ASTM E1096-86(2024)
(Specification)Standard Specification for Laboratory Glass Separatory Funnels
Standard Specification for Laboratory Glass Separatory Funnels
ABSTRACT
This specification provides standard dimensional requirements for glass separatory funnels for general laboratory use. Separatory funnels shall be in the following types: Type 1A; Type 1B; Type 1C; Type 2; Type 3; and Type 4. Separatory funnels shall be made of borosilicate glass conforming to the requirement specified. The design of each separatory funnels shall comply with the dimensions given in each requirements.
SCOPE
1.1 This specification provides standard dimensional requirements for glass separatory funnels for general laboratory use.
1.2 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
General Information
Relations
Standards Content (Sample)
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: E1096 − 86 (Reapproved 2024)
Standard Specification for
Laboratory Glass Separatory Funnels
This standard is issued under the fixed designation E1096; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
INTRODUCTION
Separatory funnels are used in laboratories primarily for liquid extractions, and are intended to
facilitate the separation of two immiscible liquids of different densities into separate layers. Some
funnels are used to add reagent solution into a reaction vessel. They are, therefore, often provided with
a tapered ground joint at the bottom of the delivery stem for joining to vessels having similarly tapered
ground necks. They may also be provided with pressure equalizing side arms.
1. Scope 3.1.1 Type 1A—Cylindrical shape with open top.
3.1.1.1 Sizes—60, 125, and 250 cm .
1.1 This specification provides standard dimensional re-
3.1.2 Type 1B—Cylindrical with stopper finish top.
quirements for glass separatory funnels for general laboratory
3.1.2.1 Sizes—60, 125, 250, 500, and 1000 cm .
use.
3.1.3 Type 1C—Cylindrical with stopper finish top, gradu-
1.2 This international standard was developed in accor-
ated.
dance with internationally recognized principles on standard-
3.1.3.1 Sizes—125, 250, 500, and 1000 cm .
ization established in the Decision on Principles for the
3.1.4 Type 2—Globe shape with stopper finish top.
Development of International Standards, Guides and Recom-
3.1.4.1 Sizes—60, 125, 250, 500, 1000, 2000, and 4000 cm .
mendations issued by the World Trade Organization Technical
3.1.5 Type 3—Globe shape, “French,” with stopper finish
Barriers to Trade (TBT) Committee.
top.
2. Referenced Documents 3
3.1.5.1 Sizes—125, 250, 500, and 1000 cm .
2.1 ASTM Standards:
3.1.6 Type 4—Pear shape, Squibb, with stopper finish top.
E438 Specification for Glasses in Laboratory Apparatus
3.1.6.1 Sizes—20, 60, 125, 250, 500, 1000, 2000, and 4000
E671 Specification for Maximum Permissible Thermal Re-
cm .
sidual Stress in Annealed Glass Laboratory Apparatus
NOTE 1—The term millilitre (mL) is commonly used as a special name
E675 Specification for Interchangeable Taper-Ground Stop-
for the cubic centimetre (cm ) and similarly the litre (L) for 1000 cubic
cocks And Stoppers
centimetres, in accordance with the International System of Units (SI).
E676 Specification for Interchangeable Taper-Ground Joints
E694 Specification for Laboratory Glass Volumetric Appa-
4. Materials and Annealing
ratus
4.1 Separatory funnels shall be made of borosilicate glass
E911 Specification for Glass Stopcocks with Polytetrafluo-
conforming to the requirement of Type 1, Class A of Specifi-
roethylene (PTFE) Plugs
cation E438.
3. Classification
4.2 Maximum residual thermal stress shall be such as to
3.1 Separatory funnels shall be in the following types and
conform to Specification E671.
sizes:
5. Design
This specification is under the jurisdiction of ASTM Committee E41 on
5.1 Type 1 cylindrical separatory funnels shall have straight
Laboratory Apparatus and is the direct responsibility of Subcommittee E41.01 on
Laboratory Ware and Supplies.
sides and comply with the dimensions given in Table 1. (Type
Current edition approved Feb. 1, 2024. Published February 2024. Originally
1A stem length shall comply with Table 2.) See Fig. 1, Type
approved in 1986. Last previous edition approved in 2018 as E1096 – 86 (2018)
1A; 1B; and 1C.
DOI: 10.1520/E1096-86R24.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
5.2 Type 2 globe-shaped separatory funnels shall be roughly
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
globular in shape and have dimensions complying with those
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website. given in Table 2. See Fig. 2, Type 2.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
E1096 − 86 (2024)
FIG. 1 Cylindrical Body Funnels
Type 2 Type 3 Type 4
FIG. 2 Globe and Squibb-Pear Shaped Body Funnels
TABLE 1 Cylindrical Body Funnels
5.4 Type 4 Squibb separatory funnels shall be pear-shaped
Nominal Capacity, or conical-shaped and meet the requirements of Table 3. See
mL 60 125 250 500 1000
Fig. 2, Type 4.
Body height, max 135 200 225 290 390
(mm) (see Fig.
5.5 Type 1 and 4 separatory funnels may be supplied with a
1 “L ”)
taper-ground joint below the stopcock, (see Specification E676
Body diameter, 34 44 53 67 79
max (mm) and Fig. 1C. Type 1 funnels with this feature may also have
Body wall thick- 0.8 0.8 0.9 0.9 1.3
pressure equalizing tubes placed at the back or opposite side of
ness, min (mm)
the funnel when it is in the position of normal use with the
Stem length, max 100 100 100 100 100
(mm) (see Fig.
handle of the stopcock on the right. See Fig. 1B).
...
Questions, Comments and Discussion