Packaging for terminally sterilized medical devices - Part 8: Re-usable sterilization containers for steam sterilizers conforming to EN 285 - Requirements and test methods

This document specifies test methods and values for re-usable containers used as sterile barrier systems that are intended to maintain sterility of terminally sterilized medical devices to the point of use. These containers are intended to be used in steam sterilizers conforming to EN 285.
NOTE 1   The need for a packaging material inside the container is determined by the manufacturers and users.
Other than the general requirements as specified in EN ISO 11607 1 and EN ISO 11607 2 this part of EN 868 specifies materials, test methods and values that are specific to the products covered by this document.
NOTE 2   When it is intended to use the containers in a steam sterilizer not conforming to EN 285 the sterilization performance of the container in the specific sterilization cycle to be used is validated by the user. Other attributes of the container are also reviewed for compatibility with the sterilizer cycle, e.g. operating temperature.
NOTE 3   The use of additional materials and/or accessories inside the sterile barrier system in order to ease the organization, drying or aseptic presentation (e.g. inner wrap, indicators, packing lists, mats, instrument organizer sets, tray liners or an additional envelope around the medical device) is not covered in this part of EN 868. However, other requirements, including the determination of the acceptability of these materials and/or accessories during validation activities, can apply.

Verpackungen für in der Endverpackung zu sterilisierende Medizinprodukte - Teil 8: Wiederverwendbare Sterilisierbehälter für Dampf-Sterilisatoren nach EN 285 - Anforderungen und Prüfverfahren

Dieses Dokument legt Prüfverfahren und Werte für wiederverwendbare Sterilisierbehälter fest, die dazu vorgesehen sind, als Sterilbarrieresysteme die Sterilität von in der Endverpackung sterilisierten Medizinpro¬dukten bis zum Anwendungszeitpunkt zu erhalten. Diese Sterilisierbehälter sind für Dampf-Sterilisatoren nach EN 285 vorgesehen.
ANMERKUNG 1   Das Erfordernis für eine zusätzliche Innenverpackung kann durch den Hersteller und den Anwender bestimmt werden.
Im Gegensatz zu den allgemeinen Anforderungen, welche durch EN ISO 11607 1 und EN ISO 11607 2 festge¬legt sind, legt dieser Teil der EN 868 Materialien, Prüfverfahren und Werte fest, welche spezifisch für die in diesem Dokument betrachteten Produkte sind.
ANMERKUNG 2   Wenn der Sterilisierbehälter in einem Dampf Sterilisator verwendet werden soll, der nicht EN 285 entspricht, so wird das Sterilisierverhalten des Sterilisierbehälters für den anzuwendenden spezifischen Sterilisa¬tionszyklus durch den Anwender validiert. Andere Merkmale des Sterilisierbehälters werden hinsichtlich der Kompa¬tibilität mit dem Sterilisationszyklus erneut überprüft, z. B. die Arbeitstemperatur.
ANMERKUNG 3   Die Verwendung von zusätzlichen Materialien und/oder von Zubehör innerhalb des Sterilbarrieresystems, um die Organisation, die Trocknung oder die aseptische Präsentation zu erleichtern (z. B. Innenver¬packung, Indikatoren, Verpackungslisten, Unterlagen, Einsätze für Instrumente, Fächer oder eine zusätzliche Umhüllung für das Medizinprodukt) ist nicht in diesem Teil der EN 868 abgedeckt. Dennoch können andere Anforderungen, einschließlich der Bestimmung der Eignung dieser Materialien und/oder Zubehör im Rahmen von Validierungsstudien gelten.

Emballages des dispositifs médicaux stérilisés au stade terminal - Partie 8: Conteneurs réutilisables de stérilisation pour stérilisateurs à la vapeur d’eau conformes à l’EN 285 - Exigences et méthodes d’essai

Le présent document spécifie des méthodes d’essai et des valeurs relatives aux conteneurs réutilisables employés comme systèmes de barrière stérile destinés à maintenir l’état de stérilité des dispositifs médicaux stérilisés au stade terminal jusqu’au point d’utilisation. Ces conteneurs sont destinés à être utilisés dans les stérilisateurs à la vapeur d’eau, conformément à l’EN 285.
NOTE 1   La nécessité d’un matériau d’emballage à l’intérieur du conteneur est déterminée par les fabricants et les utilisateurs.
Outre les exigences générales telles que spécifiées dans l’EN ISO 11607 1 et l’EN ISO 11607 2, la présente partie de l’EN 868 spécifie les matériaux, les méthodes d’essai et les valeurs spécifiques aux produits traités dans le présent document.
NOTE 2   Lorsqu’il est prévu d’utiliser les conteneurs dans un stérilisateur à la vapeur non conforme à l’EN 285, les performances de stérilisation du conteneur dans le cycle de stérilisation spécifique à utiliser sont validées par l’utilisateur. D’autres caractéristiques du conteneur sont également examinées pour contrôler la compatibilité avec le cycle du stérilisateur, par exemple, la température de consigne.
NOTE 3   L’utilisation d’accessoires et/ou de matériaux supplémentaires à l’intérieur du système de barrière stérile de manière à faciliter le rangement, le séchage ou la présentation aseptique (par exemple, enveloppe interne, indicateurs, listes d’emballage, tapis, dispositifs de rangement des instruments, fonds de paniers ou une enveloppe supplémentaire entourant le dispositif médical), n’est pas couverte par la présente partie de l’EN 868. Toutefois, d’autres exigences, notamment concernant la détermination de l’acceptabilité de ces matériaux et/ou accessoires pendant les activités de validation, peuvent s’appliquer.

Embalaža za končno sterilizirane medicinske pripomočke - 8. del: Ponovno uporabljivi vsebniki za parne sterilizatorje po EN 285 - Zahteve in preskusne metode

Ta del standarda EN 868 podaja preskusne metode in vrednosti za vrečke in zvitke iz poroznega materiala z možnostjo tesnjenja, izdelane v skladu s standardom EN 868, deli 2, 3, 6, 7, 9 ali 10, in plastične folije v skladu s točko 4, ki se uporabljajo kot sterilni pregradni sistemi in/ali sistemi embalaže, ki so namenjeni za vzdrževanje sterilnosti končno steriliziranih medicinskih pripomočkov do njihove uporabe.
OPOMBA 1: Potrebo po zaščitni embalaži lahko določita proizvajalec in uporabnik. Ta del standarda EN 868 le uvaja zahteve delovanja in preskusne metode, ki veljajo za proizvode, zajete v tem delu standarda EN 868, vendar ne dodaja ali spreminja splošnih zahtev, podanih v standardu EN ISO 11607-1.
Posebne zahteve iz točk od 4.2 do 4.5 je mogoče uporabiti za dokazovanje skladnosti z eno ali več zahtevami, vendar ne z vsemi zahtevami standarda EN ISO 11607-1.
Materiali, določeni v tem delu standarda ISO 868, so namenjeni le enkratni uporabi.
OPOMBA 2: Ko se znotraj sterilnega pregradnega sistema uporabljajo dodatni materiali za lažjo organizacijo, sušenje ali aseptičnost (npr. notranji ovoj, kazalniki filtrov, embalažni listi, podloge, kompleti organizatorjev za inštrumente, obloge za pladnje ali dodatni ovoj okoli medicinskega pripomočka), se lahko uporabljajo druge zahteve, vključno z ugotavljanjem sprejemljivosti teh materialov med preverjanjem.

General Information

Status
Published
Public Enquiry End Date
14-Oct-2017
Publication Date
03-Feb-2019
Technical Committee
VAZ - Healthcare
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
15-Jan-2019
Due Date
22-Mar-2019
Completion Date
04-Feb-2019

Relations

Effective Date
01-Mar-2019

Overview

EN 868-8:2018 - “Packaging for terminally sterilized medical devices - Part 8” specifies requirements and test methods for re-usable sterilization containers used as sterile barrier systems in steam sterilizers conforming to EN 285. The standard focuses on construction, performance and validation-related tests that demonstrate a container’s ability to maintain sterility of terminally sterilized medical devices until point of use. It complements the general packaging requirements given in EN ISO 11607‑1/2.

Key topics and technical requirements

  • Scope and purpose
    • Applies to re-usable containers intended for use in steam sterilizers conforming to EN 285.
    • If using non‑EN 285 sterilizers, users must validate sterilization performance for the specific cycle.
  • Design and construction
    • Containers are generally parallelepipedal (box‑like); internal corners should be radiused for cleaning.
    • Lids with latch mechanisms and closure gaskets to provide microbial barrier properties.
    • Tamper‑evident closure system required; if reusable, it must require a special tool or code to reset.
    • Carrying devices are required unless container can be safely hand‑carried.
  • Performance and testing (normative annexes)
    • Carrying device strength test (Annex C) - verify load support without permanent deformation.
    • Stacking tests and stacking device capability (Annexes D, E) - structural integrity under load.
    • Sterilization performance validation (Annex F) - demonstrate container allows effective steam sterilization.
    • Load dryness tests (Annex G) - ensure drying sufficient to maintain sterility and package integrity.
    • Guidance on service life and sterilization effects (Annex H).
  • Information and maintenance
    • Manufacturer must supply maintenance and cleaning instructions (e.g., gasket access, operating temperature limits).
  • Exclusions
    • Use of inner wraps, indicators, instrument organizers and other accessories inside the sterile barrier system is not covered, but may require validation.

Practical applications and users

Who uses EN 868‑8:

  • Manufacturers of re‑usable sterilization containers and sterilization equipment
  • Hospital sterile services departments (CSSD) and infection control teams evaluating container systems
  • Validation and regulatory professionals responsible for sterilization process qualification and compliance
  • Design engineers developing container gaskets, latches and carrying systems

Practical value:

  • Provides clear test methods and acceptance criteria to confirm containers maintain sterility during sterilization, handling and storage.
  • Helps align product design and documentation with European regulatory and procurement expectations.

Related standards

  • EN 285 - Steam sterilizers - Large sterilizers (required compatibility reference)
  • EN ISO 11607‑1:2017 & EN ISO 11607‑2:2017 - General requirements and validation for packaging for terminally sterilized medical devices
  • Other parts of the EN 868 series (e.g., wraps, pouches, papers) for complementary sterile barrier materials

Keywords: re-usable sterilization containers, steam sterilizers, EN 285, sterile barrier systems, EN 868-8, EN ISO 11607, packaging for terminally sterilized medical devices, test methods, sterilization validation.

Frequently Asked Questions

SIST EN 868-8:2019 is a standard published by the Slovenian Institute for Standardization (SIST). Its full title is "Packaging for terminally sterilized medical devices - Part 8: Re-usable sterilization containers for steam sterilizers conforming to EN 285 - Requirements and test methods". This standard covers: This document specifies test methods and values for re-usable containers used as sterile barrier systems that are intended to maintain sterility of terminally sterilized medical devices to the point of use. These containers are intended to be used in steam sterilizers conforming to EN 285. NOTE 1 The need for a packaging material inside the container is determined by the manufacturers and users. Other than the general requirements as specified in EN ISO 11607 1 and EN ISO 11607 2 this part of EN 868 specifies materials, test methods and values that are specific to the products covered by this document. NOTE 2 When it is intended to use the containers in a steam sterilizer not conforming to EN 285 the sterilization performance of the container in the specific sterilization cycle to be used is validated by the user. Other attributes of the container are also reviewed for compatibility with the sterilizer cycle, e.g. operating temperature. NOTE 3 The use of additional materials and/or accessories inside the sterile barrier system in order to ease the organization, drying or aseptic presentation (e.g. inner wrap, indicators, packing lists, mats, instrument organizer sets, tray liners or an additional envelope around the medical device) is not covered in this part of EN 868. However, other requirements, including the determination of the acceptability of these materials and/or accessories during validation activities, can apply.

This document specifies test methods and values for re-usable containers used as sterile barrier systems that are intended to maintain sterility of terminally sterilized medical devices to the point of use. These containers are intended to be used in steam sterilizers conforming to EN 285. NOTE 1 The need for a packaging material inside the container is determined by the manufacturers and users. Other than the general requirements as specified in EN ISO 11607 1 and EN ISO 11607 2 this part of EN 868 specifies materials, test methods and values that are specific to the products covered by this document. NOTE 2 When it is intended to use the containers in a steam sterilizer not conforming to EN 285 the sterilization performance of the container in the specific sterilization cycle to be used is validated by the user. Other attributes of the container are also reviewed for compatibility with the sterilizer cycle, e.g. operating temperature. NOTE 3 The use of additional materials and/or accessories inside the sterile barrier system in order to ease the organization, drying or aseptic presentation (e.g. inner wrap, indicators, packing lists, mats, instrument organizer sets, tray liners or an additional envelope around the medical device) is not covered in this part of EN 868. However, other requirements, including the determination of the acceptability of these materials and/or accessories during validation activities, can apply.

SIST EN 868-8:2019 is classified under the following ICS (International Classification for Standards) categories: 11.080.30 - Sterilized packaging. The ICS classification helps identify the subject area and facilitates finding related standards.

SIST EN 868-8:2019 has the following relationships with other standards: It is inter standard links to SIST EN 868-8:2009. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

You can purchase SIST EN 868-8:2019 directly from iTeh Standards. The document is available in PDF format and is delivered instantly after payment. Add the standard to your cart and complete the secure checkout process. iTeh Standards is an authorized distributor of SIST standards.

Standards Content (Sample)


SLOVENSKI STANDARD
01-marec-2019
1DGRPHãþD
SIST EN 868-8:2009
(PEDODåD]DNRQþQRVWHULOL]LUDQHPHGLFLQVNHSULSRPRþNHGHO3RQRYQR
XSRUDEOMLYLYVHEQLNL]DSDUQHVWHULOL]DWRUMHSR(1=DKWHYHLQSUHVNXVQH
PHWRGH
Packaging for terminally sterilized medical devices - Part 8: Re-usable sterilization
containers for steam sterilizers conforming to EN 285 - Requirements and test methods
Verpackungen für in der Endverpackung zu sterilisierende Medizinprodukte - Teil 8:
Wiederverwendbare Sterilisierbehälter für Dampf-Sterilisatoren nach EN 285 -
Anforderungen und Prüfverfahren
Emballages des dispositifs médicaux stérilisés au stade terminal - Partie 8: Conteneurs
réutilisables de stérilisation pour stérilisateurs à la vapeur d’eau conformes à l’EN 285 -
Exigences et méthodes d’essai
Ta slovenski standard je istoveten z: EN 868-8:2018
ICS:
11.080.30 Sterilizirana embalaža Sterilized packaging
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

EN 868-8
EUROPEAN STANDARD
NORME EUROPÉENNE
December 2018
EUROPÄISCHE NORM
ICS 11.080.30 Supersedes EN 868-8:2009
English Version
Packaging for terminally sterilized medical devices - Part
8: Re-usable sterilization containers for steam sterilizers
conforming to EN 285 - Requirements and test methods
Emballages des dispositifs médicaux stérilisés au stade Verpackungen für in der Endverpackung zu
terminal - Partie 8: Conteneurs réutilisables de sterilisierende Medizinprodukte - Teil 8:
stérilisation pour stérilisateurs à la vapeur d'eau Wiederverwendbare Sterilisierbehälter für Dampf-
conformes à l'EN 285 - Exigences et méthodes d'essai Sterilisatoren nach EN 285 - Anforderungen und
Prüfverfahren
This European Standard was approved by CEN on 20 August 2018.

CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this
European Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references
concerning such national standards may be obtained on application to the CEN-CENELEC Management Centre or to any CEN
member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by
translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC Management
Centre has the same status as the official versions.

CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania,
Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and United Kingdom.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION

EUROPÄISCHES KOMITEE FÜR NORMUNG

CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2018 CEN All rights of exploitation in any form and by any means reserved Ref. No. EN 868-8:2018 E
worldwide for CEN national Members.

Contents Page
European foreword . 4
Introduction . 6
1 Scope . 7
2 Normative references . 7
3 Terms and definitions . 7
4 Requirements . 8
5 Information to be supplied by the manufacturer . 11
Annex A (informative) Details of significant technical changes between this document and
the previous edition . 12
Annex B (informative) Guidance on dimensions . 13
Annex C (normative) Carrying device strength test . 14
C.1 Procedure. 14
C.2 Result . 14
Annex D (normative) Stacking test - Procedure . 15
Annex E (normative) Stacking device capability test . 16
E.1 Apparatus . 16
E.2 Procedure. 16
Annex F (normative) Determination of sterilization performance . 17
F.1 Apparatus . 17
F.2 Procedure. 17
F.3 Results . 17
Annex G (normative) Load dryness tests . 18
G.1 Apparatus . 18
G.2 Loads . 18
G.2.1 Metal . 18
G.2.2 Textile . 18
G.3 Procedure. 18
G.3.1 Metal Load. 18
G.3.2 Textile load . 18
G.4 Results . 19
G.4.1 Metal load . 19
G.4.2 Textile load . 19
Annex H (informative) Guidance on determination of service life with respect to
sterilization . 20
H.1 Test specimen . 20
H.2 Apparatus . 20
Bibliography . 21

European foreword
This document (EN 868-8:2018) has been prepared by Technical Committee CEN/TC 102 “Sterilizers
and associated equipment for processing of medical devices”, the secretariat of which is held by DIN.
This European Standard shall be given the status of a national standard, either by publication of an
identical text or by endorsement, at the latest by June 2019, and conflicting national standards shall be
withdrawn at the latest by June 2019.
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CEN shall not be held responsible for identifying any or all such patent rights.
This document supersedes EN 886-8:2009.
Annex A provides details of significant technical changes between this European Standard and the
previous edition.
EN 868 consists of the following parts, under the general title Packaging for terminally sterilized medical
devices:
— Part 2: Sterilization wrap — Requirements and test methods;
— Part 3: Paper for use in the manufacture of paper bags (specified in EN 868-4) and in the manufacture
of pouches and reels (specified in EN 868-5) — Requirements and test methods;
— Part 4: Paper bags — Requirements and test methods;
— Part 5: Sealable pouches and reels of porous materials and plastic film construction — Requirements
and test methods;
— Part 6: Paper for low temperature sterilization processes — Requirements and test methods;
— Part 7: Adhesive coated paper for low temperature sterilization processes — Requirements and test
methods;
— Part 8: Re-usable sterilization containers for steam sterilizers conforming to EN 285 — Requirements
and test methods;
— Part 9: Uncoated nonwoven materials of polyolefines — Requirements and test methods;
— Part 10: Adhesive coated nonwoven materials of polyolefines — Requirements and test methods.
In addition, ISO/TC 198 “Sterilization of health care products” in collaboration with CEN/TC 102
“Sterilizers and associated equipment for processing of medical devices” has prepared the
EN ISO 11607 series “Packaging for terminally sterilized medical devices”. The EN ISO 11607 series
specifies general requirements for materials, sterile barrier systems and packaging systems (Part 1)
and validation requirements for forming, sealing and assembly processes (Part 2).
According to the CEN-CENELEC Internal Regulations, the national standards organisations of the
following countries are bound to implement this European Standard: Austria, Belgium, Bulgaria,
Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia,
France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta,
Netherlands, Norway, Poland, Portugal, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and the United Kingdom.
Introduction
The EN ISO 11607 series consists of two parts under the general title “Packaging for terminally
sterilized medical devices”. Part 1 of this series specifies general requirements and test methods for
materials, preformed sterile barrier systems, sterile barrier systems and packaging systems that are
intended to maintain sterility of terminally sterilized medical devices to the point of use. Part 2 of this
series specifies validation requirements for forming, sealing and assembly processes.
General requirements for all types of sterile barrier systems are provided by EN ISO 11607-1.
The EN 868 series can be used to demonstrate compliance with one or more of the requirements
specified in EN ISO 11607-1.
CEN/TC 102/WG 4 also appreciates the initiatives of CEN with regard to the minimization of adverse
environmental impacts by standards. It was agreed that this subject should be given priority during the
next edition of the EN ISO 11607 series that is the basic reference for all parts of the EN 868 series.
1 Scope
This document specifies test methods and values for re-usable containers used as sterile barrier
systems that are intended to maintain sterility of terminally sterilized medical devices to the point of
use. These containers are intended to be used in steam sterilizers conforming to EN 285.
NOTE 1 The need for a packaging material inside the container is determined by the manufacturers and users.
Other than the general requirements as specified in EN ISO 11607-1 and EN ISO 11607-2 this part of
EN 868 specifies materials, test methods and values that are specific to the products covered by this
document.
NOTE 2 When it is intended to use the containers in a steam sterilizer not conforming to EN 285 the
sterilization performance of the container in the specific sterilization cycle to be used is validated by the user.
Other attributes of the container are also reviewed for compatibility with the sterilizer cycle, e.g. operating
temperature.
NOTE 3 The use of additional materials and/or accessories inside the sterile barrier system in order to ease the
organization, drying or aseptic presentation (e.g. inner wrap, indicators, packing lists, mats, instrument organizer
sets, tray liners or an additional envelope around the medical device) is not covered in this part of EN 868.
However, other requirements, including the determination of the acceptability of these materials and/or
accessories during validation activities, can apply.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments) applies.
EN 285:2015, Sterilization — Steam sterilizers — Large sterilizers
EN 10088-1, Stainless steels — Part 1: List of stainless steels
EN ISO 4017:2014, Fasteners — Hexagon head screws — Product grades A and B (ISO 4017:2014)
EN ISO 11607-1:2017, Packaging for terminally sterilized medical devices — Part 1: Requirements for
materials, sterile barrier systems and packaging systems (ISO 11607-1:2006, including Amd 1:2014)
ISO 4582, Plastics — Determination of changes in colour and variations in properties after exposure to
glass-filtered solar radiation, natural weathering or laboratory radiation sources
3 Terms and definitions
For the purposes of this document, the terms and definitions given in EN ISO 11607-1:2017 and
EN 285:2015 apply.
ISO and IEC maintain terminological databases for use in standardization at the following addresses:
— IEC Electropedia: available at http://www.electropedia.org/
— ISO Online browsing platform: available at http://www.iso.org/obp
4 Requirements
4.1 General
For any preformed sterile barrier system or sterile barrier system, the requirements of EN ISO 11607-1
shall apply.
This part of EN 868 only introduces performance requirements and test methods that are specific to the
products covered by this part of EN 868 but does not add or modify the general requirements specified
in EN ISO 11607-1.
As such, the particular requirements in 4.2, 4.3 and 4.4 can be used to demonstrate compliance with one
or more but not all of the requirements of EN ISO 11607-1.
NOTE 1 Compliance to EN 868-8 does not automatically mean compliance to EN ISO 11607-1.
A confirmation of compliance to EN 868-8 shall contain a statement whether EN ISO 11607-1 is
covered.
4.2 Requirements for construction and design
4.2.1 Shape and dimension
4.2.1.1 The container shall be in the general form of a parallelepipedal box.
NOTE Slight curvature or camber of the flat surfaces can be acceptable. Rounding of the corners is desirable.
4.2.1.2 The container including all connected parts, e.g. carrying devices shall fit within one
sterilization module (see EN 285).
NOTE 1 If the container does not fit within one sterilization module, but complies with all other requirements
of this part of the standard, the manufacturer can claim compliance with EN ISO 11607-1, but not with EN 868-8.
NOTE
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...

記事のタイトル:SIST EN 868-8:2019-医療用終末殺菌デバイス用パッケージング-第8部:EN 285に対応するスチーム殺菌器用再利用可能殺菌容器-要件および試験方法 記事の内容:このEN 868の一部は、使用されるまで医療用終末殺菌デバイスの殺菌を保持するために使用される再利用可能な容器である殺菌バリアシステムとしての試験方法と値を提供します。これらの容器はEN 285に対応するスチーム殺菌器で使用することを目的としています。 注1:容器内にパッケージング材料が必要かどうかは、製造業者と使用者によって決定されます。 このEN 868の一部は、EN ISO 11607-1で指定された一般的な要件を追加または修正せず、このEN 868の一部でカバーされる製品に固有の要件と試験方法のみを紹介します。 したがって、4.2から4.5の特定の要件はEN ISO 11607-1の要件の一部またはすべてを満たすことを示すために使用できます。 注2:EN 285に対応していないスチーム殺菌器で容器を使用する場合、使用者は特定の殺菌サイクルでの容器の殺菌性能を検証する必要があります。また、容器の他の属性(例:作動温度)も殺菌器のサイクルとの互換性を確認する必要があります。 注3:組織化、乾燥、または無菌な提示を容易にするために殺菌バリアシステム内に他の材料(例:内部包装、容器フィルター、指示器、梱包リスト、マット、器具整理セット、トレイライナー、または医療機器の周囲の追加の封筒)を使用する場合、追加の要件が適用される場合があります。それらの材料の妥当性を検証活動中に確認する必要もあります。

記事タイトル:SIST EN 868-8:2019 - 殺菌終末製品の包装 - 第8部:EN 285に準拠する蒸気滅菌器用の再利用可能な滅菌容器-要件と試験方法 記事内容:このEN 868の一部は、使用時点まで殺菌終末製品の無菌性を維持するために使用される再利用可能な容器の試験方法および値を提供しています。これらの容器はEN 285に準拠する蒸気滅菌器で使用することを意図しています。 注意1)容器内に包装材料が必要かどうかは、製造業者および使用者によって決定されます。 EN 868のこの部分は、この部分でカバーされる製品に特化した性能要件と試験方法を紹介するだけであり、EN ISO 11607-1で指定された一般的な要件を追加または変更しません。 したがって、4.2から4.5の特定の要件は、EN ISO 11607-1のすべての要件を満たすわけではありませんが、1つ以上の要件の遵守を示すために使用することができます。 注意2)EN 285に準拠しない蒸気滅菌器で容器を使用する場合は、使用者が特定の滅菌サイクルで容器の滅菌性能を検証する必要があります。容器の他の特性も、滅菌サイクルとの互換性のために確認する必要があります。例えば、動作温度などです。 注意3)組織化、乾燥、無菌的なプレゼンテーションのために内部に追加の材料(内包材、容器フィルタ、指示器、梱包リスト、マット、器具整理セット、トレイライナー、医療機器に追加のエンベロープなど)を使用する場合、検証活動中にこれらの材料の受容性の決定を含む他の要件が適用される場合があります。

기사 제목: SIST EN 868-8:2019 - 말단 살균 의료기기용 포장물 - 제8부: EN 285에 준하는 증기 살균기용 재사용 가능한 살균통 - 요구 사항 및 시험 방법 기사 내용: 이 EN 868의 일부는 말단 살균 의료기기의 살균되는 순간부터 사용되는 지점까지의 살균을 유지하기 위해 사용되는 재사용 가능한 살균통으로서 살균 밀폐 시스템으로 사용되는 시험 방법과 값 제공합니다. 이러한 살균통은 EN 285에 준하는 증기 살균기에 사용될 목적으로 제작되었습니다. 참고 1: 살균통 내부에 포장 재료가 필요한지는 제조업자와 사용자에 의해 결정됩니다. 이 EN 868의 일부는 EN ISO 11607-1에 명시된 일반 요구 사항을 추가하거나 수정하지 않고 이 부분에 포함된 제품에만 특정한 요구 사항과 시험 방법을 소개합니다. 따라서, 4.2에서 4.5까지의 특정 요구 사항은 EN ISO 11607-1의 일부 또는 전부를 충족함을 입증하기 위해 사용될 수 있습니다. 참고 2: EN 285에 준하지 않는 증기 살균기에서 살균통을 사용하려는 경우, 사용자는 해당 살균 주기의 살균 성능을 검증해야 합니다. 또한, 살균통의 다른 속성들 (예: 작동 온도)도 살균기 주기와의 호환성을 검토해야 합니다. 참고 3: 살균된 밀폐 시스템 내에서 조직, 건조 또는 무균적으로 제시하기 위해 기타 재료 (예: 내부 포장재, 살균통 필터, 지시자, 포장 목록, 매트, 기구 정리 세트, 트레이 라이너 또는 의료기기 주위의 추가적인 봉투)를 사용하는 경우, 유효성 검사 활동 중에 이러한 재료의 수용 가능성을 확인하기 위해 추가적인 요구 사항이 적용될 수 있습니다.

The article discusses SIST EN 868-8:2019, which provides test methods and values for re-usable sterilization containers used for maintaining the sterility of medical devices. These containers are specifically meant to be used in steam sterilizers conforming to EN 285. The article notes that the need for a packaging material inside the container is determined by the manufacturers and users. It also specifies that the requirements and test methods mentioned in this article do not modify the general requirements outlined in EN ISO 11607-1. Additional notes mention that if the containers are to be used in steam sterilizers that do not conform to EN 285, the user must validate the sterilization performance of the container in the specific sterilization cycle. Furthermore, if additional materials are used inside the sterile barrier system for organization, drying, or aseptic presentation, other requirements may apply.

기사 제목 : SIST EN 868-8:2019 - 말단 멸균된 의료기기용 포장물 - 8부: EN 285에 준하는 증기멸균기를 위한 재사용 스테릴화용기 - 요구사항 및 시험방법 기사 내용 : EN 868의 이 부분은 말단 멸균된 의료기기의 무균성을 사용 시점까지 유지하기 위한 용도로 사용되는 재사용용기에 대한 시험 방법과 값들을 제공한다. 이러한 용기는 EN 285에 준하는 증기멸균기에서 사용되도록 되어 있다. 참고1) 용기 내부에 포장 재료가 필요한지 여부는 제조업체와 사용자에 의해 결정된다. EN 868의 이 부분은 이 부분에 속하는 제품에 특화된 성능 요구사항과 시험 방법만을 소개하며 EN ISO 11607-1에 명시된 일반 요구사항을 추가하거나 수정하지는 않는다. 따라서 4.2에서 4.5까지의 특정 요구사항은 EN ISO 11607-1의 모든 요구사항을 만족시키지는 않지만 한 개 이상의 요구사항을 준수하는 데 사용될 수 있다. 참고2) EN 285에 준하지 않는 증기멸균기에서 용기를 사용할 경우 사용자는 특정 멸균주기에서 용기의 멸균 성능을 검증해야 한다. 용기의 다른 속성도 멸균주기와의 호환성을 위해 검토되어야 한다. 예를 들어 운영 온도 등을 말한다. 참고3) 내부 포장물, 용기 필터, 지시기, 포장 목록, 매트, 기구 조직 세트, 트레이 라이너, 의료기기 주위에 추가적인 봉투와 같은 조직화, 건조 또는 무균적인 제공을 위해 추가 재료를 사용할 경우, 다른 요구사항, 검증 활동 중 이러한 재료들의 수용성 결정을 포함하여, 적용될 수 있다.

The article discusses the requirements and test methods for re-usable sterilization containers used for steam sterilizers in the medical industry. These containers are designed to maintain the sterility of medical devices until they are used. The article emphasizes that the need for packaging material inside the containers is determined by the manufacturers and users. The article also notes that the requirements specified in EN ISO 11607-1 still apply, and the specific requirements mentioned in the article can be used to demonstrate compliance with some, but not all, of those requirements. If the containers are intended for use in a steam sterilizer that does not conform to EN 285, the user needs to validate the sterilization performance of the container in that specific sterilization cycle. Additionally, if other materials are used inside the sterile barrier system for organization or presentation purposes, additional requirements may apply.