Safety requirements for electrical equipment for measurement, control and laboratory use -- Part 2-081: Particular requirements for automatic and semi-automatic laboratory equipment for analysis and other purposes

Applies to automatic and semi-automatic laboratory equipment for analysis and other purposes. Automatic and semi-automatic laboratory equipment consists of instruments or systems for measuring or modifying characteristics of samples, performing the process without manual intervention. Examples of such equipment are: analytical equipment, automatic sampler (e.g. pipettor), equipment for sample replication and amplification.

Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte -- Teil 2-081: Besondere Anforderungen an automatische und semiautomatische Laborgeräte für Analysen und andere Zwecke

Règles de sécurité pour appareils électriques de mesurage, de régulation et de laboratoire -- Partie 2-081: Prescriptions particulières pour les appareils de laboratoire, automatiques et semi-automatiques, destinés à l'analyse et autres usages

S'applique aux appareils de laboratoire, automatiques et semi-automatiques, destinés à l'analyse et autres usages. Les appareils d'analyse automatiques et semi-automatiques comprennent les appareils ou systèmes utilisés pour mesurer ou modifier des caractéristiques d'échantillons, réalisant le processus sans intervention manuelle. Exemples d'équipements sont: équipement réalisant des analyses, échantillonneur automatique (p.ex. pipeteur), équipement réalisant la réplication ou l'amplification d'échantillon.

Varnostne zahteve za električno opremo za meritve, nadzorovanje in laboratorijsko uporabo - 2-081. del: Posebne zahteve za avtomatsko in polavtomatsko laboratorijsko opremo, ki se uporablja za analizo in druge namene

General Information

Status
Withdrawn
Publication Date
28-Feb-2003
Withdrawal Date
11-Dec-2017
Current Stage
9900 - Withdrawal (Adopted Project)
Start Date
12-Dec-2017
Due Date
04-Jan-2018
Completion Date
12-Dec-2017

Relations

Buy Standard

Standard
EN 61010-2-081:2003
English language
23 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

SLOVENSKI STANDARD
SIST EN 61010-2-081:2003
01-marec-2003
9DUQRVWQH]DKWHYH]DHOHNWULþQRRSUHPR]DPHULWYHQDG]RURYDQMHLQODERUDWRULMVNR
XSRUDERGHO3RVHEQH]DKWHYH]DDYWRPDWVNRLQSRODYWRPDWVNR
ODERUDWRULMVNRRSUHPRNLVHXSRUDEOMD]DDQDOL]RLQGUXJHQDPHQH
Safety requirements for electrical equipment for measurement, control and laboratory
use -- Part 2-081: Particular requirements for automatic and semi-automatic laboratory
equipment for analysis and other purposes
Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte -- Teil
2-081: Besondere Anforderungen an automatische und semiautomatische Laborgeräte
für Analysen und andere Zwecke
Règles de sécurité pour appareils électriques de mesurage, de régulation et de
laboratoire -- Partie 2-081: Prescriptions particulières pour les appareils de laboratoire,
automatiques et semi-automatiques, destinés à l'analyse et autres usages
Ta slovenski standard je istoveten z: EN 61010-2-081:2002
ICS:
19.080 (OHNWULþQRLQHOHNWURQVNR Electrical and electronic
SUHVNXãDQMH testing
71.040.10 Kemijski laboratoriji. Chemical laboratories.
Laboratorijska oprema Laboratory equipment
SIST EN 61010-2-081:2003 en
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------

SIST EN 61010-2-081:2003

---------------------- Page: 2 ----------------------

SIST EN 61010-2-081:2003
EUROPEAN STANDARD EN 61010-2-081
NORME EUROPÉENNE
EUROPÄISCHE NORM June 2002
ICS 19.080; 71.040.10
English version
Safety requirements for electrical equipment for measurement,
control and laboratory use
Part 2-081: Particular requirements for automatic and semi-automatic
laboratory equipment for analysis and other purposes
(IEC 61010-2-081:2001)
Règles de sécurité pour appareils Sicherheitsbestimmungen für elektrische
électriques de mesurage, de régulation Mess-, Steuer-, Regel- und Laborgeräte
et de laboratoire Teil 2-081: Besondere Anforderungen
Partie 2-081: Prescriptions particulières an automatische und semiautomatische
pour les appareils de laboratoire, Laborgeräte für Analysen und andere
automatiques et semi-automatiques, Zwecke
destinés à l'analyse et autres usages (IEC 61010-2-081:2001)
(CEI 61010-2-081:2001)
This European Standard was approved by CENELEC on 2002-03-01. CENELEC members are bound to
comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European
Standard the status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on
application to the Central Secretariat or to any CENELEC member.
This European Standard exists in three official versions (English, French, German). A version in any other
language made by translation under the responsibility of a CENELEC member into its own language and
notified to the Central Secretariat has the same status as the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Czech Republic,
Denmark, Finland, France, Germany, Greece, Iceland, Ireland, Italy, Luxembourg, Malta, Netherlands,
Norway, Portugal, Spain, Sweden, Switzerland and United Kingdom.
CENELEC
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
Central Secretariat: rue de Stassart 35, B - 1050 Brussels
© 2002 CENELEC - All rights of exploitation in any form and by any means reserved worldwide for CENELEC members.
Ref. No. EN 61010-2-081:2002 E

---------------------- Page: 3 ----------------------

SIST EN 61010-2-081:2003
EN 61010-2-081:2002 - 2 -
Foreword
The text of document 66/260/FDIS, future edition 1 of IEC 61010-2-081, prepared by IEC TC 66,
Safety of measuring, control, and laboratory equipment, was submitted to the IEC-CENELEC parallel
vote and was approved by CENELEC as EN 61010-2-081 on 2002-03-01.
The following dates were fixed:
– latest date by which the EN has to be implemented
at national level by publication of an identical
national standard or by endorsement (dop) 2002-12-01
– latest date by which the national standards conflicting
with the EN have to be withdrawn (dow) 2005-03-01
This Part 2-081 is to be used in conjunction with EN 61010-1:2001, Safety requirements for electrical
equipment for measurement, control and laboratory use -- Part 1: General requirements.
This Part 2-081 supplements or modifies the corresponding clauses of EN 61010-1 so as to convert it
into the European Standard: Safety requirements for automatic and semi-automatic laboratory
equipment for analysis and other purposes.
Where a particular clause or subclause of Part 1 is not mentioned in this Part 2-081, that clause or
subclause applies as far as is reasonable. Where this Part 2-081 states "addition", "modification" or
"replacement", the relevant text of Part 1 is to be adapted accordingly.
In this standard:
1) the following print types are used:
- requirements: in roman type;
- NOTES: in smaller roman type;
- conformity and test: in italic type;
- terms used throughout this standard which have been defined in clause 3: SMALL ROMAN
CAPITALS.
2) subclauses or figures which are additional to those in Part 1 are numbered starting from 101,
additional annexes are lettered AA, BB, etc.
Annexes designated "normative" are part of the body of the standard.
Annexes designated "informative" are given for information only.
In this standard, annexes AA, BB and ZA are normative.
Annex ZA has been added by CENELEC.
__________

---------------------- Page: 4 ----------------------

SIST EN 61010-2-081:2003
- 3 - «Field32»
Endorsement notice
The text of the International Standard IEC 61010-2-081:2001 was approved by CENELEC as a
European Standard without any modification.
Editorial modification to IEC 61010-2-81:
Replace figure AA.1 by the following (from EN ISO 14971:2000):
Intolerable
region
ALARP
region
Broadly acceptable
region
Increasing SEVERITY of HARM
Figure AA.1 – RISK acceptability
__________
Increasing probability of occurrence

---------------------- Page: 5 ----------------------

SIST EN 61010-2-081:2003
EN 61010-2-081:2002 - 4 -
Annex ZA
(normative)
Normative references to international publications
with their corresponding European publications
Addition to annex ZA of EN 61010-1:
Publication Year Title EN/HD Year
IEC 60068-2-75 1997 Environmental testing – EN 60068-2-75 1997
Part 2-75: Tests – Test Eh: Hammer tests

---------------------- Page: 6 ----------------------

SIST EN 61010-2-081:2003
NORME CEI
INTERNATIONALE IEC
61010-2-081
INTERNATIONAL
Première édition
STANDARD
First edition
2001-12
PUBLICATION GROUPÉE DE SÉCURITÉ
GROUP SAFETY PUBLICATION
Règles de sécurité pour appareils électriques
de mesurage, de régulation et de laboratoire –
Partie 2-081:
Prescriptions particulières pour les appareils de
laboratoire, automatiques et semi-automatiques,
destinés à l'analyse et autres usages
Safety requirements for electrical equipment
for measurement, control, and laboratory use –
Part 2-081:
Particular requirements for automatic and semi-
automatic laboratory equipment for analysis
and other purposes
 IEC 2001 Droits de reproduction réservés  Copyright - all rights reserved
Aucune partie de cette publication ne peut être reproduite ni No part of this publication may be reproduced or utilized in
utilisée sous quelque forme que ce soit et par aucun procédé, any form or by any means, electronic or mechanical,
électronique ou mécanique, y compris la photocopie et les including photocopying and microfilm, without permission in
microfilms, sans l'accord écrit de l'éditeur. writing from the publisher.
International Electrotechnical Commission 3, rue de Varembé Geneva, Switzerland
Telefax: +41 22 919 0300 e-mail: inmail@iec.ch IEC web site http://www.iec.ch
CODE PRIX
Commission Electrotechnique Internationale
R
PRICE CODE
International Electrotechnical Commission
Pour prix, voir catalogue en vigueur
For price, see current catalogue

---------------------- Page: 7 ----------------------

SIST EN 61010-2-081:2003
61010-2-081  IEC:2001 – 3 –
CONTENTS
FOREWORD.5
1 Scope and object.9
2 Normative references .11
3 Terms and definitions .11
4 Tests .13
5 Marking and documentation.13
6 Protection against electric shock .21
7 Protection against mechanical HAZARDS.21
8 Mechanical resistance to shock and impact .23
9 Protection against the spread of fire .23
10 Equipment temperature limits and resistance to heat.23
11 Protection against HAZARDS from fluids .25
12 Protection against radiation, including laser sources, and against sonic and
ultrasonic pressure .25
13 Protection against liberated gases, explosion and implosion.25
14 Components .27
15 Protection by interlocks .27
16 Test and measurement equipment.27
Annex H Index of defined terms.29
Annex AA (normative) Risk management .29
Annex BB (normative) Impact spring hammer (see 8.2) .33
Bibliography.37

---------------------- Page: 8 ----------------------

SIST EN 61010-2-081:2003
61010-2-081  IEC:2001 – 5 –
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
SAFETY REQUIREMENTS FOR ELECTRICAL EQUIPMENT FOR
MEASUREMENT, CONTROL, AND LABORATORY USE –
Part 2-081: Particular requirements for automatic and
semi-automatic laboratory equipment for analysis and other purposes
FOREWORD
1) The IEC (International Electrotechnical Commission) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of the IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, the IEC publishes International Standards. Their preparation is
entrusted to technical committees; any IEC National Committee interested in the subject dealt with may
participate in this preparatory work. International, governmental and non-governmental organizations liaising
with the IEC also participate in this preparation. The IEC collaborates closely with the International
Organization for Standardization (ISO) in accordance with conditions determined by agreement between the
two organizations.
2) The formal decisions or agreements of the IEC on technical matters express, as nearly as possible, an
international consensus of opinion on the relevant subjects since each technical committee has representation
from all interested National Committees.
3) The documents produced have the form of recommendations for international use and are published in the form
of standards, technical specifications, technical reports or guides and they are accepted by the National
Committees in that sense.
4) In order to promote international unification, IEC National Committees undertake to apply IEC International
Standards transparently to the maximum extent possible in their national and regional standards. Any
divergence between the IEC Standard and the corresponding national or regional standard shall be clearly
indicated in the latter.
5) The IEC provides no marking procedure to indicate its approval and cannot be rendered responsible for any
equipment declared to be in conformity with one of its standards.
6) Attention is drawn to the possibility that some of the elements of this International Standard may be the subject
of patent rights. The IEC shall not be held responsible for identifying any or all such patent rights.
International Standard IEC 61010-2-081 has been prepared by IEC technical committee 66:
Safety of measuring, control and laboratory equipment.
It has the status of a group safety publication in accordance with IEC Guide 104.
The text of this standard is based on the following documents:
FDIS Report on voting
66/260/FDIS 66/269/RVD
Full information on the voting for the approval of this standard can be found in the report on
voting indicated in the above table.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 3.
This part 2 is intended to be used in conjunction with IEC 61010-1. It was established on the
basis of the second edition (2001). Consideration may be given to future editions of, or
amendments to, IEC 61010-1.
This part 2 supplements or modifies the corresponding clauses in IEC 61010-1 so as to
convert that publication into the IEC standard: Safety requirements for automatic and semi-
automatic laboratory equipment for analysis and other purposes

---------------------- Page: 9 ----------------------

SIST EN 61010-2-081:2003
61010-2-081  IEC:2001 – 7 –
Where a particular subclause of part 1 is not mentioned in this part 2, that subclause applies
as far as is reasonable. Where this part states “addition”, “modification”, “replacement”, or
“deletion”, the relevant requirement, test specification or note in part 1 should be adapted
accordingly.
In this standard:
1) the following print types are used:
• requirements: in roman type;
• NOTES: in smaller roman type;
• conformity and test: in italic type;
• terms used throughout this standard which have been defined in clause 3: SMALL
ROMAN CAPITALS.
2) subclauses or figures which are additional to those in part 1 are numbered starting
from 101; additional annexes are lettered starting from AA.
Annexes AA and BB form an integral part of this standard.
The committee has decided that the contents of this publication will remain unchanged until
2004. At this date, the publication will be
• reconfirmed;
• withdrawn;
• replaced by a revised edition, or
• amended.

---------------------- Page: 10 ----------------------

SIST EN 61010-2-081:2003
61010-2-081  IEC:2001 – 9 –
SAFETY REQUIREMENTS FOR ELECTRICAL EQUIPMENT FOR
MEASUREMENT, CONTROL, AND LABORATORY USE –
Part 2-081: Particular requirements for automatic and
semi-automatic laboratory equipment for analysis and other purposes
1 Scope and object
This clause of part 1 is applicable except as follows:
1.1 Scope
Replacement:
This part 2 applies to automatic and semi-automatic laboratory equipment for analysis and
other purposes.
Automatic and semi-automatic laboratory equipment consists of instruments or systems for
measuring or modifying one or more characteristics or parameters of samples, performing the
complete process or parts of the process without manual intervention. Equipment forming part
of such a system is within the scope of this standard.
Examples of equipment within the scope of this standard include:
• analytical equipment;
• automatic sampler (pipettor, aliquoter);
• equipment for sample replication and amplification.
NOTE 1 In the case of analytical equipment the complete process usually includes the following steps:
• taking a specific quantity of the sample;
• preparing the sample by chemical, thermal, mechanical or other means;
• measurement;
• display, transmission or printing of the results of measurement.
NOTE 2 If all or part of the equipment falls within the scope of one or more other part 2 standards of IEC 61010
as well as within the scope of this standard, it will also need to meet the requirements of those other parts 2
standards.
1.2 Object
1.2.1 Aspects included in scope
Replacement:
Replace the first sentence by the following:
The purpose of this standard is to ensure that the design and methods of construction used
provide a high degree of protection at a TOLERABLE RISK for the OPERATOR and the surrounding
area, using RISK management where specified (see 7.2.101 and annex AA).

---------------------- Page: 11 ----------------------

SIST EN 61010-2-081:2003
61010-2-081  IEC:2001 – 11 –
Addition:
Add two new items:
a) biohazards;
b) hazardous chemical substances.
1.2.2 Aspects excluded from scope
Addition:
Add a new item and the following note:
g) handling or manipulation of material outside the equipment.
NOTE Requirements covering these subjects are the responsibility of committees preparing the relevant standards.
2 Normative references
This clause of part 1 is applicable except as follows:
Addition:
IEC 60068-2-75:1997, Environmental testing – Part 2-75: Test – Test Eh: Hammer tests
3 Terms and definitions
This clause of part 1 is applicable except as follows:
3.1 Equipment and states of equipment
Additions:
Additional definitions:
3.101
HARM
physical injury or damage to the health of people, or damage to property or the environment
[ISO/IEC Guide 51:1999, definition 3.3]
3.102
RISK
combination of the probability of occurrence of HARM and the severity of that HARM
[ISO/IEC Guide 51:1999, definition 3.2]
3.103
TOLERABLE RISK
RISK which is accepted in a given context based on the current values of society
[ISO/IEC Guide 51:1999, definition 3.7]
NOTE TOLERABLE RISK is the result of a balance between the ideal of absolute safety, the demands to be met by a
product, process or service, and factors such as benefit to the user, suitability of purpose, cost effectiveness, RISK
evaluation, conventions of the society concerned, and the state of the art.

---------------------- Page: 12 ----------------------

SIST EN 61010-2-081:2003
61010-2-081  IEC:2001 – 13 –
4 Tests
This clause of part 1 is applicable except as follows:
4.4.2 Application of fault conditions
Addition:
Additional subclause:
4.4.2.101 Incorrect voltage selection
Multivoltage equipment that can be set by the OPERATOR to different supply voltages shall be
set to each voltage in turn and then connected to all other RATED supply voltages in turn.
5 Marking and documentation
This clause of part 1 is applicable except as follows:
5.1.1 General
Replacement:
Replace the third paragraph by the following:
Letter symbols for quantities and units shall be in accordance with IEC 60027. Internationally
recognized symbols, including those of Table 1, shall be used as far as possible. If other
additional symbols are required, it shall not be possible to confuse them with the international
symbols. There are no colour requirements for symbols except for symbol 101 (see Table 1).
Graphic
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.