Non-invasive sphygmomanometers - Part 4: Test procedures to determine the overall system accuracy of automated non-invasive sphygmomanometers

This document describes test procedures for investigations to determine the overall system accuracy of automated non-invasive sphygmomanometers, designed for the indirect measurement of blood pressure.

Nichtinvasive Blutdruckmessgeräte - Teil 4: Prüfverfahren zur Bestimmung der Messgenauigkeit von automatischen nichtinvasiven Blutdruckmessgeräten

Dieses Dokument beschreibt Prüfverfahren für Untersuchungen zur Bestimmung der Messgenauigkeit von automatischen nichtinvasiven Blutdruckmessgeräten, die für die indirekte Blutdruckmessung bestimmt sind.

Tensiometres non invasifs - Partie 4 : Procédures pour déterminer la précision de l'ensemble du systeme des tensiometres non invasifs automatiques

Le présent document décrit les procédures d'investigation visant a déterminer la précision de l'ensemble du systeme des tensiometres non invasifs automatiques conçus pour le mesurage indirect de la pression artérielle.

Neinvazivni sfigmomanometri - 4. del: Preskusni postopki za ugotavljanje splošne točnosti sistema za avtomatizirane neinvazivne sfigmomanometre

General Information

Status
Withdrawn
Publication Date
28-Feb-2005
Withdrawal Date
18-Aug-2014
Technical Committee
Current Stage
9900 - Withdrawal (Adopted Project)
Start Date
19-Aug-2014
Due Date
11-Sep-2014
Completion Date
19-Aug-2014

Relations

Buy Standard

Standard
EN 1060-4:2005
English language
32 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.Nichtinvasive Blutdruckmessgeräte - Teil 4: Prüfverfahren zur Bestimmung der Messgenauigkeit von automatischen nichtinvasiven BlutdruckmessgerätenTensiometres non invasifs - Partie 4 : Procédures pour déterminer la précision de l'ensemble du systeme des tensiometres non invasifs automatiquesNon-invasive sphygmomanometers - Part 4: Test procedures to determine the overall system accuracy of automated non-invasive sphygmomanometers11.040.55Diagnostic equipmentICS:Ta slovenski standard je istoveten z:EN 1060-4:2004SIST EN 1060-4:2005en01-marec-2005SIST EN 1060-4:2005SLOVENSKI
STANDARD



SIST EN 1060-4:2005



EUROPEAN STANDARDNORME EUROPÉENNEEUROPÄISCHE NORMEN 1060-4September 2004ICS 11.040.55 English versionNon-invasive sphygmomanometers - Part 4: Test procedures todetermine the overall system accuracy of automated non-invasive sphygmomanometersTensiomètres non invasifs - Partie 4 : Procédures pourdéterminer la précision de l'ensemble du système destensiomètres non invasifs automatiquesNichtinvasive Blutdruckmessgeräte - Teil 4: Prüfverfahrenzur Bestimmung der Messgenauigkeit von automatischennichtinvasiven BlutdruckmessgerätenThis European Standard was approved by CEN on 9 July 2004.CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this EuropeanStandard the status of a national standard without any alteration. Up-to-date lists and bibliographical references concerning such nationalstandards may be obtained on application to the Central Secretariat or to any CEN member.This European Standard exists in three official versions (English, French, German). A version in any other language made by translationunder the responsibility of a CEN member into its own language and notified to the Central Secretariat has the same status as the officialversions.CEN members are the national standards bodies of Austria, Belgium, Cyprus, Czech Republic, Denmark, Estonia, Finland, France,Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Slovakia,Slovenia, Spain, Sweden, Switzerland and United Kingdom.EUROPEAN COMMITTEE FOR STANDARDIZATIONCOMITÉ EUROPÉEN DE NORMALISATIONEUROPÄISCHES KOMITEE FÜR NORMUNGManagement Centre: rue de Stassart, 36
B-1050 Brussels© 2004 CENAll rights of exploitation in any form and by any means reservedworldwide for CEN national Members.Ref. No. EN 1060-4:2004: ESIST EN 1060-4:2005



EN 1060-4:2004 (E) 2 Contents Foreword.4 1 Scope.5 2 Normative references.5 3 Terms and definitions.5 4 Requirements.6 4.1 General information on the non-invasive reference methods.6 4.2 General information on the invasive reference methods.6 4.3 Selection of clinical test method.6 4.4 Application of test method.7 4.4.1 Non-invasive reference measurement.7 4.4.2 Invasive reference measurement.8 4.5 Test equipment.8 4.5.1 Non-invasive reference measurement.8 4.5.2 Invasive blood pressure measurement.8 4.6 Subjects.9 4.6.1 General.9 4.6.2 Non-invasive reference measurement.9 4.6.3 Invasive reference measurement.10 4.7 Cuff size.10 4.8 Procedure.10 4.9 Evaluation of results.10 4.9.1 General.10 4.9.2 Non-invasive reference measurement.11 4.9.3 Documentation.13 5 Non-invasive test methods.13 5.1 Test method N1 - simultaneous blood pressure measurement on the same upper arm.13 5.1.1 Procedure.13 5.1.2 Evaluation.14 5.2 Test method N2 - blood pressure measurement on the opposite arms.14 5.2.1 Procedure.14 5.2.2 Evaluation.16 5.2.3 Documentation.16 5.3 Test method N3 - sequential blood pressure measurement on the same arm.17 5.3.1 Procedure.17 5.3.2 Evaluation.18 5.4 Test method N4 - simultaneous blood pressure measurement on the same upper arm at rest and under physical load.18 5.4.1 Procedure.18 5.4.2 Evaluation.18 5.4.3 Documentation.18 5.5 Test method N5 - ambulatory blood pressure measurement on the same upper arm.18 5.5.1 Procedure of measurement.18 5.5.2 Evaluation.19 5.5.3 Documentation.19 5.6 Test method N6 - ambulatory blood pressure measurement on the opposite arms.19 5.6.1 Procedure.19 5.6.2 Evaluation.20 5.6.3 Documentation.20 6 Invasive test methods.20 6.1 Test method I 1 - invasive blood pressure measurement on adults.20 6.1.1 Procedure.20 6.1.2 Evaluation.20 SIST EN 1060-4:2005



EN 1060-4:2004 (E) 3 6.1.3 Documentation.21 6.2 Test method I 2 - invasive blood pressure measurement on neonates and infants.21 6.2.1 Procedure.21 6.2.2 Evaluation.21 6.2.3 Documentation.22 Annex A (informative) Chronological order of the test method N 3 Sequential blood pressure measurement on the same arm.23 Annex B (informative)
Example of a record of a clinical investigation according to method N1 or N3.24 Annex C (informative)
Summary of the WHO recommendations for auscultatory blood pressure measurement.27 Annex D (informative)
Observer training and assessment etc.28 Annex E (informative)
A–deviations.30 Annex ZA (informative)
Relationship between this European Standard and the
Essential Requirements of EU Directive 93/42/EC Medical Devices.31 Bibliography.32
SIST EN 1060-4:2005



EN 1060-4:2004 (E) 4 Foreword This document (EN 1060-4:2004) has been prepared by Technical Committee CEN/TC 205 “Non-active medical devices”, the secretariat of which is held by BSI. This European Standard shall be given the status of a national standard, either by publication of an identical text or by endorsement, at the latest by March 2005, and conflicting national standards shall be withdrawn at the latest by March 2005. This document has been prepared under a mandate given to CEN by the European Commission and the European Free Trade Association, and supports essential requirements of EU Directive(s). For relationship with EU Directive, see informative Annex ZA, which is an integral part of this document. EN 1060 consists of the following Parts under the general title “Non-invasive sphygmomanometers”: Part 1: General requirements Part 2: Supplementary requirements for mechanical sphygmomanometers Part 3: Supplementary requirements for electro-mechanical blood pressure measuring systems Part 4: Test procedures to determine the overall system accuracy of automated non-invasive sphygmomanometers Annexes A, B, C, D, E and ZA are informative.
This document includes a Bibliography. According to the CEN/CENELEC Internal Regulations, the national standards organizations of the following countries are bound to implement this European Standard: Austria, Belgium, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Slovakia, Slovenia, Spain, Sweden, Switzerland and United Kingdom. SIST EN 1060-4:2005



EN 1060-4:2004 (E) 5 1 Scope This document describes test procedures for investigations to determine the overall system accuracy of automated non-invasive sphygmomanometers, designed for the indirect measurement of blood pressure. 2 Normative references The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
EN 1060-1:1995, Non-invasive sphygmomanometers – Part 1: General requirements. EN 1060-2, Non-invasive sphygmomanometers – Part 2: Supplementary requirements for mechanical sphygmomanometers. EN 1060-3:1997, Non-invasive sphygmomanometers - Part 3: Supplementary requirements for electro-mechanical blood pressure measuring systems. EN ISO 14155-1, Clinical investigation of medical devices for human subjects - Part 1: General requirements (ISO 14155-1:2003). EN ISO 14155-2, Clinical investigation of medical devices for human subjects - Part 2: Clinical investigation plans (ISO 14155-2:2003). 3 Terms and definitions For the purposes of this document, the terms and definitions given in EN 1060–1:1995 and EN 1060–3:1997 and the following apply. 3.1 neonatal mode mode, specified by the manufacturer of the sphygmomanometer for the measurement of neonates/infants 3.2 paired measurements parallel blood pressure measurements carried out by the observers and the device under test  at the same arm at the same time or
 on opposite arms at the same time or
 at the same arm at different times or  at one arm and another limb, at the same time or at different times. SIST EN 1060-4:2005



EN 1060-4:2004 (E) 6 4 Requirements 4.1 General information on the non-invasive reference methods The clinical investigation shall be performed in accordance with EN ISO 14155–1 and EN ISO 14155–2. The auscultatory blood pressure measurements described shall be carried out by two observers by means of a double stethoscope. The auscultatory reference value will then be the mean value of the two values determined by the observers. The difference between both values shall not exceed 4 mmHg. Any measurements with observer-to-observer differences greater than 4 mmHg shall not be included in the data set. The number of discarded measurements shall not be greater than the number of the required valid measurements. NOTE 1 Reading of the values on the reference manometer should be as accurate as possible. When reading the value on the reference manometer(s) the observers should avoid parallax errors. Rounding to zero and five has a negative effect on the result of the investigation. More information and recommendations are given in annex D.
The calibrated reference manometers shall comply with the requirements of EN 1060-1 to EN 1060-3 but shall not exceed error limits of 1 mmHg (0,1 kPa) with dropping cuff pressure prior to the start of the clinical investigation. NOTE 2
The reference manometer should be calibrated before and after the clinical investigation. When the reference manometer is linked to the pneumatic system of the tested device (e.g. test method N2), an aneroid manometer should be used. Mercury manometer may alter pneumatic characteristics (resonant frequency, damping coefficient) and should not be used in this case. 4.2 General information on the invasive reference methods The clinical investigation shall be performed in accordance with EN ISO 14155–1 and EN ISO 14155–2. The invasive blood pressure measuring system consisting of catheter, pressure transducer, monitor and recording device shall be statically calibrated so that the indicated values do not deviate from the values indicated by the reference manometer (see 4.5.1.1 a) by more than ± 2 mmHg. Prior to and after each series of measurements the resonant frequency and damping coefficient of the system shall be determined [1]. Except for devices measuring in the neonatal mode, the comparison shall be carried out by invasive and non-invasive blood pressure measurement on the same arm. The simultaneous measurement on opposite arms is permissible if no lateral difference has been determined by additional measurements. NOTE Intra-arterial methods are invasive and should not be used for patients or subjects solely for the purpose of validating instrument performance, but conducted on clinical patients in whom an intra-arterial line has already been placed for reasons other than sphygmomanometer verification. 4.3 Selection of clinical test method The selection of the clinical test method is dependent on the measurement technique, the application of the sphygmomanometers to be tested and the intended application of the device and shall be in accordance with
Table 1. SIST EN 1060-4:2005



EN 1060-4:2004 (E) 7 Table 1 — Matrix for the selection of the clinical test method Reference method Measurement technique of the device to be tested
Clinical test method as a function of application
adults neonatal mode ergo.a ABPMb Continuous pressure drop or pressure drop in steps (upper arm measurement) ≤ 3 mmHg/s or
≤ 3 mmHg/pulsec
> 3 mmHg/s or > 3 mmHg/pulsec
N1/N2/N3N2/N3
- -
N4 -
N5/N6 N6 Measurement on other sites than the upper arm
N2/N3
Auscultatory measurement at the upper arm Measurement during inflation phase N2/N3 - - N6 Invasive measurement Measurement during the pressure drop or the inflation phase I 1 I 2 - - a Ergometry (measurement under physical load) b Ambulatory blood pressure measurement c For devices adapting to the pulse rate
4.4 Application of test method 4.4.1 Non-invasive reference measurement 4.4.1.1 Simultaneous blood pressure measurement on the same upper arm; also under physical load and ambulatory The test methods (N1, N4, N5) for the simultaneous blood pressure measurement on the same upper arm are suitable for sphygmomanometers operating  with continuous pressure drop (rate of pressure drop ≤ 3 mmHg/s) and/or  with pressure drop in steps (rate of pressure drop per step ≤ 3 mmHg) and/or  with a pulse-controlled pressure drop (rate of pressure drop ≤ 3 mmHg per pulse) These values apply for the range of the systolic and diastolic blood pressures. 4.4.1.2 Simultaneous blood pressure measurement on opposite arms The test methods (N2 and N6) for the simultaneous blood pressure measurement are suitable for sphygmomanometers operating  with continuous pressure drop and/or
 with pressure drop in steps and/or  with a pulse-controlled pressure drop. This test method is also suitable for sphygmomanometers measuring during the inflation phase or measuring on other sites then upper arm. SIST EN 1060-4:2005



EN 1060-4:2004 (E) 8 4.4.1.3 Sequential blood pressure measurement The test methods (N3) for sequential blood pressure measurement are suitable for sphygmomanometers operating  with continuous pressure drop and/or
 with pressure drop in steps and/or  with a pulse-controlled pressure drop and/or  measuring on other sites than the upper arm. This test method is also suitable for sphygmomanometers measuring during the inflation phase. 4.4.2 Invasive reference measurement The test method (I1) is suitable for sphygmomanometers for measurement on adults. The test method (I2) is suitable for sphygmomanometers for measurement on
neonates and infants. 4.5 Test equipment 4.5.1 Non-invasive reference measurement
4.5.1.1 Simultaneous blood pressure measurement; also under physical load and ambulatory a) calibrated reference manometer(s) with error limit(s) of 1 mmHg (0,1 kPa); b) T-piece; c) hose; d) double stethoscope. 4.5.1.2 Opposite arm or sequential blood pressure measurement; also ambulatory a) calibrated reference manometer(s) with error limit(s) of 1 mmHg (0,1 kPa) b) double stethoscope and for method N2 additionally two stethoscopes; c) cuff(s). 4.5.2 Invasive blood pressure measurement
a) calibrated reference manometer with error limit of 1 mmHg (0,1 kPa); b) invasive blood pressure measuring system; c) device for the recording of the invasive blood pressure. SIST EN 1060-4:2005



EN 1060-4:2004 (E) 9 4.6 Subjects 4.6.1 General The selection of the subjects and their number depends on the intended purpose (In accordance with the information given in the instructions for use and/or the instructions on the device) of the device to be tested. Limits of application stated in the users manual shall be taken into account, e.g. concerning arrhythmia (see also 4.8). If no special purpose is intended, e.g. measurement during pregnancy, the following applies only for adults and children:  at least 40 % shall be male and at least 40 % shall be female;  between 50 % and 75 % shall be older than 50 years;  between 50 % and 75 % shall have a circumference of the arm, which lies within the upper half of the specified range of use of the cuff (if applicable);
 between 50 % and 75 % shall have a circumference of the wrist, which lies within the upper half of the specified range of use of the cuff (if applicable);  at least 10 % below 110 mmHg systolic blood pressure;  at least 10 % above 160 mm Hg systolic blood pressure;  at least 10 % below 70 mmHg diastolic blood pressure;  at least 10 % above 100 mm Hg diastolic blood pressure.
The different blood pressure groups (see above) shall be classified in accordance with the first valid blood pressure reference measurements. 4.6.2 Non-invasive reference measurement
4.6.2.1 General A minimum of 3 measurements shall be carried out on each of at least 85 subjects.
4.6.2.2 Additional requirements for sphygmomanometers measuring under physical load At least 6 paired measurements shall be carried out on each of at least 85 subjects. As much as possible, female and male subjects shall be evenly distributed while at most 25 % shall originate from the field of sports medicine. The different blood pressure groups (see 4.6.1) shall be classified in accordance with the first valid blood pressure reference measurements at rest. 4.6.2.3 Additional requirements for ambulatory sphygmomanometers At least 6 paired measurements shall be carried out on each of at least 85 subjects. The different blood pressure groups (see 4.6.1) shall be classified in accordance with the first valid blood pressure reference measurements at rest. SIST EN 1060-4:2005



EN 1060-4:2004 (E) 10 4.6.3 Invasive reference measurement 4.6.3.1 Adults The measurements shall be carried out on at least 15 subjects, with a minimum of 150 paired observations. A minimum of five and a maximum of ten paired measurements per subject shall be made. The measurements shall be carried out with as even as possible a distribution of male and female normotensive and hypertensive subjects. 4.6.3.2 Neonates and infants For devices measuring in the neonatal mode, at least 15 neonate/infant subjects are required, with a minimum of 100 paired observations (see Table 2). A minimum of five and a maximum of ten paired measurements per subject shall be made. Table 2 — Distribution of subjects Body weight g Subjects
< 1000 at least 3 ≥ 1000

2000
at least 3 ≥ 2000
at least 3
4.7 Cuff size The cuff of the sphygmomanometer to be tested shall be selected and applied in accordance with the recommendations of the manufacturer (see 9.2 of EN 1060-1:1995).
The measurement shall cover all cuff sizes and applications in accordance with the intended use described in the accompanying document. 4.8 Procedure The investigation shall be carried out under normal conditions which shall be within the specifications given by the manufacturer in the instructions for use. References in the instructions for use to limitations of the application shall be taken into account. The blood pressure values shall be measured carefully without interruption of the measuring series for each subject, if possible. NOTE An initial measurement should be taken. It should not be included in the evaluation (see annex A). 4.9 Evaluation of results 4.9.1 General The results of the evaluation of all measurements shall be within the limits determined in 7.9 of EN 1060-3:1997. The first 3 valid paired measurements shall be used for the evaluation. SIST EN 1060-4:2005



EN 1060-4:2004 (E) 11 4.9.2 Non-invasive reference measurement
4.9.2.1 Blood pressure measurement on the same arm (simultaneous or sequential) The evaluation of the test methods (N1, N 3 and N 5) shall be as follows: First, calculate the paired differences for all 255 (85x3) measurements in accordance with Equations (1) and (2). ∆pSYS = pSYSt – pSYSa (1) ∆pDIA = pDIAt – pDIAa (2) where pSYSt
is the systolic blood pressure value indicated by the sphygmomanometer under test; pSYSa
is the auscultatory mean systolic blood pressure value determined by the observers; pDIAt
is the diastolic blood pressure value indicated by the sphygmomanometer under test; pDIAa
is the auscultatory mean diastolic blood pressure value determined by the observers. The arithmetic mean values and the empirical standard deviations shall be calculated for the measurements at rest and under physical load/during exercise, if applicable. The mean values are
DIASYS,pp∆∆ and the related empirical standard deviations are s (∆pSYS), s (∆pDIA)
4.9.2.2 Blood pressure measurement on opposite arms (also ambulatory) To obtain correct result from the evaluation described below, compliance with 5.2.1.3 is required.
For the evaluation of the test methods (N2 and N6) six lateral differences ∆prl shall be determined separately for the systolic and diastolic blood pressure values as follows: ∆pSYSrl = pSYSr – pSYSl (3) ∆pDIArl = pDIAr – pDIAl (4) where pSYSr
is the systolic blood pressure value of
the right arm measured by the observer pSYSl
is the systolic blood pressure value of
the left arm measured by the observer pDIAr
is the diastolic blood pressure value of
the right arm measured by the observer pDIAl
is the diastolic blood pressure value of
the left arm measured by the observer SIST EN 1060-4:2005



EN 1060-4:2004 (E) 12 From these, the arithmetic mean values and the appertaining empirical standard deviations shall be calculated for all measurements and a subject shall be considered suitable if the lateral difference is reproducible, and the following shall be met: rlSYSp∆≤ 20 mmHg
rlDIAp∆≤ 15 mmHg.
and s (∆pSYSrl) ≤ 8 mmHg s (∆pDIArl) ≤ 8 mmHg The correction of the auscultatory reference blood pressure values (pSYSausc, PDIAa) of the paired measurement shall be calculated as follows: pSYSref = pSYSausc – rlSYSp (5) pDIAref = PDIAa – rlDIAp (6) where pSYSausc is the auscultatory mean systolic blood pressure value determined by the observers at the right arm PDIAa is the auscultatory mean diastolic blood pressure value determined by the observers at the right arm The arithmetic mean values and the empirical standard deviations shall be calculated for the measurements at rest and during exercise, if applicable. ∆pSYS = pSYSt – pSYSref (7) ∆pDIA = pDIAt – pDIAref (8) where pSYSt is the systolic blood pressure value indicated by the sphygmomanometer under test applied at the left arm; pDIAt is the diastolic blood pressure value indicated by the sphygmomanometer under test applied at the left arm; From these, the arithmetic mean values and the appertaining empirical standard deviations shall be calculated from the differences of the blood pressure values of all subjects and the measurements at rest and during exercise, if applicable.
DIASYS,pp∆∆ s (∆pSYS), s (∆pDIA)
SIST EN 1060-4:2005



EN 1060-4:2004 (E) 13 4.9.3 Documentation 4.9.3.1 The following shall be recorded with each series of measurements: a) serial number of the tested sphygmomanometer; b) cuff size, site and
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.