Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-032: Particular requirements for hand-held and hand-manipulated current sensors for electrical test and measurement (IEC 61010-2-032:2019 + COR1:2020)

This part of IEC 61010 specifies safety requirements for HAND-HELD and hand-manipulated
current sensors described below.
These current sensors are for measuring, detecting or injecting current, or indicating current
waveforms on circuits without physically opening the current path of the circuit being
measured. They can be stand-alone current sensors or accessories to other equipment or
parts of combined equipment (see Figure 101). These include measurement circuits which are
part of electrical test and measurement equipment, laboratory equipment, or process control
equipment. These current sensors and circuits need additional protective means between the
current sensor, the circuit and an OPERATOR.
NOTE 1 Combined equipment is equipment that is electrically connected to a current sensor by means of a
permanent connection which can be detached only by the use of a TOOL.
NOTE 2 Some current sensors are also known as current clamps, CLAMP MULTIMETERS and current probes.
Current sensors are hand-manipulated before and/or after a test or measurement, but do not
necessarily need to be HAND-HELD during the test or measurement. Current sensors used as
FIXED EQUIPMENT are not within the scope of this document.

Errichten von Niederspannungsanlagen - Teil 5-56: Auswahl und Errichtung elektrischer Betriebsmittel - Einrichtungen für Sicherheitszwecke (IEC 61010-2-032:2019 + COR1:2020)

Règles de sécurité pour appareils électriques de mesurage, de régulation et de laboratoire - Partie 2-032: Exigences particulières pour les capteurs de courant, portatifs et manipulés à la main, de test et de mesure électriques (IEC 61010-2-032:2019 + COR1:2020)

IEC 61010-2-032:2019 est disponible sous forme de IEC 61010-2-032:2019 RLV qui contient la Norme internationale et sa version Redline, illustrant les modifications du contenu technique depuis l'édition précédente.

L'IEC 61010-2-032:2019 spécifie les exigences de sécurité pour les capteurs de courant portatifs et manipulés manuellement décrits ci-dessous. Ces capteurs de courant sont conçus pour mesurer, détecter ou injecter du courant, ou afficher les formes d'onde du courant sur les circuits sans ouverture physique du chemin du courant sur le circuit mesuré. Les capteurs de courant peuvent être autonomes, ou accessoires d'autres appareils ou parties d'appareils combinés (voir la Figure 101). Cela comprend les circuits de mesure qui font partie des appareils électriques d'essai et de mesure, du matériel de laboratoire ou des appareils de contrôle de procédés industriels. Ces capteurs de courant et circuits nécessitent des moyens de protection supplémentaires entre le capteur de courant, le circuit et un opérateur. Cette quatrième édition annule et remplace la troisième édition parue en 2012. Cette édition constitue une révision technique. Cette édition inclut les modifications techniques majeures suivantes par rapport à l'édition précédente:
- Il est indiqué que les capteurs de courant utilisés comme appareils installés à poste fixe ne relèvent pas du domaine d'application du présent document.
- Les capteurs de courant à fourche ont été ajoutés.
- Les exigences de la Partie 2-033 applicables aux pinces multimétriques qui ont pour objectif principal de mesurer la tension sur des réseaux sous tension ont été incluses dans la nouvelle Annexe EE normative.
- Les distances d'isolement et les lignes de fuite applicables aux emplacements humides et aux bornes d'un circuit de mesure dont la tension alternative est supérieure à 1 000 V ou dont la tension continue est supérieure à 1 414 V ont été spécifiées.
- Il est admis d'utiliser les lignes de fuite conformes au groupe de matériaux I pour tous les matériaux isolants.
- Les exigences relatives aux circuits d'entrée/sortie des capteurs de courant de Type A, de Type B et de Type C ont été détaillées en 6.9.102.
- Les exigences relatives aux cordons des circuits de sortie ont été modifiées.
- L'essai de choc de la mâchoire a été limité à la partie avant des mâchoires.
- L'essai d'abrasion des cordons de capteurs de courant flexibles a été supprimé et remplacé par un essai de pression à haute température.
- La source de tension utilisée pour l'essai des composants ou des circuits de limitation des surtensions peut être limitée à 400 V.
- Une référence à l'IEC 61010-031 pour les sondes équipées a été ajoutée.
- Des exigences permettant d'éviter les surtensions transitoires applicables aux circuits de mesure de tension de réseau ont été ajoutées.
- Des exigences concernant les circuits de mesure entre 1 000 V et 3 000 V ont été ajoutées.
- Une Annexe CC informative relative aux dimensions des bornes "banane" a été ajoutée.
- Un organigramme de l'isolation selon le type de circuit a été ajouté dans une nouvelle Annexe DD.


Varnostne zahteve za električno opremo za meritve, nadzor in laboratorijsko uporabo - 2-032. del: Posebne zahteve za ročne in ročno vodene tokovne senzorje za električno preskušanje in meritve (IEC 61010-2-032:2019 + COR1:2020)

Ta del standarda IEC 61010 določa varnostne zahteve za spodaj navedene ROČNE in ročno vodene tokovne senzorje. Ti tokovni senzorji so namenjeni za merjenje, zaznavanje ali priključitev toka ali določanje oblik toka na tokokrogih, ne da bi bilo treba fizično odpreti tokovne poti tokokroga, na katerem se izvaja meritev. Tokovni senzorji so lahko samostojni ali priključki k drugi opremi ali delom sestavljene opreme (glej sliko 101). Zajemajo merilne tokokroge, ki so del električne preskusne in merilne opreme, laboratorijske opreme ali opreme za vodenje procesa. Pri teh tokovnih senzorjih in tokokrogih je potrebno dodatno zaščitno sredstvo med tokovnim senzorjem, tokokrogom in UPRAVLJAVCEM. OPOMBA 1: Sestavljena oprema je oprema, ki je električno povezana s tokovnim senzorjem prek trajnega priključka, ki ga je mogoče odstraniti samo z ORODJEM. OPOMBA 2: Nekateri tokovni senzorji se imenujejo tudi tokovne klešče, KLEŠČNI MULTIMETRI in tokovne sonde. Tokovne senzorje je treba pred preskusom oziroma meritvijo in/ali po tem upravljati ročno, vendar jih med preskusom oziroma meritvijo ni vedno treba DRŽATI V ROKI. Tokovni senzorji, ki se uporabljajo kot FIKSNA OPREMA, ne spadajo na področje uporabe tega dokumenta.

General Information

Status
Published
Publication Date
31-Jan-2022
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
26-Nov-2021
Due Date
31-Jan-2022
Completion Date
01-Feb-2022

Relations

Buy Standard

Standard
EN IEC 61010-2-032:2022 - BARVE
English language
67 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

SLOVENSKI STANDARD
SIST EN IEC 61010-2-032:2022
01-marec-2022
Nadomešča:
SIST EN 61010-2-032:2013
Varnostne zahteve za električno opremo za meritve, nadzor in laboratorijsko
uporabo - 2-032. del: Posebne zahteve za ročne in ročno vodene tokovne senzorje
za električno preskušanje in meritve (IEC 61010-2-032:2019 + COR1:2020)
Safety requirements for electrical equipment for measurement, control and laboratory
use - Part 2-032: Particular requirements for hand-held and hand-manipulated current
sensors for electrical test and measurement (IEC 61010-2-032:2019 + COR1:2020)
Errichten von Niederspannungsanlagen - Teil 5-56: Auswahl und Errichtung elektrischer
Betriebsmittel - Einrichtungen für Sicherheitszwecke (IEC 61010-2-032:2019 +
COR1:2020)
Règles de sécurité pour appareils électriques de mesurage, de régulation et de
laboratoire - Partie 2-032: Exigences particulières pour les capteurs de courant, portatifs
et manipulés à la main, de test et de mesure électriques (IEC 61010-2-032:2019 +
COR1:2020)
Ta slovenski standard je istoveten z: EN IEC 61010-2-032:2021
ICS:
19.080 Električno in elektronsko Electrical and electronic
preskušanje testing
71.040.10 Kemijski laboratoriji. Chemical laboratories.
Laboratorijska oprema Laboratory equipment
SIST EN IEC 61010-2-032:2022 en
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------
SIST EN IEC 61010-2-032:2022

---------------------- Page: 2 ----------------------
SIST EN IEC 61010-2-032:2022


EUROPEAN STANDARD EN IEC 61010-2-032

NORME EUROPÉENNE

EUROPÄISCHE NORM
November 2021
ICS 19.080 Supersedes EN 61010-2-032:2012 and all of its
amendments and corrigenda (if any)
English Version
Safety requirements for electrical equipment for measurement,
control and laboratory use - Part 2-032: Particular requirements
for hand-held and hand-manipulated current sensors for
electrical test and measurement
(IEC 61010-2-032:2019 + COR1:2020)
Exigences de sécurité pour appareils électriques de Sicherheitsbestimmungen für elektrische Mess-, Steuer-,
mesurage, de régulation et de laboratoire - Partie 2-032 : Regel- und Laborgeräte - Teil 2-032: Besondere
Exigences particulières pour les capteurs de courant, Anforderungen für handgehaltene und handbediente
portatifs et manipulés manuellement, pour essai électrique Stromsonden für elektrische Prüfungen und Messungen
et mesurage (IEC 61010-2-032:2019 + COR1:2020)
(IEC 61010-2-032:2019 + COR1:2020)
This European Standard was approved by CENELEC on 2019-07-26. CENELEC members are bound to comply with the CEN/CENELEC
Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC
Management Centre or to any CENELEC member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation
under the responsibility of a CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the
same status as the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the
Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and the United Kingdom.


European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2021 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
 Ref. No. EN IEC 61010-2-032:2021 E

---------------------- Page: 3 ----------------------
SIST EN IEC 61010-2-032:2022
EN IEC 61010-2-032:2021 (E)
European foreword
The text of document 66/691/FDIS, future edition 4 of IEC 61010-2-032, prepared by IEC/TC 66
"Safety of measuring, control and laboratory equipment" was submitted to the IEC-CENELEC parallel
vote and approved by CENELEC as EN IEC 61010-2-032:2021.
The following dates are fixed:
• latest date by which the document has to be implemented at national (dop) 2022-05-12
level by publication of an identical national standard or by endorsement
• latest date by which the national standards conflicting with the (dow) 2024-11-12
document have to be withdrawn
This document supersedes EN 61010-2-032:2012 and all of its amendments and corrigenda (if any).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC shall not be held responsible for identifying any or all such patent rights.
This document has been prepared under a Standardization Request given to CENELEC by the
European Commission and the European Free Trade Association, and supports essential
requirements of EU Directive(s) / Regulation(s).
For the relationship with EU Directive(s) / Regulation(s), see informative Annex ZZ, which is an
integral part of EN IEC 61010-2-032:2021/A11:2021.
Any feedback and questions on this document should be directed to the users’ national committee. A
complete listing of these bodies can be found on the CENELEC website.
Endorsement notice
The text of the International Standard IEC 61010-2-032:2019 was approved by CENELEC as a
European Standard without any modification.
In the official version, for Bibliography, the following notes have to be added for the standards
indicated:
IEC 61010-2-033 NOTE Harmonized as EN IEC 61010-2-033
IEC 61010-2-034 NOTE Harmonized as EN IEC 61010-2-034
IEC 61557-1 NOTE Harmonized as EN 61557-1
IEC 61557-2 NOTE Harmonized as EN 61557-2
IEC 61557-3 NOTE Harmonized as EN 61557-3
IEC 61557-4 NOTE Harmonized as EN 61557-4
IEC 61557-5 NOTE Harmonized as EN 61557-5
IEC 61557-6 NOTE Harmonized as EN 61557-6
IEC 61557-7 NOTE Harmonized as EN 61557-7
2

---------------------- Page: 4 ----------------------
SIST EN IEC 61010-2-032:2022
EN IEC 61010-2-032:2021 (E)
IEC 61557-8 NOTE Harmonized as EN 61557-8
IEC 61557-9 NOTE Harmonized as EN 61557-9
IEC 61557-10 NOTE Harmonized as EN 61557-10
IEC 61557-11 NOTE Harmonized as EN 61557-11
IEC 61557-12 NOTE Harmonized as EN 61557-12

3

---------------------- Page: 5 ----------------------
SIST EN IEC 61010-2-032:2022

---------------------- Page: 6 ----------------------
SIST EN IEC 61010-2-032:2022
IEC 61010-2-032
®

Edition 4.0 2019-06
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
Safety requirements for electrical equipment for measurement, control
and laboratory use –
Part 2-032: Particular requirements for HAND-HELD and hand-manipulated
current sensors for electrical test and measurement
Exigences de sécurité pour appareils électriques de mesurage, de régulation
et de laboratoire –
Partie 2-032: Exigences particulières pour les capteurs de courant, PORTATIFS
et manipulés manuellement, pour essai électrique et mesurage
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
INTERNATIONALE
ICS 19.080 ISBN 978-2-8322-6994-7
Warning! Make sure that you obtained this publication from an authorized distributor.
Attention! Veuillez vous assurer que vous avez obtenu cette publication via un distributeur agréé.
® Registered trademark of the International Electrotechnical Commission
Marque déposée de la Commission Electrotechnique Internationale

---------------------- Page: 7 ----------------------
SIST EN IEC 61010-2-032:2022
– 2 – IEC 61010-2-032:2019 © IEC 2019
CONTENTS
FOREWORD . 4
INTRODUCTION . 7
1 Scope and object . 8
2 Normative references . 10
3 Terms and definitions . 10
4 Tests . 11
5 Marking and documentation . 12
6 Protection against electric shock . 16
7 Protection against mechanical HAZARDS . 21
8 Resistance to mechanical stresses . 21
9 Protection against the spread of fire . 25
10 Equipment temperature limits and resistance to heat . 25
11 Protection against HAZARDS from fluids and solid foreign objects . 26
12 Protection against radiation, including laser sources, and against sonic and
ultrasonic pressure . 26
13 Protection against liberated gases and substances, explosion and implosion . 26
14 Components and subassemblies . 27
15 Protection by interlocks . 28
16 HAZARDS resulting from application . 28
17 RISK assessment . 28
Annexes . 36
Annex D (normative) Parts between which insulation requirements are specified (see
6.4, 6.5.3, 6.9.101 and 6.101) . 36
Annex F (normative) Routine tests . 38
Annex K (normative) Insulation requirements not covered by 6.7 . 39
Annex L (informative) Index of defined terms . 46
Annex AA (normative) MEASUREMENT CATEGORIES . 47
Annex BB (informative) HAZARDS pertaining to measurements performed in certain
environments . 49
Annex CC (informative) 4-mm "banana" TERMINALS . 52
Annex DD (informative) Flowchart for insulation according to the type of circuit . 54
Annex EE (normative) CLAMP MULTIMETER . 57
Bibliography . 60
Figure 101 – Examples of current sensors and their parts . 9
Figure 102 – CLEARANCE between the PROTECTIVE BARRIER to the JAWS and to the
HAZARDOUS LIVE conductor . 19
Figure 103 – Abrasion test of the JAW ENDS . 22
Figure 104 – Impact points for JAW impact test . 23
Figure 105 – Indentation device . 24
Figure 106 – Test probe to check protection against short-circuits . 34
Figure 107 – Use of the test probe of Figure 106 . 34

---------------------- Page: 8 ----------------------
SIST EN IEC 61010-2-032:2022
IEC 61010-2-032:2019 © IEC 2019 – 3 –
Figure D.101 – Parts of current sensors (see also Table D.101) . 36
Figure AA.1 – Example to identify the locations of measuring circuits . 48
Figure CC.1 – Recommended dimensions of 4-mm TERMINALS . 53
Figure DD.1 – Requirements for CLEARANCE, CREEPAGE DISTANCE and solid insulation . 56
Figure EE.1 – Examples of CLAMP MULTIMETERS . 57
Table 1 – Symbols . 12
Table 101 – CLEARANCES and CREEPAGE DISTANCES for measuring circuit TERMINALS
with HAZARDOUS LIVE conductive parts up to 1 000 V a.c. or 1 500 V d.c. . 17
Table 102 – Energy level for JAW impact test . 22
Table 103 – Pull forces for endcaps of flexible current sensors . 25
Table 104 – Impulse voltages . 28
Table 105 – Thickness of the test probe of Figure 106 and test voltages. 35
Table D.101 – Insulation requirements for current sensors . 37
Table K.101 – CLEARANCES of measuring circuits RATED for MEASUREMENT CATEGORIES . 40
Table K.102 – a.c. test voltages for testing electric strength of solid insulation in
measuring circuits RATED for MEASUREMENT CATEGORIES . 41
Table K.103 – Impulse test voltages for testing electric strength of solid insulation in
measuring circuits RATED for MEASUREMENT CATEGORIES . 42
Table K.104 –Test voltages for testing long-term stress of solid insulation in measuring
circuits RATED for MEASUREMENT CATEGORIES . 43
Table K.105 – Minimum values for distance or thickness of solid insulation in
measuring circuits RATED FOR MEASUREMENT CATEGORIES III and IV . 44
Table AA.1 – Characteristics of MEASUREMENT CATEGORIES . 48

---------------------- Page: 9 ----------------------
SIST EN IEC 61010-2-032:2022
– 4 – IEC 61010-2-032:2019 © IEC 2019
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
SAFETY REQUIREMENTS FOR ELECTRICAL EQUIPMENT
FOR MEASUREMENT, CONTROL, AND LABORATORY USE –
Part 2-032: Particular requirements for HAND-HELD and
hand-manipulated current sensors for electrical test and measurement
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as "IEC
Publication(s)"). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International Standard IEC 61010-2-032 has been prepared by IEC technical committee 66:
Safety of measuring, control and laboratory equipment.
This fourth edition cancels and replaces the third edition published in 2012. This edition
constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous
edition:
a) It has been indicated that current sensors used as FIXED EQUIPMENT are not within the
scope of this document.
b) Fork-style current sensors have been added.

---------------------- Page: 10 ----------------------
SIST EN IEC 61010-2-032:2022
IEC 61010-2-032:2019 © IEC 2019 – 5 –
c) Requirements from Part 2-033 applicable to CLAMP MULTIMETERS that have a primary
purpose of measuring voltage on live MAINS have been included in the new normative
Annex EE.
LEARANCES and CREEPAGE DISTANCES for measuring circuit TERMINALS exceeding
d) C
1 000 V a.c. or 1 414 V d.c. and for WET LOCATIONS have been specified.
e) Reduced CREEPAGE DISTANCES are allowed to be according to material group I for all
insulating materials.
f) Requirements for input/output circuits of Type A, Type B and Type C current sensors
have been detailed in 6.9.102.
g) Requirements for output circuit leads have been modified.
h) The JAW impact test has been limited to the front of the JAWS.
i) The abrasion test for cords of flexible current sensors has been removed and replaced by
a pressure test at high temperature.
j) The voltage source for testing overvoltage limiting components or circuits may be limited
to 400 V.
k) Reference to IEC 61010-031 for probe assemblies has been added.
l) Requirements for the prevention of TRANSIENT OVERVOLTAGES for MAINS voltage measuring
circuits have been added.
m) Requirements for measuring circuits from 1 000 V to 3 000 V have been added.
TERMINALS has been added.
n) An informative Annex CC about the dimensions of banana
o) A flowchart for insulation according to the type of circuit has been added in a new
Annex DD.
The text of this International Standard is based on the following documents:
FDIS Report on voting
66/691/FDIS 66/695/RVD
Full information on the voting for the approval of this International Standard can be found in
the report on voting indicated in the above table.
This document has been drafted in accordance with the ISO/IEC Directives, Part 2.
A list of all parts of the IEC 61010 series, under the general title Safety requirements for
electrical equipment for measurement, control, and laboratory use, can be found on the IEC
website.
This Part 2-032 is to be used in conjunction with the latest edition of IEC 61010-1. It was
established on the basis of the third edition (2010) of IEC 61010-1 and its Amendment 1
(2016), hereinafter referred to as Part 1.
This Part 2-032 supplements or modifies the corresponding clauses in IEC 61010-1 so as to
convert that publication into the IEC standard: Particular requirements for HAND-HELD and
hand-manipulated current sensors for electrical test and measurement.
Where a particular subclause of Part 1 is not mentioned in this Part 2-032, that subclause
applies as far as is reasonable. Where this Part 2-032 states "addition", "modification",
"replacement", or "deletion" the relevant requirement, test specification or note in Part 1
should be adapted accordingly.

---------------------- Page: 11 ----------------------
SIST EN IEC 61010-2-032:2022
– 6 – IEC 61010-2-032:2019 © IEC 2019
In this standard:
a) the fol
lowing print types are used:
– requirements: in roman type;
NOTES: in small roman type;

– conformity and tests: in italic type;
– terms used throughout this standard which have been defined in Clause 3: SMALL
ROMAN CAPITALS;
b) subclauses, figures, tables and notes which are additional to those in Part 1 are numbered
starting from 101. Additional annexes are lettered starting from AA and additional list
items are lettered from aa).
The committee has decided that the contents of this document will remain unchanged until the
stability date indicated on the IEC website under "http://webstore.iec.ch" in the data related to
the specific document. At this date, the document will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
IMPORTANT – The 'colour inside' logo on the cover page of this publication indicates
that it contains colours which are considered to be useful for the correct
understanding of its contents. Users should therefore print this document using a
colour printer.
The contents of the corrigendum of February 2020 have been included in this copy.

---------------------- Page: 12 ----------------------
SIST EN IEC 61010-2-032:2022
IEC 61010-2-032:2019 © IEC 2019 – 7 –
INTRODUCTION
Part 2-030 specifies the safety requirements for equipment with testing and measuring circuits
which are connected for test or measurement purposes to devices or circuits outside the
measurement equipment itself. Requirements of Part 2-030 have been included in this Part 2-
032. Equipment within the scopes of both Part 2-030 and Part 2-032 are considered to be
covered by the requirements of this Part 2-032.
Part 2-033 specifies the safety requirements for hand-held multimeters that have the primary
purpose of measuring voltage on live MAINS. For equipment within the scope of Part 2-032 and
Part 2-033, only this Part 2-032 is applicable.
Part 2-034 specifies the safety requirements for measurement equipment for insulation
resistance and test equipment for electric strength which are connected to units, lines or
circuits for test or measurement purposes. For equipment within the scope of Part 2-032 and
Part 2-034, both documents should be read in conjunction.

---------------------- Page: 13 ----------------------
SIST EN IEC 61010-2-032:2022
– 8 – IEC 61010-2-032:2019 © IEC 2019
SAFETY REQUIREMENTS FOR ELECTRICAL EQUIPMENT
FOR MEASUREMENT, CONTROL, AND LABORATORY USE –

Part 2-032: Particular requirements for HAND-HELD and
hand-manipulated current sensors for electrical test and measurement



1 Scope and object
This clause of Part 1 is applicable except as follows:
1.1.1 Equipment included in scope
Replace the existing text with the following:
HAND-HELD and hand-manipulated
This part of IEC 61010 specifies safety requirements for
current sensors described below.
These current sensors are for measuring, detecting or injecting current, or indicating current
waveforms on circuits without physically opening the current path of the circuit being
measured. They can be stand-alone current sensors or accessories to other equipment or
parts of combined equipment (see Figure 101). These include measurement circuits which are
part of electrical test and measurement equipment, laboratory equipment, or process control
equipment. These current sensors and circuits need additional protective means between the
current sensor, the circuit and an OPERATOR.
NOTE 1 Combined equipment is equipment that is electrically connected to a current sensor by means of a
permanent connection which can be detached only by the use of a TOOL.
NOTE 2 Some current sensors are also known as current clamps, CLAMP MULTIMETERS and current probes.
Current sensors are hand-manipulated before and/or after a test or measurement, but do not
necessarily need to be HAND-HELD during the test or measurement. Current sensors used as
FIXED EQUIPMENT are not within the scope of this document.
The following types of current sensors are covered:
a) Type A: a current sensor designed to be applied to or removed from HAZARDOUS LIVE
UNINSULATED CONDUCTORS. Type A current sensors have defined HAND-HELD or hand-
manipulated parts providing protection against electric shock from the conductor being
measured, and also have protection against short-circuits between wires and between
busbars during clamping.
b) Type B: a current sensor which has protection against short-circuits between wires or
busbars during clamping but without defined HAND-HELD or hand-manipulated parts which
provide protection against electric shock during clamping. Additional protective means are
necessary to avoid electric shock from HAZARDOUS LIVE conductors which cannot be de-
energised during application or removal of the current sensor.
EXAMPLE 1 Flexible current sensors.
c) Type C: a current sensor without protection against short-circuits between wires or
busbars during clamping. Type C current sensors are intended to be applied to or
removed from HAZARDOUS LIVE UNINSULATED CONDUCTORS or from non-limited-energy
circuit conductors only when they are de-energised.
EXAMPLE 2 Split-core transducers.
d) Type D: a current sensor designed to be applied to or removed from insulated conductors
or from limited-energy circuit conductors.

---------------------- Page: 14 ----------------------
SIST EN IEC 61010-2-032:2022
IEC 61010-2-032:2019 © IEC 2019 – 9 –
EXAMPLE 3 Current probes for oscilloscopes and earth leakage current detectors.
All current sensors can also be used with insulated conductors. In this case, HAZARDS are
limited to acceptable levels by the insulation of the conductors.
Additional requirements for CLAMP MULTIMETERS are given in Annex EE.
Figure 101 shows graphical representations of typical current sensors for illustration
purposes. Current sensors can look different depending on the design.


Type A current sensor as an accessory Type A current sensor with self-contained and/or
additional measuring functions


Type A current sensor with fork-style JAWS Type B flexible current sensor
part of a piece of combined equipment


Type C split-core current sensor Type D current sensor for non-HAZARDOUS LIVE
applications (shown with a sliding JAW)

Key
1 JAW END(S)
2 JAW
3 Measuring circuit TERMINALS
Figure 101 – Examples of current sensors and their parts
1.2.1 Aspects included in scope
Add the following three new paragraphs at the end of the subclause:

---------------------- Page: 15 ----------------------
SIST EN IEC 61010-2-032:2022
– 10 – IEC 61010-2-032:2019 © IEC 2019
Requirements for protection against HAZARDS resulting from NORMAL USE and REASONABLY
FORESEEABLE MISUSE of measuring circuits are given in Clause 101.
Requirements for prevention of HAZARD from arc flash and short-circuits are given in
Clause 102.
Requirements for reliance on the displayed value of CLAMP MULTIMETERS are given in
Clause EE.5 .
2 Normative references
This clause of Part 1 is applicable except as follows:
Replace "IEC 61010-031" with the following new reference:
IEC 61010-031:2015, Safety requirements for electrical equipment for measurement, control
and laboratory use – Part 031: Safety requirements for hand-held and hand-manipulated
probe assemblies for electrical test and measurement
IEC 61010-031:2015/AMD1:2
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.