les rayons de lumière traversant
un verre afin de mesurer la puissance frontale et d’identifier les principaux méridiens par un réglage et
une mise au point manuels
Note 1 à l’article: Pour les verres à puissance cylindrique, l’axe du cylindre est déterminé à l’aide de la méthode
servant à localiser les méridiens principaux du verre dans la région définie par l’ouverture du frontofocomètre.
Avec ce type de frontofocomètre, la puissance prismatique est mesurée séparément.
Note 2 à l’article: Il existe des sous-classes de frontofocomètres à mise au point manuelle. L’une possède un
oculaire, tandis que l’autre possède un écran de projection. Avec le type à oculaire, la cible de mesure est vue et
mise au point au travers d’un oculaire.
3.3.2
frontofocomètre automatique
instrument qui mesure la puissance frontale d’un verre, dans la zone définie par l’ouverture du
frontofocomètre, en une seule mesure, sans réglage de l’opérateur
3.3.3
frontofocomètre avec indication en continu
frontofocomètre avec échelle de mesure continue
Note 1 à l’article: Pour les besoins de la présente partie de l’ISO 8598, elle inclut à la fois les instruments
automatiques lorsqu’ils sont réglés par graduations de 0,06 D à 0,01 D et les instruments à mise au point manuelle
conventionnels.
3.3.4
frontofocomètre à mesure digitale arrondie
frontofocomètre qui présente des valeurs de mesures arrondies au plus près par incrément de 0,25 D ou
0,12(5) D
3.4
erreur de centrage de l’instrument
erreur prismatique résiduelle de l’instrument sans verre
3.5
indication
〈frontofocomètre〉 valeur quantitative fournie comme sortie du frontofocomètre
3.6
erreur d’indication
différence entre la valeur indiquée par le frontofocomètre et la valeur réelle du verre de référence
Note 1 à l’article: Ici, la valeur réelle du verre de référence, la puissance frontale arrière, est calculée à partir des
quatre paramètres de base connus du verre: rayon de courbure des surfaces avant et arrière (r et r ), épaisseur
1 2
centrale (t) et indice de réfraction (n) du matériau du verre de référence, et au moyen des formules données dans
l’ISO 9342-1.
Note 2 à l’article: Lors de l’utilisation d’un instrument de mesure, il convient que l’influence de l’incertitude et de
l’erreur d’indication de l’appareil soit prise en compte.
3.7
support de verre
interface mécanique de l’instrument sur laquelle est placé le verre à mesurer
Note 1 à l’article: Le frontofocomètre mesure la puissance frontale par rapport à la surface placée contre le support
de verre.
2 © ISO 2014 – Tous droits réservés

---------------------- Page: 7 ----------------------
ISO 8598-1:2014(F)

3.8
puissance de vision de près
puissance frontale mesurée au point de référence de la vision de près indiqué par le fabricant pour un
verre multifocal, progressif ou dégressif
3.9
erreur non symétrique pour la puissance cylindrique et l’axe du cylindre
erreur résiduelle dans la puissance cylindrique indiquée et/ou l’axe du cylindre indiqué d’un verre
sphéro-cylindrique pour un frontofocomètre automatique après étalonnage
3.10
erreur prismatique asymétrique d’un frontofocomètre
différence entre les indications de puissance prismatique données par les mesurages d’un prisme plan
avec sa base dans une direction et dans la direction opposée, par exemple pour des réglages de base de
180° et 360° ou 90° et 270°
3.11.1
puissance frontale arrière
inverse de la distance focale paraxiale par rapport à la face arrière mesurée en mètres
3.11.2
puissance frontale avant
inverse de la valeur paraxiale de la distance frontale avant, mesurée en mètres
Note 1 à l’article: Par convention, la puissance frontale arrière, exprimée en dioptries, est indiquée comme la
«puissance» d’un verre. La puissance frontale avant est cependant nécessaire dans certains cas, par exemple pour
le mesurage de certains verres multifocaux ou progressifs.
−1
Note 2 à l’article: L’unité servant à exprimer la puissance frontale est le mètre à la puissance moins un (m ). Le
nom de cette unité est la «dioptrie» et son symbole est «D».
4 Exigences techniques pour les frontofocomètres pour cas généraux
4.1 La plage de mesure doit au minimum comprendre des puissances frontales de −20 D à +20 D et des
puissances prismatiques de 0 Δ à au moins 5 Δ.
L’instrument doit être capable de mesurer la direction de l’axe (voir l’ISO 8429) de verres cylindriques
entre 0° et 180°. Pour les prismes, il doit être possible de déterminer la direction de la base entre 0° et
360°.
4.2 Pour les frontofocomètres à mise au point manuelle et affichage non numérique, les intervalles
entre graduations de l’échelle de puissance dioptrique ne doivent pas dépasser 0,25 D et doivent être
suffisamment clairs pour faire des interpolations à 0,12 D ou moins le plus proche.
Pour les directions d’axe (voir l’ISO 8429), l’échelon ne doit pas dépasser 5° et doit être suffisamment
clair pour permettre des interpolations au degré le plus proche.
Pour les lectures de puissance prismatique, l’échelon ne doit pas dépasser 1 Δ et doit être suffisamment
clair pour permettre des interpolations à 0,50 Δ le plus proche.
4.3 Pour les frontofocomètres à affichage numérique dans la plage de +20 D à −20 D, l’incrément
minimal de l’écran ne doit pas être supérieur à 0,06 D. L’affichage doit donner au moins deux décimales.
Pour les directions d’axe, l’incrément de l’affichage numérique doit être de 1°.
Pour les lectures de puissance prismatique, l’incrément minimal de l’affichage numérique ne doit pas
dépasser 0,06 Δ.
© ISO 2014 – Tous droits réservés 3

---------------------- Page: 8 ----------------------
ISO 8598-1:2014(F)

4.4 L’instrument conçu pour mesurer des verres de lunettes doit être capable de mesurer des verres
d’un diamètre d’au moins 80 mm et d’une épaisseur d’au moins 20 mm. Sur le support de verre, les verres
doivent pouvoir effectuer des mouvements de translation d’au moins 30 mm perpendiculairement à l’axe
optique et au rail de réglage, en démarrant à 10 mm ou moins sous ou devant ou derrière l’axe optique
de l’instrument, selon le cas. Le rail de réglage doit également pouvoir se déplacer de plus de 30 mm dans
une direction perpendiculaire à sa longueur et à l’axe optique de l’instrument. Voir Figure 1.
Dimensions en millimètres
Légende
1 support de verre
2 rail de réglage
Figure 1 — Mouvement minimal requis du rail de réglage
4.5 Il convient que le support de verre soit conçu de sorte que, lorsqu’il est utilisé suivant les instructions
du fabricant, il ne cause aucun dommage au verre soumis à essai.
Il doit être conçu et construit pour supporter un verre placé contre lui, en tenant la surface de contact
perpendiculaire à l’axe optique du frontofocomètre.
Le support de verre ne doit pas altérer la précision de mesurage par l’introduction d’une erreur sagittale.
La Figure 2 présente des exemples de supports de verre adaptés.
a) b)
4 © ISO 2014 – Tous droits réservés

---------------------- Page: 9 ----------------------
ISO 8598-1:2014(F)

Légende
d , d diamètre interne du support
1 c
d , d diamètre externe du support
2 e
d = (d + d )/2
3 1 2
h , h hauteur du support
s c
r = (d – d )/4
1 2 1
r = (d – d )/4
2 e c
Figure 2 — Exemples de supports de verre de verres de lunettes (a) et de lentilles de contact (b)
Il convient que d soit compris entre 5 mm et 8 mm pour les verres de lunettes (a), alors que pour
1
les lentilles de contact (b), il convient que d soit de 4,5 mm ± 0,5 mm. À cause de la profondeur
c
sagittale augmentée des lentilles de contact fortement incurvées, le support de lentille de contact est
habituellement plus petit en diamètre et légèrement plus fin. Il convient que la différence de hauteur
(h – h ) soit de 0,55 mm ± 0,02 mm.
s c
NOTE Les dimensions du support de lentilles de contact sont telles que spécifiées dans l’ISO 18369-3.
Le diamètre interne (d , d ) du support de verre des frontofocomètres utilisé pour le mesurage du verre
1 c
de lunettes ou pour le mesurage de lentille de contact doit être défini par le fabricant.
4.6 La marque réalisée par le marqueur d’axe doit être assez petite pour distinguer l’écart entre deux
marques.
NOTE Le diamètre recommandé est de 7 mm.
4.7 L’instrument doit être conçu pour donner des valeurs mesurées stables dans un environnement
normal (c’est-à-dire à une température de 23 °C ± 5 °C et une humidité relative inférieure à 85 %).
5 Exigences métrologiques
5.1 Généralités
Lorsqu’ils sont utilisés avec les verres de référence spécifiés dans l’ISO 9342-1, les frontofocomètres
à mise au point manuelle comme ceux à mise au point automatique doivent donner des indications
de puissances frontale et prismatique sur la totalité de leur plage de mesure revendiquée. L’erreur
d’indication pour la valeur réelle en question ne doit, en outre, pas franchir les limites énoncées dans les
Tableaux 1 et 2.
5.2 Longueur d’onde de référence
L’instrument doit être construit pour indiquer des puissances dioptriques par rapport à la raie de
mercure verte, λ = 546,07 nm, ou la raie d’hélium jaune, λ = 587,56 nm, spécifiées dans l’ISO 7944 et le
e d
fabricant doit, par ailleurs, signaler explicitement quelle longueur d’onde a été sélectionnée.
NOTE Si la source lumineuse du frontofocomètre n’est pas centrée sur la raie de mercure verte, λ = 546,07 nm,
e
ou sur la raie d’hélium jaune, λ = 587,56 nm, il convient que l’utilisateur soit en mesure de configurer l’instrument
d
ou d’appliquer aux valeurs indiquées des corrections pour les verres fabriqués avec des matériaux autres que
celui utilisé pour étalonner l’instrument.
© ISO 2014 – Tous droits réservés 5

---------------------- Page: 10 ----------------------
ISO 8598-1:2014(F)

5.3 Exigence de performance
5.3.1 Erreur d’indication
Lorsqu’il est pourvu du support de verre de lunette et utilisé avec les verres de référence de l’ISO 9342-1
et lorsqu’il est soumis à essai conformément à 6.2.1, l’instrument ne doit pas dépasser les tolérances
d’erreurs d’indication autorisées données dans les Tableaux 1 et 2.
Les erreurs d’indication maximales autorisées indiquées dans les Tableaux 1 et 2 correspondent à
l’utilisation des verres de référence de l’ISO 9342-1 dotés des valeurs réelles correspondantes.
Tableau 1 — Erreur d’indication maximale tolérée sur la puissance frontale mesurée par les
instruments pour cas généraux
Valeurs en dioptries (D)
Erreur d’indication maximale autorisée
Instruments à échelle Instruments numériques paramé-
Plage de mesure de puissance frontale
continue trés sur des incréments de
a
0,25 0,12(5)
< 0 ≥ −5 > 0 ≤ +5 ±0,06 0,0 0,0
< −5 ≥ −10 > +5 ≤ +10 ±0,09 0,0 ±0,12(5)
< −10 ≥ −15 > +10 ≤ +15 ±0,12 0,0 ±0,12(5)
< −15 ≥ −20 > +15 ≤ +20 ±0,18 ±0,25 ±0,12(5)
< −20 > +20 ±0,25 ±0,25 ±0,25
NOTE 1 L’étalonnage à des tolérances plus faibles (par exemple 10 % de la tolérance du produit) améliore l’équivalence de
l’instrument et réduit les éventuelles différences entre deux instruments.
NOTE 2 Il convient que l’utilisateur soit conscient que lorsqu’un frontofocomètre est paramétré pour lire par incréments de
0,01 D, les mesures peuvent ne pas être réelles à ce niveau de précision.
a e
0,12(5) signifie que les instruments ont été paramétrés sur des incréments de 1/8 dioptrie, affichés sous forme de
0,12 D.
Tableau 2 — Erreurs d’indication maximales tolérées sur la puissance prismatique mesurée
par les instruments pour cas généraux
Valeurs en dioptries prismatiques (Δ)
Erreur d’indication maximale autorisée
Plage de mesure de la puissance pris- Instruments à échelle Instruments numériques paramé-
matique continue trés sur des incréments de
a
0,25 0,12(5)
> 0 ≤ 5 0,25 0 0,12(5)
> 5 ≤ 10 0,25 0,25 0,25
> 10 ≤ 15 0,50 0,50 0,50
> 15 ≤ 20 0,75 0,75 0,75
> 20 1,00 1,00 1,00
NOTE L’étalonnage à des tolérances plus faibles (par exemple 10 % de la tolérance du produit) améliore l’équivalence de
l’instrument et réduit les éventuelles différences entre deux instruments.
a e
0,12(5) signifie que les instruments ont été paramétrés sur des incréments de 1/8 dioptrie, affichés sous forme de
0,12 D.
Pour les frontofocomètres automatiques, lorsqu’un des verres de référence sphériques est mesuré après
un centrage correct, il doit présenter une puissance astigmatique inférieure à 0,06 D.
6 © ISO 2014 – Tous droits réservés

---------------------- Page: 11 ----------------------
ISO 8598-1:2014(F)

5.3.2 Marqueur d’axe pour le centre optique du verre
Dans les conditions d’essai de 6.3, la position du marqueur d’axe du centre optique du verre ne doit pas
s’écarter de l’axe optique du frontofocomètre de plus de 0,4 mm.
5.3.3 Alignement du marqueur d’axe
Lors d’un essai conforme à 6.4, le désalignement de l’orientation du marqueur d’axe ne doit pas
dépasser ± 1° par rapport aux directions de 0° à 180° de l’échelle numérique pour les instruments à
mise au point manuelle ou par rapport à la valeur d’axe 0° (ou 180°) indiquée par les frontofocomètres
automatiques.
5.3.4 Rail de réglage
Lors d’un essai conformément à 6.5, le désalignement du rail de réglage ne doit pas dépasser 1° par
rapport à la parallèle aux directions de 0° à 180° de l’échelle de cadran angulaire pour les instruments à
mise au point manuelle ou par rapport à la valeur d’axe 0° (ou 180°) indiquée par les frontofocomètres
automatiques.
5.3.5 Capacité à mesurer des verres teintés
Dans le cadre d’un essai conforme à 6.9, Il convient que les frontofocomètres soient capables de mesurer
des verres teintés ayant une transmission lumineuse, τ ≥ 18 %. La valeur indiquée de mesure du verre
V
de référence et du filtre neutre de référence spécifiée en C.4 doit rester dans les erreurs d’indication
maximales tolérées du Tableau 1. La limite de capacité à mesurer des verres teintés doit être précisée
si l’instrument ne peut satisfaire aux exigences de mesurage des verres teintés dont le facteur de
transmission lumineuse, τ ≥ 18 %.
V
NOTE En pratique, il peut être difficile de mesurer des verres dont la couleur s’éloigne nettement du
neutre si le facteur de transmission du verre à la longueur d’onde de fonctionnement du frontofocomètre est
inférieur à 18 %.
5.3.6 Erreurs non symétriques des frontofocomètres automatiques
5.3.6.1 Généralités
Les exigences relatives à la puissance cylindrique, à l’axe de cylindre et à l’erreur de prisme concernent
les frontofocomètres automatiques mesurant simultanément sphère, cylindre, axe du cylindre et prisme.
5.3.6.2 Erreur non symétrique pour la puissance cylindrique
Dans le cadre d’un essai conforme à 6.6.2, l’erreur non symétrique pour les indications de puissance
d’un verre de référence de puissance sphéro-cylindrique spécifié en C.3 ne doit pas dépasser
0,06 D sph/ 0,09 D cyl.
5.3.6.3 Erreur non symétrique pour l’axe du cylindre
Dans le cadre d’un essai conforme à 6.6.3, l’erreur non symétrique de l’axe du cylindre de chaque lecture
du verre de référence de puissance sphéro-cylindrique spécifié en C.3 ne doit pas dépasser ±1° pour
chaque orientation.
5.3.6.4 Erreur non symétrique du prisme
Dans le cadre d’un essai conforme à 6.6.4, l’erreur de puissance prismatique non symétrique d’un
frontofocomètre automatique ne doit pas dépasser 0,06 Δ. Les valeurs indiquées en même temps pour
les puissances sphériques et cylindriques ne doivent pas dépasser ± 0,06 D.
© ISO 2014 – Tous droits réservés 7

---------------------- Page: 12 ----------------------
ISO 8598-1:2014(F)

5.3.7 Répétabilité des indications des frontofocomètres automatiques
Au cours d’un essai conforme à 6.7, la marge de répétabilité des indications d’un frontofocomètre
automatique ne doit pas dépasser 0,06 D sur la plage de puissance sphérique comprise entre −10 D et
+10 D et 0,12 D au-delà dans la plage de mesure au-delà de ± 10 D.
5.3.8 Erreur de centrage des frontofocomètres à mise au point manuelle
Au cours d’un essai conforme à 6.8, l’erreur prismatique résiduelle du frontofocomètre sans verre ne
doit pas dépasser 0,1 Δ.
5.3.9 Répétabilité de l’axe astigmatique des verres cylindriques à faible puissance avec les
frontofocomètres manuels
Au cours d’un essai conforme à 6.10, l’écart-type des indications d’axe pour un verre cylindrique de
0,25 D doit être inférieur à 4°.
6 Modes opératoires
6.1 Généralités
Les verres de référence décrits dans l’ISO 9342-1 doivent être utilisés pour vérifier que les exigences
de 5.1 à 5.3 sont satisfaites.
L’étalonnage initial du frontofocomètre et la vérification métrologique doivent être réalisés d’après tous
les verres de référence dans le jeu de verres de référence indiqués dans l’ISO 9342-1 qui font partie de la
plage de mesures de l’instrument.
Il convient qu’avant leur utilisation, les frontofocomètres à mise au point manuelle soient réglés
conformément à 6.11 et que la cible soit réglée conformément à 6.12.
6.2 Vérification des erreurs d’indication
6.2.1 Vérification des erreurs d’indication de puissance frontale
Afin de vérifier la conformité des erreurs
...

PROJET
NORME ISO/FDIS
FINAL
INTERNATIONALE 8598-1
ISO/TC 172/SC 7
Optique et instruments d’optique —
Secrétariat: DIN
Frontofocomètres —
Début de vote:
2012-09-13
Partie 1:
Vote clos le:
Instruments pour cas généraux
2012-11-13
Optics and optical instruments — Focimeters —
Part 1: General purpose instruments
LES DESTINATAIRES DU PRÉSENT PROJET SONT
INVITÉS À PRÉSENTER, AVEC LEURS OBSER-
Veuillez consulter les notes administratives en page iii
VATIONS, NOTIFICATION DES DROITS DE PRO-
PRIÉTÉ DONT ILS AURAIENT ÉVENTUELLEMENT
CONNAISSANCE ET À FOURNIR UNE DOCUMEN-
TATION EXPLICATIVE.
OUTRE LE FAIT D’ÊTRE EXAMINÉS POUR
ÉTABLIR S’ILS SONT ACCEPTABLES À DES FINS
INDUSTRIELLES, TECHNOLOGIQUES ET COM-
Numéro de référence
MERCIALES, AINSI QUE DU POINT DE VUE
ISO/FDIS 8598-1:2012(F)
DES UTILISATEURS, LES PROJETS DE NORMES
INTERNATIONALES DOIVENT PARFOIS ÊTRE
CONSIDÉRÉS DU POINT DE VUE DE LEUR POSSI-
BILITÉ DE DEVENIR DES NORMES POUVANT
SERVIR DE RÉFÉRENCE DANS LA RÉGLEMENTA-
©
TION NATIONALE. ISO 2012

---------------------- Page: 1 ----------------------
ISO/FDIS 8598-1:2012(F)
Notice de droit d’auteur
Ce document de l’ISO est un projet de Norme internationale qui est protégé par les droits d’auteur
de l’ISO. Sauf autorisé par les lois en matière de droits d’auteur du pays utilisateur, aucune partie de
ce projet ISO ne peut être reproduite, enregistrée dans un système d’extraction ou transmise sous
quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie,
les enregistrements ou autres, sans autorisation écrite préalable.
Les demandes d’autorisation de reproduction doivent être envoyées à l’ISO à l’adresse ci-après ou au
comité membre de l’ISO dans le pays du demandeur.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Toute reproduction est soumise au paiement de droits ou à un contrat de licence.
Les contrevenants pourront être poursuivis.
ii © ISO 2012 – Tous droits réservés

---------------------- Page: 2 ----------------------
ISO/FDIS 8598-1:2012(F)
TRAITEMENT PARALLÈLE ISO/CEN
Le présent projet final a été élaboré dans le cadre de l’Organisation internationale de normalisation (ISO) et
soumis selon le mode de collaboration sous la direction de l’ISO, tel que défini dans l’Accord de Vienne. Le
projet final a été établi sur la base des observations reçues lors de l’enquête parallèle sur le projet.
Le projet final est par conséquent soumis aux comités membres de l’ISO et aux comités membres du CEN en
parallèle à un vote d’approbation de deux mois au sein de l’ISO et à un vote formel au sein du CEN.
Les votes positifs ne doivent pas être accompagnés d’observations.
Les votes négatifs doivent être accompagnés des arguments techniques pertinents.
© ISO 2012 – Tous droits réservés iii

---------------------- Page: 3 ----------------------
ISO/FDIS 8598-1:2012(F)
Sommaire Page
Avant-propos .v
Introduction .vi
1 Domaine d’application . 1
2 Références normatives . 1
3 Termes et définitions . 1
4 Exigences techniques pour les frontofocomètres pour cas généraux .3
5 Exigences métrologiques . 5
5.1 Généralités . 5
5.2 Longueur d’onde de référence . . 5
5.3 Exigence de performance . 5
6 Modes opératoires . 8
6.1 Généralités . 8
6.2 Vérification des erreurs d’indication . 9
6.3 Vérification du marqueur d’axe pour le centre optique du verre . 9
6.4 Vérification de l’alignement du marqueur d’axe . 9
6.5 Vérification du rail de réglage .10
6.6 Vérification de l’erreur non symétrique pour les frontofocomètres automatiques .10
6.7 Vérification de la répétabilité de mesure de la puissance frontale pour
frontofocomètres automatiques .11
6.8 Vérification de l’erreur de centrage .11
6.9 Vérification de la capacité des frontofocomètres automatiques à mesurer des
verres teintés .11
6.10 Vérification de la répétabilité de l’axe astigmatique pour les verres cylindriques à
faible puissance .11
6.11 Modes opératoires spéciaux pour les frontofocomètres à oculaire .11
6.12 Critère de mise au point d’image avec un frontofocomètre manuel .12
7 Marquage .12
7.1 Référence à l’ISO 8598-1 .12
7.2 Informations générales que le fabricant doit fournir .12
7.3 informations supplémentaires que le fabricant doit fournir .12
Annexe A (informative) Utilisation de valeurs de correction lors du mesurage de verres de lunette
ou lentilles de contact .14
Annexe B (informative) Exemple d’évaluation de l’incertitude de mesure des frontofocomètres
automatiques pour cas généraux .19
Annexe C (normative) Spécifications pour les verres de référence spéciaux .25
Annexe D (informative) Informations destinées aux utilisateurs relatives aux performances des
frontofocomètres pour cas généraux couverts par la présente partie de l’ISO 8598 .27
Bibliographie .28
iv © ISO 2012 – Tous droits réservés

---------------------- Page: 4 ----------------------
ISO/FDIS 8598-1:2012(F)
Avant-propos
L’ISO (Organisation internationale de normalisation) est une fédération mondiale d’organismes
nationaux de normalisation (comités membres de l’ISO). L’élaboration des Normes internationales est
en général confiée aux comités techniques de l’ISO. Chaque comité membre intéressé par une étude
a le droit de faire partie du comité technique créé à cet effet. Les organisations internationales,
gouvernementales et non gouvernementales, en liaison avec l’ISO participent également aux travaux.
L’ISO collabore étroitement avec la Commission électrotechnique internationale (CEI) en ce qui concerne
la normalisation électrotechnique.
Les Normes internationales sont rédigées conformément aux règles données dans les Directives
ISO/CEI, Partie 2.
La tâche principale des comités techniques est d’élaborer les Normes internationales. Les projets de
Normes internationales adoptés par les comités techniques sont soumis aux comités membres pour vote.
Leur publication comme Normes internationales requiert l’approbation de 75 % au moins des comités
membres votants.
L’attention est appelée sur le fait que certains des éléments du présent document peuvent faire l’objet de
droits de propriété intellectuelle ou de droits analogues. L’ISO ne saurait être tenue pour responsable de
ne pas avoir identifié de tels droits de propriété et averti de leur existence.
L’ISO 8598-1 a été élaborée par le comité technique ISO/TC 172, Optique et instruments d’optique, sous-
comité SC 7, Optique et instruments ophtalmiques.
Cette première édition de l’ISO 8598-1 annule et remplace l’ISO 8598:1996, ainsi que
l’ISO 8598:1996/Cor.1:1998, qui a fait l’objet d’une révision technique.
L’ISO 8598 comprend les parties suivantes, présentées sous le titre général Optique et instruments
d’optique — Frontofocomètres:
— Partie 1: Instruments pour cas généraux
© ISO 2012 – Tous droits réservés v

---------------------- Page: 5 ----------------------
ISO/FDIS 8598-1:2012(F)
Introduction
Les frontofocomètres pour cas généraux sont conçus pour le mesurage des lentilles de contact, verres
de lunettes unifocaux, multifocaux et progressifs ou dégressifs, non détourés ou sur montures, ainsi que
pour l’orientation et le marquage des verres de lunettes.
vi © ISO 2012 – Tous droits réservés

---------------------- Page: 6 ----------------------
PROJET FINAL DE NORME INTERNATIONALE ISO/FDIS 8598-1:2012(F)
Optique et instruments d’optique — Frontofocomètres —
Partie 1:
Instruments pour cas généraux
1 Domaine d’application
La présente partie de l’ISO 8598 spécifie les exigences et les méthodes d’essai des frontofocomètres pour
cas généraux destinés à mesurer les puissances frontales, l’axe de cylindre, la puissance prismatique et
les réglages de la base du prisme sur une surface définie d’un point spécifique d’un verre.
Elle s’applique aux instruments typiquement destinés à la communauté ophtalmologique, dans le but de
démontrer la conformité des verres avec les Normes internationales en vigueur pour ces produits.
NOTE Elle exclut donc tout instrument qui ne peut mesurer que la totalité du verre à la fois.
2 Références normatives
Les documents de référence suivants sont indispensables pour l’application du présent document. Pour
les références datées, seule l’édition citée s’applique. Pour les références non datées, la dernière édition
du document de référence s’applique (y compris les éventuels amendements).
ISO 7944, Optique et instruments d’optique — Longueurs d’onde de référence
1)
ISO 9342-1 , Optique et instruments d’optique — Verres étalons pour l’étalonnage des frontofocomètres —
Partie 1: Verres étalons pour frontofocomètres pour le mesurage des verres de lunettes
ISO 9342-2, Optique et instruments d’optique — Verres étalons pour l’étalonnage des frontofocomètres —
Partie 2: Verres étalons pour frontofocomètres pour le mesurage des lentilles de contact
3 Termes et définitions
Pour les besoins du présent document, les termes et définitions suivants s’appliquent.
3.1
rail de réglage
rail ou barre mobile, perpendiculaire à l’axe optique du frontofocomètre et parallèle à l’axe 0° à 180°, et
servant d’axe de référence pour les lunettes pendant le mesurage
NOTE Il peut également être appelé platine verre ou appui de monture.
3.2
capacité
aptitude d’un système ou d’un processus à déterminer si un produit répond aux exigences pour son
utilisation prévue
3.3
frontofocomètre pour cas généraux
instrument servant à mesurer les puissances frontales et les effets prismatiques des verres correcteurs
et des lentilles de contact, à orienter et à marquer les verres non détourés et à vérifier le montage correct
des verres sur des montures de lunettes
1) Lors de leur prochaine révision, l’ISO 9342-1 et l’ISO 9342-2 vont être fusionnées pour ne former qu’une
seule norme, laquelle sera publiée comme étant la Partie 2 de l’ISO 8598, c’est-à-dire l’ISO 8598-2.
© ISO 2012 – Tous droits réservés 1

---------------------- Page: 7 ----------------------
ISO/FDIS 8598-1:2012(F)
3.3.1
frontofocomètre à mise au point manuelle
instrument qui permet à l’opérateur de visualiser les images formées par les rayons de lumière traversant
un verre afin de mesurer la puissance frontale et d’identifier les principaux méridiens par un réglage et
une mise au point manuels
NOTE 1 Pour les verres à puissance cylindrique, l’axe du cylindre est déterminé à l’aide de la méthode servant
à localiser les méridiens principaux du verre dans la région définie par l’ouverture du frontofocomètre. Avec ce
type de frontofocomètre, la puissance prismatique est mesurée séparément.
NOTE 2 Il existe des sous-classes de frontofocomètres à mise au point manuelle. L’une possède un oculaire, tandis
que l’autre possède un écran de projection. Avec le type à oculaire, la cible de mesure est vue et mise au point au
travers d’un oculaire.
3.3.2
frontofocomètre automatique
instrument qui mesure la puissance frontale d’un verre, dans la zone définie par l’ouverture du
frontofocomètre, en une seule mesurage, sans réglage de l’opérateur
3.3.3
frontofocomètre avec indication en continu
frontofocomètre avec échelle de mesure continue
NOTE Pour les besoins de la présente partie de l’ISO 8598, ceci inclut à la fois les instruments automatiques
lorsqu’ils sont réglés par graduations de 0,06 D à 0,01 D et les instruments à mise au point manuelle conventionnels.
3.3.4
frontofocomètre à mesure digitale arrondie
frontofocomètre qui présente des valeurs de mesures arrondies au plus près par incrément de
0,25 D ou 0,12(5) D
3.4
erreur de centrage de l’instrument
erreur prismatique résiduelle de l’instrument sans verre
3.5
indication
valeur quantitative fournie comme sortie du frontofocomètre
3.6
erreur d’indication
différence entre la valeur indiquée par le frontofocomètre et la valeur réelle du verre de référence
NOTE 1 Ici, la valeur réelle du verre de référence, la puissance frontale arrière, est calculée à partir des quatre
paramètres de base connus du verre: rayon de courbure des surfaces avant et arrière (r et r ), épaisseur centrale
1 2
(t) et indice de réfraction (n) du verre, et au moyen des formules données dans l’ISO 9342-1 et l’ISO 9342-2.
NOTE 2 Lors de l’utilisation d’un instrument de mesure, l’influence de l’incertitude et de l’erreur d’indication de
l’appareil doit être prise en compte.
3.7
support de verre
interface mécanique de l’instrument sur laquelle est placé le verre de lunette ou la lentille de contact à
mesurer
NOTE Le frontofocomètre mesure la puissance frontale par rapport à la surface placée contre le support de verre.
3.8
puissance de vision de près
puissance frontale mesurée au point de référence près indiqué par le fabricant pour un verre multifocal,
progressif ou dégressif
2 © ISO 2012 – Tous droits réservés

---------------------- Page: 8 ----------------------
ISO/FDIS 8598-1:2012(F)
3.9
erreur non symétrique pour la puissance cylindrique et l’axe du cylindre
erreur résiduelle dans la puissance cylindrique indiquée et/ou l’axe du cylindre indiqué d’un verre
sphéro-cylindrique pour un frontofocomètre automatique après étalonnage
3.10
erreur prismatique asymétrique d’un frontofocomètre
différence entre les indications de puissance prismatique données par les mesurages d’un prisme plan
avec sa base dans une direction puis dans la direction opposée; par exemple pour des réglages de base
de 180° et 360° ou 90° et 270°
3.11 Puissance frontale
3.11.1
puissance frontale arrière
inverse de la distance focale paraxiale par rapport à la face arrière mesurée en mètres
3.11.2
puissance frontale avant
inverse de la valeur paraxiale de la distance frontale avant, mesurée en mètres
NOTE 1 Par convention, la puissance frontale arrière, exprimée en dioptries, est indiquée comme la «puissance»
d’un verre de lunette ou d’une lentille de contact. La puissance frontale avant est cependant nécessaire dans
certains cas; par exemple pour le mesurage de certains verres multifocaux ou progressifs.
−1
NOTE 2 L’unité servant à exprimer la puissance frontale est le mètre à la puissance moins un (m ). Le nom de
cette unité est la «dioptrie» et son symbole est «D».
4 Exigences techniques pour les frontofocomètres pour cas généraux
4.1 La plage de mesure doit au minimum comprendre des puissances frontales de −20 D à +20 D et des
puissances prismatiques de 0 Δ à au moins 5 Δ.
L’instrument doit être capable de mesurer la direction de l’axe (voir l’ISO 8429) de verres cylindriques
entre 0° et 180°. Pour les prismes, il doit être possible de déterminer la direction de la base entre 0° et 360°.
4.2 Pour les frontofocomètres à mise au point manuelle et affichage non numérique, les intervalles
entre graduations de l’échelle de puissance dioptrique ne doivent pas dépasser 0,25 D et doivent être
suffisamment clairs pour faire des interpolations au 0,12 D ou moins le plus proche.
Pour les directions d’axe (voir l’ISO 8429), l’échelon ne doit pas dépasser 5° et doit être suffisamment
clair pour permettre des interpolations au degré le plus proche.
Pour les lectures de puissance prismatique, l’échelon ne doit pas dépasser 1 Δ et doit être suffisamment
clair pour permettre des interpolations au 0,50 Δ le plus proche.
4.3 Pour les frontofocomètres à affichage numérique dans la plage de +20 D à −20 D, l’incrément
minimal de l’écran ne doit pas être supérieur à 0,06 D. L’affichage doit donner au moins deux décimales.
Pour les directions d’axe, l’incrément de l’affichage numérique doit être de 1°.
Pour les lectures de puissance prismatique, l’incrément minimal de l’affichage numérique ne doit pas
dépasser 0,06 Δ.
4.4 L’instrument conçu pour mesurer des verres de lunette doit être capable de mesurer des verres
d’un diamètre d’au moins 80 mm et d’une épaisseur d’au moins 20 mm. Sur le support de verre, les
verres doivent pouvoir effectuer des mouvements de translation d’au moins 30 mm perpendiculairement
à l’axe optique et au rail de réglage, en démarrant à 10 mm ou moins sous ou devant l’axe optique de
© ISO 2012 – Tous droits réservés 3

---------------------- Page: 9 ----------------------
ISO/FDIS 8598-1:2012(F)
l’instrument, selon le cas. Le rail de réglage doit également pouvoir se déplacer de plus de 30 mm dans
une direction perpendiculaire à sa longueur et à l’axe optique de l’instrument. Voir Figure 1.
Dimensions en millimètres
Légende
1 support de verre
2 rail de réglage
Figure 1 — Mouvement minimal requis du rail de réglage
4.5 Il convient que le support de verre soit conçu de sorte que, lorsqu’il est utilisé suivant les instructions
du fabricant, il ne cause aucun dommage au verre soumis à essai.
Il doit être conçu et construit pour supporter un verre placé contre lui, en tenant la surface de contact
perpendiculaire à l’axe optique du frontofocomètre.
Le support de verre ne doit pas altérer l’exactitude de mesure par l’introduction d’une erreur sagittale.
La Figure 2 représente des exemples de supports de verre adaptés.
Il convient que d soit compris entre 5 mm et 8 mm pour les verres de lunettes (a) et que d soit de
1 c
4,5 mm ± 0,5 mm pour les lentilles de contact (b). Le diamètre du support de lentille de contact est
généralement inférieur et légèrement plus court en raison de la profondeur sagittale accrue des lentilles
de contact très incurvées. Il convient que la différence de hauteur (h – h ) soit de 0,55 mm ± 0,02 mm.
s c
NOTE Les dimensions du support de lentilles de contact sont telles que spécifiées dans l’ISO 18369-3.
Le diamètre interne (d d ) du support de verre des frontofocomètres utilisé pour le mesurage du verre
1, c
de lunettes ou de lentilles de contact doit être défini par le fabricant.
4.6 La marque réalisée par le marqueur d’axe doit être assez petite pour distinguer l’écart entre
deux marques.
NOTE Le diamètre recommandé est de 0,7 mm.
4.7 L’instrument doit être conçu pour donner des valeurs mesurées stables dans un environnement
normal (c’est-à-dire à une température de 23 °C ± 5 °C et une humidité relative inférieure à 85 %).
4 © ISO 2012 – Tous droits réservés

---------------------- Page: 10 ----------------------
ISO/FDIS 8598-1:2012(F)
Légende
d , d diamètre interne du support
1 c
d d diamètre externe du support
2, e
d = (d + d )/2
3 1 2
h , h hauteur du support
s c
r = (d - d )/4
1 2 1
r = (d – d )/4
2 e c
Figure 2 — Exemples de supports de verre de lunettes (a) et de lentilles de contact (b)
5 Exigences métrologiques
5.1 Généralités
Lorsqu’ils sont utilisés avec les verres de référence spécifiés dans l’ISO 9342-1 et/ou l’ISO 9342-2, les
frontofocomètres à mise au point manuelle comme ceux à mise au point automatique doivent donner des
indications de puissances frontale et prismatique sur la totalité de leur plage de mesure revendiquée.
L’erreur d’indication pour la valeur réelle en question ne doit, en outre, pas franchir les limites énoncées
dans les Tableaux 1 et 2.
5.2 Longueur d’onde de référence
L’instrument doit être construit pour indiquer des puissances dioptriques par rapport à la raie de
mercure verte λ = 546,07 nm ou la raie d’hélium jaune λ = 587,56 nm spécifiées dans l’ISO 7944 et le
e d
fabricant doit, par ailleurs, signaler explicitement quelle longueur d’onde a été sélectionnée.
NOTE Si la source lumineuse du frontofocomètre n’est pas centrée sur la raie de mercure verte λ = 546,07 nm
e
ou sur la raie d’hélium jaune λ = 587,56 nm, il convient que l’utilisateur soit en mesure de configurer l’instrument
d
ou d’appliquer aux valeurs indiquées des corrections pour les verres fabriqués avec des matériaux autres que
celui utilisé pour étalonner l’instrument.
5.3 Exigence de performance
5.3.1 Erreur d’indication
Lorsqu’il est pourvu du support de verre de lunette et utilisé avec les verres de référence de l’ISO 9342-1
et lorsqu’il est soumis à essai selon 6.2.1, l’instrument ne doit pas dépasser les tolérances d’erreurs
d’indication autorisées données dans les Tableaux 1 et 2.
© ISO 2012 – Tous droits réservés 5

---------------------- Page: 11 ----------------------
ISO/FDIS 8598-1:2012(F)
Si le fabricant a prévu de destiner l’instrument uniquement au mesurage de lentilles de contact,
l’instrument doit respecter les tolérances d’erreurs d’indication données dans le Tableau 1 quand il est
utilisé avec le support du diamètre spécifié en 4.5 par la Figure 2b) et, de préférence, avec les lentilles de
contact de référence de l’ISO 9342-2.
NOTE 1 L’exigence mentionnée ci-dessus est pour la vérification des erreurs d’indication du frontofocomètre
lorsqu’il est utilisé pour les mesures des verres de contact mais pas pour les erreurs admissibles des lentilles de
contact du commerce.
NOTE 2 Si on utilise l’instrument pour les mesurages des lentilles de contact mais qu’on l’étalonne avec les
verres de référence de lunettes indiquées dans l’ISO 9342-1, on doit l’indiquer, et il convient que l’utilisateur soit
conscient des différences significatives des résultats d’étalonnage obtenus par rapport aux résultats obtenus avec
étalonnage à partir de des lentilles de contact de référence indiquées dans l’ISO 9342-2.
Les erreurs d’indication maximales autorisées indiquées dans les Tableaux 1 et 2 correspondent
à l’utilisation des verres de référence de l’ISO 9342-1 et/ou de l’ISO 9342-2 dotés des valeurs réelles
correspondantes.
S’il est équipé de supports interchangeables, il convient que l’instrument puisse fournir des résultats
corrects pour les verres de référence quand le support de verre de lunettes est monté. En cas d’utilisation
d’une table de conversion d’étalonnage fournie par le fabricant, il convient que les lentilles de contact de
référence mesurées avec leur support donnent également des résultats corrects en utilisant la valeur de
correction fournie par la table de conversion. La valeur de correction fournie ne doit pas dépasser 0,25 D
pour une plage de mesures comprise entre −10,00 D et +10,00 D, et ne doit pas dépasser 0,50 D pour une
plage de mesures supérieure à ±10 D.
Pour les frontofocomètres automatiques, lorsqu’un des verres de référence sphériques est mesuré après
un centrage correct, il doit présenter une puissance astigmatique inférieure à 0,06 D.
Tableau 1 — Erreur d’indication maximale tolérée sur la puissance frontale mesurée par les
instruments pour cas généraux
Valeurs en dioptries (D)
Erreur d’indication maximale autorisée
Instruments à échelle Instruments numériques paramétrés sur des
Plage de mesure de puissance frontale
continue incréments de
a
0,25 0,12(5)
< 0 ≥ −5 > 0 ≤ +5 ±0,06 0,0 0,0
< −5 ≥ −10 > +5 ≤ +10 ±0,09 0,0 ±0,12(5)
< −10 ≥ −15 > +10 ≤ +15 ±0,12 0,0 ±0,12(5)
< −15 ≥ −20 > +15 ≤ +20 ±0,18 ±0,25 ±0,12(5)
< −20 > +20 ±0,25 ±0,25 ±0,25
NOTE 1 L’étalonnage à des tolérances plus faibles (par exemple: 10 % de la tolérance du produit) améliore l’équivalence de
l’instrument et réduit les éventuelles différences entre deux instruments.
NOTE 2 Il convient que l’utilisateur soit conscient que lorsqu’un frontofocomètre est paramétré pour lire par incréments de
0,01 D, les mesures peuvent ne pas être réelles à ce niveau de précision.
a ème
0,12(5) signifie que les instruments ont été paramétrés sur des incréments de 1/8 dioptrie, affichés sous forme de
0,12 D.
6 © ISO 2012 – Tous droits réservés

---------------------- Page: 12 ----------------------
ISO/FDIS 8598-1:2012(F)
Tableau 2 — Erreurs d’indication maximales tolérées sur la puissance prismatique mesurée
par les instruments pour cas généraux
Valeurs en dioptries prismatiques (Δ)
Erreur d’indication maximale autorisée
Plage de mesure de la puissance prisma- Instruments à échelle Instruments numériques paramétrés sur des
tique continue incréments de
a
0,25 0,12(5)
> 0 ≤ 5 0,25 0 0,12(5)
> 5 ≤ 10 0,25 0,25 0,25
> 10 ≤ 15 0,50 0,50 0,50
> 15 ≤ 20 0,75 0,75 0,75
> 20 1,00 1,00 1,00
NOTE L’étalonnage à des tolérances plus faibles (par exemple:10 % de la tolérance du produit) améliore l’équivalence de
l’instrument et réduit les éventuelles différences entre deux instruments.
a ème
0,12(5) signifie que les instruments ont été paramétrés sur des incréments de 1/8 dioptrie, affichés sous forme de
0,12 D.
5.3.2 Marqueur d’axe pour le centre optique du verre
Dans les conditions d’essai du paragraphe 6.3, la position du marqueur d’axe du centre optique du verre
ne doit pas s’écarter de l’axe optique du frontofocomètre de plus de 0,4 mm.
5.3.3 Alignement du marqueur d’axe
Lors d’un essai conforme à 6.4, le désalignement de l’orientation du marqueur d’axe ne doit pas dépasser
±1° par rapport aux directions 0° à 180° de l’échelle numérique ou par rapport à la valeur d’axe 0° (ou
180°) indiquée par les frontofocomètres automatiques.
5.3.4 Rail de réglage
Lors d’un essai conformément à 6.5, le désalignement du rail de réglage ne doit pas dépasser 1° par
rapport à la parallèle aux directions 0° à 180° de l’échelle de cadran angulaire pour les instruments à
mise au point manuelle ou par rapport à la valeur d’axe 0° (ou 180°) indiquée par les frontofocomètres
automatiques.
5.3.5 Erreurs non symétriques des frontofocomètres automatiques
5.3.5.1 Généralités
Les exigences relatives à la puissance cylindrique, à l’axe de cylindre et à l’erreur de prisme concernent
les frontofocomètres automatiques mesurant simultanément sphère, cylindre, axe du cylindre et prisme.
5.3.5.2 Erreur non symétrique pour la puissance cylindrique
Dans le cadre d’un essai conforme à 6.6.2, l’erreur non symétrique pour les indications de puissance
d’un verre de référence de puissance sphéro-cylindrique s
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.