Medical electrical equipment - Recurrent test and test after repair of medical electrical equipment

This International Standard applies to testing of MEDICAL ELECTRICAL EQUIPMENT and MEDICAL
ELECTRICAL SYSTEMS, hereafter referred to as ME EQUIPMENT and ME SYSTEMS, or parts of such
equipment or systems, which comply with IEC 60601-1:1988 (second edition) and its
amendments and IEC 60601-1: 2005 (third edition) and its amendments, before PUTTING INTO
SERVICE, during MAINTENANCE, INSPECTION, SERVICING and after REPAIR or on occasion of
RECURRENT TESTS to assess the safety of such ME EQUIPMENT or ME SYSTEMS or parts thereof.
For equipment not built to IEC 60601-1 these requirements may be used taking into account
the safety standards for the design and information in the instructions for use of that
equipment.
This standard contains tables with allowable values relating to different editions of
IEC 60601-1. For the purpose of this standard, the application of measuring methods is
independent of the edition according to which the ME EQUIPMENT or ME SYSTEM is designed.
This standard contains:
– "general requirements", which contain clauses of general concern, and
– "particular requirements", further clauses handling special types of ME EQUIPMENT or
ME SYSTEMS and applying in connection with the "General requirements".
NOTE At this stage, there are no particular requirements.
This standard is not suitable to assess whether ME EQUIPMENT or ME SYSTEMS or any other
equipment comply with the relevant standards for their design.
This standard is not applicable to the assembly of ME SYSTEMS. For assembling ME SYSTEMS
see Clause 16 of IEC 60601-1:2005 + IEC 60601-1:2005/AMD1:20121.
This standard does not define requirements for REPAIR, exchange of components and
MODIFICATION of ME EQUIPMENT or ME SYSTEMS.
All MAINTENANCE, INSPECTION, SERVICING, and REPAIR done in accordance with
MANUFACTURER's instructions maintain the conformity to the standard used for the design of
the equipment. Otherwise conformity to applicable requirements should be assessed and
verified, before the tests of this standard are performed.
This standard is also applicable to tests after REPAIR.
IEC 60601-1:2005 + IEC 60601-1:2005/AMD1:2012 requires that, as part of the RISK
MANAGEMENT PROCESS, the MANUFACTURER considers how the safety of ME EQUIPMENT or an
ME SYSTEM can be ensured during product lifetime. As part of the risk management process
the MANUFACTURER may have identified MAINTENANCE procedures. This includes defining the
respective tests for ME EQUIPMENT or for ME SYSTEM.The MANUFACTURER may have defined necessary measurement settings and methods
including performance assurance tests in the instructions for use or other ACCOMPANYING
DOCUMENTS. This standard provides consistent test procedures.
This standard is not intended to define time intervals for RECURRENT TESTS. If such intervals
are not defined by the MANUFACTURER, Annex F can be used to help establish such intervals.
Testing of the electrical installation, including the SUPPLY MAINS and associated wiring, in
medical locations is excluded from this standard. Such tests are covered by IEC 60364-7-710
or national equivalents,

Medizinische elektrische Geräte - Wiederholungsprüfungen und Prüfung nach Instandsetzung von medizinischen elektrischen Geräten

Appareils électromédicaux – Essai récurrent et essai après réparation d’un appareil électromédical

L'IEC 62353:2014 s'applique aux essais des appareils électromédicaux et des systèmes électromédicaux, ci-après dénommés appareils EM et systèmes EM, ou des parties de ces appareils ou systèmes, conformes à l'IEC 60601-1:1988 (deuxième édition) et à ses amendements et à l'IEC 60601-1:2005 (troisième édition) et à ses amendements, avant leur mise en service, pendant la maintenance, l'inspection, l'entretien courant et après leur réparation ou à l'occasion d'essais récurrents afin d'évaluer la sécurité de ces appareils EM ou systèmes EM ou de leurs parties. Pour les appareils non construits conformément à l'IEC 60601-1, ces exigences peuvent être utilisées en tenant compte des normes de sécurité pour la conception et les informations contenues dans les instructions d'utilisation de ces appareils. La présente norme comprend des tableaux donnant des valeurs admissibles se rapportant aux différentes éditions de l'IEC 60601-1. Pour les besoins de la présente norme, l'application des méthodes de mesure est indépendante de l'édition selon laquelle les appareils EM ou les systèmes EM sont conçus. La présente norme comprend des "exigences générales", contenant des articles de portée générale, et des "exigences particulières", d'autres articles traitant de types spéciaux d'appareils EM ou de systèmes EM et s'appliquant en conjonction avec les "exigences générales". La présente norme n'est pas appropriée pour évaluer si des appareils EM ou des systèmes EM ou d'autres appareils satisfont ou non aux normes correspondantes relatives à leur conception. La présente norme ne s'applique pas à l'assemblage des systèmes EM. Pour l'assemblage des systèmes EM, voir l'Article 16 de l'IEC 60601-1:2005 + IEC 60601-1:2005/AMD1:2012. La présente norme ne définit pas d'exigence pour la réparation, l'échange des composants et la modification des appareils EM ou des systèmes EM. Toutes les opérations de maintenance, inspection, entretien courant et réparation effectuées conformément aux instructions du fabricant conservent la conformité à la norme utilisée pour la conception de l'appareil. Sinon, il convient d'évaluer et de vérifier la conformité aux exigences applicables avant d'effectuer les essais de la présente norme. La présente norme s'applique également aux essais après réparation.
Cette deuxième édition annule et remplace la première édition de l'IEC 62353 parue en 2007. Cette édition constitue une révision technique qui contient les principales modifications suivantes:
- clarification en 5.3.4.1 que les mesurages des courants de fuite sur la base des configurations d'essai dérivées de l'IEC 60601-1 représentent une autre méthode acceptable, et ajout d'explications informatives en Annexe A;
- révision des exigences relatives à la résistance de terre de protection pour les systèmes EM utilisant des socles de prises multiples, de façon à prendre en compte l'IEC 60601-1 Amendement 1 concernant les valeurs admises sûres de la résistance de terre de protection des appareils branchés;
- ajout des valeurs minimales attendues de la résistance d'isolement dans le Tableau 2;
- et nouvel ordre pour la séquence des essais en Annexe B. La version française de cette norme n'a pas été soumise au vote.

Medicinska električna oprema - Ponavljalni preskus in preskus po popravilu medicinske električne opreme (IEC 62353:2014)

Ta mednarodni standard se uporablja za preskušanje MEDICINSKE ELEKTRIČNE OPREME in MEDICINSKIH ELEKTRIČNIH SISTEMOV, v nadaljnjem besedilu: ME OPREMA in ME SISTEMI, ali delov take opreme ali sistemov, ki so v skladu s standardom IEC 60601-1:1988 (druga izdaja) in njegovimi spremembami ter standardom IEC 60601-1:2005 (tretja izdaja) in njegovimi spremembami, pred ZAČETKOM UPORABE, med VZDRŽEVANJEM, INŠPEKCIJSKIMI PREGLEDI, SERVISIRANJEM in po POPRAVILU ali med PONAVLJALNIMI PRESKUSI za oceno varnosti take ME OPREME ali ME SISTEMOV ali njihovih delov. Za opremo, ki ni izdelana v skladu s standardom IEC 60601-1, se lahko te zahteve uporabljajo ob upoštevanju varnostnih standardov za zasnovo in informacije v navodilih za uporabo te opreme.
Ta standard vsebuje preglednice z dovoljenimi vrednostmi v zvezi z različnimi izdajami standarda IEC 60601-1. Za namen tega standarda je uporaba merilnih metod neodvisna od izdaje, v skladu s katero je ME OPREMA ali ME SISTEM zasnovan.
Ta standard vključuje:
– »splošne zahteve«, ki zajemajo splošna določila, in
– »posebne zahteve«, dodatna določila, ki obravnavajo posebne vrste ME OPREME ali ME SISTEMOV in se uporabljajo v zvezi s splošnimi zahtevami.
OPOMBA: Na tej stopnji ni posebnih zahtev.
Ta standard ni primeren za oceno, ali je ME OPREMA ali ME SISTEMI ali katera koli druga oprema skladna z ustreznimi standardi za njihovo zasnovo.
Ta standard se ne uporablja za sestav ME SISTEMOV. V zvezi s sestavljanjem ME SISTEMOV si oglejte točko 16 standarda IEC 60601-1:2005 + IEC 60601-1:2005/AMD1:20121.
Ta standard ne določa zahtev za POPRAVILO, izmenjavo komponent in SPREMINJANJE ME OPREME ali ME SISTEMOV.
Vse dejavnosti VZDRŽEVANJA, INŠPEKCIJSKIH PREGLEDOV, SERVISIRANJA in POPRAVIL, izvedene v skladu z navodili PROIZVAJALCA, ohranjajo skladnost s standardom, ki se uporablja za zasnovo opreme. Sicer bi bilo treba oceniti in preveriti skladnost z ustreznimi zahtevami pred izvedbo preskusov iz tega standarda.
Ta standard se uporablja tudi za preskuse po POPRAVILU.
V standardu IEC 60601-1:2005 + IEC 60601-1:2005/AMD1:2012 se zahteva, da PROIZVAJALEC v okviru POSTOPKA ZA OBVLADOVANJE TVEGANJA premisli o tem, kako je mogoče zagotoviti varnost ME OPREME ali ME SISTEMA med življenjsko dobo izdelka. Kot del postopka za obvladovanje tveganja lahko PROIZVAJALEC opredeli postopke VZDRŽEVANJA. To vključuje določitev ustreznih preskusov za ME OPREMO ali ME SISTEM. PROIZVAJALEC lahko določi ustrezne merilne nastavitve in metode,
vključno s preskusi za zagotavljanje učinkovitosti v navodilih za uporabo ali drugih SPREMNIH DOKUMENTIH. Ta standard zagotavlja skladne preskusne postopke.
Ta standard ni namenjen določitvi časovnih intervalov za PONAVLJALNE POSKUSE. Če takih intervalov ne določi PROIZVAJALEC, se lahko dodatek F uporabi za pomoč pri opredelitvi takih intervalov. Preskušanje električne inštalacije na medicinskih lokacijah, vključno z NAPAJALNIM OMREŽJEM in povezanim ožičenjem, je izključeno iz tega standarda. Take preskuse obravnava standard IEC 60364-7-710 ali nacionalni ekvivalenti.

General Information

Status
Published
Publication Date
23-Nov-2014
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
10-Nov-2014
Due Date
15-Jan-2015
Completion Date
24-Nov-2014

Relations

Buy Standard

Standard
EN 62353:2014
English language
65 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

SLOVENSKI STANDARD
SIST EN 62353:2014
01-december-2014
1DGRPHãþD
SIST EN 62353:2008
0HGLFLQVNDHOHNWULþQDRSUHPD3RQDYOMDOQLSUHVNXVLQSUHVNXVSRSRSUDYLOX
PHGLFLQVNHHOHNWULþQHRSUHPH ,(&
Medical electrical equipment - Recurrent test and test after repair of medical electrical
equipment
Medizinische elektrische Geräte - Wiederholungsprüfungen und Prüfung nach
Instandsetzung von medizinischen elektrischen Geräten
Appareils électromédicaux – Essai récurrent et essai après réparation d’un appareil
électromédical
Ta slovenski standard je istoveten z: EN 62353:2014
ICS:
11.040.01 Medicinska oprema na Medical equipment in general
splošno
SIST EN 62353:2014 en
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------

SIST EN 62353:2014

---------------------- Page: 2 ----------------------

SIST EN 62353:2014


EUROPEAN STANDARD EN 62353

NORME EUROPÉENNE

EUROPÄISCHE NORM
October 2014
ICS 11.040 Supersedes EN 62353:2008
English Version
Medical electrical equipment - Recurrent test and test after repair
of medical electrical equipment
(IEC 62353:2014)
Appareils électromédicaux - Essai récurrent et essai après Medizinische elektrische Geräte - Wiederholungsprüfungen
réparation d'un appareil électromédical und Prüfung nach Instandsetzung von medizinischen
(CEI 62353:2014) elektrischen Geräten
(IEC 62353:2014)
This European Standard was approved by CENELEC on 2014-10-09. CENELEC members are bound to comply with the CEN/CENELEC
Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC
Management Centre or to any CENELEC member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation
under the responsibility of a CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the
same status as the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia,
Lithuania, Luxembourg, Malta, the Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and the United Kingdom.


European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Avenue Marnix 17, B-1000 Brussels
© 2014 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
 Ref. No. EN 62353:2014 E

---------------------- Page: 3 ----------------------

SIST EN 62353:2014
EN 62353:2014 - 2 -
Foreword
The text of document 62A/942/FDIS, future edition 2 of IEC 62353 prepared by SC 62A "Common
aspects of electrical equipment used in medical practice" of IEC/TC 62 "Electrical equipment in
medical practice" was submitted to the IEC-CENELEC parallel vote and approved by CENELEC as
EN 62353:2014.

The following dates are fixed:
• latest date by which the document has to be (dop) 2015-07-09
implemented at national level by
publication of an identical national
standard or by endorsement
(dow) 2017-10-09
• latest date by which the national
standards conflicting with the
document have to be withdrawn

This document supersedes EN 62353:2008.

Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC [and/or CEN] shall not be held responsible for identifying any or all such
patent rights.

Endorsement notice
The text of the International Standard IEC 62353:2014 was approved by CENELEC as a European
Standard without any modification.
In the official version, for Bibliography, the following notes have to be added for the standards indicated:

IEC 60335 Series NOTE Harmonized as EN 60335 Series.
IEC 60950 Series NOTE Harmonized as EN 60950 Series.
IEC 60950-1 NOTE Harmonized as EN 60950-1.
IEC 61010 Series NOTE Harmonized as EN 61010 Series.
IEC 61557-2:2007 NOTE Harmonized as EN 61557-2:2007 (not modified).
IEC 61557-4:2007 NOTE Harmonized as EN 61557-4:2007 (not modified).
1) 1)
IEC 61557-16 NOTE Harmonized as EN 61557-16 (not modified).
IEC 62020 NOTE Harmonized as EN 62020.
ISO 13485:2003 NOTE Harmonized as EN ISO 13485:2012 (not modified).
ISO 14971:2007 NOTE Harmonized as EN ISO 14971:2012 (not modified).
IEC 60364-7-710 NOTE Harmonized as HD 60364-7-710.
IEC 61010-2-010 NOTE Harmonized as EN 61010-2-010.


1)
To be published.

---------------------- Page: 4 ----------------------

SIST EN 62353:2014
- 3 - EN 62353:2014
Annex ZA
(normative)

Normative references to international publications
with their corresponding European publications

The following documents, in whole or in part, are normatively referenced in this document and are
indispensable for its application. For dated references, only the edition cited applies. For undated
references, the latest edition of the referenced document (including any amendments) applies.

NOTE 1 When an International Publication has been modified by common modifications, indicated by (mod), the relevant
EN/HD applies.

NOTE 2 Up-to-date information on the latest versions of the European Standards listed in this annex is available here:
www.cenelec.eu

Publication Year Title EN/HD Year

IEC 60417-DB -  Graphical symbols for use on equipment - -
IEC 60601-1 1988 Medical electrical equipment - EN 60601-1 1990
 Part 1: General requirements for safety + corr. July 1994
+ A1 1991 + A1 1993
+ A1/corr. July 1994

+ A2 1995 + A2 1995
IEC 60601-1 2005 Medical electrical equipment - EN 60601-1 2006
 Part 1: General requirements for basic + corr. March 2010
 safety and essential performance + corr. May 2014
+A1 2012 + A1 2013
+ A1/corr. July 2014
IEC 61010-1 -  Safety requirements for electrical EN 61010-1 -
equipment for measurement, control and
laboratory use -
Part 1: General requirements
IEC 61010-031 -  Safety requirements for electrical EN 61010-031 -
equipment for measurement, control
and laboratory use -
Part 031: Safety requirements for hand-
held probe assemblies for electrical
measurement and test
IEC 61140 -  Protection against electric shock - EN 61140 -
Common aspects for installation and
equipment
IEC 61557-1 -  Electrical safety in low voltage distribution EN 61557-1 -
systems up to 1 000 V a.c. and 1 500 V
d.c. - Equipment for testing, measuring or
monitoring of protective measures -
Part 1: General requirements

---------------------- Page: 5 ----------------------

SIST EN 62353:2014

---------------------- Page: 6 ----------------------

SIST EN 62353:2014




IEC 62353

®


Edition 2.0 2014-09




INTERNATIONAL



STANDARD




NORME



INTERNATIONALE
colour

inside










Medical electrical equipment – Recurrent test and test after repair of medical

electrical equipment




Appareils électromédicaux – Essai récurrent et essai après réparation d’un

appareil électromédical















INTERNATIONAL

ELECTROTECHNICAL

COMMISSION


COMMISSION

ELECTROTECHNIQUE

PRICE CODE
INTERNATIONALE

CODE PRIX XB


ICS 11.040 ISBN 978-2-8322-1847-1



Warning! Make sure that you obtained this publication from an authorized distributor.

Attention! Veuillez vous assurer que vous avez obtenu cette publication via un distributeur agréé.

® Registered trademark of the International Electrotechnical Commission
Marque déposée de la Commission Electrotechnique Internationale

---------------------- Page: 7 ----------------------

SIST EN 62353:2014
– 2 – IEC 62353:2014 © IEC 2014
CONTENTS
FOREWORD . 5
1 Scope . 7
2 Normative references . 8
3 Terms and definitions . 8
4 Requirements . 16
4.1 * General requirements . 16
4.2 Testing before PUTTING INTO SERVICE, after MODIFICATIONS, and after REPAIR . 17
4.3 * RECURRENT TEST . 18
5 * Tests . 18
5.1 General . 18
5.2 Visual INSPECTION . 18
5.3 Measurements . 19
5.3.1 General . 19
5.3.2 Measuring of PROTECTIVE EARTH RESISTANCE . 19
5.3.3 * Measurement of insulation resistance (not mandatory) . 21
5.3.4 Leakage currents . 24
5.4 Functional test . 31
6 Results of test and evaluation . 31
6.1 Reporting of results. 31
6.2 Evaluation . 32
Annex A (informative) General guidance and rationale . 33
A.1 Intended audience . 33
A.2 Differences between IEC 60601-1 and IEC 62353 . 34
A.3 Rationale . 35
Annex B (informative) Sequence of testing . 42
Annex C (normative) Requirements for the measurement equipment and for
measurement circuits for PROTECTIVE EARTH RESISTANCE and leakage currents . 44
C.1 Requirements for the measurement equipment . 44
C.2 Measurement equipment for measurement of PROTECTIVE EARTH RESISTANCE . 44
C.3 Measurement equipment for measurements of EQUIPMENT LEAKAGE CURRENT . 45
C.4 Measurement equipment for measurements of APPLIED PART LEAKAGE
CURRENT . 45
Annex D (informative) PATIENT ENVIRONMENT . 47
Annex E (normative) Allowable values for leakage currents from IEC 60601-1 . 48
Annex F (informative) Testing intervals . 51
Annex G (informative) Example of test documentation . 52
Annex H (informative) Notes on testing ME SYSTEMS . 53
H.1 Overview . 53
H.2 Guidelines for re-testing of an ME SYSTEM . 53
H.3 Guidelines on ME SYSTEMS from the rationale annex of IEC 60601-
1:2005 /AMD1:2012 . 54
H.4 Examples of application of MULTIPLE SOCKET-OUTLETS (MSO) . 58
Bibliography . 60
Index of defined terms . 61

---------------------- Page: 8 ----------------------

SIST EN 62353:2014
IEC 62353:2014 © IEC 2014 – 3 –
Figure 1 – Measuring circuit for the measurement of PROTECTIVE EARTH RESISTANCE in
ME EQUIPMENT that is disconnected from the SUPPLY MAINS . 20
Figure 2 – Measuring circuit for the measurement of PROTECTIVE EARTH RESISTANCE in
ME EQUIPMENT or ME SYSTEMS, which for functional reasons cannot be disconnected
from the SUPPLY MAINS, or in ME EQUIPMENT or ME SYSTEMS permanently connected to
the SUPPLY MAINS . 20
Figure 3 – Measuring circuit for the measurement of the insulation resistance between
MAINS PART and protective earth for CLASS I ME EQUIPMENT and between MAINS PART and
(non-earthed) ACCESSIBLE CONDUCTIVE PARTS for CLASS I and CLASS II ME EQUIPMENT . 22
Figure 4 – Measuring circuit for measurement of the insulation resistance between
MAINS PART and APPLIED PARTS which make a patient connection for CLASS I or CLASS II
ME EQUIPMENT . 23
Figure 5 – Measuring circuit for measurement of the insulation resistance between F-
TYPE APPLIED PARTS which make a patient connection and protective earth for CLASS I
ME EQUIPMENT and between F-TYPE APPLIED PARTS which make a patient connection and
(non-earthed) ACCESSIBLE CONDUCTIVE PARTS for CLASS I and CLASS II ME EQUIPMENT . 23
Figure 6 – Measuring circuit for the measurement of ME EQUIPMENT leakage current –
alternative method . 26
Figure 7 – Measuring circuit for the measurement of EQUIPMENT LEAKAGE CURRENT–
direct method . 27
Figure 8 – Measuring circuit for the measurement EQUIPMENT LEAKAGE CURRENT–
differential method . 28
Figure 9 – Measuring circuit for the measurement of APPLIED PART LEAKAGE CURRENT “F-
TYPE APPLIED PART” – alternative method . 29
Figure 10 – Measuring circuit for the measurement of APPLIED PART LEAKAGE CURRENT –
MAINS VOLTAGE on F-TYPE APPLIED PART – direct method . 30
Figure 11 – Measuring circuit for the measurement of APPLIED PART LEAKAGE CURRENT
for equipment with an INTERNAL ELECTRICAL POWER SOURCE – direct method . 30
Figure A.1 – CLASS I ME EQUIPMENT with no earthed ACCESSIBLE CONDUCTIVE PARTS of
the enclosure . 37
Figure A.2 – Plugged-in CLASS I ME EQUIPMENT . 37
Figure A.3 – Plugged-in CLASS II ME EQUIPMENT . 38
Figure A.4 – Plugged-in CLASS I ME EQUIPMENT with mains on the APPLIED PART . 38
Figure A.5 – Plugged-in CLASS II ME EQUIPMENT with mains on the APPLIED PART . 39
Figure B.1 – Sequence of testing . 42
Figure B.2 – Measurement of LEAKAGE CURRENTS (non-PERMANENTLY INSTALLED
CLASS I ME EQUIPMENT) . 43
Figure C.1 – Example of a measuring device and its frequency characteristics . 46
Figure D.1 – Example of PATIENT ENVIRONMENT . 47
Figure G.1 – Example of test documentation . 52
Figure H.1 – Example of the construction of a MULTIPLE SOCKET-OUTLET (MSO)
(accessible only with the use of a tool). 58
Figure H 2 – Examples of application of MULTIPLE SOCKET-OUTLETS (MSO) . 59

Table 1 – Legends of symbols . 21
Table 2 – Insulation resistance values . 24
Table 3 – Allowable values for leakage currents . 31
Table A.1 – Addressees and their possible interest in this standard . 33
Table A.2 – Reasons for choosing different measuring methods . 40

---------------------- Page: 9 ----------------------

SIST EN 62353:2014
– 4 – IEC 62353:2014 © IEC 2014
Table E.1 – Allowable values for continuous leakage currents from IEC 60601-1:1988 . 48
Table E.2 – Allowable values for TOUCH CURRENTS, EARTH LEAKAGE CURRENTS, PATIENT
LEAKAGE CURRENTS and patient auxiliary currents under NORMAL CONDITION and SINGLE
FAULT CONDITION from IEC 60601-1:2005 . 49
Table E.3 – Allowable values for PATIENT LEAKAGE CURRENTS under the special test
conditions identified in 8.7.4.7 of IEC 60601-1:2005 . 50

Table H.1 – Some examples of ME SYSTEMS for illustration . 56

---------------------- Page: 10 ----------------------

SIST EN 62353:2014
IEC 62353:2014 © IEC 2014 – 5 –
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________

MEDICAL ELECTRICAL EQUIPMENT –
RECURRENT TEST AND TEST AFTER REPAIR
OF MEDICAL ELECTRICAL EQUIPMENT

FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International Standard IEC 62353 has been prepared by subcommittee 62A: Common aspects
of electrical equipment used in medical practice, of IEC technical committee 62: Electrical
equipment in medical practice.
This second edition cancels and replaces the first edition of IEC 62353 published in 2007.
This edition constitutes a technical revision. The principle revisions are:
a) clarification in 5.3.4.1 that measurements of leakage currents based on test configurations
derived from IEC 60601-1 are an allowable alternative method and the inclusion of
informative explanation in Annex A;
b) revision of the PROTECTIVE EARTH RESISTANCE requirements for MEDICAL ELECTRICAL
SYSTEMS using multiple socket outlets to take account of IEC 60601-1:2005/AMD1:2012 on
the safe allowed values of protective earth resistance of plugged-in equipment;
c) the inclusion of expected minimum insulation resistance values in Table 2; and
d) a reordering of the sequence of testing in Annex B.

---------------------- Page: 11 ----------------------

SIST EN 62353:2014
– 6 – IEC 62353:2014 © IEC 2014
The text of this standard is based on the following documents:
FDIS Report on voting
62A/942/FDIS 62A/953/RVD

Full information on the voting for the approval of this standard can be found in the report on
voting indicated in the above table.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
In this standard, the following print types are used:
– requirements and definitions: roman type;
– informative material appearing outside of tables, such as notes, examples and references: in smaller type.
Normative text of tables is also in a smaller type;
– TERMS USED THROUGHOUT THIS STANDARD THAT HAVE BEEN DEFINED IN CLAUSE 3: IN SMALL
CAPITALS.
The verbal forms used in this standard conform to usage described in Annex H of the ISO/IEC
Directives, Part 2. For the purposes of this standard, the auxiliary verb:
– “shall” means that compliance with a requirement or a test is mandatory for compliance
with this standard;
– “should” means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this standard;
– “may” is used to describe a permissible way to achieve compliance with a requirement or
test.
An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title
indicates that there is guidance or rationale related to that item in Annex A.
The committee has decided that the contents of this publication will remain unchanged until
the stability date indicated on the IEC web site under "http://webstore.iec.ch" in the data
related to the specific publication. At this date, the publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
NOTE The attention of National Committees is drawn to the fact that equipment manufacturers and testing
organizations may need a transitional period following publication of a new, amended or revised IEC or
ISO publication in which to make products in accordance with the new requirements and to equip themselves for
conducting new or revised tests. It is the recommendation of the committee that the content of this publication be
adopted for mandatory implementation nationally not earlier than 3 years from the date of publication.

IMPORTANT – The 'colour inside' logo on the cover page of this publication indicates
that it contains colours which are considered to be useful for the correct
understanding of its contents. Users should therefore print this document using a
colour printer.

---------------------- Page: 12 ----------------------

SIST EN 62353:2014
IEC 62353:2014 © IEC 2014 – 7 –
MEDICAL ELECTRICAL EQUIPMENT –
RECURRENT TEST AND TEST AFTER REPAIR
OF MEDICAL ELECTRICAL EQUIPMENT



1 Scope
This International Standard applies to testing of MEDICAL ELECTRICAL EQUIPMENT and MEDICAL
ELECTRICAL SYSTEMS, hereafter referred to as ME EQUIPMENT and ME SYSTEMS, or parts of such
equipment or systems, which comply with IEC 60601-1:1988 (second edition) and its
amendments and IEC 60601-1: 2005 (third edition) and its amendments, before PUTTING INTO
SERVICE, during MAINTENANCE, INSPECTION, SERVICING and after REPAIR or on occasion of
RECURRENT TESTS to assess the safety of such ME EQUIPMENT or ME SYSTEMS or parts thereof.
For equipment not built to IEC 60601-1 these requirements may be used taking into account
the safety standards for the design and information in the instructions for use of that
equipment.
This standard contains tables with allowable values relating to different editions of
IEC 60601-1. For the purpose of this standard, the application of measuring methods is
independent of the edition according to which the ME EQUIPMENT or ME SYSTEM is designed.
This standard contains:
– "general requirements", which contain clauses of general concern, and
– "particular requirements", further clauses handling special types of ME EQUIPMENT or
ME SYSTEMS and applying in connection with the "General requirements".
NOTE At this stage, there are no particular requirements.
This standard is not suitable to assess whether ME EQUIPMENT or ME SYSTEMS or any other
equipment comply with the relevant standards for their design.
This standard is not applicable to the assembly of ME SYSTEMS. For assembling ME SYSTEMS
1
see Clause 16 of IEC 60601-1:2005 + IEC 60601-1:2005/AMD1:2012 .
This standard does not define requirements for REPAIR, exchange of components and
MODIFICATION of ME EQUIPMENT or ME SYSTEMS.
All MAINTENANCE, INSPECTION, SERVICING, and REPAIR done in accordance with
MANUFACTURER's instructions maintain the conformity to the standard used for the design of
the equipment. Otherwise conformity to applicable requirements should be assessed and
verified, before the tests of this standard are performed.
This standard is also applicable to tests after REPAIR.
IEC 60601-1:2005 + IEC 60601-1:2005/AMD1:2012 requires that, as part of the RISK
MANAGEMENT PROCESS, the MANUFACTURER considers how the safety of ME EQUIPMENT or an
ME SYSTEM can be ensured during product lifetime. As part of the risk management process
the MANUFACTURER may have identified MAINTENANCE procedures. This includes defining the
respective tests for ME EQUIPMENT or for ME SYSTEM.
_______________
1
This citation refers to IEC 60601-1:2005 as amended by Amendment 1 published in 2012.

---------------------- Page: 13 ----------------------

SIST EN 62353:2014
– 8 – IEC 62353:2014 © IEC 2014
The MANUFACTURER may have defined necessary measurement settings and methods
including performance assurance tests in the instructions for use or other ACCOMPANYING
DOCUMENTS. This standard provides consistent test procedures.
This standard is not intended to define time intervals for RECURRENT TESTS. If such intervals
are not defined by the MANUFACTURER, Annex F can be used to help establish such intervals.
Testing of the electrical installation, including the SUPPLY MAINS and associated wiring, in
medical locations is excluded from this standard. Such tests are covered by IEC 60364-7-710
or national equiv
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.