ASTM E2251-14(2021)
(Specification)Standard Specification for Liquid-in-Glass ASTM Thermometers with Low-Hazard Precision Liquids
Standard Specification for Liquid-in-Glass ASTM Thermometers with Low-Hazard Precision Liquids
ABSTRACT
This specification covers liquid-in-glass ASTM thermometers using low hazard thermometric liquids. The gas filling above the liquid shall be nitrogen or other suitable inert gas. The filling gas shall be chosen to have very low solubility in the thermometric fluid. The stem shall be made of suitable thermometer tubing and shall have a plain front and enamel back. The bulb shall be made of glass and the following distances between graduations and the bulb, and between graduations and enlargements in the capillary, are minimum limits acceptable. All graduation lines, figures, and letters shall be clearly defined, suitably colored, and permanent. The width and the sharpness of the graduation lines shall be designed in accordance with necessary space between the graduations and the desired accuracy of interpolation. The middle of the graduation line shall be accurately determinable. In addition, the graduation lines shall be straight, of uniform width, and perpendicular to the axis of the thermometer. On partial immersion thermometers an immersion line shall be permanently marked on the front of the thermometer at the distance above the bottom of the bulb as specified. The immersion inscription shall be written in capital letters and abbreviated. The terminal number shall be in full when there are one or more numbered graduations between it and the next full number. The special inscription specified shall be marked on the thermometer in capital letters and Arabic numbers without the use of periods.
SCOPE
1.1 The purpose of this standard is to specify liquid-in-glass ASTM thermometers using low hazard thermometric liquids defined in this standard.
1.2 This standard specifies liquid-in-glass thermometers graduated in degrees Celsius or degrees Fahrenheit that are frequently identified and used in methods under the jurisdiction of the various technical committees within ASTM. The current approved thermometers are listed in Table 1.
1.3 The technical requirements for the thermometric liquids used in the thermometers in Table 1 are specified in Annex A1. Tests for conformity to the technical requirements are also found in Annex A1.
Note 1: It has been found by experience that ASTM Thermometers, although developed in general for specific tests, may also be found suitable for other applications, thus precluding the need for new thermometer specifications differing in only minor features. However, it is suggested that technical committees contact E20.05 before choosing a currently designated thermometer for a new method to be sure the thermometer will be suitable for the intended application.
1.4 For full rationale, see Appendix X1.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
General Information
Standards Content (Sample)
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation:E2251 −14 (Reapproved 2021)
Standard Specification for
Liquid-in-Glass ASTM Thermometers with Low-Hazard
1
Precision Liquids
This standard is issued under the fixed designation E2251; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 2. Referenced Documents
2
2.1 ASTM Standards:
1.1 Thepurposeofthisstandardistospecifyliquid-in-glass
E1Specification for ASTM Liquid-in-Glass Thermometers
ASTM thermometers using low hazard thermometric liquids
E77Test Method for Inspection and Verification of Ther-
defined in this standard.
mometers
1.2 This standard specifies liquid-in-glass thermometers
E344Terminology Relating to Thermometry and Hydrom-
graduated in degrees Celsius or degrees Fahrenheit that are
etry
frequentlyidentifiedandusedinmethodsunderthejurisdiction
E563Practice for Preparation and Use of an Ice-Point Bath
of the various technical committees withinASTM.The current
as a Reference Temperature
approved thermometers are listed in Table 1.
3. Terminology
1.3 The technical requirements for the thermometric liquids
3.1 Definitions—ThedefinitionsgiveninTerminologyE344
usedinthethermometersinTable1arespecifiedinAnnexA1.
apply.
Tests for conformity to the technical requirements are also
found in Annex A1.
3.2 Definitions of Terms Specific to This Standard:
NOTE 1—It has been found by experience that ASTM Thermometers,
3.2.1 bulb length, n—the distance from the bottom of the
although developed in general for specific tests, may also be found
bulb to the junction of the bulb and the stem tubing.
suitableforotherapplications,thusprecludingtheneedfornewthermom-
3.2.2 contraction chamber, n—an enlargement of the
eter specifications differing in only minor features. However, it is
suggested that technical committees contact E20.05 before choosing a
capillary, located below the main scale or between the main
currently designated thermometer for a new method to be sure the
scale and the auxiliary scale, that serves to reduce the scale
thermometer will be suitable for the intended application.
lengthortopreventcontractionofalltheliquidcolumnintothe
1.4 For full rationale, see Appendix X1.
bulb.
1.5 This standard does not purport to address all of the 3.2.3 diameter, n—thelargestoutsidedimensionoftheglass
tubing as measured with a ring gage.
safety concerns, if any, associated with its use. It is the
responsibility of the user of this standard to establish appro-
3.2.4 expansion chamber, n—an enlargement at the top of
priate safety, health, and environmental practices and deter-
the capillary to provide protection against breakage caused by
mine the applicability of regulatory limitations prior to use.
excessive gas pressure.
1.6 This international standard was developed in accor-
3.2.5 faden thermometer, n—a thermometer with a long,
dance with internationally recognized principles on standard-
thin bulb used to determine emergent stem temperatures.
ization established in the Decision on Principles for the
3.2.6 intervalerror,n—thedeviationofthenominalvalueof
Development of International Standards, Guides and Recom-
a temperature interval from its true value; either for the total
mendations issued by the World Trade Organization Technical
range(totalinterval)orforapartoftherange(partialinterval).
Barriers to Trade (TBT) Committee.
3.2.7 low-hazard liquid, n—a liquid that is biodegradable,
non-hazardousandconsiderednon-toxicinthermometerquan-
tities.
1
This specification is under the jurisdiction of ASTM Committee E20 on
Temperature Measurement and is the direct responsibility of Subcommittee E20.05
2
on Liquid-in-Glass Thermometers and Hydrometers. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved Dec. 1, 2021. Published December 2021. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
approved in 2003. Last previous edition approved in 2014 as E2251–14. DOI: Standards volume information, refer to the standard’s Document Summary page on
10.1520/E2251-14R21. the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1
---------------------- Page: 1 ----------------------
E2251−14 (2021)
NOTE 2—It is the responsibility of the manufacturer to determine the
8. Capillary Clearances
suitability of a liquid for this standard. In marking the thermometer with
8.1 The following distances between graduations
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.