Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-061: Particular requirements for laboratory atomic spectrometers with thermal atomization and ionization

2017-10-06 SW: See 41_x_or for Annex ZZ, ZA and common mods provided.

Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte - Teil 2-061: Besondere Anforderungen an Labor-Atomspektrometer mit thermischer Atomisierung und Ionisation

Exigences de sécurité pour appareils électriques de mesurage, de régulation et de laboratoire - Partie 2-061: Exigences particulières pour spectromètres atomiques de laboratoire avec vaporisation et ionisation thermiques

Varnostne zahteve za električno opremo za meritve, nadzor in laboratorijsko uporabo - 2-061. del: Posebne zahteve za laboratorijske atomske spektrometre s termično atomizacijo in ionizacijo

General Information

Status
Not Published
Current Stage
5020 - Vote circulated
Due Date
30-Apr-2021
Completion Date
30-Apr-2021

RELATIONS

Buy Standard

Draft
prEN 61010-2-061:2018/oprAA:2021
English language
6 pages
sale 10% off
Preview
sale 10% off
Preview

e-Library read for
1 day

Standards Content (sample)

SLOVENSKI STANDARD
oSIST prEN 61010-2-061:2018/oprAA:2021
01-januar-2021
Varnostne zahteve za električno opremo za meritve, nadzor in laboratorijsko
uporabo - 2-061. del: Posebne zahteve za laboratorijske atomske spektrometre s
termično atomizacijo in ionizacijo

Safety requirements for electrical equipment for measurement, control and laboratory

use - Part 2-061: Particular requirements for laboratory atomic spectrometers with

thermal atomization and ionization

Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte - Teil 2

-061: Besondere Anforderungen an Labor-Atomspektrometer mit thermischer
Atomisierung und Ionisation

Exigences de sécurité pour appareils électriques de mesurage, de régulation et de

laboratoire - Partie 2-061: Exigences particulières pour spectromètres atomiques de

laboratoire avec vaporisation et ionisation thermiques
Ta slovenski standard je istoveten z: prEN 61010-2-061:2017/prAA
ICS:
19.080 Električno in elektronsko Electrical and electronic
preskušanje testing
71.040.20 Laboratorijska posoda in Laboratory ware and related
aparati apparatus
oSIST prEN 61010-2- en,fr,de
061:2018/oprAA:2021

2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------
oSIST prEN 61010-2-061:2018/oprAA:2021
---------------------- Page: 2 ----------------------
oSIST prEN 61010-2-061:2018/oprAA:2021
EUROPEAN STANDARD DRAFT
prEN 61010-2-061:2017
NORME EUROPÉENNE
EUROPÄISCHE NORM
prAA
November 2020
ICS 19.080; 71.040.20
English Version
Safety requirements for electrical equipment for measurement,
control and laboratory use - Part 2-061: Particular requirements
for laboratory atomic spectrometers with thermal atomization and
ionization

Exigences de sécurité pour appareils électriques de Sicherheitsbestimmungen für elektrische Mess-, Steuer-,

mesurage, de régulation et de laboratoire - Partie 2-061: Regel- und Laborgeräte - Teil 2-061: Besondere

Exigences particulières pour spectromètres atomiques de Anforderungen an Labor-Atomspektrometer mit thermischer

laboratoire avec vaporisation et ionisation thermiques Atomisierung und Ionisation

This draft amendment prAA, if approved, will modify the European Standard prEN 61010-2-061:2017; it is submitted to CENELEC members

for enquiry.
Deadline for CENELEC: 2021-02-19.
It has been drawn up by CLC/TC 66X.

If this draft becomes an amendment, CENELEC members are bound to comply with the CEN/CENELEC Internal Regulations which

stipulate the conditions for giving this amendment the status of a national standard without any alteration.

This draft amendment was established by CENELEC in three official versions (English, French, German).

A version in any other language made by translation under the responsibility of a CENELEC member into its own language and notified to

the CEN-CENELEC Management Centre has the same status as the official versions.

CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,

Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the

Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,

Turkey and the United Kingdom.

Recipients of this draft are invited to submit, with their comments, notification of any relevant patent rights of which they are aware and to

provide supporting documentation.

Warning : This document is not a European Standard. It is distributed for review and comments. It is subject to change without notice and

shall not be referred to as a European Standard.
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels

© 2020 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.

Project: 65873 Ref. No. prEN 61010-2-061:2017/prAA E
---------------------- Page: 3 ----------------------
oSIST prEN 61010-2-061:2018/oprAA:2021
prEN 61010-2-061:2017/prAA:2020
1 European foreword

2 This document (prEN 61010-2-061:2017/prAA:2020) has been prepared by CLC/TC 66X “Safety of

3 measuring, control, and laboratory equipment”.
4 This document is currently submitted to the Enquiry.
5 The following dates are proposed:
• latest date by which the existence of this (doa) dor + 6 months
document has to be announced at national
level
• latest date by which this document has to be (dop) dor + 12 months
implemented at national level by publication of
an identical national standard or by
endorsement
• latest date by which the national standards (dow) dor + 36 months
conflicting with this document have to be (to be confirmed or
withdrawn modified when voting)
6 This document amends prEN 61010 2 061:2017.

7 This document has been prepared under a mandate given to CENELEC by the European Commission

8 and the European Free Trade Association, and supports essential requirements of EU Directive(s).

9 For the relationship with EU Directive(s) see informative Annex ZZ, which is an integral part of this

10 document.
---------------------- Page: 4 ----------------------
oSIST prEN 61010-2-061:2018/oprAA:2021
prEN 61010-2-061:2017/prAA:2020
11 1 Modification to 4.4.2.101, “Sampling probe tip”
12 Replace the entire subclause by the following:

13 “Any system designed to withdraw a sampling probe tip after sampling has been completed shall be

14 overridden, so as to leave the tip in its most exposed position when a sample vessel is removed.

15 Exception:

16 The withdrawal system need not be overridden if the sampling probe is designed in such a manner

17 that no SINGLE FAULT CONDITION can cause the tip to remain exposed after sampling has been

18 completed.”
19 2 Modification to 5.1.5.101, “Gas and liquid connections”
20 Replace the entire subclause by the following:

21 “The following shall be unambiguously marked adjacent to the connector on the equipment (see 5.2):

22 a) the identity of the gas or liquid;
23 b) the maximum permitted pressure,
24 c) flow direction of the gas and liquid, if applicable.

25 NOTE Such markings can be specific (for example acetylene, propane, water) or generic (for example fuel

26 gas, oxidant gas, coolant, waste liquid).

27 Where no internationally recognized symbol (such as a chemical formula) exists, the equipment shall

28 be marked with symbol 14 of Table 1.
29 Conformity is checked by inspection.”
30 3 Modification to 5.4.3, “Equipment installation”
31 Replace list item bb) with the following:
32 “bb) requi
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.