Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-010: Particular requirements for laboratory equipment for the heating of materials

This clause of Part 1 is applicable except as follows: 1.1.1 Equipment included in scope Replacement: Replace the second paragraph by the following: This part of IEC 61010 specifies particular safety requirements for the following types a) to c) of electrical equipment and their accessories, wherever they are intended to be used, whenever the heating of materials is one of the functions of the equipment. Addition: Add the following text after item c): It is possible that all or part of the equipment falls within the scope of one or more other Part 2 standards of IEC 61010 as well as within the scope of this standard. In that case, the requirements of those other Part 2 standards will also apply. In particular, if equipment is intended to be used for in vitro diagnostic (IVD) purposes, the requirements of IEC 61010-2-101 will also apply. However, when the equipment incorporates a refrigerating system and a heating function where the combination of the two introduces additional or more severe HAZARDS than if treated separately, then it is possible that IEC 61010-2-012 is applicable instead of this Part 2-010. See further information in the flow chart (Figure 102) for the selection process and the guidance in the Introduction. 1.1.2 Equipment excluded from scope Addition: Add the following items after item j): aa) equipment for the heating and ventilation of laboratories; bb) sterilizing equipment; cc) heating and/or cooling equipment which the OPERATOR is intended to enter, and which is large enough for the OPERATOR to remain inside with the door or doors closed.

Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte - Teil 2-010: Besondere Anforderungen für Laborgeräte für das Erhitzen von Stoffen

Exigences de sécurité pour appareils électriques de mesurage, de régulation et de laboratoire - Partie 2-010: Exigences particulières pour appareils de laboratoire utilisés pour l'échauffement des matières

L'article de la Partie 1 est applicable à l'exception de ce qui suit: 1.1.1 Appareils inclus dans le domaine d'application Remplacement: Remplacer le deuxième alinéa par ce qui suit: La présente partie de l'IEC 61010 spécifie les exigences de sécurité particulières pour les appareils électriques et leurs accessoires des types a) à c) suivants, quelle que soit l'utilisation à laquelle ils sont destinés, lorsque l'échauffement des matières est l'une des fonctions de l'appareil. Addition: Ajouter le texte suivant après le point c): Il est possible qu'une ou toutes les parties de l'appareil relèvent du domaine d'application d'une ou plusieurs autres Parties 2 de l'IEC 61010, ainsi que du domaine d'application de la présente norme. En particulier, si l'appareil est prévu pour être utilisé pour l'IVD, les exigences de l'IEC 61010-2-101 s'appliquent également. Toutefois, lorsque l'appareil comprend un système frigorifique et une fonction de chauffage dont la combinaison des deux introduit des dangers supplémentaires ou plus graves que lorsque le traitement est réalisé séparément, alors il est possible que l'IEC 61010-2-012 soit applicable à la place de la présente Partie 2-010. De plus amples informations sur le processus de sélection sont données dans l'organigramme (Figure 102) et des recommandations sont données dans l'Introduction. 1.1.2 Appareils exclus du domaine d'application Addition: Ajouter les points suivants après le point j): aa) appareils pour l'échauffement et la ventilation des laboratoires; bb) appareils de stérilisation; cc) appareil de chauffage et/ou de refroidissement à l'intérieur duquel il est prévu que l'OPERATEUR entre, et qui est assez grand pour que l'OPERATEUR puisse s'y tenir avec la ou les portes fermées.

Varnostne zahteve za električno opremo za meritve, nadzor in laboratorijsko uporabo - 2-010. del: Posebne zahteve za laboratorijsko opremo za gretje materialov

General Information

Status
Published
Publication Date
21-May-2020
Current Stage
6060 - Document made available - Publishing
Start Date
22-May-2020
Completion Date
22-May-2020

Relations

Buy Standard

Standard
EN IEC 61010-2-010:2020 - BARVE
English language
30 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

SLOVENSKI STANDARD
SIST EN IEC 61010-2-010:2020
01-julij-2020
Nadomešča:
SIST EN 61010-2-010:2015
Varnostne zahteve za električno opremo za meritve, nadzor in laboratorijsko
uporabo - 2-010. del: Posebne zahteve za laboratorijsko opremo za gretje
materialov
Safety requirements for electrical equipment for measurement, control and laboratory
use - Part 2-010: Particular requirements for laboratory equipment for the heating of
materials
Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte - Teil 2
-010: Besondere Anforderungen an Laborgeräte für das Erhitzen von Stoffen
Règles de sécurité pour appareils électriques de mesurage, de régulation et de
laboratoire - Partie 2-010: Exigences particulières pour appareils de laboratoire utilisés
pour l’échauffement des matières
Ta slovenski standard je istoveten z: EN IEC 61010-2-010:2020
ICS:
19.080 Električno in elektronsko Electrical and electronic
preskušanje testing
71.040.10 Kemijski laboratoriji. Chemical laboratories.
Laboratorijska oprema Laboratory equipment
SIST EN IEC 61010-2-010:2020 en,fr,de
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------
SIST EN IEC 61010-2-010:2020

---------------------- Page: 2 ----------------------
SIST EN IEC 61010-2-010:2020


EUROPEAN STANDARD EN IEC 61010-2-010

NORME EUROPÉENNE

EUROPÄISCHE NORM
May 2020
ICS 19.080; 71.040.20 Supersedes EN 61010-2-010:2014 and all of its
amendments and corrigenda (if any)
English Version
Safety requirements for electrical equipment for measurement,
control and laboratory use - Part 2-010: Particular requirements
for laboratory equipment for the heating of materials
(IEC 61010-2-010:2019)
Exigences de sécurité pour appareils électriques de Sicherheitsbestimmungen für elektrische Mess-, Steuer-,
mesurage, de régulation et de laboratoire - Partie 2-010: Regel- und Laborgeräte - Teil 2-010: Besondere
Exigences particulières pour appareils de laboratoire Anforderungen für Laborgeräte für das Erhitzen von Stoffen
utilisés pour l'échauffement des matières (IEC 61010-2-010:2019)
(IEC 61010-2-010:2019)
This European Standard was approved by CENELEC on 2019-03-19. CENELEC members are bound to comply with the CEN/CENELEC
Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC
Management Centre or to any CENELEC member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation
under the responsibility of a CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the
same status as the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the
Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and the United Kingdom.


European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2020 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
 Ref. No. EN IEC 61010-2-010:2020 E

---------------------- Page: 3 ----------------------
SIST EN IEC 61010-2-010:2020
EN IEC 61010-2-010:2020 (E)
European foreword
This document (EN IEC 61010-2-010:2020) consists of the text of IEC 61010-2-010:2019 prepared by
IEC/TC 66 "Safety of measuring, control and laboratory equipment".
The following dates are fixed:
• latest date by which the document has to be implemented at national (dop) 2020-11-22
level by publication of an identical national standard or by endorsement
• latest date by which the national standards conflicting with the (dow) 2023-05-22
document have to be withdrawn
This document supersedes EN 61010-2-010:2014 and all of its amendments and corrigenda (if any).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC shall not be held responsible for identifying any or all such patent rights.
This document has been prepared under a mandate given to CENELEC by the European Commission
and the European Free Trade Association, and supports essential requirements of EU Directive(s).
For the relationship with EU Directive(s) see informative Annex ZZ, which is an integral part of this
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.