90/385/EEC - Council Directive 90/385/EEC of 20 June 1990 on the approximation of the laws of the Member States relating to active implantable medical devices
Council Directive 90/385/EEC of 20 June 1990 on the approximation of the laws of the Member States relating to active implantable medical devices
Direktiva Sveta z dne 20. junija 1990 o približevanju zakonodaje držav članic o aktivnih medicinskih pripomočkih za vsaditev
Seznam harmoniziranih evropskih standardov in istovetnih slovenskih standardov, katerih uporaba ustvarja domnevo o skladnosti proizvoda z zahtevami direk ***Applicable until 25.05.2021
General Information
2022-06-21 - lack of compliance - publication on hold
- Amendment13 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft14 pagesEnglish languagesale 10% offe-Library read for1 day
ISO 14708-3:2008 is applicable to active implantable medical devices intended for electrical stimulation of the central or peripheral nervous system.
ISO 14708-3:2008 is also applicable to all non-implantable parts and accessories of the devices.
The tests that are specified in ISO 14708-3:2008 are type tests intended to be carried out on a sample of a device to show compliance, and are not intended to be used for the routine testing of manufactured products.
- Standard63 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft46 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies the procedure for the assessment of medical devices and their constituent materials with regard to their potential to induce skin sensitization.
This document includes:
— details of in vivo skin sensitization test procedures;
— key factors for the interpretation of the results.
NOTE Instructions for the preparation of materials specifically in relation to the above tests are given in Annex A.
- Standard60 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft59 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies the procedure for the assessment of medical devices and their constituent materials with regard to their potential to induce skin sensitization.
This document includes:
— details of in vivo skin sensitization test procedures;
— key factors for the interpretation of the results.
NOTE Instructions for the preparation of materials specifically in relation to the above tests are given in Annex A.
- Standard60 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft59 pagesEnglish languagesale 10% offe-Library read for1 day
2022-06-21 - lack of compliance - publication on hold
- Amendment13 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft14 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies requirements that are applicable to those active implantable medical devices that are intended to treat hearing impairment via electrical stimulation of the auditory pathways. Devices which treat hearing impairment via means other than electrical stimulation are not covered by this document.
The tests that are specified in this document are type tests and are to be carried out on samples of a device to show compliance.
This document is also applicable to non-implantable parts and accessories of the devices (see NOTE).
The electrical characteristics of the implantable part are determined by either the appropriate method detailed in this document or by any other method demonstrated to have an accuracy equal to, or better than, the method specified. In the case of dispute, the method detailed in this document applies.
NOTE A device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, this document specifies those requirements of non-implantable parts and accessories which could affect the safety or performance of the implantable part.
- Standard77 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft74 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies particular requirements for active implantable medical devices intended to deliver a medicinal substance to site-specific locations within the human body, to provide basic assurance of safety for both patients and users. It amends and supplements ISO 14708-1:2014. The requirements of this document take priority over those of ISO 14708-1.
This document is applicable to active implantable medical devices intended to deliver medicinal substances to site-specific locations within the human body.
This document is also applicable to some non-implantable parts and accessories of the devices defined in Clause 3.
The tests that are specified in this document are type tests intended to be carried out on a sample of a device to show compliance and are not intended to be used for the routine testing of manufactured products.
NOTE This document is not intended to apply to non-implantable infusion systems.
- Standard68 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft74 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies requirements that are applicable to those active implantable medical devices intended to treat bradyarrhythmias and devices that provide therapies for cardiac resynchronization.
The tests that are specified in this document are type tests, and are to be carried out on samples of a device to show compliance.
This document was designed for bradyarrhythmia pulse generators used with endocardial leads or epicardial leads. At the time of this edition, the authors recognized the emergence of leadless technologies for which adaptations of this part will be required. Such adaptations are left to the discretion of manufacturers incorporating these technologies.
This document is also applicable to some non-implantable parts and accessories of the devices (see Note 1).
The electrical characteristics of the implantable pulse generator or lead are determined either by the appropriate method detailed in this particular standard or by any other method demonstrated to have an accuracy equal to, or better than, the method specified. In case of dispute, the method detailed in this particular standard applies.
Any features of an active implantable medical device intended to treat tachyarrhythmias are covered by ISO 14708-6.
NOTE 1 The device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, but there is a need to specify some requirements of non-implantable parts and accessories if they could affect the safety or performance of the implantable device.
NOTE 2 In this document, terms printed in italics are used as defined in Clause 3. Where a defined term is used as a qualifier in another term, it is not printed in italics unless the concept thus qualified is also defined.
- Standard80 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft77 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies requirements that are applicable to implantable cardioverter defibrillators and CRT-Ds and the functions of active implantable medical devices intended to treat tachyarrhythmia.
The tests that are specified in ISO 14708 are type tests and are to be carried out on samples of a device to show compliance.
This document was designed for tachyarrhythmia pulse generators used with either endocardial leads or epicardial leads. At the time of this edition, the authors recognized the emergence of technologies that do not use endocardial or epicardial leads for which adaptations of this part will be required. Such adaptations are left to the discretion of manufacturers incorporating these technologies.
This document is also applicable to some non-implantable parts and accessories of the devices (see Note 1).
The characteristics of the implantable pulse generator or lead shall be determined by either the appropriate method detailed in this document or by any other method demonstrated to have accuracy equal to, or better than, the method specified. In the case of dispute, the method detailed in this document shall apply.
Any aspect of an active implantable medical device intended to treat bradyarrhythmias or cardiac resynchronization is covered by ISO 14708-2.
NOTE 1 The device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, but there is a need to specify some requirements of non-implantable parts and accessories if they could affect the safety or performance of the implantable device.
NOTE 2 In this document, terms printed in italics are used as defined in Clause 3. Where a defined term is used as a qualifier in another term, it is not printed in italics unless the concept thus qualified is also defined.
- Standard75 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft82 pagesEnglish languagesale 10% offe-Library read for1 day
ISO 14708-3:2017 is applicable to active implantable medical devices intended for electrical stimulation of the central or peripheral nervous system.
The tests that are specified in this document are type tests and are to be carried out on a sample of a device to assess device behavioural responses, and are not intended to be used for the routine testing of manufactured products.
- Standard63 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft46 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies requirements for safety and performance of active implantable circulatory support devices, including type tests, animal studies and clinical evaluation requirements.
NOTE The device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, but there is a need to specify main requirements of non-implantable parts and accessories if they could affect the safety or performance of the implantable device.
The tests that are specified in this document are type tests and are to be carried out on a sample of a device to assess device behavioural responses and are not intended to be used for the routine testing of manufactured products.
Included in the scope of this document are:
— ventricular assist devices (VAD), left or right heart support;
— total artificial hearts (TAH);
— biventricular assist devices (biVAD);
— percutaneous assist devices;
— paediatric assist devices.
- Standard79 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft82 pagesEnglish languagesale 10% offe-Library read for1 day
IEC Corrected version
- Corrigendum5 pagesFrench languagesale 10% offe-Library read for1 day
This document specifies requirements and test methods for materials, preformed sterile barrier systems, sterile barrier systems and packaging systems that are intended to maintain sterility of terminally sterilized medical devices until the point of use.
It is applicable to industry, to health care facilities, and to wherever medical devices are placed in sterile barrier systems and sterilized.
It does not cover all requirements for sterile barrier systems and packaging systems for medical devices that are manufactured aseptically. Additional requirements can be necessary for drug/device combinations.
It does not describe a quality assurance system for control of all stages of manufacture.
It does not apply to packaging materials and/or systems used to contain a contaminated medical device during transportation of the item to the site of reprocessing or disposal.
- Amendment11 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies requirements for the development and validation of processes for packaging medical devices that are terminally sterilized. These processes include forming, sealing and assembly of preformed sterile barrier systems, sterile barrier systems and packaging systems.
It is applicable to industry, to health care facilities, and to wherever medical devices are packaged and sterilized.
It does not cover all requirements for packaging medical devices that are manufactured aseptically. Additional requirements can be necessary for drug/device combinations.
- Amendment11 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment11 pagesEnglish languagesale 10% offe-Library read for1 day
IEC Corrected version
- Corrigendum5 pagesFrench languagesale 10% offe-Library read for1 day
This document specifies requirements that are applicable to those active implantable medical devices
that are intended to treat hearing impairment via electrical stimulation of the auditory pathways.
Devices which treat hearing impairment via means other than electrical stimulation are not covered by
this document.
The tests that are specified in this document are type tests and are to be carried out on samples of a
device to show compliance.
This document is also applicable to non-implantable parts and accessories of the devices (see NOTE).
The electrical characteristics of the implantable part are determined by either the appropriate method
detailed in this document or by any other method demonstrated to have an accuracy equal to, or better
than, the method specified. In the case of dispute, the method detailed in this document applies.
NOTE A device that is commonly referred to as an active implantable medical device can in fact be a single
device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of
these parts are required to be either partially or totally implantable, this document specifies those requirements
of non-implantable parts and accessories which could affect the safety or performance of the implantable part.
- Standard77 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft74 pagesEnglish languagesale 10% offe-Library read for1 day
ISO 14708-4:2008 is applicable to active implantable medical devices intended to deliver medicinal substances to site-specific locations within the human body.
ISO 14708-4:2008 is also applicable to some non-implantable parts and accessories of the devices.
The tests that are specified in ISO 14708-4:2008 are type tests intended to be carried out on a sample of a device to show compliance, and are not intended to be used for the routine testing of manufactured products.
- Standard68 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft74 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies requirements that are applicable to implantable cardioverter defibrillators and
CRT-Ds and the functions of active implantable medical devices intended to treat tachyarrhythmia.
The tests that are specified in ISO 14708 are type tests and are to be carried out on samples of a device
to show compliance.
This document was designed for tachyarrhythmia pulse generators used with either endocardial leads
or epicardial leads. At the time of this edition, the authors recognized the emergence of technologies
that do not use endocardial or epicardial leads for which adaptations of this part will be required. Such
adaptations are left to the discretion of manufacturers incorporating these technologies.
This document is also applicable to some non-implantable parts and accessories of the devices (see
Note 1).
The characteristics of the implantable pulse generator or lead shall be determined by either the
appropriate method detailed in this document or by any other method demonstrated to have accuracy
equal to, or better than, the method specified. In the case of dispute, the method detailed in this
document shall apply.
Any aspect of an active implantable medical device intended to treat bradyarrhythmias or cardiac
resynchronization is covered by ISO 14708-2.
NOTE 1 The device that is commonly referred to as an active implantable medical device can in fact be
a single device, a combination of devices, or a combination of a device or devices and one or more accessories.
Not all of these parts are required to be either partially or totally implantable, but there is a need to specify
some requirements of non-implantable parts and accessories if they could affect the safety or performance of the
implantable device.
NOTE 2 In this document, terms printed in italics are used as defined in Clause 3. Where a defined term is
used as a qualifier in another term, it is not printed in italics unless the concept thus qualified is also defined.
- Standard75 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft82 pagesEnglish languagesale 10% offe-Library read for1 day
ISO 14708-5:2010 specifies requirements for safety and performance of active implantable circulatory support devices. It is not applicable to extracorporeal perfusion devices, cardiomyoplasty, heart restraint devices and counter-pulsation devices, such as extra- or intra-aortic balloon pumps.
ISO 14708-5:2010 specifies type tests, animal studies and clinical evaluation requirements.
- Standard79 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft82 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies requirements that are applicable to those active implantable medical devices
intended to treat bradyarrhythmias and devices that provide therapies for cardiac resynchronization.
The tests that are specified in this document are type tests, and are to be carried out on samples of a
device to show compliance.
This document was designed for bradyarrhythmia pulse generators used with endocardial leads
or epicardial leads. At the time of this edition, the authors recognized the emergence of leadless
technologies for which adaptations of this part will be required. Such adaptations are left to the
discretion of manufacturers incorporating these technologies.
This document is also applicable to some non-implantable parts and accessories of the devices (see
Note 1).
The electrical characteristics of the implantable pulse generator or lead are determined either by the
appropriate method detailed in this particular standard or by any other method demonstrated to have
an accuracy equal to, or better than, the method specified. In case of dispute, the method detailed in
this particular standard applies.
Any features of an active implantable medical device intended to treat tachyarrhythmias are covered
by ISO 14708-6.
NOTE 1 The device that is commonly referred to as an active implantable medical device can in fact be
a single device, a combination of devices, or a combination of a device or devices and one or more accessories.
Not all of these parts are required to be either partially or totally implantable, but there is a need to specify
some requirements of non-implantable parts and accessories if they could affect the safety or performance of the
implantable device.
NOTE 2 In this document, terms printed in italics are used as defined in Clause 3. Where a defined term is
used as a qualifier in another term, it is not printed in italics unless the concept thus qualified is also defined
- Standard80 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft77 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies requirements and test methods for materials, preformed sterile barrier systems, sterile barrier systems and packaging systems that are intended to maintain sterility of terminally sterilized medical devices until the point of use.
It is applicable to industry, to health care facilities, and to wherever medical devices are placed in sterile barrier systems and sterilized.
It does not cover all requirements for sterile barrier systems and packaging systems for medical devices that are manufactured aseptically. Additional requirements can be necessary for drug/device combinations.
It does not describe a quality assurance system for control of all stages of manufacture.
It does not apply to packaging materials and/or systems used to contain a contaminated medical device during transportation of the item to the site of reprocessing or disposal.
- Amendment11 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies requirements for the development and validation of processes for packaging medical devices that are terminally sterilized. These processes include forming, sealing and assembly of preformed sterile barrier systems, sterile barrier systems and packaging systems.
It is applicable to industry, to health care facilities, and to wherever medical devices are packaged and sterilized.
It does not cover all requirements for packaging medical devices that are manufactured aseptically. Additional requirements can be necessary for drug/device combinations.
- Amendment11 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment11 pagesEnglish languagesale 10% offe-Library read for1 day
2021-12-09- JO- the EN was submitted to FV wiith the link to the legislation (90/385/EEC, 93/42/EEC)
however since these legilsations have been repealed, the link has been removed and the EN is to proceed to publication
- Amendment11 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft11 pagesEnglish languagesale 10% offe-Library read for1 day
This document provides general principles for the systematic evaluation of the potential and observed degradation of medical devices through the design and performance of in vitro degradation studies. Information obtained from these studies can be used in the biological evaluation described in the ISO 10993 series.
This document is applicable to both materials designed to degrade in the body as well as materials that are not intended to degrade.
This document is not applicable to:
a) the evaluation of degradation which occurs by purely mechanical processes; methodologies for the production of this type of degradation product are described in specific product standards, where available;
NOTE Purely mechanical degradation causes mostly particulate matter. Although this is excluded from the scope of this document, such degradation products can evoke a biological response and can undergo biological evaluation as described in other parts of ISO 10993.
b) leachable components which are not degradation products;
c) medical devices or components that do not contact the patient's body directly or indirectly.
- Standard22 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft20 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment27 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment27 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft52 pagesEnglish languagesale 10% offe-Library read for1 day
2021-12-09- JO- the EN was submitted to FV wiith the link to the legislation (90/385/EEC, 93/42/EEC)
however since these legilsations have been repealed, the link has been removed and the EN is to proceed to publication
- Amendment11 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft11 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies requirements for the characterization of a liquid chemical sterilizing agent and for the development, validation, process control and monitoring of sterilization by liquid chemical sterilizing agents of single-use medical devices comprising, in whole or in part, materials of animal origin.
This document covers the control of risks arising from contamination with bacteria and fungi by application of a liquid chemical sterilization process. Risks associated with other microorganisms can be assessed using other methods (see NOTE 1).
This document is not applicable to material of human origin.
This document does not describe methods for the validation of the inactivation of viruses and transmissible spongiform encephalopathy (TSE) agents (see NOTE 2 and NOTE 3).
This document does not describe methods for validation of the inactivation or elimination of protozoa and parasites.
The requirements for validation and routine control described in this document are only applicable to the defined sterilization process of a medical device, which is performed after the manufacturing process, and do not take account of the lethal effects of other bioburden reduction steps (see NOTE 4).
This document does not specify tests to establish the effects of any chosen sterilization process upon the fitness for use of the medical device (see NOTE 5).
This document does not cover the level of residual sterilizing agent within medical devices (see NOTE 6).
Guidance for the characterization of a liquid chemical sterilizing agent and for the development, validation, process control and monitoring of sterilization by liquid chemical sterilizing agents of single-use medical devices comprising, in whole or in part, materials of animal origin is provided in informative Annex A.
NOTE 1 The prior application of risk management principles to medical devices utilizing animal tissues, as described in ISO 22442-1 is important. ISO 18362 provides information on control of microbial risks during processing of cell-based health-care products.
NOTE 2 Liquid chemical sterilizing agents traditionally employed to sterilize animal tissues in medical devices might not be effective in inactivating the causative agents of TSE such as bovine spongiform encephalopathy (BSE), or scrapie. Satisfactory validation in accordance with this document does not necessarily demonstrate inactivation of infective agents of this type. Risk controls related to sourcing, collection and handling of animal materials are described in ISO 22442-2.
NOTE 3 The validation of the inactivation, elimination, or elimination and inactivation of viruses and TSE agents is described in ISO 22442-3.
NOTE 4 Manufacturing processes for medical devices containing animal tissues frequently include exposure to chemical agents which can significantly reduce the bioburden on the medical device. Following the manufacturing process, a medical device is exposed to a specified sterilization process.
NOTE 5 Such testing is a crucial part of the design and development of a medical device.
NOTE 6 ISO 10993-17 specifies a method to establish allowable limits for residues of sterilizing agents.
NOTE 7 Standards for quality management systems (see ISO 13485) can be used in the control of all stages of manufacture including the sterilization process.
- Standard50 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft45 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies requirements and gives guidance on the procedures in the preparation of samples and the selection of reference materials for medical device testing primarily in biological test systems primarily in accordance with one or more parts of the ISO 10993 series.
Specifically, this document addresses the following:
— test sample selection;
— selection of representative portions from a medical device;
— test sample preparation;
— experimental controls;
— selection of, and requirements for, reference materials;
— preparation of extracts.
This document is not applicable to live cells but can be relevant to the material or medical device components of combination products containing live cells.
Extractions for chemical characterization are covered in ISO 10993-18. Clause 7, 8, 9, 10 [with the exception of 10.3.5 and 10.3.11 b)], and 11 can apply to extractions for chemical characterization. Information given in C.1 to C.4 can also be relevant.
- Standard32 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft32 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment13 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft4 pagesEnglish languagesale 10% offe-Library read for1 day
This document provides general principles for the systematic evaluation of the potential and observed
degradation of medical devices through the design and performance of in vitro degradation studies.
Information obtained from these studies can be used in the biological evaluation described in the
ISO 10993 series.
This document is applicable to both materials designed to degrade in the body as well as materials that
are not intended to degrade.
This document is not applicable to:
a) the evaluation of degradation which occurs by purely mechanical processes; methodologies for the
production of this type of degradation product are described in specific product standards, where
available;
NOTE Purely mechanical degradation causes mostly particulate matter. Although this is excluded from
the scope of this document, such degradation products can evoke a biological response and can undergo
biological evaluation as described in other parts of ISO 10993.
b) leachable components which are not degradation products;
c) medical devices or components that do not contact the patient's body directly or indirectly
- Standard22 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft20 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies the requirements for and provides guidance on the specification, selection, qualification, bio-decontamination, validation, operation and control of isolator systems related to aseptic processing of health care products and processing of cell-based health care products.
This document does not specify requirements for restricted access barrier systems (RABS).
This document does not supersede or replace national regulatory requirements such as Good Manufacturing Practices (GMPs) and/or compendia requirements that pertain in particular to national or regional jurisdictions.
This document does not specify requirements for isolators used for sterility testing; however, some of the principles and information in this document could be applicable to this application.
This document does not define biosafety containment requirements.
- Standard39 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft44 pagesEnglish languagesale 10% offe-Library read for1 day
NOTE 1 There is guidance or rationale for this Clause contained in Clause A.2.
This document specifies the requirements for information supplied by the manufacturer for a medical device or by the manufacturer for an accessory, as defined in 3.1. This document includes the generally applicable requirements for identification and labels on a medical device or accessory, the packaging, marking of a medical device or accessory, and accompanying information. This document does not specify the means by which the information is to be supplied.
NOTE 2 Some authorities having jurisdiction impose different requirements for the identification, marking and documentation of a medical device or accessory.
Specific requirements of medical device product standards or group standards take precedence over requirements of this document.
- Standard81 pagesEnglish languagesale 10% offe-Library read for1 day
- Standard81 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft82 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment27 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment27 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft52 pagesEnglish languagesale 10% offe-Library read for1 day
2021-08-25-JO: BT Decision (BT169/DG12292/DV) to remove the link to the MDR
2021-06-24: blocked so that the link to the MDD can be removed by BT from the document
2019-12-03-JO-Assessment at ENQ uplaoded with due date of 06 Jan.2020 plus extention requested by HAS Consultant
- Amendment58 pagesEnglish languagesale 10% offe-Library read for1 day
2021-08-25-JO: BT Decision (BT169/DG12292/DV) to remove the link to the MDR
2021-06-24: blocked so that the link to the MDD can be removed by BT from the document
2019-12-03-JO-Assessment at ENQ uplaoded with due date of 06 Jan.2020 plus extention requested by HAS Consultant
- Amendment58 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies the procedure for the assessment of medical devices and their constituent materials with regard to their potential to produce irritation. The tests are designed to predict and classify the irritation potential of medical devices, materials or their extracts according to ISO 10993‑1 and ISO 10993‑2.
This document includes:
— pre-test considerations for irritation, including in silico and in vitro methods for dermal exposure;
— details of in vitro and in vivo irritation test procedures;
— key factors for the interpretation of the results.
- Standard76 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft71 pagesEnglish languagesale 10% offe-Library read for1 day
2020-11-16_JO_BT167/DG11868/DV to remove the link to (MDD 93/42/EEC) , 90/385/EEC (AIMD) and M/295
- Amendment12 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies requirements and gives guidance on the procedures in the preparation of
samples and the selection of reference materials for medical device testing primarily in biological test
systems primarily in accordance with one or more parts of the ISO 10993 series.
Specifically, this document addresses the following:
— test sample selection;
— selection of representative portions from a medical device;
— test sample preparation;
— experimental controls;
— selection of, and requirements for, reference materials;
— preparation of extracts.
This document is not applicable to live cells but can be relevant to the material or medical device
components of combination products containing live cells.
Extractions for chemical characterization are covered in ISO 10993-18. Clause 7, 8, 9, 10 [with the
exception of 10.3.5 and 10.3.11 b)], and 11 can apply to extractions for chemical characterization.
Information given in C.1 to C.4 can also be relevant.
- Standard32 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft32 pagesEnglish languagesale 10% offe-Library read for1 day
2020-11-16_JO_BT167/DG11868/DV to remove the link to (MDD 93/42/EEC) , 90/385/EEC (AIMD) and M/295
- Amendment12 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies the requirements for and provides guidance on the specification, selection, qualification, bio-decontamination, validation, operation and control of isolator systems related to aseptic processing of health care products and processing of cell-based health care products.
This document does not specify requirements for restricted access barrier systems (RABS).
This document does not supersede or replace national regulatory requirements such as Good Manufacturing Practices (GMPs) and/or compendia requirements that pertain in particular to national or regional jurisdictions.
This document does not specify requirements for isolators used for sterility testing; however, some of the principles and information in this document could be applicable to this application.
This document does not define biosafety containment requirements.
- Standard39 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft44 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies:
— the general principles governing the biological evaluation of medical devices within a risk management process;
— the general categorization of medical devices based on the nature and duration of their contact with the body;
— the evaluation of existing relevant data from all sources;
— the identification of gaps in the available data set on the basis of a risk analysis;
— the identification of additional data sets necessary to analyse the biological safety of the medical device;
— the assessment of the biological safety of the medical device.
This document applies to evaluation of materials and medical devices that are expected to have direct or indirect contact with:
— the patient's body during intended use;
— the user's body, if the medical device is intended for protection (e.g., surgical gloves, masks and others).
This document is applicable to biological evaluation of all types of medical devices including active, non-active, implantable and non-implantable medical devices.
This document also gives guidelines for the assessment of biological hazards arising from:
— risks, such as changes to the medical device over time, as a part of the overall biological safety assessment;
— breakage of a medical device or medical device component which exposes body tissue to new or novel materials.
Other parts of ISO 10993 cover specific aspects of biological assessments and related tests. Device-specific or product standards address mechanical testing.
This document excludes hazards related to bacteria, moulds, yeasts, viruses, transmissible spongiform encephalopathy (TSE) agents and other pathogens.
- Standard50 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft51 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies the procedure for the assessment of medical devices and their constituent materials with regard to their potential to produce irritation by using an in vitro reconstructed human epidermis model.
- Standard76 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft71 pagesEnglish languagesale 10% offe-Library read for1 day
This document addresses good clinical practice for the design, conduct, recording and reporting of clinical investigations carried out in human subjects to assess the clinical performance or effectiveness and safety of medical devices.
For post-market clinical investigations, the principles set forth in this document are intended to be followed as far as relevant, considering the nature of the clinical investigation (see Annex I).
This document specifies general requirements intended to
— protect the rights, safety and well-being of human subjects,
— ensure the scientific conduct of the clinical investigation and the credibility of the clinical investigation results,
— define the responsibilities of the sponsor and principal investigator, and
— assist sponsors, investigators, ethics committees, regulatory authorities and other bodies involved in the conformity assessment of medical devices.
NOTE 1 Users of this document need to consider whether other standards and/or national requirements also apply to the investigational device(s) under consideration or the clinical investigation. If differences in requirements exist, the most stringent apply.
NOTE 2 For Software as a Medical Device (SaMD) demonstration of the analytical validity (the SaMD's output is accurate for a given input), and where appropriate, the scientific validity (the SaMD's output is associated to the intended clinical condition/physiological state), and clinical performance (the SaMD's output yields a clinically meaningful association to the target use) of the SaMD, the requirements of this document apply as far as relevant (see Reference [4]). Justifications for exemptions from this document can consider the uniqueness of indirect contact between subjects and the SaMD.
This document does not apply to in vitro diagnostic medical devices. However, there can be situations, dependent on the device and national or regional requirements, where users of this document might consider whether specific sections and/or requirements of this document could be applicable.
- Standard96 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft96 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies:
— the general principles governing the biological evaluation of medical devices within a risk
management process;
— the general categorization of medical devices based on the nature and duration of their contact with
the body;
— the evaluation of existing relevant data from all sources;
— the identification of gaps in the available data set on the basis of a risk analysis;
— the identification of additional data sets necessary to analyse the biological safety of the medical
device;
— the assessment of the biological safety of the medical device.
This document applies to evaluation of materials and medical devices that are expected to have direct
or indirect contact with:
— the patient's body during intended use;
— the user’s body, if the medical device is intended for protection (e.g., surgical gloves, masks and
others).
This document is applicable to biological evaluation of all types of medical devices including active,
non-active, implantable and non-implantable medical devices.
This document also gives guidelines for the assessment of biological hazards arising from:
— risks, such as changes to the medical device over time, as a part of the overall biological safety
assessment;
— breakage of a medical device or medical device component which exposes body tissue to new or
novel materials.
Other parts of ISO 10993 cover specific aspects of biological assessments and related tests. Devicespecific
or product standards address mechanical testing.
This document excludes hazards related to bacteria, moulds, yeasts, viruses, transmissible spongiform
encephalopathy (TSE) agents and other pathogens.
- Standard50 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft51 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies a framework for the identification, and if necessary, quantification of constituents of a medical device, allowing the identification of biological hazards and the estimation and control of biological risks from material constituents, using a generally stepwise approach to the chemical characterization which can include one or more of the following:
— the identification of its materials of construction (medical device configuration);
— the characterization of the materials of construction via the identification and quantification of their chemical constituents (material composition);
— the characterization of the medical device for chemical substances that were introduced during manufacturing (e.g. mould release agents, process contaminants, sterilization residues);
— the estimation (using laboratory extraction conditions) of the potential of the medical device, or its materials of construction, to release chemical substances under clinical use conditions (extractables);
— the measurement of chemical substances released from a medical device under its clinical conditions of use (leachables).
This document can also be used for chemical characterization (e.g. the identification and/or quantification) of degradation products. Information on other aspects of degradation assessment are covered in ISO 10993-9, ISO 10993-13, ISO 10993-14 and ISO 10993-15.
The ISO 10993 series is applicable when the material or medical device has direct or indirect body contact (see ISO 10993-1 for categorization by nature of body contact).
This document is intended for suppliers of materials and manufacturers of medical devices, to support a biological evaluation.
- Standard81 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft69 pagesEnglish languagesale 10% offe-Library read for1 day
1.1 This document specifies the general criteria for tests of sterility on medical devices that have been exposed to a treatment with the sterilizing agent which has been reduced relative to that anticipated to be used in routine sterilization processing. These tests are intended to be performed when defining, validating or maintaining a sterilization process.
1.2 This document is not applicable to:
a) sterility testing for routine release of product that has been subjected to a sterilization process,
b) performing a test for sterility (see 3.12),
NOTE 1 The performance of a) or b) is not a requirement of ISO 11135, ISO 11137-1, ISO 11137-2, ISO 14160, ISO 14937, ISO 17665-1 or ISO 20857.
c) test of sterility or test for sterility for demonstration of product shelf life, stability and/or package integrity, and
d) culturing of biological indicators or inoculated products.
NOTE 2 Guidance on culturing biological indicators is included in ISO 11138-7.
- Standard35 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft28 pagesEnglish languagesale 10% offe-Library read for1 day
2020-10-13:JO-CLC/TC 62 Decision 2019-05-21/22 in Oslo to remove the link to the published standard
- Amendment9 pagesEnglish languagesale 10% offe-Library read for1 day
This document addresses good clinical practice for the design, conduct, recording and reporting of
clinical investigations carried out in human subjects to assess the clinical performance or effectiveness
and safety of medical devices.
For post-market clinical investigations, the principles set forth in this document are intended to be
followed as far as relevant, considering the nature of the clinical investigation (see Annex I).
This document specifies general requirements intended to
— protect the rights, safety and well-being of human subjects,
— ensure the scientific conduct of the clinical investigation and the credibility of the clinical
investigation results,
— define the responsibilities of the sponsor and principal investigator, and
— assist sponsors, investigators, ethics committees, regulatory authorities and other bodies involved
in the conformity assessment of medical devices.
NOTE 1 Users of this document need to consider whether other standards and/or national requirements
also apply to the investigational device(s) under consideration or the clinical investigation. If differences in
requirements exist, the most stringent apply.
NOTE 2 For Software as a Medical Device (SaMD) demonstration of the analytical validity (the SaMD’s output
is accurate for a given input), and where appropriate, the scientific validity (the SaMD’s output is associated to the
intended clinical condition/physiological state), and clinical performance (the SaMD’s output yields a clinically
meaningful association to the target use) of the SaMD, the requirements of this document apply as far as relevant
(see Reference [4]). Justifications for exemptions from this document can consider the uniqueness of indirect
contact between subjects and the SaMD.
This document does not apply to in vitro diagnostic medical devices. However, there can be situations,
dependent on the device and national or regional requirements, where users of this document might
consider whether specific sections and/or requirements of this document could be applicable.
- Standard96 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft96 pagesEnglish languagesale 10% offe-Library read for1 day
2020-10-13:JO-CLC/TC 62 Decision 2019-05-21/22 in Oslo to remove the link to the published standard
- Amendment9 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies requirements and test methods for materials, preformed sterile barrier systems, sterile barrier systems and packaging systems that are intended to maintain sterility of terminally sterilized medical devices until the point of use.
It is applicable to industry, to health care facilities, and to wherever medical devices are placed in sterile barrier systems and sterilized.
It does not cover all requirements for sterile barrier systems and packaging systems for medical devices that are manufactured aseptically. Additional requirements can be necessary for drug/device combinations.
It does not describe a quality assurance system for control of all stages of manufacture.
It does not apply to packaging materials and/or systems used to contain a contaminated medical device during transportation of the item to the site of reprocessing or disposal.
- Standard52 pagesEnglish languagesale 10% offe-Library read for1 day
- Standard52 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft52 pagesEnglish languagesale 10% offe-Library read for1 day