Evaluation and routine testing in medical imaging departments - Part 2-6: Constancy tests - Imaging performance of computed tomography X-ray equipment (IEC 61223-2-6:2006)

This part of IEC 61223 provides assistance in performing CONSTANCY TESTS on a CT SCANNER. It applies to those components of CT SCANNERS which influence the image quality, PATIENT dose and positioning. This standard - defines the essential parameters which describe the performance of CT SCANNERS with regard to image quality, PATIENT dose and positioning; the list of parameters to be tested can be found in section 4.5; - defines the methods of testing the essential parameters; - provides criteria to be applied in the evaluation of data for compliance with the tolerances of the parameters specified by the ACCOMPANYING DOCUMENTS and with respect to BASELINE VALUES. These methods rely on non-invasive measurements, using appropriate test equipment, performed to ensure that the functional performance of EQUIPMENT meets ESTABLISHED CRITERIA or to enable the early recognition of changes in the properties of components of the EQUIPMENT. The aim is to verify compliance with specifications affecting the image quality, PATIENT dose and PATIENT positioning. It is not intended to consider: - aspects of mechanical and electrical safety nor - aspects of mechanical, electrical and software performance, unless they are essential for performing the CONSTANCY TESTS and directly affect image quality, PATIENT dose and positioning.

Bewertung und routinemäßige Prüfung in Abteilungen für medizinische Bildgebung - Teil 2-6: Konstanzprüfungen - Röntgeneinrichtungen für die Computertomographie (IEC 61223-2-6:2006)

Essais d'évaluation et de routine dans les services d'imagerie médicale - Partie 2-6: Essais de constance - Performance d'imagerie des équipements de tomodensitométrie à rayonnement X (CEI 61223-2-6:2006)

La présente partie de la CEI 61223 fournit une aide à la réalisation des essais de constance des tomodensitomètres. Elle s'applique aux composants des tomodensitomètres qui influencent la qualité de l'image, la dose patient et le positionnement. La présente norme - définit les paramètres essentiels qui décrivent les performances des tomodensitomètres concernant la qualité de l'image, la dose patient et le positionnement; la liste des paramètres à soumettre aux essais peut être trouvée au paragraphe 4.5; - définit les méthodes d'essai des paramètres essentiels; - fournit les critères à appliquer lors de l'évaluation des données pour établir la conformité aux tolérances des paramètres spécifiés par les documents d'accompagnement et concernant les valeurs de base. Ces méthodes reposent sur des mesurages non invasifs, utilisant des équipements d'essai appropriés, réalisés pour s'assurer que la performance fonctionnelle de l'équipement satisfait aux critères établis ou pour permettre une reconnaissance précoce des variations des propriétés des composants de l'équipements. Le but recherché est la vérification de la conformité aux spécifications concernant la qualité de l'image, la dose patient et le positionnement du patient.

Vrednotenje in rutinsko preskušanje v oddelkih za medicinsko slikanje - 2-6. del: Preskus konstantnosti - Rentgenska oprema za računalniško tomografijo (IEC 61223-2-6:2006)

Ta del IEC 61223 predpisuje pomoč pri izvajanju PRESKUSOV KONSTANTNOSTI pri RENTGENU ZA RAČUNALNIŠKO TOMOGRAFIJO. Velja za tiste komponente RENTGENOV ZA RAČUNALNIŠKE TOMOGRAFIJE, ki vplivajo na kvaliteto slike, odmerek na PACIENTA in pozicioniranje. Ta standard opredeljuje bistvene parametre, ki opisujejo zmogljivost RENTGENOV ZA RAČUNALNIŠKO TOMOGRAFIJO glede  kvalitete slike, odmerka za PACIENTA in pozicioniranja; seznam parametrov, ki se jih preskuša, se nahaja v Razdelku 4.5.; - opredeljuje metode preskušanja bistvenih parametrov; zagotavlja merila za uporabo pri vrednotenju podatkov za skladnost z odstopanji parametrov, ki jih določajo SPREMNI DOKUMENTI in ob upoštevanju SREDNJIH VREDNOSTI. Te metode se opirajo na neinvazivna merjenja z uporabo primerne preskusne opreme in delujejo tako, da se za funkcionalno zmogljivost OPREME zagotovi, da izpolnjuje VZPOSTAVLJENA MERILA ali da omogoča zgodnje prepoznavanje sprememb lastnosti komponent OPREME. Cilj je potrditi skladnost s specifikacijami, ki vplivajo na kakovost slike, odmerka za PACIENTA in pozicioniranje PACIENTA. Njegov namen ni, da upošteva: - vidike mehanske in električne varnosti, niti - vidike mehanske in električne zmogljivosti ter zmogljivosti programske opreme, razen če so le-ti bistveni za izvajanje PRESKUSOV KONSTANTNOSTI in neposredno vplivajo na kakovost slike, odmerka za PACIENTA in njegovo pozicioniranje.

General Information

Status
Withdrawn
Publication Date
12-Jul-2010
Withdrawal Date
01-Mar-2022
Current Stage
9900 - Withdrawal (Adopted Project)
Start Date
02-Mar-2022
Due Date
25-Mar-2022
Completion Date
02-Mar-2022

Relations

Buy Standard

Standard
EN 61223-2-6:2010
English language
37 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

SLOVENSKI STANDARD
SIST EN 61223-2-6:2010
01-september-2010
1DGRPHãþD
SIST EN 61223-2-6:1998
9UHGQRWHQMHLQUXWLQVNRSUHVNXãDQMHYRGGHONLK]DPHGLFLQVNRVOLNDQMHGHO
3UHVNXVNRQVWDQWQRVWL5HQWJHQVNDRSUHPD]DUDþXQDOQLãNRWRPRJUDILMR ,(&

Evaluation and routine testing in medical imaging departments - Part 2-6: Constancy
tests - Imaging performance of computed tomography X-ray equipment (IEC 61223-2-
6:2006)
Bewertung und routinemäßige Prüfung in Abteilungen für medizinische Bildgebung - Teil
2-6: Konstanzprüfungen - Röntgeneinrichtungen für die Computertomographie (IEC
61223-2-6:2006)
Essais d'évaluation et de routine dans les services d'imagerie médicale - Partie 2-6:
Essais de constance - Performance d'imagerie des équipements de tomodensitométrie à
rayonnement X (CEI 61223-2-6:2006)
Ta slovenski standard je istoveten z: EN 61223-2-6:2007
ICS:
11.040.50 Radiografska oprema Radiographic equipment
35.240.80 Uporabniške rešitve IT v IT applications in health care
zdravstveni tehniki technology
SIST EN 61223-2-6:2010 en
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------

SIST EN 61223-2-6:2010

---------------------- Page: 2 ----------------------

SIST EN 61223-2-6:2010


EUROPEAN STANDARD
EN 61223-2-6

NORME EUROPÉENNE
May 2007
EUROPÄISCHE NORM

ICS 11.040.50 Supersedes EN 61223-2-6:1994


English version


Evaluation and routine testing in medical imaging departments -
Part 2-6: Constancy tests -
Imaging performance of computed tomography X-ray equipment
(IEC 61223-2-6:2006)


Essais d'évaluation et de routine  Bewertung und routinemäßige Prüfung
dans les services d'imagerie médicale - in Abteilungen
Partie 2-6: Essais de constance - für medizinische Bildgebung -
Performance d'imagerie des équipements Teil 2-6: Konstanzprüfungen -
de tomodensitométrie à rayonnement X Röntgeneinrichtungen
(CEI 61223-2-6:2006) für die Computertomographie
(IEC 61223-2-6:2006)




This European Standard was approved by CENELEC on 2007-03-01. CENELEC members are bound to comply
with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European Standard
the status of a national standard without any alteration.

Up-to-date lists and bibliographical references concerning such national standards may be obtained on
application to the Central Secretariat or to any CENELEC member.

This European Standard exists in three official versions (English, French, German). A version in any other
language made by translation under the responsibility of a CENELEC member into its own language and notified
to the Central Secretariat has the same status as the official versions.

CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Cyprus, the
Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia,
Lithuania, Luxembourg, Malta, the Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain,
Sweden, Switzerland and the United Kingdom.

CENELEC
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung

Central Secretariat: rue de Stassart 35, B - 1050 Brussels


© 2007 CENELEC - All rights of exploitation in any form and by any means reserved worldwide for CENELEC members.
Ref. No. EN 61223-2-6:2007 E

---------------------- Page: 3 ----------------------

SIST EN 61223-2-6:2010
EN 61223-2-6:2007 - 2 -
Foreword
The text of document 62B/629/FDIS, future edition 2 of IEC 61223-2-6, prepared by SC 62B, Diagnostic
imaging equipment, of IEC TC 62, Electrical equipment in medical practice, was submitted to the
IEC-CENELEC parallel vote and was approved by CENELEC as EN 61223-2-6 on 2007-03-01.
This European Standard supersedes EN 61223-2-6:1994.
EN 61223-2-6:2007 is harmonized with the content of EN 60601-2-44:2001 + A1:2003 and with
EN 61223-3-5:2004. Instead of harmonizing test procedures for all modalities of X-ray equipment, as
intended in EN 61223-2-6:1994, this EN 61223-2-6:2007 comprises part of a set of standards covering all
of the particular requirements for CT scanners.
The following dates were fixed:
– latest date by which the EN has to be implemented
at national level by publication of an identical
national standard or by endorsement (dop) 2007-12-01
– latest date by which the national standards conflicting
(dow) 2010-03-01
with the EN have to be withdrawn
In this standard, the following print types are used:
– requirements, compliance with which can be tested, and definitions: in roman type;
– explanations, advice, notes, general statements, exceptions and references: in smaller type;
– test specifications: in italic type;
– TERMS DEFINED IN EN 60601-1, IN EN 60788, IN EN 61223-1 OR IN OTHER PUBLICATIONS REFERENCED IN
THE INDEX OF DEFINED TERMS: SMALL CAPITALS.
Annex ZA has been added by CENELEC.
__________
Endorsement notice
The text of the International Standard IEC 61223-2-6:2006 was approved by CENELEC as a European
Standard without any modification.
In the official version, for Bibliography, the following notes have to be added for the standards indicated:
IEC 61223-2-5 NOTE  Harmonized as EN 61223-2-5:1994 (not modified).
IEC 61223-2-4 NOTE  Harmonized as EN 61223-2-4:1994 (not modified).
IEC 60336 NOTE  Harmonized as EN 60336:2005 (not modified).
IEC 60522 NOTE  Harmonized as EN 60522:1999 (not modified).
IEC 60601-1 NOTE  Harmonized as EN 60601-1:2006 (not modified).
IEC 60601-2-28 NOTE  Harmonized as EN 60601-2-28:1993 (not modified).
IEC 60601-2-32 NOTE  Harmonized as EN 60601-2-32:1994 (not modified).
IEC 61267 NOTE  Harmonized as EN 61267:2006 (not modified).
__________

---------------------- Page: 4 ----------------------

SIST EN 61223-2-6:2010
- 3 - EN 61223-2-6:2007
Annex ZA
(normative)

Normative references to international publications
with their corresponding European publications

The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.

NOTE  When an international publication has been modified by common modifications, indicated by (mod), the relevant EN/HD
applies.

Publication Year Title EN/HD Year

IEC 60601-2-44 2001 Medical electrical equipment - EN 60601-2-44 2001
A1 2002 Part 2-44: Particular requirements for the A1 2003
safety of X-ray equipment for computed
tomography


IEC/TR 60788 2004 Medical electrical equipment - Glossary of - -
defined terms


IEC 61223-3-5 2004 Evaluation and routine testing in medical EN 61223-3-5 2004
imaging departments -
Part 3-5: Acceptance tests - Imaging
performance of computed tomography X-ray
equipment

---------------------- Page: 5 ----------------------

SIST EN 61223-2-6:2010

---------------------- Page: 6 ----------------------

SIST EN 61223-2-6:2010
NORME CEI
INTERNATIONALE
IEC



61223-2-6
INTERNATIONAL


Deuxième édition
STANDARD

Second edition

2006-11


Essais d’évaluation et de routine dans
les services d’imagerie médicale –
Partie 2-6:
Essais de constance –
Performance d’imagerie des équipements
de tomodensitométrie à rayonnement X

Evaluation and routine testing in
medical imaging departments –
Part 2-6:
Constancy tests –
Imaging performance of computed
tomography X-ray equipment
© IEC 2006 Droits de reproduction réservés ⎯ Copyright - all rights reserved
Aucune partie de cette publication ne peut être reproduite ni No part of this publication may be reproduced or utilized in any
utilisée sous quelque forme que ce soit et par aucun procédé, form or by any means, electronic or mechanical, including
électronique ou mécanique, y compris la photocopie et les photocopying and microfilm, without permission in writing from
microfilms, sans l'accord écrit de l'éditeur. the publisher.
International Electrotechnical Commission, 3, rue de Varembé, PO Box 131, CH-1211 Geneva 20, Switzerland
Telephone: +41 22 919 02 11 Telefax: +41 22 919 03 00 E-mail: inmail@iec.ch Web: www.iec.ch
CODE PRIX
V
PRICE CODE
Commission Electrotechnique Internationale
International Electrotechnical Commission
МеждународнаяЭлектротехническаяКомиссия
Pour prix, voir catalogue en vigueur
For price, see current catalogue

---------------------- Page: 7 ----------------------

SIST EN 61223-2-6:2010
61223-2-6 © IEC:2006 – 3 –
CONTENTS

FOREWORD.5
INTRODUCTION.9

1 Scope and object.11
2 Normative references .11
3 Terms and definitions .13
4 General aspects of CONSTANCY TESTS .21
4.1 Preconditions .21
4.2 General conditions affecting test procedures .21
4.3 Establishment of BASELINE VALUES .23
4.4 Identification of equipment, instrumentation and test conditions .23
4.5 Scope of the CONSTANCY TESTS.25
4.6 Test equipment, including PHANTOMS and TEST DEVICES .25
4.7 Frequency of CONSTANCY TESTS .27
5 Test methods for CT SCANNERS.27
5.1 Positioning of the PATIENT SUPPORT.27
5.2 PATIENT positioning accuracy .29
5.3 TOMOGRAPHIC SECTION THICKNESS .33
5.4 Dose .37
5.5 NOISE, UNIFORMITY AND MEAN CT NUMBERS .39
5.6 SPATIAL RESOLUTION.43

Annex A (informative) Guidance on action to be taken .47
Annex B (informative) Alternate test methods for SPATIAL RESOLUTION .51
Annex C (informative) TOMOGRAPHIC SECTION THICKNESS for helical scanning .57
Annex D (informative) Visual method for LOW CONTRAST RESOLUTION .59

Bibliography.61

Index of terms.65

Figure 1 – Coordinate system used for CT SCANNERS.21

Table B.1 − Comparison of SPATIAL RESOLUTION test procedures .51

---------------------- Page: 8 ----------------------

SIST EN 61223-2-6:2010
61223-2-6 © IEC:2006 – 5 –
INTERNATIONAL ELECTROTECHNICAL COMMISSION
___________

EVALUATION AND ROUTINE TESTING
IN MEDICAL IMAGING DEPARTMENTS –

Part 2-6: Constancy tests – Imaging performance
of computed tomography X-ray equipment


FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC provides no marking procedure to indicate its approval and cannot be rendered responsible for any
equipment declared to be in conformity with an IEC Publication.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International Standard IEC 61223-2-6 has been prepared by subcommittee 62B: Diagnostic
imaging equipment, of IEC technical committee 62: Electrical equipment in medical practice.
This second edition cancels and replaces the first edition published in 1994. It constitutes a
technical revision. This second edition is harmonized with the content of IEC 60601-2-44:2001
and its Amendment 1 (2002) and with IEC 61223-3-5. Instead of harmonizing test procedures
for all modalities of X-ray equipment, as intended in the first edition, this second edition
comprises part of a set of standards covering all of the particular requirements for CT
scanners.

---------------------- Page: 9 ----------------------

SIST EN 61223-2-6:2010
61223-2-6 © IEC:2006 – 7 –
The text of this standard is based on the following documents:
FDIS Report on voting
62B/629/FDIS 62B/639/RVD

Full information on the voting for the approval of this standard can be found in the report on
voting indicated in the above table.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
A list of all parts of the IEC 61223 series, published under the general title Evaluation and routine
testing in medical imaging departments, can be found on the IEC website.
In this standard, the following print types are used:
– requirements, compliance with which can be tested, and definitions: in roman type;
– explanations, advice, notes, general statements, exceptions and references: in smaller type;
– test specifications and headings of subclauses: in italic type;
– TERMS DEFINED IN IEC 60601-1, IN IEC 60788, IN IEC 61223-1 OR IN OTHER IEC PUBLICATIONS
REFERENCED IN THE INDEX OF DEFINED TERMS: SMALL CAPITALS.
The committee has decided that the contents of this publication will remain unchanged until
the maintenance result date indicated on the IEC web site under "http://webstore.iec.ch" in
the data related to the specific publication. At this date, the publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.

---------------------- Page: 10 ----------------------

SIST EN 61223-2-6:2010
61223-2-6 © IEC:2006 – 9 –
INTRODUCTION
The second edition of this standard is harmonized with the content of
– IEC 60601-2-44:2001, Particular requirements for the safety of X-ray equipment for
computed tomography and its Amendment 1 (2002), and
– IEC 61223-3-5, Evaluation and routine testing in medical imaging departments –
Part 3-5, Acceptance tests – Imaging performance of computed tomography X-ray
equipment.
In case the CT SCANNER does not offer an integrated, automated evaluation of the test
images, attention shall be given to the proper function and setting of the IMAGE DISPLAY
DEVICE. It is strongly recommended to assure its proper functioning prior to the constancy
1)
testing of the CT SCANNER, applying IEC 61223-2-5 [1] on IMAGE DISPLAY DEVICES or the
related in-house procedure. Some provisions or statements in this standard require additional
information, which is presented in the annexes. An asterisk in the left margin of a clause or
subclause indicates the presence of such additional information.
With regard to the measurements, reference is made to methods described in related
publications, which for practical reasons should be carried out prior to the application of the
methods described in this standard.
This standard forms Part 2-6 of IEC 61223, which includes the following parts of particular
interest in the context of this standard:

Part 1: General aspects [2]
Part 2-4: Constancy tests – Hard copy cameras [3]
Part 2-5: Constancy tests – Image display devices
———————
1)
Figures in square brackets refer to the Bibliography.

---------------------- Page: 11 ----------------------

SIST EN 61223-2-6:2010
61223-2-6 © IEC:2006 – 11 –
EVALUATION AND ROUTINE TESTING
IN MEDICAL IMAGING DEPARTMENTS –

Part 2-6: Constancy tests – Imaging performance
of computed tomography X-ray equipment



1 Scope and object
This part of IEC 61223 provides assistance in performing CONSTANCY TESTS on a CT SCANNER.
It applies to those components of CT SCANNERS which influence the image quality, PATIENT
dose and positioning.
This standard
– defines the essential parameters which describe the performance of CT SCANNERS with
regard to image quality, PATIENT dose and positioning; the list of parameters to be tested
can be found in section 4.5;
– defines the methods of testing the essential parameters;
– provides criteria to be applied in the evaluation of data for compliance with the tolerances
of the parameters specified by the ACCOMPANYING DOCUMENTS and with respect to BASELINE
VALUES.
These methods rely on non-invasive measurements, using appropriate test equipment,
performed to ensure that the functional performance of EQUIPMENT meets ESTABLISHED
CRITERIA or to enable the early recognition of changes in the properties of components of the
EQUIPMENT.
The aim is to verify compliance with specifications affecting the image quality, PATIENT dose
and PATIENT positioning.
It is not intended to consider:
– aspects of mechanical and electrical safety nor
– aspects of mechanical, electrical and software performance, unless they are essential for
performing the CONSTANCY TESTS and directly affect image quality, PATIENT dose and
positioning.
2 Normative references
The following referenced documents are indispensable for the application of this document.
For dated references, only the edition cited applies. For undated references, the latest edition
of the referenced document (including any amendments) applies.
IEC 60788:2004, Medical electrical equipment – Glossary of defined terms
IEC 61223-3-5:2004, Evaluation and routine testing in medical imaging departments –
Part 3-5: Acceptance tests – Imaging performance of computed tomography X-ray equipment

---------------------- Page: 12 ----------------------

SIST EN 61223-2-6:2010
61223-2-6 © IEC:2006 – 13 –
IEC 60601-2-44:2001, Medical electrical equipment – Part 2-44: Particular requirements for
the safety of X-ray equipment for computed tomography
Amendment 1 (2002)
3 Terms and definitions
For the purposes of this document, the terms and definitions given in IEC 60788:2004 and the
following apply.
NOTE 1 An index of defined terms used in this standard is given following the annexes
NOTE 2 Attention is drawn to the fact that in cases where the concept addressed is not strongly confined to the
definition given in IEC 61223-1 or IEC 60788, a corresponding term is printed in lower case letters.
3.1
CT CONDITIONS OF OPERATION
all selectable parameters governing the operation of a CT SCANNER, for example NOMINAL
TOMOGRAPHIC SECTION THICKNESS, CT PITCH FACTOR, FILTRATION, peak X-RAY TUBE VOLTAGE and
either X-RAY TUBE CURRENT and LOADING TIME or CURRENT TIME PRODUCT
[IEC 60601-2-44:2001, definition 2.102]
3.2
CT SCANNER
X-RAY EQUIPMENT for COMPUTED TOMOGRAPHY (CT)
diagnostic X-ray system intended to generate cross-sectional images of the body by computer
reconstruction of X-ray transmission data obtained at different angles. This generic type of
device may include signal analysis and display equipment, PATIENT SUPPORT, support parts
and accessories
NOTE Secondary imaging processing is not included in the scope of this standard.
[IEC 60601-2-44, Amend.1:2002, definition 2.101]
3.3
COMPUTED TOMOGRAPHY DOSE INDEX 100
CTDI
100
integral of the DOSE PROFILE produced in a single axial scan along a line perpendicular to
the TOMOGRAPHIC PLANE from – 50 mm to + 50 mm, divided by the product of the number of
TOMOGRAPHIC SECTIONS N and the NOMINAL TOMOGRAPHIC SECTION THICKNESS T
+50 mm
D (z)
=  dz
CTDI
100

N ⋅ T
-50 mm
where
D(z) is the DOSE PROFILE along a line z perpendicular to the TOMOGRAPHIC PLANE, where
dose is reported as ABSORBED DOSE to air;
N is the number of TOMOGRAPHIC SECTIONS produced in a single axial scan of the X-ray
source;
T is the NOMINAL TOMOGRAPHIC SECTION THICKNESS.
NOTE 1 The term CTDI has been introduced as a more representative value for dose than the traditional CTDI
100
integrated from – 7 T to + 7 T as defined by the FDA in 21 CFR 1020.33 [31].
NOTE 2 The dose is reported as ABSORBED DOSE to air. This is required in order to avoid present confusion, as
some MANUFACTURERS of CT SCANNERS express dose values calculated as ABSORBED DOSE to air and others as
ABSORBED DOSE to polymethyl-methacrylate (PMMA).

---------------------- Page: 13 ----------------------

SIST EN 61223-2-6:2010
61223-2-6 © IEC:2006 – 15 –
Although CTDI refers to ABSORBED DOSE to air, for practical purposes the evaluation of ABSORBED DOSE to air
100
within a PMMA dosimetry PHANTOM is well approximated by measurement of the AIR KERMA with an ionization
chamber in the PHANTOM.
NOTE 3 This definition assumes that the DOSE PROFILE is centred on z = 0.
NOTE 4 A single axial scan is typically a 360° rotation of the X-ray source.
[IEC 60601-2-44, Amendment 1:2002, definition 2.106]
3.4
WEIGHTED CTDI
100
The WEIGHTED CTDI (CTDI ) is defined as
100 w
1 2
CTDI = CTDI + CTDI
W 100(centre ) 100(periphe ral)
3 3
where
CTDI is the value measured in the centre of the CT DOSIMETRY PHANTOM and
100(centre)

CTDI is the average value measured in the periphery of the CT DOSIMETRY
100(peripheral)
PHANTOM
3.5
COMPUTED TOMOGRAPHY NUMBER (CT NUMBER)
number used to represent the mean X-ray ATTENUATION associated with each elemental area
of the COMPUTED TOMOGRAPHY image
NOTE The CT NUMBER is normally expressed in Hounsfield units. MEASURED VALUES of the linear ATTENUATION
coefficients are transformed into CT NUMBERS using the international Hounsfield scale, using the expression:
μ − μ
material water
CT number of material = ⋅1 000
μ
water
where
μ is the linear ATTENUATION coefficient.
The CT NUMBER scale is defined so that water has a value of 0 and air a value of -1000 (μ is assumed to be 0).
air
3.6
COMPUTED TOMOGRAPHY DOSE INDEX FREE IN AIR
CTDI
free air
CTDI measured at isocentre in the absence of a PHANTOM and the PATIENT SUPPORT
100
3.7
CT PITCH FACTOR
in helical scanning the ratio of the PATIENT SUPPORT travel Δd along the z direction per rotation
of the X-RAY SOURCE divided by the product of the NOMINAL TOMOGRAPHIC SECTION THICKNESS T
and the number of TOMOGRAPHIC SECTIONS N:
Δd
CT pitch factor =
N ⋅T
where
Δd is the PATIENT SUPPORT travel along the z direction per rotation of the X-RAY SOURCE;
T is the NOMINAL TOMOGRAPHIC SECTION THICKNESS;

---------------------- Page: 14 ----------------------

SIST EN 61223-2-6:2010
61223-2-6 © IEC:2006 – 17 –
N is the number of TOMOGRAPHIC SECTIONS produced in a single axial scan of the X-RAY
SOURCE.
[IEC 60601-2-44, Amend.1:2002, definition 2.107]
3.8
DOSE PROFILE
representation of the dose as a function of position along a line.
[IEC 60601-2-44:2001, definition 2.103]
3.9
MEAN CT NUMBER
mean value of the CT NUMBERS of all pixels within a certain defined REGION OF INTEREST
3.10
NOISE
variation of CT NUMBERS from a mean value in a defined area in the image of a uniform
substance
NOTE The magnitude of NOISE is indicated by the standard deviation of the CT NUMBERS of a uniform substance in
the REGION OF INTEREST.
3.11
NOMINAL TOMOGRAPHIC SECTION THICKNESS
in CT SCANNERS the TOMOGRAPHIC SECTION THICKNESS which is selected and indicated on the
CONTROL PANEL
NOTE In helical scanning the thickness of a reconstructed image depends on the helical reconstruction algorithm
NOMINAL TOMOGRAPHIC SECTION THICKNESS. The thickness of
and pitch, and hence this thickness may not equal the
the reconstructed image may be indicated or selected prior to the helical scan.
[IEC 60601-2-44, Amend.1:2002, definition 2.110]
3.12
SENSITIVITY PROFILE
relative response of a system for COMPUTED TOMOGRAPHY as a function of position along a line
perpendicular to the TOMOGRAPHIC PLANE
[IEC 60601-2-44:2001, definition 2.104]
3.13
SPATIAL RESOLUTION
in EQUIPMENT for COMPUTED TOMOGRAPHY, the ability to resolve different objects in the
displayed image, when the difference in ATTENUATION between the objects and the back-
ground is large compared to NOISE
NOTE 1 Normally a difference in ATTENUATION coefficient between the object and the background resulting in a
difference of the respective CT NUMBERS of several hundred Hounsfield units is regarded as large enough.
NOTE 2 HIGH-CONTRAST RESOLUTION is an alternative name for SPATIAL RESOLUTION.
3.14
TOMOGRAPHIC PLANE
geometric plane perpendicular to the axis of rotation
[IEC 60601-2-44:2001, definition 2.105]

---------------------- Page: 15 ----------------------

SIST EN 61223-2-6:2010
61223-2-6 © IEC:2006 – 19 –
3.15
TOMOGRAPHIC SECTION
volume over which TRANSMISSION data of X-RADIATION are collected in a single axial scan
NOTE In a CT SCANNER with multiple detector elements along the z-axis, it is the volume over which data are
collected by a single acquisition channel (selected grouping of elements) and not the total volume irradiated.
[IEC 60601-2-44, Amendment 1:2002, definition 2.108]
3.16
TOMOGRAPHIC SECTION THICKNESS
FULL WIDTH AT HALF MAXIMUM of the SENSITIVITY PROFILE taken at the isocentre of a
TOMOGRAPHIC SECTION
[IEC 60601-2-44, Amendment 1:2002, definition 2.109]
3.17
UNIFORMITY
consistency of the CT NUMBERS of the image of a homogeneous material across the scan field
3.18
VOLUME CTDI
W
CTDI
vol
average dose over the total volume scanned for the selected CT CONDITIONS OF OPERATION
The CTDI is defined as follows:
vol
a) for axial scanning
N ⋅T
CTDI = ⋅CTDI
vol w
Δd
where
N is the number of TOMOGRAPHIC SECTIONS produced in a single axial scan of the X-RAY
SOURCE;
T is the NOMINAL TOMOGRAPHIC SECTION THICKNESS;
Δd is the PATIENT SUPPORT travel in z-direction between consecutive scans.
b) for helical scanning
CTDI
w
CTDI =
vol
CT pitch factor
c) for scanning without pre-programmed movement of the PATIENT SUPPORT
CTDI = n ⋅CTDI
vol w
where n is equal to the maximum number of pre-programmed rotations.
NOTE 1 The displayed CTDI given by the MANUFACTURER may be a representative figure for that model and not
vol
the value measured on the particular CT SCANNER
NOTE 2 The definition of CTDI in section c) will li
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.