Information supplied by the manufacturer of medical devices

This European Standard specifies requirements for information to be supplied by a manufacturer for medical devices regulated by Council Directive 90/385/EEC relating to active implantable medical devices and Council Directive 93/42/EEC concerning medical devices. It does not specify the language to be used for such information, nor does it specify the means by which the information is to be supplied. It is also intended to complement the specific requirements of the cited EU Directives on medical devices by providing guidance on means by which certain requirements can be met. If a manufacturer follows these means, they will provide a presumption of conformity with the relevant Essential Requirements regarding information to be supplied.
This standard does not cover requirements for provision of information for in vitro diagnostic medical devices, which are covered by other labelling standards (see Bibliography).
NOTE   When national transpositions of the Directives specify the means by which information shall be supplied, this standard does not provide derogation from these requirements for that country.

Bereitstellung von Informationen durch den Hersteller von Medizinprodukten

Die vorliegende !Europäische" Norm legt Anforderungen an Informationen fest, die von einem Hersteller für Medizinprodukte nach der EU-Richtlinie 90/385/EWG zu aktiven implantierbaren medizinischen Geräten und EU-Richtlinie 93/42/EWG zu Medizinprodukten zur Verfügung zu stellen sind. Es werden weder die verwendete Sprache für diese Informationen noch die Mittel festgelegt, mit denen die Informationen zu geben sind. Es ist ferner beabsichtigt, die besonderen Anforderungen der genannten EU-Richtlinien zu Medizinprodukten durch die Bereitstellung einer Anleitung zu ergänzen, wie bestimmte Anforderungen erfüllt werden können. Wenn diese Hinweise von einem Hersteller befolgt werden, so liefern sie eine Annahme der Übereinstimmung mit den relevanten grundlegenden Anforderungen hinsichtlich der bereitzustellenden Information.
Die vorliegende Norm schließt keine Anforderungen an die Bereitstellung von Informationen für In-vitro-Diag-nostika ein, die von anderen genannten Normen abgedeckt werden (siehe Literaturhinweise).
ANMERKUNG   Sofern nationale Umstellungen der Richtlinien die Mittel festlegen, mit denen Informationen zur Verfügung zu stellen sind, führt die vorliegende Norm nicht zu einer teilweisen Aufhebung dieser Anforderungen für das bestimmte Land.

Informations fournies par le fabricant de dispositifs médicaux

La !présente Norme européenne" spécifie les exigences concernant les informations à fournir par un fabricant de dispositifs médicaux d’après la réglementation dictée par la Directive du Conseil 90/385/CEE relative aux dispositifs médicaux implantables actifs et par la Directive du Conseil 93/42/CEE concernant les dispositifs médicaux. Elle ne spécifie pas la langue à utiliser pour ces informations ni les moyens par lesquels les informations sont fournies. Elle a également pour but de venir compléter les exigences spécifiques des Directives UE sur les dispositifs médicaux citées en donnant des indications sur la manière dont certaines exigences peuvent être satisfaites. Si un fabricant suit ces indications, celles-ci laissent à présumer de la conformité aux exigences essentielles relatives à la fourniture d'informations.
La présente norme ne couvre pas les exigences concernant la fourniture d’informations pour les dispositifs médicaux de diagnostic in vitro qui font l'objet d'autres normes d'étiquetage (voir Bibliographie).
NOTE   Lorsque des transpositions nationales des directives précisent les moyens par lesquels les informations doivent être fournies, la présente norme ne propose pas de dérogation à ces exigences pour un pays particulier.

Informacije, ki jih proizvajalec priloži medicinskim pripomočkom

Ta evropski standard določa zahteve za informacije, ki jih mora proizvajalec predložiti medicinskim pripomočkom, kar urejata Direktiva Sveta 90/385/EGS, ki obravnava aktivne medicinske pripomočke za vsaditev, in Direktiva Sveta 93/42/EGS o medicinskih pripomočkih. Ne določa jezika informacij in načina, na katerega je treba informacije predložiti. Namenjen je tudi dopolnitvi posebnih zahtev navedenih direktiv EU o medicinskih pripomočkih z zagotavljanjem navodil o načinu možnega izpolnjevanja določenih zahtev. Če proizvajalec upošteva te načine, priskrbi zagotovilo o skladnosti z ustreznimi bistvenimi zahtevami v zvezi z informacijami, ki jih je treba predložiti. Ta standard ne obravnava zahtev za zagotavljanje informacij za diagnostične medicinske pripomočke in vitro, ki so obravnavane v drugih standardih za označevanje (glej poglavje Literatura).

General Information

Status
Withdrawn
Publication Date
17-Nov-2013
Withdrawal Date
31-May-2021
Technical Committee
VAZ - Healthcare
Current Stage
9900 - Withdrawal (Adopted Project)
Start Date
27-May-2021
Due Date
19-Jun-2021
Completion Date
01-Jun-2021

Relations

Effective Date
01-Dec-2013
Effective Date
01-Jul-2021
Effective Date
18-Jan-2023
Effective Date
01-Dec-2013
Standard

SIST EN 1041:2008+A1:2013

English language
26 pages
Preview
Preview
e-Library read for
1 day

Frequently Asked Questions

SIST EN 1041:2008+A1:2013 is a standard published by the Slovenian Institute for Standardization (SIST). Its full title is "Information supplied by the manufacturer of medical devices". This standard covers: This European Standard specifies requirements for information to be supplied by a manufacturer for medical devices regulated by Council Directive 90/385/EEC relating to active implantable medical devices and Council Directive 93/42/EEC concerning medical devices. It does not specify the language to be used for such information, nor does it specify the means by which the information is to be supplied. It is also intended to complement the specific requirements of the cited EU Directives on medical devices by providing guidance on means by which certain requirements can be met. If a manufacturer follows these means, they will provide a presumption of conformity with the relevant Essential Requirements regarding information to be supplied. This standard does not cover requirements for provision of information for in vitro diagnostic medical devices, which are covered by other labelling standards (see Bibliography). NOTE When national transpositions of the Directives specify the means by which information shall be supplied, this standard does not provide derogation from these requirements for that country.

This European Standard specifies requirements for information to be supplied by a manufacturer for medical devices regulated by Council Directive 90/385/EEC relating to active implantable medical devices and Council Directive 93/42/EEC concerning medical devices. It does not specify the language to be used for such information, nor does it specify the means by which the information is to be supplied. It is also intended to complement the specific requirements of the cited EU Directives on medical devices by providing guidance on means by which certain requirements can be met. If a manufacturer follows these means, they will provide a presumption of conformity with the relevant Essential Requirements regarding information to be supplied. This standard does not cover requirements for provision of information for in vitro diagnostic medical devices, which are covered by other labelling standards (see Bibliography). NOTE When national transpositions of the Directives specify the means by which information shall be supplied, this standard does not provide derogation from these requirements for that country.

SIST EN 1041:2008+A1:2013 is classified under the following ICS (International Classification for Standards) categories: 01.110 - Technical product documentation; 11.040.01 - Medical equipment in general. The ICS classification helps identify the subject area and facilitates finding related standards.

SIST EN 1041:2008+A1:2013 has the following relationships with other standards: It is inter standard links to SIST EN 1041:2008, SIST EN ISO 20417:2021, SIST EN 1041:2008, SIST EN 1041:2008/kFprA1:2013. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

SIST EN 1041:2008+A1:2013 is associated with the following European legislation: EU Directives/Regulations: 90/385/EEC, 93/42/EEC; Standardization Mandates: M/023. When a standard is cited in the Official Journal of the European Union, products manufactured in conformity with it benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation.

You can purchase SIST EN 1041:2008+A1:2013 directly from iTeh Standards. The document is available in PDF format and is delivered instantly after payment. Add the standard to your cart and complete the secure checkout process. iTeh Standards is an authorized distributor of SIST standards.

Standards Content (Sample)


2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.Bereitstellung von Informationen durch den Hersteller von MedizinproduktenInformations fournies par le fabricant de dispositifs médicauxInformation supplied by the manufacturer of medical devices11.040.01Medicinska oprema na splošnoMedical equipment in general01.110L]GHONHTechnical product documentationICS:Ta slovenski standard je istoveten z:EN 1041:2008+A1:2013SIST EN 1041:2008+A1:2013en,fr,de01-december-2013SIST EN 1041:2008+A1:2013SLOVENSKI
STANDARD
EUROPEAN STANDARD NORME EUROPÉENNE EUROPÄISCHE NORM
EN 1041:2008+A1
September 2013 ICS 01.110; 11.040.01; 11.120.01 Supersedes EN 1041:2008 English version
Information supplied by the manufacturer of medical devices
Informations fournies par le fabricant de dispositifs médicaux
Bereitstellung von Informationen durch den Hersteller von Medizinprodukten This European Standard
was approved by CEN on 4 July 2008 and includes Amendment 1 approved by CEN on 11 July 2013.
CEN and CENELEC members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC Management Centre or to any CEN and CENELEC member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation under the responsibility of a CEN and CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the same status as the official versions.
CEN and CENELEC members are the national standards bodies and national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and United Kingdom.
CEN-CENELEC Management Centre: Avenue Marnix 17, B-1000 Brussels © 2013 CEN/CENELEC All rights of exploitation in any form and by any means reserved worldwide for CEN national Members and for CENELEC Members. Ref. No. EN 1041:2008+A1:2013 E
Requirements and guidance for Directives 93/42/EEC and 90/385/EEC, as amended 10 A.1 Requirements and guidance for medical devices (Directive 93/42/EEC) . 10 A.2 Requirements and guidance for active implantable medical devices (Directive 90/385/EEC) . 16 Annex B (informative)
Guidance on alternative labelling for instructions for use (IFU) . 20 B.1 Guidance on alternative labelling for medical devices (Directive 93/42/EEC) . 21 B.2 Guidance on alternative labelling for active implantable medical devices (Directive 90/385/EEC) . 22 Annex ZA (informative)
Relationship between this European Standard and the Essential Requirements of EU Directive 93/42/EC . 23 Annex ZB (informative)
Relationship between this European Standard and the Essential Requirements of EU Directive 90/385/EC . 24 Bibliography . 25
This !European Standard" specifies requirements for information to be supplied by a manufacturer for medical devices regulated by Council Directive 90/385/EEC relating to active implantable medical devices and Council Directive 93/42/EEC concerning medical devices. It does not specify the language to be used for such information, nor does it specify the means by which the information is to be supplied. It is also intended to complement the specific requirements of the cited EU Directives on medical devices by providing guidance on means by which certain requirements can be met. If a manufacturer follows these means, they will provide a presumption of conformity with the relevant Essential Requirements regarding information to be supplied. This standard does not cover requirements for provision of information for in vitro diagnostic medical devices, which are covered by other labelling standards (see Bibliography). NOTE When national transpositions of the Directives specify the means by which information shall be supplied, this standard does not provide derogation from these requirements for that country. 2 Normative references !The following documents, in whole or in part, are normatively referenced in this document and are indispensable for its application. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies." !deleted text" EN ISO 3166-1,!1)" Codes for the representation of names of countries and their subdivisions  Part 1: Country codes (ISO 3166-1:2006) ISO 639-1, Codes for the representation of names of languages  Part 1: Alpha-2 Code ISO 1000, SI units and recommendations for the use of their multiples and of certain other units ISO 8601, Data elements and interchange formats  Information interchange  Representation of dates and times CEN/TR 15133, Nomenclature  Collective terms and codes for groups of medical devices 3 Terms and definitions For the purposes of this document, the following terms and definitions apply. 3.1 batch; lot defined amount of material or a number of devices, including finished product and accessories, that is processed in one process or a series of related processes
NOTE The defined amount of material or number of devices will normally be associated with a unique statement of conformity to a defined quality specification.
! 1) EN ISO 3166-1 is currently impacted by the corrigendum EN ISO 3166-1:2006/AC:2008, Codes for the representation of names of countries and their subdivisions  Part 1: Country codes (ISO 3166-1:2006/Cor 1:2007)." SIST EN 1041:2008+A1:2013

all material, however provided, relating to the identification, technical description and use of a medical device that is intended to ensure the safe, effective and compliant use of the device NOTE Shipping documents and promotional material are excluded from this definition when identification, technical description and use of a medical device are not intended to ensure the safe, effective and compliant use of the device.
3.5 medical device any instrument, apparatus, appliance, software, material or other article, whether used alone or in combination, including the software intended by its manufacturer to be used specifically for diagnostic and/or therapeutic purposes and necessary for its proper application, intended by the manufacturer to be used for human beings for the purpose of:  diagnosis, prevention, monitoring, treatment or alleviation of disease;  diagnosis, monitoring, treatment, alleviation of or compensation for an injury or handicap;  investigation, replacement or modification of the anatomy or of a physiological process;  control of conception; and which does not achieve its principal intended action in or on the human body by pharmacological, immunological or metabolic means, but which may be assisted in its function by such means [Council Directive concerning medical devices (93/42/EEC), Article 1, paragraph 2 (a)] 3.6 user any person, legal or natural, for whom the information supplied (see 3.4) is intended 4 Requirements 4.1 General Product information and labelling shall be part of risk management procedures. NOTE 1
Due consideration should be given to the guidance provided in Annex A. NOTE 2
Product-related standards may require additional information to be supplied. 4.2 Units, symbols and colours Units used shall be SI units as specified in ISO 1000 or any other legal units. !Symbols and safety-related identification colours shall be explained in the information supplied unless they are taken from harmonised standards." SIST EN 1041:2008+A1:2013

4.4 Dates Any human-readable date shall be expressed in the format YYYY-MM-DD, YYYY-MM or YYYY, in accordance with ISO 8601. NOTE
The choice of format will be determined by the requirements of the relevant Directives and the specific nature of the devices concerned. 4.5 Device nomenclature 4.5.1 Identifiers of nomenclature When it is required to include the identification of the generic device group or the device category in the information supplied with the device, this may be done using a nomenclature that is in compliance with EN ISO 15225. NOTE
For details of nomenclatures claimed to be in compliance with EN ISO 15225, see the Bibliography. 4.5.2 Device common terms When it is appropriate to identify collective terms for medical devices in the information supplied, for example common technology or common materials of construction, this shall be done using the terms and codes set out in CEN/TR 15133. 4.5.3 Batch code; lot number; batch number; lot code These shall consist of alphanumeric characters but may also be presented by other means, for example by using machine-readable codes. 5 Requirements for provision of information 5.1 General 5.1.1 !Safe and effective use of the device" Any means of provision of information with medical devices shall take into account the intended users, the conditions of use and any issues specific to individual device types that are necessary for the safe and effective use of the device. This shall apply regardless of whether the specific requirements listed below apply to the device. The appropriate way of providing information shall be based on a risk assessment and in line with the training, experience and education of the intended users. 5.1.2 !Address required under medical devices directives
All medical devices which are placed on the market and put into service within the Community, shall contain the name or trade name and address of the manufacturer in the information supplied by the manufacturer. When the SIST EN 1041:2008+A1:2013

For devices covered by the MDD, the name or the trade name and address of the manufacturer shall appear on the label and in the instruction for use if provided with the device. When the manufacturer does not have a registered place of business in the Community, the label, or the outer packaging, or instructions for use shall contain, in addition, the name and address of the authorised representative.
For devices covered by the AIMDD, the name and address of the manufacturer shall appear on the sterile pack and the sales packaging and in the instruction for use. When the manufacturer does not have a registered place of business in the Community, the sales packaging and the instructions for use shall contain, in addition, the name and address of the authorised representative. The address to be used shall be the same as the address of the manufacturer and/or the authorised representative as their registered place of business. The address shall be the same as the address used on the declaration of conformity, in relevant certificates and in the European database for medical devices. The full address used shall contain the following elements insofar as they are available in the address system of the country where the relevant entity (manufacturer or authorised representative) is registered:  street/road;  number/house/floor;  postal code;  city;  state/region; and  country. The information regarding street/road and number/house/floor may be omitted if a postal code dedicated to the manufacturer (corporate postal code) or authorised representative is used which fully replaces the indication of street/road and number/house/floor, and is not a PO box number." 5.2 Specific requirements 5.2.1 Applicability These specific requirements shall be applicable to all devices to the extent that they are applicable to the specific device type concerned and to the means of provision of the relevant information. For example, the requirement to allow for a “use by” date is not applicable to devices that do not bear a “use by” date. 5.2.2 Accessibility The information presented with a device shall be accessible to intended users taking into account their age, education, knowledge and training. When appropriate, a specific means of provision may be restricted to users to whom it is particularly applicable. NOTE
This requirement may result in more than one means of provision being necessary. 5.2.3 Legibility Information intended for visual recognition shall be easily legible when viewed using normal vision, corrected if necessary, taking into account the specific size and conditions of use of the particular device. SIST EN 1041:2008+A1:2013

When appropriate and relevant, manufacturers should consider if there are any preventative measures that can be taken to maintain information security in relation to customer service. 5.2.6 Changes to information provided Any changes to information provided for existing users shall be clearly communicated if they are important for patient safety. 6 Documentation Documentation relating to information provided shall be maintained in the technical documentation(s) relating to the device(s) that are the subject of the information. This may take the form of a specific section holding all the documentation or, alternatively, references to parts of a larger document where the information may be found, such as a quality manual. SIST EN 1041:2008+A1:2013

Requirements and guidance for Directives 93/42/EEC and 90/385/EEC, as amended
NOTE
This annex covers those non-active and active medical devices to which Council Directives 93/42/EEC and 90/385/EEC apply, hereinafter called medical devices. A.1 Requirements and guidance for medical devices (Directive 93/42/EEC) NOTE
The left-hand column reproduces verbatim the requirements for information to be supplied by the manufacturer from Directive 93/42/EEC concerning Medical Devices given in the Essential Requirements in Annex I. The right hand column gives guidance and further explanation, as appropriate. Where no guidance is given, the Essential Requirements are not reproduced below. Requirements Information Requirements from the Council Directive concerning Medical Devices given in the Essential Requirements in Annex I Guidance General 8.7. The packaging and/or label of the device must distinguish between identical or similar products sold in both sterile and non-sterile condition.
In accordance with this standard, sterile devices are identified as such, preferably by the symbols as given in EN 980 or the word stating this condition. Sterile devices should be prominently labelled by the appropriate symbols as given in EN 980. The definitions of sterile as given in EN 556-1 and EN 556-2 apply. Where both sterile and non-sterile versions of the same device from the same manufacturer are available in similar packaging and where, in such cases, the non-sterile device could be mistaken as sterile, it may be necessary for the safety of the patient, to provide a prominent
statement of non-sterility. The similarity can either originate in the device itself or its packaging. 10.3. The measurements made by devices with a measuring function must be expressed in legal units conforming to the provisions of Council Directive 80/181/EEC. See 4.2, requirements, of this standard. 11.4.1. The operating instructions for devices emitting radiation must give detailed information as to the nature of the emitted radiation, means of protecting the patient and user and on ways of avoiding misuse and of eliminating the risks inherent in installation.
Radiation is not limited to ionizing radiation. Other examples of radiation include heat and laser radiation (see also 89/618/Euratom). 13. Information supplied by the manufacturer
13.1. Each device must be accompanied by the information needed to use it safely and properly, taking account of the training and knowledge of the potential users, and to identify the manufacturer. Any information should be made available in a way that is understandable to the intended user and/or patient. For complex equipment a user-friendly guide on how to check and operate the device in an emergency will be of SIST EN 1041:2008+A1:2013

As far as practicable and appropriate, the information needed to use the device safely must be set out on the device itself and/or on the packaging for each unit or, where appropriate, on the sales packaging. If individual packaging of each unit is not practicable, the information must be set out in the leaflet supplied with one or more devices. When instructions for use are provided by means of a leaflet, the number of leaflets in a multiple pack is determined by the manufacturer, taking into consideration the use of the device. Information may also be provided by electronic means (see Annex B). Many devices, particularly active devices and many non-active class I devices, will not be supplied with packaging, apart from transit containers. In the absence of suitable packaging, any information should be supplied on labelling, accompanying documentation, or marking of the device, as necessary. Instructions for use must be included in the packaging for every device. By way of exception, no such instructions for use are needed for devices in Class I or Class IIa if they can be used safely without any such instructions.
13.2. Where appropriate, this information should take the form of symbols. Any symbol or identification colour used must conform to harmonized standards. In areas for which no standards exist, the symbols and colours must be described in the documentation supplied with the device. Documentation can be the label and/or instructions for use. See 4.2 requirements, of this standard. 13.3. The label must bear the following particulars: National regulations may require the information referred to in sections 13.3 and 13.6 of Annex I of the medical devices directive to be in their national language(s) or in another Community language when a device reaches the final user, regardless of whether it is for professional or other use (93/42/EEC, Article 4, paragraph 4). The use of symbols that conform to harmonized standards will mean that there is no need to translate certain information. (a) the name or trade name and address of the manufacturer. For devices imported into the Community, in view of their distribution in the Community, the label, or the outer packaging, or instructions for use, shall contain in addition the name and address of the authorized representative where the manufacturer does not have a registered place of business in the Community; The full postal addresses may not be necessary if the information is sufficient to contact them, e.g., name or trade name, post code, country. However, the address needs to be sufficient to contact the physical location of the manufacturer and/or the authorized representative, if applicable. The post box alone is not sufficient.
(b) the details strictly necessary to identify the device and the contents of the packaging especially for the users; For many devices, the identity will be clearly evident to the intended user. Unpackaged devices, devices with transparent packaging or those provided only with transit or storage containers may not require further identification. For more complex devices, the identity of the product can be indicated on the device itself or on the packaging or instructions for use, as appropriate. It may be appropriate to list contents and quantity, SIST EN 1041:2008+A1:2013

The symbol STERILE should be prominent. Where only parts of the device are intended to be sterile, this should be stated e.g., sterile fluid path. (d) where appropriate, the batch code, preceded by the word "LOT", or the serial number; The word “lot” by itself is not a symbol and translation may be required. LOT as given in EN 980 is a symbol and therefore does not require translation.
The symbol LOT can be used to identify batch codes and the symbol
to identify serial numbers. (e) where appropriate, an indication of the date by which the device should be used in safety, expressed as the year and month; The symbol to identify the "use by" date is given in EN 980. This indicates the last month during which the device is intended to be used.
If it is not necessary to give a "use by" date, it may be appropriate to give the date of manufacture using the symbol given in EN 980 in the form YYYY-MM. The latter may be incorporated in the lot number
(e.g. LOT 2006-07 1234). (f) where appropriate, an indication that the device is for single use. A manufacturer's indication of single use must be consistent across the Community; The symbol ISO 7000/1051 for "Do not re-use" is reproduced in EN 980. (g) if the device is custom made, the words "custom-made device"; National language versions of the Directive also translate the words in quotes. (h) if the device is intended for clinical investigations, the words "exclusively for clinical investigations"; National language versions of the Directive also translate the words in quotes. (i) any special storage and/or handling conditions; Particulars need only be provided for unusual requirements for storage and handling conditions other than those that would normally be expected by the intended user. Information should also be given if storage or handling conditions are critical for the safe and proper performance of the device. Thus it would be generally understood without specific labelling that devices should be protected from extremes of temperature, from weather and from electro-magnetic radiation. However, if a device is required to be stored within a particular range of relative humidity and temperature, this should be specifically indicated.
Internationally recognized symbols may be used, as appropriate, for storage, handling or transport SIST EN 1041:2008+A1:2013

13.4. If the intended purpose of the device is not obvious to the user, the manufacturer must clearly state it on the label and in the instructions for use. For many devices, the intended purpose will be self-evident to the user. Unpackaged devices, or those provided only with transit or storage containers, may not require identification of their intended purpose. Transparent packaging may reduce the requirements for description. 13.5. Wherever reasonable and practicable, the devices and detachable components must be identified, where appropriate in terms of batches, to allow all appropriate action to detect any potential risk posed by the devices and detachable components. Such identification will facilitate recall of the device. Any detachable component should be identified by its batch code or by other appropriate means. 13.6. Where appropriate, the instructions for use must contain the following particulars: National regulations may require the information referred to in sections 13.3 and 13.6 of Annex I to be in their national language(s) when a device reaches the user, regardless of whether it is for professional or other use (93/42/EEC, clause 4, paragraph 4).
The use of symbols that conform to harmonized standards will mean that there is no need to translate certain information. (a) the details referred to in Section 13.3, with the exception of (d) and (e); The exceptions of d) (batch code) and e) (use by date) are not exhaustive. Section 13.6 makes it clear that the information listed under 13.3 need only be given "where appropriate". SIST EN 1041:2008+A1:2013

It would be neither appropriate nor feasible to include, for example, the date of manufacture (section 13.3 l) in the instructions for use where that date already appears on the label.
See the guidance given under this annex, 13.3 a), b), c), f), g), h), i), j), k), m) above. (b) the performances referred to in Section 3 and any undesirable side-effects; This could take the form of a reference to a relevant published standard that specifies those characteristics. (c) if the device must be installed with or connected to other medical devices or equipment in order to operate as required for its intended purpose, sufficient details of its characteristics to identify the correct devices or equipment to use in order to obtain a safe combination; Specific information will only be needed about methods of connection or the variety/types of equipment to which the device may properly be connected when these may not be expected to be common knowledge to the intended user, and are not self-evident.
Sufficient details of the characteristics (e.g., connections) can be provided by indication of compliance with a relevant published standard that specifies such characteristics. (d) all the information needed to verify whether the device is properly installed and can operate correctly and safely, plus details of the nature and frequency of the maintenance and calibration needed to ensure that the devices operate properly and safely at all times; This requirement only relates to verification by the user of installation or details of the nature and frequency of maintenance and calibration, rather than the actual steps involved. Information on installation need not be included in the instructions for use supplied to the user, although such information should be separately available if it is not self-evident and it is not intended that installation be done by the manufacturer or his agent. (e) where appropriate, information to avoid certain risks in connection with implantation of the device; This clause applies only to the instructions for use for implantable devices, and relates only to risks that are "certain", (i.e., recognized and foreseeable), as opposed to "uncertain" (i.e., unknown and/or improbable). This requirement also only relates to the risks that arise with the process of implantation, rather than those that arise after the device has been implanted. Information is not required about self-evident or trivial risks. As in 13.3 j) and 13.3 k), any special operating instructions, any warnings and/or recommended precautions should be given, taking into consideration such factors as the assumed technical knowledge and skill of the intended user. Internationally recognised symbols may be used. (h) if the device is reusable, information on the appropriate processes to allow reuse, including cleaning, disinfection, packaging and, where appropriate, the method of sterilization of the device to be resterilized, and any restriction on the number or reuses. This requirement relates only to devices intended by the manufacturer to be reusable. It does not relate to devices that a user may decide to reuse outside the manufacturer's recommendations, e.g., those devices marked as "single use". When the device is resterilizable, EN ISO 17664 may be used. Where devices are supplied with the intention that they be sterilized before use, the instructions for cleaning and sterilization must be such that, if correctly followed, the device will still comply with the requirements in Section I;
The left-hand column reproduces verbatim the requirements for information to be supplied by the manufacturer from Directive 90/385/EEC relating to Active Implantable Medical Devices given in the Essential Requirements in Annex I. The right hand column gives guidance and further explanation as appropriate. Where no guidance is given, the Essential Requirements are not reproduced below. Requirements Information Requirements from the Council Directive relating to Active Implantable Medical Devices given in the Essential Requirements in Annex I Guidance 11. The devices and, if appropriate, their component parts must be identified to allow any necessary measure to be taken following the discovery of a potential risk in connection with the devices and their component parts. Such identification will facilitate recall if necessary. 12. Devices must bear a code by which they and their manufacturer can be unequivocally identified (particularly with regard to the type of device and year of manufacture); it must be possible to read this code, if necessary, without the need for a surgical operation. An example of a means by which this code can be read without the need for a surgical operation would be by provision of radio-opaque symbols on the device, with/without further telemetry appropriate to the particular device, to obtain further identification details. 13. When a device or its accessories bear instructions required for the operation of the device or indicate operating or adjustment parameters, by means of a visual system, such information must be understandable to the user and, as appropriate, the patient. When a device is put into service, national regulations may require the information described in sections 13, 14 and 15 of Annex 1 of the Directive to be in their national language(s) (90/385/EEC, Article 4, paragraph 4). Consideration should be given to making this
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...

SIST EN 1041:2008+A1:2013 표준은 의료기기의 제조업체가 제공해야 하는 정보에 대한 요구 사항을 규정하고 있습니다. 이 표준은 90/385/EEC 지침에 따른 활성 이식 의료기기 및 93/42/EEC 지침에 대한 의료기기를 포함하여, 유럽의학 표준의 범위를 잘 정의하고 있습니다. 이 표준의 주요 강점은 제조업체가 특정 요구 사항을 충족하는 방법에 대한 지침을 제공하여, 관련 필수 요구 사항에 대한 준수를 전제로 하고 있다는 점입니다. 또한, 제공해야 하는 정보의 언어나 방법을 명시하지 않음으로써 제조업체가 자율적으로 접근할 수 있도록 허용하는 유연성을 가지고 있습니다. 이러한 유연성은 다양한 시장에서의 적용성을 높이며, 각국의 법률적 요구 사항에 맞춰 정보를 제공할 수 있는 기반을 마련합니다. SIST EN 1041:2008+A1:2013 표준은 인 비트로 진단 의료기기에 대한 정보 제공 요구 사항은 다루지 않으며, 이는 다른 라벨링 표준에 의해 규정됩니다. 이 점 역시 제조업체들이 필요할 수 있는 정보의 범위를 명확히 하고 있습니다. 결론적으로, SIST EN 1041:2008+A1:2013 표준은 유럽 의료기기 산업에서 제조업체가 필요한 정보를 적절히 제공할 수 있도록 하는 중요한 지침을 제공하며, 이는 의료기기의 안전성 및 효과성을 높이는 데 기여할 수 있습니다. 이 표준의 준수는 특정 요구 사항을 충족하여 소비자와 의료 전문가 모두에게 신뢰할 수 있는 정보를 제공하는 데 필수적입니다.

The SIST EN 1041:2008+A1:2013 standard provides a crucial framework for the information that manufacturers of medical devices must supply, addressing the requirements outlined in Council Directive 90/385/EEC for active implantable medical devices and Council Directive 93/42/EEC for medical devices. This standard's scope is particularly significant as it emphasizes the importance of providing necessary information to ensure safety and compliance in the use of medical devices. By not specifying the language or means of communication, the standard allows manufacturers the flexibility to cater to diverse markets while ensuring essential information is conveyed. This adaptability is one of its strengths, allowing manufacturers to align with local regulations while still adhering to overarching European regulatory frameworks. The relevance of this standard cannot be overstated, as it offers guidance that supports manufacturers in meeting the Essential Requirements related to informational content for medical devices. By employing the methods outlined in this standard, manufacturers can demonstrate a presumption of conformity to these requirements, which is vital for market access and patient safety. However, it is important to note that the standard does not address information provision for in vitro diagnostic medical devices, which are governed by separate labelling standards. This delineation ensures clarity and focus on active implantable and general medical devices, allowing stakeholders to seek more specific guidance as necessary. Overall, the SIST EN 1041:2008+A1:2013 standard stands out for its ability to complement existing directives with practical guidance, making it an essential reference for manufacturers seeking to deliver compliant and reliable medical devices to the market.

SIST EN 1041:2008+A1:2013は、医療機器の製造者が提供する情報に関するEU規範を示す重要な標準です。この標準は、特にCouncil Directive 90/385/EEC(アクティブインプランタブル医療機器に関する指令)およびCouncil Directive 93/42/EEC(医療機器に関する指令)に準拠するための情報提供要件を規定しています。範囲は広く、医療機器の使用者や医療関係者に対する情報の質を確保することを目的としています。 標準の強みは、そのフレキシブルな適用性にあります。具体的な言語や情報提供手段を定めていないため、製造者は自社の製品や市場に適した方法で情報を提供することができます。また、この標準は、EU指令の具体的な要件を補完する形で、特定の要件を満たすための手段についてのガイダンスを提供します。これにより、製造者が要求された情報を提供する際の効率性を高め、必要なコンプライアンスが容易になります。 さらに、SIST EN 1041は、製造者にとっての新たな信頼を生み出す手段としても機能します。製造者がこの標準に従って情報を提供することにより、関連する重要要件に対するコンプライアンスの推定が得られるため、製品の信頼性が向上することに寄与します。 ただし、この標準は、in vitro診断用医療機器に関する情報提供要件については適用されず、それに関しては他のラベリング標準が存在します。国ごとに指令の国内移植が情報提供手段を規定している場合、この標準はそれらの要求からの例外を提供するものではありません。 以上のように、SIST EN 1041:2008+A1:2013は、医療機器の製造者が適切な情報を提供するための効果的な枠組みを提供しており、医療機器市場において非常に重要な役割を果たしています。