FprEN ISO 10993-1
(Main)This document specifies:
— the general principles governing the biological evaluation of medical devices within a risk management process;
— the general categorization of medical devices based on the nature and duration of their contact with the body;
— the evaluation of existing relevant data from all sources;
— the identification of gaps in the available data set on the basis of a risk analysis;
— the identification of additional data sets necessary to analyse the biological safety of the medical device;
— the assessment of the biological safety of the medical device.
This document applies to evaluation of materials and medical devices that are expected to have direct or indirect contact with:
— the patient's body during intended use;
— the user's body, if the medical device is intended for protection (e.g., surgical gloves, masks and others).
This document is applicable to biological evaluation of all types of medical devices including active, non-active, implantable and non-implantable medical devices.
This document also gives guidelines for the assessment of biological hazards arising from:
— risks, such as changes to the medical device over time, as a part of the overall biological safety assessment;
— breakage of a medical device or medical device component which exposes body tissue to new or novel materials.
Other parts of ISO 10993 cover specific aspects of biological assessments and related tests. Device-specific or product standards address mechanical testing.
This document excludes hazards related to bacteria, moulds, yeasts, viruses, transmissible spongiform encephalopathy (TSE) agents and other pathogens.
Biologische Beurteilung von Medizinprodukten - Teil 1: Beurteilung und Prüfungen im Rahmen eines Risikomanagementsystems (ISO/FDIS 10993-1:2018)
Dieser Teil des Dokuments beschreibt:
die allgemeinen Grundsätze, die für die biologische Beurteilung von Medizinprodukten innerhalb eines Risikomanagementverfahrens gelten;
die allgemeine Einteilung von Medizinprodukten, basierend auf der Art und Dauer von deren Kontakt mit dem Körper;
die Auswertung vorhandener relevanter Daten aus allen Quellen;
die Ermittlung von Lücken in den verfügbaren Datensätzen auf der Grundlage einer Risikoanalyse;
die Ermittlung zusätzlicher Datensätze, die für die Analyse der biologischen Sicherheit des Medizinprodukts notwendig sind;
die Beurteilung der biologischen Sicherheit des Medizinprodukts.
Dieses Teildokument gilt für die Beurteilung von Materialien und Produkten, von denen erwartet wird, dass sie direkt oder indirekt während der Verwendung mit dem Körper des Patienten in Kontakt kommen. Außerdem gilt dieses Dokument für Medizinprodukte, von denen erwartet wird, dass sie direkt oder indirekt mit dem Körper des Arztes in Kontakt kommen, wenn das Produkt den Arzt schützen soll (z. B. OP-Handschuhe, Masken und andere Produkte). Dieses Dokument ist anwendbar auf die biologische Beurteilung aller Arten an Medizinprodukten, einschließlich aktiver, nicht aktiver, implantierbarer und nicht implantierbarer Medizinprodukte.
Biologische Gefährdungen aufgrund sonstiger Risiken, wie beispielsweise mechanische Ausfälle oder Veränderungen am Produkt mit der Zeit, sollten als Teil der biologischen Gesamtbeurteilung der Sicherheit betrachtet werden und können durch relevante Biokompatibilitäts-, mechanische oder andere In-vivo-Tierversuche oder andere Informationen behandelt werden. Andere Teile der ISO 10993 decken spezifische Aspekte der biologischen Beurteilungen und zugehörigen Prüfungen ab.
Évaluation biologique des dispositifs médicaux - Partie 1: Évaluation et essais au sein d'un processus de gestion du risque (ISO/FDIS 10993-1:2018)
Le présent document spécifie:
— les principes généraux sur lesquels repose l'évaluation biologique des dispositifs médicaux dans un processus de gestion du risque;
— la classification générale des dispositifs médicaux, fondée sur la nature et la durée de leur contact avec le corps humain;
— l'évaluation de toutes les données pertinentes existantes;
— l'identification de lacunes dans les ensembles de données disponibles sur la base d'une analyse de risque;
— l'identification d'ensembles de données supplémentaires nécessaires à l'analyse de la sécurité biologique du dispositif médical;
— l'évaluation de la sécurité biologique du dispositif médical.
Le présent document s'applique à l'évaluation de matériaux et dispositifs médicaux qui sont destinés à entrer en contact direct ou indirect avec:
— le corps du patient pendant leur utilisation prévue;
— le corps de l'utilisateur si le dispositif médical est destiné à la protection (par exemple gants chirurgicaux, masques et autres).
Le présent document est applicable à l'évaluation biologique de tous types de dispositifs médicaux, y compris les dispositifs médicaux actifs, non actifs, implantables et non implantables.
Le présent document donne également des lignes directrices pour l'évaluation des dangers biologiques provenant de:
— risques tels que des modifications du dispositif médical au fil du temps, dans le cadre de l'évaluation de sécurité biologique générale;
— rupture d'un dispositif médical ou d'un composant de dispositif médical qui expose des tissus de l'organisme à des matériaux nouveaux.
Les autres parties de l'ISO 10993 couvrent des aspects spécifiques des évaluations biologiques et des essais associés. Des normes de produits ou spécifiques aux dispositifs traitent des essais mécaniques.
Le présent document exclut les dangers relatifs aux bactéries, moisissures, levures, virus, agents de l'encéphalopathie spongiforme transmissible (EST) et autres agents pathogènes.
Biološko ovrednotenje medicinskih pripomočkov - 1. del: Ocena in preskušanje znotraj procesa obvladovanja tveganja (ISO/FDIS 10993-1:2018)
General Information
RELATIONS
Standards Content (sample)
SLOVENSKI STANDARD
oSIST prEN ISO 10993-1:2017
01-maj-2017
%LRORãNRRYUHGQRWHQMHPHGLFLQVNLKSULSRPRþNRYGHO2FHQDLQSUHVNXãDQMH
]QRWUDMSURFHVDREYODGRYDQMDWYHJDQMD,62',6
Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk
management process (ISO/DIS 10993-1:2017)Biologische Beurteilung von Medizinprodukten - Teil 1: Beurteilung und Prüfungen im
Rahmen eines Risikomanagementsystems (ISO/DIS 10993-1:2017)Évaluation biologique des dispositifs médicaux - Partie 1: Évaluation et essais au sein
d'un processus de gestion du risque (ISO/DIS 10993-1:2017)Ta slovenski standard je istoveten z: prEN ISO 10993-1
ICS:
11.100.20 %LRORãNRRYUHGQRWHQMH Biological evaluation of
PHGLFLQVNLKSULSRPRþNRY medical devices
oSIST prEN ISO 10993-1:2017 en
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
---------------------- Page: 1 ----------------------oSIST prEN ISO 10993-1:2017
---------------------- Page: 2 ----------------------
oSIST prEN ISO 10993-1:2017
DRAFT INTERNATIONAL STANDARD
ISO/DIS 10993-1
ISO/TC 194 Secretariat: DIN
Voting begins on: Voting terminates on:
2017-02-20 2017-05-14
Biological evaluation of medical devices —
Part 1:
Evaluation and testing within a risk management process
Évaluation biologique des dispositifs médicaux —
Partie 1: Évaluation et essais au sein d’un processus de gestion du risque
ICS: 11.100.20
THIS DOCUMENT IS A DRAFT CIRCULATED
This document is circulated as received from the committee secretariat.
FOR COMMENT AND APPROVAL. IT IS
THEREFORE SUBJECT TO CHANGE AND MAY
NOT BE REFERRED TO AS AN INTERNATIONAL
STANDARD UNTIL PUBLISHED AS SUCH.
IN ADDITION TO THEIR EVALUATION AS
ISO/CEN PARALLEL PROCESSING
BEING ACCEPTABLE FOR INDUSTRIAL,
TECHNOLOGICAL, COMMERCIAL AND
USER PURPOSES, DRAFT INTERNATIONAL
STANDARDS MAY ON OCCASION HAVE TO
BE CONSIDERED IN THE LIGHT OF THEIR
POTENTIAL TO BECOME STANDARDS TO
WHICH REFERENCE MAY BE MADE IN
Reference number
NATIONAL REGULATIONS.
ISO/DIS 10993-1:2017(E)
RECIPIENTS OF THIS DRAFT ARE INVITED
TO SUBMIT, WITH THEIR COMMENTS,
NOTIFICATION OF ANY RELEVANT PATENT
RIGHTS OF WHICH THEY ARE AWARE AND TO
PROVIDE SUPPORTING DOCUMENTATION. ISO 2017
---------------------- Page: 3 ----------------------
oSIST prEN ISO 10993-1:2017
ISO/DIS 10993-1:2017(E)
COPYRIGHT PROTECTED DOCUMENT
© ISO 2017, Published in Switzerland
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized otherwise in any form
or by any means, electronic or mechanical, including photocopying, or posting on the internet or an intranet, without prior
written permission. Permission can be requested from either ISO at the address below or ISO’s member body in the country of
the requester.ISO copyright office
Ch. de Blandonnet 8 • CP 401
CH-1214 Vernier, Geneva, Switzerland
Tel. +41 22 749 01 11
Fax +41 22 749 09 47
copyright@iso.org
www.iso.org
ii © ISO 2017 – All rights reserved
---------------------- Page: 4 ----------------------
oSIST prEN ISO 10993-1:2017
ISO/DIS 10993-1:2017(E)
Contents Page
Foreword ........................................................................................................................................................................................................................................iv
Introduction ..................................................................................................................................................................................................................................v
1 Scope ................................................................................................................................................................................................................................. 1
2 Normative references ...................................................................................................................................................................................... 1
3 Terms and definitions ..................................................................................................................................................................................... 2
4 General principles applying to biological evaluation of medical devices .................................................5
5 Categorization of medical devices ...................................................................................................................................................... 9
5.1 General ........................................................................................................................................................................................................... 9
5.2 Categorization by nature of body contact ........................................................................................................................ 9
5.2.1 Non-contacting devices .............................................................................................................................................. 9
5.2.2 Transient-contacting devices .............................................................................................................................10
5.2.3 Surface-contacting devices ..................................................................................................................................10
5.2.4 External communicating devices ....................................................................................................................10
5.2.5 Implant devices ..............................................................................................................................................................11
5.3 Categorization by duration of contact .............................................................................................................................11
6 Biocompatibility evaluation process ............................................................................................................................................12
6.1 Biological risk assessment .........................................................................................................................................................12
6.1.1 Physical and chemical information ...............................................................................................................12
6.1.2 General: Gap analysis and selection of biological endpoints for assessment ...........12
6.1.3 Biological testing...........................................................................................................................................................14
7 Interpretation of biological evaluation data and overall biological safety assessment ..........18
Annex A (informative) Endpoints to be addressed in a biological risk assessment.........................................20
Annex B (informative) Guidance on the conduct of biological evaluation within a risk
management process ....................................................................................................................................................................................25
Annex C (informative) Suggested procedure for literature review ...................................................................................38
Annex ZA (informative) Relationship between this European Standard and the essential
requirements of Directive 93/42/EEC [OJ L 169] aimed to be covered ....................................................40
Annex ZB (informative) Relationship between this European Standard and the essential
requirements of Directive 90/385/EEC [OJ L 189] aimed to be covered ................................................42
Bibliography .............................................................................................................................................................................................................................44
© ISO 2017 – All rights reserved iii---------------------- Page: 5 ----------------------
oSIST prEN ISO 10993-1:2017
ISO/DIS 10993-1:2017(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www .iso .org/ directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of
any patent rights identified during the development of the document will be in the Introduction and/or
on the ISO list of patent declarations received (see www .iso .org/ patents).Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.For an explanation on the meaning of ISO specific terms and expressions related to conformity assessment,
as well as information about ISO’s adherence to the World Trade Organization (WTO) principles in the
Technical Barriers to Trade (TBT) see the following URL: www . i so .org/ iso/ foreword .html.
The committee responsible for this document is ISO/TC 194, Biological and clinical evaluation of medical
devices.This fifth edition cancels and replaces the fourth edition (ISO 10993-1:2009), which has been technically
revised.It also incorporates the Corrigendum ISO 10993-1:2009/Cor.1.
The following technical were changes:
a) revised Annex A “Endpoints to be addressed in a biological risk assessment”;
b) replaced Annex B “Guidance on the risk management process” with “Guidance on the conduct of
biological evaluation within a risk management process” (formally TR 15499);c) editing changes in the genotoxicity section and other sections of the normative document;
d) additional definitions for terms used throughout the 10993 series of standards added;
e) additional “Non-contacting devices” and “Transient-contacting devices” categories.
A list of all parts in the ISO 10993- series can be found on the ISO website.iv © ISO 2017 – All rights reserved
---------------------- Page: 6 ----------------------
oSIST prEN ISO 10993-1:2017
ISO/DIS 10993-1:2017(E)
Introduction
The primary aim of this document is the protection of humans from potential biological risks arising
from the use of medical devices. It is compiled from numerous International and national standards
and guidelines concerning the biological evaluation of medical devices. It is intended to describe the
biological evaluation of medical devices within a risk management process, as part of the overall
evaluation and development of each device. This approach combines the review and evaluation of
existing data from all sources with, where necessary, the selection and application of additional tests,
thus enabling a full evaluation to be made of the biological responses to each medical device, relevant
to its safety in use. It must be appreciated that the term “medical device” is wide-ranging and, at one
extreme, consists of a single material, which can exist in more than one physical form, and at the other
extreme, of a medical device consisting of numerous components made of more than one material.
This document addresses the determination of the biological response to medical devices, mostly in a
general way, rather than in a specific device-type situation. Thus, for a complete biological evaluation, it
classifies medical devices according to the nature and duration of their anticipated contact with human
tissues when in use and indicates, in a matrix the biological endpoints that are thought to be relevant in
the consideration of each device category.NOTE Products which might be considered to be medical devices in some jurisdictions but for which there is
not yet a harmonized approach, are:1) aids for disabled/handicapped people;
2) devices for the treatment/diagnosis of diseases and injuries in animals;
3) accessories for medical devices;
4) disinfection substances;
5) devices incorporating animal and human tissues, which might meet the requirements of the above definition
but are subject to different controls.The range of biological hazards is wide and complex. The biological response to a constituent material
alone cannot be considered in isolation from the overall device design. Thus, in designing a device, the
choice of the best material with respect to its biocompatibility might result in a less functional device,
biocompatibility being only one of a number of characteristics to be considered in making that choice.
Where a material is intended to interact with tissue in order to perform its function, the biological
evaluation needs to address this.Biological responses that are regarded as adverse, caused by a material in one application, might not be
regarded as such in a different situation. Biological testing is based upon, among other things, in vitro
and ex vivo test methods and upon animal models, so that the anticipated behaviour when a device is
used in humans can be judged only with caution, as it cannot be unequivocally concluded that the same
biological response will also occur in this species. In addition, differences in the manner of response to
the same material among individuals indicate that some patients can have adverse reactions, even to
well-established materials.The role of this document is to serve as a framework in which to plan a biological evaluation which, as
scientific knowledge advances our understanding of the basic mechanisms of host responses, minimizes
the number and exposure of test animals by giving preference to chemical, physical, morphological,
and topographical characterization testing and in vitro models, in situations where these methods yield
equally relevant information to that obtained from in vivo models.It is not intended that this document provide a rigid set of test methods, including pass/fail criteria,
as this might result in either an unnecessary constraint on the development and use of novel medical
devices, or a false sense of security in the general use of medical devices. Where a particular application
warrants it, experts in the product or in the area of application concerned can choose to establish
specific tests and criteria, described in a product-specific vertical standard.© ISO 2017 – All rights reserved v
---------------------- Page: 7 ----------------------
oSIST prEN ISO 10993-1:2017
ISO/DIS 10993-1:2017(E)
The ISO 10993- series of standards are intended for use by professionals, appropriately qualified
by training and experience, who are able to interpret its requirements and judge the outcome of the
evaluation for each, taking into consideration all the factors relevant to the device, its intended use
and the current knowledge of the medical device provided by review of the scientific literature and
previous clinical experience.Informative Annex A contains a table that is generally helpful in identifying endpoints recommended
in the biocompatibility evaluation of medical devices, according to their category of body contact and
duration of clinical exposure. Informative Annex B contains guidance for the application of the risk
management process to medical devices which encompasses biological evaluation.vi © ISO 2017 – All rights reserved
---------------------- Page: 8 ----------------------
oSIST prEN ISO 10993-1:2017
DRAFT INTERNATIONAL STANDARD ISO/DIS 10993-1:2017(E)
Biological evaluation of medical devices —
Part 1:
Evaluation and testing within a risk management process
1 Scope
This part document describes:
— the general principles governing the biological evaluation of medical devices within a risk
management process;— the general categorization of devices based on the nature and duration of their contact with the body;
— the evaluation of existing relevant data from all sources;— the identification of gaps in the available data set on the basis of a risk analysis;
— the identification of additional data sets necessary to analyse the biological safety of the medical
device;— the assessment of the biological safety of the medical device.
This part document applies to evaluation of materials and devices that are expected to have direct
or indirect contact with the patient’s body during intended use. In addition, this document applies to
medical devices that are expected to have direct or indirect contact with the clinician’s body, if the
device is intended to protect the clinician (e.g., surgical gloves, masks and others). This document
is applicable to biological evaluation of all types of medical devices including active, non-active,
implantable and non-implantable medical devices.Biological hazards arising from other risks, such as mechanical failures or changes to the device over
time, should be considered as a part of the overall biological safety assessment, and can be addressed
by relevant biocompatibility, mechanical or other in vivo animal testing or other information. Other
parts of ISO 10993 cover specific aspects of biological assessments and related tests.
2 Normative referencesThe following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments) applies.
ISO 10993-2, Biological evaluation of medical devices — Part 2: Animal welfare requirements
ISO 10993-3, Biological evaluation of medical devices — Part 3: Tests for genotoxicity, carcinogenicity and
reproductive toxicityISO 10993-4, Biological evaluation of medical devices — Part 4: Selection of tests for interaction with blood
ISO 10993-5, Biological evaluation of medical devices — Part 5: Tests for in vitro cytotoxicity
ISO 10993-6, Biological evaluation of medical devices — Part 6: Tests for local effects after implantation
ISO 10993-7, Biological evaluation of medical devices — Part 7: Ethylene oxide sterilization residuals
ISO 10993-9, Biological evaluation of medical devices — Part 9: Framework for identification and
quantification of potential degradation products© ISO 2017 – All rights reserved 1
---------------------- Page: 9 ----------------------
oSIST prEN ISO 10993-1:2017
ISO/DIS 10993-1:2017(E)
ISO 10993-10, Biological evaluation of medical devices — Part 10: Tests for irritation and skin sensitization
ISO 10993-11, Biological evaluation of medical devices — Part 11: Tests for systemic toxicity
ISO 10993-12, Biological evaluation of medical devices — Part 12: Sample preparation and reference
materialsISO 10993-13, Biological evaluation of medical devices — Part 13: Identification and quantification of
degradation products from polymeric medical devicesISO 10993-14, Biological evaluation of medical devices — Part 14: Identification and quantification of
degradation products from ceramicsISO 10993-15, Biological evaluation of medical devices — Part 15: Identification and quantification of
degradation products from metals and alloysISO 10993-16, Biological evaluation of medical devices — Part 16: Toxicokinetic study design for
degradation products and leachablesISO 10993-17, Biological evaluation of medical devices — Part 17: Establishment of allowable limits for
leachable substancesISO 10993-18, Biological evaluation of medical devices — Part 18: Chemical characterization of materials
ISO/TS 10993-19, Biological evaluation of medical devices — Part 19: Physico-chemical, morphological and
topographical characterization of materialsISO/TS 10993-20, Biological evaluation of medical devices — Part 20: Principles and methods for
immunotoxicology testing of medical devicesISO/TR 10993-22, Biological evaluation of medical devices — Part 22: Guidance on nanomaterials
ISO 14971, Medical devices — Application of risk management to medical devices3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
ISO and IEC maintain terminological databases for use in standardization at the following addresses:
— IEC Electropedia: available at http:// www .electropedia .org/— ISO Online browsing platform: available at http:// www .iso .org/ obp
3.1
biocompatibility
ability of a medical device or material to perform with an appropriate host response in a specific
applicationNote 1 to entry: This can be demonstrated by biological testing, as well as assessment of effects of leachable
chemicals and/or morphological properties (e.g., bound chemicals, topological features) of the medical device
or materials, and device performance (e.g., maintenance of mechanical integrity) potentially impacting the
biological response.3.2
biological risk
probability of harm to health occurring as a result of medical device or material interactions
3.3biological safety
freedom from unacceptable risk in the context of the intended use
2 © ISO 2017 – All rights reserved
---------------------- Page: 10 ----------------------
oSIST prEN ISO 10993-1:2017
ISO/DIS 10993-1:2017(E)
3.4
chemical constituent
any synthetic or natural substance that is used in a process for manufacturing materials and/or medical
devices, including the base material(s), additives (antioxidants, UV stabilizers, color additives, dyes,
etc.), and processing aids (solvents, lubricants, antifoaming agents, etc.)3.5
data set
information, such as physical and/or chemical characterization, toxicity data, etc. from a variety of
sources necessary to characterize the biological response of a device3.6
direct contact
term used for a device or device component that comes into physical contact with body tissue
3.7externally communicating
term used for a device or device component that is placed within the body during a medical procedure,
with a portion of the device residing within the body, and another portion of the device outside of the body
3.8final product
medical device or device component that includes all manufacturing processes for the “to be marketed”
device including packaging and sterilization, if applicable3.9
implant
term used for a device or device component that that is placed entirely within the body during a medical
procedure3.10
indirect contact
term used for a device or device component through which a fluid or gas passes, prior to the fluid or gas
coming into physical contact with body tissue (in this case the device or device component itself does
not physically contact body tissue)3.11
material
any synthetic or natural polymer, metal, alloy, ceramic or other non-viable substance, including tissue
rendered non-viable, used as a medical device or any part thereof3.12
material characterization
the broad and general process by which a material’s chemistry, structure and properties are evaluated
and measured, to include assessment of compositional, structural, and mechanical properties, if
appropriate3.13
medical device
any instrument, apparatus, implement, machine, appliance, implant, in vitro reagent or calibrator,
software, material or other similar or related article, intended by the manufacturer to be used, alone or
in combination, for human beings for one or more of the specific purpose(s) of:— diagnosis, prevention, monitoring, treatment or alleviation of disease,
— diagnosis, monitoring, treatment, alleviation of or compensation for an injury,
— investigation, replacement, modification, or support of the anatomy or of a physiological process,
— supporting or sustaining life,© ISO 2017 – All rights reserved 3
---------------------- Page: 11 ----------------------
oSIST prEN ISO 10993-1:2017
ISO/DIS 10993-1:2017(E)
— control of conception,
— disinfection of medical devices,
— providing information for medical purposes by means of in vitro examination of specimens derived
from the human body,and which does not achieve its primary intended action in or on the human body by pharmacological,
immunological or metabolic means, but which may be assisted in its function by such means
Note 1 to entry: This definition has been developed by the Global Harmonization Task Force (GHTF).
Accessories intended specifically by manufacturers to be used together with a “parent” medical device
to enable that medical device to achieve its intended purpose, should be assessed according to the
ISO 10993- series of standards.[SOURCE: ISO 14971:2007, definition 2.9, modified to refer to ISO 10993]
Medical devices are different from drugs/biologics, and their biological evaluation requires a different
approach.Medical devices can include dental devices.
3.14
non-contact
term used to indicate that the device or device component has neither direct nor indirect contact with
body tissues3.15
physical and chemical information
knowledge regarding formulation, manufacturing processes, geometric and physical properties and
type of body contact and clinical use that is used to determine whether any additional biological or
material characterization testing is needed3.16
risk analysis
systematic use of available information to identify hazards and to estimate the risk
[SOURCE: ISO 14971:2007, definition 2.17]3.17
risk assessment
overall process comprising a risk analysis and a risk evaluation
[SOURCE: ISO 14971:2007, definition 2.18]
3.18
risk evaluation
process of comparing the estimated risk against given risk criteria to determine the acceptability of
the risk[SOURCE: ISO 14971:2007, definition 2.21]
3.19
risk management
systematic application of management policies, procedures and practices to the tasks of analysing,
evaluating, controlling and monitoring risk[SOURCE: ISO 14971:2007, definition 2.22]
3.20
toxic
capable of causing an adverse biological response
4 © ISO 2017 – All rights reserved
---------------------- Page: 12 ----------------------
oSIST prEN ISO 10993-1:2017
ISO/DIS 10993-1:2017(E)
3.21
toxicological hazard
potential for a chemical substance or material to cause an adverse biological reaction, taking into
account the nature of the reaction and the dose required to elicit it3.22
toxicological risk
probability of a specified degree of an adverse reaction occurring in response to a specified level of
exposure3.23
toxicological threshold
a limit, such as a tolerable intake (TI), tolerable exposure (TE), allowable limit (AL) value, or Threshold
of Toxicological Concern (TTC) below which adverse effects are not expected for relevant biological
endpoints3.24
transient contact
term used for a device or device component that has a very brief duration of contact with body tissue
(e.g., for less than one minute)4 General principles applying to biological evaluation of medical devices
4.1 The biological evaluation of any material or medical device intended for use in humans shall
form part of a structured biological evaluation programme within a risk management process in
accordance with ISO 14971, as given in Figure 1. This risk management process involves identification of
biological hazards, estimation of the associated biological risks, and determination of their acceptability.
Annex B provides guidance on this process. The biological evaluation shall be planned, carried out, and
documented by knowledgeable and experienced professionals.The risk management plan should identify aspects of the biological evaluation requiring specific
technical competencies and shall identify the person(s) responsible for the biological evaluation.
The evaluation programme shall include documented, informed consideration ofadvantages/disadvantages and relevance of:
a) device configuration and a listing of a device’s materials of construction (qualitative) and where
necessary, the proportion and amount (mass) of each material in the device (quantitative);
b) the physical and chemical characteristics of the various candidate materials of construction and
their composition;NOTE Where this information is already documented within the risk management for the device it can
be included by reference.c) any history of clinical use or human exposure data;
NOTE Previous regulatory approval history can be relevant.
d) any existing toxicology and other biological safety data on product and component materials,
breakdown products and metabolites;e) test procedures.
Evaluation can include both a study of relevant preclinical and clinical experience and actual testing.
Such an evaluation might result in the conclusion that no testing is needed if the material has a
demonstrable safe history of use in a specified role and physical form that is equivalent to that of the
device under design. The type of information that can be useful to demonstrate equivalence is included
in Annex B. Testing is usually not necessary when sufficient information is already available to perform
a ris...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.