FprEN ISO 21392
(Main)Cosmetics - Analytical methods - Measurement of traces of heavy metals in cosmetic finished products using ICP/MS technique (ISO/FDIS 21392:2021)
Cosmetics - Analytical methods - Measurement of traces of heavy metals in cosmetic finished products using ICP/MS technique (ISO/FDIS 21392:2021)
- Selection of the metallic elements of interest
- Selection or development of a reliable method for determination of these elements in cosmetic products which minimizes risk of interferences and allows the appropriate levels to be measured.
- Validation and characterization of the analytical method to assess its performance (global precision) at the appropriate level.
Kosmetische Mittel - Untersuchungsverfahren - Messung von Spuren von Schwermetallen in fertigen kosmetischen Mitteln mittels ICP-MS (ISO/FDIS 21392:2021)
In diesem Dokument wird ein Verfahren zur Quantifizierung von Spurenanteilen von Schwermetallen in kosmetischen Mitteln bereitgestellt.
Dieses Dokument bezieht sich nur auf Chrom, Cobalt, Nickel, Arsen, Cadmium, Antimon und Blei. Das Verfahren kann auf andere Elemente angewendet werden, es liegt jedoch in der Verantwortung des Analytikers nachzuweisen, dass es für diesen Zweck geeignet ist.
Cosmétiques - Méthodes d’analyse - Mesurage des éléments traces métalliques par ICP-MS dans les produits cosmétiques finis (ISO/FDIS 21392:2021)
Kozmetika - Analizne metode - Določevanje težkih kovin v sledovih v končnih kozmetičnih izdelkih z masno spektrometrijo z induktivno sklopljeno plazmo (ICP/MS) (ISO/FDIS 21392:2021)
General Information
Standards Content (sample)
SLOVENSKI STANDARD
oSIST prEN ISO 21392:2020
01-junij-2020
Kozmetika - Analizne metode - Določevanje sledov težkih kovin v končnih
kozmetičnih izdelkih z masno spektrometrijo z induktivno sklopljeno plazmo
(ICP/MS) (ISO/DIS 21392:2020)
Cosmetics - Analytical Methods - Measurement of traces of heavy metals in cosmetic
finished products using ICP/MS technique (ISO/DIS 21392:2020)Kosmetische Mittel - Untersuchungsverfahren - Messung von Spuren von
Schwermetallen in fertigen kosmetischen Mitteln mittels ICP-MS (ISO/DIS 21392:2020)
Cosmétiques - Méthodes analytiques - Détermination de traces de métaux lourds dans
les produits finis cosmétiques par ICP/MS (ISO/DIS 21392:2020)Ta slovenski standard je istoveten z: prEN ISO 21392
ICS:
71.100.70 Kozmetika. Toaletni Cosmetics. Toiletries
pripomočki
oSIST prEN ISO 21392:2020 en,fr,de
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
---------------------- Page: 1 ----------------------oSIST prEN ISO 21392:2020
---------------------- Page: 2 ----------------------
oSIST prEN ISO 21392:2020
DRAFT INTERNATIONAL STANDARD
ISO/DIS 21392
ISO/TC 217 Secretariat: ISIRI
Voting begins on: Voting terminates on:
2020-04-29 2020-07-22
Cosmetics — Analytical methods — Measurement of traces
of heavy metals in cosmetic finished products using ICP/
MS technique
Cosmétiques — Méthodes analytiques — Détermination de traces de métaux lourds dans les produits finis
cosmétiques par ICP/MSICS: 71.100.70
THIS DOCUMENT IS A DRAFT CIRCULATED
This document is circulated as received from the committee secretariat.
FOR COMMENT AND APPROVAL. IT IS
THEREFORE SUBJECT TO CHANGE AND MAY
NOT BE REFERRED TO AS AN INTERNATIONAL
STANDARD UNTIL PUBLISHED AS SUCH.
IN ADDITION TO THEIR EVALUATION AS
ISO/CEN PARALLEL PROCESSING
BEING ACCEPTABLE FOR INDUSTRIAL,
TECHNOLOGICAL, COMMERCIAL AND
USER PURPOSES, DRAFT INTERNATIONAL
STANDARDS MAY ON OCCASION HAVE TO
BE CONSIDERED IN THE LIGHT OF THEIR
POTENTIAL TO BECOME STANDARDS TO
WHICH REFERENCE MAY BE MADE IN
Reference number
NATIONAL REGULATIONS.
ISO/DIS 21392:2020(E)
RECIPIENTS OF THIS DRAFT ARE INVITED
TO SUBMIT, WITH THEIR COMMENTS,
NOTIFICATION OF ANY RELEVANT PATENT
RIGHTS OF WHICH THEY ARE AWARE AND TO
PROVIDE SUPPORTING DOCUMENTATION. ISO 2020
---------------------- Page: 3 ----------------------
oSIST prEN ISO 21392:2020
ISO/DIS 21392:2020(E)
COPYRIGHT PROTECTED DOCUMENT
© ISO 2020
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting
on the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address
below or ISO’s member body in the country of the requester.ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Fax: +41 22 749 09 47
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii © ISO 2020 – All rights reserved
---------------------- Page: 4 ----------------------
oSIST prEN ISO 21392:2020
ISO/DIS 21392:2020(E)
Contents Page
Foreword ........................................................................................................................................................................................................................................iv
Introduction ..................................................................................................................................................................................................................................v
1 Scope ................................................................................................................................................................................................................................. 1
2 Normative references ...................................................................................................................................................................................... 1
3 Principle ........................................................................................................................................................................................................................ 1
4 Reagents ........................................................................................................................................................................................................................ 1
5 Apparatus and equipment .......................................................................................................................................................................... 2
5.1 Digestion vessels ................................................................................................................................................................................... 2
5.2 Microwave-assisted digestion instruments ................................................................................................................... 2
5.3 Membrane filter, 0,45 µm pore size ....................................................................................................................................... 2
5.4 ICP-MS ............................................................................................................................................................................................................ 2
6 Preparation of standards solutions .................................................................................................................................................. 3
6.1 Diluted nitric acid, produced by mixing nitric acid (4.2) with pure water (4.1) at a
ratio of approximately 1+9 parts respectively. ............................................................................................................ 3
6.2 Diluting solution .................................................................................................................................................................................... 3
6.3 Internal Standard solutions. ........................................................................................................................................................ 3
6.3.1 Rhodium(*) standard solution, 1 mg/l ......................................................................................................... 3
6.3.2 Lutetium(*) standard solution, 1 mg/l ......................................................................................................... 3
6.4 Standard solutions ............................................................................................................................................................................... 4
6.4.1 High concentration mixed standard solution, 10 mg/l ................................................................... 4
6.4.2 Low concentration mixed standard solution, 0.1 mg/l .................................................................. 4
6.5 Calibration blank solution ............................................................................................................................................................. 4
6.6 Calibration solutions .......................................................................................................................................................................... 4
7 Procedure..................................................................................................................................................................................................................... 5
7.1 General ........................................................................................................................................................................................................... 5
7.2 Preparation of samples .................................................................................................................................................................... 5
7.3 Pressure assisted digestion .......................................................................................................................................................... 5
7.3.1 Preparation of Sample by Digestion – General case .......................................................................... 5
7.3.2 Preparation of Sample by Digestion – Specific cases ........................................................................ 6
7.3.3 Microwave digestion procedure ......................................................................................................................... 6
7.3.4 Preparation of measurement solutions ....................................................................................................... 6
7.4 Inductively Coupled Plasma Mass Spectrometry ...................................................................................................... 7
7.4.1 ICP-MS operating conditions ................................................................................................................................. 7
7.4.2 Quantification of the analytes by ICP-MS .................................................................................................... 7
7.5 Quality control of the analysis ................................................................................................................................................... 8
7.5.1 During the digestion ..................................................................................................................................................... 8
7.5.2 During the analysis ........................................................................................................................................................ 9
7.5.3 Example of ICP-MS sequence ................................................................................................................................ 9
8 Calculation ...............................................................................................................................................................................................................10
9 Method performance ....................................................................................................................................................................................10
Annex A Performance of the method determined by the accuracy profile methodology ........................12
Annex B (Informative) Information on the Evaluation of the method via ISO 5725 statistical
approach (*) ...........................................................................................................................................................................................................15
Bibliography .............................................................................................................................................................................................................................23
© ISO 2020 – All rights reserved iii---------------------- Page: 5 ----------------------
oSIST prEN ISO 21392:2020
ISO/DIS 21392:2020(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www .iso .org/ directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of
any patent rights identified during the development of the document will be in the Introduction and/or
on the ISO list of patent declarations received (see www .iso .org/ patents).Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and
expressions related to conformity assessment, as well as information about ISO's adherence to the
World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www .iso .org/
iso/ foreword .html.This document was prepared by Technical Committee ISO/TC 217 Cosmetics.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www .iso .org/ members .html.iv © ISO 2020 – All rights reserved
---------------------- Page: 6 ----------------------
oSIST prEN ISO 21392:2020
ISO/DIS 21392:2020(E)
Introduction
This standard specifies an analytical procedure for determination of trace elements (Chromium,
Cobalt, Nickel, Arsenic, Cadmium, Antimony and Lead) in finished cosmetics products by Inductively
Coupled Plasma Mass Spectrometry (ICP-MS) after pressure digestion of the sample. This type of
[1][2][3] [1][2][3]analytical procedure is widely described in other areas such as environment, food and
[7][8][9][10]pharmaceutical industry .
© ISO 2020 – All rights reserved v
---------------------- Page: 7 ----------------------
oSIST prEN ISO 21392:2020
---------------------- Page: 8 ----------------------
oSIST prEN ISO 21392:2020
DRAFT INTERNATIONAL STANDARD ISO/DIS 21392:2020(E)
Cosmetics — Analytical methods — Measurement of traces
of heavy metals in cosmetic finished products using ICP/
MS technique
1 Scope
The aim of this standard is to provide a method of quantification of heavy metal trace elements in
cosmetic products that consumers might be exposed to in their usage. In the sample preparation
procedure, nitric acid/hydrochloric acid mixture is used and most of the cosmetic ingredients are
digested allowing heavy metal trace elements to be solubilized for measurement. Some cosmetic
inorganic ingredients such as silica or titanium dioxide might not be completely digested under the
conditions of this standard and heavy metal trace elements confined in such ingredients might not
be fully extracted. However, the heavy metal trace elements trapped in these inorganic materials
are considered less relevant for the evaluation of the exposure level of consumers to unwanted trace
elements. The use of ICP-MS ensures reliable measurement of trace elements due to its proven high
sensitivity and selectivity.This analytical methodology can be applied to many other elements but this standard refers only to
the above listed trace elements and it is to the responsibility of the analyst to prove that it fits for
that purpose. In order to obtain comparable results, it is absolutely mandatory to comply with all the
conditions linked to the digestion of the samples.2 Normative references
There is no normative reference in this document.
3 Principle
Trace elements in cosmetic products are quantified by ICP-MS measurement of the solutions resulting
from digestion of the cosmetic products. Digestion takes place with mineral acids in sealed vessels
heated to 200ºC by microwaves, producing high pressures.4 Reagents
The reagents and the water used shall be free of the elements to be determined to such an extent
that the results are not impaired. Unless specified otherwise, solutions are understood to be aqueous
solutions.4.1 Ultrapure water, conductivity below 0,1µS/cm-1 at 25°C according to Type 1 water specifications
[11]defined in ISO 3696 standard
4.2 Nitric acid, minimum w = 60%, density = 1,38 g/ml
4.3 Hydrochloric acid, minimum w = 30 %, density = 1,15 g/ml
4.4 Internal standard stock solutions
For storage and stability conditions of the internal standard stock solutions, follow the specifications of
the suppliers.© ISO 2020 – All rights reserved 1
---------------------- Page: 9 ----------------------
oSIST prEN ISO 21392:2020
ISO/DIS 21392:2020(E)
4.4.1 Rhodium stock solution, 1 000 mg/l
4.4.2 Lutetium stock solution, 1 000 mg/l
4.5 Analytes stock solutions (Chromium, Cobalt, Nickel, Arsenic, Cadmium, Antimony and Lead),
1 000 mg/l for each elementCommercially available single element or mixed stock solutions can be used. For this 2 cases, the used
stock solutions shall not contain other elements that could interfere with the analytes to be quantified.
4.6 Commercially available ICP-MS Tune solution, containing e.g. Ce, Co, Li, Mg, Tl and Y (1 µg/l)
according to instrument manufacturer’s recommendations.5 Apparatus and equipment
All apparatus and equipment that come into direct contact with sample or solutions shall be pre-cleaned
with dilute nitric acid (6.1) and rinsed with ultrapure water (4.1) to ensure the lowest analyte background.
To prevent contamination and adsorption, do not use lab materials made with borosilicate glass.
5.1 Digestion vesselsUse commercially available, safety-tested pressure vessels and inserts made of acid-resistant, low-
contamination materials. The assembled vessels shall be able to safely withstand temperatures up to at
least 200°C and pressures up to at least 40 bar.For determination of antimony, use only digestion containers with minimal surface roughness to
(6)[12],prevent its adhesion to the container surface . Quartz containers are recommended because
they are usually more resistant to attrition. However, Teflon vessels without any scratch or damage on
their inner surface or any deposits are appropriate. Scratched or etched containers should not be used.
If there is any question regarding possible adsorption of antimony on vessels’ walls, test for surface
adsorption as described in section 7.3.3.NOTE Dedicated digestion vessels are recommended for the digestion of cosmetic samples, which may have
high levels of elements to be determined. To avoid memory effects, perform a blank digestion to clean vessels
after digesting highly loaded samples, before digesting subsequent samples.5.2 Microwave-assisted digestion instruments
Microwave-heated systems shall be equipped with a temperature measurement unit, which
simultaneously regulates the power control of the microwave. Reliable temperature measurement is
obtained e.g. through measurement sensors inserted into the pressure vessel. Only use microwave-
assisted digestion instruments equipped with temperature sensors and calibrate the temperature
sensor before use.5.3 Membrane filter, 0,45 µm pore size
The membrane filter used shall be inert with regard to the acid concentration of the measurement
solution and shall not bring any contamination into the measurement solution or adsorption of the
analytes. Several types of membrane material are commercially available (PTFE, PP…) and their fit for
purpose must be verified by means of appropriate measurements (blanks, QC samples…).
5.4 ICP-MSMass spectrometer with inductively coupled argon plasma is composed of a sample introduction and
an atomisation system, as well as an instrument control and evaluation unit. To prevent interferences
with the masses of the elements of chromium, nickel, arsenic and cadmium, use of a mass spectrometer
2 © ISO 2020 – All rights reserved---------------------- Page: 10 ----------------------
oSIST prEN ISO 21392:2020
ISO/DIS 21392:2020(E)
that is capable of compensating or minimising such interferences (e.g. collision and/or reaction cell,
resolution above 3000, alternatively corrective equations for higher concentrations).
6 Preparation of standards solutionsFor all the solutions, the terminology “part” in the standard refers to either volume or weight. That
means that standards and samples can be diluted by volume or weight. However, it should be consistent
for both standards and samples.6.1 Diluted nitric acid, produced by mixing nitric acid (4.2) with pure water (4.1) at a ratio
of approximately 1+9 parts respectively.6.2 Diluting solution
The composition of the diluting solution must have the same acid composition (total content and acid
ratio) as the Analytical Solution (the diluted digest solution). This solution should contain:
— 2,5 part of nitric acid (4.2),— 0,5 part of hydrochloric acid (4.3),
— 97 parts of water (4.1)
6.3 Internal Standard solutions.
The internal standards (IS) selected should cover all the mass range of the considered analytes and
have similar ionisation energy to the trace element for which it is used for correction purposes. It shall
also be checked that the native concentration of the internal standards in the sample to be analysed is
negligible and that they are not interfered by sample constituents.“Rhodium and Lutetium have proved to be well suited as internal standards (IS). Samples should be
checked for negligible native concentrations of the IS and that the IS are not interfered with by sample
constituents.Rhodium is suitable for determination of Chromium, Cobalt, Nickel, Arsenic, Cadmium and Antimony,
whereas Lutetium is suitable for determination of Lead. Alternatively, other elements may be used (for
example Indium or Iridium). Scandium, however, is not suitable as an IS due to Calcium interferences.
An IS with a mass (m/z) below 100 is not recommended because it may suffer from interferences from
matrix components.NOTES : - Internal standard solutions may be added in each sample and calibration solution at the same
concentration or may be added through an on line Y-fitting to a pump tube.- the concentration of the internal standard solutions must be included in the range 1 to 2 mg/l. In the following
sections, a concentration of 1 mg/l has been used for all the calculations.6.3.1 Rhodium(*) standard solution, 1 mg/l
Dilute the Rhodium stock solution (4.4.1) 1 + 999 with diluting solution (6.2). This internal standard
solution is stable at room temperature for 6 months.6.3.2 Lutetium(*) standard solution, 1 mg/l
Dilute the Lutetium stock solution (4.4.2) 1 + 999 with diluting solution (6.2). This internal standard
solution is stable at room temperature for 6 months.(*) NOTE : Indium or Iridium may also be used as internal standards
© ISO 2020 – All rights reserved 3
---------------------- Page: 11 ----------------------
oSIST prEN ISO 21392:2020
ISO/DIS 21392:2020(E)
6.4 Standard solutions
The concentrations of these standard solutions are examples and shall be adjusted to the specific
conditions in the laboratories.6.4.1 High concentration mixed standard solution, 10 mg/l
Dilute 100 times the analyte stock solution(s) (3.5) by adding:
— in the case of single analyte stock solutions, 1 part of each of these 7 solutions to 93 parts of the
diluting solution (6.2)— in the case of mixed stock solution, add 1 part of this solution to 99 parts of the diluting solution (6.2).
This high standard solution is stable at room temperature for 6 months.6.4.2 Low concentration mixed standard solution, 0.1 mg/l
Dilute 100 times the high concentration standard solution (5.4.1) by adding 1 part of this solution to
99 parts of the diluting solution (6.2). This low standard solution is stable at room temperature for
3 months.6.5 Calibration blank solution
The calibration blank solution corresponds to the matrix solution without any analyte of interest.
Generally, it corresponds to the diluting solution with the suitable concentration of the appropriate
internal standards if not added via a Y-fitting during the measurement.6.6 Calibration solutions
Mixed calibration solutions are prepared by diluting the low concentration mixed standard solution
(6.4.2) with the diluting solution (6.2) to levels in the linear range of the instrument and within the
targeted concentration range. Include a suitable concentration of the appropriate internal standards,
or add on line the internal standards by means of pumping into the sample flow through a Y-fitting.
At least 3 calibration solutions with various concentrations should be prepared. These calibration
solutions must be prepared daily.Examples of preparation procedure of calibration solutions are detailed in Table 1a (with addition of
the internal standards in all the calibration solutions) and Table 1b (with on line addition of the internal
standards via an Y-fitting).Table 1a — Example of calibration solutions of the ICP-MS – addition of the internal standards
in every calibration solutionCalibration solution Part of low Part of Rhodi- Part of Luteti- Part of the Analyte conc.
conc. mixed um standard um standard diluting solu- in the calibra-standard solution solution tion tion solution
solution
(6.3.1) (6.3.2) (6.2) (µg/l)
(6.4.2)
5.5.0 Calibration blank 0 2 2 496 0
5.5.1 Calibration solution 1 2,5 2 2 493,5 0,5
5.5.2 Calibration solution 2 5 2 2 491 1
5.5.3 Calibration solution 3 10 2 2 486 2
5.5.4 Calibration solution 4 25 2 2 471 5
5.5.5 Calibration solution 5 50 2 2 446 10
4 © ISO 2020 – All rights reserved
---------------------- Page: 12 ----------------------
oSIST prEN ISO 21392:2020
ISO/DIS 21392:2020(E)
Table 1b — Example of calibration solutions of the ICP-MS – online addition of the internal
standards via an Y-fittingPart of low Part of the di-
Analyte conc.
conc. mixed luting solution
in the calibra-
standard
Calibration solution
(6.2) tion solution
solution
(µg/l)
(6.4.2)
5.5.0 Calibration blank 0 500 0
5.5.1 Calibration solution 1 2,5 497,5 0,5
5.5.2 Calibration solution 2 5 495 1
5.5.3 Calibration solution 3 10 490 2
5.5.4 Calibration solution 4 25 475 5
5.5.5 Calibration solution 5 50 450 10
7 Procedure
7.1 General
WARNING — The use of this standard can involve hazardous materials, operations and
equipment. This standard does not address all the safety problems associated with its use. It
is the responsibility of the user of this standard to take appropriate measures for ensuring the
safety and health of the personnel prior to application of the standard and to fulfil statutory
requirements for this purpose.7.2 Preparation of samples
Homogenize the samples by means of suitable devices. After homogenization thoroughly clean the
devices in order to rule out contamination of the subsequent sample. The sample preparation step shall
ensure a homogeneous starting material for a weighed sample quantity.7.3 Pressure assisted digestion
WARNING 1 Depending on the degree or reactivity of the sample, it may be required to weigh in lower
quantities than specified in Section 7.3.1 in order to prevent extreme reactions or explosions. It shall be taken
into account that digestion of samples with high carbon contents (e.g. carbohydrates, fats, oils, waxes) may cause
explosions. Alcohols or solvents in combination with concentrated nitric acid may cause delayed severe reactions
already at room temperature.Therefore it is highly recommended to gently evaporate all volatile components
before adding the acid (Section 7.3.2).WARNING 2 Samples that are not covered by acid can cause local overheating of the digestion vessel and thus
lead to local melting and subsequent bursting of the digestion container. Prior to digestion, ensure that the entire
sample is fully covered by the acid mixture.Temperature and pressure into the vessels must be carefully controlled to ensure a proper digestion
(section 5.2.). To avoid differences in temperature and pressure among vessels, one should only digest
samples with similar composition in the same microwave-assisted digestion batch.7.3.1 Preparation of Sample by Digestion – General case
Precisely weigh ca. 200 mg of sample into a digestion container.
Add 1 ml of water (4.1) and thoroughly mix with a shaking device until the sample is completely
suspended in the water.Add 5 ml nitric acid (4.2) to the mixture and mix again. The sample should be completely covered
with the solution. Allow the mixture to rest in a closed digestion vessel to ensure that the preliminary
© ISO 2020 – All rights reserved 5---------------------- Page: 13 ----------------------
oSIST prEN ISO 21392:2020
ISO/DIS 21392:2020(E)
reaction takes place. Depending on the reactive behavior of the sample the duration of the preliminary
reaction can require resting periods of 30 minutes up to overnight.Add then 1 ml of hydrochloric acid (4.3) and briefly mix. After addition of the hydrochloric acid, the
pressure vessel shall be closed and sealed immediately to make sure that the formed chlorine gas is
available for the reaction and does not evaporate.7.3.2 Preparation of Sample by Digestion – Specific cases
• For highly water-based cosmetic products, such as lotion, milky lotion, cleanser, micellar water, a
sample mass could reach 400 mg. In this case no addition of water is required before addition of
acids (7.3.1).• For all the other specific cases, sample mass can be adapted but the ratio between sample mass and
acid volumes (7.3.1) shall not be changed.In case of high volatile content products, it is highly recommended for safety reason to completely
remove volatile portions through a caref...