Determination of particle size distribution by centrifugal liquid sedimentation methods — Part 2: Photocentrifuge method

ISO 13318-2:2007 covers methods for determining the particle size distribution of particulate materials by means of centrifugal sedimentation in a liquid. Solids concentrations are determined by the transmission of a light beam. The resulting signal enables conversion to a particle size distribution. The method of determining the particle size distribution described in ISO 13318-2:2007 is applicable to powders that can be dispersed in liquids, powders that are present in slurry form and some emulsions. Typical particle size range for analysis is from about 0,1 µm to 5 µm. The method is applicable to powders in which all particles have the same density and comparable shapes and do not undergo chemical or physical change in the suspension liquid. It is usually necessary that the particles have a density higher than that of the liquid.

Détermination de la distribution granulométrique par les méthodes de sédimentation centrifuge dans un liquide — Partie 2: Méthode photocentrifuge

General Information

Status
Published
Publication Date
17-Sep-2007
Current Stage
9093 - International Standard confirmed
Completion Date
10-Dec-2021
Ref Project

Relations

Buy Standard

Standard
ISO 13318-2:2007 - Determination of particle size distribution by centrifugal liquid sedimentation methods
English language
17 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

INTERNATIONAL ISO
STANDARD 13318-2
Second edition
2007-09-15

Determination of particle size distribution
by centrifugal liquid sedimentation
methods —
Part 2:
Photocentrifuge method
Détermination de la distribution granulométrique par les méthodes de
sédimentation centrifuge dans un liquide —
Partie 2: Méthode photocentrifuge




Reference number
ISO 13318-2:2007(E)
©
ISO 2007

---------------------- Page: 1 ----------------------
ISO 13318-2:2007(E)
PDF disclaimer
This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but
shall not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In
downloading this file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat
accepts no liability in this area.
Adobe is a trademark of Adobe Systems Incorporated.
Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation
parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In
the unlikely event that a problem relating to it is found, please inform the Central Secretariat at the address given below.


COPYRIGHT PROTECTED DOCUMENT


©  ISO 2007
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means,
electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or
ISO's member body in the country of the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland

ii © ISO 2007 – All rights reserved

---------------------- Page: 2 ----------------------
ISO 13318-2:2007(E)
Contents Page
Foreword. iv
Introduction . v
1 Scope . 1
2 Normative references . 1
3 Terms, definitions and symbols. 1
4 Principle. 2
5 Apparatus . 2
6 Sampling. 6
7 Preparation . 6
7.1 Sample preparation . 6
7.2 Temperature . 6
7.3 Dispersion . 6
8 Procedure . 6
8.1 Line-start methods. 6
8.2 Homogeneous technique. 7
9 Tests in duplicate and validation . 8
9.1 Tests in duplicate. 8
9.2 Validation. 8
10 Calculation of results . 8
10.1 General. 8
10.2 Calculation of particle size . 8
10.3 Calculation of cumulative mass percentage. 8
11 Reporting of results. 9
Annex A (informative) Worked example. 11
Annex B (informative) Extinction curve, example for titanium dioxide . 15
Annex C (informative) Effect of radial dilution . 16
Bibliography . 17

© ISO 2007 – All rights reserved iii

---------------------- Page: 3 ----------------------
ISO 13318-2:2007(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies
(ISO member bodies). The work of preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which a technical committee has been
established has the right to be represented on that committee. International organizations, governmental and
non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the
International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.
The main task of technical committees is to prepare International Standards. Draft International Standards
adopted by the technical committees are circulated to the member bodies for voting. Publication as an
International Standard requires approval by at least 75 % of the member bodies casting a vote.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. ISO shall not be held responsible for identifying any or all such patent rights.
ISO 13318-2 was prepared by Technical Committee ISO/TC 24, Sieves, sieving and other sizing methods,
Subcommittee SC 4, Sizing by methods other than sieving.
This second edition cancels and replaces ISO 13318-2:2001, of which it constitutes a minor revision, due to
the extension of Clause 4 and 5.2, and the addition of Figure 3 and the Bibliography.
ISO 13318 consists of the following parts, under the general title Determination of particle size distribution by
centrifugal liquid sedimentation methods:
⎯ Part 1: General principles and guidelines
⎯ Part 2: Photocentrifuge method
⎯ Part 3: Centrifugal X-ray method
iv © ISO 2007 – All rights reserved

---------------------- Page: 4 ----------------------
ISO 13318-2:2007(E)
Introduction
The sample suspension in a photocentrifuge may be contained in a cuvette or a disc. Sample concentration is
determined by changes in a light signal monitored at a known radius. The cuvette photocentrifuge can only be
run in the homogeneous mode whereas the disc photocentrifuge may be run in either the homogeneous or the
line-start mode. Some systems permit the coarse end of the distribution to be measured in a gravitational
mode and the fine end in the centrifugal mode. The use of light to determine particle size distribution requires
a calibration factor to be applied as the particle size approaches the wavelength of the light, due to the
inapplicability of the laws of geometric optics.
© ISO 2007 – All rights reserved v

---------------------- Page: 5 ----------------------
INTERNATIONAL STANDARD ISO 13318-2:2007(E)

Determination of particle size distribution by centrifugal liquid
sedimentation methods —
Part 2:
Photocentrifuge method
WARNING — This part of ISO 13318 may involve hazardous materials, operations and equipment. This
part of ISO 13318 does not purport to address all of the safety problems associated with its use. It is
the responsibility of the user of this part of ISO 13318 to establish appropriate safety and health
practices and determine the applicability of the regulatory limitations prior to its use.
1 Scope
This part of ISO 13318 covers methods for determining the particle size distribution of particulate materials by
means of centrifugal sedimentation in a liquid. Solids concentrations are determined by the transmission of a
light beam. The resulting signal enables conversion to a particle size distribution.
The method of determining the particle size distribution described in this part of ISO 13318 is applicable to
powders that can be dispersed in liquids, powders that are present in slurry form and some emulsions. Typical
particle size range for analysis is from about 0,1 µm to 5 µm. The method is applicable to powders in which all
particles have the same density and comparable shapes and do not undergo chemical or physical change in
the suspension liquid. It is usually necessary that the particles have a density higher than that of the liquid.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.
ISO 13318-1, Determination of particle size distribution by centrifugal liquid sedimentation methods — Part 1:
General principles and guidelines
ISO 14887, Sample preparation — Dispersing procedures for powders in liquids
3 Terms, definitions and symbols
For the purposes of this document, the terms, definitions and symbols given in ISO 13318-1, and the following
symbols, apply.
D optical density
E extinction coefficient for a particle of diameter x
i i
F(surface) frequency undersize by surface
G constant dependent upon the geometry of the system, the dimensions of the light beam and on
the shape of the particles
© ISO 2007 – All rights reserved 1

---------------------- Page: 6 ----------------------
ISO 13318-2:2007(E)
l transmission of the emergent light beam, at the time t, after the start of sedimentation
l transmission of the emergent light beam when no particles are present
0
M distance from rotation axis to measurement zone (mm)
n number of particles of diameter x in the beam
i i
R distance from rotation axis to centrifuge wall, inner disc radius (mm)
S distance from rotation axis to liquid/air interface of sample (mm)
x diameter of the smallest particle in the light beam (µm)
0
x diameter of the largest particle in the light beam, i.e. the Stokes diameter (µm)
St
4 Principle
A stable, finely collimated beam of light passes through a spinning disc or cuvette and sedimenting sample
and is detected at a known radius. Light rays, typically from either a white light source (e.g. incandescent bulb)
or a monochromatic coherent source (e.g. laser), pass through the suspension and are detected by a
photodiode or photomultiplier. The disc photocentrifuge can be operated in the line-start or homogeneous
mode whereas the cuvette photocentrifuge can be operated only in the homogeneous mode. The signal of the
light beam is monitored over the analysis time. The mass percentage of sample present in the beam is
determined by calculating the ratio of the light transmission signal, by use of a clear dispersing liquid, to the
light transmission signal with the sample present.
In the line-start mode the disc initially contains clear fill liquid to give maximum light transmission. Then the
sample is injected as a thin layer on top of the spinning fill liquid and begins to settle outward radially. When
the largest particles present reach the light beam the light transmission decreases, returning to the original
transmission value when the smallest particle present passes through the beam. A buffer layer is usually
injected over the fill liquid to prevent suspension breaking through the interface in a phenomenon known as
“streaming”.
In an alternative configuration, the determination of the particle size distribution by centrifugal liquid
sedimentation method can also be accomplished using a photocentrifuge containing a line light source and a
line sensor detector system aligned with the sample cell. In this configuration, light intensity/extinction
alterations during centrifugation are measured simultaneously over the whole sedimentation zone as a
function of both time and of position. From these data, the particle size distribution may be calculated either
from the time course of the extinction at a freely selectable position within the sample [numerical integration,
ISO 13318-1:2001, Equation (11) in 4.3.3.3] or from the extinction profile along the sample at a freely
selectable time (analytical integration, for details see Reference [1]).
5 Apparatus
5.1 Disc photocentrifuge, with a chamber consisting of a hollow disc with an entry port coaxial with the
axis of rotation (see Figure 1). Typically this is mounted vertically, or at a small angle to the vertical, on to the
−1 −1
shaft of an electric motor with a digitally variable speed typically between 500 r⋅min and 15 000 r⋅min . A
white light source and detector assembly measures transmittance through the suspension as a function of
time. The instrument can be used in either a line-start or homogeneous mode. Extinction coefficient
corrections need to be applied for the breakdown in the laws of geometric optics for both line-start and
homogeneous modes. Additionally, a correction is required for radial dilution effects when the homogeneous
mode is used. Software is provided with commercial equipment to convert the data directly into size
distributions in the form of tables or graphs of cumulative mass percentage versus particle size.
2 © ISO 2007 – All rights reserved

---------------------- Page: 7 ----------------------
ISO 13318-2:2007(E)
5.2 Cuvette photocentrifuge, in which the disc is replaced with a rectangular cell containing a
homogeneous suspension (see Figure 2). Corrections need to be made for both radial dilution and light
scattering effects as described in ISO 13318-1. Cuvette photocentrifuges can typically be run in both the
gravitational and centrifugal modes. Additionally, some systems may offer a gradient mode permitting the
centrifuge to accelerate throughout the analysis in order to reduce the measurement time.
When determining particle size distribution using an apparatus containing a line light source and a line sensor
detection system, transmittance is measured along the entire sedimentation zone simultaneously (see Figure 3).
5.3 Ancillary apparatus, consisting of:
⎯ dispersing vessel, e.g. glass beaker or bottle, of appropriate dimensions;
⎯ flexible spatula;
⎯ ultrasonic bath or probe, a bottle shaker or high speed mechanical stirrer capable of rotating at
−1 −1
500 r⋅min to 1 000 r⋅min .

Key
1 motor shaft
2 spin fluid
3 buffer layer
4 suspension
5 entry port
6 polymethylmethacrylate disc
Figure 1 — Side view of the disc of a disc photocentrifuge
© ISO 2007 – All rights reserved 3

---------------------- Page: 8 ----------------------
ISO 13318-2:2007(E)

Key
1 motor
2 motor control
3 signal process
4 analogue to digital converter (ADC)
5 central processing unit (CPU)
6 date time
7 analysis parameters
8 revolutions per minute (RPM)
9 keyboard
10 printer
11 analog interface
12 computer interface
13 photosensor revolutions per minute (RPM)
14 centrifugal cell
15 rotating disc
16 light-emitting diode (LED)
17 photocell (reference)
18 photocell (sample)
19 photocells
a
Synchro-signal (reference).
b
Analog signal.
c
Synchro-signal (sample).
Figure 2 — Schematic diagram of a typical cuvette photocentrifuge
4 © ISO 2007 – All rights reserved

---------------------- Page: 9 ----------------------
ISO 13318-2:2007(E)

Key
X position for one selected time
Y intensity
1 axis of rotation
2 incident parallel light
3 sample cell
4 centrifugal force
5 rotor
6 transmitted light
7 line detector array
Figure 3 — Schematic diagram of a cuvette photocentrifuge with line light source
and line detector array
© ISO 2007 – All rights reserved 5

---------------------- Page: 10 ----------------------
ISO 13318-2:2007(E)
6 Sampling
For sampling see ISO 13318-1.
7 Preparation
7.1 Sample preparation
An analysis sample shall be prepared as described in ISO 13318-1. The volume and concentration required
depends upon the volume of the centrifuge disc (or cuvette), the sensitivity of the optical-electronic system
and whether the line-start (disc only) or homogeneous method is to be used. In general, lower concentrations
are required than for other sedimentation methods. A mass per volume concentration typically less than
0,25 % and providing an attenuation preferably in the 20 % to 30 % range compared to
...

Questions, Comments and Discussion