ASTM ISO/ASTM51702-13
(Main)Standard Practice for Dosimetry in a Gamma Facility for Radiation Processing
Standard Practice for Dosimetry in a Gamma Facility for Radiation Processing
SIGNIFICANCE AND USE
4.1 Various products and materials routinely are irradiated at predetermined doses in gamma irradiation facilities to reduce their microbial population or to modify their characteristics. Dosimetry requirements may vary depending upon the irradiation application and end use of the product. Some examples of irradiation applications where dosimetry may be used are:
4.1.1 Sterilization of medical devices,
4.1.2 Treatment of food for the purpose of parasite and pathogen control, insect disinfestation, and shelf life extension,
4.1.3 Disinfection of consumer products,
4.1.4 Cross-linking or degradation of polymers and elastomers,
4.1.5 Polymerization of monomers and grafting of monomers onto polymers,
4.1.6 Enhancement of color in gemstones and other materials,
4.1.7 Modification of characteristics of semiconductor devices, and
4.1.8 Research on materials effects.
NOTE 3—Dosimetry is required for regulated irradiation processes such as sterilization of medical devices and the treatment of food. It may be less important for other industrial processes, for example, polymer modification, which can be evaluated by changes in the physical and chemical properties of the irradiated materials.
4.2 An irradiation process usually requires a minimum absorbed dose to achieve the intended effect. There also may be a maximum absorbed dose that the product can tolerate and still meet its functional or regulatory specifications. Dosimetry is essential to the irradiation process since it is used to determine both of these limits and to confirm that the product is routinely irradiated within these limits.
4.3 The absorbed-dose distribution within the product depends on the overall product dimensions and mass, irradiation geometry, and source activity distribution.
4.4 Before an irradiation facility can be used, it must be qualified to determine its effectiveness in reproducibly delivering known, controllable absorbed doses. This involves testing the process equipment, calibrat...
SCOPE
1.1 This practice outlines the installation qualification program for an irradiator and the dosimetric procedures to be followed during operational qualification, performance qualification, and routine processing in facilities that process products with ionizing radiation from radionuclide gamma sources to ensure that product has been treated within a predetermined range of absorbed dose. Other procedures related to operational qualification, performance qualification, and routine processing that may influence absorbed dose in the product are also discussed.
NOTE 1—Dosimetry is only one component of a total quality assurance program for adherence to good manufacturing practices used in radiation processing applications.
NOTE 2—ISO/ASTM Practices 51818 and 51649 describe dosimetric procedures for low and high enery electron beam facilities for radiation processing and ISO/ASTM Practice 51608 describes procedures for X-ray (bremsstrahlung) facilities for radiation processing.
1.2 For the radiation sterilization of health care products, see ISO 11137-1. In those areas covered by ISO 11137-1, that standard takes precedence.
1.3 This document is one of a set of standards that provides recommendations for properly implementing and utilizing dosimetry in radiation processing. It is intended to be read in conjunction with ASTM Practice E2628.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
General Information
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
ISO/ASTM 51702:2013(E)
Standard Practice for
1
Dosimetry in a Gamma Facility for Radiation Processing
This standard is issued under the fixed designation ISO/ASTM 51702; the number immediately following the designation indicates the
year of original adoption or, in the case of revision, the year of last revision.
1. Scope 2. Referenced documents
2
2.1 ASTM Standards:
1.1 This practice outlines the installation qualification pro-
E170 Terminology Relating to Radiation Measurements and
gram for an irradiator and the dosimetric procedures to be
Dosimetry
followed during operational qualification, performance
E2232 Guide for Selection and Use of Mathematical Meth-
qualification, and routine processing in facilities that process
ods for Calculating Absorbed Dose in Radiation Process-
products with ionizing radiation from radionuclide gamma
ing Applications
sources to ensure that product has been treated within a
E2303 Guide for Absorbed-Dose Mapping in Radiation
predetermined range of absorbed dose. Other procedures re-
Processing Facilities
lated to operational qualification, performance qualification,
E2628 Practice for Dosimetry in Radiation Processing
and routine processing that may influence absorbed dose in the
E2701 Guide for Performance Characterization of Dosim-
product are also discussed.
eters and Dosimetry Systems for Use in Radiation Pro-
NOTE 1—Dosimetry is only one component of a total quality assurance
cessing
program for adherence to good manufacturing practices used in radiation
2
2.2 ISO/ASTM Standards:
processing applications.
51261 Practice for Calibration of Routine Dosimetry Sys-
NOTE 2—ISO/ASTM Practices 51818 and 51649 describe dosimetric
tems for Radiation Processing
procedures for low and high enery electron beam facilities for radiation
51539 Guide for Use of Radiation-Sensitive Indicators
processing and ISO/ASTM Practice 51608 describes procedures for X-ray
51608 Practice for Dosimetry in an X-Ray (Bremsstrahlung)
(bremsstrahlung) facilities for radiation processing.
Facility for Radiation Processing
1.2 For the radiation sterilization of health care products,
51649 Practice for Dosimetry in an Electron Beam Facility
see ISO 11137-1. In those areas covered by ISO 11137-1, that
for Radiation Processing at Energies Between 300 KeV
standard takes precedence.
and 25 KeV
51707 Guide for Estimating Uncertainties in Dosimetry for
1.3 This document is one of a set of standards that provides
Radiation Processing
recommendations for properly implementing and utilizing
51818 Practice for Dosimetry in an Electron Beam Facility
dosimetry in radiation processing. It is intended to be read in
for Radiation Processing at Energies Between 80 and 300
conjunction with ASTM Practice E2628.
keV
1.4 This standard does not purport to address all of the
2.3 International Commission on Radiation Units and Mea-
3
safety concerns, if any, associated with its use. It is the
surements (ICRU) Reports:
responsibility of the user of this standard to establish appro-
ICRU Report 85a Fundamental Quantities and Units for
priate safety and health practices and determine the applica-
Ionizing Radiation
4
bility of regulatory limitations prior to use.
2.4 ISO Standards:
ISO 11137-1 Sterilization of health care products – Radia-
tion – Part 1: Requirements for development, validation,
1
This practice is under the jurisdiction of ASTM Committee E61 on Radiation
2
Processing and is the direct responsibility of Subcommittee E61.03 on Dosimetry For referenced ASTM and ISO/ASTM standards, visit the ASTM website,
Application, and is also under the jurisdiction of ISO/TC 85/WG 3. www.astm.org, or contact ASTM Customer Service at service@astm.org. For
Current edition approved Dec. 26, 2012. Published April 2013. Originally Annual Book of ASTM Standards volume information, refer to the standard’s
published as E 1702-95. Last previous ASTM edition E 1702–00. ASTM Document Summary page on the ASTM website.
ε1 3
E 1702–95 was adopted by ISO in 1998 with the intermediate designation ISO Available from the International Commission on Radiation Units and
15571:1998(E). The present International Standard ISO/ASTM 51702:2013(E) Measurements, 7910 Woodmont Ave., Suite 800, Bethesda, MD 20814, USA.
4
replaces ISO 15571 and is a major revision of the last previous edition ISO/ASTM Available from International Organization for Standardization (ISO), 1 rue de
51702–2004(E). Varembé, Case postale 56, CH-1211, Geneva 20, Switzerland, http://www.iso.ch.
© ISO/ASTM International 2022 – All rights reserved
1
---------------------- Page: 1 ----------------------
ISO/ASTM 51702:2013(E)
and routine control of a sterilization process for medical and operated in accordance with operational procedures, con-
devices sistently performs in accordance with predetermined criteria
and thereby yields pro
...








Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.