ISO/TC 121 - Anaesthetic and respiratory equipment
Standardization of anaesthetic and respiratory equipment and supplies, related devices and supply systems.
Matériel d'anesthésie et de réanimation respiratoire
Normalisation du matériel d'anesthésie et de réanimation respiratoire, accessoires et systèmes d'alimentation.
General Information
This document specifies requirements for masks and accessories, including any connecting element, that are required to connect the patient-connection port of sleep apnoea breathing therapy equipment to a patient for the application of sleep apnoea breathing therapy (e.g. nasal masks, exhaust ports and headgear). This document applies to masks and their accessories used to connect sleep apnoea breathing therapy equipment to the patient. The requirements in this document take priority over the requirements in ISO 18190. This document does not cover oral appliances. NOTE This document has been prepared to address the relevant essential principles[14] and labelling principles[15] of the International Medical Devices Regulators Forum (IMDRF) as indicated in Annex I.
- Standard36 pagesEnglish languagesale 15% off
- Standard37 pagesFrench languagesale 15% off
This document is applicable to the basic safety and essential performance of sleep apnoea breathing therapy equipment, hereafter referred to as ME equipment, intended to alleviate the symptoms of patients who suffer from obstructive sleep apnoea by delivering a therapeutic breathing pressure to the respiratory tract of the patient. Sleep apnoea breathing therapy equipment is intended for use in the home healthcare environment by lay operators as well as in professional healthcare institutions. Sleep apnoea breathing therapy equipment is not considered to utilize a physiologic closed-loop-control system unless it uses a physiological patient variable to adjust the therapy settings. This document excludes sleep apnoea breathing therapy equipment intended for use with neonates. This document is applicable to ME equipment or an ME system intended for those patients who are not dependent on artificial ventilation. This document is not applicable to ME equipment or an ME system intended for those patients who are dependent on artificial ventilation such as patients with central sleep apnoea. This document is also applicable to those accessories intended by their manufacturer to be connected to sleep apnoea breathing therapy equipment, where the characteristics of those accessories can affect the basic safety or essential performance of the sleep apnoea breathing therapy equipment. Masks and application accessories intended for use during sleep apnoea breathing therapy are additionally addressed by ISO 17510. Refer to Figure AA.1 for items covered further under this document. If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in 7.2.13 and 8.4.1 of the general standard. NOTE 2 See also 4.2 of the general standard. This document does not specify the requirements for: – ventilators or accessories intended for critical care ventilators for ventilator-dependent patients, which are given in ISO 80601‑2‑12. – ventilators or accessories intended for anaesthetic applications, which are given in ISO 80601-2-13. – ventilators or accessories intended for home care ventilators for ventilator-dependent patients, which are given in ISO 80601-2-72. – ventilators or accessories intended for emergency and transport, which are given in ISO 80601-2-84. – ventilators or accessories intended for home-care ventilatory support, which are given in ISO 80601‑2-79 and ISO 80601‑2‑80. – high-frequency ventilators[23], which are given in ISO 80601-2-87. – respiratory high flow equipment, which are given in ISO 80601‑2‑90; NOTE 3 ISO 80601-2-80 ventilatory support equipment can incorporate high-flow therapy operational mode, but such a mode is only for spontaneously breathing patients. – user-powered resuscitators, which are given in ISO 10651-4; – gas-powered emergency resuscitators, which are given in ISO 10651-5; – oxygen therapy constant flow ME equipment; and – cuirass or “iron-lung” ventilation equipment.
- Standard78 pagesEnglish languagesale 15% off
- Standard83 pagesFrench languagesale 15% off
This document specifies particular requirements for portable units which are part of a transportable liquid oxygen system and used to provide a controlled flow of oxygen for inhalation by the patient in the home-care environment. Portable units are intended to be used without professional supervision, carried by patients while moving around and during their off-site activities and refilled from a base unit via a transfilling device through the portable unit’s filling port connector. NOTE Requirements that are common to both portable units and base units are specified in ISO 18777-1.
- Standard6 pagesEnglish languagesale 15% off
- Standard6 pagesFrench languagesale 15% off
This document defines terms for respiratory care. This document is applicable: — in respiratory high-flow therapy equipment standards; — in sleep apnoea breathing therapy equipment standards; — in sleep apnoea breathing therapy masks and accessories standards; — in ventilatory support equipment standards; — in health informatics standards; — for labelling on medical electrical equipment and medical electrical systems; — in medical electrical equipment and medical electrical system instructions for use and accompanying documents; — for medical electrical equipment and medical electrical systems interoperability; — in electronic health records. This document excludes mechanical ventilation (defined in ISO 19223:2019), high-frequency and jet ventilation (defined in ISO 19223-2:2025), negative pressure ventilation, liquid ventilation, and extracorporeal membrane oxygenation.
- Standard12 pagesEnglish languagesale 15% off
This document specifies requirements for electrically powered video laryngoscopes that comprise a handle, a rigid blade, a light source, a camera and an image display that provides a clear view of the larynx to assist tracheal intubation. This document specifies requirements for accessories that are essential to the function of the video laryngoscope. This document is not applicable to the following: — laryngoscopes without a camera or image display (see ISO 7376); — endoscopes (see ISO 8600-1); — laryngoscopes designed for surgery; — image quality requirement for third party display screens, for example a phone or computer monitor, and which are not supplied by the video laryngoscope manufacturer as part of the video laryngoscope system.
- Standard22 pagesEnglish languagesale 15% off
This document applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of FUNCTIONAL NIRS EQUIPMENT, as defined in 201.3.205, intended to be used by itself, or as a part of an ME SYSTEM hereinafter referred to as ME EQUIPMENT. HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within the scope of this document are not covered by specific requirements in this document except in IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020, 7.2.13 and 8.4.1. This document is not applicable to equipment for the measurement of oxygen saturation of the haemoglobin in the micro vessels (capillaries, arterioles and venules), i.e. tissue oximeters; frequency-domain and time-domain equipment for functional near-infrared spectroscopy; equipment for the measurement of changes in the concentration of chromophores other than oxy- and deoxy-haemoglobin; equipment for the measurement of changes in the concentration of oxy- and deoxy-haemoglobin in tissues other than the brain. This document does not specify the requirements for: cerebral tissue oximeter equipment, which are given in ISO 80601-2-85; and pulse oximeter equipment, which are given in ISO 80601-2-61.
- Draft48 pagesEnglish languagesale 15% off
1.1 This document specifies requirements relating to the construction and operation of devices producing air through the blending of oxygen and nitrogen for use as sources of supply in supply systems for medical gases. 1.2 This document is applicable to proportioning units intended to produce synthetic medical air and air for driving surgical tools by mixing in defined proportions oxygen and nitrogen. 1.3 This document is applicable to proportioning units intended to be components of a medical gas supply system for medical air which supplies a medical gas pipeline distribution system complying with ISO 7396-1. 1.4 The number of proportioning units within the medical air supply system and their combination with other sources of supply (e.g. cylinder manifolds) to ensure that the supply system consists of at least three sources of supply is outside the scope of this document. Requirements for the supply systems for medical air are given in ISO 7396-1.
- Standard27 pagesEnglish languagesale 15% off
- Standard29 pagesFrench languagesale 15% off
- Standard2 pagesEnglish languagesale 15% off
- Standard2 pagesFrench languagesale 15% off
- Standard2 pagesFrench languagesale 15% off
This document defines terms for: — high-frequency oscillatory ventilation (HFOV); — percussive ventilation, including high-frequency percussive ventilation (HFPV); — jet ventilation, including high frequency jet ventilation (HFJV); — modes that combine high-frequency and physiological-rate ventilation. It is applicable: — in lung ventilator and breathing-therapy device standards, — in health informatics standards, — for labelling on medical electrical equipment and medical electrical systems, — in medical electrical equipment and medical electrical system instructions for use and accompanying documents, — for medical electrical equipment and medical electrical systems interoperability, and — in electronic health records. This document is also applicable to those accessories intended by their manufacturer to be connected to a ventilatorbreathing system or to a ventilator, where the characteristics of those accessories can affect the basic safety or essential performance of the ventilator or ventilator breathing system. NOTE This document can also be used for other applications relating to lung ventilation, including non-electrical devices and equipment, research, description of critical events, forensic analysis and adverse event (vigilance) reporting systems.
- Standard15 pagesEnglish languagesale 15% off
This document specifies the general requirements common to airway devices and related equipment.
- Standard16 pagesEnglish languagesale 15% off
- Standard17 pagesFrench languagesale 15% off
- Standard9 pagesEnglish and French languagesale 15% off
- Draft9 pagesEnglish languagesale 15% off
This document specifies requirements for fire activated oxygen shut-off devices that stop the flow of oxygen in respiratory therapy tubing when activated by fire. NOTE 1 Typical arrangements for fire activated oxygen shut-off devices are shown in Annex C. NOTE 2 Respiratory therapy tubing is covered by ISO 17256. NOTE 3 Use of fire activated oxygen shut-off devices in medical devices or accessories is not mandated in this document. The fire activated oxygen shut-off devices specified in this document are not suitable for use with oxygen therapy systems with flows in excess of 20 l/min). NOTE 4 There is rationale for this clause in A.2. The requirements in this device-specific standard take precedence over any conflicting requirements in the general standard for airway devices (ISO 18190). All the common requirements that appear in the general standard for airway devices have been removed from this document.
- Standard14 pagesEnglish languagesale 15% off
- Standard14 pagesFrench languagesale 15% off
- Standard14 pagesFrench languagesale 15% off
NOTE 1 There is guidance or rationale for this subclause contained in Clause AA.2. This document applies to the basic safety and essential performance of ventilatory support equipment, as defined in 201.3.302, for ventilatory insufficiency, as defined in 201.3.302, hereafter also referred to as ME equipment, in combination with its accessories: — intended for use in the home healthcare environment; NOTE 2 In the home healthcare environment, the supply mains driving the ventilatory support equipment is often not reliable. NOTE 3 Such ventilatory support equipment can also be used in professional health care facilities. — intended for use by a lay operator; — intended for use with patients who have ventilatory insufficiency or failure, the most fragile of which would likely experience injury with the loss of this artificial ventilation; — intended for transit-operable use; and — not intended for patients who are dependent on artificial ventilation for their immediate life support. EXAMPLE 1 Patients with moderate to severe chronic obstructive pulmonary disease (COPD), moderate amyotrophic lateral sclerosis (ALS), severe bronchopulmonary dysplasia or muscular dystrophy. Ventilatory support equipment is not considered to use a physiologic closed-loop control system unless it uses a physiological patient variable to adjust the artificial ventilation therapy settings. This document is also applicable to those accessories intended by their manufacturer to be connected to the ventilator breathing system of ventilatory support equipment for ventilatory insufficiency, where the characteristics of those accessories can affect the basic safety or essential performance of the ventilatory support equipment for ventilatory insufficiency. EXAMPLE 2 Breathing sets, connectors, water traps, expiratory valve, humidifier, breathing system filter, external electrical power source, distributed alarm system. If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601‑1:2005+AMD1:2012+AMD2:2020, 7.2.13 and 8.4.1. NOTE 4 Additional information can be found in IEC 60601‑1:2005+AMD1:2012+AMD2:2020, 4.2. NOTE 5 See ISO/TR 21954 for guidance on the selection of the appropriate ventilator for a given patient. This document does not specify the requirements for: — ventilators or accessories for ventilator-dependent patients intended for critical care applications, which are given in ISO 80601‑2‑12; — ventilators or accessories intended for anaesthetic applications, which are given in ISO 80601‑2‑13; — ventilators or accessories intended for the emergency medical services environment, which are given in ISO 80601‑2‑84; — ventilators or accessories intended for ventilator-dependent patients in the home healthcare environment, which are given in ISO 80601‑2‑72; — ventilatory support equipment or accessories intended for ventilatory impairment, which are given in ISO 80601‑2‑79; — sleep apnoea therapy ME equipment, which are given in ISO 80601‑2‑70; — high-frequency jet ventilators (HFJVs), which are given in ISO 80601‑2‑87; — high-frequency oscillatory ventilators (HFOVs); — respiratory high flow equipment, which are given in ISO 80601‑2‑90; NOTE 6 Ventilatory support equipment can incorporate high-flow therapy operational mode, but such a mode is only for spontaneously breathing patients. — user-powered resuscitators, which are given in ISO 10651-4; — gas-powered emergency resuscitators, which are given in ISO 10651-5; — oxygen therapy constant flow ME equipment; and — cuirass or “iron-lung” ventilation equipment
- Standard120 pagesEnglish languagesale 15% off
- Standard130 pagesFrench languagesale 15% off
NOTE 1 There is guidance or rationale for this subclause contained in Clause AA.2. This document applies to the basic safety and essential performance of ventilatory support equipment, as defined in 201.3.302, for ventilatory impairment, as defined in 201.3.300, hereafter also referred to as ME equipment, in combination with its accessories: — intended for use in the home healthcare environment; NOTE 2 In the home healthcare environment, the supply mains driving the ventilatory support equipment is often not reliable. NOTE 3 Such ventilatory support equipment can also be used in professional health care facilities. — intended for use by a lay operator; — intended for use with patients who have ventilatory impairment, the most fragile of these patients, would not likely experience injury with the loss of this artificial ventilation; and — not intended for patients who are dependent on artificial ventilation for their immediate life support. EXAMPLE 1 Patients with mild to moderate chronic obstructive pulmonary disease (COPD). Ventilatory support equipment is not considered to use a physiologic closed-loop control system unless it uses a physiological patient variable to adjust the artificial ventilation therapy settings. This document is also applicable to those accessories intended by their manufacturer to be connected to the breathing system of ventilatory support equipment for ventilatory impairment, where the characteristics of those accessories can affect the basic safety or essential performance of the ventilatory support equipment for ventilatory impairment. EXAMPLE 2 Breathing sets, connectors, water traps, expiratory valve, humidifier, breathing system filter, external electrical power source, distributed alarm system. If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601‑1:2005+AMD1:2012+AMD2:2020, 7.2.13 and 8.4.1. NOTE 4 Additional information can be found in IEC 60601‑1:2005+AMD1:2012+AMD2:2020, 4.2. NOTE 5 See ISO/TR 21954 for guidance on the selection of the appropriate ventilator for a given patient. This document does not specify the requirements for: — ventilators or accessories for ventilator-dependent patients intended for critical care applications, which are given in ISO 80601‑2‑12; — ventilators or accessories intended for anaesthetic applications, which are given in ISO 80601‑2‑13; — ventilators or accessories intended for the emergency medical services environment, which are given in ISO 80601‑2‑84; — ventilators or accessories intended for ventilator-dependent patients in the home healthcare environment, which are given in ISO 80601‑2‑72; — ventilatory support equipment or accessories intended for ventilatory insufficiency, which are given in ISO 80601‑2‑80; — sleep apnoea therapy ME equipment, which are given in ISO 80601‑2‑70; — high-frequency jet ventilators (HFJVs), which are given in ISO 80601‑2‑87; — high-frequency oscillatory ventilators (HFOVs); — respiratory high flow equipment, which are given in ISO 80601‑2‑90; NOTE 6 Ventilatory support equipment can incorporate high-flow therapy operational mode, but such a mode is only for spontaneously breathing patients. — user-powered resuscitators, which are given in ISO 10651-4; — gas-powered emergency resuscitators, which are given in ISO 10651-5; — oxygen therapy constant flow ME equipment; and — cuirass or “iron-lung” ventilation equipment.
- Standard107 pagesEnglish languagesale 15% off
- Standard118 pagesFrench languagesale 15% off
This document specifies requirements for anaesthetic reservoir bags for use with anaesthetic and ventilator breathing systems. It includes requirements for the design of the neck, size designation and elasticity. This document is not applicable to special-purpose bags, for example bellows, self-inflating bags and bags for use with anaesthetic gas scavenging systems. The requirements in this device-specific standard take precedence over any conflicting requirements in the general standard for airway devices (ISO 18190). All the common requirements that appear in the general standard for airway devices have been removed from this document.
- Standard15 pagesEnglish languagesale 15% off
- Standard15 pagesFrench languagesale 15% off
This document specifies requirements for the respiratory tubing and connectors used to convey respirable gases to a patient in the healthcare and homecare environments and provide a safe connection between the gas supply device and the patient interface. Respiratory tubing and connectors are mainly used for delivery of oxygen but can also be used for respirable air or oxygen/air mixtures and breathable medicinal gas mixtures such as oxygen/nitrous oxide or oxygen/helium mixtures. This document also specifies requirements for respiratory therapy extension tubing. NOTE 1 The gas supply devices referred to in this document do not include anaesthetic machines/workstations and ventilators. NOTE 2 This document does not cover breathing tubes for breathing systems. These are specified in ISO 5367. This document is written following the format of ISO 18190, General standard for airways and related equipment. The requirements in this device-specific standard take precedence over any conflicting requirements in the General standard
- Standard12 pagesEnglish languagesale 15% off
- Standard12 pagesFrench languagesale 15% off
This document specifies requirements and guidelines for systems and equipment used to evacuate plume generated by medical devices. This document applies to all types of plume evacuation systems (PESs), including a) portable; b) mobile; c) stationary, including dedicated central pipelines; d) PESs integrated into other equipment; e) PESs for endoscopic procedures (e.g., minimally invasive, laparoscopic) This document applies to all healthcare facilities where PESs are used, including, but not limited to a) surgical facilities; b) medical offices; c) cosmetic treatment facilities; d) medical teaching facilities; e) dental clinics; f) veterinary facilities. This document provides guidance on the following aspects of PESs: a) importance; b) purchasing; c) design; d) manufacture; e) documentation; f) function; g) performance; h) installation; i) commissioning; j) testing; k) training; l) use; m) risk assessment; n) servicing; o) maintenance. This document does not apply to the following: a) anaesthetic gas scavenging systems (AGSSs) which are covered in ISO 7396-2; b) medical vacuum systems which are covered in ISO 7396-1; c) heating, ventilation, and air-conditioning (HVAC) systems; d) aspects of laser safety other than airborne contamination; and e) aspects of electrosurgery, electrocautery, and mechanical surgical tools other than airborne contamination produced by such equipment resulting from interaction with tissue or materials.
- Standard37 pagesEnglish languagesale 15% off
- Standard40 pagesFrench languagesale 15% off
- Standard5 pagesEnglish languagesale 15% off
- Standard5 pagesFrench languagesale 15% off
This document specifies: — the general principles governing the biological evaluation within a risk management process of the gas pathways of a medical device, its parts or accessories, which are intended to provide respiratory care or supply substances via the respiratory tract to a patient in all environments; — the general categorization of gas pathways based on the nature and duration of their contact with the gas stream; — the evaluation of existing relevant data from all sources; — the identification of gaps in the available data set on the basis of a risk analysis; — the identification of additional data sets necessary to analyse the biological safety of the gas pathway; — the assessment of the biological safety of the gas pathway. This document covers general principles regarding biocompatibility assessment of medical device materials, which make up the gas pathway, in normal use and normal condition. This document does not cover biological hazards arising from mechanical damage. The other parts of ISO 18562 cover specific tests that address potentially hazardous substances that are added to the respirable gas stream and establish acceptance criteria for these substances. This document addresses potential contamination of the gas stream arising from the gas pathways within the medical device, which might then be conducted to the patient. This document applies over the expected lifetime of the medical device when operated according to the instructions for use. This includes degradation arising from exposure to environmental conditions as well as cleaning, disinfection and sterilisation (i.e. processing). It also includes user action or inaction (omission) that leads to an unintended or unexpected outcome (result) (i.e. use error). It does not include conscious/intentional action or inaction that violates the instructions for use and is beyond reasonable risk control by the manufacturer (i.e. abnormal use). This document does not address biological evaluation of the surfaces of medical devices that have direct contact with the patient or user. The requirements for direct contact surfaces are found in the ISO 10993 series. Medical devices, parts or accessories containing gas pathways that are addressed by this document include, but are not limited to, ventilators, anaesthesia workstations (including gas mixers), breathing systems, oxygen conserving equipment, oxygen concentrators, nebulizers, low-pressure hose assemblies, humidifiers, heat and moisture exchangers, respiratory gas monitors, respiration monitors, masks, medical respiratory personal protective equipment, mouth pieces, resuscitators, breathing tubes, breathing system filters and Y-pieces as well as any breathing accessories intended to be used with such medical devices. The enclosed chamber of an incubator, including the mattress, and the inner surface of an oxygen hood are considered to be gas pathways and are also addressed by this document. This document does not address contamination already present in the gas supplied from the gas sources while medical devices are in normal use. EXAMPLE Contamination arriving at the medical device from gas sources such as medical gas pipeline systems (including the non-return valves in the pipeline outlets), outlets of pressure regulators connected or integral to a medical gas cylinder, or room air taken into the medical device is not addressed by ISO 18562 (all parts).
- Standard34 pagesEnglish languagesale 15% off
- Standard35 pagesFrench languagesale 15% off
This document specifies tests for the emissions of particulate matter from the gas pathways of a medical device, its parts or accessories, which are intended to provide respiratory care or supply substances via the respiratory tract to a patient in all environments. The tests of this document are intended to quantify particles from 0,25 µm diameter to 10 µm diameter that are emitted by the medical device, its parts or accessories into the respirable gas stream. This document establishes acceptance criteria for these tests. This document does not address nanoparticles. Insufficient data exist to establish exposure limits for particles less than 0,25 µm diameter. This document does not address particles larger than 10 µm diameter. These particles are deposited in the nasal cavity. Additional information can be needed for medical devices or accessories that bypass the nose. This is outside the scope of this document but can be required by some authorities having jurisdiction. This document therefore adopts the same approach as the US Environmental Protection Agency (EPA) in setting limits based solely on particle size and not their chemistry. This document addresses potential contamination of the gas stream arising from the gas pathways, which is then conducted to the patient. This document applies over the expected lifetime of the medical device in normal use and takes into account the effects of any intended processing. This document does not address biological evaluation of the particles that are deliberately released by a nebulizer (i.e. the therapeutic agent). This document does not address biological evaluation of the surfaces of gas pathways that have direct contact with the patient. The requirements for direct contact surfaces are found in the ISO 10993 series. Medical devices, parts or accessories, containing gas pathways that are addressed by this document, include, but are not limited to, ventilators, anaesthesia workstations (including gas mixers), breathing systems, oxygen conserving devices, oxygen concentrators, nebulizers, low-pressure hose assemblies, humidifiers, heat and moisture exchangers, respiratory gas monitors, respiration monitors, masks, medical respiratory personal protective equipment, mouth pieces, resuscitators, breathing tubes, breathing systems filters, Y-pieces, and any breathing accessories intended to be used with such devices. The enclosed chamber of an incubator, including the mattress, and the inner surface of an oxygen hood are considered to be gas pathways and are also addressed by this document. This document does not address contamination already present in the gas supplied from the gas sources while medical devices are in normal use. EXAMPLE Contamination arriving at the medical device from gas sources such as medical gas pipeline systems (including the non-return valves in the pipeline outlets), outlets of pressure regulators connected or integral to a medical gas cylinder or room air taken into the medical device is not addressed by ISO 18562 (all parts).
- Standard16 pagesEnglish languagesale 15% off
- Standard16 pagesFrench languagesale 15% off
This document specifies tests for substances leached by liquid water condensing in gas pathways of a medical device, its parts or accessories, which are intended to provide respiratory care or supply substances via the respiratory tract to a patient in all environments. The chemical characterization methods described in this document apply to chemical substances that could leach from the medical device, its parts or accessories into the condensate. This document establishes verifiable acceptance criteria for these tests. The identity and quantity of each chemical released is intended for toxicological risk assessment as described in ISO 18562-1:2024. This document addresses potential contamination of the gas stream arising from the gas pathways, which deliver breathing gas to the patient. This document applies over the expected lifetime of the medical device in normal use and takes into account the effects of any intended processing. This document does not address biological evaluation of the surfaces of gas pathways that have direct contact with the patient. The requirements for direct contact surfaces are found in the ISO 10993 series. Medical devices, parts or accessories containing gas pathways that are addressed by this document include, but are not limited to, ventilators, anaesthesia workstations (including gas mixers), breathing systems, oxygen conserving devices, oxygen concentrators, nebulizers, low-pressure hose assemblies, humidifiers, heat and moisture exchangers, respiratory gas monitors, respiration monitors, masks, medical respiratory personal protective equipment, mouth pieces, resuscitators, breathing tubes, breathing systems filters, Y-pieces and any breathing accessories intended to be used with such devices. The enclosed chamber of an incubator, including the mattress, and the inner surface of an oxygen hood are considered to be gas pathways and are also addressed by this document. This document does not address contamination already present in the gas supplied from the gas sources while medical devices are in normal use. EXAMPLE Contamination arriving at the medical device from gas sources such as medical gas pipeline systems (including the non-return valves in the pipeline outlets), outlets of pressure regulators connected or integral to a medical gas cylinder, or room air taken into the medical device. This document does not address contact with drugs or anaesthetic agents. If a medical device or accessory is intended to be used with anaesthetic agents or drugs, then additional testing can be required. This document is intended to quantify hazardous water-soluble substances that are leached from the medical device, its parts or accessories by condensate and then conveyed by that liquid to the patient.
- Standard16 pagesEnglish languagesale 15% off
- Standard17 pagesFrench languagesale 15% off
This document specifies tests for the emissions of volatile organic substances from the gas pathways of a medical device, its parts or accessories, which are intended to provide respiratory care or supply substances via the respiratory tract to a patient in all environments. The tests of this document are intended to quantify emissions of volatile organic substances that are added to the respirable gas stream by the materials of the gas pathway. This document establishes acceptance criteria for these tests. NOTE Gaseous emission of volatile organic substances includes emissions of volatile organic compounds, semi-volatile organic compounds and very volatile organic compounds. This document addresses potential contamination of the gas stream arising from the gas pathways of medical devices or accessories, which is then conducted to the patient. This document applies over the expected lifetime of the medical device in normal use and takes into account the effects of any intended processing. This document does not address biological evaluation of the surfaces of gas pathways that are in direct contact with the patient. The requirements for direct contact surfaces are found in the ISO 10993 series. Medical devices, parts or accessories containing gas pathways that are addressed by this document include, but are not limited to, ventilators, anaesthesia workstations (including gas mixers), breathing systems, oxygen conserving devices, oxygen concentrators, nebulizers, low-pressure hose assemblies, humidifiers, heat and moisture exchangers, respiratory gas monitors, respiration monitors, masks, medical respiratory personal protective equipment, mouth pieces, resuscitators, breathing tubes, breathing systems filters, Y-pieces and any breathing accessories intended to be used with such devices. The enclosed chamber of an incubator, including the mattress, and the inner surface of an oxygen hood are considered to be gas pathways and are also addressed by this document. This document does not address contamination already present in the gas supplied from the gas sources while medical devices are in normal use. EXAMPLE Contamination arriving at the medical device from gas sources such as medical gas pipeline systems (including the non-return valves in the pipeline outlets), outlets of pressure regulators connected or integral to a medical gas cylinder or room air taken into the medical device is not addressed by ISO 18562 series. This document is intended to be read in conjunction with ISO 18562-1.
- Standard15 pagesEnglish languagesale 15% off
- Standard16 pagesFrench languagesale 15% off
- Standard6 pagesEnglish and French languagesale 15% off
- Draft6 pagesEnglish languagesale 15% off
This document provides the essential requirements for the design of supralaryngeal airways and connectors. These devices are intended to provide a distinct respiratory pathway to the top of the larynx to provide an unobstructed airway in patients during spontaneous, assisted or controlled ventilation. This document specifies the dimensions, basic properties and method of size designation of the available types of supralaryngeal airways. Airways devised for specialized applications are not specifically covered, although most may be classified by the sizing and dimensions (or other characteristics) required by this document. The following devices are outside the scope of this document: nasal and oropharyngeal airways, anesthetic masks, oro- and naso-tracheal tubes, cricothyrotomy devices, dental appliances, tracheal stents, tracheal tubes, ventilating laryngoscopes, CPAP devices, esophageal obturators, bougies and devices that require surgical placement. This document specifies dimensional disclosure so the operator will know which auxiliary devices, such as tracheal tubes and bronchoscopes will be size-compatible. Flammability of airways, for example if used with certain flammable anesthetic gases, electrosurgical units or lasers, is a well-recognized hazard that is outside the scope of this document. (See E.1.7).
- Standard29 pagesEnglish languagesale 15% off
This document applies to the basic safety and essential performance of a critical care ventilator in combination with its accessories, hereafter referred to as ME equipment: ¾ intended for use in an environment that provides specialized care for patients whose conditions can be life-threatening and who can require comprehensive care and constant monitoring in a professional healthcare facility; NOTE 2 For the purposes of this document, such an environment is referred to as a critical care environment. Ventilators for this environment are considered life-sustaining. NOTE 3 For the purposes of this document, such a critical care ventilator can provide ventilation during transport within a professional healthcare facility (i.e. be a transit-operable ventilator). NOTE 4 A critical care ventilator intended for use in transport within a professional healthcare facility is not considered as an emergency medical services environment ventilator. ¾ intended to be operated by a healthcare professional operator; and ¾ intended for those patients who need differing levels of support from artificial ventilation including for ventilator-dependent patients. A critical care ventilator is not considered to use a physiologic closed-loop-control system unless it uses a physiological patient variable to adjust the artificial ventilation therapy settings. This document is also applicable to those accessories intended by their manufacturer to be connected to a ventilator breathing system, or to a ventilator, where the characteristics of those accessories can affect the basic safety or essential performance of the ventilator. NOTE 5 If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601-1:2005+AMD1:2012+AMD2:2020, 7.2.13 and 8.4.1. NOTE 6 Additional information can be found in IEC 60601-1:2005+AMD1:2012+AMD2:2020, 4.2. This document is not applicable to ME equipment or an ME system operating in a ventilator-operational mode solely intended for patients who are not dependent on artificial ventilation. NOTE 7 A critical care ventilator, when operating in such a ventilator-operational mode, is not considered life-sustaining. This document is not applicable to ME equipment that is intended solely to augment the ventilation of spontaneously breathing patients within a professional healthcare facility. This document does not specify the requirements for: NOTE 8 See ISO/TR 21954 for guidance on the selection of the appropriate ventilator for a given patient. ¾ ventilators or accessories intended for anaesthetic applications, which are given in ISO 80601‑2‑13; ¾ ventilators or accessories intended for the emergency medical services environment, which are given in ISO 80601-2-84; ¾ ventilators or accessories intended for ventilator-dependent patients in the home healthcare environment, which are given in ISO 80601‑2-72; ¾ ventilators or accessories intended for home-care ventilatory support devices, which are given in ISO 80601-2-79 and ISO 80601-2-80; ¾ obstructive sleep apnoea therapy ME equipment, which are given in ISO 80601‑2‑70; ¾ continuous positive airway pressure (CPAP) ME equipment. ¾ high-frequency ventilators, which are given in ISO 80601‑2‑87; NOTE 9 A critical care ventilator can incorporate high-frequency jet or high-frequency oscillatory ventilator-operational modes. ¾ respiratory high-flow therapy equipment, which are given in ISO 80601‑2‑90; NOTE 10 A critical care ventilator can incorporate high-flow therapy operational mode, but such a mode is only for spontaneously breathing p
- Standard158 pagesEnglish languagesale 15% off
- Standard171 pagesFrench languagesale 15% off
This document applies to the basic safety and essential performance of an EMS ventilator in combination with its accessories, hereafter also referred to as ME equipment: ¾ intended for patients who need differing levels of support from artificial ventilation including ventilator-dependent patients; ¾ intended to be operated by a healthcare professional operator; ¾ intended for use in the EMS environment; and ¾ intended for invasive or non-invasive ventilation. NOTE 2 An EMS ventilator can also be used for transport within a professional healthcare facility. An EMS ventilator is not considered to use a physiologic closed loop-control system unless it uses a physiological patient variable to adjust the artificial ventilation therapy settings. This document is also applicable to those accessories intended by their manufacturer to be connected to the ventilator breathing system, or to an EMS ventilator, where the characteristics of those accessories can affect the basic safety or essential performance of the EMS ventilator. NOTE 3 If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601‑1:2005+AMD1:2012+AMD2:2020, 7.2.13 and 8.4.1. NOTE 4 Additional information can be found in IEC 60601-1:2005+AMD1:2012+AMD2:2020, 4.2. This document does not specify the requirements for the following: NOTE 5 See ISO/TR 21954 for guidance on the selection of the appropriate ventilator for a given patient. ¾ ventilators or accessories intended for ventilator-dependent patients in critical care applications, which are given in ISO 80601-2-12. ¾ ventilators or accessories intended for ventilator-dependent patients in the home healthcare environment, which are given in ISO 80601-2-72. ¾ ventilators or accessories intended for anaesthetic applications, which are given in ISO 80601‑2‑13. ¾ ventilators or accessories intended for ventilatory support equipment (intended only to augment the ventilation of spontaneously breathing patients), which are given in ISO 80601‑2-79 and ISO 80601-2-80. ¾ obstructive sleep apnoea therapy ME equipment, which are given in ISO 80601‐2‐70. ¾ user-powered resuscitators, which are given in ISO 10651‐4. ¾ gas-powered emergency resuscitators, which are given in ISO 10651‐5. ¾ continuous positive airway pressure (CPAP) ME equipment. ¾ high‐frequency jet ventilators (HFJVs), which are given in ISO 80601-2-87. ¾ high‐frequency oscillatory ventilators (HFOVs) REF _Ref69393406 \r \h \* MERGEFORMAT [44] 08D0C9EA79F9BACE118C8200AA004BA90B02000000080000000D0000005F00520065006600360039003300390033003400300036000000 , which are given in ISO 80601-2-87. NOTE 6 An EMS ventilator can incorporate high-frequency jet or high-frequency oscillatory ventilation-modes. ¾ respiratory high-flow therapy equipment, which are given in ISO 80601-2-90. NOTE 7 An EMS ventilator can incorporate high-flow therapy operational mode, but such a mode is only for spontaneously breathing patients. ¾ oxygen therapy constant flow ME equipment. ¾ cuirass or “iron‐lung” ventilators.
- Standard138 pagesEnglish languagesale 15% off
- Standard148 pagesFrench languagesale 15% off
- Standard1 pageEnglish languagesale 15% off
- Standard1 pageFrench languagesale 15% off
This document specifies requirements for flow control devices that can be connected by the user either directly, by means of a probe or a gas-specific connector, or indirectly by means of a low-pressure hose assembly conforming with ISO 5359 to: a) a terminal unit conforming with ISO 9170-1 of a medical gas pipeline system conforming with ISO 7396-1:2016; b) the pressure outlet of a regulator conforming with ISO 10524-1:2018; or c) to the pressure outlet of a valve integrated pressure regulator (VIPR) conforming with ISO 10524-3 (see 5.2 gas inlets). This document applies to the following types of flow control devices (FCDs): a) flowmeters; b) flowgauge FCDs; and c) fixed orifice FCDs. NOTE Flow control devices that are classed as medical electrical equipment can be subject to additional requirements of IEC 60601-1. This document applies to flow control devices for the following gases: — oxygen; — oxygen 93 %; — nitrous oxide; — medical air; — carbon dioxide; — oxygen/nitrous oxide mixture 50/50 (% volume fraction); — oxygen-enriched air; — helium; — xenon; and — specified mixtures of the gases listed above. NOTE Flow control devices can be available for other gases. This document does not apply to flow control devices that are: a) for use with gases for driving surgical tools; b) an integral part of a regulator (see ISO 10524-1:2018); or c) an integral part of a valve with integrated pressure regulator (VIPR) (see ISO 10524-3).
- Standard22 pagesEnglish languagesale 15% off
- Standard23 pagesFrench languagesale 15% off
This document specifies requirements for the safety and performance testing of general-purpose nebulizing systems intended for continuous or breath-actuated delivery of liquids, in aerosol form, to humans through the respiratory system. This document includes gas-powered nebulizers (which can be powered by, e.g., compressors, pipeline systems, cylinders, etc.) and electrically powered nebulizers [e.g. spinning disc, ultrasonic, vibrating mesh (active and passive), and capillary devices] or manually powered nebulizers. This document does not specify the electrical requirements of electrically powered nebulizers. This document does not specify the minimum performance of nebulizing systems. This document does not apply to: a) devices intended for nasal deposition; b) devices intended solely to provide humidification or hydration by providing water in aerosol form. NOTE 1 ISO 80601-2-74 and ISO 20789 cover these devices. c) drug-specific nebulizers or their components (e.g. metered dose inhalers, metered liquid inhalers, dry powder inhalers). NOTE 2 ISO 20072 covers these devices. NOTE 3 See Annex A for rationale.
- Standard33 pagesEnglish languagesale 15% off
- Standard35 pagesFrench languagesale 15% off
This document specifies minimum requirements for breathing sets and breathing tubes intended to be used with anaesthetic breathing systems, ventilator breathing systems, humidifiers or nebulizers. It applies to breathing sets and breathing tubes and patient end adaptors supplied already assembled and to those supplied as components and assembled in accordance with the manufacturer’s instructions. This document is applicable to breathing sets which include special components (e.g. water traps) between the patient end and machine end. Provision is made for coaxial and related bifurcated, double-lumen, or multiple-lumen breathing sets and breathing tubes suitable for use with patient end adaptors.
- Standard27 pagesEnglish languagesale 15% off
- Standard28 pagesFrench languagesale 15% off
This document applies to the basic safety and essential performance of a ventilator in combination with its accessories, hereafter referred to as ME equipment: — intended for use in the home healthcare environment; — intended for use by a lay operator; and — intended for those patients who need differing levels of support from artificial ventilation including for ventilator-dependent patients. This document is also applicable to those accessories intended by their manufacturer to be connected to a ventilator breathing system or to a ventilator where the characteristics of those accessories can affect the basic safety or essential performance of the ventilator.
- Standard142 pagesEnglish languagesale 15% off
- Standard153 pagesFrench languagesale 15% off
This document specifies requirements for user-powered resuscitators intended for use with all age groups and which are intended to provide lung ventilation to patients whose breathing is inadequate. User-powered resuscitators are designated according to ideal body mass range. Example user-powered resuscitators include: — self-inflating bag resuscitators intended to be squeezed by the user’s hand and refilled by elastic recoil; and NOTE 1 Self-inflating bag resuscitators are generally transit-operable and can be used in a wide range of environmental and emergency situations. — flow-inflating bag resuscitators intended to be squeezed by the user’s hand and refilled by a flow from a medical gas source. This document is also applicable to those accessories that are intended for use with resuscitators where the characteristics of those accessories can affect the safety of the user-powered resuscitator. Examples of such accessories include face masks, PEEP valves, capnometric indicators, manometers, metronomes, flow restrictors, filters, gas refill valves, oxygen gas mixers, connectors, electronic feedback devices, electronic sensors and transmission of data to other equipment. This document is also applicable to point-of-use packaging. This document does not specify the requirements for: — gas-powered emergency resuscitators, which are given in ISO 10651-5; — electrically-powered resuscitators; — gas powered resuscitators for professional healthcare facilities; and — anaesthetic reservoir bags, which are given in ISO 5362. NOTE 2 This document has been prepared to address the relevant essential principles[24] and labelling[25] guidances of the International Medical Devices Regulators Forum (IMDRF) as indicated in Annex D. NOTE 3 This document has been prepared to address the relevant essential principles of safety and performance of ISO 16142-1:2016 as indicated in Annex E. NOTE 4 This document has been prepared to address the relevant general safety and performance requirements of European regulation (EU) 2017/745[23] as indicated in Annex F.
- Standard66 pagesEnglish languagesale 15% off
- Standard80 pagesFrench languagesale 15% off
This document provides specific requirements for the basic safety and essential performance for oro-tracheal and naso-tracheal tubes and tracheal tube connectors, tracheal tubes with walls reinforced with metal or plastic, tracheal tubes with shoulders, tapered tracheal tubes, tracheal tubes with means for suctioning, monitoring or delivery of drugs or other gases, and the many other types of tracheal tubes devised for specialized applications. Tracheobronchial (including endobronchial) tubes (see ISO 16628), tracheostomy tubes (see ISO 5366), and supralaryngeal airways (see ISO 11712) are excluded from the scope of this document. Tracheal tubes intended for use with flammable anaesthetic gases or agents, lasers, or electrosurgical equipment are outside the scope of this document. NOTE 1 There is guidance or rationale for this clause contained in Annex A.2. NOTE 2 ISO 11990-1, ISO 11990-2, and ISO 14408 deal with laser surgery of the airway.
- Standard45 pagesEnglish languagesale 15% off
- Standard46 pagesFrench languagesale 15% off
This document specifies the requirements and methods for the clinical investigation of continuous automated non-invasive sphygmomanometers used for the measurement of the blood pressure of a patient. This document does not cover usability aspects such as the form and manner of the data display or output. This document does not specify a numerical threshold on the minimum output period. A continuous automated non-invasive sphygmomanometer providing blood pressure parameters (e.g., systolic blood pressure, diastolic blood pressure or mean arterial pressure) with an output period considerably larger than 30 s is not typically considered a continuous automated non-invasive sphygmomanometer. This document covers both trending continuous automated non-invasive sphygmomanometers and absolute accuracy continuous automated non-invasive sphygmomanometers and focuses solely on requirements for the clinical investigation. Representation of output is not covered by this document. NOTE 1 IEC 62366-1 provides requirements on the application of usability engineering to medical devices. The usability engineering process can be used to clarify for the intended user whether the displayed data concerns absolute accurate values or trending values. The requirements and methods for the clinical investigation of continuous automated non-invasive sphygmomanometers provided in this document are applicable to any subject population, and any condition of use of the continuous automated non-invasive sphygmomanometers. NOTE 2 Subject populations can, for example, be represented by age or weight ranges. NOTE 3 This document does not provide a method to assess the effect of artefacts during the clinical investigation (e.g. motion artefacts induced by the movement of the subject or the movement of the platform supporting the subject). This document specifies additional disclosure requirements for the accompanying documents of continuous automated non-invasive sphygmomanometers that have undergone clinical investigation according to this document. This document is not applicable to: — the clinical investigation of a non-automated sphygmomanometer as given in ISO 81060-1, — the clinical investigation of an intermittent automated non-invasive sphygmomanometer as given in ISO 81060-2, — an automated non-invasive sphygmomanometer as given in IEC 80601-2-30, or — invasive blood pressure monitoring equipment as given in IEC 60601‑2‑34.
- Standard36 pagesEnglish languagesale 15% off
- Standard40 pagesFrench languagesale 15% off
This document specifies requirements for low-flow nasal cannulae, used in both home care and hospital environments for the administration of oxygen therapy. This document does not include requirements to prevent the proliferation of fire within the tubing but does specify a user-detachable connection that can be used to fit a fire-activated oxygen shut-off device.
- Standard9 pagesEnglish languagesale 15% off
- Standard9 pagesFrench languagesale 15% off
This document applies to the basic safety and essential performance of an infant cardiorespiratory monitor, as defined in 3.10, hereafter also referred to as ME equipment, in combination with its accessories: — intended for use in the home healthcare environment; — intended for use by a lay operator; — intended to monitor cardiorespiratory parameters in sleeping or resting children under three years of age; and — intended for transit-operable use. NOTE An infant cardiorespiratory monitor can also be used in professional health care facilities. This document is also applicable to those accessories intended by their manufacturer to be connected to the infant cardiorespiratory monitor, where the characteristics of those accessories can affect the basic safety or essential performance of the infant cardiorespiratory monitor. EXAMPLE probes, cables distributed alarm system
- Standard57 pagesEnglish languagesale 15% off
- Standard60 pagesFrench languagesale 15% off
This document specifies requirements for safety, materials, design and information supplied with tracheobronchial tubes. These devices are used when isolation of the airways of one or both lungs is required. Tracheal tubes that include bronchus blockers are excluded from the scope of this document
- Standard13 pagesEnglish languagesale 15% off
- Standard13 pagesFrench languagesale 15% off
This document specifies essential performance and safety requirements for cuff pressure indicators used to indicate the intracuff pressure of airway devices, such as supralaryngeal airways, tracheal tubes or tracheostomy tubes. This document is also applicable to devices that combine intracuff pressure indication with a method of cuff inflation (such as a syringe or pump). The device can also provide a method of automatically maintaining cuff inflation at a specific pressure or within a pressure range. The requirements specified in this document apply to stand-alone cuff pressure indicators and those integrated into other medical devices e.g. ventilators anaesthesia workstations etc.
- Standard9 pagesEnglish languagesale 15% off
- Standard10 pagesFrench languagesale 15% off
This document specifies minimum performance and safety requirements for air entrainment devices used for delivery of designated oxygen concentrations to patients. It provides a test method to check the accuracy of the oxygen concentration in the air/oxygen mixture generated by the air entrainment devices. Air entrainment devices can be fixed to deliver a single oxygen concentration or adjustable, to deliver a range of oxygen concentration outputs. This document also specifies marking requirements and recommends an optional system of colour coding to assist the user in identifying the designated oxygen concentration. This document does not cover air entrainment devices which are integral with medical devices specified in other standards (e.g. emergency lung ventilators, humidifiers, nebulizers).
- Standard8 pagesEnglish languagesale 15% off
- Standard8 pagesFrench languagesale 15% off
This document is applicable to the basic safety and essential performance of an anaesthetic workstation for administering inhalational anaesthesia whilst continuously attended by a professional operator. This document specifies particular requirements for a complete anaesthetic workstation and the following anaesthetic workstation components which, although considered as individual devices in their own right, may be utilized, in conjunction with other relevant anaesthetic workstation components, to form an anaesthetic workstation to a given specification: anaesthetic gas delivery system; anaesthetic breathing system; anaesthetic gas scavenging system (AGSS); anaesthetic vapour delivery system; anaesthetic ventilator; monitoring equipment; alarm system; protection device. NOTE 1 Monitoring equipment, alarm systems and protection devices are summarized in Table AA.1. An anaesthetic workstation supplied complete and its individual components are considered as ME equipment or ME systems with regard to the general standard. NOTE 2 The applicability of this document is indicated in Table AA.2. This document is also applicable to those accessories intended by their manufacturer to be connected to an anaesthetic workstation where the characteristics of those accessories can affect the basic safety and essential performance of the anaesthetic workstation. If a clause or subclause is specifically intended to be applicable to anaesthetic workstation components or its accessories only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to an anaesthetic workstation and its individual components including accessories, as relevant. Hazards inherent in the intended physiological function of an anaesthetic workstation and its individual components including accessories within the scope of this document are not covered by specific requirements in this document except in IEC 60601-1:2005+AMD1:2012+AMD2:2020, 7.2.13 and 8.4.1. NOTE 3 See also IEC 60601-1:2005+AMD1:2012+AMD2:2020, 4.2. This document is not applicable to any anaesthetic workstation intended for use with flammable anaesthetic agents, as determined by Annex BB.
- Standard103 pagesEnglish languagesale 15% off
- Standard122 pagesFrench languagesale 15% off
This document specifies safety and performance requirements for electrically powered medical and surgical suction equipment. It applies to equipment used in health care facilities such as hospitals, for domiciliary care of patients and for field use and transport use.
- Standard3 pagesEnglish languagesale 15% off
- Standard3 pagesFrench languagesale 15% off
This document specifies safety and performance requirements for manually powered suction equipment intended for oro-pharyngeal suction. It applies to equipment operated by foot or by hand or both. The commonest use of manually powered suction is in situations outside of healthcare settings often described as field use or transport use. Use in these situations may involve extreme conditions of weather or terrain. Additional/alternative requirements for manually powered suction equipment intended for field use or transport use are included in this document. This document does not apply to mucus extractors.
- Standard2 pagesEnglish languagesale 15% off
- Standard2 pagesFrench languagesale 15% off
This document specifies basic safety and performance requirements for medical suction equipment powered from a vacuum or positive pressure gas source generating venturi suction. It applies to suction equipment connected to medical gas pipeline systems or cylinders and venturi attachments and can be standalone or part of an integrated system.
- Standard3 pagesEnglish languagesale 15% off
- Standard3 pagesFrench languagesale 15% off
This document establishes a vocabulary of terms used for anaesthetic and respiratory equipment and supplies, related devices and supply systems. NOTE 1 To avoid multiple definitions of the same term in different categories, this document attempts to ensure consistency by the inclusion of a ‘general’ category, and by use of domain specifiers and unique pre-coordinated domain-specific term names. NOTE 2 In addition to terms and definitions used in two of the three official ISO languages (English and French), this document gives the equivalent terms in the German language; these are published under the responsibility of the member body for Germany. However, only the terms and definitions given in the official languages can be considered as ISO terms and definitions.
- Standard6 pagesEnglish languagesale 15% off
- Standard195 pagesSlovenian, English, German and French languagesale 15% off
This document specifies general requirements for medical suction equipment that are common to all parts of the ISO 10079 series. This document is not applicable to the following: a) end-pieces such as suction catheters, drains, curettes, Yankauer suckers and suction tips; b) syringes; c) dental suction equipment; d) anaesthetic gas scavenging systems; e) laboratory suction; f) autotransfusion systems; g) mucus extractors including neonatal mucus extractors; h) suction equipment where the collection container is downstream of the vacuum pump; i) ventouse (obstetric) equipment; j) suction equipment marked for endoscopic use only; and k) plume evacuation systems.
- Standard35 pagesEnglish languagesale 15% off
- Standard36 pagesFrench languagesale 15% off
This document applies to the basic safety and essential performance of respiratory high-flow therapy equipment, as defined in 201.3.220, hereafter also referred to as ME equipment or ME system, in combination with its accessories: — intended for use with patients who can breathe spontaneously; and — intended for patients who would benefit from improved alveolar gas exchange; and who would benefit from receiving high-flow humidified respiratory gases, which can include a patient whose upper airway is bypassed. EXAMPLE 1 Patients with Type 1 Respiratory Failure who exhibit a reduction in arterial blood oxygenation. EXAMPLE 2 Patients who would benefit from reduced work of breathing, as needed in Type 2 Respiratory Failure, where arterial carbon dioxide is high. EXAMPLE 3 Patients requiring humidification to improve mucociliary clearance. Respiratory high-flow therapy equipment can be intended for use in the home healthcare environment or intended for use in professional healthcare facilities. NOTE 1 In the home healthcare environment, the supply mains is often not reliable. Respiratory high-flow therapy equipment can be: — fully integrated ME equipment; or — a combination of separate items forming a ME system. This standard also applies to other types of respiratory equipment when that equipment includes a respiratory high-flow therapy mode. NOTE 2 This standard and ISO 80601-2-12[14] are applicable to a critical care ventilator with a high-flow therapy mode. Respiratory high-flow therapy equipment can be transit-operable. This document is also applicable to those accessories intended by their manufacturer to be connected to the respiratory high-flow therapy equipment, where the characteristics of those accessories can affect the basic safety or essential performance of the respiratory high-flow therapy equipment. EXAMPLE 4 Breathing sets, connectors, humidifier, breathing system filter, external electrical power source, distributed alarm system, high-flow nasal cannula, tracheal tube, tracheostomy tube, face mask and supra-laryngeal airway. If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in the general standard, 7.2.13 and 8.4.1. NOTE 3 Additional information can be found in the general standard, 4.2. This document does not specify the requirements for: — ventilators or accessories for ventilator-dependent patients intended for critical care applications, which are given in ISO 80601‑2‑12[14]; — ventilators or accessories intended for anaesthetic applications, which are given in ISO 80601‑2‑13[15]; — ventilators or accessories intended for the emergency medical services environment, which are given in ISO 80601‑2‑84[20]; — ventilators or accessories intended for ventilator-dependent patients in the home healthcare environment, which are given in ISO 80601‑2‑72[17]; — ventilatory support equipment or accessories intended for patients with ventilatory impairment, which are given in ISO 80601‑2‑79[18]; — ventilatory support equipment or accessories intended for patients with ventilatory insufficiency, which are given in ISO 80601‑2‑80[19]; — sleep apnoea therapy ME equipment, which are given in ISO 80601‑2‑70[16]; — continuous positive airway pressure (CPAP) ME equipment; — high-frequency jet ventilators (HFJVs)[31], which are given in ISO 80601‑2‑87[21]; — gas mixers for medical use, which are given in ISO 11195[9]; — flowmeters, which are given in ISO 15002[11]; — high-frequency oscillatory ventilators (HFOVs), which are given in ISO 80601‑2‑87[21]; and — cuirass or “iron-lung” venti
- Standard89 pagesEnglish languagesale 15% off
- Standard100 pagesFrench languagesale 15% off
This document specifies performance requirements for voice prostheses including requirements for marking, packaging and information to be provided by the manufacturer as well as test methods for the evaluation of physical characteristics of voice prostheses. NOTE There is guidance or rationale for this list item contained in A.2.
- Standard12 pagesEnglish languagesale 15% off
- Standard13 pagesFrench languagesale 15% off
- Standard13 pagesFrench languagesale 15% off
This document applies to the basic safety and essential performance of a humidifier, also hereafter referred to as ME equipment, in combination with its accessories, the combination also hereafter referred to as ME system. This document is also applicable to those accessories intended by their manufacturer to be connected to a humidifier where the characteristics of those accessories can affect the basic safety or essential performance of the humidifier. EXAMPLE 1 Heated breathing tubes (heated-wire breathing tubes) or ME equipment intended to control these heated breathing tubes (heated breathing tube controllers). NOTE 1 Heated breathing tubes and their controllers are ME equipment and are subject to the requirements of IEC 60601‑1. NOTE 2 ISO 5367 specifies other safety and performance requirements for breathing tubes. This document includes requirements for the different medical uses of humidification, such as invasive ventilation, non-invasive ventilation, nasal high-flow therapy, and obstructive sleep apnoea therapy, as well as humidification therapy for tracheostomy patients. NOTE 3 A humidifier can be integrated into other equipment. When this is the case, the requirements of the other equipment also apply to the humidifier. EXAMPLE 2 Heated humidifier incorporated into a critical care ventilator where ISO 80601‑2-12[10] also applies. EXAMPLE 3 Heated humidifier incorporated into a homecare ventilator for dependent patients where ISO 80601‑2-72[12] also applies. EXAMPLE 4 Heated humidifier incorporated into sleep apnoea therapy equipment where ISO 80601‑2‑70[11] also applies. EXAMPLE 5 Heated humidifier incorporated into ventilatory support equipment where either ISO 80601-2-79[13] or ISO 80601-2-80[14] also apply. EXAMPLE 6 Heated humidifier incorporated into respiratory high-flow therapy equipment where ISO 80601‑2‑90[15] also applies. This document also includes requirements for an active HME (heat and moisture exchanger), ME equipment which actively adds heat and moisture to increase the humidity level of the gas delivered from the HME to the patient. This document is not applicable to a passive HME, which returns a portion of the expired moisture and heat of the patient to the respiratory tract during inspiration without adding heat or moisture. NOTE 4 ISO 9360‑1 and ISO 9360‑2[4] specify safety and performance requirements for a passive HME. NOTE 5 If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601‑1:2005+AMD1:2012+AMD2:2020, 7.2.13 and 8.4.1. NOTE 6 Additional information can be found in IEC 60601‑1:2005+AMD1:2012+AMD2:2020, 4.2. This document does not specify the requirements for cold pass-over or cold bubble-through humidification devices, the requirements for which are given in ISO 20789[6]. This document is not applicable to equipment commonly referred to as “room humidifiers” or humidifiers used in heating, ventilation and air conditioning systems, or humidifiers incorporated into infant incubators. This document is not applicable to nebulizers used for the delivery of a drug to patients. NOTE 7 ISO 27427[7] specifies the safety and performance requirements for nebulizers.
- Standard99 pagesEnglish languagesale 15% off
- Standard108 pagesFrench languagesale 15% off





