1.1       This document specifies requirements for the development, validation and routine control of a radiation sterilization process for medical devices.
NOTE         Although the scope is limited to medical devices, this document can be applicable to other products and equipment.
This document covers radiation processes employing irradiators using:
a)      the radionuclide 60Co or 137Cs;
b)      a beam from an electron generator; or
c)       a beam from an X-ray generator.
1.2       This document is not applicable to processes for inactivating viruses or the causative agents of spongiform encephalopathies, such as scrapie, bovine spongiform encephalopathy and Creutzfeldt-Jakob disease.
NOTE         For information on such processes, see ISO 22442-1, ISO 22442-2, ISO 22442-3, ISO 13022 and ICH Q5A.
1.2.1       This document does not specify requirements for designating a medical device as sterile.
NOTE         Regional and national requirements can designate medical devices as sterile. See, for example, EN 556-1 or ANSI/AAMI ST67.
1.2.2       This document does not specify a quality management system for the control of all stages of production of medical devices.
NOTE         It is not a requirement of this document to have a complete quality management system during manufacture, but the elements of a quality management system that are the minimum necessary to control the sterilization process are normatively referenced at appropriate places in the text (see, in particular, Clause 4). Attention is drawn to the standards for quality management systems (see ISO 13485) that control all stages of production of medical devices, including the sterilization process. Regional and national regulations for the provision of medical devices can require implementation of a complete quality management system and the assessment of that system by a third party.
1.2.3       This document does not require that biological indicators be used for validation or monitoring of radiation sterilization, nor does it require that a pharmacopoeial test for sterility be carried out for product release.
1.2.4       This document does not specify requirements for occupational safety associated with the design and operation of irradiation facilities.
NOTE         Regulations on safety requirements for occupational safety related to radiation can exist in some countries.
1.2.5       This document does not specify requirements for the sterilization of used or reprocessed devices.

  • Standard
    61 pages
    English language
    e-Library read for
    1 day

1.1       This document specifies requirements for the development, validation and routine control of a radiation sterilization process for medical devices.
NOTE         Although the scope is limited to medical devices, this document can be applicable to other products and equipment.
This document covers radiation processes employing irradiators using:
a)      the radionuclide 60Co or 137Cs;
b)      a beam from an electron generator; or
c)       a beam from an X-ray generator.
1.2       This document is not applicable to processes for inactivating viruses or the causative agents of spongiform encephalopathies, such as scrapie, bovine spongiform encephalopathy and Creutzfeldt-Jakob disease.
NOTE         For information on such processes, see ISO 22442-1, ISO 22442-2, ISO 22442-3, ISO 13022 and ICH Q5A.
1.2.1       This document does not specify requirements for designating a medical device as sterile.
NOTE         Regional and national requirements can designate medical devices as sterile. See, for example, EN 556-1 or ANSI/AAMI ST67.
1.2.2       This document does not specify a quality management system for the control of all stages of production of medical devices.
NOTE         It is not a requirement of this document to have a complete quality management system during manufacture, but the elements of a quality management system that are the minimum necessary to control the sterilization process are normatively referenced at appropriate places in the text (see, in particular, Clause 4). Attention is drawn to the standards for quality management systems (see ISO 13485) that control all stages of production of medical devices, including the sterilization process. Regional and national regulations for the provision of medical devices can require implementation of a complete quality management system and the assessment of that system by a third party.
1.2.3       This document does not require that biological indicators be used for validation or monitoring of radiation sterilization, nor does it require that a pharmacopoeial test for sterility be carried out for product release.
1.2.4       This document does not specify requirements for occupational safety associated with the design and operation of irradiation facilities.
NOTE         Regulations on safety requirements for occupational safety related to radiation can exist in some countries.
1.2.5       This document does not specify requirements for the sterilization of used or reprocessed devices.

  • Standard
    61 pages
    English language
    e-Library read for
    1 day

This document specifies a framework for the identification, and if necessary, quantification of constituents of a medical device, allowing the identification of biological hazards and the estimation and control of biological risks from material constituents, using a generally stepwise approach to the chemical characterization which can include one or more of the following:
—          the identification of its materials of construction (medical device configuration);
—          the characterization of the materials of construction via the identification and quantification of their chemical constituents (material composition);
—          the characterization of the medical device for chemical substances that were introduced during manufacturing (e.g. mould release agents, process contaminants, sterilization residues);
—          the estimation (using laboratory extraction conditions) of the potential of the medical device, or its materials of construction, to release chemical substances under clinical use conditions (extractables);
—          the measurement of chemical substances released from a medical device under its clinical conditions of use (leachables).
This document can also be used for chemical characterization (e.g. the identification and/or quantification) of degradation products. Information on other aspects of degradation assessment are covered in ISO 10993-9, ISO 10993-13, ISO 10993-14 and ISO 10993-15.
The ISO 10993 series is applicable when the material or medical device has direct or indirect body contact (see ISO 10993-1 for categorization by nature of body contact).
This document is intended for suppliers of materials and manufacturers of medical devices, to support a biological evaluation.

  • Standard
    81 pages
    English language
    e-Library read for
    1 day

ISO 14708-3:2017 is applicable to active implantable medical devices intended for electrical stimulation of the central or peripheral nervous system.
The tests that are specified in this document are type tests and are to be carried out on a sample of a device to assess device behavioural responses, and are not intended to be used for the routine testing of manufactured products.

  • Standard
    63 pages
    English language
    e-Library read for
    1 day

This document specifies the procedure for the assessment of medical devices and their constituent materials with regard to their potential to induce skin sensitization.
This document includes:
—    details of in vivo skin sensitization test procedures;
—    key factors for the interpretation of the results.
NOTE      Instructions for the preparation of materials specifically in relation to the above tests are given in Annex A.

  • Standard
    60 pages
    English language
    e-Library read for
    1 day

This document specifies the procedure for the assessment of medical devices and their constituent materials with regard to their potential to induce skin sensitization.
This document includes:
—    details of in vivo skin sensitization test procedures;
—    key factors for the interpretation of the results.
NOTE      Instructions for the preparation of materials specifically in relation to the above tests are given in Annex A.

  • Standard
    60 pages
    English language
    e-Library read for
    1 day

ISO 14708-3:2017 is applicable to active implantable medical devices intended for electrical stimulation of the central or peripheral nervous system.
The tests that are specified in this document are type tests and are to be carried out on a sample of a device to assess device behavioural responses, and are not intended to be used for the routine testing of manufactured products.

  • Standard
    63 pages
    English language
    e-Library read for
    1 day

This document specifies requirements for safety and performance of active implantable circulatory support devices, including type tests, animal studies and clinical evaluation requirements.
NOTE    The device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, but there is a need to specify main requirements of non-implantable parts and accessories if they could affect the safety or performance of the implantable device.
The tests that are specified in this document are type tests and are to be carried out on a sample of a device to assess device behavioural responses and are not intended to be used for the routine testing of manufactured products.
Included in the scope of this document are:
—     ventricular assist devices (VAD), left or right heart support;
—     total artificial hearts (TAH);
—     biventricular assist devices (biVAD);
—     percutaneous assist devices;
—     paediatric assist devices.

  • Standard
    79 pages
    English language
    e-Library read for
    1 day

This document specifies requirements that are applicable to implantable cardioverter defibrillators and CRT-Ds and the functions of active implantable medical devices intended to treat tachyarrhythmia.
The tests that are specified in ISO 14708 are type tests and are to be carried out on samples of a device to show compliance.
This document was designed for tachyarrhythmia pulse generators used with either endocardial leads or epicardial leads. At the time of this edition, the authors recognized the emergence of technologies that do not use endocardial or epicardial leads for which adaptations of this part will be required. Such adaptations are left to the discretion of manufacturers incorporating these technologies.
This document is also applicable to some non-implantable parts and accessories of the devices (see Note 1).
The characteristics of the implantable pulse generator or lead shall be determined by either the appropriate method detailed in this document or by any other method demonstrated to have accuracy equal to, or better than, the method specified. In the case of dispute, the method detailed in this document shall apply.
Any aspect of an active implantable medical device intended to treat bradyarrhythmias or cardiac resynchronization is covered by ISO 14708-2.
NOTE 1    The device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, but there is a need to specify some requirements of non-implantable parts and accessories if they could affect the safety or performance of the implantable device.
NOTE 2    In this document, terms printed in italics are used as defined in Clause 3.  Where a defined term is used as a qualifier in another term, it is not printed in italics unless the concept thus qualified is also defined.

  • Standard
    75 pages
    English language
    e-Library read for
    1 day

This document specifies requirements that are applicable to those active implantable medical devices intended to treat bradyarrhythmias and devices that provide therapies for cardiac resynchronization.
The tests that are specified in this document are type tests, and are to be carried out on samples of a device to show compliance.
This document was designed for bradyarrhythmia pulse generators used with endocardial leads or epicardial leads. At the time of this edition, the authors recognized the emergence of leadless technologies for which adaptations of this part will be required. Such adaptations are left to the discretion of manufacturers incorporating these technologies.
This document is also applicable to some non-implantable parts and accessories of the devices (see Note 1).
The electrical characteristics of the implantable pulse generator or lead are determined either by the appropriate method detailed in this particular standard or by any other method demonstrated to have an accuracy equal to, or better than, the method specified. In case of dispute, the method detailed in this particular standard applies.
Any features of an active implantable medical device intended to treat tachyarrhythmias are covered by ISO 14708-6.
NOTE 1    The device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, but there is a need to specify some requirements of non-implantable parts and accessories if they could affect the safety or performance of the implantable device.
NOTE 2    In this document, terms printed in italics are used as defined in Clause 3. Where a defined term is used as a qualifier in another term, it is not printed in italics unless the concept thus qualified is also defined.

  • Standard
    80 pages
    English language
    e-Library read for
    1 day

This document specifies particular requirements for active implantable medical devices intended to deliver a medicinal substance to site-specific locations within the human body, to provide basic assurance of safety for both patients and users. It amends and supplements ISO 14708-1:2014. The requirements of this document take priority over those of ISO 14708-1.
This document is applicable to active implantable medical devices intended to deliver medicinal substances to site-specific locations within the human body.
This document is also applicable to some non-implantable parts and accessories of the devices defined in Clause 3.
The tests that are specified in this document are type tests intended to be carried out on a sample of a device to show compliance and are not intended to be used for the routine testing of manufactured products.
NOTE       This document is not intended to apply to non-implantable infusion systems.

  • Standard
    68 pages
    English language
    e-Library read for
    1 day

This document specifies requirements that are applicable to those active implantable medical devices that are intended to treat hearing impairment via electrical stimulation of the auditory pathways. Devices which treat hearing impairment via means other than electrical stimulation are not covered by this document.
The tests that are specified in this document are type tests and are to be carried out on samples of a device to show compliance.
This document is also applicable to non-implantable parts and accessories of the devices (see NOTE).
The electrical characteristics of the implantable part are determined by either the appropriate method detailed in this document or by any other method demonstrated to have an accuracy equal to, or better than, the method specified. In the case of dispute, the method detailed in this document applies.
NOTE       A device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, this document specifies those requirements of non-implantable parts and accessories which could affect the safety or performance of the implantable part.

  • Standard
    77 pages
    English language
    e-Library read for
    1 day

This document specifies requirements and test methods for materials, preformed sterile barrier systems, sterile barrier systems and packaging systems that are intended to maintain sterility of terminally sterilized medical devices until the point of use.
It is applicable to industry, to health care facilities, and to wherever medical devices are placed in sterile barrier systems and sterilized.
It does not cover all requirements for sterile barrier systems and packaging systems for medical devices that are manufactured aseptically. Additional requirements can be necessary for drug/device combinations.
It does not describe a quality assurance system for control of all stages of manufacture.
It does not apply to packaging materials and/or systems used to contain a contaminated medical device during transportation of the item to the site of reprocessing or disposal.

  • Amendment
    11 pages
    English language
    e-Library read for
    1 day

This document specifies requirements for the development and validation of processes for packaging medical devices that are terminally sterilized. These processes include forming, sealing and assembly of preformed sterile barrier systems, sterile barrier systems and packaging systems.
It is applicable to industry, to health care facilities, and to wherever medical devices are packaged and sterilized.
It does not cover all requirements for packaging medical devices that are manufactured aseptically. Additional requirements can be necessary for drug/device combinations.

  • Amendment
    11 pages
    English language
    e-Library read for
    1 day

This document specifies particular requirements for active implantable medical devices intended to deliver a medicinal substance to site-specific locations within the human body, to provide basic assurance of safety for both patients and users. It amends and supplements ISO 14708-1:2014. The requirements of this document take priority over those of ISO 14708-1.
This document is applicable to active implantable medical devices intended to deliver medicinal substances to site-specific locations within the human body.
This document is also applicable to some non-implantable parts and accessories of the devices defined in Clause 3.
The tests that are specified in this document are type tests intended to be carried out on a sample of a device to show compliance and are not intended to be used for the routine testing of manufactured products.
NOTE       This document is not intended to apply to non-implantable infusion systems.

  • Standard
    68 pages
    English language
    e-Library read for
    1 day

This document specifies requirements for safety and performance of active implantable circulatory support devices, including type tests, animal studies and clinical evaluation requirements.
NOTE    The device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, but there is a need to specify main requirements of non-implantable parts and accessories if they could affect the safety or performance of the implantable device.
The tests that are specified in this document are type tests and are to be carried out on a sample of a device to assess device behavioural responses and are not intended to be used for the routine testing of manufactured products.
Included in the scope of this document are:
—     ventricular assist devices (VAD), left or right heart support;
—     total artificial hearts (TAH);
—     biventricular assist devices (biVAD);
—     percutaneous assist devices;
—     paediatric assist devices.

  • Standard
    79 pages
    English language
    e-Library read for
    1 day

This document specifies requirements that are applicable to implantable cardioverter defibrillators and CRT-Ds and the functions of active implantable medical devices intended to treat tachyarrhythmia.
The tests that are specified in ISO 14708 are type tests and are to be carried out on samples of a device to show compliance.
This document was designed for tachyarrhythmia pulse generators used with either endocardial leads or epicardial leads. At the time of this edition, the authors recognized the emergence of technologies that do not use endocardial or epicardial leads for which adaptations of this part will be required. Such adaptations are left to the discretion of manufacturers incorporating these technologies.
This document is also applicable to some non-implantable parts and accessories of the devices (see Note 1).
The characteristics of the implantable pulse generator or lead shall be determined by either the appropriate method detailed in this document or by any other method demonstrated to have accuracy equal to, or better than, the method specified. In the case of dispute, the method detailed in this document shall apply.
Any aspect of an active implantable medical device intended to treat bradyarrhythmias or cardiac resynchronization is covered by ISO 14708-2.
NOTE 1    The device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, but there is a need to specify some requirements of non-implantable parts and accessories if they could affect the safety or performance of the implantable device.
NOTE 2    In this document, terms printed in italics are used as defined in Clause 3.  Where a defined term is used as a qualifier in another term, it is not printed in italics unless the concept thus qualified is also defined.

  • Standard
    75 pages
    English language
    e-Library read for
    1 day

This document specifies requirements that are applicable to those active implantable medical devices intended to treat bradyarrhythmias and devices that provide therapies for cardiac resynchronization.
The tests that are specified in this document are type tests, and are to be carried out on samples of a device to show compliance.
This document was designed for bradyarrhythmia pulse generators used with endocardial leads or epicardial leads. At the time of this edition, the authors recognized the emergence of leadless technologies for which adaptations of this part will be required. Such adaptations are left to the discretion of manufacturers incorporating these technologies.
This document is also applicable to some non-implantable parts and accessories of the devices (see Note 1).
The electrical characteristics of the implantable pulse generator or lead are determined either by the appropriate method detailed in this particular standard or by any other method demonstrated to have an accuracy equal to, or better than, the method specified. In case of dispute, the method detailed in this particular standard applies.
Any features of an active implantable medical device intended to treat tachyarrhythmias are covered by ISO 14708-6.
NOTE 1    The device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, but there is a need to specify some requirements of non-implantable parts and accessories if they could affect the safety or performance of the implantable device.
NOTE 2    In this document, terms printed in italics are used as defined in Clause 3. Where a defined term is used as a qualifier in another term, it is not printed in italics unless the concept thus qualified is also defined.

  • Standard
    80 pages
    English language
    e-Library read for
    1 day

This document specifies requirements that are applicable to those active implantable medical devices that are intended to treat hearing impairment via electrical stimulation of the auditory pathways. Devices which treat hearing impairment via means other than electrical stimulation are not covered by this document.
The tests that are specified in this document are type tests and are to be carried out on samples of a device to show compliance.
This document is also applicable to non-implantable parts and accessories of the devices (see NOTE).
The electrical characteristics of the implantable part are determined by either the appropriate method detailed in this document or by any other method demonstrated to have an accuracy equal to, or better than, the method specified. In the case of dispute, the method detailed in this document applies.
NOTE       A device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, this document specifies those requirements of non-implantable parts and accessories which could affect the safety or performance of the implantable part.

  • Standard
    77 pages
    English language
    e-Library read for
    1 day

This document specifies requirements for the development and validation of processes for packaging medical devices that are terminally sterilized. These processes include forming, sealing and assembly of preformed sterile barrier systems, sterile barrier systems and packaging systems.
It is applicable to industry, to health care facilities, and to wherever medical devices are packaged and sterilized.
It does not cover all requirements for packaging medical devices that are manufactured aseptically. Additional requirements can be necessary for drug/device combinations.

  • Amendment
    11 pages
    English language
    e-Library read for
    1 day

This document specifies requirements and test methods for materials, preformed sterile barrier systems, sterile barrier systems and packaging systems that are intended to maintain sterility of terminally sterilized medical devices until the point of use.
It is applicable to industry, to health care facilities, and to wherever medical devices are placed in sterile barrier systems and sterilized.
It does not cover all requirements for sterile barrier systems and packaging systems for medical devices that are manufactured aseptically. Additional requirements can be necessary for drug/device combinations.
It does not describe a quality assurance system for control of all stages of manufacture.
It does not apply to packaging materials and/or systems used to contain a contaminated medical device during transportation of the item to the site of reprocessing or disposal.

  • Amendment
    11 pages
    English language
    e-Library read for
    1 day

This document provides general principles for the systematic evaluation of the potential and observed degradation of medical devices through the design and performance of in vitro degradation studies. Information obtained from these studies can be used in the biological evaluation described in the ISO 10993 series.
This document is applicable to both materials designed to degrade in the body as well as materials that are not intended to degrade.
This document is not applicable to:
a)   the evaluation of degradation which occurs by purely mechanical processes; methodologies for the production of this type of degradation product are described in specific product standards, where available;
NOTE    Purely mechanical degradation causes mostly particulate matter. Although this is excluded from the scope of this document, such degradation products can evoke a biological response and can undergo biological evaluation as described in other parts of ISO 10993.
b)   leachable components which are not degradation products;
c)   medical devices or components that do not contact the patient's body directly or indirectly.

  • Standard
    22 pages
    English language
    e-Library read for
    1 day

This document specifies requirements for the characterization of a liquid chemical sterilizing agent and for the development, validation, process control and monitoring of sterilization by liquid chemical sterilizing agents of single-use medical devices comprising, in whole or in part, materials of animal origin.
This document covers the control of risks arising from contamination with bacteria and fungi by application of a liquid chemical sterilization process. Risks associated with other microorganisms can be assessed using other methods (see NOTE 1).
This document is not applicable to material of human origin.
This document does not describe methods for the validation of the inactivation of viruses and transmissible spongiform encephalopathy (TSE) agents (see NOTE 2 and NOTE 3).
This document does not describe methods for validation of the inactivation or elimination of protozoa and parasites.
The requirements for validation and routine control described in this document are only applicable to the defined sterilization process of a medical device, which is performed after the manufacturing process, and do not take account of the lethal effects of other bioburden reduction steps (see NOTE 4).
This document does not specify tests to establish the effects of any chosen sterilization process upon the fitness for use of the medical device (see NOTE 5).
This document does not cover the level of residual sterilizing agent within medical devices (see NOTE 6).
Guidance for the characterization of a liquid chemical sterilizing agent and for the development, validation, process control and monitoring of sterilization by liquid chemical sterilizing agents of single-use medical devices comprising, in whole or in part, materials of animal origin is provided in informative Annex A.
NOTE 1  The prior application of risk management principles to medical devices utilizing animal tissues, as described in ISO 22442-1 is important. ISO 18362 provides information on control of microbial risks during processing of cell-based health-care products.
NOTE 2  Liquid chemical sterilizing agents traditionally employed to sterilize animal tissues in medical devices might not be effective in inactivating the causative agents of TSE such as bovine spongiform encephalopathy (BSE), or scrapie. Satisfactory validation in accordance with this document does not necessarily demonstrate inactivation of infective agents of this type. Risk controls related to sourcing, collection and handling of animal materials are described in ISO 22442-2.
NOTE 3  The validation of the inactivation, elimination, or elimination and inactivation of viruses and TSE agents is described in ISO 22442-3.
NOTE 4  Manufacturing processes for medical devices containing animal tissues frequently include exposure to chemical agents which can significantly reduce the bioburden on the medical device. Following the manufacturing process, a medical device is exposed to a specified sterilization process.
NOTE 5  Such testing is a crucial part of the design and development of a medical device.
NOTE 6  ISO 10993-17 specifies a method to establish allowable limits for residues of sterilizing agents.
NOTE 7  Standards for quality management systems (see ISO 13485) can be used in the control of all stages of manufacture including the sterilization process.

  • Standard
    50 pages
    English language
    e-Library read for
    1 day

This document specifies requirements and gives guidance on the procedures in the preparation of samples and the selection of reference materials for medical device testing primarily in biological test systems primarily in accordance with one or more parts of the ISO 10993 series.
Specifically, this document addresses the following:
—     test sample selection;
—     selection of representative portions from a medical device;
—     test sample preparation;
—     experimental controls;
—     selection of, and requirements for, reference materials;
—     preparation of extracts.
This document is not applicable to live cells but can be relevant to the material or medical device components of combination products containing live cells.
Extractions for chemical characterization are covered in ISO 10993-18. Clause 7, 8, 9, 10 [with the exception of 10.3.5 and 10.3.11 b)], and 11 can apply to extractions for chemical characterization. Information given in C.1 to C.4 can also be relevant.

  • Standard
    32 pages
    English language
    e-Library read for
    1 day

This document provides general principles for the systematic evaluation of the potential and observed degradation of medical devices through the design and performance of in vitro degradation studies. Information obtained from these studies can be used in the biological evaluation described in the ISO 10993 series.
This document is applicable to both materials designed to degrade in the body as well as materials that are not intended to degrade.
This document is not applicable to:
a)   the evaluation of degradation which occurs by purely mechanical processes; methodologies for the production of this type of degradation product are described in specific product standards, where available;
NOTE    Purely mechanical degradation causes mostly particulate matter. Although this is excluded from the scope of this document, such degradation products can evoke a biological response and can undergo biological evaluation as described in other parts of ISO 10993.
b)   leachable components which are not degradation products;
c)   medical devices or components that do not contact the patient's body directly or indirectly.

  • Standard
    22 pages
    English language
    e-Library read for
    1 day

This document specifies the requirements for and provides guidance on the specification, selection, qualification, bio-decontamination, validation, operation and control of isolator systems related to aseptic processing of health care products and processing of cell-based health care products.
This document does not specify requirements for restricted access barrier systems (RABS).
This document does not supersede or replace national regulatory requirements such as Good Manufacturing Practices (GMPs) and/or compendia requirements that pertain in particular to national or regional jurisdictions.
This document does not specify requirements for isolators used for sterility testing; however, some of the principles and information in this document could be applicable to this application.
This document does not define biosafety containment requirements.

  • Standard
    39 pages
    English language
    e-Library read for
    1 day

2021-08-25-JO: BT Decision (BT169/DG12292/DV) to remove the link to the MDR
2021-06-24: blocked so that the link to the MDD can be removed by BT from the document
2019-12-03-JO-Assessment at ENQ uplaoded with due date of 06 Jan.2020 plus extention requested by HAS Consultant

  • Amendment
    58 pages
    English language
    e-Library read for
    1 day

2021-08-25-JO: BT Decision (BT169/DG12292/DV) to remove the link to the MDR
2021-06-24: blocked so that the link to the MDD can be removed by BT from the document
2019-12-03-JO-Assessment at ENQ uplaoded with due date of 06 Jan.2020 plus extention requested by HAS Consultant

  • Amendment
    58 pages
    English language
    e-Library read for
    1 day

This document specifies the procedure for the assessment of medical devices and their constituent materials with regard to their potential to produce irritation. The tests are designed to predict and classify the irritation potential of medical devices, materials or their extracts according to ISO 10993‑1 and ISO 10993‑2.
This document includes:
—     pre-test considerations for irritation, including in silico and in vitro methods for dermal exposure;
—     details of in vitro and in vivo irritation test procedures;
—     key factors for the interpretation of the results.

  • Standard
    76 pages
    English language
    e-Library read for
    1 day

This document specifies requirements and gives guidance on the procedures in the preparation of samples and the selection of reference materials for medical device testing primarily in biological test systems primarily in accordance with one or more parts of the ISO 10993 series.
Specifically, this document addresses the following:
—     test sample selection;
—     selection of representative portions from a medical device;
—     test sample preparation;
—     experimental controls;
—     selection of, and requirements for, reference materials;
—     preparation of extracts.
This document is not applicable to live cells but can be relevant to the material or medical device components of combination products containing live cells.
Extractions for chemical characterization are covered in ISO 10993-18. Clause 7, 8, 9, 10 [with the exception of 10.3.5 and 10.3.11 b)], and 11 can apply to extractions for chemical characterization. Information given in C.1 to C.4 can also be relevant.

  • Standard
    32 pages
    English language
    e-Library read for
    1 day

2020-11-16_JO_BT167/DG11868/DV to remove the link to (MDD 93/42/EEC) , 90/385/EEC (AIMD) and M/295

  • Amendment
    12 pages
    English language
    e-Library read for
    1 day

This document specifies the requirements for and provides guidance on the specification, selection, qualification, bio-decontamination, validation, operation and control of isolator systems related to aseptic processing of health care products and processing of cell-based health care products.
This document does not specify requirements for restricted access barrier systems (RABS).
This document does not supersede or replace national regulatory requirements such as Good Manufacturing Practices (GMPs) and/or compendia requirements that pertain in particular to national or regional jurisdictions.
This document does not specify requirements for isolators used for sterility testing; however, some of the principles and information in this document could be applicable to this application.
This document does not define biosafety containment requirements.

  • Standard
    39 pages
    English language
    e-Library read for
    1 day

This document specifies the procedure for the assessment of medical devices and their constituent materials with regard to their potential to produce irritation. The tests are designed to predict and classify the irritation potential of medical devices, materials or their extracts according to ISO 10993‑1 and ISO 10993‑2.
This document includes:
—     pre-test considerations for irritation, including in silico and in vitro methods for dermal exposure;
—     details of in vitro and in vivo irritation test procedures;
—     key factors for the interpretation of the results.

  • Standard
    76 pages
    English language
    e-Library read for
    1 day

This document specifies a framework for the identification, and if necessary, quantification of constituents of a medical device, allowing the identification of biological hazards and the estimation and control of biological risks from material constituents, using a generally stepwise approach to the chemical characterization which can include one or more of the following:
—          the identification of its materials of construction (medical device configuration);
—          the characterization of the materials of construction via the identification and quantification of their chemical constituents (material composition);
—          the characterization of the medical device for chemical substances that were introduced during manufacturing (e.g. mould release agents, process contaminants, sterilization residues);
—          the estimation (using laboratory extraction conditions) of the potential of the medical device, or its materials of construction, to release chemical substances under clinical use conditions (extractables);
—          the measurement of chemical substances released from a medical device under its clinical conditions of use (leachables).
This document can also be used for chemical characterization (e.g. the identification and/or quantification) of degradation products. Information on other aspects of degradation assessment are covered in ISO 10993-9, ISO 10993-13, ISO 10993-14 and ISO 10993-15.
The ISO 10993 series is applicable when the material or medical device has direct or indirect body contact (see ISO 10993-1 for categorization by nature of body contact).
This document is intended for suppliers of materials and manufacturers of medical devices, to support a biological evaluation.

  • Standard
    81 pages
    English language
    e-Library read for
    1 day

1.1    This document specifies the general criteria for tests of sterility on medical devices that have been exposed to a treatment with the sterilizing agent which has been reduced relative to that anticipated to be used in routine sterilization processing. These tests are intended to be performed when defining, validating or maintaining a sterilization process.
1.2    This document is not applicable to:
a)    sterility testing for routine release of product that has been subjected to a sterilization process,
b)    performing a test for sterility (see 3.12),
NOTE 1    The performance of a) or b) is not a requirement of ISO 11135, ISO 11137-1, ISO 11137-2, ISO 14160, ISO 14937, ISO 17665-1 or ISO 20857.
c)    test of sterility or test for sterility for demonstration of product shelf life, stability and/or package integrity, and
d)    culturing of biological indicators or inoculated products.
NOTE 2    Guidance on culturing biological indicators is included in ISO 11138-7.

  • Standard
    35 pages
    English language
    e-Library read for
    1 day

This document specifies requirements and test methods for materials, preformed sterile barrier systems, sterile barrier systems and packaging systems that are intended to maintain sterility of terminally sterilized medical devices until the point of use.
It is applicable to industry, to health care facilities, and to wherever medical devices are placed in sterile barrier systems and sterilized.
It does not cover all requirements for sterile barrier systems and packaging systems for medical devices that are manufactured aseptically. Additional requirements can be necessary for drug/device combinations.
It does not describe a quality assurance system for control of all stages of manufacture.
It does not apply to packaging materials and/or systems used to contain a contaminated medical device during transportation of the item to the site of reprocessing or disposal.

  • Standard
    52 pages
    English language
    e-Library read for
    1 day

This document specifies requirements for the development and validation of processes for packaging medical devices that are terminally sterilized. These processes include forming, sealing and assembly of preformed sterile barrier systems, sterile barrier systems and packaging systems.
It is applicable to industry, to health care facilities, and to wherever medical devices are packaged and sterilized.
It does not cover all requirements for packaging medical devices that are manufactured aseptically. Additional requirements can be necessary for drug/device combinations.

  • Standard
    21 pages
    English language
    e-Library read for
    1 day

This document specifies terminology, principles and a process for risk management of medical devices, including software as a medical device and in vitro diagnostic medical devices. The process described in this document intends to assist manufacturers of medical devices to identify the hazards associated with the medical device, to estimate and evaluate the associated risks, to control these risks, and to monitor the effectiveness of the controls.
The requirements of this document are applicable to all phases of the life cycle of a medical device. The process described in this document applies to risks associated with a medical device, such as risks related to biocompatibility, data and systems security, electricity, moving parts, radiation, and usability.
The process described in this document can also be applied to products that are not necessarily medical devices in some jurisdictions and can also be used by others involved in the medical device life cycle.
This document does not apply to:
—          decisions on the use of a medical device in the context of any particular clinical procedure; or
—          business risk management.
This document requires manufacturers to establish objective criteria for risk acceptability but does not specify acceptable risk levels.
Risk management can be an integral part of a quality management system. However, this document does not require the manufacturer to have a quality management system in place.
NOTE       Guidance on the application of this document can be found in ISO/TR 24971[9].

  • Standard
    46 pages
    English language
    e-Library read for
    1 day
  • Standard – translation
    44 pages
    Slovenian language
    e-Library read for
    1 day

Frequently Asked Questions

An EU Directive is a legislative act of the European Union that sets out goals that all EU member states must achieve. However, it is up to each member state to devise their own laws on how to reach these goals through national transposition. Directives are used to harmonize laws across the EU, particularly for the functioning of the single market.

Directive 90/385/EEC covers "Active implantable medical devices". There are 432 standards associated with this directive.

Harmonized standards under 90/385/EEC are European standards (ENs) developed by CEN, CENELEC, or ETSI in response to a mandate from the European Commission. When these standards are cited in the Official Journal of the European Union, products manufactured in conformity with them benefit from a presumption of conformity with the essential requirements of 90/385/EEC, facilitating CE marking and free movement within the European Economic Area.