Microbeam analysis — Electron probe microanalysis — Guidelines for the determination of experimental parameters for wavelength dispersive spectroscopy

This document gives general guidelines for the determination of experimental parameters relating to the electron probe, the wavelength spectrometer, and the specimen that need to be taken into account when carrying out electron probe microanalysis. It also defines procedures for the determination of probe current, probe diameter, dead time, wavelength resolution, background, analysis area, analysis depth, and analysis volume. This document is applicable for the analysis of a well-polished specimen using normal beam incidence. This document does not apply to energy dispersive X-ray spectroscopy.

Analyse par microfaisceaux — Analyse par microsonde électronique (Microsonde de Castaing) — Lignes directrices pour la détermination des paramètres expérimentaux pour la spectrométrie à dispersion de longueur d'onde

General Information

Status
Published
Publication Date
06-Jun-2024
Current Stage
6060 - International Standard published
Start Date
07-Jun-2024
Due Date
10-May-2024
Completion Date
07-Jun-2024
Ref Project

Relations

Standard
ISO 14594:2024 - Microbeam analysis — Electron probe microanalysis — Guidelines for the determination of experimental parameters for wavelength dispersive spectroscopy Released:7. 06. 2024
English language
18 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


International
Standard
ISO 14594
Third edition
Microbeam analysis — Electron
2024-06
probe microanalysis — Guidelines
for the determination of
experimental parameters for
wavelength dispersive spectroscopy
Analyse par microfaisceaux — Analyse par microsonde
électronique (Microsonde de Castaing) — Lignes directrices
pour la détermination des paramètres expérimentaux pour la
spectrométrie à dispersion de longueur d'onde
Reference number
© ISO 2024
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on
the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address below
or ISO’s member body in the country of the requester.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii
Contents Page
Foreword .iv
Introduction .v
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 1
4 Abbreviated terms . 3
5 Experimental parameters . 3
5.1 General .3
5.2 Parameters related to the electron probe .3
5.2.1 Accelerating voltage .3
5.2.2 Probe current .3
5.2.3 Magnification and field of view .3
5.3 Parameters related to wavelength dispersive X-ray spectrometers .3
5.3.1 General .3
5.3.2 Take-off angle .4
5.3.3 Wavelength resolution .4
5.3.4 X-ray detector and Pulse height analyser .4
5.3.5 Peak location (wavelength) .4
5.3.6 Background .5
5.4 Parameters related to the specimen .5
5.4.1 Specimen stage .5
5.4.2 Surface roughness .5
5.4.3 X-ray line .5
5.4.4 Analysis volume .5
6 Procedures and measurements . 6
6.1 General .6
6.2 Electron probe .6
6.2.1 Probe current .6
6.2.2 Probe diameter .6
6.3 Parameters related to measured peaks .7
6.3.1 Dead time correction .7
6.3.2 Wavelength resolution of detected characteristic X-ray peaks .7
6.3.3 Background subtraction .8
6.4 Parameters related to the specimen .9
6.4.1 General .9
6.4.2 Analysis area . .9
6.4.3 Analysis depth .9
6.4.4 Analysis volume .9
7 Test report . 9
Annex A (informative) Methods of estimating analysis area .11
Annex B (informative) Methods of estimating analysis depth .13
Annex C (informative) Method of estimating X-ray analysis volume by applying the Monte Carlo
(MC) simulation . 14
Bibliography .18

iii
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out through
ISO technical committees. Each member body interested in a subject for which a technical committee
has been established has the right to be represented on that committee. International organizations,
governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely
with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are described
in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the different types
of ISO document should be noted. This document was drafted in accordance with the editorial rules of the
ISO/IEC Directives, Part 2 (see www.iso.org/directives).
ISO draws attention to the possibility that the implementation of this document may involve the use of (a)
patent(s). ISO takes no position concerning the evidence, validity or applicability of any claimed patent
rights in respect thereof. As of the date of publication of this document, ISO had not received notice of (a)
patent(s) which may be required to implement this document. However, implementers are cautioned that
this may not represent the latest information, which may be obtained from the patent database available at
www.iso.org/patents. ISO shall not be held responsible for identifying any or all such patent rights.
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and expressions
related to conformity assessment, as well as information about ISO's adherence to the World Trade
Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www.iso.org/iso/foreword.html.
This document was prepared by Technical Committee ISO/TC 202, Microbeam analysis, Subcommittee SC 2,
Electron probe microanalysis.
This third edition cancels and replaces the second edition (ISO 14594:2014), which has been technically
revised.
The main changes are as follows:
— Introduction has been added;
— Terms in Clause 3 have been updated;
— Technical terms in Clause 5 have been updated and the clause has been restructured;
— Content of the Test report in Clause 7 has been revised.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www.iso.org/members.html.

iv
Introduction
To ensure reliability and reproducibility during electron probe microanalysis (EPMA), the experimental
parameters that include beam current, current density, dead time, wavelength resolution, background,
analysis area, analysis depth, and analysis volume should be carefully considered. To reliably consider EPMA
results, guidelines standardizing the decision procedure of an experimental parameter are important.

v
International Standard ISO 14594:2024(en)
Microbeam analysis — Electron probe microanalysis —
Guidelines for the determination of experimental parameters
for wavelength dispersive spectroscopy
1 Scope
This document gives general guidelines for the determination of experimental parameters relating to the electron
probe, the wavelength spectrometer, and the specimen that need to be taken into account when carrying out
electron probe microanalysis. It also defines procedures for the determination of probe current, probe diameter,
dead time, wavelength resolution, background, analysis area, analysis depth, and analysis volume.
This document is applicable for the analysis of a well-polished specimen using normal beam incidence.
This document does not apply to energy dispersive X-ray spectroscopy.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes
requirements of this document. For dated references, only the edition cited applies. For undated references,
the latest edition of the referenced document (including any amendments) applies.
ISO/IEC 17025:2017, General requirements for the competence of testing and testing laboratories
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
ISO and IEC maintain terminology databases for use in standardization at the following addresses:
— ISO Online browsing platform: available at https:// www .iso .org/ obp
— IEC Electropedia: available at https:// www .electropedia .org/
3.1
analysis area
area, projected by the interaction volume on the beam entrance surface, from which X-rays are emitted (at a
defined fraction, e.g. 95 % of the total) and collected by the spectrometer
[SOURCE: ISO 23833:2013, 5.7.1.1]
3.2
analysis depth
maximum depth from which a defined fraction (e.g. 95 % of the total) of the X-rays are emitted from the
interaction volume after absorption
[SOURCE: ISO 23833:2013, 5.7.1.2]
3.3
analysis volume
volume from which a defined fraction (e.g. 95 % of the total) of the X-rays are emitted after generation and
absorption
[SOURCE: ISO 23833:2013, 5.7.1.3]

3.4
background
non-characteristic component of an X-ray spectrum arising (ideally) from the X-ray continuum
[SOURCE: ISO 23833:2013, 5.7.1.2]
3.5
probe current
electron current contained within the electron probe
[SOURCE: ISO 23833:2013, 4.3.1]
3.6
probe diameter
diameter of the probe containing a specified fraction of the total current, for example 0,8 (80 %) of the total
[SOURCE: ISO 23833:2013, 4.3.2]
3.7
dead time
time that the system is unable to record a photon measurement because it is busy processing a previous
event and frequently expressed as a percentage of the total time
[SOURCE: ISO 23833:2013, 4.5.6]
3.8
wavelength resolution
full peak width at half maximum of a peak in terms of wavelength (Δλ) obtained from a single X-ray transition
by a WDS
[SOURCE: ISO 23833:2013, 4.6.16]
3.9
field of view
length in the X- and Y-direction of the image or mapping area
3.10
overvoltage ratio
ratio of the incident beam energy to the critical excitation energy for a particular atomic shell
Note 1 to entry: This factor must be greater than unity for characteristic X-ray production to occur from that atomic shell.
[SOURCE: ISO 23833:2013, 5.1.3]
3.11
Johansson optics
wavelength-dispersive X-ray spectrometer in which the diffractor is bent to a radius twice that of the
Rowland circle and then its surface ground to the radius, achieving a fully focussing situation
[SOURCE: ISO 23833:2013, 4.6.14.8]
3.12
Johann optics
wavelength-dispersive X-ray spectrometer in which the diffractor is bent to a radius twice that of the
Rowland circle, achieving a “semi focussing” situation
[SOURCE: ISO 23833:2013, 4.6.14.7]

4 Abbreviated terms
EPMA electron probe microanalysis or electron probe microanalyser
FWHM full width at half maximum
WDS wavelength-dispersive spectrometer
WDX wavelength-dispersive X-ray spectrometry
5 Experimental parameters
5.1 General
The parameters given in 5.2.1, 5.2.2, and 5.2.3 should be setup properly according to the purpose of the
experiment and recorded. Checking the calibration of probe current, and magnification together with
counter dead time should be included in the maintenance schedule of the instrument.
5.2 Parameters related to the electron probe
5.2.1 Accelerating voltage
The beam energy accelerating voltage typically ranges from 2 keV to 30 keV. Since the sensitivity and
spatial resolution (or analysis volume) of analysis depend on the accelerating voltage, the optimization of
accelerating voltage can be critical in some cases. However, these performances are also depending on the
material of the specimen and the measured X-ray lines. The optimization of accelerating voltage is specified
in the guidance for the element and line (see 5.4.3) and the analysis volume (See 5.4.4).
5.2.2 Probe current
Because X-ray peak intensity is directly proportional to the probe current, the precision of the measurement
of the probe current should be better than the precision required for quantitative analysis.
Probe current stability over long periods of time is essential for consistent quantitative analysis. The probe
current stability should be tested periodically, especially prior to quantitative calibration and analysis. It
is possible to compensate for small changes in probe current if this is recorded prior to and following each
measurement. Then all X-ray peak and background measurements should be scaled appropriately by I /I
i m,
where I is the initial probe current and I is the probe current at the time of the measurement.
i m
5.2.3 Magnification and field of view
To properly define the field of view for line-scans and images acquired by deflecting the electron probe, it is
essential to calibrate the magnification scale while operating in the scanning electron mode.
5.3 Parameters related to wavelength dispersive X-ray spectrometers
5.3.1 General
An instrument may be fitted with one or more WDX spectrometers, each with a number of diffracting
crystals-depending on the line of the analysed element. The following parameters are important for the
proper operation of WDX spectrometers.

5.3.2 Take-off angle
The take-off angle affects quantitative analysis. Any comparison of measurements from instruments with
different take-off angles should be taken into account and the procedures used be noted in the analysis report.
NOTE The value of this angle, which is normally fixed, is provided by the instrument manufacturer.
5.3.3 Wavelength resolution
The spectral resolution depends on the following parameters:
— crystal material (and Miller indices of the crystal planes);
— the radius of curvature of the diffracting crystal (Johansson optics vs. Johann optics);
— the size and position of the counter entrance window or of the entrance slit if present.
All these settings determine the wavelength resolution of the measured X-ray spectrum and the observed
line-width (FWHM) of the characteristic X-ray peaks. X-ray lines of analysed elements and X-ray order shall
be recorded in the test report.
Resolution can also influence the ability of the system to discriminate against overlapping peaks, background
signals, and the sensitivity of measurements to specimen height and beam position on the specimen.
5.3.4 X-ray detector and Pulse height analyser
Many spectrometers use a gas-filled proportional counter to detect X-rays. The magnitude of the output
pulses from these detectors is determined by the incident X-ray energy and/or the voltage applied to the
counters. Two discriminators are used to select the pulse of interest. A low discriminator setting is used to
eliminate pulses due to noise, while a high discriminator setting excludes pulses from high order reflections
of more energetic X-rays. Optimum settings depend on the X-ray lines of interest.
It is important to set the discriminator to ensure that any unintended shift in pulse amplitude, for example,
due to high count rates or changes in atmospheric temperature and pressure (flow counter), has no
significant effect on the measured count rate.
Because X-ray counting efficiency decreases with increasing count rate, it is important to correct the
measured count rate for the effect of the dead time. In an automated system, the discriminator settings can
be set automatically. These settings should be routinely checked to ensure proper automatic operation.
5.3.5 Peak location (wavelength)
Under normal circumstances, the wavelength which has the maximum peak intensity is used to define
the location of an X-ray peak. It is necessary, using suitable reference materials, to periodically check
and correct for the difference in a peak’s theoretical position and its actual measured position on a given
spectrometer and diffraction crystal. The time between checks will depend on the stability of the instrument
spectrometers.
The measured maximum intensities of peaks which have narrow FWHM values are strongly affected by the
errors in peak location. The peak intensity can be changed due to the chemical state and polarization effects.
NOTE 1 If the element in the specimen of interest is in a different chemical state than that of the reference
material, then the shape of the characteristic X-ray peak can be different for specimen and standard. In this case, it
is possible that the peak maximum does not provide a reliable measure of the total peak intensity and an
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...