Sterilization of health care products — Biological indicators — Part 1: General requirements

 ISO 11138-1:2017 specifies general requirements for production, labelling, test methods and performance characteristics of biological indicators, including inoculated carriers and suspensions, and their components, to be used in the validation and routine monitoring of sterilization processes. ISO 11138-1:2017 specifies basic and common requirements that are applicable to all parts of ISO 11138. Requirements for biological indicators for particular specified processes are provided in the relevant parts of ISO 11138. If no specific subsequent part is provided, this document applies. NOTE National or regional regulations can apply. ISO 11138-1:2017 does not apply to microbiological test systems for processes that rely on physical removal of microorganisms, e.g. filtration processes or processes that combine physical and/or mechanical removal with microbiological inactivation, such as use of washer disinfectors or flushing and steaming of pipelines. This document, however, can contain elements relevant to such microbiological test systems.

Stérilisation des produits de santé — Indicateurs biologiques — Partie 1: Exigences générales

 ISO 11138-1:2017 spécifie les exigences générales relatives à la production, à l'étiquetage, aux méthodes d'essai et fournit les caractéristiques de performance des indicateurs biologiques, notamment les porte-germes inoculés et les suspensions et leurs composants, à utiliser dans la validation et la surveillance de routine des procédés de stérilisation. ISO 11138-1:2017 spécifie les exigences fondamentales et communes applicables à toutes les parties pertinentes de l'ISO 11138. Les exigences relatives aux indicateurs biologiques pour des procédés particuliers spécifiés sont précisées dans les parties pertinentes de l'ISO 11138. En l'absence d'autre partie spécifique, le présent document s'applique. NOTE Les réglementations nationales ou régionales sont susceptibles de s'appliquer. ISO 11138-1:2017 ne s'applique pas aux systèmes d'essais microbiologiques utilisés dans les procédés fondés sur une extraction physique de micro-organismes, par exemple les procédés de filtration ou les procédés combinant une extraction physique et/ou mécanique et une inactivation microbiologique, par exemple l'utilisation de laveurs-désinfecteurs ou l'injection d'eau et de vapeur dans les conduites. Le présent document peut, toutefois, comporter des éléments applicables à ce type de systèmes d'essai microbiologiques.

General Information

Status
Published
Publication Date
02-Mar-2017
Current Stage
9093 - International Standard confirmed
Completion Date
09-Dec-2022
Ref Project

Relations

Buy Standard

Standard
ISO 11138-1:2017 - Sterilization of health care products -- Biological indicators
English language
41 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
ISO 11138-1:2017 - Stérilisation des produits de santé -- Indicateurs biologiques
French language
43 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

INTERNATIONAL ISO
STANDARD 11138-1
Third edition
2017-03
Sterilization of health care products —
Biological indicators —
Part 1:
General requirements
Stérilisation des produits de santé — Indicateurs biologiques —
Partie 1: Exigences générales
Reference number
ISO 11138-1:2017(E)
©
ISO 2017

---------------------- Page: 1 ----------------------
ISO 11138-1:2017(E)

COPYRIGHT PROTECTED DOCUMENT
© ISO 2017, Published in Switzerland
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized otherwise in any form
or by any means, electronic or mechanical, including photocopying, or posting on the internet or an intranet, without prior
written permission. Permission can be requested from either ISO at the address below or ISO’s member body in the country of
the requester.
ISO copyright office
Ch. de Blandonnet 8 • CP 401
CH-1214 Vernier, Geneva, Switzerland
Tel. +41 22 749 01 11
Fax +41 22 749 09 47
copyright@iso.org
www.iso.org
ii © ISO 2017 – All rights reserved

---------------------- Page: 2 ----------------------
ISO 11138-1:2017(E)

Contents Page
Foreword .iv
Introduction .v
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 1
4 General manufacturing requirements . 4
4.1 Manufacturing controls . 4
4.1.1 Quality management systems . 4
4.1.2 Traceability . 4
4.1.3 Finished product requirements . 4
4.1.4 Personnel . 4
4.2 Test organism . 4
4.2.1 Strain . 4
4.2.2 Originating inoculum for suspension. 5
4.2.3 Test organism count . 5
4.3 Information to be provided by the manufacturer (labelling) . 5
4.4 Storage and transport . 6
5 Specific manufacturing requirements . 7
5.1 Suspensions . 7
5.2 Carrier, primary and secondary packaging . 7
5.3 Inoculated carrier . 7
5.4 Biological indicators . 8
5.5 Self-contained biological indicators . 8
6 Determination of population and resistance . 8
6.1 General resistance requirements . 8
6.2 Test organism . 8
6.3 Population of test organisms . 8
6.4 Resistance characteristics . 9
6.5 Test conditions . 9
7 Culture conditions.10
7.1 Incubator .10
7.2 Growth medium .10
7.3 Incubation .10
7.4 Software validation .11
7.5 Incubation time using detection system .11
Annex A (normative) Determination of viable count .12
Annex B (normative) Determination of growth inhibition by carriers and primary
packaging materials exposed to sterilization processes .14
Annex C (normative) D value determination by survivor curve method .17
Annex D (normative) D value determination by fraction negative method .21
Annex E (normative) Survival-kill response characteristics .37
Annex F (informative) Relationship between components of biological indicators.39
Bibliography .40
© ISO 2017 – All rights reserved iii

---------------------- Page: 3 ----------------------
ISO 11138-1:2017(E)

Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www .iso .org/ directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of
any patent rights identified during the development of the document will be in the Introduction and/or
on the ISO list of patent declarations received (see www .iso .org/ patents).
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation on the voluntary nature of standards, the meaning of ISO specific terms and
expressions related to conformity assessment, as well as information about ISO’s adherence to the
World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT) see the following
URL: w w w . i s o .org/ iso/ foreword .html.
This document was prepared by Technical Committee ISO/TC 198, Sterilization of health care products.
This third edition cancels and replaces the second edition (ISO 11138-1:2006), which has been
technically revised.
A list of all parts of ISO 11138 can be found on the ISO website.
iv © ISO 2017 – All rights reserved

---------------------- Page: 4 ----------------------
ISO 11138-1:2017(E)

Introduction
This document specifies general requirements for production, labelling, test methods and performance
requirements for the manufacture of biological indicators including inoculated carriers and suspensions
intended for use in validation and monitoring of sterilization processes. Other parts of ISO 11138
provide additional specific requirements for biological indicators for defined sterilization processes.
A graphic description of a biological indicator and its components is presented in Table F.1. The
presentation includes the two types of biological indicators which are covered by ISO 11138 (all parts).
This shows that inoculated carriers can be presented directly to the sterilizing agent without prior
packaging, or included in a primary package that permits access by the sterilizing agent.
The resistance characteristics depend on the type of test organism, its numbers, the method of
preparation, the substrate upon which it is inoculated, environmental conditions during inoculation
and drying and the effects of the primary package. Advice on selection, use and interpretation of results
of biological indicators can be found in ISO 14161.
For any individual sterilization process, including those covered in relevant parts of ISO 11138, the
resistance of the biological indicator will also depend on its microenvironment during testing. In
theory, this could lead to an infinite variation in the preparation of biological indicators. Moreover, a
sterilization process could be manipulated in infinite variety to suit each possible set of conditions to
which products could be exposed. It has, therefore, been a routine practice to manufacture biological
indicators that, when exposed to a set of conditions in a defined sterilization process, provide resistance
characteristics expressed as D values and, where relevant, z values. Such values are set out in the
relevant parts of ISO 11138.
The ISO 11138 series represents the current “state-of-the-art” according to the experts representing
manufacturers, users and regulatory authorities involved in developing this document.
Biological indicators for specific sterilization processes not covered by reference test conditions in
relevant parts of ISO 11138 should comply with the general requirements in this document, including
the resistance testing procedures. Such biological indicators might not be well enough described,
or might be used for novel sterilization processes, or might be represented by isolated bioburden
microorganisms. If microorganisms other than risk group 1 (WHO 2004) are included in these biological
indicators, appropriate safety measures (e.g. containment) are necessary.
Standards exist providing requirements for the validation and control of sterilization processes
(see Bibliography).
NOTE It is possible that some countries or regions have published other standards covering requirements
for sterilization or biological indicators (see Bibliography).
© ISO 2017 – All rights reserved v

---------------------- Page: 5 ----------------------
INTERNATIONAL STANDARD ISO 11138-1:2017(E)
Sterilization of health care products — Biological
indicators —
Part 1:
General requirements
1 Scope
This document specifies general requirements for production, labelling, test methods and performance
characteristics of biological indicators, including inoculated carriers and suspensions, and their
components, to be used in the validation and routine monitoring of sterilization processes.
This document specifies basic and common requirements that are applicable to all parts of ISO 11138.
Requirements for biological indicators for particular specified processes are provided in the relevant
parts of ISO 11138. If no specific subsequent part is provided, this document applies.
NOTE National or regional regulations can apply.
This document does not apply to microbiological test systems for processes that rely on physical removal
of microorganisms, e.g. filtration processes or processes that combine physical and/or mechanical
removal with microbiological inactivation, such as use of washer disinfectors or flushing and steaming
of pipelines. This document, however, can contain elements relevant to such microbiological test
systems.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments) applies.
ISO 11135, Sterilization of health-care products — Ethylene oxide — Requirements for the development,
validation and routine control of a sterilization process for medical devices
ISO 11737-1:2006, Sterilization of medical devices — Microbiological methods — Part 1: Determination of
a population of microorganisms on products
ISO 14937, Sterilization of health care products — General requirements for characterization of a sterilizing
agent and the development, validation and routine control of a sterilization process for medical devices
ISO 17665-1, Sterilization of health care products — Moist heat — Part 1: Requirements for the development,
validation and routine control of a sterilization process for medical devices
ISO 18472, Sterilization of health care products — Biological and chemical indicators — Test equipment
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
ISO and IEC maintain terminological databases for use in standardization at the following addresses:
— IEC Electropedia: available at http:// www .electropedia .org/
— ISO Online browsing platform: available at https:// www .iso .org/ obp/
© ISO 2017 – All rights reserved 1

---------------------- Page: 6 ----------------------
ISO 11138-1:2017(E)

3.1
biological indicator
test system containing viable microorganisms providing a defined resistance to a specified
sterilization process
[SOURCE: ISO/TS 11139:2006, 2.3]
3.2
carrier
supporting material on or in which test microorganisms are deposited
3.3
colony forming unit
CFU
individual visible units of growth of microorganisms arising from a single cell or multiple cells
3.4
culture collection number
unique identification of the test organism allocated by a scientifically recognized service culture
collection
3.5
culture conditions
combination of growth media and manner of incubation used to promote germination, growth and/or
multiplication of microorganisms
Note 1 to entry: The manner of incubation can include the temperature, time and any other conditions specified
for incubation.
[SOURCE: ISO/TS 11139:2006, 2.10]
3.6
D value
D value
10
time or dose required to achieve inactivation of 90 % of a population of the test microorganism under
stated conditions
Note 1 to entry: Other critical process variable(s) exhibiting first order inactivation kinetics can achieve an
inactivation of 90 % of a population of the test microorganism under stated conditions.
[SOURCE: ISO/TS 11139:2006, 2.11, modified]
3.7
inactivation
loss of ability of microorganisms to grow and/or multiply
[SOURCE: ISO/TS 11139:2006, 2.21]
3.8
inoculated carrier
supporting material on or in which a defined number of viable test organisms have been deposited
Note 1 to entry: See Annex F.
3.9
nominal population
manufacturer’s stated number of viable microorganisms
6
Note 1 to entry: This is generally expressed in log function (e.g. 10 ).
10
2 © ISO 2017 – All rights reserved

---------------------- Page: 7 ----------------------
ISO 11138-1:2017(E)

3.10
primary package
element of the packaging system which maintains the integrity of the product
Note 1 to entry: The packaging system protects the inoculated carrier from damage and contamination without
preventing penetration of the sterilizing agent.
3.11
process challenge device
PCD
item designed to constitute a defined resistance to a sterilization process and used to assess
performance of the process
[SOURCE: ISO/TS 11139:2006, 2.33]
3.12
resistometer
test equipment designed to create defined reference combinations of the physical and/or chemical
variables of a sterilization process
[SOURCE: ISO 18472:2006, 3.11, modified]
3.13
secondary package
container in which biological indicators are packed for transport and storage
3.14
self-contained biological indicator
biological indicator presented in such a way that the primary package, intended for incubation, contains
the incubation medium required for recovery of the test organism
3.15
survival-kill window
extent of exposure to a sterilization process under defined conditions where there is a transition from
all biological indicators showing growth to all biological indicators showing no growth
3.16
survivor curve
graphical representation of the inactivation of a population of microorganisms with increasing
exposure to a microbicidal agent under stated conditions
3.17
suspension
viable test organisms suspended in a fluid
Note 1 to entry: Suspension can be a biological indicator if ready to use in a sealed glass ampoule or may be an
intermediate component used to produce an inoculated carrier or biological indicator.
3.18
viable count
actual number of recoverable colony-forming units or other appropriate units
Note 1 to entry: See Annex A.
3.19
z value
change in exposure temperature of a thermal sterilization process, which corresponds to a tenfold
change in D value
Note 1 to entry: See ISO 11138-3 and ISO 11138-4.
© ISO 2017 – All rights reserved 3

---------------------- Page: 8 ----------------------
ISO 11138-1:2017(E)

4 General manufacturing requirements
4.1 Manufacturing controls
4.1.1 Quality management systems
A formal quality system (e.g. ISO 13485, GMPs or other national or regional requirements) to cover all
operations required by this document shall be established, documented and maintained. In particular,
precautions at all stages of production to minimize contamination that would adversely affect the
performance of the biological indicator shall be taken.
4.1.2 Traceability
4.1.2.1 Traceability of manufacturing components shall be maintained.
4.1.2.2 Manufacturing components shall include all materials incorporated in, or coming into direct
contact with, the test organism suspension, the inoculated carrier and/or its primary package.
4.1.3 Finished product requirements
The finished product shall comply with the following requirements:
a) labelling (4.3);
b) manufacturing (Clause 5);
c) resistance characteristics (6.4);
d) storage and transport (4.4);
e) incubation (7.3).
NOTE 1 Advice on methods for the use of biological indicators is provided in ISO 14161.
NOTE 2 National and/or regional requirements might exist, for example, in the various national or regional
pharmacopoeias.
4.1.4 Personnel
The procedures and methods in this document shall be carried out by suitably trained and experienced
laboratory personnel (see e.g. ISO 13485).
4.2 Test organism
4.2.1 Strain
4.2.1.1 Test organisms shall be of a defined strain, available through a recognized culture collection,
and shall be identified by appropriate test methods. A statement of traceability shall be provided to the
purchaser upon request.
4.2.1.2 The test organism shall be a strain that is
a) suitable for handling without special containment facilities, does not need specific containment
procedures for handling and does not have specific transport or mailing requirements (e.g. Risk
Group 1, WHO 2004), and
4 © ISO 2017 – All rights reserved

---------------------- Page: 9 ----------------------
ISO 11138-1:2017(E)

b) sufficiently stable to maintain its resistance characteristics for the duration of the stated shelf-life
when transported and stored in accordance with label directions.
NOTE Traditionally, the test organisms of biological indicators have been bacterial spores, usually derived
from Bacillus or Geobacillus species.
4.2.1.3 Test organisms other than bacterial spores may be used if they have been shown to provide
appropriate resistance to the sterilization process.
4.2.2 Originating inoculum for suspension
4.2.2.1 The initial inoculum for each batch of test organism suspension shall be
a) traceable to the reference culture and available through a recognized culture collection, and
b) verified as to its identity and purity.
4.2.2.2 The methods used for maintaining test organism cultures shall be designed to protect
them from contamination and to minimize any induced changes in the inherent properties of the test
organisms.
4.2.2.3 Verification tests are specific for each strain of test organism and shall be documented and
validated.
4.2.3 Test organism count
The viable test organism count of the suspension shall be determined in accordance with Annex A.
4.3 Information to be provided by the manufacturer (labelling)
4.3.1 The following information shall be provided on the label of each individual unit of suspension,
inoculated carrier packaging and biological indicator:
— a unique code by which the manufacturing history can be traced;
— the name of the test organism;
— an indication of the sterilization process for which the suspension, inoculated carriers or biological
indicators are suitable;
— the expiry date, expressed according to ISO 8601, e.g. YYYY-MM-DD;
— the manufacturer’s name, trademark, address or other means of identification.
Internationally recognized symbols may be used where appropriate (see 4.1.3, ISO 15223-1, and
ISO 11140-1).
© ISO 2017 – All rights reserved 5

---------------------- Page: 10 ----------------------
ISO 11138-1:2017(E)

4.3.2 In addition to the information provided in 4.3.1, the information given in Table 1 shall be provided
on or within the secondary packaging of each batch of product.
Table 1 — Information to be provided by the manufacturer
Inoculated Biological
Information requirement Suspension
carrier indicator
The nominal volume of suspension, in ml Required — —
The process for which the product is suitable for use,
the resistance and the procedure and carrier used to Required Required Required
a
determine the resistance
The specified storage conditions Required Required Required
Disposal instructions Required Required Required
Instructions for use, especially data about the medium,
incubation and other conditions to be used for recov-
Required Required Required
ery of test organisms after exposure to the steriliza-
tion process
The number of test organisms per ml (suspension), or
Required Required Required
a
per unit (inoculated carrier or biological indicator)
The number of product units in the secondary pack — Required Required
A reference to this document Required Required Required
a
Test methodology used to determine resistance and population should be supplied by the manufacturer upon request.
4.3.3 If a specific instrument is required to detect outgrowth of the biological indicator, the instrument
manufacturer shall specify use of that instrument and provide a certificate of calibration and information
on the required recalibration procedures.
4.3.4 If applicable, information shall be provided on required maintenance service of the specified
instrument including detailed instructions of how to perform the maintenance, service and its frequency.
4.4 Storage and transport
4.4.1 Storage and transport conditions for the test organism suspension shall be maintained such that
the test organism suspension complies with the requirements of this document and, where relevant, a
subsequent part of ISO 11138.
4.4.2 If inoculated carriers are packaged for storage or transport, they shall be packaged in a way that
does not adversely affect the nominal population or performance of individual inoculated carriers.
4.4.3 Storage and transport conditions for inoculated carriers shall be maintained such that the
inoculated carriers comply with the requirements of this document and where relevant, a subsequent
part of ISO 11138.
4.4.4 Individually-packaged biological indicators shall be placed in a secondary package for transport
and storage. Packaging for transport and storage shall ensure that biological indicators comply with this
document and, where relevant, a subsequent part of ISO 11138.
6 © ISO 2017 – All rights reserved

---------------------- Page: 11 ----------------------
ISO 11138-1:2017(E)

5 Specific manufacturing requirements
5.1 Suspensions
5.1.1 Culture medium and incubation conditions shall consistently produce test organism suspensions
that meet the performance requirements of this document and any relevant subsequent part of
ISO 11138.
5.1.2 The suspending medium for the test organism suspension shall not adversely affect the stability
of the test organism and shall be compatible with the procedures and materials employed in the
manufacture of inoculated carriers and biological indicators.
5.1.3 The method of harvesting and subsequent treatment of suspensions to be used in the inoculation
of carriers shall ensure that residues do not adversely influence the performance of the inoculated carrier
or biological indicator.
5.2 Carrier, primary and secondary packaging
5.2.1 The materials of the carrier and the primary and secondary packaging shall not contain any
contamination (physical, chemical or microbial) that would adversely affect the performance of the
biological indicator.
5.2.2 The carrier, the primary and secondary packaging, and the specified storage conditions shall be
designed so that the performance characteristics of the biological indicator meet the requirements of
this document throughout the stated shelf life of the product. The purchaser shall be provided with a
statement of the maximum and minimum values of each dimension of the carrier on request.
5.2.3 During and after the sterilization process, the carrier and the primary packaging shall not retain
or release any substance to such an extent that, on transfer to the incubation medium, under the culture
conditions, the growth of low numbers of surviving test organisms will be inhibited. Testing shall comply
with Annex B.
5.2.4 The carrier, the primary packaging, and the secondary packaging shall withstand planned
transport and handling at the point of use, without damage or breakage.
5.2.5 Raw materials used for the carrier and the primary packaging shall withstand exposure to the
sterilization pr
...

NORME ISO
INTERNATIONALE 11138-1
Troisième édition
2017-03
Version corrigée
2018-02
Stérilisation des produits de santé —
Indicateurs biologiques —
Partie 1:
Exigences générales
Sterilization of health care products — Biological indicators —
Part 1: General requirements
Numéro de référence
ISO 11138-1:2017(F)
©
ISO 2017

---------------------- Page: 1 ----------------------
ISO 11138-1:2017(F)

DOCUMENT PROTÉGÉ PAR COPYRIGHT
© ISO 2017, Publié en Suisse
Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite ni utilisée
sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie, l’affichage sur
l’internet ou sur un Intranet, sans autorisation écrite préalable. Les demandes d’autorisation peuvent être adressées à l’ISO à
l’adresse ci-après ou au comité membre de l’ISO dans le pays du demandeur.
ISO copyright office
Ch. de Blandonnet 8 • CP 401
CH-1214 Vernier, Geneva, Switzerland
Tel. +41 22 749 01 11
Fax +41 22 749 09 47
copyright@iso.org
www.iso.org
ii © ISO 2017 – Tous droits réservés

---------------------- Page: 2 ----------------------
ISO 11138-1:2017(F)

Sommaire Page
Avant-propos .iv
Introduction .v
1 Domaine d'application . 1
2 Références normatives . 1
3 Termes et définitions . 1
4 Exigences générales de fabrication . 4
4.1 Contrôles de fabrication . 4
4.1.1 Systèmes de management de la qualité . 4
4.1.2 Traçabilité . 4
4.1.3 Exigences relatives au produit fini . 4
4.1.4 Personnel . 4
4.2 Organisme d’essai . 5
4.2.1 Souche . 5
4.2.2 Inoculum de départ utilisé pour la suspension . 5
4.2.3 Dénombrement des organismes d’essai . 5
4.3 Informations à fournir par le fabricant (étiquetage) . 5
4.4 Stockage et transport . 6
5 Exigences spécifiques relatives à la fabrication. 7
5.1 Suspensions . 7
5.2 Porte-germes, emballage primaire et secondaire . 7
5.3 Porte-germes inoculé . 8
5.4 Indicateurs biologiques . 8
5.5 Indicateurs biologiques autonomes . 8
6 Détermination de la population et de la résistance. 8
6.1 Exigences générales applicables à la résistance . 8
6.2 Organisme d’essai . 9
6.3 Population d’organismes d’essai . 9
6.4 Caractéristiques de résistance . 9
6.5 Conditions d’essai .10
7 Conditions de culture .10
7.1 Étuve bactériologique .10
7.2 Milieu de culture .10
7.3 Incubation .11
7.4 Validation des logiciels .11
7.5 Temps d’incubation au moyen d’un système de détection .11
Annexe A (normative) Détermination du nombre d’organismes viables .12
Annexe B (normative) Détermination de l’inhibition de la croissance par les matériaux des
porte-germes et de l’emballage primaire exposés aux procédés de stérilisation .14
Annexe C (normative) Détermination de la valeur de D selon la méthode de la courbe de survie .17
Annexe D (normative) Détermination de la valeur de D selon la méthode de la fraction négative .22
Annexe E (normative) Caractéristiques de réponse survie-destruction .38
Annexe F (informative) Relations entre les composants d’indicateurs biologiques .40
Bibliographie .42
© ISO 2017 – Tous droits réservés iii

---------------------- Page: 3 ----------------------
ISO 11138-1:2017(F)

Avant-propos
L'ISO (Organisation internationale de normalisation) est une fédération mondiale d'organismes
nationaux de normalisation (comités membres de l'ISO). L'élaboration des Normes internationales est
en général confiée aux comités techniques de l'ISO. Chaque comité membre intéressé par une étude
a le droit de faire partie du comité technique créé à cet effet. Les organisations internationales,
gouvernementales et non gouvernementales, en liaison avec l'ISO participent également aux travaux.
L'ISO collabore étroitement avec la Commission électrotechnique internationale (IEC) en ce qui
concerne la normalisation électrotechnique.
Les procédures utilisées pour élaborer le présent document et celles destinées à sa mise à jour sont
décrites dans les Directives ISO/IEC, Partie 1. Il convient, en particulier de prendre note des différents
critères d'approbation requis pour les différents types de documents ISO. Le présent document a été
rédigé conformément aux règles de rédaction données dans les Directives ISO/IEC, Partie 2 (voir www
.iso .org/ directives).
L'attention est appelée sur le fait que certains des éléments du présent document peuvent faire l'objet de
droits de propriété intellectuelle ou de droits analogues. L'ISO ne saurait être tenue pour responsable
de ne pas avoir identifié de tels droits de propriété et averti de leur existence. Les détails concernant
les références aux droits de propriété intellectuelle ou autres droits analogues identifiés lors de
l'élaboration du document sont indiqués dans l'Introduction et/ou dans la liste des déclarations de
brevets reçues par l'ISO (voir www .iso .org/ brevets).
Les appellations commerciales éventuellement mentionnées dans le présent document sont données
pour information, par souci de commodité, à l’intention des utilisateurs et ne sauraient constituer un
engagement.
Pour une explication de la nature volontaire des normes, la signification des termes et expressions
spécifiques de l'ISO liés à l'évaluation de la conformité, ou pour toute information au sujet de l'adhésion
de l'ISO aux principes de l’Organisation mondiale du commerce (OMC) concernant les obstacles
techniques au commerce (OTC), voir le lien suivant: www .iso .org/ iso/ fr/ avant -propos .html
Le présent document a été élaboré par le comité technique ISO/TC 198, Stérilisation des produits de santé.
Cette troisième édition annule et remplace la deuxième édition (ISO 11138-1:2006), qui a fait l’objet
d’une révision technique.
Une liste de toutes les parties de la série ISO 11138, se trouve sur le site Web de l’ISO.
La présente version corrigée de l’ISO 11138-1:2017 inclut la correction suivante:
— En B.4.4, le texte «une absence de croissance» a été modifié en «croissance».
iv © ISO 2017 – Tous droits réservés

---------------------- Page: 4 ----------------------
ISO 11138-1:2017(F)

Introduction
Le présent document spécifie les exigences générales relatives à la production, à l’étiquetage, aux
méthodes d’essai et les exigences de performance applicables à la fabrication d’indicateurs biologiques
y compris les porte-germes inoculés et les suspensions utilisés dans la validation et la surveillance
des procédés de stérilisation. Les autres parties de l’ISO 11138 présentent des exigences spécifiques
supplémentaires relatives aux indicateurs biologiques utilisés dans des procédés de stérilisation
définis.
Une description schématique d’un indicateur biologique et de ses composants est présentée dans le
Tableau F.1. Elle présente les deux types d’indicateurs biologiques couverts par l’ISO 11138 (toutes
les parties). Elle montre que les porte-germes inoculés peuvent être directement mis en contact avec
l’agent stérilisant sans emballage préalable ou conditionnés dans un emballage primaire dans lequel
l’agent stérilisant peut pénétrer.
Les caractéristiques de résistance dépendent de l’organisme d’essai, de sa quantité, de la méthode de
préparation, du substrat sur lequel il est inoculé, des conditions environnementales durant l’inoculation
et de l’influence de l’emballage primaire. Des conseils concernant le choix, l’utilisation des indicateurs
biologiques et l’interprétation des résultats peuvent être obtenus dans l’ISO 14161.
Dans tout procédé de stérilisation individuel, y compris ceux traités dans les parties pertinentes de
l’ISO 11138, la résistance de l’indicateur biologique dépendra aussi de son micro-environnement
pendant l’essai. En théorie, ceci pourrait conduire à une multitude infinie de variations dans la
préparation des indicateurs biologiques. De plus, les manipulations effectuées lors du procédé de
stérilisation pourraient varier à l’infini afin d’adapter le procédé concerné à chaque ensemble de
conditions auxquelles le produit est susceptible d’être exposé. La pratique courante a donc consisté à
fabriquer des indicateurs biologiques qui, lorsqu’ils sont exposés à un ensemble de conditions dans le
cadre d’un procédé de stérilisation défini, présentent des caractéristiques de résistance exprimées sous
la forme de valeurs de D et, le cas échéant, de valeurs z. Ces valeurs sont présentées dans les parties
pertinentes de l’ISO 11138.
La série ISO 11138 représente l’état actuel des connaissances, selon les spécialistes chargés de
représenter les fabricants, les utilisateurs et les autorités réglementaires impliqués dans l’élaboration
du présent document.
Il convient que les indicateurs biologiques des procédés spécifiques de stérilisation non couverts par
les conditions d’essai de référence décrites dans les parties pertinentes de l’ISO 11138 soient conformes
aux exigences générales exposées dans le présent document, y compris les modes opératoires d’essai
de résistance. Ces indicateurs biologiques peuvent être décrits de manière incomplète, ou être
utilisés pour des procédés de stérilisation nouveaux ou représentés par des micro-organismes isolés
de la charge biologique. Si des micro-organismes autres que ceux appartenant au groupe de risque 1
(OMS 2004) sont inclus dans ces indicateurs biologiques, des mesures de sécurité appropriées (par
exemple confinement) sont nécessaires.
Il existe des normes décrivant des exigences applicables à la validation et au contrôle des procédés de
stérilisation (voir Bibliographie).
NOTE Il est possible que certains pays ou certaines zones géographiques aient publié d’autres normes
couvrant des exigences relatives à la stérilisation ou aux indicateurs biologiques (voir Bibliographie).
© ISO 2017 – Tous droits réservés v

---------------------- Page: 5 ----------------------
NORME INTERNATIONALE ISO 11138-1:2017(F)
Stérilisation des produits de santé — Indicateurs
biologiques —
Partie 1:
Exigences générales
1 Domaine d'application
Le présent document spécifie les exigences générales relatives à la production, à l’étiquetage, aux
méthodes d’essai et fournit les caractéristiques de performance des indicateurs biologiques, notamment
les porte-germes inoculés et les suspensions et leurs composants, à utiliser dans la validation et la
surveillance de routine des procédés de stérilisation.
Le présent document spécifie les exigences fondamentales et communes applicables à toutes les parties
pertinentes de l’ISO 11138. Les exigences relatives aux indicateurs biologiques pour des procédés
particuliers spécifiés sont précisées dans les parties pertinentes de l’ISO 11138. En l’absence d’autre
partie spécifique, le présent document s’applique.
NOTE Les réglementations nationales ou régionales sont susceptibles de s’appliquer.
Le présent document ne s’applique pas aux systèmes d’essais microbiologiques utilisés dans les procédés
fondés sur une extraction physique de micro-organismes, par exemple les procédés de filtration ou les
procédés combinant une extraction physique et/ou mécanique et une inactivation microbiologique, par
exemple l’utilisation de laveurs-désinfecteurs ou l’injection d’eau et de vapeur dans les conduites. Le
présent document peut, toutefois, comporter des éléments applicables à ce type de systèmes d’essai
microbiologiques.
2 Références normatives
Les documents suivants cités dans le texte constituent, pour tout ou partie de leur contenu, des
exigences du présent document. Pour les références datées, seule l’édition citée s’applique. Pour les
références non datées, la dernière édition du document de référence s'applique (y compris les éventuels
amendements).
ISO 11135, Stérilisation des produits de santé — Oxyde d'éthylène — Exigences de développement, de
validation et de contrôle de routine d'un processus de stérilisation pour des dispositifs médicaux
ISO 11737-1:2006, Stérilisation des dispositifs médicaux — Méthodes microbiologiques — Partie 1:
Détermination d'une population de micro-organismes sur des produits
ISO 14937, Stérilisation des produits de santé — Exigences générales pour la caractérisation d'un agent
stérilisant et pour la mise au point, la validation et la vérification de routine d'un processus de stérilisation
pour dispositifs médicaux
ISO 17665-1, Stérilisation des produits de santé — Chaleur humide — Partie 1: Exigences pour le
développement, la validation et le contrôle de routine d'un procédé de stérilisation des dispositifs médicaux
ISO 18472, Stérilisation des produits de santé — Indicateurs biologiques et chimiques — Appareillage d’essai
3 Termes et définitions
Pour les besoins du présent document, les termes et définitions suivants s’appliquent.
© ISO 2017 – Tous droits réservés 1

---------------------- Page: 6 ----------------------
ISO 11138-1:2017(F)

L’ISO et l’IEC tiennent à jour des bases de données terminologiques destinées à la normalisation,
disponibles aux adresses suivantes:
— IEC Electropedia: disponible à l'adresse http:// www .electropedia .org/
— ISO Online browsing platform: disponible à l'adresse https:// www .iso .org/ obp/
3.1
indicateur biologique
système d’essai contenant des micro-organismes viables, garantissant une résistance définie à un
procédé de stérilisation spécifié
[SOURCE: ISO/TS 11139:2006, 2.3]
3.2
porte-germes
support sur ou dans lequel des micro-organismes d’essai sont déposés
3.3
unité formant colonie
UFC
unités individuelles visibles de la croissance de micro-organismes se développant à partir d’une ou de
plusieurs cellules
3.4
numéro de collection de culture
identification unique de l’organisme d’essai allouée par un centre de collection de culture de micro-
organismes scientifiquement reconnu
3.5
conditions de culture
association de milieux de culture et de conditions d’incubation pour favoriser la germination, la
croissance et/ou la multiplication de micro-organismes
Note 1 à l'article: Les conditions d’incubation peuvent inclure la température, le temps et toute autre condition
d’incubation spécifiée.
[SOURCE: ISO/TS 11139:2006, 2.10]
3.6
valeur de D
valeur de D
10
temps ou dose nécessaire pour inactiver 90 % d’une population de micro-organismes d’essai dans des
conditions établies
Note 1 à l'article: Une ou plusieurs autres variables de procédé présentant une cinétique d’inactivation de premier
ordre peuvent atteindre une inactivation de 90 % d’une population de micro-organismes d’essai dans des
conditions établies.
[SOURCE: ISO/TS 11139:2006, 2.11, modifiée]
3.7
inactivation
perte de la capacité des micro-organismes à croître et/ou à se multiplier
[SOURCE: ISO/TS 11139:2006, 2.21]
3.8
porte-germes inoculé
support sur ou dans lequel a été déposé un nombre défini d’organismes d’essai viables
Note 1 à l'article: Voir Annexe F.
2 © ISO 2017 – Tous droits réservés

---------------------- Page: 7 ----------------------
ISO 11138-1:2017(F)

3.9
population nominale
nombre de micro-organismes viables, déclaré par le fabricant
6
Note 1 à l'article: Ce nombre est généralement exprimé à l’aide d’une fonction log (par exemple 10 ).
10
3.10
emballage primaire
élément du système d’emballage qui maintient l’intégrité du produit
Note 1 à l'article: Le système d’emballage protège le porte-germes inoculé de l’endommagement et de la
contamination sans empêcher la pénétration de l’agent stérilisant.
3.11
dispositif d’épreuve de procédé
PCD
article conçu pour constituer une résistance définie à un procédé de stérilisation et destiné à évaluer
l’efficacité du procédé
[SOURCE: ISO/TS 11139:2006, 2.33]
3.12
résistomètre
matériel d’essai conçu pour créer des combinaisons de référence définies des paramètres physiques
et/ou chimiques d’un procédé de stérilisation
[SOURCE: ISO 18472:2006, 3.11, modifiée]
3.13
emballage secondaire
récipient dans lequel les indicateurs biologiques sont emballés en vue de leur transport et de leur
stockage
3.14
indicateur biologique autonome
indicateur biologique présenté de telle manière que l’emballage primaire, destiné à l’incubation,
contienne le milieu d’incubation nécessaire à la récupération des organismes d’essai
3.15
fenêtre de survie/destruction
étendue de l’exposition à un procédé de stérilisation dans des conditions définies, entre la situation
où tous les indicateurs biologiques présentent une croissance et la situation où tous les indicateurs
biologiques ne présentent aucune croissance
3.16
courbe de survie
représentation graphique de l’inactivation d’une population de micro-organismes avec l’augmentation
de l’exposition à un agent microbicide dans des conditions établies
3.17
suspension
organismes d’essai viables en suspension dans un fluide
Note 1 à l'article: La suspension peut constituer un indicateur biologique si elle est prête à l’emploi dans une
ampoule de verre scellée ou peut constituer un composant intermédiaire utilisé pour la préparation d’un porte-
germes inoculé ou d’un indicateur biologique.
3.18
dénombrement d’organismes viables
nombre réel d’unités formant colonies ou d’autres unités appropriées récupérables
Note 1 à l'article: Voir Annexe A.
© ISO 2017 – Tous droits réservés 3

---------------------- Page: 8 ----------------------
ISO 11138-1:2017(F)

3.19
valeur de z
variation de la température d’exposition dans un procédé de stérilisation thermique, qui correspond à
une variation décuplée de la valeur de D
Note 1 à l'article: Voir l’ISO 11138-3 et l’ISO 11138-4.
4 Exigences générales de fabrication
4.1 Contrôles de fabrication
4.1.1 Systèmes de management de la qualité
Un système qualité officiel (par exemple l’ISO 13485, les BPF ou d’autres exigences nationales ou
régionales) permettant de couvrir toutes les opérations exigées par le présent document doit être mis
en place, documenté et maintenu à jour. Des précautions doivent en particulier être prises à toutes les
étapes de la production pour réduire le plus possible la contamination qui pourrait léser l’indicateur
biologique.
4.1.2 Traçabilité
4.1.2.1 La traçabilité des composants de fabrication doit être assurée.
4.1.2.2 Les composants de fabrication doivent inclure tous les matériaux incorporés ou en contact
direct avec la suspension d’organismes d’essai, le porte-germes inoculé et/ou son emballage primaire.
4.1.3 Exigences relatives au produit fini
Le produit fini doit être conforme aux exigences suivantes:
a) étiquetage (4.3);
b) fabrication (Article 5);
c) caractéristiques de résistance (6.4);
d) stockage et transport (4.4);
e) incubation (7.3).
NOTE 1 Pour des conseils concernant les méthodes d’utilisation des indicateurs biologiques, consulter
l’ISO 14161.
NOTE 2 Il peut exister des exigences nationales et/ou régionales, par exemple dans les différentes
pharmacopées nationales ou régionales.
4.1.4 Personnel
Les modes opératoires et les méthodes décrits dans le présent document doivent être exécutés par un
personnel de laboratoire dûment formé et expérimenté (voir par exemple ISO 13485).
4 © ISO 2017 – Tous droits réservés

---------------------- Page: 9 ----------------------
ISO 11138-1:2017(F)

4.2 Organisme d’essai
4.2.1 Souche
4.2.1.1 Les organismes d’essai doivent être issus d’une souche définie, approvisionnable auprès d’un
centre de collection de culture de micro-organismes reconnu et doivent être identifiés par des méthodes
d’essai appropriées. Une déclaration de traçabilité doit être fournie à l’acheteur sur demande.
4.2.1.2 L’organisme d’essai doit être une souche qui est:
a) adaptée à la manipulation sans dispositifs de confinement spécifique, qui ne nécessite pas de
modes opératoires de confinement spécifique pour la manipulation et n’implique aucune exigence
spécifique pour le transport ou l’expédition (par exemple Groupe de Risque 1, OMS, 2004); et
b) suffisamment stable pour conserver ses caractéristiques de résistance pendant la durée de
conservation stipulée, lorsque les conditions de transport et de stockage sont conformes aux
indications mentionnées sur l’étiquetage.
NOTE Traditionnellement, les organismes d’essai des indicateurs biologiques sont des spores bactériennes,
généralement dérivées des espèces Bacillus ou Geobacillus.
4.2.1.3 Les organismes d’essai autres que les spores bactériennes peuvent être utilisés s’il a été
démontré qu’ils opposent une résistance appropriée au procédé de stérilisation.
4.2.2 Inoculum de départ utilisé pour la suspension
4.2.2.1 L’inoculum de départ utilisé pour chaque lot de suspension d’organismes d’essai doit satisfaire
aux conditions suivantes:
a) être traçable par rapport à la culture de référence et obtenu auprès d’une collection de culture
reconnue; et
b) être vérifié quant à son identité et à sa pureté.
4.2.2.2 Les méthodes de préservation des cultures d’organismes d’essai utilisées doivent être étudiées
pour garantir la protection de ces cultures contre la contamination et pour réduire le plus possible des
modifications induites dans les propriétés intrinsèques des organismes d’essai.
4.2.2.3 Les essais de vérification sont spécifiques à chaque souche d’organisme d’essai et doivent être
documentés et validés.
4.2.3 Dénombrement des organismes d’essai
Le nombre des organismes d’essai viables en suspension doit être déterminé conformément à l’Annexe A.
4.3 Informations à fournir par le fabricant (étiquetage)
4.3.1 Les informations suivantes doivent être indiquées sur l’étiquetage de chaque unité de suspension,
de chaque emballage de porte-germes inoculé et de chaque indicateur biologique:
— un code unique permettant de retracer l’historique de la fabrication;
— le nom de l’organisme d’essai;
— une indication du procédé de stérilisation pour lequel la suspension, les porte-germes inoculés ou
les indicateurs biologiques sont appropriés;
© ISO 2017 – Tous droits réservés 5

---------------------- Page: 10 ----------------------
ISO 11138-1:2017(F)

— la date d’expiration, exprimée conformément à l’ISO 8601, par exemple AAAA-MM-JJ;
— le nom du fabricant, la marque commerciale, l’adresse ou tout autre moyen d’identification.
Des symboles reconnus sur le plan international peuvent être utilisés si cela s’avère approprié (voir
4.1.3, ISO 15223-1, et ISO 11140-1).
4.3.2 Outre les informations fournies en 4.3.1, les informations suivantes, données dans le Tableau 1,
doivent être fournies sur ou à l’intérieur de l’emballage secondaire de chaque lot de produit.
Tableau 1 — Informations devant être fournies par le fabricant
Porte-germes Indicateur
Informations exigées Suspension
inoculé biologique
Le volume nominal de la suspension, en ml Exigé — —
Le procédé pour lequel le produit peut être employé,
la résistance, le mode opératoire et le porte-germes Exigé Exigé Exigé
a
utilisé pour déterminer la résistance
Les conditions de stockage spécifiées Exigé Exigé Exigé
Les instructions pour la mise au rebut Exigé Exigé Exigé
Les instructions d’utilisation, en particulier les don-
nées concernant le milieu, l’incubation et les autres
conditions à respecter pour la récupération des Exigé Exigé Exigé
organismes d’essai après exposition au procédé de
stérilisation
Le nombre d’organismes d’essai par ml (suspension)
ou par unité (porte-germes inoculé ou indicateur Exigé Exigé Exigé
a
biologique)
Le nombre d’unités de produit dans l’emballage secondaire — Exigé Exigé
Une référence au présent document Exigé Exigé Exigé
a
Il convient que les méthodes d’essai utilisées pour déterminer la résistance et la population soient fournies par le
fabricant sur demande.
4.3.3 Si un instrument spécifique est requis pour détecter la croissance de l’indicateur biologique, le
fabricant de l’instrument doit préciser l’utilisation de cet instrument et fournir un certificat d’étalonnage
et des informations sur les modes opératoires de réétalonnage requis.
4.3.4 Le cas échéant, des informations doivent être fournies sur le service de maintenance requis
pour les instruments indiqués, y compris des instructions détaillées sur le mode de réalisation de la
maintenance, du service et leur fréquence.
4.4 Stockage et transport
4.4.1 Les conditions de stockage et de transport de la suspension d’organismes d’essai doivent être
préservées pour garantir la conformité de la suspension d’organismes d’essai avec les exigences du
présent document et, le cas échéant, d’une autre partie d
...