ASTM E2108-00
(Practice)Standard Practice for Calibration of the Electron Binding-Energy Scale of an X-Ray Photoelectron Spectrometer
Standard Practice for Calibration of the Electron Binding-Energy Scale of an X-Ray Photoelectron Spectrometer
SCOPE
1.1 This practice describes a procedure for calibrating the electron binding-energy (BE) scale of an X-ray photoelectron spectrometer that is to be used for surface analysis with unmonochromated aluminum or magnesium K X-rays or monochromated aluminum K X rays.
1.2 It is recommended that the BE scale be calibrated after the instrument is installed or modified in any substantive way. Also, it is recommended that the instrumental BE scale be checked, and if necessary, recalibrated at intervals chosen to ensure that BE measurements are statistically unlikely to be made with greater uncertainty than a tolerance limit, specified by the analyst, based on the instrumental stability and the analyst's needs. Information is provided by which an analyst can select an appropriate tolerance limit for the BE measurements and the frequency of calibration checks.
1.3 This practice is based on the assumption that the BE scale of the spectrometer is sufficiently close to linear that the BE scale can be calibrated by measurements of reference photoelectron lines made near the extremes of the working BE scale. In most commercial instruments, X-ray sources with aluminum or magnesium anodes are employed and BEs are typically measured over the 0-1000 eV range. This practice can be used for the BE range from 0 eV to 1040 eV.
1.4 The assumption that the BE scale is linear is checked by a measurement made with a reference photoelectron line or Auger-electron line that appears at an intermediate position. A single check is a necessary but not sufficient condition for establishing linearity of the BE scale. Additional checks can be made with specified reference lines on instruments equipped with magnesium or unmonochromated aluminum X-ray sources, with secondary BE standards, or by following the procedures of the instrument manufacturer. Deviations from BE-scale linearity can occur because of mechanical misalignments, excessive magnetic fields in the region of the analyzer, or imperfections or malfunctions in the power supplies. This practice does not check for, nor identify, problems of this type.
1.5 After an initial check of the BE-scale linearity and measurements of the repeatability standard deviation for the main calibration lines for a particular instrument, a simplified procedure is given for routine checks of the calibration at subsequent times.
1.6 This practice is recommended for use with X-ray photoelectron spectrometers operated in the constant-pass-energy or fixed-analyzer-transmission mode and for which the pass energy is less than 200 eV; otherwise, depending on the configuration of the instrument, a relativistic equation could be needed for the calibration equation. The practice should not be used for instruments operated in the constant-retardation-ratio mode at retardation ratios less than 10, for instruments with an energy resolution worse than 1.5 eV, or in applications for which BE measurements are desired with tolerance limits of 0.03 eV or less.
1.7 On instruments equipped with a monochromated aluminum K X-ray source, a measurement of the position of a specified Auger-electron line can be used, if desired, to determine the average energy of the X rays incident on the specimen. This information is needed for the determination of modified Auger parameters.
1.8 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. ^REFERENCE: ^REFDOCID: E 456^REFDOCTITLE: Terminology Relating to Quality and Statistics^REFDOCID: E 673^REFDOCTITLE: Terminology Relating to Surface Analysis^REFDOCID: E 902^REFDOCTITLE: Practice for Checking the Operating Characteristics of X-Ray Photoelectron Spectrometers^REFDOCID: E 1016^REFDOCTITLE: Guide for Literature Describing Properties of Electrostatic Electron Spe...
General Information
Relations
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: E 2108 – 00
Standard Practice for
Calibration of the Electron Binding-Energy Scale of an
1
X-Ray Photoelectron Spectrometer
This standard is issued under the fixed designation E2108; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
1. Scope 1.5 After an initial check of the BE-scale linearity and
measurements of the repeatability standard deviation for the
1.1 This practice describes a procedure for calibrating the
main calibration lines for a particular instrument, a simplified
electron binding-energy (BE) scale of an X-ray photoelectron
procedure is given for routine checks of the calibration at
spectrometer that is to be used for surface analysis with
subsequent times.
unmonochromated aluminum or magnesium Ka X-rays or
1.6 This practice is recommended for use with X-ray
monochromated aluminum Ka X rays.
photoelectron spectrometers operated in the constant-pass-
1.2 It is recommended that the BE scale be calibrated after
energy or fixed-analyzer-transmission mode and for which the
the instrument is installed or modified in any substantive way.
pass energy is less than 200 eV; otherwise, depending on the
Also, it is recommended that the instrumental BE scale be
configurationoftheinstrument,arelativisticequationcouldbe
checked, and if necessary, recalibrated at intervals chosen to
needed for the calibration equation.The practice should not be
ensure that BE measurements are statistically unlikely to be
used for instruments operated in the constant-retardation-ratio
made with greater uncertainty than a tolerance limit, specified
mode at retardation ratios less than 10, for instruments with an
by the analyst, based on the instrumental stability and the
energy resolution worse than 1.5 eV, or in applications for
analyst’s needs. Information is provided by which an analyst
whichBEmeasurementsaredesiredwithtolerancelimitsof 6
can select an appropriate tolerance limit for the BE measure-
0.03 eV or less.
ments and the frequency of calibration checks.
1.7 On instruments equipped with a monochromated alumi-
1.3 This practice is based on the assumption that the BE
num Ka X-ray source, a measurement of the position of a
scale of the spectrometer is sufficiently close to linear that the
specified Auger-electron line can be used, if desired, to
BE scale can be calibrated by measurements of reference
determine the average energy of the X rays incident on the
photoelectron lines made near the extremes of the working BE
specimen. This information is needed for the determination of
scale. In most commercial instruments, X-ray sources with
modified Auger parameters.
aluminum or magnesium anodes are employed and BEs are
1.8 This standard does not purport to address all of the
typically measured over the 0–1000 eV range. This practice
safety concerns, if any, associated with its use. It is the
can be used for the BE range from 0 eV to 1040 eV.
responsibility of the user of this standard to establish appro-
1.4 TheassumptionthattheBEscaleislinearischeckedby
priate safety and health practices and determine the applica-
a measurement made with a reference photoelectron line or
bility of regulatory limitations prior to use.
Auger-electron line that appears at an intermediate position.A
single check is a necessary but not sufficient condition for
2. Referenced Documents
establishinglinearityoftheBEscale.Additionalcheckscanbe
2.1 ASTM Standards:
made with specified reference lines on instruments equipped
2
E456 Terminology Relating to Quality and Statistics
with magnesium or unmonochromated aluminum X-ray
3
E673 Terminology Relating to Surface Analysis
sources, with secondary BE standards, or by following the
E902 Practice for Checking the Operating Characteristics
procedures of the instrument manufacturer. Deviations from
3
of X-Ray Photoelectron Spectrometers
BE-scale linearity can occur because of mechanical misalign-
E1016 Guide for Literature Describing Properties of Elec-
ments, excessive magnetic fields in the region of the analyzer,
3
trostatic Electron Spectrometers
or imperfections or malfunctions in the power supplies. This
E1078 Guide for Specimen Preparation and Mounting in
practice does not check for, nor identify, problems of this type.
3
Surface Analysis
E1523 Guide to Charge Control and Charge Referencing
1
This practice is under the jurisdiction of ASTM Committee E42 on Surface
AnalysisandisthedirectresponsibilityofSubcommitteeE42.03onAugerElectron
2
Spectroscopy and X-Ray Photoelectron Spectroscopy. Annual Book of ASTM Standards, Vol 14.02.
3
Current edition approved Sept. 10, 2000. Published December 20
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.