Safety requirements for electrical equipment for measurement, control and laboratory use - Part 031: Safety requirements for hand-held and hand-manipulated probe assemblies for electrical test and measurement (IEC 61010-031:2015/A1:2018)

NEXT ACTION: TB ACTION BY 2021-09-14: common mods, currently at ENQ
2019-10-03 JF: NWI for A11 (PR=70187) was approved by the BTthrough decision D163/098. The work on the amendment is under way
2019-03-11 TC is addressing the negative assessment and is invited to provide the information whether the changes will be needed in the main body of the amendment or only in Annex ZZ.
Negative assessment. IEC standard is published.

Siicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte - Teil 031: Sicherheitsbestimmungen für handgehaltenes Messzubehör zum Messen und Prüfen (IEC 61010-031:2015/A1:2018)

Exigences de sécurité pour appareils électriques de mesurage, de régulation et de laboratoire - Partie 031: Exigences de sécurité pour sondes équipées portatives et manipulées à la main pour mesurage et essais électriques (IEC 61010-031:2015/A1:2018)

Varnostne zahteve za električno opremo za meritve, nadzor in laboratorijsko uporabo - 031. del: Varnostne zahteve za ročne sonde za električne meritve in preskušanje - Dopolnilo A1 (IEC 61010-031:2015/A1:2018)

General Information

Status
Published
Public Enquiry End Date
30-Jul-2017
Publication Date
01-Feb-2022
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
26-Nov-2021
Due Date
31-Jan-2022
Completion Date
02-Feb-2022

RELATIONS

Buy Standard

Amendment
SIST EN 61010-031:2015/A1:2022
English language
18 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day
Amendment
SIST EN 61010-031:2015/oprA1:2017
English language
12 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (sample)

SLOVENSKI STANDARD
SIST EN 61010-031:2015/A1:2022
01-marec-2022
Varnostne zahteve za električno opremo za meritve, nadzor in laboratorijsko
uporabo - 031. del: Varnostne zahteve za ročne sonde za električne meritve in
preskušanje - Dopolnilo A1 (IEC 61010-031:2015/A1:2018)

Safety requirements for electrical equipment for measurement, control and laboratory

use - Part 031: Safety requirements for hand-held and hand-manipulated probe
assemblies for electrical test and measurement (IEC 61010-031:2015/A1:2018)

Siicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte - Teil

031: Sicherheitsbestimmungen für handgehaltenes Messzubehör zum Messen und
Prüfen (IEC 61010-031:2015/A1:2018)

Exigences de sécurité pour appareils électriques de mesurage, de régulation et de

laboratoire - Partie 031: Exigences de sécurité pour sondes équipées portatives et

manipulées à la main pour mesurage et essais électriques (IEC 61010-
031:2015/A1:2018)
Ta slovenski standard je istoveten z: EN 61010-031:2015/A1:2021
ICS:
19.080 Električno in elektronsko Electrical and electronic
preskušanje testing
71.040.10 Kemijski laboratoriji. Chemical laboratories.
Laboratorijska oprema Laboratory equipment
SIST EN 61010-031:2015/A1:2022 en,fr,de

2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------
SIST EN 61010-031:2015/A1:2022
---------------------- Page: 2 ----------------------
SIST EN 61010-031:2015/A1:2022
EUROPEAN STANDARD EN 61010-031:2015/A1
NORME EUROPÉENNE
EUROPÄISCHE NORM
November 2021
ICS 19.080
English Version
Safety requirements for electrical equipment for measurement,
control and laboratory use - Part 031: Safety requirements for
hand-held and hand-manipulated probe assemblies for electrical
test and measurement
(IEC 61010-031:2015/A1:2018)

Exigences de sécurité pour appareils électriques de Sicherheitsbestimmungen für elektrische Mess-, Steuer-,

mesurage, de régulation et de laboratoire - Partie 031: Regel- und Laborgeräte - Teil 031:

Exigences de sécurité pour sondes équipées portatives et Sicherheitsbestimmungen für handgehaltenes

manipulées à la main pour mesurage et essais électriques Messzubehör zum Messen und Prüfen

(IEC 61010-031:2015/A1:2018) (IEC 61010-031:2015/A1:2018)

This amendment A1 modifies the European Standard EN 61010-031:2015; it was approved by CENELEC on 2018-07-03. CENELEC

members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this amendment the

status of a national standard without any alteration.

Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC

Management Centre or to any CENELEC member.

This amendment exists in three official versions (English, French, German). A version in any other language made by translation under the

responsibility of a CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the same status as

the official versions.

CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,

Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the

Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,

Turkey and the United Kingdom.
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels

© 2021 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.

Ref. No. EN 61010-031:2015/A1:2021 E
---------------------- Page: 3 ----------------------
SIST EN 61010-031:2015/A1:2022
EN 61010-031:2015/A1:2021 (E)
European foreword

The text of document 66/664/FDIS, future IEC 61010-031/A1, prepared by IEC/TC 66 "Safety of

measuring, control and laboratory equipment" was submitted to the IEC-CENELEC parallel vote and

approved by CENELEC as EN 61010-031:2015/A1:2021.
The following dates are fixed:

• latest date by which the document has to be implemented at national (dop) 2022-05-12

level by publication of an identical national standard or by endorsement

• latest date by which the national standards conflicting with the (dow) 2024-11-12

document have to be withdrawn

Attention is drawn to the possibility that some of the elements of this document may be the subject of

patent rights. CENELEC shall not be held responsible for identifying any or all such patent rights.

This document has been prepared under a Standardization Request given to CENELEC by the

European Commission and the European Free Trade Association, and supports essential

requirements of EU Directive(s) / Regulation(s).

For the relationship with EU Directive(s) / Regulation(s), see informative Annex ZZ, which is an

integral part of EN 61010-031:2015/A11:2021.

Any feedback and questions on this document should be directed to the users’ national committee. A

complete listing of these bodies can be found on the CENELEC website.
Endorsement notice

The text of the International Standard IEC 61010-031/A1:2018 was approved by CENELEC as a

European Standard without any modification.
---------------------- Page: 4 ----------------------
SIST EN 61010-031:2015/A1:2022
IEC 61010-031
Edition 2.0 2018-05
INTERNATIONAL
STANDARD
GROUP SAFETY PUBLICATION
AMENDMENT 1
Safety requirements for electrical equipment for measurement, control and
laboratory use –
Part 031: Safety requirements for hand-held and hand-manipulated probe
assemblies for electrical test and measurement
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
ICS 19.080
ISBN 978-2-8322-5745-6

Warning! Make sure that you obtained this publication from an authorized distributor.

® Registered trademark of the International Electrotechnical Commission
---------------------- Page: 5 ----------------------
SIST EN 61010-031:2015/A1:2022
– 2 – IEC 61010-031:2015/AMD1:2018
© IEC 2018
FOREWORD

This amendment has been prepared by IEC technical committee 66: Safety of measuring,

control and laboratory equipment.
The text of this standard is based on the following documents:
FDIS Report on voting
66/664/FDIS 66/670/RVD

Full information on the voting for the approval of this amendment can be found in the report

on voting indicated in the above table.

The committee has decided that the contents of this amendment and the base publication will

remain unchanged until the stability date indicated on the IEC website under

"http://webstore.iec.ch" in the data related to the specific publication. At this date, the

publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
A bilingual version of this publication may be issued at a later date.
_____________
Title
Replace the part title as follows:
Part 031: Safety requirements for hand-held and hand-manipulated
probe assemblies for electrical test and measurement
1 Scope and object
1.1.1 Probe assemblies included in scope
Figure 4 – Examples of Type D probe assemblies
Delete, in key 3, the words “or clamp”.
---------------------- Page: 6 ----------------------
SIST EN 61010-031:2015/A1:2022
IEC 61010-031:2015/AMD1:2018 – 3 –
© IEC 2018
3 Terms and definitions
3.1.1
TERMINAL
Note 1 to entry:
Delete the word “connectors,”.
3.1.5
CONNECTOR

Delete, in the definition, the words “a CONNECTOR of” so that the end of the sentence will read:

“…to connect to a TERMINAL of the equipment or to another probe assembly”.
3.4.11
MEASUREMENT CATEGORY
Replace the existing text of the definition with the following:

classification of testing and measuring circuits according to the types of mains to which they

are intended to be connected
4 Tests
4.3.9 Duty cycle
Replace the existing title of 4.3.9 with the following:
4.3.9 Short-term or intermittent operation
4.4.4.2 Temperature
Replace the existing second paragraph with the following:

This temperature is determined by measuring the temperature rise of the surface or part and

adding it to the ambient temperature of 40 °C, or to the maximum RATED ambient temperature

if higher.
4.5.2 Fuses

Replace, in the fourth sentence of the second paragraph, the word “MAINS” (in SMALL CAPS)

with “mains” (in regular font).
5 Marking and documentation
5.1.5 RATING

Replace, in list item a) of the first paragraph, the words “(see also 5.4.3 f) and g))” with “(see

also 5.4.3 k))”.
---------------------- Page: 7 ----------------------
SIST EN 61010-031:2015/A1:2022
– 4 – IEC 61010-031:2015/AMD1:2018
© IEC 2018
6 Protection against electric shock
6.2.2 Examination

Figure 6 – Methods for determination of ACCESSIBLE parts (see 6.2) and for voltage tests

(see 6.4.2)
Replace the existing subfigures 6c and 6d with the following new subfigures:

Connecting parts are partially mated so as just to make electrical contact while allowing maximum access to the

test finger.
Note the two possible positions of the test finger.
Figure 6c – Partially mated probe assemblies (see 6.2 and 6.4.2 b))
Key
F rigid test finger (see Figure B.1)
H potentially HAZARDOUS LIVE part
Note the two possible positions of the test finger.
Figure 6d – Unmated parts of a probe assembly (see 6.2 and 6.4.2 c))
6.3.4.1 General
Replace, in the first paragraph, the existing list with the following:
a) the probe body;
b) hand-held or hand-manipulated parts of each CONNECTOR;
± 20 mm of the PROBE WIRE or the maximum length of the cable whichever is
c) 150 mm
shorter;
d) other hand-held or hand-manipulated parts.
---------------------- Page: 8 ----------------------
SIST EN 61010-031:2015/A1:2022
IEC 61010-031:2015/AMD1:2018 – 5 –
© IEC 2018
Figure 8 – Voltage and touch current measurement
Delete, in the text of key item 2c, the words “(see 12.3.2)”.
6.3.4.2 Probe assemblies with floating outer conductors
Figure 10 – Voltage and touch current measurement with shielded test probe
Delete, in the text of key item 2c, the words “(see 12.3.2)”.
6.3.4.3 High frequency test probes

Replace, in the fourth paragraph, “the circuit from A.3” with “the circuit from Figure A.3”.

6.4.1 General
Replace, in the third paragraph, the existing list with the following list:

a) DOUBLE INSULATION, consisting of BASIC INSULATION plus SUPPLEMENTARY INSULATION (see

6.4.6)
b) BASIC INSULATION plus impedance (see 6.4.4);
c) REINFORCED INSULATION (see 6.4.6);
d) PROTECTIVE IMPEDANCE (see 6.4.5).
6.4.2 CONNECTORS
Add, at the end of the sentence of 6.4.2 c) 2), “(see Figure 5)”.
6.4.3.2 Protection by a PROTECTIVE FINGERGUARD
Replace the existing third paragraph with the following:

The height of the PROTECTIVE FINGERGUARD from the side where the fingers are intended to be

applied shall be at least 2 mm.
6.4.3.4 Protection by tactile indicator
Replace the existing first paragraph with the following:

SPRING-LOADED CLIPS RATED for MEASUREMENT CATEGORY II or without MEASUREMENT CATEGORY

for maximum 300 V which require finger pressure at about 90° to the axis of the clip are

acceptable without a PROTECTIVE FINGERGUARD, provided that there is a tactile indicator to

indicate the limit of safe access for the OPERATOR.
---------------------- Page: 9 ----------------------
SIST EN 61010-031:2015/A1:2022
– 6 – IEC 61010-031:2015/AMD1:2018
© IEC 2018
6.4.3.5 PROBE TIPS used as CONNECTORS
Replace the text with the following text:

PROBE TIPS which can be used as CONNECTORS and are intended to be connected to specified

accessories (for example to a SPRING-LOADED CLIP) shall, in combination with those

accessories, also meet the requirements for CONNECTORS in fully-mated position and partially-

mated position (see 6.4.2 a) and b)).
6.5.1.2.4 SOLID INSULATION

Delete the entire subclause including Table 4, Table 5, Figure 15, Figure 16 and Figure 17.

Add the following new subclause:
6.5.1.3 Solid insulation

The term “solid insulation” is used to describe many different types of construction, including

monolithic blocks of insulating material and insulation subsystems composed of multiple

insulating materials, organized in layers or otherwise.

The electric strength of a thickness of solid insulation is considerably greater than that of the

same thickness of air. The insulating distances through solid insulation are therefore typically

smaller than the distances through air. As a result, electric fields in solid insulation are

typically higher, and often are less homogeneous.

Solid insulation material may contain gaps or voids. When a solid insulation system is

constructed from layers of solid materials, there are also likely to be gaps or voids between

layers. These voids will perturb the electric field so that a disproportionately large part of the

electric field is located in the void, potentially causing ionization within the void, resulting in

partial discharge. These partial discharges will influence the adjacent solid insulation and may

reduce its service life.

Solid insulation is not a renewable medium: damage is cumulative over the life of the

equipment. Solid insulation is also subject to ageing and to degradation from repeated high

voltage testing.
Conformity is checked as specified in 6.5.2.5 and 6.5.2.6.
6.5.2.2 CLEARANCES for probe assemblies of MEASUREMENT CATEGORIES II, III and IV
Replace the existing Table 6 with the following:
---------------------- Page: 10 ----------------------
SIST EN 61010-031:2015/A1:2022
IEC 61010-031:2015/AMD1:2018 – 7 –
© IEC 2018
Table 6 – CLEARANCES of probe assemblies RATED for MEASUREMENT CATEGORIES
Nominal a.c. r.m.s. CLEARANCE
line-to-neutral or
d.c. voltage of mm
mains to which the
BASIC INSULATION
probe assembly is
REINFORCED INSULATION
and SUPPLEMENTARY INSULATION
designed to be
connected
MEASUREMENT MEASUREMENT MEASUREMENT MEASUREMENT MEASUREMENT MEASUREMENT
CATEGORY II CATEGORY III CATEGORY IV CATEGORY II CATEGORY III CATEGORY IV
0,04 0,1 0,5 0,10 0,32 1,4
≤ 50
> 50 ≤ 100 0,1 0,5 1,5 0,32 1,4 3,0
> 100 ≤ 150 0,5 1,5 3,0 1,4 3,0 6,0
1,5 3,0 5,5 3,0 6,0 10,4
> 150 ≤ 300
3,0 5,5 8 6,0 10,4 15
> 300 ≤ 600
> 600 ≤ 1 000 5,5 8 14 10,4 15 23,9
8 11 18 16 22 36
> 1 000 ≤ 1 500
14 18 22 28 36 44
> 1 500 ≤ 2 000
> 2 000 ≤ 3 000 18 22 25 36 44 50

Replace, in the conformity statement, the word “clearance” with “CLEARANCE” (in SMALL CAPS).

6.5.2.3.1 General

Replace, in the second paragraph, the words “required clearance” with “required CLEARANCE”.

Add the following two new subclauses, Table 4, Table 5, Table 14, Figure 15, Figure 16 and

Figure 17:
6.5.2.5 Solid insulation of probe assemblies RATED for MEASUREMENT CATEGORIES
6.5.2.5.1 General

6.5.2.5.1.1 Solid insulation of probe assemblies RATED for MEASUREMENT CATEGORIES shall

withstand the electrical and mechanical stresses that may occur in NORMAL USE and in all

RATED environmental conditions (see 1.4) during the intended life of the probe assembly.

The manufacturer should take the expected life of the probe assembly into account when

selecting insulating materials.
Conformity is checked by both of the following tests:

a) the a.c. voltage test of 6.6.5.1 with a duration of at least 5 s using the applicable test

voltage of Table 4 or the impulse voltage test of 6.6.5.3 using the applicable test voltage

of Table 14;

b) the a.c. voltage test of 6.6.5.1 or if stressed only by d.c., the d.c. voltage test of 6.6.5.2,

with a duration of at least 1 min using the test voltage determined by 6.5.2.5.1.2.

NOTE Test a) checks the effects of transient overvoltages, while test b) checks the effects of long-term stress of

solid insulation.
---------------------- Page: 11 ----------------------
SIST EN 61010-031:2015/A1:2022
– 8 – IEC 61010-031:2015/AMD1:2018
© IEC 2018
Table 4 – a.c. test voltages for testing electric strength
of solid insulation in probe assemblies RATED for MEASUREMENT CATEGORIES
a.c. test voltage
Nominal a.c. r.m.s.
line-to-neutral or
V r.m.s
d.c. voltage of
mains being BASIC INSULATION
REINFORCED INSULATION
measured and SUPPLEMENTARY INSULATION
MEASUREMENT MEASUREMENT MEASUREMENT MEASUREMENT MEASUREMENT MEASUREMENT
CATEGORY II CATEGORY III CATEGORY IV
...

SLOVENSKI STANDARD
SIST EN 61010-031:2015/oprA1:2017
01-julij-2017

9DUQRVWQH]DKWHYH]DHOHNWULþQRRSUHPR]DPHULWYHQDG]RURYDQMHLQODERUDWRULMVNR

XSRUDERGHO9DUQRVWQH]DKWHYH]DURþQHVRQGH]DHOHNWULþQHPHULWYHLQ
SUHVNXãDQMH

Safety requirements for electrical equipment for measurement, control and laboratory

use - Part 031: Safety requirements for hand-held and hand-manipulated probe
assemblies for electrical test and measurement
Ta slovenski standard je istoveten z: EN 61010-031:2015/prA1:2017
ICS:
19.080 (OHNWULþQRLQHOHNWURQVNR Electrical and electronic
SUHVNXãDQMH testing
71.040.10 Kemijski laboratoriji. Chemical laboratories.
Laboratorijska oprema Laboratory equipment
SIST EN 61010-031:2015/oprA1:2017 en,fr,de

2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------
SIST EN 61010-031:2015/oprA1:2017
---------------------- Page: 2 ----------------------
SIST EN 61010-031:2015/oprA1:2017
66/632/CDV
COMMITTEE DRAFT FOR VOTE (CDV)
PROJECT NUMBER:
IEC 61010-031/AMD1 ED2
DATE OF CIRCULATION: CLOSING DATE FOR VOTING:
2017-05-19 2017-08-11
SUPERSEDES DOCUMENTS:
IEC TC 66 : SAFETY OF MEASURING, CONTROL AND LABORATORY EQUIPMENT
SECRETARIAT: SECRETARY:
United Kingdom Mr David Hyde
OF INTEREST TO THE FOLLOWING COMMITTEES: PROPOSED HORIZONTAL STANDARD:
GROUP SAFETY FUNCTION
TC 78,TC 85
Other TC/SCs are requested to indicate their interest, if
any, in this CDV to the secretary.
FUNCTIONS CONCERNED:
EMC ENVIRONMENT QUALITY ASSURANCE SAFETY

SUBMITTED FOR CENELEC PARALLEL VOTING NOT SUBMITTED FOR CENELEC PARALLEL VOTING

This document is still under study and subject to change. It should not be used for reference purposes.

Recipients of this document are invited to submit, with their comments, notification of any relevant patent rights of

which they are aware and to provide supporting documentation.
TITLE:

Safety requirements for electrical equipment for measurement, control and laboratory use – Part

031: Safety requirements for hand-held and hand-manipulated probe assemblies for electrical

test and measurement.
NOTE FROM TC/SC OFFICERS:

Copyright © 2017 International Electrotechnical Commission, IEC. All rights reserved. It is permitted to

download this electronic file, to make a copy and to print out the content for the sole purpose of preparing

National Committee positions. You may not copy or "mirror" the file or printed version of the document, or any part

of it, for any other purpose without permission in writing from IEC.
---------------------- Page: 3 ----------------------
SIST EN 61010-031:2015/oprA1:2017
61010-031/Ed2/A1/CDV © IEC (E) 2 66/632/CDV
1 FOREWORD

2 This amendment has been prepared by IEC technical committee 66: Safety of measuring,

3 control and laboratory equipment.
4 The text of this standard is based on the following documents:
FDIS Report on voting
66/XX/FDIS 66/XX/RVD

6 Full information on the voting for the approval of this standard can be found in the report on

7 voting indicated in the above table.

8 The committee has decided that the contents of this publication will remain unchanged until

9 the stability date indicated on the IEC web site under "http://webstore.iec.ch" in the data

10 related to the specific publication. At this date, the publication will be
11  reconfirmed,
12  withdrawn,
13  replaced by a revised edition, or
14  amended.

16 The National Committees are requested to note that for this publication the stability date

17 is 2021.

18 THIS TEXT IS INCLUDED FOR THE INFORMATION OF THE NATIONAL COMMITTEES AND WILL BE DELETED

19 AT THE PUBLICATION STAGE.
21 Change the sub-title to read:
23 Part 031: Safety requirements for hand-held and hand-manipulated probe
24 assemblies for electrical test and measurement
26 1 Scope and object
27 1.1.1 Probe assemblies included in Scope
28 Figure 4, Key 3
29 Deletion
30 Delete the words “or clamp”
31 2 Normative references
32 3 Terms and definitions
33 3.1.1
34 TERMINAL
35 Note 1 to entry
36 Deletion
37 Delete the word “connectors”
38 3.1.5
39 CONNECTOR
---------------------- Page: 4 ----------------------
SIST EN 61010-031:2015/oprA1:2017
61010-031/Ed2/A1/CDV © IEC (E) 3 66/632/CDV
40 Deletion
41 Delete the word “connector” so the sentence will read:
42 “…to connect to a TERMINAL of the equipment or to another probe assembly”
43 3.4.11
44 MEASUREMENT CATEGORY
45 Replacement
46 Replace the text with the following:

47 classification of testing and measuring circuits according to the types of mains to which they

48 are intended to be connected
49 4 Tests
50 4.3.9 Duty cycle
51 Replacement
52 Replace the title with the following:
53 4.3.9 Short-term or intermittent operation
54 4.4.4.2 Temperature
55 Replacement
56 Replace the second paragraph with the following:

57 This temperature is determined by measuring the temperature rise of the surface or part and

58 adding it to the ambient temperature of 40 °C, or to the maximum RATED ambient temperature

59 if higher.
60 4.5.2 Fuses
61 Replacement
62 Replace the fourth sentence of the second paragraph with the following:

63 With respect to prospective short circuit currents associated with mains installations, the fuse

64 shall be RATED according to 12.2 and no additional testing related to the interrupt current

65 RATING is necessary.
66 5 Marking and documentation
67 5.1.5 RATING
68 Replacement
69 Replace the second paragraph by the following:

70 a) Probe assemblies which do not have a RATING for MEASUREMENT CATEGORIES II, III or IV

71 (see 6.5.2) shall be marked with the RATED voltage to earth and with symbol 7 of Table 1

72 (see also 5.4.3 k)).
73 6 Protection against electric shock
74 6.2.2 Examination
75 Replacement
76 Replace figures 6c and 6d by the following figures
---------------------- Page: 5 ----------------------
SIST EN 61010-031:2015/oprA1:2017
61010-031/Ed2/A1/CDV © IEC (E) 4 66/632/CDV

78 Connecting parts are partially mated so as just to make electrical contact while allowing maximum access to the

79 test finger. Note the two possible positions of the testfinger.
80 Figure 6c – Partially mated probe assemblies (see 6.2 and 6.4.2 b))
---------------------- Page: 6 ----------------------
SIST EN 61010-031:2015/oprA1:2017
61010-031/Ed2/A1/CDV © IEC (E) 5 66/632/CDV
82 Key
83 F rigid test finger (see Figure B.1)
84 H potentially HAZARDOUS LIVE part
85 Note the two possible positions of the testfinger
86 Figure 6d – Unmated parts of a probe assembly (see 6.2 and 6.4.2 c))
---------------------- Page: 7 ----------------------
SIST EN 61010-031:2015/oprA1:2017
61010-031/Ed2/A1/CDV © IEC (E) 6 66/632/CDV
87 6.3.4.1 General
88 Figure 8
89 Replacement
90 Replace the list with the following:
91 a) the probe body;
92 b) hand-held or hand-manipulated parts of each CONNECTOR;

93 c) 150 mm  20 mm of the PROBE WIRE or the maximum length of the cable whichever is

94 shorter;
95 d) other hand-held or hand-manipulated parts.
97 Deletion
98 Delete in the key 2c) the following:
99 (see 12.3.2)
100
101 6.3.4.2 Probe assemblies with floating outer connection
102 Figure 10
103 Deletion
104 Delete in the key 2c) the following:
105 (see 12.3.2)
106 6.4.1 General
107 Replacement
108 Replace the list in the fourth paragraph by the following:

109 a) DOUBLE INSULATION, consisting of BASIC INSULATION plus SUPPLEMENTARY INSULATION (see

110 6.4.6)
111 b) BASIC INSULATION plus impedance (see 6.4.4)
112 c) REINFORCED INSULATION (see 6.4.6)
113 d) PROTECTIVE IMPEDANCE (see 6.4.5)
114 6.4.3.2 Protection by a PROTECTIVE FINGERGUARD
115 Replacement
116 Replace the third paragraph by the following:

117 The height of the protective FINGERGUARD from the side where the fingers are intended to be

118 applied shall be at least 2 mm.
119 6.4.3.4 Protection by tactile indicator
120 Replacement
121 Replace the first paragraph by the following:

122 SPRING-LOADED CLIPS RATED for MEASUREMENT CATEGORY II or without MEASUREMENT CATEGORY

123 for maximum 300 V which require finger pressure at about 90° to the axis of the clip are

124 acceptable without a PROTECTIVE FINGERGUARD, provided that there is a tactile indicator to

125 indicate the limit of safe access for the OPERATOR.
126 6.4.3.5 Probe tips used as connectors
127 Replacement
128 Replace the text with the following text:
---------------------- Page: 8 ----------------------
SIST EN 61010-031:2015/oprA1:2017
61010-031/Ed2/A1/CDV © IEC (E) 7 66/632/CDV

129 PROBE TIPS which can be used as CONNECTORS and are intended to be connected to specified

130 accessories (for example to a SPRING-LOADED CLIP) shall, in combination with those

131 accessories, also meet the requirements for CONNECTORS in fully-mated position and partially-

132 mated position (see 6.4.2 a) and b)).
133 6.5.1.2.4 Solid insulation
134 6.5.1.2.4.1 General
135 Replacement
136 Replace the text and the Table 4 by the following:

137 Solid insulation shall withstand the electrical and mechanical stresses that may occur in

138 NORMAL USE, in all RATED environmental conditions (see 1.4), during the intended life of the

139 equipment.

140 The manufacturer should take the expected life of the equipment into account when selecting

141 insulating materials.

142 Conformity is checked by both of the following tests for probe assemblies RATED for use in

143 MEASUREMENT CATEGORIES II, III or IV:

144 a) the a.c. voltage test of 6.6.5.1 with a duration of at least 5 s or the impulse voltage test of

145 6.6.5.3 or, using the applicable test voltage of Table 4 or Table 4a;

146 b) the a.c. voltage test of 6.6.5.1 with a duration of at least 1 min or, for measuring circuits

147 stressed only b
...

Questions, Comments and Discussion