90/385/EEC - Council Directive 90/385/EEC of 20 June 1990 on the approximation of the laws of the Member States relating to active implantable medical devices
Council Directive 90/385/EEC of 20 June 1990 on the approximation of the laws of the Member States relating to active implantable medical devices
Direktiva Sveta z dne 20. junija 1990 o približevanju zakonodaje držav članic o aktivnih medicinskih pripomočkih za vsaditev
Seznam harmoniziranih evropskih standardov in istovetnih slovenskih standardov, katerih uporaba ustvarja domnevo o skladnosti proizvoda z zahtevami direktive Sveta 90/385/EEC
General Information
This document specifies:
— the general principles governing the biological evaluation of medical devices within a risk
management process;
— the general categorization of medical devices based on the nature and duration of their contact with
the body;
— the evaluation of existing relevant data from all sources;
— the identification of gaps in the available data set on the basis of a risk analysis;
— the identification of additional data sets necessary to analyse the biological safety of t...view more
- Standard50 pagesEnglish language
sale 10% off- e-Library read for1 day
- Draft51 pagesEnglish language
sale 10% off- e-Library read for1 day
This document specifies a framework for the identification, and if necessary, quantification of
constituents of a medical device, allowing the identification of biological hazards and the estimation
and control of biological risks from material constituents, using a generally stepwise approach to the
chemical characterization which can include one or more of the following:
— the identification of its materials of construction (medical device configuration);
— the characterization of the mat...view more
- Standard81 pagesEnglish language
sale 10% off- e-Library read for1 day
- Draft69 pagesEnglish language
sale 10% off- e-Library read for1 day
2020-10-13:JO-CLC/TC 62 Decision 2019-05-21/22 in Oslo to remove the link to the published standard
- Amendment9 pagesEnglish language
sale 10% off- e-Library read for1 day
This document addresses good clinical practice for the design, conduct, recording and reporting of
clinical investigations carried out in human subjects to assess the clinical performance or effectiveness
and safety of medical devices.
For post-market clinical investigations, the principles set forth in this document are intended to be
followed as far as relevant, considering the nature of the clinical investigation (see Annex I).
This document specifies general requirements intended to
— ...view more
- Standard96 pagesEnglish language
sale 10% off- e-Library read for1 day
- Draft96 pagesEnglish language
sale 10% off- e-Library read for1 day
EN-ISO 11737-2 specifies the general criteria for tests of sterility on medical devices that have been exposed to a treatment with the sterilizing agent which has been reduced relative to that anticipated to be used in routine sterilization processing. These tests are intended to be performed when defining, validating or maintaining a sterilization process. This document is not applicable to: a) sterility testing for routine release of product that has been subjected to a sterilization process, ...view more
- Standard35 pagesEnglish language
sale 10% off- e-Library read for1 day
- Draft28 pagesEnglish language
sale 10% off- e-Library read for1 day
This document specifies terminology, principles and a process for risk management of medical devices,
including software as a medical device and in vitro diagnostic medical devices. The process described in
this document intends to assist manufacturers of medical devices to identify the hazards associated with
the medical device, to estimate and evaluate the associated risks, to control these risks, and to monitor
the effectiveness of the controls.
The requirements of this document are appl...view more
- Standard46 pagesEnglish language
sale 10% off- e-Library read for1 day
- Draft51 pagesEnglish language
sale 10% off- e-Library read for1 day
This document specifies requirements and test methods for materials, preformed sterile barrier systems, sterile barrier systems and packaging systems that are intended to maintain sterility of terminally sterilized medical devices until the point of use. It is applicable to industry, to health care facilities, and to wherever medical devices are placed in sterile barrier systems and sterilized. It does not cover all requirements for sterile barrier systems and packaging systems for medical devic...view more
- Standard52 pagesEnglish language
sale 10% off- e-Library read for1 day
- Standard52 pagesEnglish language
sale 10% off- e-Library read for1 day
- Draft52 pagesEnglish language
sale 10% off- e-Library read for1 day
This document specifies requirements for the development and validation of processes for packaging medical devices that are terminally sterilized. These processes include forming, sealing and assembly of preformed sterile barrier systems, sterile barrier systems and packaging systems.
It is applicable to industry, to health care facilities, and to wherever medical devices are packaged and sterilized.
It does not cover all requirements for packaging medical devices that are manufactured aseptic...view more
- Standard21 pagesEnglish language
sale 10% off- e-Library read for1 day
- Draft26 pagesEnglish language
sale 10% off- e-Library read for1 day
This document specifies requirements and test methods for materials, preformed sterile barrier systems, sterile barrier systems and packaging systems that are intended to maintain sterility of terminally sterilized medical devices until the point of use.
It is applicable to industry, to health care facilities, and to wherever medical devices are placed in sterile barrier systems and sterilized.
It does not cover all requirements for sterile barrier systems and packaging systems for medical dev...view more
- Standard52 pagesEnglish language
sale 10% off- e-Library read for1 day
- Standard52 pagesEnglish language
sale 10% off- e-Library read for1 day
- Draft52 pagesEnglish language
sale 10% off- e-Library read for1 day
This Standard specifies requirements for the development, validation and routine control of a low temperature steam and formaldehyde (LTSF) sterilization process for medical devices using a mixture of low temperature steam and formaldehyde as sterilizing agent and which operates below ambient pressure. This document is intended to be applied by process developers, manufacturers of sterilization equipment, manufacturers of medical devices to be sterilized and the organizations with responsibility...view more
- Standard63 pagesEnglish language
sale 10% off- e-Library read for1 day
- Draft49 pagesEnglish language
sale 10% off- e-Library read for1 day
20191119 - Negative assessment addressed through BT decision C168/2019 (SV)
2019-03-07-JO- under HAS assessment at PUB stage. E&Y Report was due on 03 March 2019- Awaiting for the assessment report E&Y Report
- Amendment19 pagesEnglish language
sale 10% off- e-Library read for1 day
- Amendment4 pagesEnglish language
sale 10% off- e-Library read for1 day
20191119 - Negative assessment addressed through BT decision C168/2019 (SV)
2019-03-07-JO- under HAS assessment at PUB stage. E&Y Report was due on 03 March 2019. Awaiting for assessement report from E&Y.
2018-10-17 - TAN : Lack of compliance
- Amendment22 pagesEnglish language
sale 10% off- e-Library read for1 day
- Amendment7 pagesEnglish language
sale 10% off- e-Library read for1 day
This document specifies terminology, principles and a process for risk management of medical devices, including software as a medical device and in vitro diagnostic medical devices. The process described in this document intends to assist manufacturers of medical devices to identify the hazards associated with the medical device, to estimate and evaluate the associated risks, to control these risks, and to monitor the effectiveness of the controls.
The requirements of this document are applicab...view more
- Standard46 pagesEnglish language
sale 10% off- e-Library read for1 day
- Draft51 pagesEnglish language
sale 10% off- e-Library read for1 day
20191119 - Negative assessment addressed through BT decision C168/2019 (SV)
2019-03-07-JO- under HAS assessment at PUB stage. E&Y Report was due on 03 March 2019- Awaiting for the assessment report E&Y Report
- Amendment19 pagesEnglish language
sale 10% off- e-Library read for1 day
- Amendment4 pagesEnglish language
sale 10% off- e-Library read for1 day
20191119 - Negative assessment addressed through BT decision C168/2019 (SV)
2019-03-07-JO- under HAS assessment at PUB stage. E&Y Report was due on 03 March 2019. Awaiting for assessement report from E&Y.
2018-10-17 - TAN : Lack of compliance
- Amendment22 pagesEnglish language
sale 10% off- e-Library read for1 day
- Amendment7 pagesEnglish language
sale 10% off- e-Library read for1 day
1.1 Inclusions
1.1.1 This document specifies requirements for the development, validation and routine control of a low temperature steam and formaldehyde (LTSF) sterilization process for medical devices using a mixture of low temperature steam and formaldehyde as sterilizing agent and which operates below ambient pressure.
NOTE Although the scope of this document is limited to medical devices, it specifies requirements and provides guidance that can be applicable to other products and equipmen...view more
- Standard63 pagesEnglish language
sale 10% off- e-Library read for1 day
- Draft49 pagesEnglish language
sale 10% off- e-Library read for1 day
This part of ISO 10993 describes the procedure for the assessment of medical devices and their constituent
materials with regard to their potential to produce irritation and skin sensitization.
This part of ISO 10993 includes:
a) pretest considerations for irritation, including in silico and in vitro methods for dermal exposure;
b) details of in vivo (irritation and sensitization) test procedures;
c) key factors for the interpretation of the results.
Instructions are given in Annex A for t...view more
- Standard81 pagesEnglish language
sale 10% off- e-Library read for1 day
This document specifies:
— the general principles governing the biological evaluation of medical devices within a risk management process;
— the general categorization of medical devices based on the nature and duration of their contact with the body;
— the evaluation of existing relevant data from all sources;
— the identification of gaps in the available data set on the basis of a risk analysis;
— the identification of additional data sets necessary to analyse the biological safety of the...view more
- Standard50 pagesEnglish language
sale 10% off- e-Library read for1 day
- Draft51 pagesEnglish language
sale 10% off- e-Library read for1 day
This document specifies requirements for sterilizing filtration as part of aseptic processing of health
care products conducted in accordance with ISO 13408-1. It also offers guidance to filter users
concerning general requirements for set-up, validation and routine operation of a sterilizing filtration
process.
This document is not applicable to removal of viruses.
Sterilizing filtration is not applicable to fluids that intentionally contain particles larger than the pore
size of the filt...view more
- Standard47 pagesEnglish language
sale 10% off- e-Library read for1 day
ISO 10993-11:2017 specifies requirements and gives guidance on procedures to be followed in the evaluation of the potential for medical device materials to cause adverse systemic reactions.
- Standard81 pagesEnglish language
sale 10% off- e-Library read for1 day
CCMC - creation of a 2nd corrigendum as the instructions in the 1st corrigendum were incomplete and unclear
- Corrigendum9 pagesEnglish language
sale 10% off- e-Library read for1 day
CCMC - creation of a 2nd corrigendum as the instructions in the 1st corrigendum were incomplete and unclear
- Corrigendum9 pagesEnglish language
sale 10% off- e-Library read for1 day
ISO 13408-2:2018 specifies requirements for sterilizing filtration as part of aseptic processing of health care products conducted in accordance with ISO 13408‑1. It also offers guidance to filter users concerning general requirements for set-up, validation and routine operation of a sterilizing filtration process.
ISO 13408-2:2018 is not applicable to removal of viruses.
Sterilizing filtration is not applicable to fluids that intentionally contain particles larger than the pore size of the fi...view more
- Standard47 pagesEnglish language
sale 10% off- e-Library read for1 day
This document specifies requirements and provides guidance on the enumeration and microbial
characterization of the population of viable microorganisms on or in a health care product, component,
raw material or package.
NOTE 1 The nature and extent of microbial characterization is dependent on the intended use of bioburden data.
NOTE 2 See Annex A for guidance on Clauses 1 to 9.
This document does not apply to the enumeration or identification of viral, prion or protozoan
contaminants. Thi...view more
- Standard58 pagesEnglish language
sale 10% off- e-Library read for1 day
ISO 11737-1:2018 specifies requirements and provides guidance on the enumeration and microbial characterization of the population of viable microorganisms on or in a health care product, component, raw material or package.
NOTE 1 The nature and extent of microbial characterization is dependent on the intended use of bioburden data.
NOTE 2 See Annex A for guidance on Clauses 1 to 9.
ISO 11737-1:2018 does not apply to the enumeration or identification of viral, prion or protozoan contaminants. ...view more
- Standard58 pagesEnglish language
sale 10% off- e-Library read for1 day
This document provides principles on designing and performing toxicokinetic studies relevant to
medical devices. Annex A describes the considerations for inclusion of toxicokinetic studies in the
biological evaluation of medical devices.
- Standard26 pagesEnglish language
sale 10% off- e-Library read for1 day
ISO 10993-4:2017 specifies general requirements for evaluating the interactions of medical devices with blood.
It describes
a) a classification of medical devices that are intended for use in contact with blood, based on the intended use and duration of contact as defined in ISO 10993‑1,
b) the fundamental principles governing the evaluation of the interaction of devices with blood,
c) the rationale for structured selection of tests according to specific categories, together with the princip...view more
- Standard83 pagesEnglish language
sale 10% off- e-Library read for1 day
ISO 10993-16:2017 provides principles on designing and performing toxicokinetic studies relevant to medical devices. Annex A describes the considerations for inclusion of toxicokinetic studies in the biological evaluation of medical devices.
- Standard26 pagesEnglish language
sale 10% off- e-Library read for1 day
ISO 10993-4:2017 specifies general requirements for evaluating the interactions of medical devices with blood.
It describes
a) a classification of medical devices that are intended for use in contact with blood, based on the intended use and duration of contact as defined in ISO 10993‑1,
b) the fundamental principles governing the evaluation of the interaction of devices with blood,
c) the rationale for structured selection of tests according to specific categories, together with the princip...view more
- Standard83 pagesEnglish language
sale 10% off- e-Library read for1 day
IEC 82304-1:2016 applies to the safety and security of health software products designed to operate on general computing platforms and intended to be placed on the market without dedicated hardware, and its primary focus is on the requirements for manufacturers. It covers the entire lifecycle including design, development, validation, installation, maintenance, and disposal of health software products.
- Standard30 pagesEnglish language
sale 10% off- e-Library read for1 day
This document specifies requirements for test organisms, suspensions, inoculated carriers, biological
indicators and test methods intended for use in assessing the performance of sterilization processes
employing moist heat as the sterilizing agent.
NOTE 1 Requirements for validation and control of moist heat sterilization processes are provided by the
ISO 17665 series.
NOTE 2 National or regional regulations can provide requirements for work place safety.
- Standard16 pagesEnglish language
sale 10% off- e-Library read for1 day
This document specifies general requirements for production, labelling, test methods and performance
characteristics of biological indicators, including inoculated carriers and suspensions, and their
components, to be used in the validation and routine monitoring of sterilization processes.
This document specifies basic and common requirements that are applicable to all parts of ISO 11138.
Requirements for biological indicators for particular specified processes are provided in the relevant
...view more
- Standard50 pagesEnglish language
sale 10% off- e-Library read for1 day
ISO 11138-2:2017 specifies requirements for test organisms, suspensions, inoculated carriers, biological indicators and test methods intended for use in assessing the performance of sterilizers and sterilization processes employing ethylene oxide gas as the sterilizing agent, either as pure ethylene oxide gas or mixtures of this gas with diluent gases, at sterilizing temperatures within the range of 29 °C to 65 °C.
NOTE 1 Requirements for validation and control of ethylene oxide sterilization p...view more
- Standard15 pagesEnglish language
sale 10% off- e-Library read for1 day
ISO 11138-3:2017 specifies requirements for test organisms, suspensions, inoculated carriers, biological indicators and test methods intended for use in assessing the performance of sterilization processes employing moist heat as the sterilizing agent.
NOTE 1 Requirements for validation and control of moist heat sterilization processes are provided by the ISO 17665 series.
NOTE 2 National or regional regulations can provide requirements for work place safety.
- Standard16 pagesEnglish language
sale 10% off- e-Library read for1 day
ISO 11138-1:2017 specifies general requirements for production, labelling, test methods and performance characteristics of biological indicators, including inoculated carriers and suspensions, and their components, to be used in the validation and routine monitoring of sterilization processes.
ISO 11138-1:2017 specifies basic and common requirements that are applicable to all parts of ISO 11138. Requirements for biological indicators for particular specified processes are provided in the releva...view more
- Standard50 pagesEnglish language
sale 10% off- e-Library read for1 day
ISO 11138-2:2017 specifies requirements for test organisms, suspensions, inoculated carriers, biological indicators and test methods intended for use in assessing the performance of sterilizers and sterilization processes employing ethylene oxide gas as the sterilizing agent, either as pure ethylene oxide gas or mixtures of this gas with diluent gases, at sterilizing temperatures within the range of 29 °C to 65 °C.
NOTE 1 Requirements for validation and control of ethylene oxide sterilization p...view more
- Standard15 pagesEnglish language
sale 10% off- e-Library read for1 day
This document identifies requirements for symbols used in medical device labelling that convey
information on the safe and effective use of medical devices. It also lists symbols that satisfy the
requirements of this document.
This document is applicable to symbols used in a broad spectrum of medical devices, which are
marketed globally and therefore need to meet different regulatory requirements.
These symbols may be used on the medical device itself, on its packaging or in the associated
...view more
- Standard41 pagesEnglish language
sale 10% off- e-Library read for1 day
This part of ISO 10993 specifies test methods for the assessment of the local effects after implantation
of biomaterials intended for use in medical devices.
This part of ISO 10993 applies to materials that are
— solid and non-absorbable,
— non-solid, such as porous materials, liquids, gels, pastes, and particulates, and
— degradable and/or absorbable, which may be solid or non-solid.
The test sample is implanted into a site and animal species appropriate for the evaluation of the
biologi...view more
- Standard42 pagesEnglish language
sale 10% off- e-Library read for1 day
ISO 10993-6:2016 specifies test methods for the assessment of the local effects after implantation of biomaterials intended for use in medical devices.
ISO 10993-6:2016 applies to materials that are
- solid and non-absorbable,
- non-solid, such as porous materials, liquids, gels, pastes, and particulates, and
- degradable and/or absorbable, which may be solid or non-solid.
The test sample is implanted into a site and animal species appropriate for the evaluation of the biological safety of ...view more
- Standard42 pagesEnglish language
sale 10% off- e-Library read for1 day
ISO 15223-1:2016 identifies requirements for symbols used in medical device labelling that convey information on the safe and effective use of medical devices. It also lists symbols that satisfy the requirements of this document.
ISO 15223-1:2016 is applicable to symbols used in a broad spectrum of medical devices, which are marketed globally and therefore need to meet different regulatory requirements.
These symbols may be used on the medical device itself, on its packaging or in the associat...view more
- Standard41 pagesEnglish language
sale 10% off- e-Library read for1 day
EN ISO 13485 specifies requirements for a quality management system where an organization needs to demonstrate its ability to provide medical devices and related services that consistently meet customer and applicable regulatory requirements. Such organizations can be involved in one or more stages of the life-cycle, including design and development, production, storage and distribution, installation, or servicing of a medical device and design and development or provision of associated activiti...view more
- Standard68 pagesEnglish language
sale 10% off- e-Library read for1 day
- Standard – translation131 pagesSlovenian and English language
sale 10% off- e-Library read for1 day
- Standard78 pagesGerman language
sale 10% off- e-Library read for1 day
ISO 13485:2016 specifies requirements for a quality management system where an organization needs to demonstrate its ability to provide medical devices and related services that consistently meet customer and applicable regulatory requirements. Such organizations can be involved in one or more stages of the life-cycle, including design and development, production, storage and distribution, installation, or servicing of a medical device and design and development or provision of associated activi...view more
- Standard68 pagesEnglish language
sale 10% off- e-Library read for1 day
- Standard – translation131 pagesSlovenian and English language
sale 10% off- e-Library read for1 day
- Standard78 pagesGerman language
sale 10% off- e-Library read for1 day
PWI created for possible future // procedures
- Amendment45 pagesEnglish language
sale 10% off- e-Library read for1 day
This European Standard specifies the requirements for an aseptically processed medical device to be designated 'STERILE'.
NOTE For the purpose of the EU Directive(s) for medical devices (see Bibliography), designating that a medical device is 'STERILE' is permissible when a validated manufacturing and sterilization process has been applied. Requirements for validation and routine control of aseptic processes are specified in EN ISO 13408-1. Specific requirements for the aseptic processing of s...view more
- Standard16 pagesEnglish language
sale 10% off- e-Library read for1 day
PWI created for possible future // procedures
- Amendment45 pagesEnglish language
sale 10% off- e-Library read for1 day
This part of ISO 13408 specifies requirements and provides guidance on alternative approaches to process
simulations for the qualification of the aseptic processing of medical devices and combination products that cannot
be terminally sterilized and where the process simulation approach according to ISO 13408-1 cannot be applied.
This part of ISO 13408 describes how risk assessment can be used during the development of an aseptic
process to design a process simulation study for medical devices a...view more
- Standard30 pagesEnglish language
sale 10% off- e-Library read for1 day
This part of EN 45502 specifies requirements that are generally applicable to ACTIVE IMPLANTABLE MEDICAL
DEVICES.
NOTE 1 For particular types of ACTIVE IMPLANTABLE MEDICAL DEVICES, these general requirements are supplemented or
modified by the requirements of particular standards which form additional parts of this European Standard.
The tests that are specified in EN 45502 are type tests and are to be carried out on samples of an ACTIVE
IMPLANTABLE MEDICAL DEVICE to show compliance.
This ...view more
- Standard60 pagesEnglish language
sale 10% off- e-Library read for1 day
This European Standard specifies the requirements for an aseptically processed medical device to be designated 'STERILE'.
NOTE For the purpose of the EU Directive(s) for medical devices (see Bibliography), designating that a medical device is 'STERILE' is permissible when a validated manufacturing and sterilization process has been applied. Requirements for validation and routine control of aseptic processes are specified in EN ISO 13408 1. Specific requirements for the aseptic processing of so...view more
- Standard16 pagesEnglish language
sale 10% off- e-Library read for1 day
2013-10-02 - BT146_DG9211 extended DOW to 2018-12-31
2013-08-13 - Negative assessment from Consultant
- Amendment10 pagesEnglish language
sale 10% off- e-Library read for1 day
ISO 11137-1:2006 specifies requirements for the development, validation and routine control of a radiation sterilization process for medical devices. Although the scope of ISO 11137-1:2006 is limited to medical devices, it specifies requirements and provides guidance that may be applicable to other products and equipment.
ISO 11137-1:2006 covers radiation processes employing irradiators using the radionuclide 60Co or 137Cs, a beam from an electron generator or a beam from an X-ray generator.
I...view more
- Standard58 pagesEnglish language
sale 10% off- e-Library read for1 day