This document specifies a method of testing the laser-induced ignition and damage of medical beam delivery systems to allow checking of suitable products according to the classification system.
NOTE 1 Take care when interpreting these results, since the direct applicability of the results of this test method to the clinical situation has not been fully established.
NOTE 2 Users of products tested by this method are cautioned that the laser will be wavelength sensitive and tested at the wavelength for which it is intended to be used. If tested using other wavelengths, the power settings and modes of beam delivery need to be explicitly stated.
CAUTION — This test method can involve hazardous materials, operations and equipment. This document provides advice on minimizing some of the risks associated with its use but does not purport to address all such risks. It is the responsibility of the user of this document to establish appropriate safety and health practices and to determine the applicability of regulatory limitations prior to use.

  • Standard
    26 pages
    English language
    e-Library read for
    1 day

1.1       This document specifies requirements for the development, validation and routine control of a radiation sterilization process for medical devices.
NOTE         Although the scope is limited to medical devices, this document can be applicable to other products and equipment.
This document covers radiation processes employing irradiators using:
a)      the radionuclide 60Co or 137Cs;
b)      a beam from an electron generator; or
c)       a beam from an X-ray generator.
1.2       This document is not applicable to processes for inactivating viruses or the causative agents of spongiform encephalopathies, such as scrapie, bovine spongiform encephalopathy and Creutzfeldt-Jakob disease.
NOTE         For information on such processes, see ISO 22442-1, ISO 22442-2, ISO 22442-3, ISO 13022 and ICH Q5A.
1.2.1       This document does not specify requirements for designating a medical device as sterile.
NOTE         Regional and national requirements can designate medical devices as sterile. See, for example, EN 556-1 or ANSI/AAMI ST67.
1.2.2       This document does not specify a quality management system for the control of all stages of production of medical devices.
NOTE         It is not a requirement of this document to have a complete quality management system during manufacture, but the elements of a quality management system that are the minimum necessary to control the sterilization process are normatively referenced at appropriate places in the text (see, in particular, Clause 4). Attention is drawn to the standards for quality management systems (see ISO 13485) that control all stages of production of medical devices, including the sterilization process. Regional and national regulations for the provision of medical devices can require implementation of a complete quality management system and the assessment of that system by a third party.
1.2.3       This document does not require that biological indicators be used for validation or monitoring of radiation sterilization, nor does it require that a pharmacopoeial test for sterility be carried out for product release.
1.2.4       This document does not specify requirements for occupational safety associated with the design and operation of irradiation facilities.
NOTE         Regulations on safety requirements for occupational safety related to radiation can exist in some countries.
1.2.5       This document does not specify requirements for the sterilization of used or reprocessed devices.

  • Standard
    61 pages
    English language
    e-Library read for
    1 day

This document specifies general interface requirements for small-bore connectors that form part of a medical device or accessory that conveys liquids or gases to a patient.
This document also identifies the applications for which these small-bore connectors are intended to be used, which include, but are not limited to:
—     respiratory;
—     enteral;
—     limb cuff inflation;
—     neural;
—     intravascular or hypodermic;
—     other use cases utilizing an ISO 80369-7 small-bore connector.
This document provides the methodology to assess non-interconnectable characteristics of small-bore connectors based on their inherent design in order to reduce the risk of misconnections between medical devices or between accessories for different applications as specified in this document as well as those that can be developed under future parts of the ISO and IEC 80369 series.
NOTE            Clause A.2 contains guidance or rationale for this Clause.

  • Standard
    55 pages
    English language
    e-Library read for
    1 day

This document specifies the requirements and methods for the clinical investigation of medical electrical (ME) equipment used to measure the body temperature in indirect measurement mode.
This document covers both intermittently and continuously measuring clinical thermometers.
NOTE 1        This document does not apply to clinical thermometers measuring the body temperature in direct measurement mode.
NOTE 2        For clinical thermometers in direct measurement mode determining the technical accuracy in accordance with ISO 80601-2-56:—1) is considered sufficient.
This document is applicable to clinical thermometers with claimed measurement time shorter than 60 seconds (for methods such as oral or rectal measurement), or shorter than 5 minutes (for methods such as axillary measurement), and which are treated as predictive type thermometers and fall under the scope of this document.
This document specifies additional disclosure requirements.
This document does not apply to the clinical investigation of a screening thermographs for human febrile temperature screening whose laboratory accuracy requirements are described in IEC 80601-2-59.
This document does not apply to pulmonary artery catheter for the determination of cardiac output by thermodilution.
NOTE 3        ISO 80601-2-56:—1) does include pulmonary artery catheter for the determination of cardiac output by thermodilution.

  • Standard
    33 pages
    English language
    e-Library read for
    1 day

1.1       This document specifies requirements for the development, validation and routine control of a radiation sterilization process for medical devices.
NOTE         Although the scope is limited to medical devices, this document can be applicable to other products and equipment.
This document covers radiation processes employing irradiators using:
a)      the radionuclide 60Co or 137Cs;
b)      a beam from an electron generator; or
c)       a beam from an X-ray generator.
1.2       This document is not applicable to processes for inactivating viruses or the causative agents of spongiform encephalopathies, such as scrapie, bovine spongiform encephalopathy and Creutzfeldt-Jakob disease.
NOTE         For information on such processes, see ISO 22442-1, ISO 22442-2, ISO 22442-3, ISO 13022 and ICH Q5A.
1.2.1       This document does not specify requirements for designating a medical device as sterile.
NOTE         Regional and national requirements can designate medical devices as sterile. See, for example, EN 556-1 or ANSI/AAMI ST67.
1.2.2       This document does not specify a quality management system for the control of all stages of production of medical devices.
NOTE         It is not a requirement of this document to have a complete quality management system during manufacture, but the elements of a quality management system that are the minimum necessary to control the sterilization process are normatively referenced at appropriate places in the text (see, in particular, Clause 4). Attention is drawn to the standards for quality management systems (see ISO 13485) that control all stages of production of medical devices, including the sterilization process. Regional and national regulations for the provision of medical devices can require implementation of a complete quality management system and the assessment of that system by a third party.
1.2.3       This document does not require that biological indicators be used for validation or monitoring of radiation sterilization, nor does it require that a pharmacopoeial test for sterility be carried out for product release.
1.2.4       This document does not specify requirements for occupational safety associated with the design and operation of irradiation facilities.
NOTE         Regulations on safety requirements for occupational safety related to radiation can exist in some countries.
1.2.5       This document does not specify requirements for the sterilization of used or reprocessed devices.

  • Standard
    61 pages
    English language
    e-Library read for
    1 day

This document specifies a method of testing the laser-induced ignition and damage of medical beam delivery systems to allow checking of suitable products according to the classification system.
NOTE 1 Take care when interpreting these results, since the direct applicability of the results of this test method to the clinical situation has not been fully established.
NOTE 2 Users of products tested by this method are cautioned that the laser will be wavelength sensitive and tested at the wavelength for which it is intended to be used. If tested using other wavelengths, the power settings and modes of beam delivery need to be explicitly stated.
CAUTION — This test method can involve hazardous materials, operations and equipment. This document provides advice on minimizing some of the risks associated with its use but does not purport to address all such risks. It is the responsibility of the user of this document to establish appropriate safety and health practices and to determine the applicability of regulatory limitations prior to use.

  • Standard
    26 pages
    English language
    e-Library read for
    1 day

This document specifies allowable limits (AL) for residual ethylene oxide (EO) and ethylene chlorohydrin (ECH) in EO-sterilized medical devices, procedures for the measurement of EO and ECH, and methods for determining conformity so that devices can be released. Additional background, including guidance and a flowchart showing how this document is applied, are also included in Annexes A, B, C, D, E, F, G, H, I, J and K.
EO-sterilized devices or components that have neither direct nor indirect body or user contact (e.g. in vitro diagnostic devices) are out of scope of this document. This document does not apply to devices that have been demonstrated to not absorb or retain EO or its degradation product ECH, such as medical devices made exclusively of metal alloys and glass, see Clause C.5[228].
NOTE            This document does not specify limits for ethylene glycol (EG). No device limits are specified for EG because the risk assessment in Annex F indicates that calculated allowable levels are higher than those likely to occur in a medical device.

  • Standard
    135 pages
    English language
    e-Library read for
    1 day

This document specifies requirements for implantation test methods for preclinical assessment of the local effects after implantation of medical devices or materials intended for use in medical devices. This document is applicable to the evaluation of local tissue responses from medical devices that are intended to be used where skin or mucosal tissue is breached, when required.
This document is applicable to medical device or materials that require implantation evaluation and can be solid or non-solid (such as porous materials, liquids, gels, pastes, powders, and particulates), absorbable, degradable, non- absorbable, or can be tissue-engineered medical products (TEMPs).
These implantation tests are not intended to evaluate or determine the performance of the test sample in terms of mechanical loading or functional performance. This document also does not provide guidance on methods and study design to satisfy requirements for systemic toxicity, carcinogenicity, teratogenicity or mutagenicity. However, the study designs can be modified to also assess other biological effects.

  • Standard
    53 pages
    English language
    e-Library read for
    1 day

This document specifies general interface requirements for small-bore connectors that form part of a medical device or accessory that conveys liquids or gases to a patient.
This document also identifies the applications for which these small-bore connectors are intended to be used, which include, but are not limited to:
—     respiratory;
—     enteral;
—     limb cuff inflation;
—     neural;
—     intravascular or hypodermic;
—     other use cases utilizing an ISO 80369-7 small-bore connector.
This document provides the methodology to assess non-interconnectable characteristics of small-bore connectors based on their inherent design in order to reduce the risk of misconnections between medical devices or between accessories for different applications as specified in this document as well as those that can be developed under future parts of the ISO and IEC 80369 series.
NOTE            Clause A.2 contains guidance or rationale for this Clause.

  • Standard
    55 pages
    English language
    e-Library read for
    1 day

This document specifies the requirements and methods for the clinical investigation of medical electrical (ME) equipment used to measure the body temperature in indirect measurement mode.
This document covers both intermittently and continuously measuring clinical thermometers.
NOTE 1        This document does not apply to clinical thermometers measuring the body temperature in direct measurement mode.
NOTE 2        For clinical thermometers in direct measurement mode determining the technical accuracy in accordance with ISO 80601-2-56:—1) is considered sufficient.
This document is applicable to clinical thermometers with claimed measurement time shorter than 60 seconds (for methods such as oral or rectal measurement), or shorter than 5 minutes (for methods such as axillary measurement), and which are treated as predictive type thermometers and fall under the scope of this document.
This document specifies additional disclosure requirements.
This document does not apply to the clinical investigation of a screening thermographs for human febrile temperature screening whose laboratory accuracy requirements are described in IEC 80601-2-59.
This document does not apply to pulmonary artery catheter for the determination of cardiac output by thermodilution.
NOTE 3        ISO 80601-2-56:—1) does include pulmonary artery catheter for the determination of cardiac output by thermodilution.

  • Standard
    33 pages
    English language
    e-Library read for
    1 day

This document specifies requirements for implantation test methods for preclinical assessment of the local effects after implantation of medical devices or materials intended for use in medical devices. This document is applicable to the evaluation of local tissue responses from medical devices that are intended to be used where skin or mucosal tissue is breached, when required.
This document is applicable to medical device or materials that require implantation evaluation and can be solid or non-solid (such as porous materials, liquids, gels, pastes, powders, and particulates), absorbable, degradable, non- absorbable, or can be tissue-engineered medical products (TEMPs).
These implantation tests are not intended to evaluate or determine the performance of the test sample in terms of mechanical loading or functional performance. This document also does not provide guidance on methods and study design to satisfy requirements for systemic toxicity, carcinogenicity, teratogenicity or mutagenicity. However, the study designs can be modified to also assess other biological effects.

  • Standard
    53 pages
    English language
    e-Library read for
    1 day

This document specifies allowable limits (AL) for residual ethylene oxide (EO) and ethylene chlorohydrin (ECH) in EO-sterilized medical devices, procedures for the measurement of EO and ECH, and methods for determining conformity so that devices can be released. Additional background, including guidance and a flowchart showing how this document is applied, are also included in Annexes A, B, C, D, E, F, G, H, I, J and K.
EO-sterilized devices or components that have neither direct nor indirect body or user contact (e.g. in vitro diagnostic devices) are out of scope of this document. This document does not apply to devices that have been demonstrated to not absorb or retain EO or its degradation product ECH, such as medical devices made exclusively of metal alloys and glass, see Clause C.5[228].
NOTE            This document does not specify limits for ethylene glycol (EG). No device limits are specified for EG because the risk assessment in Annex F indicates that calculated allowable levels are higher than those likely to occur in a medical device.

  • Standard
    135 pages
    English language
    e-Library read for
    1 day

This document specifies good clinical practice (GCP) for the design, conduct, recording and reporting of clinical investigations carried out in human subjects to assess the clinical performance or effectiveness and safety of medical devices.
For post-market clinical investigations, the principles set forth in this document are intended to be followed as far as relevant, considering the nature of the clinical investigation (see Annex I).
This document specifies the general requirements intended to
protect the rights, safety and well-being of human subjects, users or other persons,
ensure the scientific conduct of the clinical investigation and the credibility of the clinical investigation results,
define the responsibilities of the sponsor and principal investigator, and
assist sponsors, investigators, ethics committees, regulatory authorities and other bodies involved in the conformity assessment of medical devices.
Other standards or national requirements can also apply to the investigational device(s) under consideration or the clinical investigation(s).
NOTE            For Software as a Medical Device (SaMD), where appropriate, demonstration of the analytical validity (the SaMD’s output is accurate for a given input), the scientific validity (the SaMD’s output is associated to the intended clinical condition/physiological state), and clinical performance (the SaMD’s output yields a clinically meaningful association to the target use) of the SaMD, the requirements of this document apply as far as relevant (see Reference [5]). Justifications for exemptions from this document can consider the uniqueness of indirect contact between subjects and the SaMD.
This document does not apply to in vitro diagnostic medical devices. However, there can be situations, dependent on the device and national or regional requirements, where users of this document can consider whether either specific sections or requirements of this document, or both, can be applicable.

  • Standard
    106 pages
    English language
    e-Library read for
    1 day

This document specifies requirements and test methods for male condoms made from natural rubber latex.
This document does not specify requirements related to any medicinal substances applied to or delivered by the condom.
NOTE            The safety and effectiveness of any medicinal substance are assessed according to national and regional regulations.

  • Standard
    93 pages
    English language
    e-Library read for
    1 day

NOTE 1        There is guidance or rationale for this Clause in A.2.1.
This document specifies the requirements for information supplied by the manufacturer for a medical device or an accessory, as defined in 3.1. This document includes the generally applicable requirements for identification and labels on a medical device or accessory, the packaging, marking of a medical device or accessory, and accompanying information. This document does not specify the means by which the information is to be supplied.
NOTE 2        Some authorities having jurisdiction impose different requirements for the identification, marking and documentation of a medical device or accessory.
Specific requirements of medical device product standards or group standards take precedence over requirements of this document.

  • Standard
    60 pages
    English language
    e-Library read for
    1 day

This document specifies good clinical practice (GCP) for the design, conduct, recording and reporting of clinical investigations carried out in human subjects to assess the clinical performance or effectiveness and safety of medical devices.
For post-market clinical investigations, the principles set forth in this document are intended to be followed as far as relevant, considering the nature of the clinical investigation (see Annex I).
This document specifies the general requirements intended to
protect the rights, safety and well-being of human subjects, users or other persons,
ensure the scientific conduct of the clinical investigation and the credibility of the clinical investigation results,
define the responsibilities of the sponsor and principal investigator, and
assist sponsors, investigators, ethics committees, regulatory authorities and other bodies involved in the conformity assessment of medical devices.
Other standards or national requirements can also apply to the investigational device(s) under consideration or the clinical investigation(s).
NOTE            For Software as a Medical Device (SaMD), where appropriate, demonstration of the analytical validity (the SaMD’s output is accurate for a given input), the scientific validity (the SaMD’s output is associated to the intended clinical condition/physiological state), and clinical performance (the SaMD’s output yields a clinically meaningful association to the target use) of the SaMD, the requirements of this document apply as far as relevant (see Reference [5]). Justifications for exemptions from this document can consider the uniqueness of indirect contact between subjects and the SaMD.
This document does not apply to in vitro diagnostic medical devices. However, there can be situations, dependent on the device and national or regional requirements, where users of this document can consider whether either specific sections or requirements of this document, or both, can be applicable.

  • Standard
    106 pages
    English language
    e-Library read for
    1 day

This document specifies requirements and test methods for male condoms made from natural rubber latex.
This document does not specify requirements related to any medicinal substances applied to or delivered by the condom.
NOTE            The safety and effectiveness of any medicinal substance are assessed according to national and regional regulations.

  • Standard
    93 pages
    English language
    e-Library read for
    1 day

This document specifies particular requirements for washer-disinfectors (WD) intended for use when the level of assurance of disinfection that is necessary can be achieved by cleaning and thermal disinfection (A0 not less than 60) and does not require an independent automated record of critical processes to be kept. It is intended to be used in conjunction with ISO 15883-1, which specifies general requirements for WD.
The range of products on which WD of this particular type can be used is restricted to non-invasive and non-critical devices and equipment (i.e. not penetrating skin or contacting mucosal surfaces).
NOTE            Thermal disinfection can be achieved by rinsing the load with hot water, exposure to steam, or combination of the two.
This document does not cover powered devices, lumened devices, and other semi-critical and critical medical devices.
Devices identified within the scopes of ISO 15883-2, ISO 15883-3, ISO 15883-4, and ISO 15883-7 do not fall within the scope of this document.

  • Standard
    20 pages
    English language
    e-Library read for
    1 day

NOTE 1        There is guidance or rationale for this Clause in A.2.1.
This document specifies the requirements for information supplied by the manufacturer for a medical device or an accessory, as defined in 3.1. This document includes the generally applicable requirements for identification and labels on a medical device or accessory, the packaging, marking of a medical device or accessory, and accompanying information. This document does not specify the means by which the information is to be supplied.
NOTE 2        Some authorities having jurisdiction impose different requirements for the identification, marking and documentation of a medical device or accessory.
Specific requirements of medical device product standards or group standards take precedence over requirements of this document.

  • Standard
    60 pages
    English language
    e-Library read for
    1 day

This document specifies particular requirements for washer-disinfectors (WD) intended for use when the level of assurance of disinfection that is necessary can be achieved by cleaning and thermal disinfection (A0 not less than 60) and does not require an independent automated record of critical processes to be kept. It is intended to be used in conjunction with ISO 15883-1, which specifies general requirements for WD.
The range of products on which WD of this particular type can be used is restricted to non-invasive and non-critical devices and equipment (i.e. not penetrating skin or contacting mucosal surfaces).
NOTE            Thermal disinfection can be achieved by rinsing the load with hot water, exposure to steam, or combination of the two.
This document does not cover powered devices, lumened devices, and other semi-critical and critical medical devices.
Devices identified within the scopes of ISO 15883-2, ISO 15883-3, ISO 15883-4, and ISO 15883-7 do not fall within the scope of this document.

  • Standard
    20 pages
    English language
    e-Library read for
    1 day

IEC 62083:2025, with the inclusion of type tests and site tests, applies to the design, manufacture, installation, and maintenance of the radiotherapy treatment planning system.
This document applies to the communication of the radiotherapy treatment planning system with other devices
– used in medical practice,
– that imports data either through input by the operator or from other devices,
– that outputs data to other devices, and
– that is intended to be
- for normal use, under the authority of appropriately qualified persons, by operators having the required skills and training,
- used and maintained in accordance with the recommendations given in the instructions for use, and
– used within the environmental conditions specified in the technical description.
This document applies to any software application that is used for the development, evaluation, or approval of a treatment plan, whether stand-alone or part of another system.
IEC 62083:2025 cancels and replaces the second edition published in 2009. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
– modification of the title from Medical electrical system - Requirements for the safety of radiotherapy treatment planning systems, to Medical device software - Requirements for the safety of radiotherapy treatment planning systems;
– Adaptive radiotherapy is added with Clause 16;
– The title reflects different implementations of radiotherapy treatment planning systems.

  • Standard
    62 pages
    English language
    e-Library read for
    1 day

IEC 60601-2-57:2023 applies to basic safety and essential performance of equipment incorporating one or more sources of optical radiation in the wavelength range 200 nm to 3 000 nm, with the exception of laser radiation, and intended to create photobiological effects in humans for therapeutic, diagnostic, monitoring, and cosmetic or aesthetic applications; hereafter referred to as light source equipment (ls equipment).

  • Standard
    31 pages
    English language
    e-Library read for
    1 day

IEC 60601-2-64:2025 applies to the BASIC SAFETY and essential performance of LIGHT ION BEAM ME EQUIPMENT, hereafter referred to as ME EQUIPMENT, used for treatment of patients. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
This document, with the inclusion of TYPE TESTS and SITE TESTS, applies respectively to the manufacturer and specified installation aspects of LIGHT ION BEAM ME EQUIPMENT
– intended for RADIOTHERAPY in human medical practice, including those in which the selection and DISPLAY of operating parameters can be controlled automatically by PROGRAMMABLE ELECTRONIC SUBSYSTEMS (PESS),
– that, in NORMAL USE, deliver a RADIATION BEAM of LIGHT IONS having ENERGY PER NUCLEON in the range 10 MeV/n to 500 MeV/n,
and
– intended to be
• for NORMAL USE, operated under the authority of appropriately licensed or QUALIFIED PERSONS by OPERATORS having the required skills for a particular medical application, for particular SPECIFIED clinical purposes maintained in accordance with the recommendations given in the INSTRUCTIONS FOR USE,
• subject to regular quality assurance performance and calibration checks by a QUALIFIED PERSON.
IEC 60601-2-64:2025 cancels and replaces the first edition published in 2014. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a) harmonization with IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2011 and IEC 60601-1:2005/AMD2:2020;
b) harmonization with IEC 62667:2017 for defined terms and definitions;
c) address revision to neutrons outside the field of irradiation.

  • Standard
    63 pages
    English language
    e-Library read for
    1 day

The amendment to EN IEC 60601-2-22:2020 contains the Annexes ZA (Normative references to international publications with their corresponding European publications) and ZZ (Relationship between this European standard and the General Safety and Performance Requirements of Regulation (EU) 2017/745 aimed to be covered).
These two Annexes are necessary for the harmonization of the standard to the Regulation (EU) 2017/745.

  • Amendment
    8 pages
    English language
    e-Library read for
    1 day

The amendment to EN IEC 60601-2-57 contains the Annexes ZA (Normative references to international publications with their corresponding European publications) and ZZ (Relationship between this European standard and the General Safety and Performance Requirements of Regulation (EU) 2017/745 aimed to be covered).
These two Annexes are necessary for the harmonization of the standard to the Regulation (EU) 2017/745.

  • Amendment
    7 pages
    English language
    e-Library read for
    1 day

IEC 62083:2025, with the inclusion of type tests and site tests, applies to the design, manufacture, installation, and maintenance of the radiotherapy treatment planning system. This document applies to the communication of the radiotherapy treatment planning system with other devices – used in medical practice, – that imports data either through input by the operator or from other devices, – that outputs data to other devices, and – that is intended to be - for normal use, under the authority of appropriately qualified persons, by operators having the required skills and training, - used and maintained in accordance with the recommendations given in the instructions for use, and – used within the environmental conditions specified in the technical description. This document applies to any software application that is used for the development, evaluation, or approval of a treatment plan, whether stand-alone or part of another system. IEC 62083:2025 cancels and replaces the second edition published in 2009. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: – modification of the title from Medical electrical system - Requirements for the safety of radiotherapy treatment planning systems, to Medical device software - Requirements for the safety of radiotherapy treatment planning systems; – Adaptive radiotherapy is added with Clause 16; – The title reflects different implementations of radiotherapy treatment planning systems.

  • Standard
    62 pages
    English language
    e-Library read for
    1 day

IEC 60601-2-64:2025 applies to the BASIC SAFETY and essential performance of LIGHT ION BEAM ME EQUIPMENT, hereafter referred to as ME EQUIPMENT, used for treatment of patients. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. This document, with the inclusion of TYPE TESTS and SITE TESTS, applies respectively to the manufacturer and specified installation aspects of LIGHT ION BEAM ME EQUIPMENT – intended for RADIOTHERAPY in human medical practice, including those in which the selection and DISPLAY of operating parameters can be controlled automatically by PROGRAMMABLE ELECTRONIC SUBSYSTEMS (PESS), – that, in NORMAL USE, deliver a RADIATION BEAM of LIGHT IONS having ENERGY PER NUCLEON in the range 10 MeV/n to 500 MeV/n, and – intended to be • for NORMAL USE, operated under the authority of appropriately licensed or QUALIFIED PERSONS by OPERATORS having the required skills for a particular medical application, for particular SPECIFIED clinical purposes maintained in accordance with the recommendations given in the INSTRUCTIONS FOR USE, • subject to regular quality assurance performance and calibration checks by a QUALIFIED PERSON. IEC 60601-2-64:2025 cancels and replaces the first edition published in 2014. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) harmonization with IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2011 and IEC 60601-1:2005/AMD2:2020; b) harmonization with IEC 62667:2017 for defined terms and definitions; c) address revision to neutrons outside the field of irradiation.

  • Standard
    63 pages
    English language
    e-Library read for
    1 day

IEC 60601-2-57:2023 applies to basic safety and essential performance of equipment incorporating one or more sources of optical radiation in the wavelength range 200 nm to 3 000 nm, with the exception of laser radiation, and intended to create photobiological effects in humans for therapeutic, diagnostic, monitoring, and cosmetic or aesthetic applications; hereafter referred to as light source equipment (ls equipment).

  • Standard
    31 pages
    English language
    e-Library read for
    1 day

The amendment to EN IEC 60601-2-57 contains the Annexes ZA (Normative references to international publications with their corresponding European publications) and ZZ (Relationship between this European standard and the General Safety and Performance Requirements of Regulation (EU) 2017/745 aimed to be covered). These two Annexes are necessary for the harmonization of the standard to the Regulation (EU) 2017/745.

  • Amendment
    7 pages
    English language
    e-Library read for
    1 day

The amendment to EN IEC 60601-2-22:2020 contains the Annexes ZA (Normative references to international publications with their corresponding European publications) and ZZ (Relationship between this European standard and the General Safety and Performance Requirements of Regulation (EU) 2017/745 aimed to be covered). These two Annexes are necessary for the harmonization of the standard to the Regulation (EU) 2017/745.

  • Amendment
    8 pages
    English language
    e-Library read for
    1 day

This document specifies the requirements and general principles governing the biological evaluation of medical devices within a risk management process according to ISO 14971.
This document applies to the biological evaluation of medical devices that have direct contact or indirect contact with either:
—     a patient's body during intended use or reasonably foreseeable misuse; or
—     the body of other users who are not patients, if the medical device is intended for personal protection (e.g. medical gloves, surgical masks).
Biological evaluation assesses the biological safety of the medical device by considering the biological risks associated with:
—     constituents of a medical device; and
—     tissue-device interactions (including physical effects).
The biological evaluation specified in this document can address the biological safety of the medical device, considering the life cycle from design and development through initial use of the finished medical device to final decommissioning or withdrawal from use. The biological evaluation considers both the biological safety of the finished device in first use, and the significance of any changes to the medical device which can occur throughout the life cycle. However, the evaluation of risks related to environmental impacts of decommissioning of medical devices are not within the scope of this document. This document does not mandate re-testing of medical devices that are already on the market and have established and acceptable safety profiles (see 6.6.2).
This document can be useful to support clinical or usability evaluations of medical devices. For example, a biological evaluation is a pre-requisite for conducting a clinical trial. This means that principles outlined in this document can be applied to the evaluation of prototype or development stage devices, as well as to finished medical devices.
Other parts of the ISO 10993 series cover specific aspects of biological evaluation, such as chemical characterization, biological testing, sample preparation, animal welfare and toxicological risk assessment.
For some types of medical devices, specific requirements from other standards (outside the ISO 10993 series) can be considered with a justification for the approach taken if there are differences between the requirements of the ISO 10993 series and those provided in other standards. For example, the ISO 18562 series provides specific requirements for biological evaluation of breathing gas pathway medical devices and ISO 7405 provides specific requirements for biological evaluation of dental devices.
The evaluation of risks related to infectious agents [e.g. bacteria, moulds, yeasts, viruses, transmissible spongiform encephalopathy (TSE) agents] is not within the scope of this document.
NOTE 1        The evaluation of bacterial endotoxins is addressed by ISO 11737-3.
NOTE 2        The evaluation of risks related to viruses, TSE agents and other pathogens originating from materials of animal origin is addressed by the ISO 22442 series.

  • Standard
    59 pages
    English language
    e-Library read for
    1 day

This document specifies requirements for single use transfusion sets for use with pressure infusion equipment capable of generating pressures. It ensures compatibility with containers for blood and blood components as well as intravenous equipment.
This document also provides guidance on specifications relating to the quality and performance of materials used in transfusion sets, to present designations for transfusion set components, and to ensure the compatibility of sets with red cell and plasma blood components.
NOTE            In some countries, the national pharmacopoeia or other national regulations are legally binding and take precedence over this document.

  • Standard
    36 pages
    English language
    e-Library read for
    1 day

This document specifies requirements for single use transfusion gravity sets for medical use to ensure their compatibility with containers for blood and blood components as well as with intravenous equipment.
It also provides guidance on specifications relating to the quality and performance of materials used in transfusion sets, presents designations for transfusion set components, and ensures the compatibility of sets with a range of cellular and plasma blood components.
NOTE            In some countries, the national pharmacopoeia or other national regulations are legally binding and take precedence over this document.

  • Standard
    31 pages
    English language
    e-Library read for
    1 day

This document specifies the requirements and general principles governing the biological evaluation of medical devices within a risk management process according to ISO 14971.
This document applies to the biological evaluation of medical devices that have direct contact or indirect contact with either:
—     a patient's body during intended use or reasonably foreseeable misuse; or
—     the body of other users who are not patients, if the medical device is intended for personal protection (e.g. medical gloves, surgical masks).
Biological evaluation assesses the biological safety of the medical device by considering the biological risks associated with:
—     constituents of a medical device; and
—     tissue-device interactions (including physical effects).
The biological evaluation specified in this document can address the biological safety of the medical device, considering the life cycle from design and development through initial use of the finished medical device to final decommissioning or withdrawal from use. The biological evaluation considers both the biological safety of the finished device in first use, and the significance of any changes to the medical device which can occur throughout the life cycle. However, the evaluation of risks related to environmental impacts of decommissioning of medical devices are not within the scope of this document. This document does not mandate re-testing of medical devices that are already on the market and have established and acceptable safety profiles (see 6.6.2).
This document can be useful to support clinical or usability evaluations of medical devices. For example, a biological evaluation is a pre-requisite for conducting a clinical trial. This means that principles outlined in this document can be applied to the evaluation of prototype or development stage devices, as well as to finished medical devices.
Other parts of the ISO 10993 series cover specific aspects of biological evaluation, such as chemical characterization, biological testing, sample preparation, animal welfare and toxicological risk assessment.
For some types of medical devices, specific requirements from other standards (outside the ISO 10993 series) can be considered with a justification for the approach taken if there are differences between the requirements of the ISO 10993 series and those provided in other standards. For example, the ISO 18562 series provides specific requirements for biological evaluation of breathing gas pathway medical devices and ISO 7405 provides specific requirements for biological evaluation of dental devices.
The evaluation of risks related to infectious agents [e.g. bacteria, moulds, yeasts, viruses, transmissible spongiform encephalopathy (TSE) agents] is not within the scope of this document.
NOTE 1        The evaluation of bacterial endotoxins is addressed by ISO 11737-3.
NOTE 2        The evaluation of risks related to viruses, TSE agents and other pathogens originating from materials of animal origin is addressed by the ISO 22442 series.

  • Standard
    59 pages
    English language
    e-Library read for
    1 day

This document specifies a test method and the minimum requirements for virucidal activity of chemical disinfectant and antiseptic products that form a homogeneous physically stable preparation when diluted with hard water – or in the case of ready-to-use products, i.e. products that are not diluted when applied, – with water. Ready-to-use-products can only be tested at a concentration up to 80 % (97 %, with a modified method for special cases) as some dilution is always produced by adding the test organisms and interfering substance.
This document is applicable to products that are used in the medical area in the fields of hygienic handrub, hygienic handwash, instrument disinfection by immersion, surface disinfection by wiping, spraying, flooding or other means and textile disinfection.
This document is applicable to areas and situations where disinfection is medically indicated. Such indications occur in patient care, for example:
— in hospitals, in community medical facilities, and in dental institutions;
— in clinics of schools, of kindergartens, and of nursing homes;
and can occur in the workplace and in the home. It can also include services such as laundries and kitchens supplying products directly for the patient.
NOTE 1 The method described is intended to determine the activity of commercial formulations or active substances under the conditions in which they are used.
NOTE 2 This method corresponds to a phase 2, step 1 test.
NOTE 3 EN 14885 specifies in detail the relationship of the various tests to one another and to “use recommendations”.

  • Standard
    51 pages
    English language
    e-Library read for
    1 day

This document specifies requirements for single use transfusion gravity sets for medical use to ensure their compatibility with containers for blood and blood components as well as with intravenous equipment.
It also provides guidance on specifications relating to the quality and performance of materials used in transfusion sets, presents designations for transfusion set components, and ensures the compatibility of sets with a range of cellular and plasma blood components.
NOTE            In some countries, the national pharmacopoeia or other national regulations are legally binding and take precedence over this document.

  • Standard
    31 pages
    English language
    e-Library read for
    1 day

This document specifies requirements for single use transfusion sets for use with pressure infusion equipment capable of generating pressures. It ensures compatibility with containers for blood and blood components as well as intravenous equipment.
This document also provides guidance on specifications relating to the quality and performance of materials used in transfusion sets, to present designations for transfusion set components, and to ensure the compatibility of sets with red cell and plasma blood components.
NOTE            In some countries, the national pharmacopoeia or other national regulations are legally binding and take precedence over this document.

  • Standard
    36 pages
    English language
    e-Library read for
    1 day

TThis document specifies a test method and the minimum requirements for virucidal activity of chemical disinfectant and antiseptic products that form a homogeneous physically stable preparation when diluted with hard water - or in the case of ready-to-use products, i.e, products that are not diluted when applied, - with water. Ready-to-use-products can only be tested at a concentration up to 80 % (97 %, with a modified method for special cases) as some dilution is always produced by adding the test organisms and interfering substance.
This document applies to products that are used in the medical area in the fields of hygienic handrub, hygienic handwash, instrument disinfection by immersion, surface disinfection by wiping, spraying, flooding or other means and textile disinfection.
This document applies to areas and situations where disinfection is medically indicated. Such indications occur in patient care, for example:
-   in hospitals, in community medical facilities, and in dental institutions;
-   in clinics of schools, of kindergartens, and of nursing homes;
and can occur in the workplace and in the home. It can also include services such as laundries and kitchens supplying products directly for the patient.
NOTE 1   The method described is intended to determine the activity of commercial formulations or active substances under the conditions in which they are used.
NOTE 2   This method corresponds to a phase 2, step 1 test.
NOTE 3   EN 14885 specifies in detail the relationship of the various tests to one another and to "use recommendations".

  • Standard
    51 pages
    English language
    e-Library read for
    1 day

This document specifies the requirements for sterilized, single-use, gravity feed infusion sets, used together with the volumetric infusion controllers of IEC 60601-2-24.

  • Standard
    19 pages
    English language
    e-Library read for
    1 day

This document specifies requirements for electrically heated steam boilers manufactured from stainless steel specifically dedicated for generating steam for sterilizers and disinfectors.
This document covers only steam boilers that are heated by immersion heaters and which have a maximum allowable pressure (PS) of not greater than 6 bar, a maximum volume (V) of 1 000 litres and a product of PS ∙ V not greater than 3 000 bar ∙ l.

  • Standard
    20 pages
    English language
    e-Library read for
    1 day

This document specifies specific requirements for mammary implants. With regard to safety, this document specifies requirements for intended performance, design attributes, materials, design evaluation, manufacturing, packaging, sterilization and information supplied by the manufacturer.

  • Standard
    68 pages
    English language
    e-Library read for
    1 day

This document specifies a test method for the measurement of residual chemicals used in product manufacture, particularly potentially Type IV allergenic substances employed and remaining in medical gloves. It also provides information on the extraction medium, the method of extraction and quantitative assay of residual chemicals.
This document does not provide information on the allergenic potential, biological evaluation, or safety to the user of any product.

  • Standard
    23 pages
    English language
    e-Library read for
    1 day

This document specifies fundamental requirements and their test methods for unglazed spectacle frames designed for use with prescription lenses. It is applicable to spectacle frames at the point of sale by the manufacturer or supplier to the retailer.
This document is applicable to:
—     all mass-produced spectacle frame types, including rimless mounts, semi-rimless mounts and folding spectacle frames;
—     spectacle frames made with additive manufacturing, for example, 3D printing;
—     spectacle frames made from natural organic materials;
—     the frame or mount of clip-ons designed specifically for attachment to particular models of spectacle frame, but not to their lenses or filters to which ISO 16034 or ISO 12312-1 apply;
—     prescription inserts designed for attachment to particular models of, for example, eye protector, sunglass or diving mask.
Parts of this document are applicable to custom-made frames – see 3.1.3 and Table 1.
NOTE            See Annex A for recommendations on the design of spectacle frames and terms to be used when describing metal frames.
This document is not applicable to spectacle frames used in eye protection, where ISO 16321-1 applies, or to sunglasses with afocal filters, where ISO 12312-1 applies.

  • Standard
    50 pages
    English language
    e-Library read for
    1 day

Frequently Asked Questions

An EU Regulation is a binding legislative act that must be applied in its entirety across the European Union. Unlike directives, regulations do not need to be transposed into national law and are directly applicable in all member states. Regulations are used when uniform application across all EU countries is essential.

Regulation 2017/745 covers "Regulation (EU) 2017/745 of the European Parliament and of the Council of 5 April 2017 on medical devices, amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 and repealing Council Directives 90/385/EEC and 93/42/EEC". This legislation has harmonized European standards (hENs) that provide a presumption of conformity with its essential requirements. There are 400 standards associated with this regulation.

Harmonized standards under 2017/745 are European standards (ENs) developed by CEN, CENELEC, or ETSI in response to a mandate from the European Commission. When these standards are cited in the Official Journal of the European Union, products manufactured in conformity with them benefit from a presumption of conformity with the essential requirements of 2017/745, facilitating CE marking and free movement within the European Economic Area.