Representation of results of particle size analysis - Part 2: Calculation of average particle sizes/diameters and moments from particle size distributions

This part of ISO 9276 provides relevant equations and coherent nomenclatures for the calculation of
moments, mean particle sizes and standard deviations from a given particle size distribution. Two
notation systems in common use are described. One is the method of moments while the second
describes the moment-ratio method. The size distribution may be available as a histogram or as an
analytical function.
The equivalent diameter of a particle of any shape is taken as the size of that particle. Particle shape
factors are not taken into account. It is essential that the measurement technique is stated in the report
in view of the dependency of sizing results of measurement principle. Samples of particles measured are
intended to be representative of the population of particles.
For both notation systems, numerical examples of the calculation of mean particle sizes and standard
deviation from histogram data are presented in an annex.
The accuracy of the mean particle size may be reduced if an incomplete distribution is evaluated. The
accuracy may also be reduced when very limited numbers of size classes are employed.

Représentation de données obtenues par analyse granulométrique - Partie 2: Calcul des tailles/diamètres moyens des particules et des moments à partir de distributions granulométriques

Predstavitev podatkov, dobljenih z granulometrijsko analizo - 2. del: Izračun povprečnih velikosti/premerov in momentov iz granulacij

Ta del standarda ISO 9276 navaja ustrezne enačbe in skladne nomenklature za izračune momentov in povprečnih velikosti iz granulacij ter standardnih odklonov. Opisana sta dva sistema zapisa, ki se običajno uporabljata. Eden je metoda momentov, drugi pa opisuje metodo razmerja momenta. Granulacija je lahko na voljo kot histogram ali analitična funkcija.
Ekvivalentni premer delca kakršne koli oblike je vzet kot velikost tega delca. Dejavniki oblike delca niso upoštevani. Bistveno je, da je tehnika merjenja navedena v poročilu v smislu odvisnosti rezultatov določanja velikosti od načela merjenja. Vzorci izmerjenih delcev naj bi bili reprezentativni za populacijo delcev.
Za oba sistema zapisa so v dodatku predstavljeni numerični primeri izračuna povprečnih velikosti delcev in standardnega odklona od podatkov histograma.
Natančnost povprečne velikosti delca se lahko zmanjša, če se ocenjuje nepopolna granulacija. Natančnost se lahko zmanjša tudi, kadar se uporablja zelo omejeno število razredov velikosti.

General Information

Status
Published
Publication Date
19-Feb-2015
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
10-Feb-2015
Due Date
17-Apr-2015
Completion Date
20-Feb-2015

Relations

Buy Standard

Standard
ISO 9276-2:2014
English language
35 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
ISO 9276-2:2015
English language
32 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day
Standard
ISO 9276-2:2014 - Representation of results of particle size analysis
English language
27 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

МЕЖДУНАРОДНЫЙ ISO
СТАНДАРТ 9276-2
Второе издание
2014-05-15

Гранулометрический анализ.
Представление результатов.
Часть 2:
Расчет средних размеров/диаметров
частиц и моментов по распределению
частиц по размерам
Representation of results of particle size analysis —
Part 2: Calculation of average particle sizes/diameters and moments
from particle size distributions



Ответственность за подготовку русской версии несёт GOST R
(Российская Федерация) в соответствии со статьёй 18.1 Устава ISO
Ссылочный номер
ISO 9276-2:2014(R)
©
ISO 2014

---------------------- Page: 1 ----------------------
ISO 9276-2:2014(R)

ДОКУМЕНТ ЗАЩИЩЕН АВТОРСКИМ ПРАВОМ


©  ISO 2014
Все права сохраняются. Если не указано иное, никакую часть настоящей публикации нельзя копировать или использовать в
какой-либо форме или каким-либо электронным или механическим способом, включая фотокопии и микрофильмы, без
предварительного письменного согласия ISO по адресу, указанному ниже, или членов ISO в стране регистрации пребывания.
ISO copyright office
Case postale 56  CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Опубликовано в Швейцарии

ii ISO 2014 – Все права сохраняются

---------------------- Page: 2 ----------------------
ISO 9276-2:2014(R)
Содержание Страница
Предисловие . iv
Введение . v
1  Область применения . 1
2  Нормативные ссылки . 1
3  Термины и определения. 1
4  Представление моментов . 3
4.1  Определение моментов согласно методу представления моментов . 3
4.2  Определение среднего размеров частиц по методу представления моментов . 4
4.3  Расчет моментов и средние размеры частиц по данному распределению размеров . 7
4.4  Дисперсия и стандартное отклонение распределения частиц по размерам . 9
4.5  Расчет моментов и средние размеры частиц из логарифмически нормального
распределения . 10
4.6  Расчет объемной удельной площади поверхности и среднего диаметра Соутера . 10
5  Метод представления отношения моментов . 10
5.1  Определение моментов согласно методу представления отношений моментов . 11
5.2  Определение средних размеров частиц методом представления отношений
моментов . 11
5.3  Расчет средних размеров частиц по данному распределению размеров . 13
5.4  Дисперсия и стандартное отклонение распределения частиц по размерам . 14
5.5  Соотношение между средними размерами частиц . 15
5.6  Расчет объемной удельной площади поверхности и среднего диаметра Соутера . 16
6  Соотношение между представлением моментов и представлением отношений
моментов . 16
7  Точность рассчитанных параметров распределения частиц по размерам . 19
Приложение А (информативное) Числовой пример для расчета средних размеров частиц и
стандартное отклонение по гистограмме распределения частиц по объему . 20
Приложение В (информативное) Числовой пример расчета средних размеров частиц и
стандартного отклонения по гистограмме распределения частиц по объему . 23
Приложение С (информативное) Точность рассчитанных параметров распределения частиц
по размеру . 27
Библиография . 29

ISO 2014 – Все права сохраняются iii

---------------------- Page: 3 ----------------------
ISO 9276-2:2014(R)
Предисловие
Международная организация по стандартизации (ISO) всемирная федерация национальных органов
по стандартизации (комитеты-члены ISO). Работа по подготовке международных стандартов обычно
ведется через технические комитеты ISO. Каждый комитет-член ISO, проявляющий интерес к
тематике, по которой учрежден технический комитет, имеет право быть представленным в этом
комитете. Международные организации, государственные и негосударственные, имеющие связи с ISO,
также принимают участие в работе. ISO тесно сотрудничает с Международной электротехнической
комиссией (IEC) по всем вопросам стандартизации в области электротехники.
Процедуры, используемые для разработки данного документа, и процедуры, предусмотренные для его
дальнейшего ведения, описаны в Части 1 Директив ISO/IEC. В частности, следует отметить различные
критерии утверждения, требуемые для различных типов документов ISO. Проект данного документа
был разработан в соответствии с редакционными правилами Части 2 Директив ISO/IEC.
www.iso.org/directives.
Необходимо обратить внимание на возможность того, что ряд элементов данного документа могут
быть предметом патентных прав. Международная организация ISO не должна нести ответственность
за идентификацию таких прав, частично или полностью. Сведения о патентных правах,
идентифицированных при разработке документа, будут указаны во Введении и/или в перечне
полученных ISO объявлениях о патентном праве. www.iso.org/patents .
Любое торговое название, использованное в данном документе, является информацией,
предоставляемой для удобства пользователей, а не свидетельством в пользу того или иного товара
или той или иной компании.
Для пояснения значений конкретных терминов и выражений ISO, относящихся к оценке соответствия, а
также информация о соблюдении Международной организацией ISO принципов ВТО по техническим
барьерам в торговле (TБT), см. следующий унифицированный локатор ресурса (URL): Foreword -
Supplementary information.
За данный документ несет ответственность технический комитет ISO/TC 24, Определение
характеристик частиц, включая ситовый анализ, Подкомитетом SC 4, Определение характеристик
частиц.
Настоящее второе издание отменяет и заменяет первое издание (ISO 9276-2:2001) после технического
пересмотра.
ISO 9276 состоит из следующих частей под общим названием Гранулометрический анализ.
Представление результатов:
— Часть 1: Графическое представление
— Часть 2: Расчет средних размеров/диаметров частиц и моментов по распределению частиц по
размерам
— Часть 3: Подгонка экспериментальной кривой в соответствии с эталонной моделью
— Часть 4: Определение процесса классификации
— Часть 5: Методы расчета при гранулометрическом анализе с использованием
логарифмического нормального распределения вероятностей
— Часть 6: Описательное и количественное представление формы и морфологии частиц
iv ISO 2014 – Все права сохраняются

---------------------- Page: 4 ----------------------
ISO 9276-2:2014(R)
Введение
Для характеризации твердых частиц часто пользуются гранулометрическим анализом. Взаимосвязь
между физическими свойствами материала в виде частиц, такие как прочность, текучесть, скорость
растворения, стабильность эмульсии/суспензии и размер частиц порошка всегда дает основание для
такой характеризации. Для материалов, имеющих распределение частиц по размерам, важно
использовать соответствующий параметр, определенный средний размер частиц, взвешенный,
например, по количеству, площади или объему, во взаимосвязи с физическими свойствами.
Данная часть ISO 9276 описывает две процедуры для использования моментов для расчета средних
размеров, разброс и другие статистические меры распределения частиц по размерам
(гранулометрического состава).
Первый метод называется метод символического представления моментов. Конкретная польза
символического представления моментов заключается в характеризации гранулометрического состава
с помощью моментов и средних размеров. Символическое представление моментов черпает принципы
взвешивания из физики, особенно из механики, и включает среднее арифметическое от
[1][2]
количественных распределений по размерам только как одну часть .
Второй метод называется методом символического представления отношения моментов. Этот метод
основан на статистических критериях и частотах, а также включает преобразование в другие типы
[3][4]
величин .
Важно отметить, что значение подстрочных индексов средних размеров, определенных в методе
символического представления моментов отличается от подстрочных индексов средних размеров в
методе символического представления отношения моментов. Оба представления связаны простым
отношением, приведенным в Разделе 6.
Оба представления подходят для выведения и /или выбора средних размеров, относящихся к
физическим свойствам продукта или процесса для так называемых функций свойства или функций
процесса. Тип предпочтительного среднего размера должен иметь причинно-следственную связь с
соответствующим физическим свойством продукции или процесса.
Множество характеристик частиц включает очень широкий спектр научных дисциплин. На обозначение
задействованного распределения по размерам влияет отрасль промышленности и применение, и,
таким образом, ни одно обозначение не получило всеобщего предпочтения.
Существует несколько свойств, зависящих от размера частиц, например, рассеяние света в
определенных диапазонах размеров частиц, которые невозможно характеризовать с помощью средних
[5]
размеров частиц, выведенных из простых уравнений экспоненциального закона систем обозначения .

ISO 2014 – Все права сохраняются v

---------------------- Page: 5 ----------------------
МЕЖДУНАРОДНЫЙ СТАНДАРТ ISO 9276-2:2014(R)

Гранулометрический анализ. Представление результатов.
Часть 2.
Расчет средних размеров/диаметров частиц и моментов по
распределению частиц по размерам
1 Область применения
Данная часть ISO 9276 представляет соответствующие формулы и логически обоснованные
номенклатуры для вычисления моментов, средних размеров частиц и стандартных отклонений от
данного распределения частиц по размерам. Описаны две обычно использующиеся системы
обозначения. Одна из метода моментов, вторая система описывает метод отношения моментов.
Гранулометрический состав можно получить в виде гистограмме или аналитической функции.
Эквивалентный диаметр частицы любой формы берется как размер этой частицы. Коэффициенты
формы частицы не учитываются. Важно указать в протоколе технику измерения в свете зависимости
результатов определения размеров от принципов измерения. Предполагается, что пробы измеренных
частиц являются репрезентативными для совокупности частиц.
Для обеих систем обозначения в приложении представлены числовые примеры расчета средних
размеров частиц и стандартного отклонения по данным гистограммы.
Точность среднего размера частиц может снизиться, если оценивают неполное распределение.
Точность может также снизиться, если задействованы очень ограниченные количества классов
распределения размеров.
2 Нормативные ссылки
Следующие ссылочные нормативные документы являются обязательными при применении данного
документа. Для жестких ссылок применяется только цитированное издание документа. Для плавающих
ссылок необходимо использовать самое последнее издание нормативного ссылочного документа
(включая любые изменения).
ISO 9276-1:1998, Гранулометрический анализ. Представление размеров. Часть 1: графическое
представление
ISO 9276-5:2005, Гранулометрический анализ. Представление размеров. Часть 5: Методы расчета,
относящиеся к гранулометрическому анализу с помощью нормального логарифмического
распределения вероятностей
3 Термины и определения
При необходимости различные обозначения даются в представлении моментов (М) и представлении
отношения моментов (M-R). Это служит цели четкой дифференциации между двумя системами. Для
обеих систем обозначения терминология конкретных средних размеров частиц вставлена в
соответствующие разделы: Раздел 4 и Раздел 5, соответственно.
ISO 2014 – Все права сохраняются 1

---------------------- Page: 6 ----------------------
ISO 9276-2:2014(R)
Представ- Представ-
Описание
ление М ление M-R
i i номер класса распределения размеров с максимальным размером
частиц: x (M)
i
или средним размером частиц Di(M-R)
k степень x
m m число классов распределения размеров
r r тип количественного распределения (общее описание)
 r = 0,  тип распределения: по количеству
 r = 1,  тип распределения: по длине (диаметру)
 r = 2,  тип распределения: по поверхностиь или спроектированной
площади
 r = 3,  тип распределения: по объему или по массе
M полный k-тый момент выборочного распределения q (x)
k,r r
m полный k-тый центральный момент выборочного распределения q (x)
k,r r
M p-тый момент плотности количественного распределения
p
m p- тый центральный момент плотности количественного распределения
p
N общее число частиц в пробе
O порядок среднего размера частиц (O = p + q)
p, q степени D в моментах или подстрочные индексы, указывающие то же
самое
q (x) q (D) плотность распределения по типу распределения частиц r
r r
средний уровень плотности распределения на интервале размеров i-той
q
ri,
частицы, ∆x
i
Q (x) Q (D) интегральная функция распределения типа величины r
r r
∆Q разность двух значений интегральной функции, т.e. относительная
r,i
величина на интервале размеров i-той частицы, ∆x
i
s s стандартное отклонение распределения q (x) и q (D)
r r r r
s s геометрическое стандартное отклонение распределения
g g
s s стандартное отклонение логарифмически нормального распределения
(s = ln s )
g
S S площадь поверхности
S S объемная удельная площадь поверхности
V V
V V объем частицы
средний объем частицы

V
x D размер частицы, диаметр эквивалентной сферы
x верхний размер интервала размеров i-той частицы
i
x нижний размер интервала размеров i-той частицы
i-1
D средний размер в i-том размерном классе
i
2 ISO 2014 – Все права сохраняются

---------------------- Page: 7 ----------------------
ISO 9276-2:2014(R)
x размер частиц минимальный в данном распределении по размерам
min
(гранулометрическом составе)
x размер частиц максимальный в данном распределении по размерам
max
(гранулометрическом составе)
средние размеры частиц (общее описание)
x D
kr, pq,
средние геометрические размеры частиц
D
pp,
средний арифметический размер частиц
x
k,0
средний взвешенный размер частиц
x
kr,
средний геометрический размер частиц
x
geo,r
средний гармонический диаметр
x
har,r
x медианный размер частиц интегральной функции объемного
50,3
распределения
∆x = x – x ширина интервала размеров i-той частицы
i i i-1
4 Представление моментов
Моменты являются основой для определения средних размеров и стандартных отклонений
распределения частиц по размерам. Случайная выборка, содержащая ограниченное число частиц из
большой совокупности размеров частиц, используется для оценки моментов распределения по
размерам этой совокупности. Оценка касается предположения о числовых значениях неизвестной
совокупности по значениям в выборке. Измерения размеров частиц всегда выполняют на дискретных
выборках т включают ряд дискретных размерных классов. Поэтому, данная часть ISO 9276 касается
только моментов, относящихся к выборкам.
4.1 Определение моментов согласно методу представления моментов
[ ]
1
Полный k-тый момент плотности распределения представляют интегралами в соответствии с
формулой (1). M означает момент. Первый подстрочный индекс, k, для M, указывает степень размера x
частицы, второй индекс, r, для M описывает тип величины плотности распределения.
x
max
k
Mx q()xdx (1)
k,r r

x
min
Если r = 0, q (x) представляет плотность количественного распределения, если r = 3, q (x)
0 3
представляет плотность распределения объемов или масс.
Формула (1) описывает полный момент, если границы интеграла представлены минимальным (x ) и
min
максимальным размером частиц (x ).
max
Частный полный момент представлен M :
0,r
xx
max max
0
Mxq()xdx q()xdxQ(x )Q(x )1 (2)
0m,r r r rax rmin

xx
min min
при
ISO 2014 – Все права сохраняются 3

---------------------- Page: 8 ----------------------
ISO 9276-2:2014(R)
x
i
Qx() qx()dx (3)
ri r

x
min
Момент является неполным, если интегрирование выполняется между двумя случайными диаметрами
частиц x и x в данном диапазоне размеров распределения:
i-1 i
x < x < x < x < x
min i-1 i max
x
i
k
Mx(,x) xq(x)dx (4)
k,r i1 i r

x
i1
Помимо моментов, связанных с началом оси размеров частиц и показанных в формуле (1) и (4), так
называемый k-тый центральный момент плотности распределения q (x), m , можно вывести из
r k,r
данной плотности распределения. Он связан с взвешенным средним размером частиц x [см.
kr,
формулу (11)].
Полный k-тый центральный момент определяют как:
x
max
k
mx()xq(x)dx (5)
1,r
k,r r

x
min
4.2 Определение среднего размеров частиц по методу представления моментов
Все средние размеров частиц определяют по формуле (6):
k
xM (6)
k,r
k,r
В зависимости от численных значений, выбранных для подстрочных индексов, k и r, можно определить
различные средние размеров частиц. Поскольку средние значения, вычисленные по формуле (6),
могут значительно отличаться, всегда необходимо указывать значения индексов k и r.
Существует две группы средних размеров частиц, которые предпочтительно использовать, как то:
средние арифметические размеры частиц и взвешенные средние размеры частиц.
4.2.1 Терминология для средних размеров частиц в методе представления моментов x
kr,
Таблица 1 представляет примеры терминологии в отношении средних размеров.
Таблица 1 — Терминология для средних размеров частиц x
kr,
Систематический код Терминология
средний арифметический размер
x
1,0
средняя арифметическая площадь
x
2,0
x средний арифметический объем
3,0
взвешенный по (линейному) размеру (диаметру) средний размер
x
1,1
Взвешенный по площади средний размер, средний диаметр по
x
1,2
Соутеру (Sauter)
x Взвешенный по объему средний размер
1,3
4 ISO 2014 – Все права сохраняются

---------------------- Page: 9 ----------------------
ISO 9276-2:2014(R)
4.2.2 Средние арифметические размеры
Средний арифметический размер является взвешенным по количеству средний размер, рассчитанный
по плотности количественного распределения, q (x):
0
k
xM (7)
k,0
k,0
Подсчет отдельных частиц на изображении под микроскопом является типичным примером получения
процентов количеств (r = 0) как основу для усреднения.
В соответствии со Ссылкой [2], рекомендуемыми средними размерами частиц являются:
средний арифметический размер (соответствует средней арифметической длине):
xM (8)
1,0
1,0
[ ]
2
средняя арифметическая площадь (средний арифметический диаметр Хейвуда (Heywood ): средний
диаметр поверхности):
2
xM (9)
2,0
2,0
[ ]
2
средний арифметический объем (средневзвешенный диаметр Хейвуда (Heywood ):
3
xM (10)
3,0
3,0
4.2.3 Взвешенные средние размеры частиц
Взвешенные средние размеры частиц определяют по формуле:
k
xM (11)
k,r
k,r
Взвешивание сит до и после просеивания является типичным примером получения процентов масс
(r = 3) как основы для усреднения. Взвешенные средние размеры частиц представляют абсциссу
центра тяжести q (x) – распределения. Рекомендованные взвешенные средние размеры частиц
r
представлены формулами (12) – (15).
Взвешенный средний размер частиц плотности количественного распределения, q (x), эквивалентен
0
средней арифметической длине [см. формулу (8)]. Он представлен следующим образом:
[ ]
2
Средний арифметический размер (средний диаметр Хейвуда (по количеству) (Heywood ):
xM (12)
1,0
1,0
Взвешенный средний размер частиц плотности распределения длин, q (x), представляют как:
1
взвешенный по (линейному) размеру средний размер (линейный средний диаметр Хейвуда
[ ]
6
(Heywood ):
xM (13)
1,1
1,1
Взвешенный средний размер частиц плотности поверхностного распределения, q (x), представляют
2
как:
ISO 2014 – Все права сохраняются 5

---------------------- Page: 10 ----------------------
ISO 9276-2:2014(R)
[ ]
6
взвешенный по площади средний размер (средний диаметр поверхности Хейвуда (Heywood ):
xM (14)
1,2
1,2
Взвешенный средний размер частиц плотности распределения по объему, q (x), представляют как:
3
[ ]
6
взвешенный по объему средний размер (средний взвешенный диаметр Хейвуда (Heywood ):
xM (15)
1,3
1,3
4.2.4 Средние геометрическиеразмеры частиц
Если распределение размеров частиц удовлетворительно соответствует логарифмически
нормальному распределению размеров (см. ISO 9276-5), то средний геометрический размер частиц
характеризует среднее значение логарифма x. Медиана логарифмически нормального распределения
имеет то же значение, что и средний геометрический размер.
Вместо среднего арифметического, вычисленного как сумма n значений, деленная на их число n,
среднее геометрическое представляет собой корень n-той степени из произведения n значений. В
пересчете на логарифмы, логарифм среднего геометрического рассчитывают как сумму логарифмов
n значений, деленную на их число n. Среднее арифметическое больше, чем среднее геометрическое,
причем расхождение увеличивается с увеличением разброса среди значений.
Математический анализ предельных условий формулы (6) при k стремящимся к нулю (см. также вывод
[ ]
3
для p = q в представлении отношений моментов ) ведет к среднему геометрическому размеру:
x
max
Inxq (x)dx

r
x
min
xex (16)
0,rrgeo,
Или в пересчете на логарифмы:
x
max
InxxIn Inxq(x)dx (17)
0,rrgeo,
r

x
min
На основе данных от гистограммы получаем r-взвешенный средний геометрический размер:
mm
 
 
xxexp Inqxexp InxQx (18)
0,ri i geo,r
ri,ir,i
 
ii11
 
4.2.5 Средние гармонические размеры частиц
Среднее гармоническое ряда значений является обратной величиной среднего арифметического их
[ ]
7
обратных значений . Среднее гармоническое меньше, чем среднее геометрическое, причем
расхождение возрастает с увеличением разброса среди значений. Поэтому r-взвешенный средний
гармонический размер можно рассчитать по формуле:
11
x (19)
har,r
x
min
M
1,r
1
qx()dx
r

x
x
max
6 ISO 2014 – Все права сохраняются

---------------------- Page: 11 ----------------------
ISO 9276-2:2014(R)
или на основе данных гистограммы:
11
x (20)
har,r
mm
11
qxQ
ir,i
ri,
xx
ii
ii11
4.3 Расчет моментов и средние размеры частиц по данному распределению размеров
4.3.1 Расчет M и средние размеры частиц по распределению количеств или объемов
k,r
В большинстве случаев практического применения данные измерений представлены либо плотностью
количественного распределения, q0(x), или плотностью объемного распределения, q3(x). Расчет
[1]
средних размеров частиц, описанный выше, тогда можно выполнить по формуле (21) :
MM
kr,0 k r3,3
k
xMkk (21)
k,r
k,r
MM
rr,0 3,3
Откуда:
M
1,3
xM (22)
2,0
2,0
M
3,3
1
3
xM (23)
3,0 3
3,0
M
3,3
MM
2,0 1,3
xM  (24)
1,1
1,1
MM
1,0 2,3
M
1
3,0
xM  (25)
1,2
1,2
MM
2,0 1,3
M
4,0
xM (26)
1,3
1,3
M
3,0
Из форму (21) – (26) очевидно, что следующие моменты требуются, если необходимо вычислить
средние размеры частиц, определенные выше:
по данной плотности количественного распределения, q (x):  M ;  M ;  M ;  M
0 1,0 2,0 3,0 4,0
по данной плотности объемного распределения, q (x):  M  M ;  M ;  M
3 1,3 –1,3 –2,3 –3,3
Формула (21) показывает, что каждый средний размер частиц или момент данного типа величины
можно выразить как отношение двух количественных моментов, которое используется как основа
метода представления отношений моментовe.
4.3.2 Расчет M по распределению размеров частиц в виде гистограммы
k,r
Если плотность распределения представлена как гистограмма, то q (x , x ) является константой на
r i-1 i
интервале размеров частиц ∆x = x − x .
i i i-1
ISO 2014 – Все права сохраняются 7

---------------------- Page: 12 ----------------------
ISO 9276-2:2014(R)
Поэтому формулу (1) можно переписать следующим образом:
x
max m
kk
Mxqxdx xqx (27)

kr,,r  i ri i

i1
x
min
при
Q
ri,
q  (28)
ri,
xx
i i1
x
max m
kk
Mxqxdx xQ (29)
kr,,r  i ri

i1
x
min
k
Приближенное среднее x в пределах класса распределения размеров можно рассчитать как среднее
i
арифметическое в каждом классе распределения размеров
k

xx
k
ii1
x   (30)
i

2

Альтернативные приближенные средние типа среднего геометрического или интегрального среднего
особых преимуществ не имеют. Ряд исследований показывает (см., например, Ссылку [8]), что не
существует общего преимущества для всех возможных типов распределений.
Дискретный характер данных гистограммы приводит к неопределенностям до нескольких процентов в
вычисленных моментах и средних размерах, связанным с шириной классов распределений размеров,
и соответствующей неопределенности в количестве частиц, выведенном по оцененному среднему
размеру в каждом классе. Методы улучшения разрешения размера в представлении измерения при
наблюдении большего числа классов распределения размеров описаны в ISO 9276-3.
Следовательно, моменты M , M , M , M , M , M , M и M можно рассчитать по
1,0 2,0 3,0 4,0 1,3 –1,3 –2,3 –3,3
формулам (31) – (38):
mm
11
MxqxxQ (31)
1,0ii0, i i 0,i
ii11
mm
22
MxqxxQ (32)
2,0ii0, i i 0,i
ii11
mm
33
MxqxxQ (33)
3,0ii0, i i 0,i
ii11
mm
44
MxqxxQ (34)
4,0ii0, i i 0,i
ii11
8 ISO 2014 – Все права сохраняются

---------------------- Page: 13 ----------------------
ISO 9276-2:2014(R)
mm
11
(35)
MxqxxQ
1,3ii3, i i 3,i
ii11
mm
11
Mqx Q (36)
1,3 3,ii 3,i

11
xx
ii11ii
mm
11
(37)
Mqx Q
2,33,ii 3,i
22
xx
ii11ii
mm
11
(38)
Mqx Q
3,33,ii 3,i
33
xx
ii11ii
4.4 Дисперсия и стандартное отклонение распределения частиц по размерам
Распространение распределения размеров может быть представлено его дисперсией, которая
2
представляет квадрат стандартного отклонения, s . Дисперсия, sr , распределения q (x) - определяют
r r
как:
x
max
2
2
sxx qxdx (39)
1,r

rr

x
min
Вводя полные моменты, можно рассчитать дисперсию (Ссылка [3]) по формуле:
2
2
smMM (40)

rr2, 2,r 1,r
или на основе данных из гистограммы:
22
mm m m
  
22 1 2 1
  
sxq x xqxxQ xQ (41)
ri ri,,i i ri i i r,i i ri,
  
ii11 i1 i1
  
2
Числовой пример расчета sr приведен в Приложении A.
Для логарифмически нормального распределения q (x) стандартное отклонение s можно рассчитать
r
по формуле:
sxln( /x ) ln(x /x ) (42)
84,rr50, 50,r 16,r
Геометрическое стандартное отклонение s получают по формуле:
g
s = exp(s) (43)
g
Следовательно,
sx//x x x (44)
g 84,rr50, 50,r 16,r
ISO 2014 – Все права сохраняются 9

---------------------- Page: 14 ----------------------
ISO 9276-2:2014(R)
4.5 Расчет моментов и средние размеры частиц из логарифмически нормального
распределения
Полный k-тый момент логарифмически нормального распределения вероятностей, qr(x), рассчитывают
по формуле
22
22
kxln 0,5ks
kk0,5s
50,r
Mxee (45)
kr,50,r
Ряд средних размеров частиц можно рассчитать из корня k-того момента в формуле (45) или из
медианы (и среднего геометрического) x , и стандартного отклонения этого распределения по
50,r
формуле (46):
2
2
lnxk0,5s
0,5ks
50,r
xxee (46)
kr,50,r
Медиана x логарифмически нормального распределения имеет такое же значение, как средний
50,r
геометрический размер x .
0,r
4.6 Расчет объемной удельной площади поверхности и среднего диаметра Соутера
Из любого типа распределений можно использовать моменты в расчете объемной удельной площади
поверхности, S , поскольку S обратно пропорциональна среднему взвешенному размеру
V V
поверхности, среднему диаметру Соутера, x (формула 14). Задается выражением:
1,2
6
S  (47)
V
x
1,2
Принимая во внимание формулу (25), из распределений по поверхности, количеству или объему
вытекает:
M
6
2,0
SM66 (48)
V 1,3
MM
1,2 3,0
Для частиц, отличающихся от сфер необходимо вводить коэффициент формы.
5 Метод представления отношения моментов
Моменты являются основой для определения средних размеров и стандартных отклонений
распределений частиц по размерам. Для оценки моментов распределения по размерам большой
совокупности частиц используют случайную выборку, включающую ограниченное число частиц из
этой совокупности. Оценка касается предположения о численных значениях неизвестной совокупности
по значениям выборки. Измерения размеров частиц всегда выполняют на дискретных выборках и
включают ряд дискретных классов распределения размеров. Следовательно, в данной части ISO 9276
рассматриваются только моменты, связанные с выборками.
ПРИМЕЧАНИЕ Плотность распределения и интегральная функция распределения представлены
обозначениями q и Q в серии ISO 9276, но в литературе также используются (Ссылки [3],[4],[7]–[9]) обозначения f и F.
10 ISO 2014 – Все права сохраняются

---------------------- Page: 15 ----------------------
ISO 9276-2:2014(R)
5.1 Определение моментов согласно методу представления отношений мо
...

SLOVENSKI STANDARD
SIST ISO 9276-2:2015
01-marec-2015
1DGRPHãþD
SIST ISO 9276-2:2002
3UHGVWDYLWHYSRGDWNRYGREOMHQLK]JUDQXORPHWULMVNRDQDOL]RGHO,]UDþXQ
SRYSUHþQLKYHOLNRVWLSUHPHURYLQPRPHQWRYL]JUDQXODFLM
Representation of results of particle size analysis - Part 2: Calculation of average particle
sizes/diameters and moments from particle size distributions
Représentation de données obtenues par analyse granulométrique - Partie 2: Calcul des
tailles/diamètres moyens des particules et des moments à partir de distributions
granulométriques
Ta slovenski standard je istoveten z: ISO 9276-2:2014
ICS:
19.120 Analiza velikosti delcev. Particle size analysis. Sieving
Sejanje
SIST ISO 9276-2:2015 en
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------

SIST ISO 9276-2:2015

---------------------- Page: 2 ----------------------

SIST ISO 9276-2:2015
INTERNATIONAL ISO
STANDARD 9276-2
Second edition
2014-05-15
Representation of results of particle
size analysis —
Part 2:
Calculation of average particle sizes/
diameters and moments from particle
size distributions
Représentation de données obtenues par analyse granulométrique —
Partie 2: Calcul des tailles/diamètres moyens des particules et des
moments à partir de distributions granulométriques
Reference number
ISO 9276-2:2014(E)
©
ISO 2014

---------------------- Page: 3 ----------------------

SIST ISO 9276-2:2015
ISO 9276-2:2014(E)

COPYRIGHT PROTECTED DOCUMENT
© ISO 2014
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized otherwise in any form
or by any means, electronic or mechanical, including photocopying, or posting on the internet or an intranet, without prior
written permission. Permission can be requested from either ISO at the address below or ISO’s member body in the country of
the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland
ii © ISO 2014 – All rights reserved

---------------------- Page: 4 ----------------------

SIST ISO 9276-2:2015
ISO 9276-2:2014(E)

Contents Page
Foreword .iv
Introduction .v
1 Scope . 1
2 Normative references . 1
3 Definitions, symbols and abbreviated terms . 1
4 The moment-notation . 3
4.1 Definition of moments according to the moment-notation . 3
4.2 Definition of mean particle sizes according to the moment-notation . 4
4.3 Calculation of moments and mean particle sizes from a given size distribution . 7
4.4 Variance and standard deviation of a particle size distribution . 9
4.5 Calculation of moments and mean particle sizes from a lognormal distribution . 9
4.6 Calculation of volume specific surface area and the Sauter mean diameter .10
5 The moment-ratio-notation .10
5.1 Definition of moments according to the moment-ratio-notation .10
5.2 Definition of mean particle sizes according to the moment-ratio-notation.11
5.3 Calculation of mean particle sizes from a given size distribution .13
5.4 Variance and standard deviation of a particle size distribution .14
5.5 Relationships between mean particle sizes .15
5.6 Calculation of volume specific surface area and the Sauter mean diameter .16
6 Relationship between moment-notation and moment-ratio-notation .16
7 Accuracy of calculated particle size distribution parameters .18
Annex A (informative) Numerical example for calculation of mean particle sizes and standard
deviation from a histogram of a volume based size distribution .19
Annex B (informative) Numerical example for calculation of mean particle sizes and standard
deviation from a histogram of a volume based size distribution .22
Annex C (informative) Accuracy of calculated particle size distribution parameters .25
Bibliography .27
© ISO 2014 – All rights reserved iii

---------------------- Page: 5 ----------------------

SIST ISO 9276-2:2015
ISO 9276-2:2014(E)

Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of any
patent rights identified during the development of the document will be in the Introduction and/or on
the ISO list of patent declarations received (see www.iso.org/patents).
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation on the meaning of ISO specific terms and expressions related to conformity
assessment, as well as information about ISO’s adherence to the WTO principles in the Technical Barriers
to Trade (TBT) see the following URL: Foreword - Supplementary information.
The committee responsible for this document is ISO/TC 24, Particle characterization including sieving,
Subcommittee SC 4, Particle characterization.
This second edition cancels and replaces the first edition (ISO 9276-2:2001), which has been technically
revised.
ISO 9276 consists of the following parts, under the general title Representation of results of particle size
analysis:
— Part 1: Graphical representation
— Part 2: Calculation of average particle sizes/diameters and moments from particle size distributions
— Part 3: Adjustment of an experimental curve to a reference model
— Part 4: Characterization of a classification process
— Part 5: Methods of calculation relating to particle size analyses using logarithmic normal probability
distribution
— Part 6: Descriptive and quantitative representation of particle shape and morphology
iv © ISO 2014 – All rights reserved

---------------------- Page: 6 ----------------------

SIST ISO 9276-2:2015
ISO 9276-2:2014(E)

Introduction
Particle size analysis is often used for characterization of particulate matter. The relationship between
the physical properties of particulate matter, such as powder strength, flowability, dissolution rate,
emulsion/suspension stability and particle size forms always the reason for such characterization. For
materials having a particle size distribution, it is important to use the relevant parameter, a certain
mean particle size, weighted for example by number, area or volume, in the relationship with physical
properties.
This part of ISO 9276 describes two procedures for the use of moments for the calculation of mean sizes,
the spread and other statistical measures of a particle size distribution.
The first method is named moment-notation. The specific utility of the moment-notation is to characterize
size distributions by moments and mean sizes. The moment-notation addresses weighting principles
from physics, especially mechanical engineering, and includes arithmetic means from number based
[1][2]
distributions only as one part .
The second method is named moment-ratio-notation. The moment-ratio-notation is based on a number
[3][4]
statistics and frequencies approach, but includes also conversion to other types of quantities .
Important is that the meaning of the subscripts of mean sizes defined in the moment-notation differs
from the subscripts of mean sizes defined in the moment-ratio-notation. Both notations are linked by a
simple relationship, given in Clause 6.
Both notations are suited for derivation and/or selection of mean sizes related to physical product and
process properties for so-called property functions and process functions. The type of mean size to be
preferred should have a causal relationship with the relevant physical product or process property.
The particle characterization community embraces a very broad spectrum of science disciplines.
The notation of the size distribution employed has been influenced by the branch of industry and the
application and thus no single notation has found universal favour.
There are some particle size dependent properties, like light scattering in certain particle size ranges,
which cannot be characterized by mean particles sizes, derived from simple power law equations of the
[5]
notation systems .
© ISO 2014 – All rights reserved v

---------------------- Page: 7 ----------------------

SIST ISO 9276-2:2015

---------------------- Page: 8 ----------------------

SIST ISO 9276-2:2015
INTERNATIONAL STANDARD ISO 9276-2:2014(E)
Representation of results of particle size analysis —
Part 2:
Calculation of average particle sizes/diameters and
moments from particle size distributions
1 Scope
This part of ISO 9276 provides relevant equations and coherent nomenclatures for the calculation of
moments, mean particle sizes and standard deviations from a given particle size distribution. Two
notation systems in common use are described. One is the method of moments while the second
describes the moment-ratio method. The size distribution may be available as a histogram or as an
analytical function.
The equivalent diameter of a particle of any shape is taken as the size of that particle. Particle shape
factors are not taken into account. It is essential that the measurement technique is stated in the report
in view of the dependency of sizing results of measurement principle. Samples of particles measured are
intended to be representative of the population of particles.
For both notation systems, numerical examples of the calculation of mean particle sizes and standard
deviation from histogram data are presented in an annex.
The accuracy of the mean particle size may be reduced if an incomplete distribution is evaluated. The
accuracy may also be reduced when very limited numbers of size classes are employed.
2 Normative references
The following documents, in whole or in part, are normatively referenced in this document and are
indispensable for its application. For dated references, only the edition cited applies. For undated
references, the latest edition of the referenced document (including any amendments) applies.
ISO 9276-1:1998, Representation of results of particle size analysis — Part 1: Graphical representation
ISO 9276-5:2005, Representation of results of particle size analysis — Part 5: Methods of calculation relating
to particle size analyses using logarithmic normal probability distribution
3 Definitions, symbols and abbreviated terms
If necessary, different symbols are given to the moment-notation (M) and the moment-ratio-notation
(M-R). This serves the purpose of a clear differentiation between the two systems. For both notation
systems, a terminology of specific mean particle sizes is inserted in the corresponding clauses: Clause 4
and Clause 5, respectively.
M-notation M-R-notation Description
i i number of the size class with upper particle size: x (M)
i
or midpoint particle size D (M-R)
i
k power of x
m m number of size classes
© ISO 2014 – All rights reserved 1

---------------------- Page: 9 ----------------------

SIST ISO 9276-2:2015
ISO 9276-2:2014(E)

M-notation M-R-notation Description
r r type of quantity of a distribution (general description)
r = 0,  type of quantity: number
r = 1,  type of quantity: length
r = 2,  type of quantity: surface or projected area
r = 3,  type of quantity: volume or mass
M complete k-th moment of a q (x) – sample distribution
k,r r
m complete k-th central moment of a q (x) – sample distribution
k,r r
M p-th moment of a number distribution density
p
m p-th central moment of a number distribution density
p
N total number of particles in a sample
O order of a mean particle size (O = p + q)
p, q powers of D in moments or subscripts indicating the same
q (x) q (D) distribution density of type of particle quantity r
r r
mean height of a distribution density in the i-th particle size interval,
q
ri,
Δx
i
Q (x) Q (D) cumulative distribution of type of quantity r
r r
ΔQ difference of two values of the cumulative distribution, i.e. relative
r,i
amount in the i-th particle size interval, Δx
i
s s standard deviation of a q (x) and q (D) distribution
r r r r
s s geometric standard deviation of a distribution
g g
s s standard deviation of lognormal distribution (s = ln s )
g
S S surface area
S S volume specific surface area
V V
V V particle volume
mean particle volume
V
x D particle size, diameter of an equivalent sphere
x upper particle size of the i-th particle size interval
i
x lower particle size of the i-th particle size interval
i-1
D midpoint size of the i-th size class
i
x particle size below which there are no particles in a given size distri-
min
bution
2 © ISO 2014 – All rights reserved

---------------------- Page: 10 ----------------------

SIST ISO 9276-2:2015
ISO 9276-2:2014(E)

M-notation M-R-notation Description
x particle size above which there are no particles in a given size distri-
max
bution
mean particle sizes (general description)
x D
kr, pq,
geometric mean particle sizes
D
pp,
arithmetic mean particle size
x
k,0
weighted mean particle size
x
kr,
geometric mean particle size
x
geo,r
harmonic mean particle size
x
har,r
x median particle size of a cumulative volume distribution
50,3
Δx = x – x width of the i-th particle size interval
i i i-1
4 The moment-notation
Moments are the basis for defining mean sizes and standard deviations of particle size distributions.
A random sample, containing a limited number of particles from a large population of particle sizes, is
used for estimation of the moments of the size distribution of that population. Estimation is concerned
with inference about the numerical values of the unknown population from those of the sample. Particle
size measurements are always done on discrete samples and involve a number of discrete size classes.
Therefore, only moments related to samples are dealt with in this part of ISO 9276.
4.1 Definition of moments according to the moment-notation
[1]
The complete k-th moment of a distribution density is represented by integrals as defined in
Formula (1). M stands for moment. The first subscript, k, of M indicates the power of the particle size x,
the second subscript, r, of M describes the type of quantity of the distribution density.
x
max
k
Mx= qx()dx (1)
k,r r

x
min
If r = 0, q (x) represents a number distribution density, if r = 3, q (x) represents a volume or mass
0 3
distribution density.
Formula (1) describes a complete moment if the integral boundaries are represented by the minimum
particle size (x ) and the maximum particle size (x ).
min max
© ISO 2014 – All rights reserved 3

---------------------- Page: 11 ----------------------

SIST ISO 9276-2:2015
ISO 9276-2:2014(E)

A special complete moment is represented by M :
0,r
x x
maxx max
0
Mx==qx()ddxq ()xx =−Qx() Qx()=1 (2)
0,r rr rrmaxmin
∫ ∫
x x
min min
with
x
i
Qx()= qx()dx (3)
ri r

x
min
A moment is incomplete, if the integration is performed between two arbitrary particle diameters x
i-1
and x within the given size range of a distribution:
i
x < x < x < x < x
min i-1 i max
x
i
k
Mx(,xx)(= qx)dx (4)
k,ri−1 i r

x
i−1
Apart from the moments related to the origin of the particle size axis and shown in Formulae (1) and (4),
the so-called k-th central moment of a q (x) – distribution density, m , can be derived from a given
r k,r
distribution density. It is related to the weighted mean particle size x [see Formula (11)].
kr,
The complete k-th central moment is defined as:
x
max
k
mx=−()xq ()xxd (5)
1,r
k,r r

x
min
4.2 Definition of mean particle sizes according to the moment-notation
All mean particle sizes are defined by Formula (6):
k
xM= (6)
k,r
k,r
Depending on the numbers chosen for the subscripts, k and r, different mean particle sizes may be defined.
Since the mean particle sizes calculated from Formula (6) may differ considerably, the subscripts k and
r should always be quoted.
There are two groups of mean particle sizes, which should preferably be used, viz. arithmetic mean
particle sizes and weighted mean particle sizes.
4.2.1 Terminology for mean particle sizes in the moment-notation x
kr,
Table 1 presents terminology examples of mean sizes.
4 © ISO 2014 – All rights reserved

---------------------- Page: 12 ----------------------

SIST ISO 9276-2:2015
ISO 9276-2:2014(E)

Table 1 — Terminology for mean particle sizes x
kr,
Systematic code Terminology
x arithmetic mean size
1,0

x arithmetic mean area size
2,0

x arithmetic mean volume size
3,0

x size-weighted mean size
1,1

x area-weighted mean size, Sauter mean diameter
1,2

x volume-weighted mean size
1,3

4.2.2 Arithmetic mean particle sizes
Arithmetic mean size is a number-weighted mean size, calculated from a number distribution density,
q (x):
0
k
xM= (7)
k,0
k,0
Counting single particles in a microscope image is a typical example to obtain number (r = 0) percentages
as basis of averaging.
In accordance with Reference [2], the recommended mean particle sizes are:
arithmetic mean size (corresponds to arithmetic mean length size):
xM= (8)
10,
10,
[2]
arithmetic mean area size (Heywood : mean surface diameter):
2
xM= (9)
20,
20,
[2]
arithmetic mean volume size (Heywood : mean-weight diameter):
3
xM= (10)
30,
30,
4.2.3 Weighted mean particle sizes
Weighted mean particle sizes are defined by:
k
xM= (11)
k,r
k,r
Weighing sieves before and after sieving is a typical example to obtain mass (r = 3) percentages as basis
of averaging. Weighted mean particle sizes represent the abscissa of the centre of gravity of a q (x) –
r
distribution. The recommended weighted mean particle sizes are represented by Formulae (12) to (15).
The weighted mean particle size of a number distribution density, q (x), is equivalent to the arithmetic
0
mean length size [see Formula (8)]. It is represented by:
© ISO 2014 – All rights reserved 5

---------------------- Page: 13 ----------------------

SIST ISO 9276-2:2015
ISO 9276-2:2014(E)

[6]
arithmetic mean size (Heywood : numerical mean diameter):
xM= (12)
10,
10,
The weighted mean particle size of a length distribution density, q (x), is given by:
1
[6]
size-weighted mean size (Heywood : linear mean diameter):
xM= (13)
11,
11,
The weighted mean particle size of a surface distribution density, q (x), is represented by:
2
[6]
area-weighted mean size (Heywood : surface mean diameter):
xM= (14)
12,
12,
The weighted mean particle size of a volume distribution density, q (x), is given by:
3
[6]
volume-weighted mean size (Heywood : weight mean diameter):
xM= (15)
13,
13,
4.2.4 Geometric mean particle sizes
If a particle size distribution conforms satisfactorily to a lognormal size distribution (see ISO 9276-5),
the geometric mean particle size characterizes the mean value of the logarithm of x. The median of a
lognormal distribution has the same value as the geometric mean size.
Instead of the arithmetic mean, calculated from the sum of n values, divided by their number n, the
geometric mean is the n-th root of the product of n values. In terms of logarithms the logarithm of the
geometric mean is calculated from the sum of the logarithms of n values, divided by their number n.
The arithmetic mean is greater than the geometric, the inequality increasing the greater the dispersion
among the values.
Mathematical limit analysis of Formula (6) with k approaching zero (see also derivation for p = q in
[3]
moment-ratio-notation ) leads to the geometric mean size:
x
max
In xq ()xdx

r
x
min
xe==x (16)
0,r geo,r
or in terms of logarithms:
x
max
Inxx==In Inxq ()xdx (17)
0,rrgeo,
r

x
min
Based on data from a histogram one obtains the r-weighted geometric mean size:
m m
   
xx=expe In qxΔΔ= xp InxQ ==x (18)
0,ri i geo,r
∑∑ri, i ri,
   
i==11i
   
6 © ISO 2014 – All rights reserved

---------------------- Page: 14 ----------------------

SIST ISO 9276-2:2015
ISO 9276-2:2014(E)

4.2.5 Harmonic mean particle sizes
[7]
The harmonic mean of a series of values is the reciprocal of the arithmetic mean of their reciprocals .
The harmonic mean is smaller than the geometric, the inequality increasing the greater the dispersion
among the values. Therefore the r-weighted harmonic mean size can be calculated from:
1 1
x == (19)
harr,
x
min M
−1,r
1
qx()dx
r

x
x
max
or based on data from a histogram:
1 1
x == (20)
harr,
m m
1 1
qxΔΔQ
i ri,
∑∑ri,
x x
i i
i==11i
4.3 Calculation of moments and mean particle sizes from a given size distribution
4.3.1 The calculation of M and mean particle sizes from a number or a volume based size
k,r
distribution
In many cases of practical application, the measured data are either represented by a number distribution
density, q (x), or a volume distribution density, q (x). The calculation of the mean particle sizes described
0 3
[1]
above can then be performed according to Formula (21) :
M M
kr++,0 kr−3,3
k
xM== k =k (21)
k,r
k,r
M M
r,0 r−3,3
This leads to:
M
−13,
xM== (22)
20,
20,
M
−33,
1
3
xM== 3 (23)
30,
30,
M
−33,
M M
20, −13,
xM== = (24)
11,
11,
M M
10, −23,
M
1
30,
xM== = (25)
12,
12,
MM
20,,−13
M
40,
xM== (26)
13,
13,
M
30,
One realizes from Formulae (21) to (26), that the following moments are needed if the mean particle
sizes defined above are to be calculated:
from a given number distribution density, q (x):  M ;  M ;  M ;  M
0 1,0 2,0 3,0 4,0
from a given volume distribution density, q (x):  M  M ;  M ;  M
3 1,3 –1,3 –2,3 –3,3
Formula (21) shows, that each mean particle size or moment of a given type of quantity can be expressed
as a ratio of two number based moments, which is used as the main approach of the moment-ratio-
notation.
© ISO 2014 – All rights reserved 7

---------------------- Page: 15 ----------------------

SIST ISO 9276-2:2015
ISO 9276-2:2014(E)

4.3.2 Calculation of M from a particle size distribution, given as a histogram
k,r
If a distribution density is given as a histogram, q (x , x ) is constant in the particle size interval
r i-1 i
Δx = x − x .
i i i-1
Formula (1) may therefore be rewritten as follows:
x
max m
k k
Mx= qx dx = xq Δx (27)
()
kr,,r ∑ i ri i

i=1
x
min
One obtains with
ΔQ
ri,
q = (28)
ri,
xx−
i i1−
x
max
m
k k
Mx= qx dx = xQΔ (29)
()
kr,,r i ri


i=1
x
min
k
The approximate mean x within a size class can be calculated as arithmetic mean in each size class
i
k
 
xx+
k
ii−1
x =   (30)
i
 
2
 
Alternative approximate means like geometric mean or an integral mean have not specific advantage.
Several investigations have shown (see e.g. Reference [8]), that there is no general preference for all
types of distributions possible.
The discrete nature of histogram data causes uncertainties up to a few percent in the calculated moments
and mean sizes, which relate to the width of the size classes and the corresponding uncertainty in
the amount of particles derived from the estimated mean size in each class. Methods to improve the
size resolution of the representation of measurement by observation of more size classes are given in
ISO 9276-3.
The moments M , M , M , M , M , M , M and M can therefore be calculated from
1,0 2,0 3,0 4,0 1,3 –1,3 –2,3 –3,3
Formulae (31) to (38):
m m
1 1
Mx==qxΔΔxQ (31)
10,,ii0 i ii0,
∑∑
i==1 i 1
m m
2 2
Mx==qxΔΔxQ (32)
20,,∑∑ii0 i ii0,
i==1 i 1
m m
3 3
Mx==qxΔΔxQ (33)
30,,ii0 i ii0,
∑∑
i==1 i 1
m m
4 4
Mx==qxΔΔxQ (34)
40,,∑∑ii0 i ii0,
i==1 i 1
m m
1 1
Mx==qxΔΔxQ (35)
13,,ii3 i ii3,
∑∑
i==1 i 1
m m
11
M ==qxΔΔQ (36)
−13,,∑∑3ii 3,i
1 1
x x
i==1 i i 1 i
8 © ISO 2014 – All rights reserved

---------------------- Page: 16 ----------------------

SIST ISO 9276-2:2015
ISO 9276-2:2014(E)

m m
11
M ==qxΔΔQ (37)
−23,,∑∑3ii 3,i
2 2
x x
i==1 i i 1 i
m m
11
M ==qxΔΔQ (38)
−33,,∑∑3ii 3,i
3 3
x x
i==1 i i 1 i
4.4 Variance and standard deviation of a particle size distribution
The spread of a size distribution may be represented by its variance, which represents the square of the
2
standard deviation, s . The variance, s , of a q (x) – distribution is defined as:
r r r
x
max
2
2
sx=−xq xdx (39)
1,r ()
()
r r

x
min
Introducing complete moments, the variance can be calculated (Reference [3]) from:
2
2
sm==MM− (40)
()
rr22,,rr1,
or based on data from a histogram:
2 2
m m m m
   
22 1 2 1
   
sx=−qxΔ xq ΔΔxx= Q −− xQΔ (41)
ri∑∑ri,,i ir ii ∑ ir,i ∑ ir,i
   
i==1 i 1 i=1 i=1
   
2
A numerical example of the calculation of s is given in Annex A.
r
For a lognormal q (x) – distribution, the standard deviation s can be calculated from:
r
sx==ln(/xx)ln( /)x (42)
84,,rr50 50,,rr16
The geometrical standard deviation s is obtained from:
g
s = exp(s)
g
(43)
Hence,
sx==//xx x (44)
g 84,,rr50 50,,rr16
4.5 Calculation of moments and mean particle sizes from a lognormal distribution
The complete k-th moment of a lognormal probability distribution, q (x), calculates to
r
22
22
kxln +05, ks
kk05, s
50,r
Mx==ee (45)
kr,,50r
A series of mean particle sizes can be calculated from the k-th root of the k-th moment in Formula (45)
or from the median (and geometric mean) x and the standard deviation of that distribution using
50,r
Formula (46):
2
2
ln xk+05, s
05, ks
50,r
xx==ee (46)
kr,,50 r
The median x of a lognormal distribution has the same value as the geometric mean size x .
50,r
0,r
© ISO 2014 – All rights reserved 9

---------------------- Page: 17 ----------------------

SIST ISO 9276-2:2015
ISO 9276-2:2014(E)

4.6 Calculation of volume specific surface area and the Sauter mean diameter
From distributions of any type of quantity moments can be used in the calculation of the volume specific
surface area, S , since S is inversely proportional to the weighted mean surface size, the Sauter mean
V V
diameter, x (Formula 14). It is given by:
12,
6
S = (47)
V
x
12,
Taking into account Formula (25), one arrives from surface, number or volume distributions at:
M
6
20,
S ==66=⋅M (48)
V −13,
M M
12, 30,
For particles other than spheres a shape factor shall be introduced.
5 The moment-ratio-notation
Moments are the basis for defining mean sizes and standard deviations of particle size distributions.
A random sample, containing a limited number of particles from a large population of particle sizes, is
used for estimation of the moments of the size distribution of that population. Estimation is concerned
with inference about the numerical values of the unknown population from those of the sample. Particle
size measurements are always done on discrete samples and involve a number of discrete size classes.
Therefore, only moments related to
...

INTERNATIONAL ISO
STANDARD 9276-2
Second edition
2014-05-15
Representation of results of particle
size analysis —
Part 2:
Calculation of average particle sizes/
diameters and moments from particle
size distributions
Représentation de données obtenues par analyse granulométrique —
Partie 2: Calcul des tailles/diamètres moyens des particules et des
moments à partir de distributions granulométriques
Reference number
ISO 9276-2:2014(E)
©
ISO 2014

---------------------- Page: 1 ----------------------
ISO 9276-2:2014(E)

COPYRIGHT PROTECTED DOCUMENT
© ISO 2014
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized otherwise in any form
or by any means, electronic or mechanical, including photocopying, or posting on the internet or an intranet, without prior
written permission. Permission can be requested from either ISO at the address below or ISO’s member body in the country of
the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland
ii © ISO 2014 – All rights reserved

---------------------- Page: 2 ----------------------
ISO 9276-2:2014(E)

Contents Page
Foreword .iv
Introduction .v
1 Scope . 1
2 Normative references . 1
3 Definitions, symbols and abbreviated terms . 1
4 The moment-notation . 3
4.1 Definition of moments according to the moment-notation . 3
4.2 Definition of mean particle sizes according to the moment-notation . 4
4.3 Calculation of moments and mean particle sizes from a given size distribution . 7
4.4 Variance and standard deviation of a particle size distribution . 9
4.5 Calculation of moments and mean particle sizes from a lognormal distribution . 9
4.6 Calculation of volume specific surface area and the Sauter mean diameter .10
5 The moment-ratio-notation .10
5.1 Definition of moments according to the moment-ratio-notation .10
5.2 Definition of mean particle sizes according to the moment-ratio-notation.11
5.3 Calculation of mean particle sizes from a given size distribution .13
5.4 Variance and standard deviation of a particle size distribution .14
5.5 Relationships between mean particle sizes .15
5.6 Calculation of volume specific surface area and the Sauter mean diameter .16
6 Relationship between moment-notation and moment-ratio-notation .16
7 Accuracy of calculated particle size distribution parameters .18
Annex A (informative) Numerical example for calculation of mean particle sizes and standard
deviation from a histogram of a volume based size distribution .19
Annex B (informative) Numerical example for calculation of mean particle sizes and standard
deviation from a histogram of a volume based size distribution .22
Annex C (informative) Accuracy of calculated particle size distribution parameters .25
Bibliography .27
© ISO 2014 – All rights reserved iii

---------------------- Page: 3 ----------------------
ISO 9276-2:2014(E)

Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of any
patent rights identified during the development of the document will be in the Introduction and/or on
the ISO list of patent declarations received (see www.iso.org/patents).
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation on the meaning of ISO specific terms and expressions related to conformity
assessment, as well as information about ISO’s adherence to the WTO principles in the Technical Barriers
to Trade (TBT) see the following URL: Foreword - Supplementary information.
The committee responsible for this document is ISO/TC 24, Particle characterization including sieving,
Subcommittee SC 4, Particle characterization.
This second edition cancels and replaces the first edition (ISO 9276-2:2001), which has been technically
revised.
ISO 9276 consists of the following parts, under the general title Representation of results of particle size
analysis:
— Part 1: Graphical representation
— Part 2: Calculation of average particle sizes/diameters and moments from particle size distributions
— Part 3: Adjustment of an experimental curve to a reference model
— Part 4: Characterization of a classification process
— Part 5: Methods of calculation relating to particle size analyses using logarithmic normal probability
distribution
— Part 6: Descriptive and quantitative representation of particle shape and morphology
iv © ISO 2014 – All rights reserved

---------------------- Page: 4 ----------------------
ISO 9276-2:2014(E)

Introduction
Particle size analysis is often used for characterization of particulate matter. The relationship between
the physical properties of particulate matter, such as powder strength, flowability, dissolution rate,
emulsion/suspension stability and particle size forms always the reason for such characterization. For
materials having a particle size distribution, it is important to use the relevant parameter, a certain
mean particle size, weighted for example by number, area or volume, in the relationship with physical
properties.
This part of ISO 9276 describes two procedures for the use of moments for the calculation of mean sizes,
the spread and other statistical measures of a particle size distribution.
The first method is named moment-notation. The specific utility of the moment-notation is to characterize
size distributions by moments and mean sizes. The moment-notation addresses weighting principles
from physics, especially mechanical engineering, and includes arithmetic means from number based
[1][2]
distributions only as one part .
The second method is named moment-ratio-notation. The moment-ratio-notation is based on a number
[3][4]
statistics and frequencies approach, but includes also conversion to other types of quantities .
Important is that the meaning of the subscripts of mean sizes defined in the moment-notation differs
from the subscripts of mean sizes defined in the moment-ratio-notation. Both notations are linked by a
simple relationship, given in Clause 6.
Both notations are suited for derivation and/or selection of mean sizes related to physical product and
process properties for so-called property functions and process functions. The type of mean size to be
preferred should have a causal relationship with the relevant physical product or process property.
The particle characterization community embraces a very broad spectrum of science disciplines.
The notation of the size distribution employed has been influenced by the branch of industry and the
application and thus no single notation has found universal favour.
There are some particle size dependent properties, like light scattering in certain particle size ranges,
which cannot be characterized by mean particles sizes, derived from simple power law equations of the
[5]
notation systems .
© ISO 2014 – All rights reserved v

---------------------- Page: 5 ----------------------
INTERNATIONAL STANDARD ISO 9276-2:2014(E)
Representation of results of particle size analysis —
Part 2:
Calculation of average particle sizes/diameters and
moments from particle size distributions
1 Scope
This part of ISO 9276 provides relevant equations and coherent nomenclatures for the calculation of
moments, mean particle sizes and standard deviations from a given particle size distribution. Two
notation systems in common use are described. One is the method of moments while the second
describes the moment-ratio method. The size distribution may be available as a histogram or as an
analytical function.
The equivalent diameter of a particle of any shape is taken as the size of that particle. Particle shape
factors are not taken into account. It is essential that the measurement technique is stated in the report
in view of the dependency of sizing results of measurement principle. Samples of particles measured are
intended to be representative of the population of particles.
For both notation systems, numerical examples of the calculation of mean particle sizes and standard
deviation from histogram data are presented in an annex.
The accuracy of the mean particle size may be reduced if an incomplete distribution is evaluated. The
accuracy may also be reduced when very limited numbers of size classes are employed.
2 Normative references
The following documents, in whole or in part, are normatively referenced in this document and are
indispensable for its application. For dated references, only the edition cited applies. For undated
references, the latest edition of the referenced document (including any amendments) applies.
ISO 9276-1:1998, Representation of results of particle size analysis — Part 1: Graphical representation
ISO 9276-5:2005, Representation of results of particle size analysis — Part 5: Methods of calculation relating
to particle size analyses using logarithmic normal probability distribution
3 Definitions, symbols and abbreviated terms
If necessary, different symbols are given to the moment-notation (M) and the moment-ratio-notation
(M-R). This serves the purpose of a clear differentiation between the two systems. For both notation
systems, a terminology of specific mean particle sizes is inserted in the corresponding clauses: Clause 4
and Clause 5, respectively.
M-notation M-R-notation Description
i i number of the size class with upper particle size: x (M)
i
or midpoint particle size D (M-R)
i
k power of x
m m number of size classes
© ISO 2014 – All rights reserved 1

---------------------- Page: 6 ----------------------
ISO 9276-2:2014(E)

M-notation M-R-notation Description
r r type of quantity of a distribution (general description)
r = 0,  type of quantity: number
r = 1,  type of quantity: length
r = 2,  type of quantity: surface or projected area
r = 3,  type of quantity: volume or mass
M complete k-th moment of a q (x) – sample distribution
k,r r
m complete k-th central moment of a q (x) – sample distribution
k,r r
M p-th moment of a number distribution density
p
m p-th central moment of a number distribution density
p
N total number of particles in a sample
O order of a mean particle size (O = p + q)
p, q powers of D in moments or subscripts indicating the same
q (x) q (D) distribution density of type of particle quantity r
r r
mean height of a distribution density in the i-th particle size interval,
q
ri,
Δx
i
Q (x) Q (D) cumulative distribution of type of quantity r
r r
ΔQ difference of two values of the cumulative distribution, i.e. relative
r,i
amount in the i-th particle size interval, Δx
i
s s standard deviation of a q (x) and q (D) distribution
r r r r
s s geometric standard deviation of a distribution
g g
s s standard deviation of lognormal distribution (s = ln s )
g
S S surface area
S S volume specific surface area
V V
V V particle volume
mean particle volume
V
x D particle size, diameter of an equivalent sphere
x upper particle size of the i-th particle size interval
i
x lower particle size of the i-th particle size interval
i-1
D midpoint size of the i-th size class
i
x particle size below which there are no particles in a given size distri-
min
bution
2 © ISO 2014 – All rights reserved

---------------------- Page: 7 ----------------------
ISO 9276-2:2014(E)

M-notation M-R-notation Description
x particle size above which there are no particles in a given size distri-
max
bution
mean particle sizes (general description)
x D
kr, pq,
geometric mean particle sizes
D
pp,
arithmetic mean particle size
x
k,0
weighted mean particle size
x
kr,
geometric mean particle size
x
geo,r
harmonic mean particle size
x
har,r
x median particle size of a cumulative volume distribution
50,3
Δx = x – x width of the i-th particle size interval
i i i-1
4 The moment-notation
Moments are the basis for defining mean sizes and standard deviations of particle size distributions.
A random sample, containing a limited number of particles from a large population of particle sizes, is
used for estimation of the moments of the size distribution of that population. Estimation is concerned
with inference about the numerical values of the unknown population from those of the sample. Particle
size measurements are always done on discrete samples and involve a number of discrete size classes.
Therefore, only moments related to samples are dealt with in this part of ISO 9276.
4.1 Definition of moments according to the moment-notation
[1]
The complete k-th moment of a distribution density is represented by integrals as defined in
Formula (1). M stands for moment. The first subscript, k, of M indicates the power of the particle size x,
the second subscript, r, of M describes the type of quantity of the distribution density.
x
max
k
Mx= qx()dx (1)
k,r r

x
min
If r = 0, q (x) represents a number distribution density, if r = 3, q (x) represents a volume or mass
0 3
distribution density.
Formula (1) describes a complete moment if the integral boundaries are represented by the minimum
particle size (x ) and the maximum particle size (x ).
min max
© ISO 2014 – All rights reserved 3

---------------------- Page: 8 ----------------------
ISO 9276-2:2014(E)

A special complete moment is represented by M :
0,r
x x
maxx max
0
Mx==qx()ddxq ()xx =−Qx() Qx()=1 (2)
0,r rr rrmaxmin
∫ ∫
x x
min min
with
x
i
Qx()= qx()dx (3)
ri r

x
min
A moment is incomplete, if the integration is performed between two arbitrary particle diameters x
i-1
and x within the given size range of a distribution:
i
x < x < x < x < x
min i-1 i max
x
i
k
Mx(,xx)(= qx)dx (4)
k,ri−1 i r

x
i−1
Apart from the moments related to the origin of the particle size axis and shown in Formulae (1) and (4),
the so-called k-th central moment of a q (x) – distribution density, m , can be derived from a given
r k,r
distribution density. It is related to the weighted mean particle size x [see Formula (11)].
kr,
The complete k-th central moment is defined as:
x
max
k
mx=−()xq ()xxd (5)
1,r
k,r r

x
min
4.2 Definition of mean particle sizes according to the moment-notation
All mean particle sizes are defined by Formula (6):
k
xM= (6)
k,r
k,r
Depending on the numbers chosen for the subscripts, k and r, different mean particle sizes may be defined.
Since the mean particle sizes calculated from Formula (6) may differ considerably, the subscripts k and
r should always be quoted.
There are two groups of mean particle sizes, which should preferably be used, viz. arithmetic mean
particle sizes and weighted mean particle sizes.
4.2.1 Terminology for mean particle sizes in the moment-notation x
kr,
Table 1 presents terminology examples of mean sizes.
4 © ISO 2014 – All rights reserved

---------------------- Page: 9 ----------------------
ISO 9276-2:2014(E)

Table 1 — Terminology for mean particle sizes x
kr,
Systematic code Terminology
x arithmetic mean size
1,0

x arithmetic mean area size
2,0

x arithmetic mean volume size
3,0

x size-weighted mean size
1,1

x area-weighted mean size, Sauter mean diameter
1,2

x volume-weighted mean size
1,3

4.2.2 Arithmetic mean particle sizes
Arithmetic mean size is a number-weighted mean size, calculated from a number distribution density,
q (x):
0
k
xM= (7)
k,0
k,0
Counting single particles in a microscope image is a typical example to obtain number (r = 0) percentages
as basis of averaging.
In accordance with Reference [2], the recommended mean particle sizes are:
arithmetic mean size (corresponds to arithmetic mean length size):
xM= (8)
10,
10,
[2]
arithmetic mean area size (Heywood : mean surface diameter):
2
xM= (9)
20,
20,
[2]
arithmetic mean volume size (Heywood : mean-weight diameter):
3
xM= (10)
30,
30,
4.2.3 Weighted mean particle sizes
Weighted mean particle sizes are defined by:
k
xM= (11)
k,r
k,r
Weighing sieves before and after sieving is a typical example to obtain mass (r = 3) percentages as basis
of averaging. Weighted mean particle sizes represent the abscissa of the centre of gravity of a q (x) –
r
distribution. The recommended weighted mean particle sizes are represented by Formulae (12) to (15).
The weighted mean particle size of a number distribution density, q (x), is equivalent to the arithmetic
0
mean length size [see Formula (8)]. It is represented by:
© ISO 2014 – All rights reserved 5

---------------------- Page: 10 ----------------------
ISO 9276-2:2014(E)

[6]
arithmetic mean size (Heywood : numerical mean diameter):
xM= (12)
10,
10,
The weighted mean particle size of a length distribution density, q (x), is given by:
1
[6]
size-weighted mean size (Heywood : linear mean diameter):
xM= (13)
11,
11,
The weighted mean particle size of a surface distribution density, q (x), is represented by:
2
[6]
area-weighted mean size (Heywood : surface mean diameter):
xM= (14)
12,
12,
The weighted mean particle size of a volume distribution density, q (x), is given by:
3
[6]
volume-weighted mean size (Heywood : weight mean diameter):
xM= (15)
13,
13,
4.2.4 Geometric mean particle sizes
If a particle size distribution conforms satisfactorily to a lognormal size distribution (see ISO 9276-5),
the geometric mean particle size characterizes the mean value of the logarithm of x. The median of a
lognormal distribution has the same value as the geometric mean size.
Instead of the arithmetic mean, calculated from the sum of n values, divided by their number n, the
geometric mean is the n-th root of the product of n values. In terms of logarithms the logarithm of the
geometric mean is calculated from the sum of the logarithms of n values, divided by their number n.
The arithmetic mean is greater than the geometric, the inequality increasing the greater the dispersion
among the values.
Mathematical limit analysis of Formula (6) with k approaching zero (see also derivation for p = q in
[3]
moment-ratio-notation ) leads to the geometric mean size:
x
max
In xq ()xdx

r
x
min
xe==x (16)
0,r geo,r
or in terms of logarithms:
x
max
Inxx==In Inxq ()xdx (17)
0,rrgeo,
r

x
min
Based on data from a histogram one obtains the r-weighted geometric mean size:
m m
   
xx=expe In qxΔΔ= xp InxQ ==x (18)
0,ri i geo,r
∑∑ri, i ri,
   
i==11i
   
6 © ISO 2014 – All rights reserved

---------------------- Page: 11 ----------------------
ISO 9276-2:2014(E)

4.2.5 Harmonic mean particle sizes
[7]
The harmonic mean of a series of values is the reciprocal of the arithmetic mean of their reciprocals .
The harmonic mean is smaller than the geometric, the inequality increasing the greater the dispersion
among the values. Therefore the r-weighted harmonic mean size can be calculated from:
1 1
x == (19)
harr,
x
min M
−1,r
1
qx()dx
r

x
x
max
or based on data from a histogram:
1 1
x == (20)
harr,
m m
1 1
qxΔΔQ
i ri,
∑∑ri,
x x
i i
i==11i
4.3 Calculation of moments and mean particle sizes from a given size distribution
4.3.1 The calculation of M and mean particle sizes from a number or a volume based size
k,r
distribution
In many cases of practical application, the measured data are either represented by a number distribution
density, q (x), or a volume distribution density, q (x). The calculation of the mean particle sizes described
0 3
[1]
above can then be performed according to Formula (21) :
M M
kr++,0 kr−3,3
k
xM== k =k (21)
k,r
k,r
M M
r,0 r−3,3
This leads to:
M
−13,
xM== (22)
20,
20,
M
−33,
1
3
xM== 3 (23)
30,
30,
M
−33,
M M
20, −13,
xM== = (24)
11,
11,
M M
10, −23,
M
1
30,
xM== = (25)
12,
12,
MM
20,,−13
M
40,
xM== (26)
13,
13,
M
30,
One realizes from Formulae (21) to (26), that the following moments are needed if the mean particle
sizes defined above are to be calculated:
from a given number distribution density, q (x):  M ;  M ;  M ;  M
0 1,0 2,0 3,0 4,0
from a given volume distribution density, q (x):  M  M ;  M ;  M
3 1,3 –1,3 –2,3 –3,3
Formula (21) shows, that each mean particle size or moment of a given type of quantity can be expressed
as a ratio of two number based moments, which is used as the main approach of the moment-ratio-
notation.
© ISO 2014 – All rights reserved 7

---------------------- Page: 12 ----------------------
ISO 9276-2:2014(E)

4.3.2 Calculation of M from a particle size distribution, given as a histogram
k,r
If a distribution density is given as a histogram, q (x , x ) is constant in the particle size interval
r i-1 i
Δx = x − x .
i i i-1
Formula (1) may therefore be rewritten as follows:
x
max m
k k
Mx= qx dx = xq Δx (27)
()
kr,,r ∑ i ri i

i=1
x
min
One obtains with
ΔQ
ri,
q = (28)
ri,
xx−
i i1−
x
max
m
k k
Mx= qx dx = xQΔ (29)
()
kr,,r i ri


i=1
x
min
k
The approximate mean x within a size class can be calculated as arithmetic mean in each size class
i
k
 
xx+
k
ii−1
x =   (30)
i
 
2
 
Alternative approximate means like geometric mean or an integral mean have not specific advantage.
Several investigations have shown (see e.g. Reference [8]), that there is no general preference for all
types of distributions possible.
The discrete nature of histogram data causes uncertainties up to a few percent in the calculated moments
and mean sizes, which relate to the width of the size classes and the corresponding uncertainty in
the amount of particles derived from the estimated mean size in each class. Methods to improve the
size resolution of the representation of measurement by observation of more size classes are given in
ISO 9276-3.
The moments M , M , M , M , M , M , M and M can therefore be calculated from
1,0 2,0 3,0 4,0 1,3 –1,3 –2,3 –3,3
Formulae (31) to (38):
m m
1 1
Mx==qxΔΔxQ (31)
10,,ii0 i ii0,
∑∑
i==1 i 1
m m
2 2
Mx==qxΔΔxQ (32)
20,,∑∑ii0 i ii0,
i==1 i 1
m m
3 3
Mx==qxΔΔxQ (33)
30,,ii0 i ii0,
∑∑
i==1 i 1
m m
4 4
Mx==qxΔΔxQ (34)
40,,∑∑ii0 i ii0,
i==1 i 1
m m
1 1
Mx==qxΔΔxQ (35)
13,,ii3 i ii3,
∑∑
i==1 i 1
m m
11
M ==qxΔΔQ (36)
−13,,∑∑3ii 3,i
1 1
x x
i==1 i i 1 i
8 © ISO 2014 – All rights reserved

---------------------- Page: 13 ----------------------
ISO 9276-2:2014(E)

m m
11
M ==qxΔΔQ (37)
−23,,∑∑3ii 3,i
2 2
x x
i==1 i i 1 i
m m
11
M ==qxΔΔQ (38)
−33,,∑∑3ii 3,i
3 3
x x
i==1 i i 1 i
4.4 Variance and standard deviation of a particle size distribution
The spread of a size distribution may be represented by its variance, which represents the square of the
2
standard deviation, s . The variance, s , of a q (x) – distribution is defined as:
r r r
x
max
2
2
sx=−xq xdx (39)
1,r ()
()
r r

x
min
Introducing complete moments, the variance can be calculated (Reference [3]) from:
2
2
sm==MM− (40)
()
rr22,,rr1,
or based on data from a histogram:
2 2
m m m m
   
22 1 2 1
   
sx=−qxΔ xq ΔΔxx= Q −− xQΔ (41)
ri∑∑ri,,i ir ii ∑ ir,i ∑ ir,i
   
i==1 i 1 i=1 i=1
   
2
A numerical example of the calculation of s is given in Annex A.
r
For a lognormal q (x) – distribution, the standard deviation s can be calculated from:
r
sx==ln(/xx)ln( /)x (42)
84,,rr50 50,,rr16
The geometrical standard deviation s is obtained from:
g
s = exp(s)
g
(43)
Hence,
sx==//xx x (44)
g 84,,rr50 50,,rr16
4.5 Calculation of moments and mean particle sizes from a lognormal distribution
The complete k-th moment of a lognormal probability distribution, q (x), calculates to
r
22
22
kxln +05, ks
kk05, s
50,r
Mx==ee (45)
kr,,50r
A series of mean particle sizes can be calculated from the k-th root of the k-th moment in Formula (45)
or from the median (and geometric mean) x and the standard deviation of that distribution using
50,r
Formula (46):
2
2
ln xk+05, s
05, ks
50,r
xx==ee (46)
kr,,50 r
The median x of a lognormal distribution has the same value as the geometric mean size x .
50,r
0,r
© ISO 2014 – All rights reserved 9

---------------------- Page: 14 ----------------------
ISO 9276-2:2014(E)

4.6 Calculation of volume specific surface area and the Sauter mean diameter
From distributions of any type of quantity moments can be used in the calculation of the volume specific
surface area, S , since S is inversely proportional to the weighted mean surface size, the Sauter mean
V V
diameter, x (Formula 14). It is given by:
12,
6
S = (47)
V
x
12,
Taking into account Formula (25), one arrives from surface, number or volume distributions at:
M
6
20,
S ==66=⋅M (48)
V −13,
M M
12, 30,
For particles other than spheres a shape factor shall be introduced.
5 The moment-ratio-notation
Moments are the basis for defining mean sizes and standard deviations of particle size distributions.
A random sample, containing a limited number of particles from a large population of particle sizes, is
used for estimation of the moments of the size distribution of that population. Estimation is concerned
with inference about the numerical values of the unknown population from those of the sample. Particle
size measurements are always done on discrete samples and involve a number of discrete size classes.
Therefore, only moments related to samples are dealt with in this part of ISO 9276.
NOTE Distribution density and cumulative distribution are represented by q and Q in the ISO 9276 series, but
in literature (References [3],[4],[7]–[9]) the symbols f and F are also used.
5.1 Definition of moments according to the moment-ratio-notation
Two different types of moments may be used, viz., moments and central moments. Moments are centred
around the origin of the particle size axis and central moments around the arithmetic mean particle
size.
The p-th moment of a sample, denoted as M , is defined as:
p
−1 p
MN= nD (49)
pi∑ i
i
where
Nn=
i is the total number of particles involved in the measurement;

i
D is the midpoint of the i-th size class [see also comments to Formula (30)];
i
n is the number of particles in the i-th size class (i.e. the class frequency of the number distri-
i
bution density).
The first moment, the (arithmetic) sample mean M of the particle sizes D, is mostly represented by D .
1
The second and third moments are proportional to the mean surface area and the mean volume
respectively, of the particles in a sample.
10 © ISO 2014 – All rights reserved

---------------------- Page: 15 ----------------------
ISO 9276-2:2014(E)

The p-th central moment around the mean D , denoted as m , is defined by:
p
−1 p
mN=−nD()D (50)
pi∑ i
i
Central moments of particle sizes D, are related to differ
...